(367 days)
The 740 Ventilator System is used to provide continuous ventilation to patient's requiring respiratory support. This device is used for a wide range of patients from pediatric to adult and for a wide variety of clinical conditions.
The device is a low-cost, critical care ventilator intended to provide continuous ventilation for pediatric and adult patients. The 740 Ventilator supplies mandatory or spontaneous breaths. A mandatory (or assisted) breath provides the patient with a preset tidal volume, peak flow, waveform, and oxygen concentration. A spontaneous breath allows the patient inspiratory flows of up to 300 L/min, with or without pressure support. The 740 Ventilator offers three modes of ventilation: Assist/control (A/C), Spontaneous (SPONT), and Synchronous intermittent mandatory ventilation (SIMV). The 740 Ventilator offers two breathing types: Volume-Controlled Ventilation (VCV) and Pressure Support Ventilation (PSV). The 740 Ventilator includes two microcontrollers: 1) the breath delivery microcontroller which controls ventilation and 2) the user interface microcontroller which manages the user interface and monitors ventilator and patient data. This ventilator utilizes a piston-based breath delivery system. A proprietary technique is emploved in which oxygen under a regulated and constant high pressure is pulsed for precisely timed intervals via one of two solenoid valves as the piston is retracting. Piston retraction entrains room air plus the pulsed oxygen into the cylinder where the two gases mix resulting in the desired F.O2. The ventilator's mixing technique allows ventilation without the need for a blender, compressor, or hospital-grade wall air. The operator communicates with the ventilator electronics via membrane keys on the user interface. An optical rotary encoder and accept key (ACCEPT) are used to change and accept settings. An LCD message display window enhances or augments visual display of alarms, settings, modes, menus, conditions, technical alerts and specialized testing procedures.
The provided document is a 510(k) summary for the Puritan-Bennett 740 Ventilator System. It describes the device, its intended use, and claims substantial equivalence to predicate devices. However, the document does not contain a dedicated section outlining specific acceptance criteria or a detailed study designed to prove the device meets such criteria with reported performance metrics, statistical analyses, or specific outcome measures.
Instead, the document states generally that: "The combined testing and analysis of results provides assurance that the device meets it's specifications and is safe and effective for its intended use." It also mentions various testing conducted in accordance with FDA guidance and international standards (e.g., software design and development, environmental and electrical testing, performance testing).
Therefore, I cannot populate the table or answer all the detailed questions as the specific information requested is not present in the provided text.
Here's a breakdown of what can be inferred or stated based on the provided text, and what information is missing:
Acceptance Criteria and Device Performance
| Acceptance Criteria (Inferred from general statements) | Reported Device Performance (General statements) |
|---|---|
| Device meets specifications | "The combined testing and analysis of results provides assurance that the device meets its specifications..." |
| Device is safe and effective for intended use | "...and is safe and effective for its intended use." |
| Compliance with various standards (e.g., EN60601-1:1990, ISO 10651-1) | "The 740 Ventilator device design and testing are also compliant with various voluntary, international standards..." |
| Biocompatibility | "Biocompatibility analysis... demonstrate Puritan-Bennett's 740 Ventilator product to be safe for its intended use." |
| Software design and development per FDA guidance | "Software design and development... was conducted using FDA's Reviewers Guidance of Medical Device Software Submissions. 15 Dec. 1995 draft as a guidance and per internal company requirements." |
| Environmental and electrical testing per FDA guidance | "Environmental and electrical testing was conducted using FDA's Reviewers Guidance for Premarket Notification Submissions, Nov. 1993 draft as a quideline." |
| Performance testing per FDA guidance | "Performance testing was conducted using FDA's Reviewer Guidance for Ventilators draft as a guidance and per internal, company requirements." |
| Substantial Equivalence to predicate devices | "Puritan-Bennett Corp. has provided evidence that shows the 740 Ventilator to be safe and effective. This device is considered to be substantially equivalent to currently marketed devices which have been previously cleared by FDA." |
Study Details (Based on available information)
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Information Not Provided: The document does not specify any sample sizes for a "test set" in the context of clinical or performance data. The testing mentioned appears to be engineering validation (e.g., software, electrical, environmental) rather than a clinical study with a patient test set. Data provenance is also not mentioned.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Information Not Provided: As no specific "test set" or clinical study for establishing ground truth is described, this information is not available.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Information Not Provided: No clinical study or test set requiring adjudication is described.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC study was not done. This device is a continuous ventilator, not an imaging device or diagnostic tool that would typically involve human "readers" or AI assistance in a diagnostic context. The document describes engineering testing and compliance with standards rather than clinical comparative effectiveness studies against an "AI."
