Search Results
Found 327 results
510(k) Data Aggregation
(127 days)
Trade/Device Name: extriCARE® 1000 Negative Pressure Wound Therapy System
Regulation Number: 21 CFR 878.4780
TRADE NAME: | extriCARE® 1000 NPWT System |
| DEVICE CLASSIFICATION: | Class II per 21 CFR 878.4780
The extriCARE® 1000 NPWT System is intended to generate negative pressure to remove wound exudates, infectious material, and tissue debris from the wound bed.
The extriCARE® 1000 NPWT System is intended for use in wounds with low to moderate levels of exudate. Appropriate wound types include:
- Chronic wounds
- Acute wounds
- Traumatic wounds
- Subacute and dehisced wounds
- Partial-thickness burns
- Ulcer wounds (such as diabetic or pressure)
- Flaps and grafts
- Closed surgical incisions
The extriCARE® 1000 NPWT System is intended for use in healthcare facilities. The extriCARE® 1000 NPWT System is intended for use with extriCARE® 1000 Bandage Dressings.
The extriCARE® 1000 Negative Pressure Wound Therapy (NPWT) System consists of a portable pump unit intended to generate vacuum pressure to remove low to moderate level of wound exudates, infectious material, and tissue debris from the wound bed. Three preset vacuum pressures can be selected by the user – 80, 100, and 125mmHg. The extriCARE® 1000 NPWT System is packaged and provided with the following components:
- extriCARE® 1000 Negative Pressure Wound Therapy Pump
- extriCARE® 1000 50cc Collection Canister
- Adapter Power Cord(5VDC)
- extriCARE® 1000 Bandage Dressing/K140634 (this component is provided separately)
N/A
Ask a specific question about this device
(90 days)
/Device Name:** RENASYS WOUND+ Dressing Kit with AIRLOCK Technology
Regulation Number: 21 CFR 878.4780
The RENASYS WOUND+ Dressing Kit with AIRLOCK Technology is intended to be used in conjunction with Smith+Nephew traditional Negative Pressure Wound Therapy (tNPWT) RENASYS systems. The Smith+Nephew RENASYS NPWT system is indicated for patients who would benefit from a suction pump (Negative Pressure Wound Therapy), as it allows for wound management via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials. Appropriate wound types for RENASYS WOUND+ Dressing include:
- Chronic
- Acute
- Traumatic
- Sub-Acute and dehisced wounds
- Ulcers (such as pressure or diabetic)
- Partial-thickness burns
- Flaps
- Grafts
RENASYS WOUND+ Dressing without filler can be used on the following additional wound type:
- Closed surgical incisions
The RENASYS WOUND+ Dressing Kit with AIRLOCK Technology in Small, Medium, and Large sizes is intended to be used with Smith+Nephew RENASYS TOUCH, RENASYS GO, and RENASYS EDGE NPWT systems.
RENASYS WOUND+ Dressing kits with AIRLOCK Technology can be used without a wound filler, with foam or with gauze wound filler. The kits are accessories intended for use in conjunction with RENASYS TOUCH, RENASYS GO and RENASYS EDGE Negative Pressure Wound Therapy systems. RENASYS Negative Pressure Wound Therapy allows for wound management via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.
N/A
Ask a specific question about this device
(270 days)
Trade/Device Name:** Nisus ONE Negative Pressure Wound Therapy System
Regulation Number: 21 CFR 878.4780
Therapy System
COMMON NAME: Nisus ONE NPWT System
DEVICE CLASSIFICATION: Class II per 21 CFR 878.4780
(k) Number:** K243187 | 510(k) Number: K140022 |
| Regulation Number / Product Code: 21 CFR 878.4780
/ OMP | Regulation Number / Product Code: 21 CFR 878.4780 / OMP |
Indications for Use:
| *
The Nisus ONE Negative Pressure Wound Therapy System is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device allows wound management.
The Nisus ONE Negative Pressure Wound Therapy System is only intended to be used with the Cork NPWT Wound Dressing Kit (K132004)
The Nisus ONE Negative Pressure Wound Therapy System is suitable for use in both a professional healthcare facility and home use environment.
Cork Medical has developed a negative pressure wound therapy (NPWT) system with the same intended use as the predicate device (K140022). The subject device features a modified user interface, and an adjusted feature set intended to support ease of use, while maintaining the same fundamental therapeutic function and performance characteristics as the predicate device. The Nisus ONE Negative Pressure Wound Therapy System interface utilizes a single membrane switch keypad to power the device and switch between continuous and intermittent therapy modes.
The device display is limited to presenting critical operational parameters, including therapy mode, battery status, and target pressure settings. Visual and audible alarms are consistent with the predicate device (K140022) and alert critical battery, pressure leakage, system blockage, and full canister occur. While the Nisus ONE NPWT System's interface differentiates from the predicate Cork NPWT system, the mechanical components and therapy application are identical. The new model is intended to enhance accessibility by aligning with the needs of a broad and diverse patient population across various care settings.
The components included within the Nisus ONE NPWT System are:
- Nisus ONE Negative Pressure Wound Therapy Pump (OCMPP-100)
- Nisus ONE Pump Battery Charger (NIS-CHRG-A-18)
- Nisus NPWT Canister 500-mL (CPC-500)
- Nisus NPWT Canister 250-mL (CPC-250, previously cleared in K140022)
Accessory components are required to operate the device. Injection molded components are sonically welded to form a canister designed to mate with the pump and collect excess exudates, infectious material, and tissue debris. The Nisus NPWT Canister 250-mL design was previously cleared in the Nisus NPWT system 510k application (K140022). The Nisus NPWT Canister 500-mL is based on the same design as the Nisus NPWT Canister 250-mL. The dimensions were enlarged to provide a larger reservoir for patient convenience. Thus, the subject and predicate canisters have substantially equivalent characteristics.
