Search Results
Found 3 results
510(k) Data Aggregation
(167 days)
Filter CareStar Plus Intended use Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and respiratory use. Indications All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in pediatric (with a tidal volume between 100 and adult patients, depending on the respective device. Filter SafeStar Plus Intended use Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and respiratory use. Indications All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in adult patients. Filter/HME TwinStar Plus Intended use Bidirectionally breathing system filter against bacterial and viral contamination for anesthetic and for respiratory use, as well as heat and moisture exchanger for humidifying respired gases for the patient. Indications All devices are intended for single use up to 24 hours and must be used by trained medical personnel only. The devices are designed for use with ventilators and anesthesia machines. They are intended for use in adult, pediatric and neonatal patients, depending on the respective device.
The devices are breathing circuit filters used to filter the inhaled and/or the exhaled air of the patient against microbiological and particulate matter from the gases in the breathing circuit. They enclose a filter material in a housing that fits to standard breathing system connectors. Additionally, there are breathing system filters combined with a foam to function as HME (Heat and Moisture Exchangers) for passively humidifying the inspired air. The portfolio contains the following types of breathing circuit filters: - Filter CareStar Plus are electrostatic filters for use against contamination with microorganisms - . Filter SafeStar Plus are mechanical filters for use against contamination with microorganisms - Filter/HME TwinStar Plus are filters for use against contamination with microorganisms and for passive humidification of breathing gases
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. A table of acceptance criteria and the reported device performance
The provided document describes three devices: Filter CareStar Plus, Filter SafeStar Plus, and Filter/HME TwinStar Plus. The table below compiles the acceptance criteria and reported performance for these devices based on the "Summary of non-clinical testing" section (pages 18-19).
Test Method & Purpose | Acceptance Criteria | Reported Performance |
---|---|---|
ISO 9360-1:2000 - Determination and Evaluation of Pneumatic Compliance | Compliance is less than or equal to 1mL/kPa at 15, 30, 60, and 70 hPa | PASSED |
ISO 9360-1:2000 - Determination and Evaluation of Pneumatic Leakage | Pneumatic leakage is less than or equal to 50mL/min at 70hPa. | PASSED |
ISO 9360-1:2000 - Determination and Evaluation of Pressure Drop (Pneumatic Resistance) | Pneumatic resistance/pressure drop is acc. to IfU value | PASSED |
ISO 5356-1:2015 - Evaluation of Conical Connectors (ISO 5356-1) | Cone dimensions comply with ISO 5356-1. | PASSED |
ISO 80369-7:2021 - Luer Lock Connector (ISO 80369-7) | Luer-Lock connector fulfills the requirements laid out in ISO 80369-7. | PASSED |
IEC 60601-1:2005 - Product's Ability to Withstand Damage from Dropping | When dropped, the product should not suffer any damage which influences its function. | PASSED |
ISTA 3A - Product Durability During Transport, Mechanical Aspects | - The packaging shows no or minor damage |
- The DUT (Device Under Test) shows no signs of damage and retains functionality after simulated transport. | PASSED |
| ISO 23328-1:2003 - Filtration Efficiency (Particulate Matter) incl. Usage Time | - Adult and pediatric electrostatic filters achieve a filtration efficiency of >90% before and after the specified usage time - Neonatal filters achieve a filtration efficiency of >75% before and after the specified usage time
- Mechanical filters achieve a filtration efficiency of >99% before and after the specified usage time
- Mechanical filters achieve a HEPA classification ≥ class H13 | PASSED |
| ASTM F2101:2019 - Filtration Efficiency (Viral and Bacterial) | - Electrostatic filters achieve 99.99% (bacterial) and 99.9% (viral) filtration efficiency - Electrostatic filters for neo applications achieve 99.98% (bacterial) and 99.9% (viral) filtration efficiency. | PASSED |
| ISO 10993:2018 and ISO 18562-1:2017 - Evaluation of Product's Biological Compatibility | Evaluation according to ISO 10993:2018 and/or ISO 18562-1:2017 | PASSED |
| ISO 9360-1:2000 - Evaluation of HME Water Loss, Resistance | · pediatric/neonatal: The moisture loss shall be
Ask a specific question about this device
(197 days)
For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and/or expired gases is desired and to add, maintain, retain moisture for the exhaled breathe of the patient. Use up to 24 hours.