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable in the context of this device. The "algorithm" here refers to the software controlling the ventilator. Its performance is inherent to the device's function, not a separate standalone diagnostic algorithm. The device itself is designed to provide continuous ventilation, implying a human-in-the-loop (clinician setting parameters) for its operation. The performance testing ensured the device (including its internal algorithms) met specifications.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Information Not Provided: For the engineering and compliance testing described, "ground truth" would likely relate to established engineering standards, physical measurements, and functional requirements rather than clinical ground truth like pathology or outcomes data. No specific clinical ground truth establishment is mentioned.
-
The sample size for the training set:
- Not applicable in the context of this device as described. The device's operation is based on programmed logic and control systems, not machine learning or AI models that require a "training set" in the modern sense.
-
How the ground truth for the training set was established:
- Not applicable (see point 7).
{0}------------------------------------------------
SUMMARY OF SAFETY AND EFFECTIVENESS
NON 1 4 1997
| SUBMITTER: | Puritan-Bennett Corp. |
|---|---|
| DATE: | November 1, 1996 |
| COMMON NAME: | Continuous Ventilator |
| PROPRIETARY NAME: | 740 Ventilator System (740 Ventilator) |
| CONTACT: | Ann-Marie ButlerRegulatory Affairs Project ManagerPuritan-Bennett Corp.2200 Faraday Ave.Carlsbad, CA 92008(619) 603-5300 (phone)(619) 603-5901 (fax) |
| CLASSIFICATION: | Class II per 21 CFR 868.5895Continuous Ventilator |
PREDICATED DEVICES:
Puritan-Bennett Corp. is claiming substantial equivalence to the following three predicate medical devices:
| Predicate Device | 510(k) Number | Classification |
|---|---|---|
| Puritan Bennett Corp.,7200 Series Ventilator | K902506B | Class II,Continuous Ventilatorper 21 CFR 868.5895 |
| Emerson Company, J.H.,Emerson 3-MV IMV | K864394 | Class IIContinuous Ventilatorper 21 CFR 868.5895 |
1. DEVICE DESCRIPTION:
The device is a low-cost, critical care ventilator intended to provide continuous ventilation for pediatric and adult patients.
The 740 Ventilator supplies mandatory or spontaneous breaths. A mandatory (or assisted) breath provides the patient with a preset tidal volume, peak flow, waveform, and oxygen concentration. A spontaneous breath allows the patient inspiratory flows of up to 300 L/min, with or without pressure support.
{1}------------------------------------------------
The 740 Ventilator offers three modes of ventilation:
- Assist/control (A/C), which consists entirely of mandatory breaths. .
- Spontaneous (SPONT), which consists entirely of spontaneous breaths. .
- Synchronous intermittent mandatory ventilation (SIMV), which can include . both mandatory and spontaneous breaths.
The 740 Ventilator offers two breathing types:
- Volume-Controlled Ventilation (VCV), in which the ventilator delivers a preset . tidal volume at a preset peak flow.
- Pressure Support Ventilation (PSV), in which spontaneous effort is . augmented by a preset PSV value.
The 740 Ventilator includes two microcontrollers: 1) the breath delivery microcontroller which controls ventilation and 2) the user interface microcontroller which manages the user interface and monitors ventilator and patient data. Each microcontroller verifies that the other's instructions are being carried out properly. Using two independent microcontrollers prevents a single fault from causing a simultaneous failure of controlling and monitoring functions.
This ventilator utilizes a piston-based breath delivery system. A proprietary technique is emploved in which oxygen under a regulated and constant high pressure is pulsed for precisely timed intervals via one of two solenoid valves as the piston is retracting. Piston retraction entrains room air plus the pulsed oxygen into the cylinder where the two gases mix resulting in the desired F.O2. The ventilator's mixing technique allows ventilation without the need for a blender, compressor, or hospital-grade wall air.
The operator communicates with the ventilator electronics via membrane keys on the user interface. An optical rotary encoder and accept key (ACCEPT) are used to change and accept settings. An LCD message display window enhances or augments visual display of alarms, settings, modes, menus, conditions, technical alerts and specialized testing procedures. Safety is enhanced through the requirement that all breath related settings be acknowledged prior to successfully changing from one breath type to another.
2. INTENDEDUSE:
Purpose and function of device:
- · The 740 Series Ventilator is intended to provide continuous ventilation to patient's requiring respiratory support.
- · This product is intended for a wide range of patients from pediatric to adult and for a wide variety of clinical conditions.