The Wound Kit comes in multiple iterations containing a minimum of one port pad, wound foam and peel and stick drape. Integrated with the pump mechanism, these components ensure the safe and efficient removal of wound exudate. NPWT Accessories were previously cleared in K132004. Our application for the Nisus ONE NPWT system includes no revisions to these previously cleared components.
The provided FDA 510(k) clearance letter for the Nisus ONE Negative Pressure Wound Therapy System does not contain information related to an AI/ML-based device performance study.
The document details the substantial equivalence of the Nisus ONE system to a predicate device, focusing on changes to the user interface, physical dimensions, and technical specifications of a powered suction pump. The "Nonclinical Tests" section describes bench performance testing of the pump's functionality (e.g., pressure, alarms, flow rate) using simulated wound exudate and a test bed fixture, as well as software verification and validation, electrical safety, EMC, and usability testing. These are standard engineering and regulatory tests for medical devices, not performance studies of an AI/ML algorithm.
Therefore, I cannot fulfill your request for:
- A table of acceptance criteria and reported device performance for an AI/ML algorithm.
- Sample size, data provenance, expert qualifications, and adjudication methods for a test set related to an AI/ML algorithm.
- Details on MRMC studies or standalone AI performance.
- Information on training set size or ground truth establishment for an AI/ML algorithm.
The document explicitly states: "No Clinical Testing was required to support these 510(k) submissions. No clinical testing has been performed." This further indicates that no studies involving human data or expert review for AI/ML performance were conducted or reported in this submission.
If you have a document describing an AI/ML device, please provide that, and I would be happy to analyze it according to your criteria.
Ask a specific question about this device
(200 days)
K243638**
Trade/Device Name: MV Flow Aspiration Pump (MVFLOWPMP01)
Regulation Number: 21 CFR 878.4780
| Classification Name | Apparatus, Suction, Ward Use, Portable, Ac-Powered |
| Regulation Number | 878.4780
The MV Flow Aspiration Pump is intended for general suction use in hospitals or clinics.
The MV Flow Aspiration Pump is an electrically powered (100 VAC/50 Hz, 240 VAC/60 Hz) device designed to generate vacuum pressure and serve as the vacuum source intended for general suction use in hospitals or clinics. The MV Flow Aspiration Pump is not intended for use in transport or field settings. The pump delivers 29 inHg of maximum vacuum pressure with a maximum flow rate of 36 L/min. The power button and vacuum-relief button provide pressure control. The brushless DC, dual-diaphragm pump design includes reverse-flow protection. Compatible pump accessories, supplied separately, include the MV Flow Canister. The MV Flow Canister consists of a single-use collection canister with integrated bacteria filter, overflow protection, and a pre-installed intermediate tubing.
The provided 510(k) clearance letter is for the MV Flow Aspiration Pump, which is a medical device, not an AI/ML-driven diagnostic or therapeutic device. As such, the information you've requested regarding acceptance criteria, study methodologies, and ground truth establishment (which are typically relevant for AI/ML device clearances) is not applicable to this particular device.
The clearance for the MV Flow Aspiration Pump focuses on demonstrating substantial equivalence to a predicate device (Penumbra Pump MAX) based on non-clinical performance testing, electrical safety, EMC/EMI compliance, and device-specific pump standards. There are no mentions of AI/ML components, image analysis, human reader studies, or complex ground truth establishment processes as would be found in clearances for AI-powered diagnostic tools.
Therefore, I cannot provide the requested information in the format of a table of acceptance criteria for AI performance, sample sizes for test/training sets, expert adjudication methods, or MRMC studies, as these types of studies were not conducted or required for the clearance of this specific device.
If you have a document related to an AI/ML medical device, I would be happy to analyze it according to your requested criteria.
Ask a specific question about this device
(28 days)
(BN007); Baby Nasal Aspirator (BN008); Baby Nasal Aspirator (BN009)
Regulation Number: 21 CFR 878.4780
(BN007); Baby Nasal Aspirator (BN008); Baby Nasal Aspirator (BN009)
Regulation Number: 21 CFR 878.4780
The Baby Nasal Aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.
Not Found
I am sorry, but the provided FDA 510(k) Clearance Letter for the "Baby Nasal Aspirator" does not contain information about acceptance criteria and a study proving the device meets those criteria.
This document is a formal clearance letter issued by the FDA, confirming that the device is substantially equivalent to legally marketed predicate devices. It focuses on regulatory aspects such as:
- Device Identification: Trade name, regulation number, product code, regulatory class.
- Substantial Equivalence Determination: The FDA's finding that the device is equivalent to existing legal devices.
- Regulatory Requirements: Information about general controls (registration, listing, GMP, labeling, misbranding, adulteration), additional controls for Class II/III devices, Quality System (QS) regulation, medical device reporting, Unique Device Identification (UDI) system, and contact information for regulatory advice.
- Indications for Use: The stated purpose and target user group for the device.
The letter explicitly states: "We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..."
The letter does not include details on:
- Specific performance acceptance criteria for the device itself (e.g., suction pressure, battery life, noise levels).
