510(k) Data Aggregation
(165 days)
The Osteon Precision Milled Suprastructure is indicated for attachments in the treatment of partially or fully edentulous jaws for the purpose of restoring function. The Osteon Precision Milled Suprastructures are intended for attachment to a minimum of two (2) abutments.
The Osteon Milled Suprastructure is indicated for compatibility with the following abutment systems:
- · Astra Tech Implant System® Multi Base Abutment EV, 4.8mm, max 30°
- · BioHorizons Multi Unit Abutment, 4.8mm, max 30°
- · CONELOG® Implant System
- · Biomet 3 i Multi Unit Abutments, 4.8mm, max 30°
- · TSXTM Implants
- · Tapered Screw-Vent Implant
- · DESS Dental Multi Unit Abutments, 3.4-5.7 mm, 0°
- · 3i OSSEOTITE®
- · Astra Tech OsseoSpeed™
- · Neodent Grand Morse
- NobelReplace® Trilobe
- · NobelReplace® Conical
- · Nobel Brånemark System®
- · Straumann BLX Implants
- · DESS Dental Multi Unit Abutments. Angled. 3.4-6.5 mm, max 30°
- · NobelActive® NobelParallel Conical
- Straumann® Bone Level
- · Zimmer Screw Vent® and Tapered Screw-Vent®
- · Dentium SuperLine® Abutments, 4.5-5.5 mm, max30°
- · Genesis ACTIVE™ Multi-Unit Abutments, 4.8mm, max 30°
- · Implant Direct GPS® Angled Abutment, 5.0mm, max 30°
- KDG Abutments, 4.8mm, max 30°
- · Keystone Multi Unit Abutment, 4.8mm, 0°
- · Medentika Multi Unit Abutments, 4.8mm, max 30°
- · EV Series Dentsply® Implants Astratech Osseospeed®
- · F Series Nobel Biocare NobelActive® NobelReplace® Conical
- · H Series Biomet 3i Certain®
- L Series Straumann Bone Level
- · N Series Straumann Soft tissue Level
- · R Series Zimmer Dental Tapered Screwvent®
- Medentika Multi Unit Abutments, 4.8mm, 0°
- E Series Nobel Biocare Replace™ Select
- · I Series Biomet 3i Osseotite®
- K Series Nobel Biocare™ Branemark
- · S Series Astra Tech OsseoSpeedTM
- · T Series Dentsply Friadent® Frialit/Xive®
- · MegaGen Multi Unit Abutments, 4.8mm, max 30°
- · Xpeed® AnyRidge® Internal Implant System
- AnyOne® Internal Implant System
- AnyRidge® Octa 1 Implant System
- AnyOne® External Implant System
- AnyRidge® Octa 1 Implant System
- AnyOne® Internal Implant System
- Rescue Internal Implant System
- MIS Multi-unit Abutments, 4.8mm
- · C1 Conical Connection Implant System, max 30°
- · V3 Conical Connection Implant System, max 30°
- · Internal Hex Implant System, max 30°
- · Conical Connection, max 30°
- · Neodent GM Mini Conical Abutment, 4.8 mm, max 30°
- · Nobel Biocare™ Brånemark Multi Unit Abutment, 4.8 mm, max 17°
- · Nobel Biocare™ Multi Unit Abutment Plus, 4.8 mm, max 30°
- · Nobel Biocare™ Multi Unit Abutment, 4.8 mm, max 30°
- Nobel Biocare™ Multi Unit Abutments for Straumann and Astra Tech System, 4.8 mm, max 30°
- Nobel Biocare™ Multi Unit Abutments for Astra Tech, Camlog and Ankylos Implant Systems, 4.8 mm, max 30°
- · Nobel Biocare Xeal Abutments, 4.8 mm, max 30°
- · OSSTEM Multi Unit Abutment, 4.8mm, max 30°
- SS SA Fixture Implants
- SA Implant System
- · ET US SSS Prosthetic System
- · Paltop Multi Unit Abutment, 5.0 mm, max 17°
- · Southern Compact Conical Abutments, 4.8 mm
- MAX Implant System, 0°
- · Provata Implant System, max 30°
- · Deep Conical (DC) Implants, 0°
- · Piccolo Implants, 0°
- · External Hex Implants, max 30°
- Straumann® BLX Screw Retained Abutment, 4.6 mm, max 30°
- · Straumann® Screw Retained Abutment, 4.6 mm, max 30°
- · Zimmer Angled Tapered Abutments, 4.5 mm, max 30°
The Osteon Precision Milled Suprastructures (also referred to as superstructures) are metallic dental restorative devices that are intended to be attached by screw retention to dental implant abutments to aid in the treatment of partial and totally edentulous patients for the purpose of restoring chewing function. These suprastructures attach to dental implant abutments using the prosthetic screws from the original equipment manufacturers (OEM) and are used to support the final multi-unit restoration.