The Models of the filters are A501-1, A501-2, A501-3, and A501-4. The filters are intended for patient population that weighs greater than 10Kgs.
Breathing circuit bacterial/viral filter has into two models: - 1) Bacterial/ viral filters; 2) Bacterial/viral filters HME. Breathing circuit bacterial/viral filters are available in multiple sizes and shapes, rectangular and round, and incorporate standard 15/22 mm connectors with a gas sampling luer port. The depth filter use electrostatic media for filtration and a foam media for the HME media.
The provided document is a 510(k) premarket notification for a medical device: a Breathing Circuit Bacterial/Viral Filter manufactured by Zhejiang Haisheng Medical Device Co., Ltd. The document asserts that the device is substantially equivalent to a legally marketed predicate device (EMS Electra Filter and Filter/HME, K013122).
Based on the provided text, here's information regarding the acceptance criteria and the study that proves the device meets those criteria:
1. Table of Acceptance Criteria and Reported Device Performance
The device's performance is compared directly to the predicate device, implying that the predicate's performance metrics serve as the acceptance criteria for achieving substantial equivalence.
Acceptance Criteria (from Predicate) | Reported Device Performance (Zhejiang Haisheng Breathing Circuit Bacterial/Viral Filter) | Met/Not Met |
---|---|---|
Resistance to Flow: | ||
Ask a specific question about this device
(230 days)
Anesthesia / Respiratory Filters
For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is desired.
Model – HEPA filter 3000/04 – Single patient use for exhalation limb of circuit on NPB 700 series ventilators
Models – HEPA filters - 6500/01, 6888/01, 8222/01, 8444/01 Model – Non-HEPA filters - 4000/01 Single patient use up to 24 hours. Patient tidal volumes > 150 ml, when applicable.
The Air Safety HEPA and Non-HEPA filters are available in multiple sizes and shapes, and incorporate standard 15 / 22 mm connectors with or without a gas sampling luer port. Some models adapt to fit ventilator exhalation limb only. The depth (HEPA filtration) filter uses a pleated paper fiber for filtration. Filters are tested for rating performance according to EN 13328 Salt for Breathing System filtration performance. The "HEPA" performance was also tested in accordance to DOE-3025-99, DOE-3020-97 and ASTM D2986 - DOP. The electrostatic (non-HEPA filtration) filters are tested by Nelson Laboratories for BFE and VFE.
This document describes the safety and effectiveness of Air Safety HEPA and Non-HEPA Filters. However, it does not fit the typical format of a study designed to prove a device meets acceptance criteria, especially in the context of an AI/ML powered device. This is a 510(k) premarket notification for traditional medical devices (filters), and therefore, many of the requested categories are not applicable.
Here's an attempt to answer the questions based on the provided text, with clear indications where the information is not available or not relevant to this type of device submission.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present "acceptance criteria" in a pass/fail format with direct device performance comparison in a single table, as might be found in a study for an AI/ML device. Instead, it lists technical characteristics and performance standards that the device meets.