- Any studies or test results demonstrating the device meets such criteria.
- Information about test set sample sizes, data provenance, expert ground truth establishment, or clinical study methodologies (like MRMC studies) as these are typically associated with performance data for more complex or AI/imaging devices, not simple mechanical devices like a nasal aspirator.
For a device like a baby nasal aspirator, the "acceptance criteria" and "proof of meeting them" would typically involve:
- Bench testing: Measuring suction pressure, flow rate, durability, material biocompatibility, and safety (e.g., absence of sharp edges, proper battery compartment sealing).
- Biocompatibility testing: To ensure materials are safe for contact with human tissue.
- Electrical safety testing: If powered, to ensure it meets relevant electrical safety standards.
- Usability testing: To ensure it can be safely and effectively used by the intended users (parents or caregivers).
However, none of these specific test results or their underlying methodologies are detailed in the provided FDA clearance letter. The letter signifies regulatory clearance based on a review of the submitted documentation, which would have included these types of test reports, but it does not reproduce them within the letter itself.
Ask a specific question about this device
(137 days)
Aspirator (NASA601,NASA602,NASA603,NASA605,NASA606,NASA608,NASA610)
Regulation Number: 21 CFR 878.4780
Powered Suction Pump
Regulatory Class: II
Product Code: BTA
Regulation Number: 21 CRF 878.4780
Desida Technology Co., Ltd. | Shenzhen Desida Technology Co., Ltd. | / |
| Regulation number | 21 CFR § 878.4780
| 21 CFR § 878.4780 | Same |
| Product code | BTA | BTA | Same |
| Device classification | Class II
The Electric Nasal Aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2~12 years old). This device is used in a home environment.
Electric Nasal Aspirator consists of silicone nozzle which has three models (conical nozzle, flat nozzle, angled nozzle), collection cup, air tube, tweezers, charging cable-USB and aspirator body. The Electric Nasal Aspirator is a portable device which is intended for suction of nasal passages in children 2-12 years of age. The motor pump provides a negative pressure which removes nasal secretions. The motor pump operates on a rechargeable battery. The rechargeable battery can be charged from the external power adapter (not included in this device) through the provided charging line.
The user interface consists of buttons and LED display, and the user can control the vacuum pressure through the button. Three different shapes of silicone nasal tips are provided to enable easier and more effective removal of the nasal mucus.
The provided text is a 510(k) clearance letter and summary for an Electric Nasal Aspirator. It details the device's characteristics and its comparison to a predicate device, along with the performance data submitted to demonstrate substantial equivalence.
However, the nature of this medical device (a nasal aspirator, essentially a suction pump) means the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are focused on engineering performance parameters (e.g., vacuum pressure, flow rate, safety standards) rather than clinical efficacy studies with human subjects that measure the device's ability to diagnose or treat a condition, which is what is typically discussed in the context of "human-in-the-loop performance," "ground truth," "expert consensus," and "MRMC studies" for AI/software-as-a-medical-device (SaMD).
Therefore, the requested information elements related to AI/SaMD (e.g., training set, test set, expert adjudication, MRMC studies, standalone performance, ground truth for image analysis) are not applicable to this specific device submission. The "study" here is a series of bench tests and electrical/biocompatibility assessments.
Below is a breakdown of the applicable acceptance criteria and the "study" (bench testing) as described in the provided text, while explicitly stating when an element is not applicable.
Acceptance Criteria and Device Performance for Electric Nasal Aspirator (K244033)
1. Table of Acceptance Criteria and Reported Device Performance
| Feature/Test | Acceptance Criteria (Predicate Device Performance) | Reported Device Performance (Subject Device) | Compliance/Remark |
|---|---|---|---|
| Intended Use | Intermittent removal of nasal secretions and mucus from children (age 2~12 years old) in a home environment. | Intermittent removal of nasal secretions and mucus from children (age 2~12 years old) in a home environment. | Same, Meets. |
| Patient Population | Age 2-12 years old | Age 2-12 years old | Same, Meets. |
| Vacuum Pressure | 52-60 KPa | 52-60 KPa | Same, Meets. |
| Music Function | Yes | Yes | Same, Meets. |
| Light Function | Yes | Yes | Same, Meets. |
| Motor Type | 3.7V DC | 3.7V DC | Same, Meets. |
| Operating Condition | 5℃(41℉) to 40℃(104℉); 15% to 93% R.H. | 5℃(41℉) to 40℃(104℉); 15% to 93% R.H. | Same, Meets. |
| Storage Condition | -10°C(-23℉) to 70°C (158℉); 10% to 95% R.H. | -10°C(-23℉) to 70°C (158℉); 10% to 95% R.H. | Same, Meets. |
| Expected Service Life | 2 years | 2 years | Same, Meets. |
| Type BF Applied Part | Type BF applied part | Type BF applied part | Same, Meets. |
| Water-Resistance | IP22 | IP22 | Same, Meets. |
| Biocompatibility | Tested to ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Skin Sensitization), ISO 10993-23 (Skin Irritation) for silicone nasal aspiration tips (contact < 24 hrs). | Passed ISO 10993-5, ISO 10993-10, ISO 10993-23 for silicone nasal aspiration tips. | Passed tests, Meets. |
| Electrical Safety | Tested to IEC 60601-1, IEC 60601-1-11. | Passed IEC 60601-1, IEC 60601-1-11. | Passed tests, Meets. |
| Electromagnetic Compatibility (EMC) | Tested to IEC 60601-1-2. | Passed IEC 60601-1-2. | Passed tests, Meets. |
| Battery Safety | Tested to IEC 62133-2. | Passed IEC 62133-2. | Passed tests, Meets. |
| Software Verification & Validation | Consistent with basic documentation level; all software requirements met, hazards mitigated. | System validation testing completed, requirements met, hazards mitigated. | Meets. |
