K Number
K181959

Validate with FDA (Live)

Date Cleared
2019-03-22

(242 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HALYARD SMART-FOLD* Sterlization Wap (H450, H650) is intended to enclose and is to be sterlized by a healthare provider using Steris V-PRO Low Temperature Sterilization system. The wrap was validated in the following pre-programmed cycles:

  • STERIS V-PRO 1 low temperature sterilization system (lumen cycle),
  • STERIS V-PRO 1 Plus low temperature sterilization system (Lumen and Non-lumen cycle),
  • STERIS V-PRO maX low temperature sterilization system (Lumen cycle, flexible cycle), .
  • STERIS V-PRO 60 sterilizer (Lumen, Non-Lumen cycle, flexible cycle)
  • STERIS V-PRO maX 2 low temperature sterilization system (Lumen cycle, flexible cycle)
    HALYARD SMART-FOLD* Sterlization Wrap (H450, H650) is intended to allow terminal sterlization of medical devices using vaporized hydrogen peroxide (VHP) and to maintain sterilly of the enclosed within the period of time for which performance data demonstrating maintenance of sterility has been provided.
Device Description

HALYARD SMART-FOLD* Sterlization Wrap is supplied in bulk to the customer as pre-shaped sterilization wap which is then used to wrap a medical device or a collection of medical devices for sterilization Wrap is compised of two preshaped sheets of HAL YARD* Sequential Sterilization Wrap. Each sheet is composed of a three-laver SMS (spunbond-meltblownspunbond) polypropylene fabric treated with an antistatic treatment.
The SMART-FOLD* Sterilization Wrap features reinforcement reference line, a white inner layer, sidetabs with closure strips and pull-tabs which allow for aseptic presentation of the sterilized medical device. The white same material composition but contains no blue pigment.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device (HALYARD SMART-FOLD* Sterilization Wrap), and it describes the device's acceptance criteria and the studies conducted to prove it meets those criteria. However, it is not an AI/ML device, so many of the requested fields related to AI/ML model validation (e.g., sample size for training set, number of experts for ground truth, MRMC studies) are not applicable and therefore not present in the document.

Here's a breakdown of the information that is available, formatted to address your specific questions where possible.


Device Name: HALYARD SMART-FOLD* Sterilization Wrap (Models: H450, H650)
Device Type: Sterilization Wrap (not an AI/ML device)
Intended Use: To enclose medical devices for sterilization using Steris V-PRO Low Temperature Sterilization systems and maintain sterility until used.


1. Table of Acceptance Criteria and Reported Device Performance

The document describes several performance tests and their results, which essentially serve as the acceptance criteria and proof of performance for this non-AI medical device. The "acceptance criteria" are implied by the "passing results" of these tests.

Acceptance Criteria (Implied by Test Objective)Reported Device Performance (Results)
Sterilant Penetration/Efficacy: Demonstrate effective sterilant penetration into instrument trays wrapped with the device, under half-cycle and worst-case lumen conditions, achieving a 10⁻⁶ SAL.The Halyard SMART-FOLD* (H450, H650) Sterilization wrap demonstrated effective sterilant penetration into instrument trays wrapped with Halyard Health SMART-FOLD* (H450, H650) Sterilization Wrap after processing in the V-PRO maX low temperature sterilizer under half cycle condition in the V-PRO maX low temperature sterilizer, worst-case lumen cycle. (And achieved a 10⁻⁶ SAL following processing in a worst-case half-cycle, as noted in the Technological Characteristics Comparison section for both proposed and predicate devices).
Maintenance of 365-Day Package Integrity: Demonstrate that the device is an effective barrier for maintaining sterility of various sterilization loads for 30, 180, and 365 days after processing in the worst-case V-PRO 60 Low Temperature Sterilizer Lumen cycle, followed by an event-related shelf life study.The Halyard SMART-FOLD* (H450, H650) Sterilization wrap demonstrated effective barrier using the simulated test pack for the best and worst-case weight wraps for maintaining the sterility of their contents for 30, 180, 365 days after processing in the worst-case V-PRO 60 Low Temperature Sterilizer Lumen cycle and followed by an event-related shelf life study. (Also explicitly stated 365 days passing results for all V-PRO models in Technological Characteristics Comparison section).
Material Biocompatibility: Demonstrate non-irritating and non-cytotoxic properties in accordance with ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).The Halyard SMART-FOLD* (H450, H650) Sterilization wrap was found to be non-cytotoxic and non-irritant based on the conditions of these tests.
Physical Properties: Demonstrate passing results for various physical properties (bursting, air permeability, abrasion resistance, mass per unit area, flame resistance, water resistance, linting/cleanliness, tearing strength, breaking strength, seam failure, bacterial filtration efficiency).The physical properties testing met the acceptance criteria and demonstrated passing results for the Halyard SMART-FOLD* (H450, H650) Sterilization wrap.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes (e.g., number of sterilization wraps or test packages) used for each test. For example, for "Maintenance of 365-Day Package Integrity," it mentions "simulated test pack for the best and worst-case weight wraps" but not the quantity of these packs.

Data Provenance: The studies were conducted by the manufacturer, Owens & Minor Halyard, Inc., based in Alpharetta, GA (USA). The document does not specify if the testing was performed retrospectively or prospectively, but standard medical device validation studies are typically prospective.


3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not Applicable. This is not an AI/ML device that requires human expert annotation for ground truth. The acceptance criteria are based on physical and biological laboratory tests.


4. Adjudication Method for the Test Set

Not Applicable. As this is not an AI/ML device, there are no "interpretations" that would require adjudication.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

Not Applicable. This is not an AI/ML device, and no human reader effectiveness is being evaluated.


6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

Not Applicable. This is a physical medical device, not an algorithm.


