K Number
K243418

Validate with FDA (Live)

Date Cleared
2024-12-02

(28 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Riptide™ Aspiration Pump is intended as a vacuum source for Medtronic aspiration devices or systems for use in hospitals or clinics. Refer to the Instructions for Use included with the accessory aspiration device for indications related to the procedure for which this device will be used.

Device Description

The Riptide™ Aspiration Pump is an externally powered electromechanical device capable of generating a vacuum designed for drawing fluids and thrombus into the Riptide™ Collection Canister during interventional procedures. It is intended for general suction use in hospitals or clinics and is not intended for transport or field applications. The Riptide™ Aspiration Pump includes a receptacle that holds the Riptide™ Collection Canister with Intermediate Tubing in place. The intermediate tubing is connected to the vacuum inlet port.

The Riptide™ Aspiration Pump package includes the following:

  • Riptide™ Aspiration Pump (LMT-RAP)
  • Region specific power cord for connection to earthed receptacle.
  • Riptide™ Aspiration Pump User's Manual

The following item is required for use with Riptide™ Aspiration Pump (packaged separately):

  • Riptide™ Collection Canister with Intermediate Tubing (LMT-RCT)

The Riptide™ Collection Canister is provided non-sterile and is pre-assembled with the Intermediate Tubing. The Riptide™ Collection Canister with Intermediate Tubing is single use. The Riptide™ Collection Canister is the repository for aspirated material. The Riptide™ Canister is placed into the receptacle of the Riptide™ Aspiration Pump as shown in Figure 3. The Intermediate Tubing is then connected to the vacuum inlet port.

AI/ML Overview

Here's an analysis of the Riptide™ Aspiration Pump's acceptance criteria and the study proving it, based on the provided document:

This document is a 510(k) premarket notification summary, which focuses on demonstrating substantial equivalence to a predicate device rather than comprehensive de novo clinical trial data. Therefore, the device's acceptance criteria are framed in terms of meeting established standards and matching the performance of a legally marketed predicate device.


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Riptide™ Aspiration Pump are primarily based on compliance with international standards for medical electrical equipment and medical suction equipment, as well as demonstrating equivalent performance to the predicate device.

Acceptance Criteria (Standard / Characteristic)Reported Device Performance
Electrical Safety (IEC 60601-1, Edition 3.2 2020-08)Pass
Electromagnetic Compatibility (IEC 60601-1-2, Edition 4.1 2020-09)Pass
User Manual Design Validation (FDA Guidance & ISO 14971, ANSI/AAMI HE75)Pass (evaluated per FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices" and recognized consensus standards ISO 14971 and ANSI/AAMI HE75)
Medical Suction Equipment (ISO 10079-1, Fourth Edition 2022-03)Pass
Medical Suction Equipment (ISO 10079-4, First Edition 2021-08)Pass
Vacuum Range0-29 inHg (Same as predicate device Penumbra Pump MAX™)
Flow Rate (60 Hz US)0-0.8 SCFM (0-23 LPM) (Same as predicate device Penumbra Pump MAX™)
Voltage110-115 Vac (Comparable to predicate device 100-115 Vac)
Frequency60 Hz (US) (Same as predicate device Penumbra Pump MAX™)
Duty CycleNon-continuous: 97% (58.2 minutes on, 1.8 minutes off) (Compared to predicate device 97.8% (45 minutes on, 1 minute off) - changed to match reference device Riptide™ Aspiration System)
Applied Part ClassificationType CF (Same as predicate device Penumbra Pump MAX™; change from Reference Device Type BF to match predicate)
Noise<70 dBa (Compared to predicate device <60 dBa and reference device 60.5 dBa)
Materials of Construction (Collection Canister)Commonly used medical grade plastics (Same as predicate device Penumbra Pump MAX™)
Volume (Collection Canister)1200 mL (Compared to predicate device 1000 mL - matched to reference device Riptide™ Aspiration System)
Shelf Life (Collection Canister)3 years (Compared to predicate device "Unknown" - matched to reference device Riptide™ Aspiration System)
Operating Life500 hours (Same as predicate device Penumbra Pump MAX™)
BiocompatibilityNo impact to biocompatibility (due to no change in materials associated with the proposed changes)

2. Sample Size Used for the Test Set and Data Provenance

The document indicates that the determination of substantial equivalence is based upon non-clinical bench testing. It does not specify a "test set" in the context of clinical data, as no clinical studies were performed.

  • Sample Size for Test Set: Not applicable as no clinical test set was used. The performance evaluation was based on bench tests of the device itself.
  • Data Provenance: The data provenance is from bench testing conducted by Micro Therapeutics Inc. d/b/a ev3 Neurovascular (the applicant). This is internal testing, not data from a specific country or clinical setting, and is prospective in nature (tests performed on the device to establish its performance).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. For this type of 510(k) submission based on non-clinical bench testing, "ground truth" is not established by medical experts for a test set of patient cases. Instead, compliance with established engineering and medical device standards (e.g., IEC, ISO, FDA guidance) serves as the "ground truth" for technical performance.


