(30 days)
The EVS 4343A / EVS 4343AG / EVS 3643AG Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
The EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG is a flat-panel type digital X-ray detector that captures projection radiographic images in digital format within seconds, eliminating the need for an entire x-rav film or an image plate as an image capture medium. EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG differs from traditional X-ray systems in that, instead of exposing a film and chemically processing it to create a hard copy image, a device called a Detector is used to capture the image in electronic form.
The EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG Detector is an indirect conversion device in the form of a square plate in which converts the incoming X-rays into visible light. This visible light is then collected by an optical sensor, which generates an electric charges representation of the spatial distribution of the incoming X-ray quanta.
The charges are converted to a modulated electrical signal through thin film transistors. The amplified signal is converted to a voltage signal and is then converted from an analog to digital signal which can be transmitted to a viewed image print out, transmitted to remote viewing or stored as an electronic data file for later viewing.
The provided FDA 510(k) summary (K192400) for the DRTECH EVS 4343A, EVS 4343AG, EVS 3643A, and EVS 3643AG digital X-ray detectors focuses on demonstrating substantial equivalence to a predicate device (K162555). Therefore, the "acceptance criteria" discussed are primarily related to showing that the new devices perform as well as or better than the predicate, particularly in key physical performance metrics.
Here's an analysis of the acceptance criteria and the study details based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the predicate device's performance and the expectation that the new devices should meet or exceed these values for key metrics like DQE and MTF.
| Performance Parameter | Predicate Device (EVS 4343 / EVS 4343G) Acceptance Criteria | Subject Device (EVS 4343A, EVS 3643A, EVS 4343AG, EVS 3643AG) Reported Performance |
|---|---|---|
| DQE | EVS 4343: 43.9 % at 1.0 lp/mm EVS 4343G: 23.6 % at 1.0 lp/mm | EVS 4343A: 52.9 % at 1.0 lp/mm EVS 3643A: 50.5 % at 1.0 lp/mm EVS 4343AG: 27.2 % at 1.0 lp/mm EVS 3643AG: 26.3 % at 1.0 lp/mm |
| MTF | EVS 4343: 37.7 % at 2.0 lp/mm EVS 4343G: 34.0 % at 2.0 lp/mm | EVS 4343A: 44.1 % at 2.0 lp/mm EVS 3643A: 44.5 % at 2.0 lp/mm EVS 4343AG: 49.2 % at 2.0 lp/mm EVS 3643AG: 46.3 % at 2.0 lp/mm |
| Resolution | 3.5 lp/mm | 3.5 lp/mm |
The document states: "it is proved that the DQE and MTF of predicated device and subject device are basically equal or worth than the predicate device." and "As a result, subject devices performance is equal or worth than the predicate device." However, the presented data shows that the subject devices exceed the DQE and MTF values of the predicate device, indicating superior performance in these measured aspects.
2. Sample Size Used for the Test Set and Data Provenance
The provided summary does not specify a sample size for a clinical test set involving patients or images. The "non-clinical data" discussed pertains to bench testing of the detector's physical performance (DQE, MTF, Resolution). Therefore, the concepts of "test set" in the context of clinical images, "country of origin," and "retrospective/prospective" are not applicable to the non-clinical performance evaluation described.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not applicable as the evaluation was a non-clinical, bench-top performance assessment of the device's physical imaging characteristics (DQE, MTF, resolution), not a clinical study requiring expert interpretation of medical images.
4. Adjudication Method for the Test Set
This information is not applicable for the same reasons as point 3.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or described in this 510(k) summary. The submission focuses on demonstrating substantial equivalence based on technical specifications and non-clinical performance, not on a comparison of human reader performance with or without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is an X-ray detector, a hardware component that captures images. It does not contain an AI algorithm for image analysis in isolation (standalone) or for human-in-the-loop performance. Its "performance" refers to how well it acquires images, not how well it interprets them.
7. The Type of Ground Truth Used
The ground truth for the non-clinical performance evaluation (DQE, MTF, Resolution) would be based on physical phantom measurements and established international standards (e.g., IEC 62220-1) for characterizing X-ray detector performance. It is not expert consensus, pathology, or outcomes data, as those relate to clinical diagnostic accuracy.
8. The Sample Size for the Training Set
This 510(k) summary does not describe a training set. The device is a digital X-ray detector, which is a hardware component. There is no mention of machine learning or AI algorithms requiring a training set for this particular submission. The "study" here is a technical performance assessment of the detector itself.
9. How the Ground Truth for the Training Set was Established
This information is not applicable as no training set for an AI algorithm is described.