7. The Type of Ground Truth Used

The "ground truth" for this device's performance is established by:

  • Laboratory Testing Standards: Adherence to established scientific and regulatory standards (e.g., ISO 10993 for biocompatibility, various ASTM and ISO standards for physical properties, and internal validation protocols for sterilization efficacy).
  • Biological Indicators: For sterilization efficacy tests, the "ground truth" of sterility is typically determined by the absence of microbial growth in biological indicators after processing.
  • Physical Measurements: For physical properties, the ground truth is direct measurement against pre-defined specifications.

8. The Sample Size for the Training Set

Not Applicable. This is not an AI/ML device and does not have a "training set."


9. How the Ground Truth for the Training Set Was Established

Not Applicable. As there is no training set for an AI/ML model, this question is irrelevant to the provided document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 22, 2019

Owens & Minor Halyard, Inc Angela L. Bunn, RAC Director, Regulatory Affairs, Products Division 5405 Windward Parkway Alpharetta, GA 30004

Re: K181959

Trade/Device Name: HALYARD SMART-FOLD* Sterilization Wrap (Models: H450, H650) Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: February 19, 2019 Received: February 21, 2019

Dear Angela L. Bunn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Clarence W. Murray lii III -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMANSERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K181959 Device Name

HALYARD SMART-FOLD* Sterilization Wrap (Models: H450, H650)

Indications for Use (Describe)

HALYARD SMART-FOLD* Sterlization Wap (H450, H650) is intended to enclose and is to be sterlized by a healthare provider using Steris V-PRO Low Temperature Sterilization system. The wrap was validated in the following pre-programmed cycles:

  • STERIS V-PRO 1 low temperature sterilization system (lumen cycle), ●
  • STERIS V-PRO 1 Plus low temperature sterilization system (Lumen and Non-lumen cycle), ●
  • STERIS V-PRO maX low temperature sterilization system (Lumen cycle, flexible cycle), .
  • STERIS V-PRO 60 sterilizer (Lumen, Non-Lumen cycle, flexible cycle) ●
  • STERIS V-PRO maX 2 low temperature sterilization system (Lumen cycle, flexible cycle) ●

HALYARD SMART-FOLD* Sterlization Wrap (H450, H650) is intended to allow terminal sterlization of medical devices using vaporized hydrogen peroxide (VHP) and to maintain sterilly of the enclosed within the period of time for which performance data demonstrating maintenance of sterility has been provided.

Table 1 Summary of V-PRO Low Temperature Sterilizer System Cycles
SterilizationCycleSterilizer Model
V-PRO 1V-PRO 1 PlusV-PRO maXV-PRO 60V-PRO maX 2
LumenYesYesYesYesYes
Non-LumenN/AYesYesYesYes
FlexibleN/AN/AYesYesYes

Test results validated that HALYARD SMART-FOLD $160, H650 allowed sterlization of the enclosed devices using Steris V-PRO Love Temperature Sterlization system.All models of the HALYARD Sterilization Wrap (H60, H550) have been valided for use with the STERIS V-PRO Lov Temperature Sterilization System cycles (Table 2, Table 3, and Table 4).