| Appearance (Color, Dimension, Weight) | - (Compared to predicate, but specific criteria not given as numerical range) | Different from predicate, but product performance test met requirements. | Met requirements through performance testing. |
| Flow Rate | 2.5-2.7 L/min (Predicate) | 3.3 - 6.0 L/min (Subject Device) | Higher flow rate in subject device, but same vacuum pressure (52-60 kPa). Performance testing verified requirements. Does not raise issues. |
| Power Consumption | 2W (max) (Predicate) | 5W (Subject Device) | Different, but complies with safety standards (IEC 62133-2, IEC 60601-1, IEC 60601-1-2) so it does not raise safety/effectiveness issues. |
| Power Source | DC 3.7 V / 700mAh Rechargeable Li-ion battery (Predicate) | DC 3.7 V / 2500mAh Rechargeable Li-ion battery (Subject Device) | Different, but complies with safety standards (IEC 62133-2, IEC 60601-1, IEC 60601-1-2) so it does not raise safety/effectiveness issues. |
| Structural Composition | Silicone nozzle, collection cup, charging cable-USB (Predicate) | Silicone nozzle, collection cup, air tube, tweezers, charging cable-USB and aspirator main body. | Different (subject device has air tube and tweezers). Performance tests support safety/effectiveness of air tube. Tweezers are an accessory and do not raise safety/effectiveness concerns. |
| Tubing Bending Resistance | (Implied standard for performance test) | Assured performance (Product Performance Test). | Met. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated as a number of devices tested. The studies described are type testing (verification and validation) against engineering standards (e.g., IEC, ISO). This implies a representative sample of manufactured devices were tested to ensure design compliance. For biocompatibility, electrical safety, EMC, battery safety, and software V&V, these are typically conducted on a limited number of devices or components to demonstrate compliance with the standards' requirements for design verification.
- Data Provenance: The device manufacturer, Shenzhen Desida Technology Co., Ltd. (China), performed these tests. The data provenance is internal testing to support the 510(k) submission. These are bench tests, not clinical data.
- Retrospective/Prospective: Not applicable as these are bench tests and not clinical studies involving patient data collected over time.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
- Not Applicable. For this type of medical device (Electric Nasal Aspirator), "ground truth" as typically applied to clinical or diagnostic AI/SaMD studies (e.g., for image interpretation) is not established through expert consensus on a test set. The validation relies on meeting pre-defined engineering and safety standards. The "experts" involved would be accredited test houses and engineers specializing in the relevant standards (e.g., electrical safety, biocompatibility).
4. Adjudication Method for the Test Set
- Not Applicable. Adjudication methods (e.g., 2+1, 3+1) are used in clinical studies, particularly for establishing ground truth in diagnostic accuracy assessments involving human interpretation. This device's validation focuses on engineering performance, not interpretative accuracy.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is not an AI/SaMD device, and no MRMC study was conducted. The device is a physical product for suction, not a diagnostic or interpretive tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is not an algorithm-based device. Performance testing for this device is "standalone" in the sense that the device itself is tested against its specifications and relevant safety standards, without continuous human operation being part of the assessment of device performance (though a human operates it to perform the tests).
7. The Type of Ground Truth Used
- Engineering and Safety Standards: The "ground truth" for this device's acceptance is adherence to established international and national engineering standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62133-2, ISO 10993 series) and the manufacturer's own verified product requirements derived from these standards and the predicate device's performance. For biocompatibility, this typically involves laboratory assays confirming the absence of cytotoxicity, skin sensitization, and irritation based on ISO 10993 protocols. For physical properties, it involves measurements against specified ranges (e.g., 52-60 KPa vacuum pressure).
- Not clinical outcomes data, pathology, or expert consensus (in the traditional sense for diagnostic/AI devices).
8. The Sample Size for the Training Set
- Not Applicable. This device is a physical product, not an AI/Software device requiring a "training set."
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. As there is no "training set" for this physical device, there is no ground truth established for it in the context of machine learning. The device's design and manufacturing processes are validated against engineering principles and standards.
Ask a specific question about this device
(26 days)
Blvd
Warren, New Jersey 07059
Re: K251111
Trade/Device Name: Cary
Regulation Number: 21 CFR 878.4780
Classification Name:* Apparatus, Suction, Ward Use, Portable, Ac-Powered
Regulation Number: 878.4780
Cary is indicated for removal of fluids from the upper airway. The device creates a negative pressure (vacuum) that draws fluids through disposable suction tube that is connected to a disposable collection canister. The fluids are captured in the collection canister for proper disposal. Cary is for use only on the order of a physician or other licensed practitioner (e.g. Emergency Medical Technician in field use).
[Cary] is a handheld, electrically powered medical suction device used to clear the upper airway during emergency care. It consists of a reusable drive unit and disposable collection container (pumpister) and tubing. The drive unit powers a piston pump, creating negative pressure (vacuum) in the disposable collection container. The vacuum draws fluids and other matter through a disposable tubing connected to the container, where they are collected for disposal. The size and shape of the device enable use by one hand, and a rechargeable lithium-ion battery allows field and transport use of the device.
