Search Results
Found 2 results
510(k) Data Aggregation
(219 days)
The Slinky Catheter is indicated for general intravascular use. including the neuro and peripheral vasculature. It can be used to facilitate introduction of diagnostic agents and therapeutic devices. It is not intended for use in coronary arteries.
The Slinky® Catheter is a single lumen, variable stiffness catheter with a coil and braid reinforcement. It has a radiopaque marker at the distal tip for enhanced fluoroscopic visualization, a lubricious inner, and a hydrophilic coating on the distal 35 cm to enhance lubricity and vascular navigation. The Slinky® Catheter is provided sterile via Ethylene Oxide (EO) sterilization, non-pyrogenic, and is for single use only.
This document describes the premarket notification (510(k)) for the Anoxia Medical, Inc. Slinky Catheter. The primary purpose of the submission is to demonstrate substantial equivalence to a predicate device, the SOFIA PLUS/Distal Access Catheters (K150366). The document does not describe a study involving an AI component or the improvement of human readers with AI assistance. Instead, it focuses on the engineering and biocompatibility performance of the Slinky Catheter.
Since the prompt asks for details related to AI performance, multi-reader multi-case (MRMC) studies, and the use of expert consensus for ground truth in an AI context, and the provided text describes a medical device (a catheter) and its physical and biological performance, I will interpret the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, adapting the categories to fit the provided information on a physical medical device rather than an AI or software product.
Here's an analysis based on the provided text for the Slinky Catheter, focusing on its performance criteria and the studies conducted:
1. A table of acceptance criteria and the reported device performance
The document provides a table of various tests, their methods, and conclusions. The "Conclusions" column effectively serves as the "reported device performance" and implicitly indicates that the acceptance criteria were met, as all conclusions state the device performed as intended, met established criteria, or was comparable to the predicate.
Test | Test Method Summary | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|---|
Bench Performance Testing | |||
Simulated Use | Test articles used by three trained interventional neuroradiologists in a tortuous anatomical bench test model, demonstrating compatibility with ancillary devices. | Device to perform as intended under simulated use conditions. | Device performed as intended under simulated use conditions. |
Particulate Test | Test articles tracked multiple times through tortuous anatomical model with ancillary devices. Inspected for coating anomalies pre- and post-testing. | Particulate generation of subject device to be similar to predicate device. | Particulate generation of the subject device was similar to the predicate device. |
Product Compatibility | Simulated use testing by three physicians using full-length silicone anatomical tortuous path model. Included use with worst-case ancillary devices. | Device to be compatible with ancillary devices. | The device was found to be compatible with ancillary devices. |
Dimensional and Visual Attributes | Physical and dimensional attributes evaluated and measured. | Device to meet established dimensional and visual specifications. | Device met the established dimensional and visual specifications. |
Kink Resistance | Distal, mid, and proximal sections wrapped around varying size pin gauges. | Device to meet established kink resistance criteria. | Device met the established kink resistance criteria. |
Radio Detectability | Distal marker band and catheter body visibility evaluated under fluoroscopy. | Device radiopacity to be comparable to predicate device. | Device radiopacity comparable to the predicate device. |
Catheter Hub | Hub functionality evaluated during simulated use, freedom from leakage, flow rate, dynamic burst, particulate, and other bench testing. | Device hub to meet requirements of ISO 80369-7. | Device hub meets the requirements of ISO 80369-7. |
Durability/Lubricity of Hydrophilic Coating | Coating evaluated after simulated use for defects and friction force. | Device to meet established friction and lubricity criteria; friction force similar to predicate. | Device met the established friction and lubricity criteria. The friction force was similar to the predicate. |
Tip Flexibility | Device tip stiffness characterized on a cantilever bend test. | Tip stiffness to be similar to predicate. | The tip stiffness was similar to the predicate. |
Torque Strength | Devices pre-conditioned by simulated use and torqued to failure inside a tortuous anatomical model with distal tip fixed. | Device torque strength to be same as predicate device. | Device torque strength is the same as the predicate device. |
Force at Break (Distal and Hub) | Catheter force at break averaged 8 lbf for hub and 5-6 lbf for mid and distal segments, exceeding acceptance criteria from withdrawal force measurements. | Tensile strength to meet acceptance criteria determined from withdrawal force measurements. | Tensile strength meets the test acceptance criteria. |
Flow Rate | Characterized at 100 psi with diagnostic agents (e.g., saline, contrast media). | Device to meet specified requirements for delivery of diagnostic agents. | Device meets the specified requirements for delivery of diagnostic agents. |
Static Burst Pressure | Device withstand static pressure above labeled 100 psi maximum with margin of safety. | Device integrity to be suitable for intended clinical use and meet requirements of ISO 10555-1. | Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1. |
Freedom from Leakage | Testing followed methods and criteria defined in EN ISO 10555-1 after pre-conditioning by simulated use. | Device integrity to be suitable for intended clinical use and meet requirements of ISO 10555-1. | Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1. |
