Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    K Number
    K193160
    Date Cleared
    2019-12-03

    (18 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Glidesheath Slender Tibial Pedal Kit

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glidesheath Slender Tibial Pedal Kit is indicated to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee.

    Device Description

    Both the predicate Glidesheath Slender Tibial Pedal Kit (7cm needle) and the modified Glidesheath Slender Tibial Pedal Kit (4cm needle) are used to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee. It consists of an introducer (sheath and dilator), which are packaged together with a mini guide wire, an entry needle and a guide inserter.

    During a diagnostic or interventional procedure in a cath lab, the stainless steel entry needle (cannula) is used to gain access to the vein or artery for placement of the mini guide wire. The nitinol mini guide wire is inserted through the cannula into the patient's blood vessel. The wire is used for placement of the sheath and dilator into the vein or artery. The Guide Inserter which is attached to the Mini Guide Wire holder is used to assist the placement of the wire into the needle.

    Following guide wire insertion, the cannula is removed and the sheath and dilator are then inserted over the mini guide wire and into the blood vessel. The mini guide wire is then withdrawn from the vessel. The dilator maintains the integrity of the sheath and dilates the blood vessel during insertion. Once the sheath is situated in the vessel, the dilator is removed and an appropriate catheter can then be inserted through the sheath.

    The sheath incorporates a 1-way valve and a 3-way stopcock connected by a side tube. The sheath is coated with hydrophilic coating to minimize frictional resistance when inserting or removing the sheath from the patient's blood vessel. In addition, the sheath and dilator contain bismuth, making these devices visible under fluoroscopy. The sheath, dilator, entry needle, mini guide wire and guide inserter are provided in a single package and sterilized together.

    The Glidesheath Slender Tibial Pedal Kit is a disposable, ethylene oxide gas sterilized device intended for single use only.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called the Glidesheath Slender Tibial Pedal Kit. This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria through a comprehensive study involving human readers or algorithm-only performance.

    Therefore, many of the requested details such as acceptance criteria, sample sizes for test and training sets, expert qualifications, and adjudication methods are not applicable nor present in this type of submission.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of acceptance criteria or specific performance metrics as typically seen in efficacy studies for AI/imaging devices. Instead, the "performance" section refers to a "Verification by Analysis" that confirms the modified device meets "predetermined criteria (Product Specifications)." These product specifications are general design and material characteristics as detailed in the comparison table on page 6, rather than clinical performance metrics.

    Acceptance Criteria (Inferred from "Product Specifications")Reported Device Performance
    Maintain existing physical characteristics of components beyond the needle length, e.g., Sheath Size, Sheath Length, Hydrophilic Coating, Dilator applicable to Guide Wire OD, Dilator Length, Guide Wire OD, Guide Wire Length."Meets the predetermined criteria (Product Specifications)."
    Maintain existing material compatibility and integrity."All materials for components are identical to the predicate, with the exception of the entry needle length."
    Maintain existing sterilization efficacy."No changes have been made to the sterilization processes, packaging or shelf-life of the device relative to predicate."
    Biological safety (biocompatibility)."No additional biocompatibility testing was completed... No changes to materials or components for the Glidesheath Slender Tibial Pedal Kit, other than the transitory/transient patient contacting shorter needle length."
    Mechanical integrity and functionality."The differences between the predicate and proposed devices do not raise any new issues regarding safety and effectiveness. Therefore, no additional physical testing is required to ensure the safety and effectiveness of the proposed Glidesheath Slender Tibial Pedal kit throughout the shelf life."
    Risk assessment (ISO 14971:2012 compliance)."A Product Risk Analysis was conducted... it was determined that there were no new or increased risks associated with the change."
    Entry Needle LengthThe new device adds a 40mm entry needle length option to the existing 70mm.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable. The submission is a Special 510(k) for a modification (different needle length) to an already cleared device. It relies on a "Verification by Analysis" and comparison to the predicate device, not a new clinical study with a test set of data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as no test set requiring ground truth established by experts was used.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as no test set requiring expert adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a catheter introducer kit, not an AI or imaging diagnostic device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The device is a physical medical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable, as the evaluation was based on engineering verification and comparison to a predicate device, not clinical ground truth.

    8. The sample size for the training set

    This information is not applicable as no training set was used.

    9. How the ground truth for the training set was established

    This information is not applicable as no training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181237
    Date Cleared
    2018-08-03

    (85 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Glidesheath Slender Tibial Pedal Kit

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glidesheath Slender Tibial Pedal Kit is indicated to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee.

    Device Description

    The Glidesheath Slender Tibial Pedal Kit is used to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee. It consists of an introducer (sheath and dilator), which are packaged together with a mini guide wire, an entry needle and a guide inserter.