{3}------------------------------------------------

TABLE 2: Validated Low Temperature Sterilization Products STERIS V-PRO 1, STERIS V-PRO 1 Plus, STERIS V-PRO maX, STERIS V-PRO maX 2,and STERIS V-PRO 60, Lumen¹ Cycles.
Note: Refer to the Manufacturer's User's Guide for complete instructions on load and cycle for each STERIS V-PRO System. The instructionsprovided below are not intended to replace the detailed Instructions for Use provided with the STERIS – VPRO
STERIS V-PRO SystemLumen CycleIntended Load
STERIS V-PRO 1STERIS V-PRO 1 PlusSTERIS V-PRO maXSTERIS V-PRO maX 2STERIS V-PRO maX,Lumened and non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations: Single channeled devices with a stainless lumen that is $\ge$ 0.77mm internal diameter (ID) and $\le$ 500mm in length Dual channeled devices with stainless steel lumens that are $\ge$ 0.77mm ID and $\le$ 527mm in length Triple channeled devices with stainless steel lumens that are $\ge$ 1.2 mm ID and $\le$ 275 mm in length $\ge$ 1.8 mm ID and $\le$ 310 mm in length or $\ge$ 2.8 mm ID and $\le$ 317 mm in length STERIS V-PRO maX 2 Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations: Single channeled devices with a stainless lumen that is $\ge$ 0.77mm internal diameter (ID) and $\le$ 500mm in length Dual channeled devices with stainless steel lumens that are $\ge$ 0.77mm ID and $\le$ 527mm in length Triple channeled devices with stainless steel lumens that are $\ge$ 1.2 mm ID and $\le$ 275 mm in length $\ge$ 1.8 mm ID and $\le$ 310 mm in length or $\ge$ 2.8 mm ID and $\le$ 317 mm in length STERIS V-PRO 1 STERIS V-PRO 1 Plus The V-PRO 1 Cycle and the V-PRO 1 Plus Lumen Cycle can sterilize* instruments/devices with the following features: Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Medical devices with a single lumen with: an inside diameter of 3 mm or larger and a length of 400 mm or shorter an inside diameter of 2 mm or larger and a length of 250 mm or shorter an inside diameter of 1 mm or larger and a length of 125 mm or shorter
STERIS V-PRO 60The V-PRO 60 sterilizer's Lumen cycle can sterilize: Instruments with diffusion- restricted spaces such as the hinged portion of forceps and scissors. Non- Lumened devices including non- lumened rigid and semi-rigid endoscopes Medical Devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations: $\Upsilon$ Single or dual lumen devices with stainless steel lumens that are $\ge$ 0.77 mm (~1/32") internal diameter (ID) and $\le$ 410 mm (16-9/64") in length $\Upsilon$ Triple lumen devices with stainless steel lumens that are $\ge$ 1.2 mm (~3/64") ID and $\le$ 275 mm (~10 – 55/64") in length $\ge$ 1.8 mm (~5/64") ID and $\le$ 310 mm (~12 – 13/64") in length or $\ge$ 2.8 mm (~7/64") ID and $\le$ 317 mm (~12 – 31/64") in length
TABLE 3: Validated Low Temperature Sterilization Products STERIS V-PRO 1 Plus, STERIS V-PRO maX, STERIS V-PRO maX 2, andSTERIS V-PRO 60, Non-Lumen CyclesNote: Refer to the User's Guide for complete instructions on load and cycle for each STERIS V-PRO System. The instructions provided below are notintended to replace the detailed Instructions for Use provided with the STERIS – VPRO
STERIS V-PRO System Non-LumenCycleIntended Load
STERIS V-PRO 1 PlusSTERIS V-PRO maXSTERIS V-PRO maX 2STERIS V-PRO maX 2Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.STERIS V-PRO maXThe Non-Lumen Cycle can sterilize instruments/devices with the following features:Non-lumened instruments including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors. STERIS V-PRO 1 PlusThe V-PRO 1 Plus Non-Lumen Cycle can sterilize instruments/devices with the following features:Non-Lumen instruments including non-lumened instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors.
STERIS V-PRO 60The V-PRO 60 Non-Lumen Cycle can sterilize instruments/devices with the following features:Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with stainless steel or titanium diffusion-restricted spaces such as the hinged portion of forceps and scissors.
TABLE 4: Validated Low Temperature Sterilization Products STERIS V-PRO maX. STERIS V-PRO maX 2, and STERIS V-PRO 60, Flexible CyclesNote: Refer to the User's Guide for complete instructions on load and cycle for each STERIS V-PRO System. The instructions provided below are notintended to replace the detailed Instructions for Use provided with the STERIS – VPRO
STERIS V-PRO System FlexibleCycleIntended Load
STERIS V-PRO maXSTERIS V-PRO maX 2STERIS V-PRO maX 2The Flexible Cycle can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and SurgicalCare) or bronchoscopes in either of two load configurations:1. Two flexible endoscopes with a light cord (if not integral to the endoscope) and mat with no additional load. The flexible endoscopes maycontain either:• A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length• Or two lumens with:- One lumen that is ≥ 1 mm ID and ≤ 998 mm in length- And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments. The flexibleendoscopes may contain either:• A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length• Or two lumens with:- One lumen that is ≥ 1 mm ID and ≤ 990 mm in length- And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in lengthSTERIS V-PRO maXThe Flexible Cycle can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) or bronchoscopesin either of two load configurations:1. Two flexible endoscopes with a light cord (if not integral to the endoscope) and mat with no additional load3. The flexible endoscopes my containeither:• A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length• Or two lumens with:- One lumen that is ≥ 1 mm ID and ≤ 998 mm in length- And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments including instruments withdiffusion-restricted areas such as the hinged portion of forceps or scissors4. The flexible endoscopes my contain either:• A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length• Or two lumens with:- One lumen that is ≥ 1 mm ID and ≤ 998 mm in length- And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length
STERIS V-PRO 60The V-PRO 60 sterilizer's 60 flexible cycle can sterilize:One flexible surgical endoscope or bronchoscope with a light cord (if not integral to endoscope) and mat without any additional load. The flexible endoscopemay be a:• Single or dual lumen device with lumens that are $≥ 1 mm (~3/64") ID and ≤990 mm (38-63/64")$ in length

{4}------------------------------------------------

{5}------------------------------------------------

{6}------------------------------------------------

HALYARD SMART-FOLD* Sterlization Wraps (H450, H650) Recommendations for Use with STERS V-PRO Lov Temperature Sterlization (Table 5).

TABLE 5: Halyard Smart-Fold* Sterilization Wrap Recommendations2 for Use with Steris V-PRO Sterilizer
SMART-FOLD*SterilizationWrap ModelIntendedLoads3Maximum Wrapped Package Content Weights 4
STERISV-PRO® 1,V-PRO® 1PlusV-PRO®maXV-PRO®maX 2LumenSTERISV-PRO® 1Plus V-PRO®maXV-PRO®maX 2Non-LumenSTERISV-PRO®maXV-PRO®maX 2FlexibleSTERISV-PRO®60 LumenSTERISV-PRO®60 Non-LumenSTERISV-PRO® 60 FlexibleSTERISV-PRO®maX 2Fast Non-Lumen
H450Moderate toHeavyweightPackage (forexample:general usemedicalinstruments)13 lbs13 lbs.13 lbs.11 lbs.12 lbs.The validation studies were conducted with one flexibleendoscope packaged into a tray with silicone wrap, andinstrument organizers and light cord ( if not integral toscope) and no additional load.N/A
H650Moderate toHeavyweightPackage (forexample:general usemedicalinstruments)19.65 lbs.25 lbs.-max2,19.65 - max,1 and 1 plus24 lbs.11 lbs.12 lbs.The validation studies were conducted with one flexibleendoscope packaged into a tray with silicone wrap, andinstrument organizers and light cord (if not integral toscope) and no additional load.N/A
2Individual results may differ due to factors such as variations in handling practices, wrapping techniques, and folding methods. Results may also differ due to the use of irregularly shapedcontents, which may put added stress on the wrap. Each healthcare facility should determine for itself which wrap model is most appropriate for each intended use.
3Intended loads include: Medical Instruments with and without lumens that include telescopes, endoscopes, cameras, light cords and general use medical instruments.
4It is recommended to not exceed the maximum wrapped package content weights indicated for each wrap model. Furthermore, it is recommended to not exceed the number, weight, andsize of individual content types that were validated for HALYARD SMART-FOLD* Sterilization Wraps.
5The Fast-Non-Lumen cycle is intended to sterilize pouched instruments trays only and is therefore not intended to be used for sterilization of HALYARD SMART-FOLD* Sterilizationwraps.