{0}------------------------------------------------
January 26, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
DRTECH Corporation Choul-Woo Shin Vice President Suite No.1, 1 Floor / Suite No.2, 3 Floor, 29, Dunchon-daero541 beon-gil, Jungwon-gu, Seongam-si, Gyeonggi-do 13216 REPUBLIC OF KOREA
Re: K192400
Trade/Device Name: EVS 4343A, EVS 4343AG, EVS 3643A, EVS 3643AG Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB
Dear Choul-Woo Shin:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 3, 2019. Specifically, FDA is updating this SE Letter as an administrative correction to update the 510(k) Summary document, where the dimensions of several detector models were previously misstated as "13" instead of "14".
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Dr. Lu Jiang, OHT8: Office of Radiological Health, 240-402-5779, lu.jiang@fda.hhs.gov.
Sincerely,
Laurel Burk, Ph.D. Director DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 3, 2019
DRTECH Corporation % Choul-Woo Shin Vice President Suite No.1. 1 Floor / Suite No.2. 3 Floor, 29. Dunchon-daero541 beon-gil, Jungwon-gu, Seongam-si, Gyeonggi-do 13216 REPUBLIC OF KOREA
Re: K192400
Trade/Device Name: EVS 4343A, EVS 4343AG, EVS 3643A, EVS 3643AG Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MOB Dated: August 27, 2019 Received: September 3, 2019
Dear Choul-Woo Shin:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@;tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hole 2. Nils
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K192400
Device Name EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG
Indications for Use (Describe)
The EVS 4343A / EVS 4343AG / EVS 3643AG Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
DRTECH Corporation Suite No.1, 1 Floor / Suite No. 2, 3 Floor, 29, Dunchon-Daero 541beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, Republic of Korea
510(k) Summary
[As required by 21 CFR 807.92]
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92
1. Date Prepared [21 CFR 807.92(a) (1)]
08/27/2019
2. Submitter's Information [21 CFR 807.92(a) (1)]
- Name of Sponsor: DRTECH Corporation ● ● Address: Suite No.1, 1 Floor / Suite No. 2, 3 Floor, 29, Dunchon-daero541 beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, Republic of Korea
- Contact Name: ● Choul-Woo Shin
- Telephone No.: + 82-31-779-7783 ●
- Fax No.: + 82-31-779-7790 ●
- Email Address : ● dwshin@drtech.co.kr
- Registration Number: 3005172103 ●
- Name of Manufacturer: Same as Sponsor ●
3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
- . Trade Name: EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG
- Common Name: Digital Flat Panel X-ray Detector ●
- Classification Name: Stationary X-ray System ●
- Classification Panel: Radiology
- Classification Regulation: 21 CFR 892.1680
- Product Code: MOB ●
- Device Class: ● II
4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
- 510(k) Number: K162555 ●
- . Applicant: DRTECH Corporation
- Trade Name: EVS 4343 / EVS 4343G ●
- Classification Name: Stationary X-ray System ●
- Classification Panel: Radiology
- Classification Regulation: 21 CFR 892.1680
- Product Code: MQB ●
- Device Class: ● II
006 1/7 510(k) Summary
{5}------------------------------------------------
DRTECH Corporation Suite No.1, 1 Floor / Suite No. 2, 3 Floor, 29, Dunchon-Daero 541beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, Republic of Korea
5. Description of the Modified Device [21 CFR 807.92(a) (4)]
- Addition of EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG: The differences . between the subject devices and the predicate devices are adaptor, cable type and performance (MTF & DQE). Predicate devices used tether cable, Lan cable and SSU, however, subject devices use 2 Lan cables and PoE Adaptor. In case of EVS 3643A and EVS 3643AG, the size is smaller than EVS 4343 and EVS 4343G.
- . Optional components were removed such as interface cable, generator interface cable, USB switch box and hand switch compared to predicated devices components.
The EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG is a flat-panel type digital X-ray detector that captures projection radiographic images in digital format within seconds, eliminating the need for an entire x-rav film or an image plate as an image capture medium. EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG differs from traditional X-ray systems in that, instead of exposing a film and chemically processing it to create a hard copy image, a device called a Detector is used to capture the image in electronic form.
6. Intended Use [21 CFR 807.92(a)(5)]
The intended use has not been changed as a result of the modification and is as follows:
The EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
7. Technological Characteristics [21 CFR 807.92(a)(6)]
The EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG Detector is an indirect conversion device in the form of a square plate in which converts the incoming X-rays into visible light. This visible light is then collected by an optical sensor, which generates an electric charges representation of the spatial distribution of the incoming X-ray quanta.
The charges are converted to a modulated electrical signal through thin film transistors. The amplified signal is converted to a voltage signal and is then converted from an analog to digital signal which can be transmitted to a viewed image print out, transmitted to remote viewing or stored as an electronic data file for later viewing.