    During a diagnostic or interventional procedure in a cath lab, the stainless steel entry needle (cannula) is used to gain access to the vein or artery for placement of the mini guide wire. The nitinol mini guide wire is inserted through the cannula into the patient's blood vessel. The wire is used for placement of the sheath and dilator into the vein or artery. The Guide Inserter which is attached to the Mini Guide Wire holder is used to assist the placement of the wire into the needle.

    Following guide wire insertion, the cannula is removed and the sheath and dilator are then inserted over the mini guide wire and into the blood vessel. The mini guide wire is then withdrawn from the vessel. The dilator maintains the integrity of the sheath and dilates the blood vessel during insertion. Once the sheath is situated in the vessel, the dilator is removed and an appropriate catheter can then be inserted through the sheath.

    The sheath incorporates a 1-way valve and a 3-way stopcock connected by a side tube. The sheath is coated with hydrophilic coating to minimize frictional resistance when inserting or removing the sheath from the patient's blood vessel. In addition, the sheath and dilator contain bismuth, making these devices visible under fluoroscopy.

    The sheath, dilator, entry needle, mini guide wire and guide inserter are provided in a single package and sterilized together.

    The Glidesheath Slender Tibial Pedal Kit is a disposable, ethylene oxide gas sterilized device intended for single use only.

    AI/ML Overview

    The Glidesheath Slender Tibial Pedal Kit is a medical device, and the provided document is a 510(k) summary for its premarket notification to the FDA. As such, the concept of "acceptance criteria" and "device performance" in this context refers to demonstrating substantial equivalence to a legally marketed predicate device rather than achieving specific quantitative performance metrics typically seen in AI/software medical devices.

    The "study" that proves the device meets the acceptance criteria is a combination of non-clinical tests (packaging, cadaver testing) and a comparison to existing PRIME Registry data, rather than a clinical trial directly evaluating the device's efficacy against a predefined standard.

    Here's an attempt to extract the requested information, understanding that some categories might not be directly applicable or phrased differently for a medical device clearance like this.


    1. Table of Acceptance Criteria and Reported Device Performance

    Parameter / Acceptance CriteriaReported Device Performance / Evaluation Finding
    Substantial Equivalence (Overall Acceptance Criteria): Demonstrate that the Glidesheath Slender Tibial Pedal Kit is safe and effective and substantially equivalent to legally marketed predicate devices in its intended use/indications for use, technology/principal of operation, materials, and performance.The Glidesheath Slender Tibial Pedal Kit is considered substantially equivalent to the predicate device (K142183 Glidesheath Slender) and reference devices (K152173 Glidesheath, K111606 Pinnacle Precision Access System) based on:
    • Identical materials, formulation, geometry, source, processing, and sterilization method for individual components.
    • Successful packaging verification testing.
    • Demonstrated clinical feasibility in cadaver testing.
    • Favorable comparison to PRIME Registry data. |
      | Packaging Durability: Packaging must ensure the durability of the device throughout distribution. | Packaging verification testing was performed, including Visual Inspection (Post Environmental Conditioning and Distribution Simulation), Detecting Seal Leaks by Dye Penetration (ASTM F1929-15), Seal Strength (ASTM F88/F88M-15), and Climatic Stressing (ASTM F2825-10). The packaging was found to be durable. |
      | Clinical Feasibility (Cadaver Testing): Device compatible with insertion angles and acceptable for use in Below-The-Knee (BTK) access; protocols used by physicians in standard practice are equivalent to the test protocol. | Cadaver testing using lower leg models demonstrated that the Glidesheath Slender sheath is compatible with insertion angles for ultrasound-guided BTK access in four BTK vessels. The kit and components were deemed acceptable for BTK access. A follow-up survey confirmed physician protocols align with the study's access protocol. |
      | Biocompatibility: Meet biological safety requirements for an external communicating device, circulating blood with limited contact (up to 24 hours), per EN ISO 10993-1 and FDA Guidance. | No additional biocompatibility testing was performed as all components are identical to predicate/reference devices in materials, formulation, etc. The device is considered to have substantially equivalent biocompatibility. |
      | Sterilization (SAL 10-6): Achieve a Sterility Assurance Level (SAL) of 10-6 in accordance with ISO 11135:2014. | The device was adopted into an existing ethylene oxide sterilization process validated via the overkill half-cycle approach to achieve an SAL of 10-6. |
      | Sterilant Residuals: Ethylene oxide (EO) and ethylene chlorohydrin (ECH) levels must not exceed average daily doses of 4 mg and 9 mg, respectively, after 24 hours of heated aeration, per EN ISO 10993-7:2008. | Sterilant residual testing demonstrated that EO and ECH levels did not exceed the specified limits after 24 hours of heated aeration. |