*Registered Trademark or Trademark of O&M Halyard or its affiliates. ©2018. All rights reserved.

{7}------------------------------------------------

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *

The burden time for this collection of information is a response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{8}------------------------------------------------

510(K) SUMMARY HALYARD SMART-FOLD* Sterilization Wrap (H450, H650)

Applicant's Name, Address, Telephone, FAX, Contact Person

Owens & Minor Halyard, Inc 5405 Windward Parkway Alpharetta, GA 30004

Contact Name: Angela L. Bunn, RAC

Director, Regulatory Affairs Products Division Tel: 470.448.5856 Fax: 470.200.3177 Email: angela.bunn@hyh.com

510(k) Number: K181959

Establishment Registration Number: 3014421917

Date: March 22, 2019

CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME

Classification: Class II per 21 CFR 880.6850 Classification Name: Sterilization Wrap Common/Usual Name: Sterilization Wrap Product Code: FRG Proprietary Name: HALYARD SMART-FOLD* Sterilization Wrap (Models: H450, H650)

PREDICATE DEVICE:

K142782- KIMGUARD ONE-STEP* Sterilization Wrap (Models: H100, H200, H300, H400, H500, H600)

REFERENCE DEVICE:

K140963 – KIMGUARD* SMART-FOLD* sterilization wrap (Models: KC450, KC650)

{9}------------------------------------------------

INDICATIONS FOR USE

HALYARD SMART-FOLD* Sterilization Wrap (H450, H650) is intended to enclose that is to be sterlized by a healthcare provider using Steris V-PRO Low Temperature Sterilization system. The following pre-programmed cycles:

  • STERIS V-PRO 1 low temperature sterilization system (lumen cycle), .
  • STERIS V-PRO 1 Plus low temperature sterilization system (Lumen and Non-lumen cycle), ●
  • STERIS V-PRO maX low temperature sterilization system (Lumen cycle, flexible cycle), ●
  • STERIS V-PRO 60 low temperature sterilization system (Lumen cycle, flexible cycle) ●
  • STERIS V-PRO maX 2 low temperature sterilization system (Lumen cycle, flexible cycle) ●

HALYARD SMART-FOLD* Sterilization Wrap (H450, H650) is intended to allow terminal sterilization of medical devices using vaporized hydrogen peroxide (VHP) and to maintain sterility of the enclosed device(s) until used within the period of time for which performance data demonstrating maintenance of sterility has been provided.

Table 1 Summary of V-PRO Low Temperature Sterilizer System Cycles
Sterilization CycleSterilizer Model
V-PRO 1V-PRO 1 PlusV-PRO maXV-PRO 60V-PRO maX 2
LumenYesYesYesYesYes
Non-LumenN/AYesYesYesYes
FlexibleN/AN/AYesYesYes

Test results validated that HALYARD SMART-FOLD* Sterilization Wrap (H450) allowed sterilization of the enclosed devices using Steris V-PRO Low Temperature Sterilization system. All models of the HAL YARD Smart Fold Sterilization Wrap (H450, H650) have been validated for use with the Steris V-PRO Low Temperature Sterilization System cycles (Table 1.2, Table 1.3).

{10}------------------------------------------------

TABLE 2: Validated Low Temperature Sterilization Products STERIS V-PRO 1, STERIS V-PRO 1 Plus, STERIS V-PRO maX, STERIS V-PRO maX 2, andSTERIS V-PRO 60, Lumen1 Cycles.
Note: Refer to the Manufacturer's User's Guide for complete instructions on load and cycle for each STERIS V-PRO System. The instructionsprovided below are not intended to replace the detailed Instructions for Use provided with the STERIS – VPRO
STERIS V-PRO System LumenCycleIntended Load
STERIS V-PRO 1STERIS V-PRO 1 PlusSTERIS V-PRO maXSTERIS V-PRO maX 2STERIS V-PRO maX,Lumened and non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations: Single channeled devices with a stainless lumen that is ≥ 0.77mm internal diameter (ID) and ≤500mm in length Dual channeled devices with stainless steel lumens that are ≥ 0.77mm ID and ≤ 527mm in length Triple channeled devices with stainless steel lumens that are ≥ 1.2 mm ID and ≤ 275 mm in length ≥1.8 mm ID and ≤ 310 mm in length or ≥ 2.8 mm ID and ≤ 317 mm in length STERIS V-PRO maX 2 Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations: Single channeled devices with a stainless lumen that is ≥ 0.77mm internal diameter (ID) and ≤500mm in length Dual channeled devices with stainless steel lumens that are ≥ 0.77mm ID and ≤ 527mm in length Triple channeled devices with stainless steel lumens that are ≥ 1.2 mm ID and ≤ 275 mm in length ≥1.8 mm ID and ≤ 310 mm in length or ≥ 2.8 mm ID and ≤ 317 mm in length STERIS V-PRO 1 STERIS V-PRO 1 Plus The V-PRO 1 Cycle and the V-PRO 1 Plus Lumen Cycle can sterilize* instruments/devices with the following features: Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Medical devices with a single lumen with: an inside diameter of 3 mm or larger and a length of 400 mm or shorter an inside diameter of 2 mm or larger and a length of 250 mm or shorter an inside diameter of 1 mm or larger and a length of 125 mm or shorter
TABLE 2: Validated Low Temperature Sterilization Products STERIS V-PRO 1, STERIS V-PRO 1 Plus, STERIS V-PRO maX, STERIS V-PRO maX 2, andSTERIS V-PRO 60, Lumen¹ Cycles.
Note: Refer to the Manufacturer's User's Guide for complete instructions on load and cycle for each STERIS V-PRO System. The instructionsprovided below are not intended to replace the detailed Instructions for Use provided with the STERIS – VPRO
STERIS V-PRO System Lumen CycleIntended Load
STERIS V-PRO 60The V-PRO 60 sterilizer's Lumen cycle can sterilize :Instruments with diffusion- restricted spaces such as the hinged portion of forceps and scissors. Non- Lumened devices including non- lumened rigid and semi-rigid endoscopes Medical Devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations: Y Single or dual lumen devices with stainless steel lumens that are Ο ≥ 0.77 mm (1/32") internal diameter (ID) and ≤ 410 mm (16-9/64") in length Y Triple lumen devices with stainless steel lumens that are Ο ≥ 1.2 mm ( 3/64") ID and ≤ 275 mm (~ 10-55/64") in length Ο ≥ 1.8 mm (~ 5/64") ID and ≤ 310 mm (~ 12-13/64") in length Ο orΟ ≥ 2.8 mm (~ 7/64") ID and ≤ 317 mm (~12-31/64") in length
1 The Non-Lumen articles can be processed in the Lumen cycle as it takes mixed loads

