(80 days)
Not Found
No
The summary describes a standard digital X-ray detector and its image capture process, with no mention of AI or ML for image analysis, processing, or interpretation.
No.
A therapeutic device is one that treats or prevents a disease or condition. This device is an X-ray detector used for diagnostic imaging, not for treatment.
Yes
The "Intended Use / Indications for Use" section explicitly states that the device is "intended for digital imaging solution designed for providing general radiographic diagnosis of human anatomy."
No
The device description explicitly details a physical flat-panel X-ray detector that converts X-rays into electrical signals, indicating it is a hardware device, not software-only.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information about a person's health.
- Device Function: The EVS 3643 and EVS 3643G Digital X-ray detector is used to capture projection radiographic images of the human anatomy. This is an in vivo imaging technique, meaning it is performed on a living organism.
- Intended Use: The intended use clearly states "general radiographic diagnosis of human anatomy," which involves imaging the body directly, not analyzing samples taken from the body.
The description of the device and its function aligns with a medical imaging device used for diagnostic purposes on the patient, not with samples from the patient.
N/A
Intended Use / Indications for Use
The EVS 3643 and EVS 3643G Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Product codes (comma separated list FDA assigned to the subject device)
MOB
Device Description
The EVS 3643(G) is a wired/wireless flat-panel type digital X-ray detector that captures projection radiographic images in digital format within seconds, eliminating the need for an entire x-ray film or an image plate as an image capture medium. EVS 3643(G) differs from traditional X-ray systems in that, instead of exposing a film and chemically processing it to create a hard copy image, a device called a Detector is used to capture the image in electronic form.
The EVS 3643(G) Detector is an indirect conversion device in the form of a square plate in which converts the incoming X-rays into visible light. This visible light is then collected by an optical sensor, which generates an electric charges representation of the spatial distribution of the incoming X-ray quanta.
The charges are converted to a modulated electrical signal thin film transistors. The amplified signal is converted to a voltage signal and is then converted from an analog to digital signal which can be transmitted to a viewed image print out, transmitted to remote viewing or stored as an electronic data file for later viewing.
The subject device integrates with X-ray system through three modes such as Auto Trigger mode (AT), Sync. Trigger mode and USB SW mode. The Auto Trigger Mode (AT) is available for acquiring images without any connection to X-ray generator interface cable is not required. The Sync. Trigger mode is the most common and recommended exposure mode at B2B scopes. User can achieve the high quality images with Sync. Tigger Mode. The USB SW Mode is the most common and recommended exposure mode at a retrofit scope. User can achieve high quality images with USB SW Mode.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-ray
Anatomical Site
human anatomy
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Summary of Non-Clinical Data:
The non-clinical performance testing constrains that the main physical values for comparison of Xray devices like DQE and MTF are basically equal or higher the predicate device as following table:
Parameter | Modified Device | Predicate Device K151942 | Predicate Device K150766 | Remark |
---|---|---|---|---|
DQE | EVS 3643: 36.8 % at 1 lp/mm, 25.56% at 3 lp/mm; EVS 3643G: 34.6 % at 1 lp/mm, 5 % at 3 lp/mm | 36.8 % at 1 lp/mm, 25.56 % at 3 lp/mm | 29.7 % at 1 lp/mm, 5.0 % at 3 lp/mm | Same |
MTF | EVS 3643: 35 % at 2 lp/mm; EVS 3643G: 30 % at 2 lp/mm | 35 % at 2.0 lp/mm | 17.4% at 2.0 lp/mm | Same |
The EVS 3643(G) complies with the following international and FDA-recognized consensus standards:
- Medical Electrical Equipment -- Part 1: General Requirements for Basic Safety And Essential Performance (IEC 60601-1:2005, Mod)
- Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance – Collateral Standard: Electromagnetic Compatibility - Requirements And Tests (Edition 3)
- Medical Devices - Application of Risk Management to Medical Devices. (General I (QS/RM))
- Medical electrical equipment - Characteristics of digital X-ray imaging devices - Part 1: Determination of the detective quantum efficiency
- Digital Imaging and Communications in Medicine (DICOM) Set
Summary of Clinical Data:
A single-blinded concurrence study according to CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices was conducted. The study confirmed that the new x-ray detectors EVS 3643G provide images of equivalent diagnostic capability to the predicate devices (K151942 and K150766), and its results demonstrate substantial equivalence.
Clinical images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device (note X-ray detector technology that is identical to the predicate device) but they provide further evidence in addition to the laboratory performance data to show that the subject device works as intended.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.1680 Stationary x-ray system.