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set for Packaging: The document does not specify the exact sample size for each packaging test (Visual Inspection, Seal Leaks, Seal Strength, Climatic Stressing).
    • Test Set for Cadaver Testing: The document mentions "lower leg models" but does not specify the number of cadavers or individual access attempts.
    • Test Set for Clinical Comparison (PRIME Registry): "The first 71 patients treated with the Terumo Glidesheath Slender when undergoing an index endovascular procedure using a tibial access point."
      • Data Provenance: The document states this was compared "to a literature based dataset of procedures using radial and femoral access," implying this data might be a mix of prospective collection (for the 71 patients) and retrospective (for the literature comparison). No country of origin is specified for the PRIME Registry data, but generally, medical device registries can be multinational.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Packaging Testing: No external experts are mentioned for ground truth establishment; testing was done against established ASTM/ISO standards.
    • Cadaver Testing: The study involved "physicians." A "follow-up survey" was conducted to confirm protocols. The number and specific qualifications of these physicians are not detailed beyond being "physicians" involved in BTK access.
    • Clinical Comparison (PRIME Registry): The PRIME Registry data itself is likely based on clinical outcomes and assessments made by healthcare professionals during actual procedures. The document does not specify experts involved in establishing "ground truth" for comparative purposes, rather, it uses existing clinical data.

    4. Adjudication Method for the Test Set

    • Packaging Testing: Adjudication is inherently built into meeting the criteria of the referenced ASTM/ISO standards. No human adjudication is specified beyond standard quality control procedures.
    • Cadaver Testing: The document implies assessment by physicians performing the procedures. No formal adjudication method (e.g., 2+1, 3+1 consensus) for the cadaver test results is described.
    • Clinical Comparison (PRIME Registry): The document does not describe any specific adjudication method for the PRIME Registry data or the literature-based dataset. Registry data typically relies on established clinical reporting.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is a catheter introducer (hardware), not an AI/software device that would typically involve human readers interpreting output with or without AI assistance. The clinical evaluation focused on comparing the device's performance to predicate devices and existing clinical practices using registry data.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    No, this question is not applicable. The device is a physical medical instrument (catheter introducer kit), not an algorithm or software. Its performance is tied to its physical characteristics and how it functions during a medical procedure, not an algorithm's output.


    7. The Type of Ground Truth Used

    • Packaging Testing: Ground truth was based on adherence to established international standards (ASTM and ISO).
    • Cadaver Testing: Ground truth was based on direct observation of the device's performance and compatibility with BTK access in simulated use, assessed by physicians. This can be considered a form of "expert assessment" in a simulated environment.
    • Clinical Comparison (PRIME Registry): Ground truth was established through real-world clinical outcomes and procedural observations recorded in the PRIME Registry and a literature-based dataset. This could be considered a form of "outcomes data" and "clinical assessment."

    8. The Sample Size for the Training Set

    This question is not applicable. This is a hardware medical device, not a machine learning or AI model, thus there is no "training set" in the context of algorithm development.


    9. How the Ground Truth for the Training Set was Established

    This question is not applicable as there is no training set for a hardware device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173831
    Date Cleared
    2018-05-08

    (141 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Glidesheath Slender

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glidesheath Slender is indicated to facilitate placing a catheter through the skin into the radial artery.

    Device Description

    The Glidesheath Slender consists of an introducer (sheath and a dilator), which are packaged together with an entry needle, mini guide wire and guide wire inserter. The Glidesheath Slender devices (all sheath sizes) are used to facilitate placing a catheter through the skin into the radial artery. The sheath and dilator contain bismuth, making these devices visible under fluoroscopy. The sheath is coated with hydrophilic coating to minimize frictional resistance when inserting or removing the sheath from the patient's blood vessel.

    The entry needle (cannula) is used to gain access to the radial artery for placement of the mini guide wire. The entry needle is offered in either the stainless steel (SS) entry needle version or the Surflo (SR) IV catheter (which includes a needle). There are two different types of stainless-steel entry needles that are available in various gauges and lengths. The two needle types are referenced in the documentation as TPC and TRI. These initials are only used in internal Terumo documents.

    The mini guide wire is used for placement of the sheath and dilator into the radial artery. The mini guide wire is offered in three versions made out of three materials, either a stainless-steel spring coil model (stainless steel), a nitinol model with palladium tip (nitinol) or a polyurethane plastic model with a nitinol core (plastic).

    During either a diagnostic or interventional catheterization procedure, a physician will perform the following procedure. The mini guide wire is inserted through a cannula placed in the patient's blood vessel. A guide wire inserter is provided to assist in insertion of the mini guide wire into the cannula. Following guide wire insertion, the cannula is removed, and the sheath and dilator are then inserted over the mini guide wire and into the blood vessel. The mini guide wire is then withdrawn from the vessel. The dilator maintains the integrity of the sheath and dilates the blood vessel during insertion. Once the sheath is situated in the vessel, the dilator is removed, and an appropriate catheter can then be inserted through the sheath.