{11}------------------------------------------------

{12}------------------------------------------------

TABLE 3: Validated Low Temperature Sterilization Products STERIS V-PRO 1 Plus, STERIS V-PRO maX, STERIS V-PRO maX 2, and STERIS V-PRO60, Non-Lumen CyclesNote: Refer to the User's Guide for complete instructions on load and cycle for each STERIS V-PRO System. The instructions provided below are notintended to replace the detailed Instructions for Use provided with the STERIS – VPRO
STERIS V-PRO System Non-LumenCycleIntended Load
STERIS V-PRO 1 PlusSTERIS V-PRO maXSTERIS V-PRO maX 2STERIS V-PRO maX 2Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.STERIS V-PRO maXThe Non-Lumen Cycle can sterilize instruments/devices with the following features:Non-lumened instruments including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors. STERIS V-PRO 1 PlusThe V-PRO 1 Plus Non-Lumen Cycle can sterilize instruments/devices with the following features:Non-Lumen instruments including non-lumened instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors.
STERIS V-PRO 60The V-PRO 60 Non-Lumen Cycle can sterilize instruments/devices with the following features:Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with stainless steel or titanium diffusion restricted spaces such as the hinged portion of forceps and scissors.
TABLE 4: Validated Low Temperature Sterilization Products STERIS V-PRO maX. STERIS V-PRO maX 2, and STERIS V-PRO 60, Flexible Cycles Note:Refer to the User's Guide for complete instructions on load and cycle for each STERIS V-PRO System. The instructions provided below are notintended to replace the detailed Instructions for Use provided with the STERIS – VPRO
STERIS V-PRO System FlexibleCycleIntended Load
STERIS V-PRO maXSTERIS V-PRO maX 2STERIS V-PRO maX 2The Flexible Cycle can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and SurgicalCare) or bronchoscopes in either of two load configurations:1. Two flexible endoscopes with a light cord (if not integral to the endoscope) and mat with no additional load. The flexibleendoscopes may contain either:A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length Or two lumens with: One lumen that is ≥ 1 mm ID and ≤ 998 mm in length And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length 2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments. Theflexible endoscopes may contain either:A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length Or two lumens with: One lumen that is ≥ 1 mm ID and ≤ 990 mm in length And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length STERIS V-PRO maXThe Flexible Cycle can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) orbronchoscopes in either of two load configurations:1. Two flexible endoscopes with a light cord (if not integral to the endoscope) and mat with no additional load3. The flexible endoscopes my contain either:A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length Or two lumens with: One lumen that is ≥ 1 mm ID and ≤ 998 mm in length And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length 2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments includinginstruments with diffusion-restricted areas such as the hinged portion of forceps or scissors4. The flexible endoscopes my contain either:A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length Or two lumens with: One lumen that is ≥ 1 mm ID and ≤ 998 mm in length And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length
STERIS V-PRO 60The V-PRO 60 sterilizer's 60 flexible cycle can sterilize:
One flexible surgical endoscope or bronchoscope with a light cord (if not integral to endoscope) and mat without any additional load. The flexibleendoscope may be a:
• Single or dual lumen device with lumens that are > 1 mm (~3/64") ID and <990 mm (38-63/34") in length

{13}------------------------------------------------

{14}------------------------------------------------

HALYARD Smart-Fold* Sterlization Wrap Recommendations for Use with Steris V-PRO low temperature sterilizer is provided in Table 1.5:

TABLE 5: Halyard Smart-Fold* Sterilization Wrap Recommendations2 for Use with Steris V-PRO Sterilizer
SMART-FOLD*SterilizationWrap ModelIntendedLoads3Maximum Wrapped Package Content Weights 4
STERISV-PRO® 1,V-PRO® 1PlusV-PRO®maXV-PRO®maX 2LumenSTERISV-PRO® 1Plus V-PRO®maXV-PRO®maX 2Non-LumenSTERISV-PRO®maXV-PRO®maX 2FlexibleSTERISV-PRO®60LumenSTERISV-PRO® 60Non-LumenSTERISV-PRO® 60FlexibleSTERISV-PRO®maX 2Fast Non-Lumen5
H450Moderate toHeavyweightPackage (forexample:general usemedicalinstruments)13 lbs13 lbs.13 lbs.11 lbs.12 lbs.The validation studieswere conducted with oneflexible endoscopepackaged into a tray withsilicone wrap, andinstrument organizersand light cord ( if notintegral to scope) and noadditional load.N/A