Air Leakage | No air leakage at hub into syringe for 15 seconds. | Device integrity to be suitable for intended clinical use and meet requirements of ISO 10555-1. | Device integrity is suitable for intended clinical use and met requirements of ISO 10555-1. |
Dynamic Burst | Devices pre-conditioned by simulated use. Test devices connected to pressurized fluid source, starting with 250 psi and increased every 10 seconds up to 600 psi. | Device to meet labeled maximum infusion pressure of 100 psi. | Device met labeled maximum infusion pressure of 100 psi. |
Peak Tensile Force | Devices pre-conditioned by simulated use. Tensile testing conducted per methods defined in EN ISO 10555-1. | Device to meet acceptance criteria determined from withdrawal force testing. | Device met acceptance criteria determined from withdrawal force testing. |
Biocompatibility Testing | |||
Cytotoxicity (ISO 10993-5) - MEM Elution Assay | Cell culture observed for cytotoxic reactivity following exposure to test extract with Grade , ISO 10993-11) | Study animals observed for individual temperature rise following intravenous injection of test extracts. | Non-pyrogenic. |
Hemolysis (ASTM F756, ISO10993-4) – Direct and Indirect (extract) Hemolysis | Difference between hemolytic indexes of subject device and negative control evaluated after exposure to test extracts and direct exposure to test article. | Non-hemolytic. | Direct contact - Non-hemolytic; Indirect contact - Non-hemolytic |
SC5b-9 Complement Activation Assay (ISO 10993-4) | Using an enzyme immunoassay, SC5b-9 concentration following exposure of test article extracts to normal human serum (NHS) compared to SC5b-9 of control article exposed to NHS. | Not considered to be a potential activator of the complement system. | Not considered to be a potential activator of the complement system |
Thrombogenicity (ASTM F2382) - Partial Thromboplastin Time | Thrombogenicity potential of test article compared to control articles. | Acceptable results. | Acceptable results |
Platelet and Leukocyte Count (ASTM F2888-19, ISO 10993-4) In vitro Blood Loop | (Test Method Summary Blank) | (Implied acceptable performance for blood compatibility) | (Conclusions blank, but generally implied to be acceptable based on overall summary) |
Sterilization and Shelf-Life | |||
Sterilization | Validated process to achieve a minimum sterility assurance level (SAL) of 10^-6. | SAL of 10^-6. | Achieved minimum sterility assurance level (SAL) of 10^-6. |
Shelf-Life | Studies conducted to demonstrate product and packaging remain functional and sterile. | Product and packaging remain functional and sterile for 12 months. | Demonstrates functionality and sterility for 12 months. |
2. Sample size used for the test set and the data provenance
- Sample Size for Bench Tests:
- Simulated Use, Product Compatibility, Torque Strength, Dynamic Burst, Peak Tensile Force, Freedom from Leakage, Durability/Lubricity: Not explicitly stated as a number, but "Test articles" and "Devices" implies multiple units were used for each test.
- Simulated Use, Product Compatibility: "Test articles were used by three trained interventional neuroradiologists."
- Sample Size for Biocompatibility Tests: Not explicitly stated as numbers (e.g., how many guinea pigs, how many cell cultures). The phrase "Study animals" is used.
- Data Provenance: The document does not specify the country of origin for the data. The tests performed are standard bench tests and regulated biocompatibility studies, following ISO and ASTM standards. The studies were conducted in accordance with 21 CFR Part 58 Good Laboratory Practices, which are US regulations. It's a prospective test in the sense that these tests are performed on the manufactured device models to demonstrate their performance characteristics for regulatory submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- For Simulated Use and Product Compatibility bench testing: three trained interventional neuroradiologists were used. Their specific years of experience are not mentioned, but they are identified as "trained interventional neuroradiologists."
- For other tests, "ground truth" is established by adherence to specified test methods (e.g., ISO, ASTM standards) and their quantitative or qualitative pass/fail criteria, rather than expert consensus on, for example, image interpretation.
4. Adjudication method for the test set
- The document does not describe an adjudication method in the context of multiple expert readers or interpretations. For the "Simulated Use" and "Product Compatibility" tests where three neuroradiologists participated, it's not stated how disagreements, if any, were resolved; however, the conclusion states the device "performed as intended" and "was found to be compatible," suggesting a consensus or successful outcome observed by all. This is not a formal adjudication process as would be seen in an imaging study.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. This submission is for a physical medical device (a catheter), not an AI or software device. Therefore, the concept of human readers improving with AI assistance is not applicable to this document. The "three trained interventional neuroradiologists" participated in simulated use bench testing to evaluate the physical interaction and performance of the catheter in a model, not to read images or interpret data generated by the device in terms of AI output.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- This question is not applicable as the device is a physical catheter, not an algorithm or AI product.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" for the catheter's performance is established through bench testing against established engineering and biocompatibility standards (ISO, ASTM) and comparison to the predicate device's characteristics.