    The entry needle, the mini guide wire and the guide wire inserter are all accessories to the Glidesheath Slender sheath/dilator. The accessories for a given product code are provided with the Glidesheat Slender sheath/dilator in an individual package and sterilized together.

    AI/ML Overview

    The acceptance criteria and study proving the device meets them are detailed for the Glidesheath Slender. The device is a catheter introducer for radial artery access.

    Here's a breakdown of the information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document provides a comprehensive list of performance tests conducted on various components of the Glidesheath Slender. The "Reported Device Performance" column implicitly states that the "Glidesheath Slender tested met the predetermined acceptance criteria and results support a determination of substantial equivalence." This means that for each test item listed, the device's performance fell within the acceptable parameters defined by the cited standards (ISO, ASTM, in-house) or internal requirements.

    ComponentTest ItemReferenceAcceptance CriteriaReported Device Performance
    SheathFluoroscopyISO 11070:2014, Section 4.5
    ASTM F640-12(Implicit: Device must be visible under fluoroscopy as per standards)Met predetermined acceptance criteria.
    Sheath visual inspectionIn-house standard(Implicit: Visual quality meets internal specifications)Met predetermined acceptance criteria.
    Sheath Effective LengthIn-house standard(Implicit: Length within specified tolerances)Met predetermined acceptance criteria.
    Side Tube LengthIn-house standard(Implicit: Length within specified tolerances)Met predetermined acceptance criteria.
    Sheath Tip IDIn-house standard(Implicit: Inner diameter at tip within specified tolerances)Met predetermined acceptance criteria.
    Sheath, Hemostatic Valve LeakISO 11070:2014, Annex E(Implicit: Hemostatic valve must prevent leakage as per standard)Met predetermined acceptance criteria.
    Sheath Pressure TestISO 11070:2014, Annex D
    ISO 11070:2013, Annex C(Implicit: Withstands specified pressure without failure as per standards)Met predetermined acceptance criteria.
    Sheath PenetrationIn-house standard(Implicit: Achieves appropriate penetration characteristics)Met predetermined acceptance criteria.
    Visual inspection after penetration testingISO 11070:2014 - Annex A section A3(Implicit: No significant damage or alteration after penetration as per standard)Met predetermined acceptance criteria.
    Sheath Tip Rollback TestISO 11070:2014 - Annex A section A3(Implicit: Tip integrity maintained during rollback as per standard)Met predetermined acceptance criteria.
    Dilator Hub to Sheath Hub Snap Fit StrengthIn-house standard(Implicit: Connection strength meets internal specifications)Met predetermined acceptance criteria.
    Valve Mobility ResistanceIn-house standard(Implicit: Valve operates smoothly within specified resistance)Met predetermined acceptance criteria.
    Sheath Kink resistanceISO 11070:2014 Annex A section A.1(Implicit: Resists kinking under specified conditions as per standard)Met predetermined acceptance criteria.
    Sheath Tubing/Housing Joint StrengthISO 11070:2014, Section 7.6(Implicit: Joint strength meets specified requirements as per standard)Met predetermined acceptance criteria.
    Cap to Housing Joint StrengthIn-house standard(Implicit: Connection strength meets internal specifications)Met predetermined acceptance criteria.
    Sheath Support to Housing StrengthIn-house standard(Implicit: Support strength meets internal specifications)Met predetermined acceptance criteria.
    Sheath Tubing Tensile StrengthISO 11070:2014, Section 7.6(Implicit: Tensile strength meets specified requirements as per standard)Met predetermined acceptance criteria.
    Sheath Lubricity and DurabilityIn-house standard(Implicit: Maintains lubricity and durability over intended use)Met predetermined acceptance criteria.
    Particle capture during simulated useFDA PTCA Guidance(Implicit: Particle generation during simulated use is within acceptable limits as per FDA guidance)Met predetermined acceptance criteria.
    Coating integrity after simulated useIn-house standard(Implicit: Hydrophilic coating maintains integrity after simulated use)Met predetermined acceptance criteria.
    DilatorFluoroscopyISO 11070:2014, Section 4.5
    ASTM F640-12(Implicit: Device must be visible under fluoroscopy as per standards)Met predetermined acceptance criteria.
    Dilator visual inspectionIn-house standard(Implicit: Visual quality meets internal specifications)Met predetermined acceptance criteria.
    Dilator Useable LengthIn-house standard(Implicit: Length within specified tolerances)Met predetermined acceptance criteria.
    Dilator Tip IDIn-house standard(Implicit: Inner diameter at tip within specified tolerances)Met predetermined acceptance criteria.
    Dilator OD at Sheath TipIn-house standard(Implicit: Outer diameter at sheath tip within specified tolerances)Met predetermined acceptance criteria.
    Dilator PenetrationIn-house standard(Implicit: Achieves appropriate penetration characteristics)Met predetermined acceptance criteria.
    Visual inspection after penetration testingISO 11070:2014 - Annex A section A3(Implicit: No significant damage or alteration after penetration as per standard)Met predetermined acceptance criteria.
    Dilator Tip Rollback TestISO 11070:2014 - Annex A section A3(Implicit: Tip integrity maintained during rollback as per standard)Met predetermined acceptance criteria.
    Dilator to Hub TensileISO 11070:2014, Annex C
    ISO 11070:2013, Annex B(Implicit: Tensile strength of dilator to hub connection meets specified requirements as per standards)Met predetermined acceptance criteria.
    Dilator Hub to Sheath Hub Snap Fit StrengthIn-house standard(Implicit: Connection strength meets internal specifications)Met predetermined acceptance criteria.
    Particle capture during simulated useFDA PTCA Guidance(Implicit: Particle generation during simulated use is within acceptable limits as per FDA guidance)Met predetermined acceptance criteria.
    BiocompatibilityCytotoxicityEN ISO 10993-1, FDA Guidance(Implicit: No cytotoxic effects detected)Met predetermined acceptance criteria.
    SensitizationEN ISO 10993-1, FDA Guidance(Implicit: No sensitization detected)Met predetermined acceptance criteria.
    Intracutaneous ReactivityEN ISO 10993-1, FDA Guidance(Implicit: No significant intracutaneous reaction)Met predetermined acceptance criteria.
    Systemic Toxicity (Acute)EN ISO 10993-1, FDA Guidance(Implicit: No acute systemic toxicity)Met predetermined acceptance criteria.
    PyrogenicityEN ISO 10993-1, FDA Guidance(Implicit: No pyrogenic response)Met predetermined acceptance criteria.
    HemocompatibilityEN ISO 10993-1, FDA Guidance(Implicit: Compatible with blood, does not cause adverse hematological effects)Met predetermined acceptance criteria.
    SterilizationSterility AssayISO 11135:2014Sterility Assurance Level (SAL) of 10-6Achieved SAL of 10-6.
    Residual Ethylene Oxide (EO)EN ISO 10993-7:2008Not exceeding an average daily dose of 4 mg after 24 hours of heated aerationMet predetermined acceptance criteria.
    Residual Ethylene Chlorohydrin (ECH)EN ISO 10993-7:2008Not exceeding an average daily dose of 9 mg after 24 hours of heated aerationMet predetermined acceptance criteria.
    Shelf LifeShelf life validation(Implied in submission)(Implicit: Device maintains performance and safety for 30 months)Demonstrated 30 months shelf life, with performance maintained throughout.

