Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192713
    Date Cleared
    2020-03-16

    (171 days)

    Product Code
    Regulation Number
    868.5260
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Filter only - For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and/or expired gases is desired. Use up to 24 hours.

    HME / Filter - For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and/or expired gases is desired and to maintain moisture levels in the patient's respiratory tract during anesthesia, artificial respiration and other types of assisted ventilation. Use up to 24 hours.

    Environment of use - clinical setting including hospital, sub-acute, pre-hospital and home

    For patient with Tidal Volumes > 300 ml.

    Device Description

    The subject device is provided in two (2) configurations.

    • Filter only ●
    • . HME / Filter combination
    • The common features are:
      • Standard conical 15 mm / 22 mm fittings for connections
      • Female luer lock port for gas sampling for end-tidal CO2 ●
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "Altera Filter and HME/Filter". It details the device's characteristics, indications for use, and a comparison with a predicate device, including performance data from nonclinical testing.

    Here's the breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Test Method (Attribute)Acceptance CriteriaReported Device Performance
    ISO 9360-1 (Resistance to Flow)Equivalent to a predicate<1.7 cmH2O @ 60 lpm
    ISO 9360-1 (Dead Space)Equivalent to a predicate45 ml for filter only 55 ml for HME
    ISO 9360-1 (Leakage)No leak @ 1 psi for 2 min0.002 ml leak
    ISO 9360-1 (Housing Burst Pressure)No pass/fail criteria (reported value)>76 kPa
    ISO 80369-7 Luer / ISO 5356-1 Conical (Connectors)Passes RequirementsMeets the Standards
    ASTM F2101 (Bacterial Filtration Efficiency - BFE)Equivalent to predicate>99.998%
    ASTM F2101 (Viral Filtration Efficiency - VFE)Equivalent to predicate>99.96%
    ISO 9360-1 (HME Performance - Moisture Output)Moisture Output equivalent to predicate39 mg/L (Vt = 1000 ml @ 24 hours)
    ISO 10993 and ISO 18562 (Cytotoxicity)Non-cytotoxicNon-cytotoxic
    ISO 10993 and ISO 18562 (Sensitization)Non-sensitizerNon-sensitizer
    ISO 10993 and ISO 18562 (Irritation)Non-irritantNon-irritant
    ISO 10993 and ISO 18562 (Chemical Characterization TRA)Margin of Safety > 1Margin of Safety > 1
    ISO 10993 and ISO 18562 (Acute Systemic Toxicity)Non-systemicNon-systemic
    ISO 10993 and ISO 18562 (Volatile Organic Compound)Identified VOCIdentified VOC
    ISO 10993 and ISO 18562 (Particulate Matter)<12 micrograms/ m³PM<12 micrograms/ m³
    ASTM 1980-16 (Accelerated Aging)Device meets performance specification post-conditioningDevice met performance specifications

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document relates to nonclinical testing of a physical medical device. It does not mention clinical trials or studies involving human subjects in the typical sense of a "test set" for AI algorithms. The "tests" refer to laboratory-based performance and biocompatibility assessments of the device itself.

    • Sample Size for Test Set: The sample sizes for each specific test (e.g., number of filters tested for flow resistance, BFE, etc.) are not explicitly stated in the provided text.
    • Data Provenance: The document is a submission to the U.S. Food & Drug Administration (FDA) by Meditera Tibbi Malzeme San. ve Tic. A.S., located in Turkey. The tests are nonclinical, laboratory-based. The text does not specify where the testing laboratories are located, though it implies the data was generated to support a submission in the US. The data is nonclinical/laboratory test data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable to the provided document. The "tests" performed are objective measurements of the device's physical and performance characteristics against established international standards (ISO, ASTM). Ground truth in this context is defined by the requirements of these standards and the measured physical properties, not by expert human interpretation like in medical imaging.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods are typically associated with resolving discrepancies in expert interpretations (e.g., when multiple radiologists review images). For objective nonclinical device performance tests, the results are measured directly against defined criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. The document describes a physical medical device (Altera Filter and HME/Filter) and its nonclinical performance testing. It does not involve any AI components, human readers, or image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a physical breathing circuit filter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this nonclinical device, the "ground truth" for the test results is based on:

    • Established International Standards: ISO (International Organization for Standardization) and ASTM (American Society for Testing and Materials) standards define the test methodologies and acceptable performance ranges.
    • Predicate Device Equivalence: Many acceptance criteria are stated as "Equivalent to a predicate," indicating that the performance of the new device is compared against the known performance of a legally marketed, similar device (Zhejiang Haisheng Medical Device Co. Ltd - K151498).
    • Direct Physical Measurements: The performance criteria (e.g., flow resistance, filtration efficiency, moisture output, burst pressure) are determined through direct, objective physical measurements in a laboratory setting.

    8. The sample size for the training set

    This question is not applicable. The device is a physical medical device, not an AI algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1