(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle or bird with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 2, 2016
DRTECH Corporation % Mr. Choul-Woo Shin Suite no. 2, 3td floor, 29, Dunchon-daero 541 beon-gil Jungwon-gu. Seongnam-si. Gyeonggi-do. 13230 REPUBLIC OF KOREA
Re: K162552
Trade/Device Name: EVS 3643, EVS 3643G Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: October 31, 2016 Received: November 2, 2016
Dear Mr. Choul-Woo Shin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K162552
Device Name EVS 3643, EVS 3643G
Indications for Use (Describe)
The EVS 3643 and EVS 3643G Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
DRIECH
510(k) Summary
[As required by 21 CFR 807.92]
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92
1. Date Prepared [21 CFR 807.92(a) (1)]
10/28/2016
2. Submitter's Information [21 CFR 807.92(a) (1)]
Name of Sponsor: | DRTECH Corporation | |
---|---|---|
• | Address: | Suit No. 2, 3 Floor, 29, Dunchon-daero541 beon-gil, |
Jungwon-gu, Seongnam-si, Gyeonggi-do, 13230, Republic of Korea | ||
• | Contact Name: | Choul-Woo Shin |
• | Telephone No.: | + 82-31-779-7783 |
• | Fax No.: | + 82-31-779-7790 |
• | Email Address : | cwshin@drtech.co.kr |
• | Registration Number: | 3005172103 |
• | Name of Manufacturer: | Same as Sponsor |
3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
● | Trade Name: | EVS 3643, EVS 3643G |
---|---|---|
● | Common Name: | Digital Flat Panel X-ray Detector |
● | Classification Name: | Stationary X-ray System |
● | Classification Panel: | Radiology |
● | Classification Regulation: | 21 CFR 892.1680 |
● | Product Code: | MQB |
● | Device Class: | II |
● | 501(k) Number: | K162552 |
4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
Parameter | Predicate Device |
---|---|
• 510(k) Number: | K150766 |
• Applicant: | Carestream Health, Inc. |
• Trade Name: | Carestream DRX-1 System w/ DRX |
Plus Detectors | |
• Classification Name: | |
• Classification Panel: | |
• Classification Regulation: | |
• Product Code: | |
• Device Class: |
4
5. Description of the Modified Device [21 CFR 807.92(a) (4)]
- A wireless charging function and accessories (EVS-WPCS) were added to EVS 3643(G) for wireless charging.
- Addition of EVS 3643G: This is to notify that modified EVS 3643 / EVS 3643G are same Hardware, Software and components. But scintillator layer are different. Scintillator is a phosphor that produces scintillations.
Model Name | Scintillator layer |
---|---|
EVS 3643 | CsI (Cesium Iodide) |
EVS 3643G | GoS (Gadolinium Oxysulfide) |
The EVS 3643(G) is a wired/wireless flat-panel type digital X-ray detector that captures projection radiographic images in digital format within seconds, eliminating the need for an entire x-ray film or an image plate as an image capture medium. EVS 3643(G) differs from traditional X-ray systems in that, instead of exposing a film and chemically processing it to create a hard copy image, a device called a Detector is used to capture the image in electronic form.
6. Intended Use [21 CFR 807.92(a)(5)]
The intended use has not changed as a result of the modification and is as follows:
The EVS 3643 and EVS 3643G Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
7. Technological Characteristics [21 CFR 807.92(a)(6)]
The EVS 3643(G) Detector is an indirect conversion device in the form of a square plate in which converts the incoming X-rays into visible light. This visible light is then collected by an optical sensor, which generates an electric charges representation of the spatial distribution of the incoming X-ray quanta.
The charges are converted to a modulated electrical signal thin film transistors. The amplified signal is converted to a voltage signal and is then converted from an analog to digital signal which can be transmitted to a viewed image print out, transmitted to remote viewing or stored as an electronic data file for later viewing.
Comparisons with the predicate, devices show the technological characteristics of the EVS 3643(G) to be same to the predicate devices. The EVS 3643(G) is functionally identical to the predicate devices.
The subject device integrates with X-ray system through three modes such as Auto Trigger mode (AT), Sync. Trigger mode and USB SW mode. The Auto Trigger Mode (AT) is available for acquiring images without any connection to X-ray generator interface cable is not required. The Sync. Trigger mode is the most common and recommended exposure mode at B2B scopes. User can achieve the high quality images with Sync. Tigger Mode. The USB SW Mode is the most common and recommended exposure mode at a retrofit scope. User can achieve high quality images with USB SW Mode.
5
Image /page/5/Picture/0 description: The image shows the word "DRTECH" in blue font. The letters are all capitalized and sans-serif. There is a small blue square above and to the left of the "D". A cursor is pointing at the "D".