    2. Sample Size for the Test Set and Data Provenance:

    The document describes non-clinical performance testing, which typically involves testing of device units according to engineering standards rather than a "test set" of patient data. The specific number of units tested for each performance test item is not provided in this summary. The data provenance is internal to the manufacturer (Terumo Medical Corporation) and follows recognized international and in-house standards. This is not patient data, so "country of origin" and "retrospective/prospective" are not applicable in the usual sense for clinical studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    This information is not applicable as the document explicitly states: "This 510(k) does not include data from clinical tests." The performance testing described is engineering and biocompatibility testing, where "ground truth" is established by adherence to specified physical and chemical properties and compliance with recognized standards.

    4. Adjudication Method for the Test Set:

    This information is not applicable as clinical data or expert review of cases is not part of this 510(k) summary.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "This 510(k) does not include data from clinical tests." Therefore, there is no information on how much human readers improve with AI vs. without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable as the device is a physical medical instrument (catheter introducer), not an AI algorithm or software.

    7. The Type of Ground Truth Used:

    For the performance testing, the "ground truth" is established by:

    • International Standards: ISO 11070:2014, ASTM F640-12, ISO 11135:2014, EN ISO 10993-1, EN ISO 10993-7:2008.
    • FDA Guidance: FDA PTCA Guidance, FDA Guidance on Use of International Standard ISO 10993-1.
    • In-house standards: Internal specifications developed by Terumo Medical Corporation.

    These standards and guidances define acceptable physical, mechanical, chemical, and biological properties for device components.

    8. The Sample Size for the Training Set:

    This is not applicable as the device is a physical medical instrument and does not involve a "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set was Established:

    This is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152173
    Device Name
    Glidesheath
    Date Cleared
    2015-12-01

    (119 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Glidesheath

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glidesheath is indicated to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.