{15}------------------------------------------------

SMART-FOLD*SterilizationWrap ModelIntendedLoads³STERISV-PRO® 1,V-PRO® 1PlusV-PRO®maXV-PRO®maX 2LumenSTERISV-PRO® 1Plus V-PRO®maXV-PRO®maX 2Non-LumenSTERISV-PRO®maXV-PRO®maX 2FlexibleSTERISV-PRO® 60LumenSTERISV-PRO® 60Non-LumenSTERISV-PRO® 60FlexibleSTERISV-PRO®maX 2Fast Non-Lumen⁵
H650Moderate toHeavyweightPackage (forexample:general usemedicalinstruments)19.65 lbs.25 lbs.-max2,19.65 – max,1 and 1 plus24 lbs.11 lbs.12 lbs.The validation studieswere conducted with oneflexible endoscopepackaged into a tray withsilicone wrap, andinstrument organizersand light cord (if notintegral to scope) and noadditional load.N/A

which may of added stress on the wrop eaching to risell which was model is most appropriate for each intended use.

? Intended bads include: Medical Instruments without uners that include telescopes, cameras. Inth cords and general instruments.

"It is recommented to not exceed the nazimum wapped with wither to each wap model. Futhernor, it is recommended to not size of

individual content types that were validated for HALYARD SMART-FOLD Sterilization Wraps.

"The Fas-Non-Lumen cycle is niended to seriize pour and is thereove not intended to be used for steriization of ALL YAD SMART-FO.D" Steriitzation wags.

DESCRIPTION OF DEVICE

HALYARD SMART-FOLD* Sterlization Wrap is supplied in bulk to the customer as pre-shaped sterilization wap which is then used to wrap a medical device or a collection of medical devices for sterilization Wrap is compised of two preshaped sheets of HAL YARD* Sequential Sterilization Wrap. Each sheet is composed of a three-laver SMS (spunbond-meltblownspunbond) polypropylene fabric treated with an antistatic treatment.

The SMART-FOLD* Sterilization Wrap features reinforcement reference line, a white inner layer, sidetabs with closure strips and pull-tabs which allow for aseptic presentation of the sterilized medical device. The white same material composition but contains no blue pigment.

The HALYARD SMART-FOLD* models and product codes for clearance with Sterilization system (STERIS V-PRO® 1, V-PRO® 1 Plus®, V-PRO® maX, V-PRO® maX 2, STERIS V-PRO® 60) are listed below,

{16}------------------------------------------------

HALYARD 450 SMART-FOLD* Sterilization Wrap
Product CodeSize (in.)Case Ct
1432422"x45"72
1427928"x46"72
1426840"x47"48
1427540"x55"48
1431348"x61"48
HALYARD 650 SMART-FOLD* Sterilization Wrap
1432022"x45"48
1428128"x46"48
1427140"x47"48
1427740"x55"48
1430948"x61"36

{17}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS COMPARISION:

TechnologicalCharacteristicsProposed:HALYARD SMART-FOLD* SterilizationWrapPredicate:KIMGUARD* ONE-STEP*Sterilization Wrap(K142782)Comparison toPredicate
ManufacturerO&M Halyard. IncHalyard Health. IncSimilar
Device ModelsH450, H650KC100, KC200, KC300, KC400, KC500, KC600Similar
Regulation/ProductCodeSterilization Wrap: 880.6850 / FRGSterilization Wrap: 880.6850 / FRGSame
Indications for UseHALYARD SMART-FOLD* Sterilization Wrap (H450, H650) is intended to enclose another medical device that is to besterilized by a healthcare provider using Steris V-PRO®Low Temperature Sterilization system that includes:V-PRO® 1 (Lumen Cycle) V- PRO® 1 Plus (Lumen and Non-Lumen Cycle) V-PRO® maX, V-PRO® max 2 and V-PRO® 60(Lumen, Non-Lumen, and Flexible Cycle) The HALYARD SMART-FOLD* Sterilization Wrap (H450,H650) is intended to allow sterilization of the enclosedmedical device(s) and to maintain sterility of the encloseddevice(s) until used.KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200,KC300, KC400, KC500 and KC600) are intended to encloseanother medical device that is to be sterilized by a healthcareprovider using:V-PRO® 60 Low Temperature Sterilization System that includes:Lumen Cycle Non-Lumen Cycle Flexible Cycle The KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200,KC300, KC400, KC500 and KC600) is intended to allowsterilization of the enclosed medical device(s) and maintain sterilityof the enclosed device(s) until used.Similar
Device DesignTwo pre-shaped sheets of Halyard Sterilization Wrap,which include reinforcement zones, medical deviceplacement reference line, a white inner layer, side-tabs withclosure strips and pull-tabs to allow convenient wrappingwith two sheets simultaneouslyTwo rectangular sheets of KIMGUARD* Sterilization Wrap seemedon edges to allow convenient wrapping with two sheetssimultaneously.Similar
SterilantVaporized Hydrogen PeroxideVaporized Hydrogen PeroxideSame
TechnologicalCharacteristicsProposed:HALYARD SMART-FOLD* SterilizationWrapPredicate:KIMGUARD* ONE-STEP*Sterilization Wrap(K142782)Comparison toPredicate
Maintenance of PackageSterilityPassing 365 days Maintenance of Package Sterilityresults for all models of HALYARD SMART-FOLD*Sterilization wrap (H650, H450).STERIS V-PRO 1 = 365 daysSTERIS V-PRO 1 Plus = 365 daysSTERIS V-PRO maX = 365 daysSTERIS V-PRO 60 = 365 daysSTERIS V-PRO maX 2 = 365 daysPassing 365 days Maintenance of Package Sterility results forall models of KIMGUARD* ONE-STEP* Sterilization Wrap(KC100, KC200, KC300, KC400, KC500 and KC600).STERIS V-PRO 60 = 365 daysSame
Sterilant PenetrationPassing results of sterilant penetration testing verified thesterilant penetrated the wrap effectively reached into thecontents or instrument trays at half cycle condition.Achieved a 10-6 SAL following processing in a worst-casehalf-cycle.Passing results of sterilant penetration testing verified the sterilantpenetrated the wrap effectively reached into the contents orinstrument trays at half cycle condition. Achieved a 10-6 SALfollowing processing in a worst-case half-cycle.Same
Material Biocompatibility10993-5: 2009Biological evaluation of medical devices - Tests for in vitrocytotoxicity. Based on the conditions of this test: The deviceextract was found not to be cytotoxic.10993-10:2010Biological evaluation of medical devices -Tests for irritationand skin sensitization. Based on the conditions of this test:The polar and non-polar device extract was found not to be anirritant to the animal model used in the study.ISO 10993-5: 2009Biological evaluation of medical devices - Tests for in vitrocytotoxicity. Based on the conditions of this test: The device extract was found not to be cytotoxic.10993-10:2010 BiologicalSame. The proposed devicewas additionally tested withevaluation of medical devicesTests for irritation and skinsensitization.
Physical PropertiesThe physical properties testing met the acceptance criteria and demonstrated passing results for the Halyard SMART-FOLD* (H450, H650) Sterilization wrap:Bursting Properties of Fabrics - Part 2 - Pneumatic Method for Determination of Bursting Strength and Bursting Distension, ISO 13828-2:1999 Determination of the permeability of fabrics (Air Permeability), ISO 9237:1995 E Standard Test Method for Abrasion Resistance of Textile Fabrics (Martindale Abrasion Tester Method), ASTM D4966-12 Standard Test Method for Mass Per Unit Area (Weight) of Fabric, ASTM D3776/D3776M-09a Flame resistance requirements for Class I products of > 3.5 seconds flame travel, CPSC 16 CFR 1610:2007 Water Resistance: Hydrostatic Pressure Test, AATCC-127 Textiles - Test methods for nonwovens – Part 10: Lint and other particles generation in the dry state – Linting/Cleanliness, ISO 9073-10 Standard Test Method for Tearing Strength of Nonwoven Fabrics by the Trapezoid Procedure, ASTM D5733-99 Standard Test Method for Breaking Strength and Elongation of Textile Fabrics (Grab Test), ASTM D5034-95 Test Method of Failure in Sewn Seams of Woven Fabrics, ASTM D1683/D1683M-11 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus, ASTMThe physical properties testing met the acceptance criteria and demonstrated passing results for the KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200, KC300, KC400, KC500 and KC600) Sterilization wrap:Bacterial Filtration Efficiency (%) ASTM F2101:07 Flammability CPSC 1610.4 (16 CFR Part 1610) Hydrostatic Resistance (mbar) AATCC127-2008; WSP 80.6:2005 Basis Weight (osy) ASTM D3776-07 Mullen Burst (psi) ASTM D3786-06 Grab Tensile Strength (lbf) ASTM D5034-09 Trapezoid Tear Strength (lbf) ASTM D5733-99 Abrasion Resistance - Martindale ASTM D4966-98 (re-approved 2004) Air Permeability Test (cfm) ISO 9237:1995 E Tensile Strength and Elongation - Seam Test (gf) ASTM D1683-90a Gelbo Lint (particles > 10 microns) WSP 160-1:2005Similar
Technology/ Chemical compositionPolypropylene SMS material that allows sterilant to penetrate and maintain sterile barrierPolypropylene SMS material that allows sterilant to penetrate and maintain sterile barrierSame
MaterialsEach sheet is composed of a three-layer SMS (spunbond-meltblown-spunbond) polypropylene fabric treated with an antistatic treatment.Each sheet is composed of a three-layer SMS (spunbond-meltblown-spunbond) polypropylene fabric treated with an antistatic treatment. [SAME]Same
Single Use DeviceYesYes [SAME]Same

{18}------------------------------------------------

{19}------------------------------------------------

{20}------------------------------------------------

Table 1.7 Technological Characteristics Comparison Table
ComponentSubjectDevice,K181959ReferenceDevice, K140963Comparison to the Reference
Intended UseIntended to allow sterilization of the enclosedmedical device(s) and also maintain sterility of theenclosed device(s) until used.Intended to allow sterilization of the enclosedmedical device(s) and also maintain sterility ofthe enclosed device(s) until used.Same
PackagingBulk and Non-sterileBulk and Non-sterileSame
Base Materials of ConstructionThree-layer SMS (spunbond-meltblown-spunbond)fabricThree-layer SMS (spunbond-meltblown-spunbond) fabric.Same
DesignTwo pre-shaped sheets of KIMGUARD*Sterilization Wrap, which include reinforcementstrips, and seamed on edges to allow convenientwrapping with two sheets simultaneouslyTwo pre-shaped sheets of KIMGUARDSterilization Wrap, which include reinforcementstrips, and seamed on edges to allowconvenient wrapping with two sheetssimultaneously.Same
Technology/ ChemicalcompositionPolypropylene SMS material that allows sterilant topenetrate and maintain sterile barrierPolypropylene SMS material that allowssterilant to penetrate and maintain sterilebarrierSame
UseSingle UseSingle UseSame

Both the subject device and predicate device have similar in use statement, which is to enclose another medical device that is to be sterilized by a heathcare provider using Steris V-PRO Low sterilization system. The predicate 510(k) was cleared for use with Steris V-PRO 60 sterlizer under K142782. The remaining models, Steris V-PRO 1 Plus and Steris V-PRO 1 naX were cleared for use with KIMGUARD* One -Step sterilization by FDA under 510(k) submissions, K092167 and K112805.