4. Adjudication Method for the Test Set

Not applicable. There was no clinical test set requiring expert adjudication.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "The determination of substantial equivalence is based upon non-clinical bench testing as there is no change to the intended use, fundamental scientific technology, or materials of construction." Therefore, there is no effect size of human readers improving with or without AI assistance.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

While the device is an electromechanical pump (not an AI algorithm), the performance data presented is standalone performance of the device itself, evaluated through bench testing without human-in-the-loop performance as part of its primary performance evaluation for this submission. The "User Manual Design Validation" does involve human factors, but this is about the usability of the manual, not the device's functional performance itself.


7. The Type of Ground Truth Used

The "ground truth" for this submission is:

  • Compliance with recognized international standards: IEC 60601-1, IEC 60601-1-2, ISO 14971, ANSI/AAMI HE75, ISO 10079-1, and ISO 10079-4.
  • Equivalence to a legally marketed predicate device: Penumbra Pump MAX™ (K122756) in terms of fundamental technological characteristics and performance attributes (vacuum range, flow rate, materials, operating life).
  • Equivalence to a reference device: Riptide™ Aspiration System (K172448) for certain updated characteristics like duty cycle, volume, and shelf life of the collection canister, where these characteristics are deemed equivalent to or improve upon the predicate without raising new safety/effectiveness concerns.

There is no pathology, outcomes data, or expert consensus on patient cases used as ground truth for this mechanical device submission.


8. The Sample Size for the Training Set

Not applicable. This is a physical medical device, an aspiration pump, not a machine learning or AI algorithm. Therefore, there is no "training set."


9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 2, 2024

Micro Therapeutics Inc. d/b/a ev3 Neurovascular Tony Rizzardi Principal Regulatory Specialist 9775 Toledo Way Irvine, California 92618

Re: K243418

Trade/Device Name: Riptide™ Aspiration Pump; Riptide™ Collection Canister with Intermediate Tubing Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: JCX Dated: November 1, 2024 Received: November 4, 2024

Dear Tonv Rizzardi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Long H. Chen
Long H. Chen -S-s
Date: 2024.12.02 09:42:22 -05'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K243418

Device Name

Riptide TM Aspiration Pump;

Riptide™ Collection Canister with Intermediate Tubing

ndications for Use (Describe)

The Riptide™ Aspiration Pump is intended as a vacuum source for Medtronic aspiration devices or systems for use in hospitals or clinics. Refer to the Instructions for Use included with the accessory aspiration device for indications related to the procedure for which this device will be used.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

510(k) OwnerMicro Therapeutics, Inc. d/b/a ev3 Neurovascular9775 Toledo WayIrvine, CA 92618, USAEstablishment Registration: 2029214
Contact PersonTony RizzardiPrincipal Regulatory Affairs SpecialistTelephone: (763) 505-3017Email: tony.rizzardi@medtronic.com
Date Summary Prepared31 October 2024
Trade Name of DeviceRiptide™ Aspiration Pump
Common Name of DeviceApparatus, suction, ward use, portable, ac-powered
Review PanelGeneral & Plastic Surgery
Medical SpecialtyGeneral & Plastic Surgery
Product CodeJCX
Regulation Number21 CFR 878.4780
Regulation NameApparatus, suction, ward use, portable, ac-powered
Device ClassificationClass II
Predicate DevicePenumbra Pump MAX™ (K122756)
Reference DeviceRiptide™ Aspiration System (K172448)

Device Description

Riptide™ Aspiration Pump

The Riptide™ Aspiration Pump is an externally powered electromechanical device capable of generating a vacuum designed for drawing fluids and thrombus into the Riptide™ Collection Canister during interventional procedures. It is intended for general suction use in hospitals or clinics and is not intended for transport or field applications. The Riptide™ Aspiration Pump includes a receptacle that holds the Riptide™ Collection Canister with Intermediate Tubing in place. The intermediate tubing is connected to the vacuum inlet port.

The Riptide™ Aspiration Pump package includes the following:

  • Riptide™ Aspiration Pump (LMT-RAP) ●
  • Region specific power cord for connection to earthed receptacle.
  • . Riptide™ Aspiration Pump User's Manual

The following item is required for use with Riptide™ Aspiration Pump (packaged separately):

  • Riptide™ Collection Canister with Intermediate Tubing (LMT-RCT)

{5}------------------------------------------------

Image /page/5/Figure/1 description: The image shows a diagram of a vacuum pump with labels pointing to different parts. The labels include vacuum control knob, vacuum inlet, vacuum gauge, on/off button, and canister holder. The vacuum pump has a canister holder on the bottom and a vacuum control knob and gauge on the top. The vacuum inlet is on the side of the pump.