Comparisons with the predicate, devices show the technological characteristics of the EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG to be same to the predicate devices. EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG is functionally identical to the predicate devices.
{6}------------------------------------------------
Contents
| Requirements | ||
|---|---|---|
| 80kHz ~ 240kHz | ||
| 60kVp ~ 150kVp |
| Power frequency | 30kHz ~ 240kHz | ||
|---|---|---|---|
| Generator | KV | 40kVp ~ 150kVp | |
| mA Range | 10mA ~ 1000mA | ||
| Exposure Time | 0.001~10sec | ||
| mAs Range | 0.1~1000mAs | ||
| Accuracy | ± 5% | ||
| Operating Type | moving | Stepping Motor | |
| Bucky | Spring | ||
| CAM Motor type | |||
| Static(Fixed) | |||
| Tray size(mm) | 460 mm x 483 mm x 15.5 mm or higher | ||
| Ratio | 5:1, 6:1, 8:1, 10:1, 12:1, 15:1 | ||
| Grid | Line | 85 ~ 215 Line | |
| SID | 100 ~ 180 cm |
8. Hardware and Software Requirements
- X-ray System Requirement
- The EVS 4343A / EVS 4343AG / EVS 3643AG detector is not compatible with trays that are less than 460 mm x 483 mm x 15.5 mm.
- Software Requirement
The EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG detector is compatible with Econsole1 (K152172).
{7}------------------------------------------------
9. Substantial Equivalence [21 CFR 807.92(b)]
| Parameter | Subject Device | Predicate Device | |||||
|---|---|---|---|---|---|---|---|
| 510(K) Number | Unknown | K162555 | |||||
| Manufacturer | DRTECH Corporation | DRTECH Corporation | |||||
| Model Name | EVS 4343A / EVS 4343AGEVS 3643A / EVS 3643AG | EVS 4343 / EVS 4343G | |||||
| Classification Name | Stationary X-ray System | ||||||
| Classification Panel | Radiology | ||||||
| ClassificationRegulation | 21 CFR 892.1680 | ||||||
| Product Code | MQB | ||||||
| Device Class | Class II | ||||||
| Intended Use | The EVS 4343A / EVS 4343AG / EVS3643A / EVS 3643AG Digital X-raydetector is indicated for digital imagingsolution designed for providing generalradiographic diagnosis of humananatomy. This device is intended toreplace film or screen basedradiographic systems in all generalpurpose diagnostic procedures. Thisdevice is not intended formammography applications | The EVS 4343A / EVS 4343AG / EVS3643A / EVS 3643AG Digital X-raydetector is indicated for digital imagingsolution designed for providing generalradiographic diagnosis of humananatomy. This device is intended toreplace film or screen basedradiographic systems in all generalpurpose diagnostic procedures. Thisdevice is not intended formammography applications. | |||||
| Panel Shape | EVS 4343A: Square PanelEVS 4343AG: Square PanelEVS 3643A: Rectangular PanelEVS 3643AG: Rectangular Panel | Square Panel | |||||
| DetectorSize | EVS 4343A: 17" X 17"EVS 4343AG: 17" X 17"EVS 3643A: 14" X 17"EVS 3643AG: 14" X 17" | 17" X 17" | |||||
| DesignDimensions | EVS 4343A:460(W) x 483(L) x 15.5(H)EVS 4343AG:460(W) x 483(L) x 15.5(H)EVS 3643A:460(W) × 409(L) × 15.5(H)EVS 3643AG:460(W) × 409(L) × 15.5(H) | 460(W) x 460(L) x 15(H) | |||||
| Pixel Pitch | |||||||
| Image Size | EVS 4343A: 3,072 x 3,072EVS 4343AG: 3,072 x 3,072EVS 3643A: 2,560 x 3,072EVS 3643AG: 2,560 x 3,072 | 3,072 x 3,072 | |||||
| Materials Scintillator | TFT –amorphous SiliconEVS 4343A/EVS 3643A: CsIEVS 4343AG/ EVS 3643AG: GOS | TFT –amorphous SiliconEVS 4343: CsIEVS 4343G: GOS | |||||
| PerformanceDQE | EVS 4343A: 52.9% at 1.0 lp/mmEVS 3643A: 50.5 % at 1.0 lp/mmEVS 4343AG: 27.2 % at 1.0 lp/mm | EVS 4343: 43.9 % at 1.0 lp/mmEVS 4343G: 23.6 % at 1.0 lp/mm | |||||
| MTF | EVS 3643AG: 26.3 % at 1.0 lp/mmEVS 4343A: 44.1 % at 2.0 lp/mmEVS 3643A: 44.5 % at 2.0 lp/mmEVS 4343AG: 49.2 % at 2.0 lp/mmEVS 3643AG: 46.3 % at 2.0 lp/mm | EVS 4343: 37.7 % at 2.0 lp/mmEVS 4343G: 34.0 % at 2.0 lp/mm | |||||
| Resolution | 3.5 lp/mm | 3.5 lp/mm | |||||
| Anatomical Sites | General Radiography | ||||||
| Power Supply | 100 | ||||||
| Component | PoE AdaptorLan cables (2 ea)AC Power CableEthernet Connector Bracket | ||||||
| CommunicationInterface | Wired: Ethernet | Wired: Ethernet |
{8}------------------------------------------------
DRTECH Corporation Suite No.1, 1 Floor / Suite No. 2, 3 Floor, 29, Dunchon-Daero 541beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, Republic of Korea
When compared to the predicate devices (K162555), the EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG presented in this submission have the same:
- Intended Use
- Technological characteristics •
- . Operating principle
- Performance (Resolution)
- Communication Method
A few differences are as follows
- Adaptor / cable type
- Components
- Size
- Performance (DQE and MTF)
There are no significant differences between the EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG and the predicate device(s) that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.