However, based on the General Technical Characteristics table, we can infer performance criteria and compare them to the device's reported performance:
Attribute/Performance Standard | Acceptance Criteria (Inferred from Predicate/Standards) | Reported Device Performance (Air Safety Filters) |
---|---|---|
Dead Space (ml) | Comparable to predicate / within acceptable range for intended use | 45 to 84 ml (various models), 209 mm for Model 3000/04 |
Resistance to flow | 99.97% of 0.3 micron DOP particle @ 60 Lpm) | 99.99999% (Models 6500/01, 6888/01, 8222/01, 8444/01) |
Model 3000/04 - 99.9999% | ||
Bacterial Filtration Efficiency (BFE) - Nelson (for listed HEPA models) | Standard requirement for bacterial filters (implied by predicate and testing) | 99.99975% (Models 6500/01, 6888/01, 8222/01, 8444/01) |
Model 3000/04 - 99.9999% | ||
Viral Filtration Efficiency (VFE) - Nelson (for listed HEPA models) | Standard requirement for viral filters (implied by predicate and testing) | 99.99996% (Models 6500/01, 6888/01, 8222/01, 8444/01) |
Bacterial Filtration Efficiency (BFE) - Nelson (for Non-HEPA Model 4000/01) | Standard requirement for bacterial filters (implied) | 99.99996% |
Viral Filtration Efficiency (VFE) - Nelson (for Non-HEPA Model 4000/01) | Standard requirement for viral filters (implied) | 99.99925% |
Filtration Performance (General) | According to EN 13328 Salt for Breathing System filtration performance. | "Filters are tested for rating performance according to EN 13328 Salt for Breathing System filtration performance." (Implied compliance) |
DOP Test Performance (for HEPA) | > 99.97% of 0.3 micron DOP particle @ 60 Lpm | > 99.97% of 0.3 micron DOP particle @ 60 Lpm |
Standard 15/22 mm connectors | Required for compatibility | Yes |
Gas sampling port | Optional | Optional |
Single patient use up to 24 hours | Yes | Yes |
ISO 5356-1 Conical 15/22 | Compliance with standard | Yes |
ISO 594-2 Luer Fittings | Compliance with standard | Yes |
Summary of "Study" to Prove Device Meets Criteria:
The document details the device characteristics and states that the devices were tested to specific performance standards. This is not a "study" in the clinical trial sense but rather a series of engineering and laboratory performance tests.
- Testing Methodologies:
- Filtration Performance: Tested according to EN 13328 Salt for Breathing System filtration performance.
- HEPA Performance (Specific models): Tested in accordance with DOE-3025-99, DOE-3020-97, and ASTM D2986 - DOP.
- Electrostatic (non-HEPA) Filtration (BFE and VFE): Tested by Nelson Laboratories.
- Conclusion: The submission concludes that "The data within the submission demonstrates that the proposed devices when compared to the legally marketed predicate devices are safe and effective and substantially equivalent." This implies that the performance characteristics met or exceeded those of the predicate devices and applicable standards.
Information Not Applicable / Not Provided for this Device Type:
The following requested information is primarily relevant for AI/ML-powered diagnostic or predictive devices, or those requiring clinical studies with human assessors. The provided document concerns a physical filter, for which these criteria are generally not applicable or not detailed in this type of 510(k) submission.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not Applicable / Not provided in this context. For physical filters, testing typically involves laboratory samples of the filters themselves, not patient data in the sense of AI/ML. The number of filter units tested for each specific performance claim (e.g., BFE, VFE, resistance) is not specified. Data provenance like "country of origin" for test data is not a typical requirement for device performance testing in this manner.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not Applicable. "Ground truth" established by experts (like radiologists for an image analysis AI) is not relevant for a breathing circuit filter. Performance is measured objectively through standardized laboratory tests (e.g., particle counts, pressure measurements).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. Adjudication methods are used to resolve discrepancies in expert opinions, typically for establishing ground truth in clinical datasets. This is not relevant for the objective performance testing of a physical filter.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This is a physical filter, not an AI/ML-powered diagnostic tool that assists human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a physical filter, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not Applicable. The "ground truth" for a filter's performance is derived from standardized physical and biological challenge tests (e.g., aerosolized particles, bacterial/viral suspensions) and measurements, not clinical expert consensus or pathology.
8. The sample size for the training set:
- Not Applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established:
- Not Applicable. As there is no training set for an AI/ML algorithm, this question is not relevant.
Ask a specific question about this device
Page 1 of 1