8. Hardware and Software Requirements
- X-ray System Requirement
Contents | Requirements | |
---|---|---|
Generator | Power frequency | 30kHz ~ 240kHz |
KV | 40kVp ~ 150kVp | |
mA Range | 10mA ~ 1000mA | |
Exposure Time | 0.001~10sec | |
mAs Range | 0.1~1000mAs | |
Accuracy | $\pm$ 5% | |
Bucky | Operating Type | moving: Stepping Motor |
Spring | ||
CAM Motor type | ||
Static(Fixed) | ||
Trey size (mm) | 386 x 460 x 15 or higher | |
Grid | Ratio | 5:1, 6:1, 8:1, 10:1, 12:1, 15:1 |
Line | 85 ~ 215 Line | |
SID | 100 ~ 180 cm |
- The EVS 3643(G) detector is not compatible with trays that are less than 386 mm x 15 mm.
- Software Requirement
The EVS 3643(G) detector is compatible with Econsole1 (K152172).
9. Substantial Equivalence [21 CFR 807.92(b)]
Parameter | Subject Device | Predicate Device | Remark | ||
---|---|---|---|---|---|
510(K) Number | K162552 | K151942 | K150766 | - | |
Manufacturer | DRTECH | ||||
Corporation | DRTECH | ||||
Corporation | Carestream | ||||
Health, Inc. | - | ||||
Model Name | EVS 3643, EVS | ||||
3643G | EVS 3643 | Carestream DRX-1 | |||
System w/ DRX | |||||
Plus Detectors | - | ||||
Classification Name | Stationary X-ray System | same | |||
Classification Panel | Radiology | same | |||
Classification | |||||
Regulation | 21 CFR 892.1680 | same | |||
Product Code | MQB | same | |||
Device Class | Class II | same | |||
Intended Use | The EVS 3643, | ||||
EVS 3643G Digital | |||||
X-ray detector is | |||||
indicated | |||||
for digital imaging | |||||
solution designed | |||||
for providing | |||||
general | The EVS 3643 | ||||
Digital X-ray | |||||
detector is | |||||
indicated | |||||
for digital imaging | |||||
solution designed | |||||
for providing | |||||
general | The device is | ||||
intended to | |||||
capture for | |||||
display | |||||
radiographic | |||||
images of human | |||||
anatomy | same | ||||
radiographic | |||||
diagnosis of human | |||||
anatomy. This | |||||
device is intended | |||||
to replace film or | |||||
screen based | |||||
radiographic | |||||
systems in all | |||||
general purpose | |||||
diagnostic | |||||
procedures. This | |||||
device is not | |||||
intended for | |||||
mammography | |||||
applications | radiographic | ||||
diagnosis of human | |||||
anatomy. This | |||||
device is intended | |||||
to replace film or | |||||
screen based | |||||
radiographic | |||||
systems in all | |||||
general purpose | |||||
diagnostic | |||||
procedures. This | |||||
device is not | |||||
intended for | |||||
mammography | |||||
applications | including both | ||||
pediatric and | |||||
adult patients. | |||||
The device is | |||||
intended for use | |||||
in general | |||||
projection | |||||
radiographic | |||||
applications | |||||
wherever | |||||
conventional | |||||
screen-film | |||||
systems or CR | |||||
systems may be | |||||
used. Excluded | |||||
from the | |||||
indications for | |||||
use are | |||||
mammography, | |||||
fluoroscopy, and | |||||
angiography | |||||
applications. | |||||
Design | Panel | ||||
Shape | Rectangular Panel | Rectangular Panel | Rectangular Panel | same | |
Detector | |||||
Size | 13" X 17" | 13" X 17" | 13" X 17" | same | |
Pixel Pitch | $140μm$ | $140µm$ | $139μm$ | same | |
Materials Scintillator | TFT -amorphous | ||||
Silicon | |||||
EVS 3643: CsI | |||||
EVS 3643G: GOS | TFT -amorphous | ||||
Silicon | |||||
EVS 3643: CsI | TFT -amorphous | ||||
Silicon | |||||
GOS | same | ||||
Performance | DQE | EVS 3643: | |||
36.8 % at 1 lp/mm | |||||
25.56 % at 3 lp/mm | |||||
EVS 3643G: | |||||
34.6 % at 1 lp/mm | |||||
5 % at 3 lp/mm | 36.8 % at 1 lp/mm | ||||
25.56 % at 3 lp/mm | 29.7 % at 1 lp/mm | ||||
5.0 % at 3 lp/mm | same | ||||
MTF | EVS 3643: | ||||
37.8 % at 2 lp/mm | |||||
EVS 3643G: | |||||
34 % at 2 lp/mm | 37.8 % at 2 lp/mm | 17.4% at 2 lp/mm | same | ||
Resolution | 3.5 lp/mm | 3.5 lp/mm | 3.5 lp/mm | same | |
Wireless | |||||
Charging | available | non-available | non-available | different | |
Anatomical Sites | General | ||||
Radiography | General | ||||
Radiography | General | ||||
Radiography | same | ||||
Power Supply | 100 | ||||
50/60 Hz | 100 | ||||
Hz | 100 | ||||
50/60 Hz | same | ||||
Communication | Wire | Wire | same |
6
Image /page/6/Picture/0 description: The image shows the logo for IDRTECH. The logo is blue and features a small square above the "I" in IDRTECH. A mouse cursor is also visible in the image, pointing towards the "I" in IDRTECH.