According to bench test report, it is proved that the DQE and MTF of predicated device and subject device are basically equal or worth than the predicate device. As a result, subject devices performance is equal or worth than the predicate device.
{9}------------------------------------------------
DRTECH
10. Summary of Non-Clinical Data [21 CFR 807.92(b)(1)]
The non-clinical performance testing constrains that the main physical values for comparison of Xray devices like DQE and MTF are basically equal or worth than the predicate device as following table:
| Parameter | Modified Device | Predicate Device | Remark |
|---|---|---|---|
| DQE | EVS 4343A: 52.9% at 1.0 lp/mm | EVS 4343: 43.9 % at 1.0 lp/mm | |
| EVS 3643A: 50.5 % at 1.0 lp/mm | |||
| EVS 4343AG: 27.2 % at 1.0 lp/mm | EVS 4343G: 23.6 % at 1.0 lp/mm | ||
| EVS 3643AG: 26.3 % at 1.0 lp/mm | |||
| MTF | EVS 4343A: 44.1 % at 2.0 lp/mm | EVS 4343: 37.7 % at 2.0 lp/mm | |
| EVS 3643A: 44.5 % at 2.0 lp/mm | |||
| EVS 4343AG: 49.2 % at 2.0 lp/mm | EVS 4343G: 34.0 % at 2.0 lp/mm | ||
| EVS 3643AG: 46.3 % at 2.0 lp/mm |
The EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG detector complies with the following international and FDA-recognized consensus standards:
| Recognition No. | Standard No. | Title of Standard | Remark |
|---|---|---|---|
| 19-4 | ANSI AAMI ES60601-1:2005/(R)2012 andA1:2012 | Medical electrical equipment- Part 1: Generalrequirements for basic safetyand essential performance(IEC 60601-1:2005, MOD) | |
| 12-289 | IEC 62220-1 Edition 1.02015-03 | Medical electrical equipment-Characteristics of digital X-ray imaging devices Part 1-1:Determination of thedetective quantum efficiencyDetectors used inradiographic imaging | |
| 5-40 | ISO 14971 Second edition2007-03-01 | Medical devices - Applicationof risk management tomedical devices | |
| 5-89 | IEC 60601-1-6 Edition 3.1 | Medical electrical equipment- Part 1-6: Generalrequirements for basic safetyand essential performance -Collateral standard: Usability | |
| 12-300 | PS 3.1 - 3.20 (2016) | Digital Imaging andCommunications in Medicine(DICOM) Set | |
| 19-8 | IEC 60601-1-2 Edition 4.0 | Medical electrical equipment- Part 1-2: Generalrequirements for basic safetyand essential performance -Collateral Standard:Electromagnetic disturbances- Requirements and tests | |
| 13-32 | ANSI AAMI IEC62304:2006 | Medical device software -Software life cycle processes |
{10}------------------------------------------------
11. Conclusion [21 CFR 807.92(b)(3)]
The modified EVS 4343A / EVS 4343AG / EVS 3643AG detector is substantially equivalent to the currently marketed and predicate device (EVS 4343(G), K162555) in terms of fundamental scientific technology, indications for use, and safety and effectiveness.
Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC 60601-1-2:2014, ANSI AAMI IEC 62304:2006 and IEC 62220-1-1 which complied with the requirements specified in the CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices.
The results of these tests demonstrate that The EVS 4343AG / EVS 3643A / EVS 3643AG meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data and safety testing demonstrates that the device is as safe, as effective, and performs as well or better than the predicate devices.
§ 892.1680 Stationary x-ray system.
(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.