The provided FDA 510(k) clearance letter and summary for the "Cary" device ([K251111](https://510k.innolitics.com/search/K251111)) primarily discuss non-clinical performance testing to demonstrate substantial equivalence to a predicate device. This document does not describe a clinical study (e.g., MRMC or standalone human-in-the-loop study) involving human readers or expert panels for establishing ground truth, as would typically be seen for AI/ML-driven diagnostic devices. Instead, the testing focuses on the physical and functional performance of the suction pump itself against recognized consensus standards.
Therefore, many of the requested points regarding clinical study design, expert panels, ground truth, and MRMC studies are not applicable based on the provided text.
Here's an analysis based on the information provided:
1. Table of Acceptance Criteria and Reported Device Performance
The document describes several non-clinical tests with implied or explicit acceptance criteria based on industry standards and comparison to the predicate device.
| Test Category | Acceptance Criteria (Implied/Explicit from Standards/Predicate) | Reported Device Performance |
|---|---|---|
| Operating Position | Ability to operate in any spatial orientation. | Confirmed. |
| Protection Devices | Compliance with overfill protection and reverse flow protection. | Results within acceptance criteria. |
| Noise Level | Unspecified standard, implied safe/acceptable level. | 70.5 dB. "Within the acceptance criteria apart from 0.5dB excess when the suction tip was blocked. This is considered as an acceptable deviation from the standard requirement as it does not decrease the safety or effectiveness of the subject device when compared to the predicate device." |
| Air Leakage | As per standard requirements for general use suction equipment. | Results within acceptance criteria. |
| Vacuum Levels & Free Air Flow | Within acceptance criteria for High vacuum/high flow and Low vacuum/low flow (similar to predicate: High vacuum <600 mmHg vs <550 mmHg; Low vacuum ≤100 mmHg vs 50 mmHg; High flow ≥ 20 l/min vs 30 l/m). | Results within acceptance criteria. Suction performance "similar to LCSU 4". |
| Pharyngeal Suction Equipment | Ability to evacuate ≥ 200 ml of simulated vomitus in ≤ 10 s. | Results within acceptance criteria. |
| Reprocessing Validation | Appropriateness of manual reprocessing process; user ability to carry out instructions. | Outcome within acceptance criteria. |
| Biocompatibility | Compliance with ISO 10993 standards for cytotoxicity, sensitization, irritation, acute systemic toxicity, and material mediated pyrogenicity. | Confirmed for direct-contact materials. |
| Software | Compliance with IEC 62304 and FDA guidance for software functions, classified as Class B. | Verification and validation conducted, documentation provided as recommended. (Specific performance metrics are not applicable here as it's about compliance, not a clinical outcome). |
| EMC and Electrical Safety | Compliance with IEC 60601-1-2 and IEC 60601-1 series. | Confirmed. |
2. Sample Size Used for the Test Set and Data Provenance
The document describes non-clinical, laboratory-based testing of the device's physical and functional performance, not a clinical study involving a dataset of patient cases. Therefore, "sample size for the test set" and "data provenance" in the context of patient data are not applicable. The tests were performed on the device prototypes/production samples in controlled laboratory environments.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. Ground truth, in the context of device performance testing for a suction pump, refers to established engineering specifications, physical laws, and recognized standards (e.g., ISO 10079-1). There are no "experts" in the sense of medical professionals interpreting patient data to establish ground truth for this type of device.
4. Adjudication Method for the Test Set
Not applicable. As this is non-clinical performance testing against engineering standards, an adjudication method for conflicting interpretations of medical data is not relevant.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. The document explicitly describes non-clinical testing and comparison to a predicate device based on device performance metrics (e.g., vacuum levels, flow rates, noise), not on human reader performance with or without AI assistance. This device is a powered suction pump, not an AI diagnostic/imaging device that would typically undergo MRMC studies.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Not applicable. The device is a mechanical suction pump, not an AI algorithm. Its performance is inherent to its physical design and operation, not an output from an algorithm in the typical AI sense.
7. The Type of Ground Truth Used
The ground truth used for these non-clinical tests is based on:
- Recognized Consensus Standards: e.g., ISO 10079-1, ISO 10079-4, IEC 62304, ISO 10993, IEC 60601 series.
- Engineering Specifications: Explicit performance requirements (e.g., ≥ 20 l/min flow rate, < 1 kPa air leakage).
- Comparison to Predicate Device Performance: Establishing equivalence or superiority in key performance areas.
8. The Sample Size for the Training Set
Not applicable. There is no "training set" as this is not an AI/ML device that requires machine learning training.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for an AI/ML algorithm.
Ask a specific question about this device
(15 days)
07059
Re: K251101
Trade/Device Name: AIRO Suction Unit (AIRO-01)
Regulation Number: 21 CFR 878.4780
07059
Re: K251101
Trade/Device Name: AIRO Suction Unit (AIRO-01)
Regulation Number: 21 CFR 878.4780
Classification Name: Pump, Portable, Aspiration (Manual Or Powered)
Regulation Number: 878.4780
The AIRO Suction Unit, Model AIRO-01, is a portable, battery-powered medical suction device intended for use in emergency airway management procedures by professional medical providers. It is intended for intermittent operation to remove secretions, such as blood or vomit, during airway management procedures and other procedures for which an intermittent suction medical device may be applicable. The device is indicated for the suctioning of adult patients.