002-4/6
510(k) Summary
7
DR TECH
Method | |||
---|---|---|---|
Wireless | |||
• IEEE | |||
802.11a//g/n | |||
(2.4 GHz / 5 GHz) | |||
• Security: | |||
WEP/WPA/WP | |||
A2 | Wireless | ||
• IEEE | |||
802.11a//g/n | |||
(2.4 GHz / 5 GHz) | |||
• Security: | |||
WEP/WPA/WP | |||
A2 | - | ||
Exposure Mode | USB SW Mode | ||
Sync Trigger Mode | |||
Auto Trigger Mode | USB SW Mode | ||
Sync Trigger Mode | |||
Auto Trigger Mode | same |
When compared to the predicate devices (K151942 and K150766), the EVS 3643(G) presented in this submission has the same:
- Intended Use
- Technological characteristics
- · Operating principle
- Materials Scintillator
- · Design features
- · Performance (DQE, MTF, Resolution)
A few differences are as follows
- · Performance (Wireless charging)
There are no significant differences between the EVS 3643(G) and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.
10. Summary of Non-Clinical Data [21 CFR 807.92(b)(1)]
The non-clinical performance testing constrains that the main physical values for comparison of Xray devices like DQE and MTF are basically equal or higher the predicate device as following table:
Parameter | Modified Device | Predicate Device | Remark | |
---|---|---|---|---|
K151942 | K150766 | |||
DQE | EVS 3643: | |||
36.8 % at 1 lp/mm | ||||
25.56% at 3 lp/mm | ||||
EVS 3643G: | ||||
34.6 % at 1 lp/mm | ||||
5 % at 3 lp/mm | 36.8 % at 1 lp/mm | |||
25.56 % at 3 lp/mm | 29.7 % at 1 lp/mm | |||
5.0 % at 3 lp/mm | Same | |||
MTF | EVS 3643: | |||
35 % at 2 lp/mm | ||||
EVS 3643G: | ||||
30 % at 2 lp/mm | 35 % at 2.0 lp/mm | 17.4% at 2.0 lp/mm | Same |
8
The EVS 3643(G) complies with the following international and FDA-recognized consensus standards:
| AAMI ANSI ES60601-1: | Medical Electrical Equipment -- Part 1: General Requirements for Basic
Safety And Essential Performance (IEC 60601-1:2005, Mod) |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2: | Medical Electrical Equipment - Part 1-2: General Requirements For Basic
Safety And Essential Performance – Collateral Standard: Electromagnetic
Compatibility - Requirements And Tests (Edition 3) |
| ISO 14971: | Medical Devices - Application of Risk Management to Medical Devices.
(General I (QS/RM)) |
| IEC 62220-1: | Medical electrical equipment - Characteristics of digital X-ray imaging
devices - Part 1: Determination of the detective quantum efficiency |
| NEMA PS 3.1 - 3.20: | Digital Imaging and Communications in Medicine (DICOM) Set |
11. Summary of Clinical Data [21 CFR 807.92(b)(2)]
A single-blinded concurrence study according to CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices was conducted, and the study confirmed that the new x-ray detectors EVS 3643G provide images of equivalent diagnostic capability to the predicate devices (K151942 and K150766) and its results demonstrate substantial equivalence.
Clinical images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device (note X-ray detector technology that is identical to the predicate device) but they provide further evidence in addition to the laboratory performance data to show that the subject device works as intended.
12. Conclusion [21 CFR 807.92(b)(3)]
The modified EVS 3643(G) is substantially equivalent to the currently marketed and predicate device (EVS 3643(K151942) and Carestream DRX-1 System w/ DRX Plus Detectors(K150766)) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC60601-1, IEC 60601-1-2, IEC6220-1 and IEC 62133 and the clinical test, which complied with the requirements specified in the CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices.
The results of these tests demonstrate that EVS 3643(G) meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing, and clinical image concurrence data demonstrates that the device is as safe, as effective, and performs as well or better than the predicate devices.