Figure 1. Riptide™ Aspiration Pump (Front)

Image /page/5/Figure/3 description: The image shows the back of a machine with labels pointing to different parts. The top of the machine has a "COOLING AIR INTAKE". The bottom of the machine has a "POWER CORD CONNECTION" on the left and a "POTENTIAL EQUALIZATION PLUG" on the right.

Figure 2. Riptide™ Aspiration Pump (Back)

Riptide™ Collection Canister with Intermediate Tubing

The Riptide™ Collection Canister is provided non-sterile and is pre-assembled with the Intermediate Tubing. The Riptide™ Collection Canister with Intermediate Tubing is single use. The Riptide™ Collection Canister is the repository for aspirated material. The Riptide™ Canister is placed into the receptacle of the Riptide™ Aspiration Pump as shown in Figure 3. The Intermediate Tubing is then connected to the vacuum inlet port.

Image /page/5/Picture/7 description: The image shows two medical suction devices. The device on the left has a Riptide Collection Canister connected to the top of the machine via tubing. The device on the right has Intermediate Tubing connected to the top of the machine. Both devices have a gauge on the top and a collection canister on the side.

Figure 3. Riptide™ Collection Canister with Intermediate Tubing

{6}------------------------------------------------

Indications for Use

The Riptide™ Aspiration Pump is intended as a vacuum source for Medtronic aspiration devices or systems for use in hospitals or clinics. Refer to the Instructions for Use included with the accessory aspiration device for indications related to the procedure for which this device will be used.

Comparison of Technological Characteristics

AttributePredicate Device:Penumbra Pump MAX™(PMX110 - K122756)Reference Device:Riptide™ Aspiration System(MAP-1000 - K172448)Subject Device:Riptide™ Aspiration Pump(LMT-RAP)
General
Product Code/RegulationJCX (21 CFR 878.4780)NRY (21 CFR 870.1250)Same as Predicate (K122756)
ClassificationClass IIClass IISame as Predicate (K122756)
ComponentsComprising theSystemPump, collection canister with filter, intermediate tubingPump, collection canister with filter, intermediate tubingSame as Predicate (K122756)
Pump User Manual Characteristics
Indications forUseThe Penumbra Pump MAX™ isintended for general suction use inhospitals or clinics.The Riptide™ Aspiration System isintended for use in therevascularization of patients withacute ischemic stroke secondary tointracranial large vessel occlusivedisease (within the internal carotid,middle cerebral M1 and M2segments, basilar, and vertebralarteries) within 8 hours of symptomonset. Patients who are ineligible forintravenous tissue plasminogenactivator (IV t-PA) or who fail IV t-PAtherapy are candidates for treatment.The Riptide™ Aspiration Pump isintended as a vacuum source forMedtronic aspiration devices orsystems for use in hospitals or clinics.Refer to the Instructions for Useincluded with the accessoryaspiration device for indicationsrelated to the procedure for whichthis device will be used.
User ManualUser manual with safety informationand directions for the PenumbraPump MAX™ for general suction use.User manual with safety informationand directions for the Riptide™Aspiration Pump for neurovascularuse.User manual with safety informationand directions for the Riptide™Aspiration Pump for general suctionuse.
Pump Characteristics
Vacuum Range0-29 inHgSame as Predicate (K122756)Same as Predicate (K122756)
Flow Rate60 Hz (US):0-0.8 SCFM (0-23 LPM)Same as Predicate (K122756)Same as Predicate (K122756)
Voltage100-115 Vac110-115 Vac110-115 Vac
Frequency60 Hz (US)Same as Predicate (K122756)Same as Predicate (K122756)
Duty CycleNon-continuous97.8% (45 minutes on, 1 minuteoff)Non-continuous97% (58.2 minutes on, 1.8minutes off)Non-continuous97% (58.2 minutes on, 1.8minutes off)
Applied PartClassificationType CFType BFSame as Predicate (K122756)
IEC 60601-1andIEC 60601-1-2ComplianceYes:IEC 60601-1 andIEC 60601-1-2Yes:IEC 60601-1 (Edition 3.1) andIEC 60601-1-2 (Edition 4.0)Yes:IEC 60601-1 (Edition 3.2 2020-08)and IEC 60601-1-2 (Edition 4.12020-09)
AttributePredicate Device:Penumbra Pump MAX™(PMX110 - K122756)Reference Device:Riptide™ Aspiration System(MAP-1000 - K172448)Subject Device:Riptide™ Aspiration Pump(LMT-RAP)
ISO 10079-1andISO 10079-4ComplianceYes:EN ISO 10079-1NoDesign changes to meet IEC60601-1 (Edition 3.2 2020-08)compared to the ReferenceDevice required displaying dualscale units for pressure as inHgand kPa and included addition ofa new component:• Vacuum gauge (P/N 140643-01)Yes:ISO 10079-1 (Fourth edition2022-03) and ISO 10079-4 (Firstedition 2021-08)Design changes to meet IP22ingress protection requirement ofISO 10079-1 (Fourth edition2022-03) required addition ofnew components including:• Sealing washer (P/N 12130-0003) for pressure regulator• Silicon seal (P/N 12130-0001)under power button• Silicon seal inside (P/N12130-0002)• O-ring (P/N 12130-0004) toaid in sealing around gauge• SS plate (P/N 05131-0001) tostop plastic deformationfrom point loading at clampand apply even pressure toseal and O-ring• Ring clamp (P/N 08130-0001)• Epoxy (P/N 140518-01)applied to seal handle screws(existing component) toimprove sealing aroundinterfaces
PhysicalDimensionsLength: 15.5"Depth: 11.2"Height: 13.2"Length: 16.1"Depth: 13.2"Height: 12.3"Length: 16.1"Depth: 13.2"Height: 12.3"
Weight22.3 lbs (10.1 kg)Approximately 24 lbs (11 kg)Approximately 24 lbs (11 kg)
Noise<60 dBa60.5 dBa<70 dBa
InternalComponentsCommonly used electricalcomponents, fittings, and tubing.Same as Predicate (K122756)Same as Predicate (K122756)
SterilizationNon-sterileSame as Predicate (K122756)Same as Predicate (K122756)
Operating Life500 hoursSame as Predicate (K122756)Same as Predicate (K122756)
AttributePredicate Device:Penumbra Pump MAX™(PMX110 - K122756)Reference Device:Riptide™ Aspiration System(MAP-1000 - K172448)Subject Device:Riptide™ Aspiration Pump(LMT-RAP)
Collection Canister with Intermediate Tubing Characteristics
Materials of ConstructionCommonly used medical grade plasticsSame as Predicate (K122756)Same as Predicate (K122756)
Volume1000 mL1200 mL1200 mL
Shelf LifeUnknown3 years3 years

