Search Results
Found 1 results
510(k) Data Aggregation
(185 days)
Besmed HME Filter (Heat and Moisture Exchange) is intended to decrease the transmission of bacteria and viruses to/from a patient, and to maintain moisture levels in the patient's respiratory tract while administering anesthesia gas, providing artificial respiration, or assisting with ventilation.
The Besmed HME Filter is typically placed at the patient's end of the breathing system, specifically between the circuits Y - piece and either the catheter mount or patient airway device. It is essential to note that the Besmed HME Filter is intended for single-use only and should be used on a single patient for up to 24 hours.
Besmed HME Filter is intended to be used within critical care and hospital/ institutional environments by qualified personnel.
Besmed Bacterial Filter
The purpose of the Besmed Bacterial Filter is to diminish the passage of bacteria and viruses to patients receiving anesthesia gas. Besmed Bacterial Filter is designed for utilization with ventilators, anesthesia machines, and open flow systems where there is a need for filtration of inhaled gases.
The Besmed Bacterial Filter is a disposable device intended for the exclusive use of a single patient within a 24-hour timeframe. Bacterial Filter is designed for utilization in hospital environments by medical professionals who possess adequate training in the operation of mechanical ventilators, respiratory systems, and humidification systems.
Besmed HMEF
The Besmed HME Filter absorbs the heat and moisture exhaled by the patients. Upon inhalation, it releases the previously retained heat and humidity to the dry inhalation gases. The HME filter is designed with electrostatic cotton for filtration, as well as a heat-preserved and moisture-absorbed paper roll for maintaining the warmth and humidity of the airway in a respiratory system. Additionally, it eliminates bacterial and viral contaminants from the breathing circuit gas. To ensure passive humidification treatment, it is recommended to install the Heat and Moisture Exchanger Filter (HMEF) on the patient's end. Besmed HME Filter is intended for adult and child patients with the tidal volume between 150ml-1000ml.
Besmed Bacterial Filter
Besmed Bacterial Filter is for single-use purpose. The intended use is achieved through the implementation of an electrostatic filtration mechanism. Transparent and compact housing material makes it easy to check the surface of the filter media for a foreign substance.
The provided text describes the acceptance criteria and the results of non-clinical testing for the "Besmed Bacterial Filter and HMEF." This summary focuses on the non-clinical test results, as no clinical testing was performed or required.
Acceptance Criteria and Reported Device Performance
The device's performance was evaluated against a set of predetermined acceptance criteria across various standards. The table below summarizes these criteria and the reported results.
Test Standards | Test Performed | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Packaging for Terminally Sterilized Medical Devices | |||
ISO 11737-2:2020 | Sterility test | Negative | Pass |
ISO 10993-7:2008 | EO/ ECH/ EG residual test | Non-Detected | Pass |
ASTM F1929-23 | Dye Penetration Test | No Penetration | Pass |
ASTM F1140/F1140M-13(2020)e1 | Burst test | Package successfully held for creep duration | Pass |
Creep test | It should maintain the specified pressure for 30 seconds | Pass | |
ASTM F88/F88M-23 | Seal strength of flexible barrier materials test | Seal strength shall not be greater than ± 2SD compare to pre transit package | Pass |
ASTM F1608-21 | Microbial Ranking Test | LRV>3.0 (99.9%) | Pass |
ISTA 2A:2011 | Packaged products test | No visible damage | Pass |
Shelf Life Test | |||
ASTM F1980-21 | Accelerated aging test | No visible damage after aging simulation | Device meets its performance specification |
Performance Test | |||
Measurement of moisture loss | 99.9% @15 LPM and 30 LPM under 0 hour and 24 hour conditioning | Pass | |
ISO 80369-7:2021 | Leakage by pressure decay | Leakage rate should not exceed 0.005 Pa·m³/s @ 330 kPa for 20 s | Pass |
Sub-atmospheric pressure air leakage | Leakage rate should not exceed 0.005 Pa·m³/s @ 101kPa for 20 s | Pass | |
Stress cracking | No signs of seal break and leakage @101 kPa after being stress | Pass | |
Resistance to separation from axial load | No disconnections observed at an axial force of 35 N | Pass | |
Resistance to separation from unscrewing | No disconnections observed at unscrewing torque of 0.020Nm | Pass | |
Resistance to overriding | No override the threads observed at torque of 0.17 Nm | Pass | |
ISO 5356-1:2015 | Security of engagement | No disconnections observed at axial separation force of (50 ± 5) N for 10 s | Pass |
Leakage from 22 mm latching sockets | No signs of leakage at (8 ± 0.5) kPa | Pass | |
Drop procedure for 22 mm latching sockets | No signs of damage during the test | Pass | |
ASTM F2101-19 | Bacterial / Virus Filtration Efficiency test | Evaluate BFE/VFE filter efficiency shall be greater than 90% | Pass |
Biocompatibility Test | |||
ISO 10993-5:2009 | In Vitro Cytotoxicity test | Not more than 50 % of the cells are round | Pass |
ISO 10993-10:2021 | Skin sensitization tests (Maximization Test) | Did not produce skin sensitization on Guinea pig | Pass |
ISO 10993-23:2021 | Tests for irritation | Did not cause intracutaneous irritation | Pass |
ISO 10993-11:2017 | Acute systemic toxicity study | Did not cause systemic toxicity reaction or death | Pass |
USP 151 | Pyrogenicity Test | Did not induce pyrogenic response | Pass |
ISO 18562-2:2024 | Emissions of particulate matter | Measured concentrations of particulate matter PM2.5 ≤ 35µg/m³; PM10 ≤ 150µg/m³ | Pass |
ISO 18562-3:2024 | Emissions of VOCs | The MOS (Margin Of Safety) value shall be higher than 1 for identified substances. | Pass |
Study Details
The provided document describes a non-clinical testing summary to support the substantial equivalence of the "Besmed Bacterial Filter and HMEF."
-
Sample Size used for the test set and the data provenance: The document does not specify the exact sample sizes (number of devices) used for each individual test. The data provenance is implied to be from internal testing conducted by Besmed Health Business Corp., likely in Taiwan (New Taipei City, Taiwan is their address). The nature of the tests (performance, packaging, biocompatibility) indicates that these were prospective evaluations of manufactured devices.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not applicable as the studies are non-clinical, laboratory-based tests against established international and ASTM standards. Ground truth would be defined by the criteria within these standards, which are met if the device passes the specified thresholds. No human expert consensus for "ground truth" in clinical interpretation is relevant here.
-
Adjudication method for the test set: Not applicable for non-clinical, objective laboratory tests against defined standards.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This is not applicable as the device is a physical medical device (breathing circuit bacterial filter and HMEF), not an AI-powered diagnostic or assistive technology.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: This is not applicable as the device is a physical medical device, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for these tests is based on the acceptance criteria defined by international standards (ISO) and ASTM standards. These standards provide objective, measurable thresholds for performance, safety, and biocompatibility.
-
The sample size for the training set: This is not applicable as the device is a physical medical device, not a machine learning model that requires a training set.
-
How the ground truth for the training set was established: This is not applicable for the same reason as above.
Ask a specific question about this device
Page 1 of 1