- For dimensional and physical properties: direct measurement and adherence to specifications.
- For functional properties (e.g., kink resistance, torque strength, flow rate): performance against defined quantitative criteria or qualitative observations (e.g., performs as intended).
- For biocompatibility: results from standardized biological assays performed on animal models, cell cultures, or blood samples, with pass/fail criteria defined by the ISO 10993 series and other relevant standards.
- For simulated use: observation by trained physicians against expected performance.
- The overall "ground truth" for regulatory clearance is substantial equivalence to the predicate device, demonstrated by meeting comparable performance characteristics through these tests.
8. The sample size for the training set
- This question is not applicable. This document describes performance testing for a physical medical device, not the development or validation of an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
- This question is not applicable for the same reason as above.
Ask a specific question about this device
(251 days)
The CEREGLIDE™ 71 Intermediate Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. The CEREGLIDE™ 71 Intermediate Catheter is also indicated for use as a conduit for retrieval devices.
The CEREGLIDE™ 71 Intermediate Catheter is a variable stiffness, single lumen catheter designed to be introduced over a steerable guide wire or microcatheter into the neuro vasculature. The catheter shaft is composed of a stainless steel variable pitch braid with a PTFE inner liner to facilitate movement of guide wires and other devices. The exterior of the catheter shaft is covered with polymer materials, which encapsulate the stainless steel braid construction. The catheter has a stiff proximal shaft which transitions into the flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The distal end of the catheter has a radiopaque marker band to facilitate fluoroscopic visualization and has a hydrophilic coating to provide lubricity to reduce friction during navigation in the vasculature. The proximal end of the catheter has a luer fitting located on the end of the catheter hub. An ID band is placed at the distal end of the hub over a strain relief. The catheter is packaged with a Tuohy Borst rotating hemostasis valve (RHV) with a side port and two slit introducers as accessories. The RHV with side port is used for flushing and insertion of catheters. The slit introducers are designed to introduce the catheter into the base catheter and protect the distal tip of the catheter during insertion into the RHV of the base catheter.
The CEREGLIDE™ 71 Intermediate Catheter is a medical device for facilitating neurovascular interventions. The document provided outlines the non-clinical performance data to demonstrate its substantial equivalence to a legally marketed predicate device, the CERENOVUS Large Bore Catheter (K191237).
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the CEREGLIDE™ 71 Intermediate Catheter are implicitly defined by a series of bench tests and biocompatibility tests, with the reported performance consistently being "PASS" or "Samples met the established acceptance criteria," sometimes exceeding comparator devices.
Test Category | Specific Test | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|---|
Design Verification | Visual Inspection | Catheter meets visual requirements described in ISO 10555-1 Section 4.4. | PASS: Samples met the established acceptance criteria |
Catheter ID | Catheter internal diameters meet requirements. | PASS: Samples met the established acceptance criteria | |
Catheter OD | Catheter outer diameters meet requirements. | PASS: Samples met the established acceptance criteria | |
Catheter Working Length | Working length of the catheter as defined in ISO 10555-1 Section 3.6. | PASS: Samples met the established acceptance criteria | |
Catheter Tip Length | Catheter tip length meets requirements. | PASS: Samples met the established acceptance criteria | |
System Air Leakage | No air leak into the hub subassembly. | PASS: Samples met the established acceptance criteria | |
System Liquid Leakage | Catheter joint strength meets the freedom from leakage (liquid during pressurization) requirements of ISO 10555-1:2013, section 4.7. | PASS: Samples met the established acceptance criteria | |
Delamination of PTFE Liner | PTFE appropriately adhered to the inner lumen of the Catheter with braid reinforcement. | PASS: Samples met the established acceptance criteria | |
Kink (Distal & Proximal) | Catheter remains stable and does not kink during use. | PASS: Samples met the established acceptance criteria | |
Tip Movement | Catheter meets the tip column stiffness requirement. | PASS: Samples met the established acceptance criteria | |
Tip Linear Stiffness | Tip flexibility of the Catheter is acceptable, relative to other devices of similar design. | PASS: Samples met the established acceptance criteria | |
Coating Lubricity and Durability | Lubriciousness and durability of the Catheter's hydrophilic coating are verified. | PASS: Samples met the established acceptance criteria | |
Coating Length | Catheter hydrophilic coating length meets the design requirements. | PASS: Samples met the established acceptance criteria | |
Peak Tensile Strength | Catheter joint strength meets the requirements of Section 4.5 of ISO 10555-1. | PASS: Samples met the established acceptance criteria | |
Introducer ID | Introducer internal diameters meet the requirements. | PASS: Samples met the established acceptance criteria | |