The AIRO Suction Unit, Model AIRO-01, is a compact, handheld, portable battery-powered medical suction device intended for use in emergency airway management procedures by professional medical providers such as emergency medical technicians, paramedics, and hospital personnel. The device is not intended for contact with patient tissue during procedures. It is intended for intermittent operation to remove secretions, such as blood or vomit, during airway management procedures and other procedures for which an intermittent suction medical device may be applicable. The device is indicated for the suctioning of adult patients.
The provided FDA 510(k) clearance letter for the AIRO Suction Unit (AIRO-01) is for a powered suction pump. Such devices are typically evaluated based on their physical and functional performance rather than AI/ML algorithm performance.
Therefore, many of the requested categories related to AI/ML specific studies (like MRMC studies, sample sizes for AI training sets, ground truth establishment for AI, number of experts for ground truth, adjudication methods) are not applicable to this type of medical device submission.
Here's a breakdown of the available information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document states the device was tested to ISO 10079-1 and ISO 10079-4, and all applicable clauses were successfully executed, meeting the established criteria. However, the specific quantitative acceptance criteria and detailed performance results are not explicitly provided in the excerpt. We can infer the performance from the device description and the ISO standards it complies with.
| Acceptance Criterion (Inferred from ISO 10079-1:2015 and device description) | AIRO Suction Unit (AIRO-01) Reported Performance |
|---|---|
| Vacuum Pressure Range: | 20-550 mmHg |
| Low Vacuum Level (≤ 150 mmHg / ≤ 20 kPa) compliance: | Compliant (device achieves a continuous range of vacuum pressures from 20 mmHg) |
| Medium Vacuum Level (150 - 450 mmHg / 20 - 60 kPa) compliance: | Compliant (device achieves a continuous range of vacuum pressures between 150-450 mmHg) |
| High Vacuum Level (≥ 450 mmHg / ≥ 60 kPa) compliance: | Compliant (device achieves a continuous range of vacuum pressures up to 550 mmHg) |
| Low Flowrate Device (Free air flowrate < 20 LPM) compliance: | Achieves all vacuum levels while maintaining a low flowrate of less than 20 LPM (nominal airflow rate given as 0.75 LPM in comparison table, which is < 20 LPM) |
| Electrical Safety/Electromagnetic Compatibility (ES/EMC): | Tested to applicable standards (not specified which ones, but implied by ES/EMC testing mentioned) |
| Biocompatibility: | Tested to applicable standards (not specified which ones, but implied by biocompatibility testing mentioned) |
| Use-Life: | Tested to applicable standards (implied by Use-Life testing mentioned) |
| Reprocessing: | Tested to applicable standards (implied by reprocessing testing mentioned) |
| Software Verification & Validation (V&V): | Tested to applicable standards (implied by software V&V testing mentioned) |
| Packaging Integrity (ISTA standards): | Packaging is sufficient to maintain device functionality according to ISTA standards |
| Usability (Performance for target user group): | Results indicate the device performs adequately for the target user group (professional medical providers in emergency airway management) |
| Water Protection (IP rating): | IP33 (Identical to reference device TACVAC K153459, similar to predicate S-SCORT Jr Quickdraw Model 2400 K041154 which has IPX4) |
| Canister Capacity: | 500 mL (Same as reference device TACVAC K153459, different from predicate S-SCORT Jr Quickdraw Model 2400 K041154 which has 300 mL) |
| Compatibility with standard third-party disposable tubing/catheters: | Compatible with Standard Disposable Suction Tubing (9/32" ID x 36" L) and Standard Disposable Suction Catheters (e.g., Yankauer, DuCanto, others) – This is a functional requirement for usability and integration with existing medical practice. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not specified in the provided text for any of the non-clinical tests (e.g., ES/EMC, biocompatibility, use-life, reprocessing, software V&V, distribution, usability). For physical device testing, sample sizes are often determined by relevant standards (e.g., ISO, ASTM) and internal quality protocols.
- Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. These are typically internal corporate studies.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable for this type of device (powered suction pump). Ground truth is not established by expert consensus in the way it is for diagnostic AI devices. Performance is measured against engineering specifications and international standards (like ISO 10079-1).
For the usability study, it "employed users expected to operate the device." This implies medical professionals, but their number and specific qualifications are not detailed.
4. Adjudication Method for the Test Set
Not applicable for this type of device. Adjudication methods (like 2+1, 3+1) are relevant for subjective interpretations, often in diagnostic studies with human readers. Device performance is measured objectively against defined engineering and safety standards.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. MRMC studies are used to assess the impact of AI on human reader performance, typically in diagnostic imaging. This device is a physical medical instrument, not a diagnostic AI tool.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Not applicable. This device is a physical medical instrument, not an AI algorithm. While it has "software V&V" mentioned, this likely refers to embedded firmware controlling the physical function of the pump, not a diagnostic or decision-support algorithm.
7. Type of Ground Truth Used
Engineering Specifications and International Standards: The "ground truth" for this device's performance is compliance with established engineering specifications (e.g., vacuum pressure range, airflow rate, battery life) and recognized international standards such as ISO 10079-1 (for medical suction equipment), ISO 10079-4 (likely for specific aspects like safety or cleaning), and ISTA (for packaging validation).
8. Sample Size for the Training Set
Not applicable. This device is not an AI/ML system that requires a "training set" in the conventional sense.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As there is no "training set" for an AI/ML algorithm in this context, the method for establishing its ground truth is irrelevant.