{21}------------------------------------------------

The subject device, under this submission, HALYARD SMART-FOLD* Sterilization Wrap is the reference devices which is cleared under K140963. Additionally, the HALYARD SMART-FOLD* Sterilization Wrap is comprised of the same material and composition as the base sheet of the Halyard ONE-STEP* sterilization wrap, which is the predicate device.

SUMMARY OF NONCLINICAL TESTS

Performance testing was conducted in accordance to the applicable requirements. Test results demonstrate that HALYARD SMART-FOLD* Sterilization Wrap (H450, H650), after steris V-PRO Low Temperature system (Models: V-PRO 1, V-PRO 1 Plus, V-PRO maX, V-PRO maX2 and V-PRO 60) maintains sterility until used.

Table 1.8: Sterilization Wrap Performance Tests
StudyDescriptionResults
Sterilant Penetration/EfficacyThe purpose of the sterilant penetration/Efficacy testing was to demonstrate sterilantpenetration into instrument trays wrapped with Halyard SMART-FOLD* (H450, H650)Sterilization Wrap. The testing was performed using end of shelf life sterilant todemonstrate efficacy of the sterilant under half cycle condition in the V- PRO maX lowtemperature sterilizer, worst case lumen cycle.The Halyard SMART-FOLD* (H450, H650) Sterilization wrapdemonstrated effective sterilant penetration into instrument trayswrapped with Halyard Health SMART-FOLD* (H450, H650)Sterilization Wrap after processing in the V-PRO maX lowtemperature sterilizer under half cycle condition in the V- PROmaX low temperature sterilizer, worst case lumen cycle.
Maintenance of 365-DayPackage IntegrityThe purpose of the maintenance of Sterility testing was to demonstrate that theHalyard SMART-FOLD* (H450, H650) is an effective barrier for maintaining thesterility of various sterilization load for 30, 180, 365 days after processing in the worst-case V-Pro 60 Low Temperature Sterilizer Lumen cycle and followed by an eventrelated shelf life study.The Halyard SMART-FOLD* (H450, H650) Sterilization wrapdemonstrated effective barrier using the simulated test pack forthe best and worst-case weight wraps for maintaining the sterilityof their contents for 30, 180, 365 days after processing in theworst-case V-PRO 60 Low Temperature Sterilizer Lumen cycleand followed by an event related shelf life study.
Material BiocompatibilityThe purpose of the biocompatibility testing was to demonstrate that the HalyardSMART-FOLD* (H450, H650) is non-irritating and non- cytotoxic in accordance toCytotoxicity, 10993-5, Biological Evaluation of medical devices - Tests for invitrocytotoxicity and 10993 -10:2010, Biological Evaluation of medical devices - Tests forIrritation and Sensitization.The Halyard SMART-FOLD* (H450, H650) Sterilization wrap wasfound to be non-cytotoxic and non-irritant based on the conditionsof these tests.
StudyDescriptionResults
Physical propertiesThe purpose of the physical propertiestesting was to demonstrate passingresults for the following physicalproperties for Halyard SMART-FOLD*(H450, H650) wrap:Bursting Properties ofFabrics - Part 2 - PneumaticMethod for Determination ofBursting Strength andBursting Distension, ISO13828-2:1999 Determination of the permeability of fabrics (AirPermeability), ISO 9237:1995 E Standard Test Method for Abrasion Resistanceof Textile Fabrics (Martindale Abrasion TesterMethod),ASTM D4966-12 Standard Test Method for Mass Per Unit Area(Weight) of Fabric,ASTM D3776/D3776M-09a Flame resistance requirements for Class Iproducts of > 3.5 secondsflame travel, CPSC 16 CFR 1610:2007 Water Resistance: Hydrostatic Pressure Test,AATCC-127 Textiles - Test methods for nonwovens - Part10: Lint andother particles generation in the dry state -Linting/Cleanliness, ISO 9073-10 Standard Test Method for Tearing Strength ofNonwoven Fabrics by the TrapezoidProcedure, ASTM D5733-99 Standard Test Method for Breaking Strengthand Elongation of Textile Fabrics (Grab Test),ASTM D5034-95 Test Method of Failure in Sewn Seams ofWoven Fabrics, ASTM D1683/D1683M-11 Standard Test Method forEvaluating the BacterialFiltration Efficiency (BFE) ofThe physical properties testing met the acceptance criteriaand demonstrated passing results for the Halyard SMART-FOLD* (H450, H650) Sterilization wrap.

{22}------------------------------------------------

{23}------------------------------------------------

Medical Face MaskMaterials, Using a BiologicalAerosol of Staphylococcusaureus, ASTM F2101

In summary, testing was conducted to evaluate the performance of the HALYARD SMART-FOLD* when used with Steris Low temperature system, including chemical composition, physical propertation leficacy, maintenance of package integrity and biocompatibility tests that have passed and met the acceptance criteria. The passing results from above sterilant penetration, maintenance of package integrity, biocompatibility and the physical properties demonstrate that the acceptance criteria and the specification for the performance test for the subject device, HALYARD SMART-FOLD* when used with Steris V-PRO Low sterilization system (Models STERIS V-PRO maX, STERIS V-PRO maX 2, STERIS V-PRO 1, STERIS V-PRO 1 Plus).

Conclusion

Based on the intended use, technological characteristics, performance data and nonclinical tests performed, the subject HALYARD SMART-FOLD Sterilization Wrap (H450, H650) is as safe, as effective, and performs as well as or better than the legally marketed prodicate device, K142782.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).