    Device Description

    The Glidesheath consists of an introducer (sheath and dilator), which are packaged together with a mini guide wire. Some product configurations also include an entry needle, guide wire inserter and a flushing syringe. The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery. The sheath is coated with hydrophilic coating to minimize frictional resistance when inserting or removing the sheath from the patient's blood vessel. In addition, the sheath and dilator contain bismuth, making these devices visible under fluoroscopy.

    The entry needle (cannula) is used to gain access to the vein or artery for placement of the mini guide wire. The entry needle is offered in either stainless steel (SS) version or the Surflo (SR) IV catheter (which includes a needle) (K891087). There are two different types of stainless steel entry needles that are available in various gauges and lengths.

    The mini guide wire is inserted through the cannula into the patient's blood vessel and is used for placement of the sheath and dilator into the vein or artery. A guide wire inserter is provided to assist in insertion of the mini guide wire into the cannula. The mini guide wire is offered in three versions, either a stainless steel spring coil model (stainless steel), a nitinol model with palladium tipped (nitinol) or a polyurethane plastic model with a nitinol core (plastic).

    Following guide wire insertion, the cannula is removed and the sheath and dilator are then inserted over the mini guide wire and into the blood vessel. The mini guide wire is then withdrawn from the vessel. The dilator maintains the integrity of the sheath and dilates the blood vessel during insertion. Once the sheath is situated in the vessel, the dilator is removed and an appropriate catheter can then be inserted through the sheath. The flushing syringe is used to prime the dilator with heparinized saline.

    The entry needle, the mini guide wire, the guide wire inserter and flushing syringe, are all accessories to the Glidesheath. Depending on the product code, certain accessories may or may not be contained within the kit. All of the accessories for a given product code are provided in an individual package with the Glidesheath and sterilized together.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called Glidesheath™. This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets a specific set of acceptance criteria through a clinical study with primary performance endpoints.

    For medical devices seeking 510(k) clearance, the primary goal is to show that the new device is as safe and effective as a legally marketed predicate device. This is typically done through non-clinical performance testing and comparison of design, materials, and indications for use. Clinical studies (with human subjects) are often not required for 510(k) submissions if non-clinical data is sufficient to demonstrate substantial equivalence.

    Based on the provided text:

    1. A table of acceptance criteria and the reported device performance

      The document provides extensive tables (Table 5.4 to 5.9) detailing the non-clinical performance tests conducted on the proposed Glidesheath™. These tables list the component, test description, and the standard (e.g., ISO, ASTM, or Internal) against which the test was performed. The blanket statement "The components tested met the predetermined acceptance criteria and results support a determination of substantial equivalence" (Page 16) serves as the reported device performance, indicating that all listed tests passed their respective internal or external standard requirements.

      Since this is a 510(k) for substantial equivalence, the "acceptance criteria" are generally the standards set forth by the referenced ISO/ASTM documents or internal specifications, with the performance being that the device passed these tests.

      Example Excerpt from Table 5.4 (Sheath Performance Testing):

      ComponentTest DescriptionStandardReported Device Performance (Implied)
      SheathVisual Inspection – Extraneous

    matter | ISO 11070: 1998, Sect. 4.3 | Met acceptance criteria |
    | Sheath | Liquid leakage through sheath | ISO 11070:1998, Sect. 7.3 & Annex D | Met acceptance criteria |
    | Sheath | External Lubricity | Internal | Met acceptance criteria |

    *(This table would be much longer if all tests were listed, but this demonstrates the format.)*
    

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does **not explicitly state the sample sizes** used for each non-clinical performance test (e.g., number of sheaths, dilators, or needles tested). It refers to the tests being conducted on "the proposed TMC Glidesheath."
    
    The data provenance is from **non-clinical tests** performed by Terumo Medical Corporation (Elkton, MD, USA) as the manufacturing site for the proposed device. The tests are prospective in the sense that they were specifically conducted for this 510(k) submission, testing the new device manufactured at the new site. The country of origin for the data generation would primarily be the USA (Elkton, MD).
    

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is **not applicable** to this 510(k) submission. The document describes non-clinical performance and biocompatibility testing of a physical medical device. "Ground truth" and expert adjudication are concepts typically associated with diagnostic AI/imaging studies where human interpretation is being compared against an AI system. For the physical performance tests listed, the "ground truth" is defined by the objective metrics and standards referenced (e.g., a certain force for tensile strength, or a specific leakage rate). The experts involved would be the engineers and technicians performing and verifying these laboratory tests, following the specified standards.
    

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is **not applicable** as the document describes non-clinical performance testing of a physical medical device, not a diagnostic study requiring adjudication of interpretations.
    

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is **not applicable**. The document does not describe an AI device or a diagnostic imaging study. It pertains to a catheter introducer sheath.
    

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is **not applicable**. The device is a physical catheter introducer sheath, not an algorithm or AI system.
    