Ask a specific question about this device
(47 days)
China
Re: K250697
Trade/Device Name: Electric Nasal Aspirator
Regulation Number: 21 CFR 878.4780
suction pump
Classification Name: Pump, portable, aspiration (powered)
Regulation: 21 CFR § 878.4780
BC023 | Electric nasal aspirator, model: BC026 | / |
| Classification | Class II Device, BTA (21 CFR § 878.4780
) | Class II Device, BTA (21 CFR § 878.4780) | Identical |
| Classification Panel | General & Plastic
The Electric Nasal Aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.
Electric nasal aspirator (model: BC026, BC025 and BC023) consists of main unit, and suction portion working together as one unit. The Electric nasal aspirator is a portable device which is intended for suction of nasal passages in children 2-12 years of age. The motor pump provides a negative pressure which removes nasal secretions. The motor pump operates on a rechargeable battery. The rechargeable battery can be charged from the external power adapter (not included in this device) through the provided charging line. The user interface consists of buttons and LED display, and the user can control the vacuum pressure through the button.
The differences between BC026, BC025, and BC023 are as follows:
- BC026 has 5 suction levels, while BC023 and BC025 have 3 suction levels.
- BC026 and BC023 have "Light" function for distracting and pacifying the child. BC025 does not have this function.
- BC026 and BC023 can pause suction during the suction process. BC025 does not have this feature.
- BC026 and BC023 have 4 buttons, BC025 has 2 buttons.
- In addition, BC026, BC025, and BC023 have different Display, Size, Weight, and Appearance.
The provided FDA 510(k) clearance letter and summary for the Electric Nasal Aspirator do not contain specific acceptance criteria for device performance such as sensitivity, specificity, accuracy, or other diagnostic metrics. This device is a powered suction pump and not a diagnostic device that would typically have such performance criteria. The clearance is based on demonstrating substantial equivalence to a legally marketed predicate device, primarily through non-clinical safety and performance testing.
Therefore, many of the requested fields cannot be extracted directly from the provided text as they pertain to clinical performance studies (e.g., sample size for test set, number of experts, adjudication method, MRMC studies, standalone performance, ground truth establishment) that were explicitly not performed.
Here's the information that can be extracted, and an explanation for what cannot:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Test Type) | Standard Designation Number / Description | Reported Device Performance |
|---|---|---|
| Safety | ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] | Conforms |
| EMC (Electromagnetic Compatibility) | IEC 60601-1-2:2020 IEC TR 60601-4-2: 2016 | Conforms |
| Home healthcare environment | IEC 60601-1-11:2020 | Conforms |
| Performance | Enterprise standard | Conforms |
| Biocompatibility | ISO 10993-1:2018 | Conforms |
| Software | ANSI AAMI IEC 62304:2006/A1:2016 | Conforms |
| Safety of Lithium battery | IEC 62133-2:2017 | Conforms |
| Safety of lamps | IEC 62471: 2006 | Conforms |
| Vacuum pressure | 52-60 Kpa (Predicate comparison) | 52-60 Kpa (Subject device) - Identical to predicate |
| Noise Level | <80 dBA (Predicate comparison) | <80 dBA (Subject device) - Identical to predicate |
| Power consumption | 2.2W (Predicate comparison) | 2.2W (Subject device) - Identical to predicate |
| Motor Type | 3.7V DC (Predicate comparison) | 3.7V DC (Subject device) - Identical to predicate |
| Power Source | DC 3.7 V / 700mAh Rechargeable Li-ion battery (Predicate comparison) | DC 3.7 V / 700mAh Rechargeable Li-ion battery (Subject device) - Identical to predicate |
| Expected service life | 2 years (Predicate comparison) | 2 years (Subject device) - Identical to predicate |
| Water-resistance | IP22 (Predicate comparison) | IP22 (Subject device) - Identical to predicate |
The study that proves the device meets the acceptance criteria is the Non-clinical testing which included product service life, software validation, electromagnetic compatibility and electrical safety, and function tests. The results demonstrated that the Electric Nasal Aspirator meets the requirements of its predefined acceptance criteria and intended use.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable / not provided. The submission explicitly states: "No clinical test data was used to support the decision of substantial equivalence." The testing performed was non-clinical (engineering and safety standards conformance) on the device itself, not on patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable / not provided. As above, no clinical studies with human "ground truth" were conducted.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable / not provided. No clinical studies with adjudication were performed.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable / not provided. The device is an Electric Nasal Aspirator, a hardware device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable / not provided. The device is a physical product, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
This information is not applicable / not provided. "Ground truth" in the context of clinical studies or AI algorithm evaluation is not relevant as no such studies were conducted. The "truth" for this device's performance is derived from conformance to engineering and safety standards.
8. The sample size for the training set
This information is not applicable / not provided. This device is not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
This information is not applicable / not provided. As above, no training set or ground truth in that context was established.
Ask a specific question about this device
(259 days)
Disposable strap for ATMOS C 051 Thorax (316.1200.0): Hose clamp (061.0079.0) Regulation Number: 21 CFR 878.4780
Classification Name: | Pump, Portable, Aspiration (Manual Or Powered) |
| ■ | Regulation Number: | 878.4780
The ATMOS C 051 Thorax is indicated for aspiration and removal of surgical fluids, tissue, gases, bodily fluids, or infectious material from a patient's respiratory system after surgery. The general indication of the ATMOS C 051 Thorax is thoracic drainage.