{7}------------------------------------------------

{8}------------------------------------------------

Biocompatibility

There is no impact to the biocompatibility of the Riptide™ Aspiration Pump associated with the proposed changes.

Performance Data - Bench

The following non-clinical bench tests were conducted for the Riptide™ Aspiration Pump:

TestTest Method SummaryResults
Riptide™ Aspiration Pump
Electrical Safety(IEC 60601-1)Electrical safety testing conducted per IEC 60601-1Edition 3.2 2020-08.Pass
Electromagnetic Compatibility(IEC 60601-1-2)Electromagnetic compatibility (EMC) testing conductedper IEC 60601-1-2 Edition 4.1 2020-09.Pass
User Manual Design Validation(for general suction use)The subject device was evaluated per FDA guidance"Applying Human Factors and Usability Engineering toMedical Devices, Guidance for Industry and Food andDrug Administration Staff" (February 2016) and per FDArecognized consensus standards ISO 14971 (Third Edition2019-12) and ANSI/AAMI HE75:2009/(R)2018 HumanFactors Engineering - Design of Medical Devices.Pass
Riptide™ Aspiration Pump and Riptide™ Collection Canister with Intermediate Tubing
Medical Suction Equipment(ISO 10079-1/10079-4)Medical suction equipment testing conducted per ISO10079-1 Fourth Edition 2022-03 and ISO 10079-4 FirstEdition 2021-08.Pass

Performance Data – Animal

The determination of substantial equivalence is based upon non-clinical bench testing as there is no change to the intended use, fundamental scientific technology, or materials of construction.

Performance Data - Clinical

The determination of substantial equivalence is based upon non-clinical bench testing as there is no change to the intended use, fundamental scientific technology, or materials of construction.

{9}------------------------------------------------

Conclusion

There is no change to the intended use, fundamental scientific technology, principle of operation, and materials of construction of the subject Riptide™ Aspiration Pump (LMT-RAP) in comparison to the legally marketed predicate Penumbra Pump MAX™ (K122756). The design and labeling changes for Type CF applied part classification, meeting latest electrical safety and medical suction equipment standards (including IEC 60601-1, IEC 60601-1-2, ISO 10079-1, and ISO 10079-4), and updated indications for general use in hospitals or clinics under product code JCX do not raise different questions of safety and effectiveness and have been tested through non-clinical bench testing using established scientific methods. The information provided in this submission supports a determination of substantial equivalence for the subject Riptide™ Aspiration Pump.

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.