The Osteon Precision Milled Suprastructure is designed for an individual patient from scans of the patient's dental impression. The suprastructure is manufactured with the aid of Computer Aided Design (CAD) and Computer Aided Manufacturing (CAM) technology. All CAD/CAM fabrication is performed by Osteon Medical.
Osteon Precision Milled Suprastructures facilitate the attachment of both removable and fixed dental prosthesis and hence are categorized as Type A and Type B.
- Type A: Intended to act as a supporting structure to facilitate the attachment of a removable dental prosthesis and include Primary Bar and Nexus Removable Bar.
- . Type B: Intended to act as a supporting structure to facilitate the attachment of a fixed dental prosthesis and include Melbourne Bar and Nexus Fixed Bar (including Nexus Hybrid, Nexus Bridge, Micro Nexus, and Nexus Wraparound).
The provided FDA 510(k) summary (K233083) describes the acceptance criteria and study for the Osteon Precision Milled Suprastructure.
1. Table of Acceptance Criteria and Reported Device Performance
This submission is a 510(k) for an expansion of an existing device (K221019) to add compatibility with new OEM abutment systems and introduce three new Type B Nexus Fixed Bars. The acceptance criteria are primarily based on demonstrating substantial equivalence to the predicate devices in terms of intended use, technological characteristics, and material composition.
The tables below synthesize the design specifications (acceptance criteria) and the device's performance (reported as meeting these specifications) by demonstrating substantial equivalence to predicate devices.
Table 1: Comparison of General Characteristics and Materials (summarized for clarity)
Characteristic | Acceptance Criteria (Predicate) | Reported Device Performance (Subject Device) |
---|---|---|
Intended Use | Dental implant abutment for restoring chewing function in partially or fully edentulous jaws, attached to a minimum of two abutments. Specific compatible OEM abutment systems listed. | Substantially Equivalent. The subject device shares the same intended use. The primary difference is an expanded list of compatible OEM abutment systems. |
Device Material | Titanium alloy Ti-6Al-4V (ASTM F136) | Meets/Substantially Equivalent. Manufactured from titanium alloy conforming to ASTM F136. |
Design/Technology | CAD/CAM milling from single milling blanks | Meets/Substantially Equivalent. CAD/CAM milling from single milling blanks. |
Design/Construction | Patient specific/machined | Meets/Substantially Equivalent. Patient specific/machined. |
Sterility | Supplied Nonsterile | Meets/Substantially Equivalent. Supplied Nonsterile (to be sterilized by end-user). |
Prescription/OTC | Prescription only | Meets/Substantially Equivalent. Prescription only. |
Recommended Cleaning & Maintenance | Proper oral hygiene | Meets/Substantially Equivalent. Proper oral hygiene. |
Table 2: Design Specifications (Acceptance Criteria vs. Reported Performance)
Design Parameters | Type A (Removable Prosthesis) Acceptance Criteria (Predicate) | Type B (Fixed Prosthesis) Acceptance Criteria (Predicate) | Type A (Removable Prosthesis) Reported Performance (Subject Device) | Type B (Fixed Prosthesis) Reported Performance (Subject Device) |
---|---|---|---|---|
Total Cylinders | 2-10 | 2-10 | 2-10 | 2-10 |
Suprastructure Span Between Cylinders (mm) | 1-30 mm | 1-30 mm | 1-30 mm | 1-30 mm |
Suprastructure Height (mm) | 3-12 mm | 3*-22 mm | 3-12 mm | 3*-22 mm |
Suprastructure Width (mm) | 3.4-12 mm | 3.4**-12 mm | 3.4-12 mm | 3.4**-12 mm |
Distal Cantilever Section (mm) | 0-15 mm | 0-15 mm | 0-15 mm | 0-15 mm |
Cylinder Height (mm) | 0-4.6 mm | 0-4.6 mm | 0-4.6 mm | 0-4.6 mm |
Cylinder Diameter (mm) | 4.5-8 mm | 4.5-8 mm | 4.5-8 mm | 4.5-8 mm |
*Note for Suprastructure Height: The Minimum Suprastructure Height for the Micro Nexus Bar (a new Type B bar in the subject device) is 2.5 mm, which is within the stated range of 3*-22 mm, where * indicates variability for this new bar.