Intended patient population:
- · The intended patient population includes pediatric and adult patients (tidal volume 0.04 - 2 L) who require continuous respiratory support.
- · Intended for patient who require either invasive or non-invasive ventilation.
{2}------------------------------------------------
Intended environment of use:
- The 740 Ventilator is intended for use in hospitals and hospital type facilities which provide respiratory care for patients requiring respiratory support.
- · The 740 Ventilator may be used during hospital and hospiral type facility transport.
- · The 740 Ventilator is not to be used in the presence of flammable anesthetics.
- The 740 Ventilator is intended for sale by or on the order of a physician only.
- This product is intended for operation by trained and qualified clinicians only and is intended for servicing by trained and qualified persons only.
3. SUBSTANTIALEQUIVALENCE:
The intended use of the 740 Ventilator is the same as that for standard, currently marketed critical care ventilators. The materials and design of this device are similar to those of the predicate devices. A piston/cylinder pneumatic system is also used in the Emerson 3-MV critical care ventilator. The technical characteristics of 740 Ventilator do not introduce new questions regarding safety or effectiveness of critical care ventilators. Furthermore, the labeling associated with the 740 Ventilator provides similar information as the predicate devices.
Information provided in the 510(k) submission supports the determination of substantial equivalence. Biocompatibility analysis and laboratory testing demonstrate Puritan-Bennett's 740 Ventilator product to be safe for its intended use. Software design and development, (including verification and validation testing, test and software quality procedures) was conducted using FDA's Reviewers Guidance of Medical Device Software Submissions. 15 Dec. 1995 draft as a guidance and per internal company requirements. Environmental and electrical testing was conducted using FDA's Reviewers Guidance for Premarket Notification Submissions, Nov. 1993 draft as a quideline. Performance testing was conducted using FDA's Reviewer Guidance for Ventilators draft as a guidance and per internal, company requirements. The 740 Ventilator device design and testing are also compliant with various voluntary, international standards including: EN60601-1:1990, EN 60601-1-2:1993, CAN/CSA C22.2 No. 601-1M90:1994, UL 2601-1:1994, prEN 794-1, ISO 10651-1, and 93/42/EEC MDD.
The combined testing and analysis of results provides assurance that the device meets it's specifications and is safe and effective for its intended use.
In summary Puritan-Bennett Corp. has provided evidence that shows the 740 Ventilator to be safe and effective. This device is considered to be substantially equivalent to currently marketed devices which have been previously cleared by FDA.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUM
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 4 1997
Ms. Ann-Marie Butler Puritan-Bennett Corp. 2200 Faraday Avenue Carlsbad, California 92008
K964540 Re: 740 Ventilator System -----------------Regulatory Class: II (two) Product Code: 73 CBK --Dated: August 15, 1997 Received: August 20, 1997
Dear Ms. Bulter:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Ms. Ann-Marie Butler
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callehan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
INTENDED USE, INDICATION FOR USE 2.
The device intended use is described in sections 2.1 through 2.3 below. The device indication for use is provided separately on the next page.
Intended Use:
-
2.1 Purpose and function of device:
The 740 Ventilator is intended to provide continuous ventilation to patient's requiring respiratory support. TThis product is intended for a wide range of patients from pediatric to adult and for a wide variety of clinical conditions. -
2.2 Intended patient population: -----------
The intended patient population includes pediatric and adult patients (tidal volume 0.04 - 2 L) who require continuous respiratory support. -
2.3 Intended environment of use:
The 740 Ventilator is intended for use in hospitals and hospital type facilities which provide respiratory care for patients requiring respiratory support.
The 740 Ventilator may be used during hospital and hospital type facility transport.
The 740 Ventilator is not to be used in the presence of flammable anesthetics.
The 740 Ventilator is intended for sale by or on the order of a physician only. This product is intended for operation by trained and qualified clinicians only and is intended for servicing by trained and qualified persons only.
A.A. Liddell.
(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number K964540
Puritan-Bennett Corp. 740 Ventilator System 510(K) Submission
{6}------------------------------------------------
INDICATION FOR USE
| 510(k) Number: | |
|---|---|
| Device Name: | 740 Ventilator System |
| Indication for Use: | The 740 Ventilator System is used to provide continuous ventilation to patient's requiring respiratory support. This device is used for a wide range of patients from pediatric to adult and for a wide variety of clinical conditions. |
Perscription Use: Yes (Per 21 CFR 801.109)
Puritan-Bennett Corp.
740 Ventilator System 510(K) Submission
·
ﻴﻌﻮ
22
§ 868.5895 Continuous ventilator.
(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).