Particulate Count | Coating integrity of the Catheter's outer surface meets the requirements for content of Particle Matter in alignment with USP counting methods and compared to the reference predicate device. | PASS: Samples met the established acceptance criteria | |
Burst Pressure | Catheter can withstand the maximum hydrostatic pressure. | PASS: Samples met the established acceptance criteria | |
Introducer Working Length | Working length of the introducer is confirmed. | PASS: Samples met the established acceptance criteria | |
Introducer Separation Force | Force required to separate the introducer is confirmed. | PASS: Samples met the established acceptance criteria | |
Torque Test | Number of revolutions to failure of the Catheter in simulated anatomy meets or exceeds established criteria. | PASS: Samples exceeded comparator devices in revolutions to failure | |
Design Validation | In Vitro Usability Studies | Usability parameters such as trackability, tip stability, durability, and (ancillary) device compatibility with tracking of the Catheter to target sites and delivery of a stent-retriever in the neurovascular model are acceptable. | PASS: Samples met the established acceptance criteria |
Biocompatibility | Chemical Characterization | Determination of extractable species from the test article in purified water, isopropyl alcohol, and hexane is acceptable. | PASS |
Cytotoxicity | Lack of potential cytotoxic effects using monolayers of L-929 mouse fibroblast cells when evaluated with 1X MEM extract. | PASS | |
Sensitization – Guinea Pig Maximization | No allergenic potential or sensitizing capacity. | PASS | |
Irritation - Rabbit Intracutaneous Reactivity | No chemicals that may leach or be extracted capable of causing local irritation in the dermal tissues of rabbits. | PASS | |
Acute Systemic Toxicity in Mice | No potential toxic effects as a result of a single-dose systemic injection in mice. | PASS | |
Pyrogenicity – Materials Mediated Rabbit Pyrogen | No febrile response in rabbits caused by a saline extract of the test article. | PASS | |
ASTM Hemolysis Study: Direct and Extraction Methods | Acceptable hemolytic potential of the test articles and test article extract according to ASTM method (F756-17). | PASS | |
SC5b-9 Complement Activation Assay | Test article is not capable of generating the activation fragment, SC5b-9, which contributes to inflammatory immune response in humans, as measured in Normal Human Serum (NHS). | PASS | |
ASTM Heparinized Platelet and Leukocyte Count | Medical materials exposed to human whole blood do not adversely affect the platelet and leukocyte ratios in whole blood. | PASS | |
ASTM Partial Thromboplastin Time (PTT) | No coagulation abnormalities in the intrinsic coagulation pathway that would indicate an adverse effect. | PASS | |
In Vivo Thromboresistance | Acceptable thrombogenic potential of a blood contacting medical device in comparison to a predicate device. | PASS | |
Shelf-Life | Stability Testing | Device maintains its performance and safety for 1 year when subjected to standard transportation conditioning and aging. | PASS: Results of testing on the subject device all met established acceptance criteria. |
2. Sample Sizes Used for the Test Set and Data Provenance
The document states: "All testing was conducted using sampling methods as required by internal procedure." Specific sample sizes for each test are not explicitly provided.
The data provenance is non-clinical bench testing and in-vitro testing. There is no mention of human clinical data or patient data. Therefore, there is no country of origin or retrospective/prospective distinction for patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
Not applicable. This device's evaluation relies on objective engineering and material science tests (bench and in-vitro, and biocompatibility tests), not expert interpretation of diagnostic images or clinical outcomes. Therefore, there is no "ground truth" established by experts in this context.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods (like 2+1, 3+1) are typically used for establishing ground truth in clinical or image-based studies where expert consensus is required for ambiguous cases. As the studies described are bench and biocompatibility tests, human adjudication is not relevant.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted as this is a device clearance based on substantial equivalence using non-clinical performance data, not an AI or diagnostic imaging device requiring human reader interaction.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This pertains to an intermediate catheter, which is a physical medical device, not a software algorithm or AI.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The "ground truth" for the non-clinical tests is established by pre-defined engineering specifications, material properties, and regulatory standards (e.g., ISO standards, ASTM methods, USP guidance). The device's performance is compared against these objective criteria. For the biocompatibility tests, the "ground truth" is typically the absence of adverse biological reactions as defined by the specific test (e.g., no cytotoxicity, no sensitization, no irritation).
8. The Sample Size for the Training Set
Not applicable. There is no "training set" as this is not a machine learning or AI-based device.
9. How the Ground Truth for the Training Set Was Established**
Not applicable. As there is no training set, there is no ground truth establishment for it.
Ask a specific question about this device
Page 1 of 1