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance tests, the "ground truth" is established by adherence to **predefined, objective metrics and standards** such as:
    *   ISO 11070: 1998 (Sterile single-use intravascular introducers, dilators and accessories)
    *   ASTM F640-12 (Standard Practice for Radiopacity of Medical Photographic Films)
    *   ISO 594-1:1986, ISO 594-2: 1998 (Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment)
    *   ISO 7886-1: 1993 (Sterile hypodermic syringes for single use – Part 1: Syringes for manual use)
    *   ISO 10993 series (Biological evaluation of medical devices)
    
    These standards define what constitutes acceptable performance for various physical and biological properties of the device components. Visual inspections would also follow internal quality standards based on these principles.
    

    8. The sample size for the training set

    This is **not applicable**. The Glidesheath™ is a physical medical device, not an AI or machine learning model that requires a "training set."
    

    9. How the ground truth for the training set was established

    This is **not applicable** for the same reason as point 8.
    
    Ask a Question

    Ask a specific question about this device

    K Number
    K142183
    Date Cleared
    2014-11-21

    (105 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    GLIDESHEATH SLENDER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glidesheath Slender is used to facilitate placing a catheter through the skin into the radial artery.

    Device Description

    Both the predicate Glidesheath Slender 6Fr (GSS 6Fr) and the modified Glidesheath Slender 5Fr & 7Fr (GSS 5Fr & 7Fr) consist of an introducer sheath and a dilator which are packaged together with an entry needle, mini guide wire and guide wire inserter. The GSS devices (all sheath sizes) are used to facilitate placing a catheter through the skin into the radial artery. The sheath and dilator contain bismuth, making these devices visible under fluoroscopy. The sheath is coated with hydrophilic coating to minimize frictional resistance when inserting or removing the sheath from the patient's blood vessel.

    The entry needle (cannula) is used to gain access to the radial artery for placement of the mini guide wire. The entry needle is offered in two versions, either a stainless steel entry needle or a Surflo IV catheter (K891087).

    The mini guide wire is used for placement of the sheath and dilator into the radial artery. The mini guide wire is offered in two versions, either a stainless steel (spring coil) model or a polyurethane (nitinol core) plastic model.

    The mini guide wire is inserted through a cannula placed in the patient's blood vessel. A guide wire inserter is provided to assist in insertion of the mini guide wire into the cannula. Following guide wire insertion, the cannula is removed and the sheath and dilator are then inserted over the mini guide wire and into the blood vessel. The mini guide wire is then withdrawn from the vessel. The dilator maintains the integrity of the sheath and dilates the blood vessel during insertion. Once the sheath is situated in the vessel, the dilator is removed and an appropriate catheter can then be inserted through the sheath.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Glidesheath Slender (GSS) device, specifically for the 5Fr and 7Fr sizes. The submission claims substantial equivalence to the predicate GSS (6Fr) device (K122980). Therefore, the study is a non-clinical performance study comparing the modified device to the predicate.

    Here's an analysis of the provided information concerning acceptance criteria and the study:

    1. Table of Acceptance Criteria and the Reported Device Performance:

    The document primarily relies on adherence to established international and internal standards as acceptance criteria. The device performance is generally stated as "met the requirements" or "results were within the predetermined acceptance criteria." Quantitative data for specific performance metrics are not provided in this summary.