The ATMOS C 051 Thorax (317.0200.0) drainage system is a device for mobile, digital thoracic drainage. It is portable, mains-independent and has an electronic monitoring system with optical and acoustic display. The device takes medical effect as a system in combination with a secretion canister and a hose system. The system creates the (natural) vacuum in the pleural cavity by draining off air and secretion. The ATMOS C 051 Thorax is equipped with a rechargeable battery. A charging unit which is located within the suction device guarantees the secure charging of the battery. The hose system is rinsed with air at regular intervals to prevent the collection of debris in the hose. Bacterial and viral filters in the secretion canister and measuring hose prevent contaminated secretions from entering the device. The device is equipped with a disposable strap and a carrying handle. These enable mobility. A universal bracket or the standard rail bracket can be ordered separately as accessories. The ATMOS C 051 Thorax is a reusable device delivered in non-sterile state. The secretion canister and hose systems are single-use devices and delivered sterile. All other accessories subject to this submission are reusable and delivered non-sterile.
The provided document is a 510(k) clearance letter from the FDA for a medical device called "ATMOS C 051 Thorax," a powered suction pump for thoracic drainage. While it discusses performance tests and adherence to various standards, it does not contain the detailed clinical study data typically found in a clinical trial report or a comprehensive technical submission.
Therefore, I cannot extract answers for many of your specific questions related to acceptance criteria, ground truth, expert qualifications, adjudication methods, MRMC studies, or training set details from this document. The document primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance bench testing, software validation, biocompatibility testing, human factors validation, and risk management.
However, I can provide information on what is explicitly stated or can be inferred from the document regarding the acceptance criteria and the nature of the study.
Here's what can be extracted:
1. A table of acceptance criteria and the reported device performance
The document mentions that "Performance tests were carried out to ensure that the subject device functions as intended and meet design specifications." and "Risk analysis was carried out based on acceptance criteria established in accordance with ISO 14971." However, the specific numerical acceptance criteria or the direct reported performance values for each criterion are NOT provided in this summary.
The document lists the following categories of performance tests conducted with "no unexpected results or significant deviations":
| Acceptance Criteria Category (Details Not Provided) | Reported Device Performance (Qualitative) |
|---|---|
| Electrical, Mechanical, and Thermal Safety (in accordance with AAMI/ANSI and IEC 60601 standards) | Successfully tested and complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-4-2. |
| Software Validation (in accordance with FDA guidance "Content of Premarket Submissions for Device Software Functions," June 14, 2023) | Performed, documentation level "enhanced," cybersecurity aspects followed FDA guidance. |
| Biocompatibility Testing (according to ISO 10993-1 and FDA guidance) | Performed (details not explicitly stated). |
| Performance Bench Testing: | "No unexpected results or significant deviations." |
| - Maximum flow rate | (Specific values not provided) |
| - Function and accuracy of flow display | (Specific values not provided) |
| - Minimum to maximum vacuum | (Specific values not provided) |
| - Function, accuracy and control delay of vacuum display | (Specific values not provided) |
| - Battery operation at full power mode | (Specific values not provided) |
| - Battery charge time | (Specific values not provided) |
| - Over suction protection | (Specific values not provided) |
| - Weight | Subject device: 1.16 kg (predicate: 3.27 kg) |
| - Alarm sound and noise level | (Specific values not provided) |
| - Length of hose system | (Specific values not provided) |
| Human Factors Validation (in accordance with IEC 62366-1, IEC 60601-1-6, and FDA guidance) | "Use of the device is safe, especially with regard to the use by lay persons. The residual risks that remain...have been reviewed and determined to be acceptable." |
| Risk Management (in accordance with ISO 14971:2019) | "Overall benefits of using the device outweigh the risks. The device is safe and effective for use in the intended patient population." |
| Reprocessing, Sterility, and Shelf-Life (for certain components) | Cleaning and disinfecting procedures described. Sterile validation performed for secretion canister and hose systems according to ISO 11135. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document primarily describes bench testing and validation activities rather than a typical clinical study with a "test set" of patient data. Therefore, the concept of sample size as it applies to patient data is not relevant here. The "tests" refer to engineering and quality assurance activities. No country of origin for data or retrospective/prospective nature is mentioned as it's not a clinical data study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided as there is no mention of clinical "ground truth" derived from expert interpretation in this document. The tests performed are technical and engineering-focused (e.g., flow rates, vacuum levels, battery performance), with human factors validation likely involving usability experts or human factors engineers, but their specific qualifications and number are not detailed.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided as it's relevant for clinical adjudication of medical images or diagnoses, not for the technical bench testing described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was conducted or mentioned. This device is a powered suction pump, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. The device is a physical medical device (suction pump), not a software algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the tests mentioned are engineering specifications, international standards (IEC, ISO), and regulatory guidance documents (FDA guidance). For example, for flow rate, the "ground truth" would be the measured flow rate compared against the predefined design specification. For human factors, it would be the assessment of safety and effectiveness of use against defined usability goals and risk analysis.
8. The sample size for the training set
This question is not applicable. This is not an AI/machine learning device, so there is no "training set."
9. How the ground truth for the training set was established
This question is not applicable for the reason stated above.
In summary, the provided document details a 510(k) submission for a non-AI medical device (powered suction pump). The "studies" proving it meets acceptance criteria are predominantly non-clinical bench tests and validation activities against established engineering specifications, international standards, and regulatory guidelines, rather than clinical trials that generate patient data and require expert ground truth establishment.
Ask a specific question about this device
Page 1 of 33