**Note for Suprastructure Width: The Minimum Suprastructure Width for the Micro Nexus Bar (a new Type B bar in the subject device) is 2.5 mm, which is within the stated range of 3.4**-12 mm, where ** indicates variability for this new bar.
2. Sample size used for the test set and the data provenance
The submission is a 510(k) for an expansion of an existing device (K221019). The "study" proving the device meets acceptance criteria primarily relies on non-clinical data and demonstration of substantial equivalence to predicate devices.
- Test set/Sample size: The document does not describe a "test set" in the traditional sense of a clinical or imaging study with a defined sample size for statistical analysis. Instead, it relies on:
- Reverse engineering analysis of OEM abutments and OEM abutment screws: This was done to confirm compatibility. The sample size for this analysis is not explicitly stated but would involve the specific OEM abutments listed for compatibility with the subject device.
- Fatigue testing: The document states, "Fatigue testing was not performed since the Subject devices are abutment-borne and are not intended to compensate for angulation in excess of the maximum angulation of OEM angled abutments..." This indicates that engineering analysis and design specifications, rather than physical fatigue testing, were used to demonstrate performance in this regard.
- Data provenance: Not directly applicable as there isn't a "test set" from patients. The data relies on engineering analysis, existing predicate device data, and compliance with standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is not a study requiring expert readers to establish ground truth. Substantial equivalence claims are based on engineering, material, and design comparisons to legally marketed devices.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. There is no "test set" or diagnostic performance study that would require an adjudication method.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a physical dental suprastructure, not an AI-powered diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is a physical dental suprastructure, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" equivalent for this submission is the established safety and effectiveness of the primary predicate device (K221019) and various reference devices, supported by:
- Compliance with recognized standards: Biocompatibility (ISO 10993-5, ISO 10993-12), sterilization validation (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2, ANSI/AAMI/ISO 14937).
- Material specifications: Conformance to ASTM F136 for titanium alloy.
- Engineering analysis and design specifications: Comparison of design features to predicate devices and confirmation of compatibility through reverse engineering.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. As above, there is no training set for this device.
Ask a specific question about this device
(374 days)
The Osstem Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-units restorations including; cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Ultra wide Implant System is intended to be used in the molar region.
Products with diameter of less than 3.25mm should be used exclusively for the lateral incisor in the maxilla and a central or lateral incisor in the mandible.
The Osstem Implant System is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches. The Ultra-Wide implants are intended to be used only to replace molar teeth and angled abutments are not to be used with the Ultra-Wide implants.
Osstem Implant System is similar to other commercially available products based on the intended use, technology used, claims, material composition employed and performance characteristics.
Here's an analysis of the provided text to extract information about acceptance criteria and the supporting study:
The document is a 510(k) summary for the Osstem Implant System, seeking substantial equivalence to predicate devices. As such, it focuses on demonstrating that the new device modifications (primarily additional dimensions and a non-hex connection for some implants) are as safe and effective as the previously cleared predicate devices. Therefore, the "acceptance criteria" discussed are primarily about meeting existing performance standards (like ISO 14801 for fatigue testing) and demonstrating consistency with the predicate devices in terms of materials, manufacturing, indications for use, etc.
Since this is a substantial equivalence submission for a dental implant system involving physical product variations rather than a novel AI/software device, many of the typical AI-specific criteria (like MRMC studies, effect size of AI assistance, standalone algorithm performance, number of experts for ground truth, adjudication methods, training set size/ground truth establishment) are generally not applicable or explicitly stated in the context of this traditional medical device submission.
1. A table of acceptance criteria and the reported device performance
For the specific modifications in the proposed device, the primary acceptance criterion for mechanical performance is successful fatigue testing according to ISO 14801:2016 for the worst-case scenario. Other performance criteria are met by leveraging tests performed on predicate devices, demonstrating equivalence in materials, manufacturing, and design principles.
Acceptance Criteria (Study Performed) | Reported Device Performance |
---|---|
Fatigue Testing (ISO 14801:2016) for worst-case implant-abutment combination (specifically for TSIII SA Implant (Non-Hex)) | Test results demonstrate that the proposed devices perform as intended and support substantial equivalence to the predicate devices. (No specific numerical values are given in this summary.) |
Gamma Sterilization Validation (ISO 11137-1,2,3) (leveraged from K121585) | Met the standards. No change to manufacturing or sterilization processes. |
Shelf-Life Testing (ASTM F1980) (leveraged from K121585) | Met the standards. |
Biocompatibility Testing (ISO 10993-1, -6, -10, -11) (leveraged from K121995) | Met the standards. Subject device is biocompatible and substantially equivalent due to same materials and manufacturing. |
Bacterial Endotoxin Test (ISO 10993-11, USP) (leveraged from K161604) | Met the standards. |
MR Compatibility Review (magnetic displacement force and torque evaluation per FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment") | Non-clinical worst-case MRI review performed using scientific rationale and published literature for the entire system and material composition. Rationale addressed parameters per FDA guidance. (No specific results stated beyond review was performed.) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Fatigue Testing (ISO 14801:2016): The document mentions "worst-case implant-abutment combination" was chosen. ISO 14801 typically specifies minimum sample sizes (e.g., 5 or 10 samples per test condition for static or dynamic testing). The exact number of samples tested for the new Non-Hex implants is not explicitly stated in this summary, nor is the provenance of the test data (e.g., conducted in-house, by a certified lab). However, such tests are generally conducted prospectively on newly manufactured devices or representative samples.
- Other tests (Sterilization, Shelf-Life, Biocompatibility, Bacterial Endotoxin): These were leveraged from predicate devices (K121585, K121995, K161604). The sample sizes and provenance for those original predicate device studies are not detailed in this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable for this type of device submission. This is a mechanical and material performance evaluation, not a diagnostic or interpretative device requiring clinical expert ground truth for a test set. Design requirements and performance standards (like ISO 14801) define the "ground truth" to be met.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable for this type of device submission. Adjudication methods are typically used in clinical or image-based studies to resolve discrepancies among experts.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is a dental implant system, not a diagnostic or AI-assisted device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a hardware medical device; there is no algorithm or AI component for standalone performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For mechanical performance (like fatigue testing), the "ground truth" is defined by international standards (ISO 14801:2016) and applicable FDA guidance documents. These standards specify test methods, failure criteria, and acceptable performance limits for dental implants.
- For material and biological compatibility, the "ground truth" is established by adherence to ISO 10993 standards for biocompatibility, ISO 11137 for sterilization, and ASTM F1980 for shelf-life. These are laboratory-based, objective performance measures.
8. The sample size for the training set
- Not applicable. This is not an AI/ML device that requires a training set. The device design and materials are based on established engineering principles and prior predicate devices.
9. How the ground truth for the training set was established
- Not applicable.