    Acceptance Criteria (Standard/Test)Reported Device Performance
    Sheath:
    Surface (ISO 11070: Sec. 4.3)Met requirements of applicable standards
    Corrosion Resistance (ISO 11070: Sec. 4.4)Met requirements of applicable standards
    Radiodetectability (ISO 11070: Sec.4.5)Met requirements of applicable standards
    Size Designation (ISO 11070: Sec. 7.2)Met requirements of applicable standards
    Freedom from Leakage from Sheath Introducer (ISO 11070: Sec. 7.3)Met requirements of applicable standards
    Freedom from Leakage through Haemostasis Valve (ISO 11070: Sec. 7.4)Met requirements of applicable standards
    Force at Break (sheath) (ISO 11070: Sec. 7.6)Met requirements of applicable standards
    Force at Break (sheath to hub) (ISO 11070: Sec. 7.6)Met requirements of applicable standards
    Sheath to Dilator Fit (ISO 11070: Annex A.1)Met requirements of applicable standards
    Rollback Test (ISO 11070: Annex A.1)Met requirements of applicable standards
    Puncture model test (ISO 11070: Annex A.1)Met requirements of applicable standards
    Flexibility (Kink Angle) (ISO 11070: Annex A.1)Met requirements of applicable standards
    Flexibility (Radius of Curvature) (ISO 11070: Annex A.1)Met requirements of applicable standards
    Catheter Insertion and Removal Resistance (Internal Standard)Met predetermined acceptance criteria
    Penetration Resistance (Internal Standard)Met predetermined acceptance criteria
    External Surface Sliding Performance (Internal Standard)Met predetermined acceptance criteria
    Hydrophilic coating Separation Resistance (Internal Standard)Met predetermined acceptance criteria
    Hydrophilic coating Particulate Evaluation (FDA Guidance/USP 788)Met predetermined acceptance criteria
    Dilator:
    Surface (ISO 11070: Sec. 4.3)Met requirements of applicable standards
    Size Designation (ISO 11070: Sec. 9.2)Met requirements of applicable standards
    Conical Fitting (ISO 11070: Sec 9.3.2/ISO594-1)Met requirements of applicable standards
    Strength of Union between Hub and Dilator (ISO 11070: Sec 9.3.3)Met requirements of applicable standards
    Biocompatibility:
    Cytotoxicity (ISO 10993-5: 2009)Met requirements of applicable standards
    Sensitization (ISO 10993-10: 2010)Met requirements of applicable standards
    Irritation/Intracutaneous Reactivity (ISO 10993-10: 2010)Met requirements of applicable standards
    Acute Systemic Toxicity (ISO 10993-11: 2006)Met requirements of applicable standards
    Pyrogenicity (ISO 10993-11: 2006)Met requirements of applicable standards
    Hemolysis (ASTM F 756: 2008)Met requirements of applicable standards
    Complement Activity (ISO 10993-4: 2002/A1:2006)Met requirements of applicable standards
    Thrombosis (ISO 10993-4: 2002/A1:2006)Met requirements of applicable standards
    Physiochemical (USP )Physiochemical properties do not change throughout shelf life
    FT-IR (USP )Physiochemical properties do not change throughout shelf life
    Sterilization:
    Sterility Assurance Level (ANSI/AAMI/ISO 11135-1)SAL of 10^-6 achieved
    Residual Ethylene Oxide (ISO 10993-7)Levels met requirements for limited exposure devices ( Bacterial Endotoxins Test)

    2. Sample Size Used for the Test Set and the Data Provenance:

    • Sample Size: The document does not explicitly state the sample sizes used for each specific non-clinical test.
    • Data Provenance: The tests are "non-clinical tests" and were conducted to "ensure safety and effectiveness of the modified GSS (5Fr & 7Fr) device throughout the shelf-life, verify conformity to applicable ISO and internal standards and acceptance criteria, and demonstrate substantial equivalence to the predicate device." The manufacturer is Terumo Medical Corporation, with production facilities in Japan (Ashitaka Factory of Terumo Corporation). This indicates the data is likely from internal testing by the manufacturer, not from a patient population. It is retrospective in the sense that it evaluates the manufactured device, not prospective in a clinical trial sense. The "country of origin of the data" would be Japan, where the manufacturing and potentially some testing occurs, and potentially the US, where Terumo Medical Corporation is located for regulatory affairs.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    This is a non-clinical, performance testing study focused on engineering and material properties. There is no human "ground truth" establishment in the traditional sense, as these tests are based on objective physical standards (e.g., ISO, ASTM, USP, internal engineering standards). Therefore, this question is not applicable to the provided study description.

    4. Adjudication Method for the Test Set:

    Not applicable, as this is objective non-clinical performance testing against defined standards, not subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a non-clinical performance study of a physical medical device (an introducer sheath), not an AI algorithm or an imaging device involving human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical medical device.

    7. The type of ground truth used:

    The "ground truth" in this context refers to the defined specifications and requirements set forth by international standards (ISO, ASTM, USP) and the manufacturer's own internal standards. These are objective engineering and material science metrics. It is not an expert consensus, pathology, or outcomes data in the biological sense.

    8. The sample size for the training set:

    Not applicable. This is not a machine learning or artificial intelligence study, so there is no "training set."

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set in this context.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    GLIDESHEATH SLENDER MODEL RM* ES6J10HQS, GLIDESHEATH SLENDER MODEL RM* ES6F16HQ, GLIDESHEATH SLENDER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glidesheath Slender is used to facilitate placing a catheter through the skin into the radial artery.

    Device Description

    The Glidesheath Slender consists of an introducer sheath and a dilator which are packaged together with an entry needle and mini guide wire. The Glidesheath Slender is used to facilitate placing a catheter through the skin into the radial artery. The sheath and dilator contain bismuth. making these devices visible under fluoroscopy. The sheath is coated with hydrophilic coating to minimize frictional resistance when inserting or removing the sheath from the patient's blood vessel.

    The entry needle (cannula) is used to gain access to the radial artery for placement of the mini guide wire. The entry needle is offered in two versions, either a stainless steel needle or a Surflo IV catheter (K891087).

    The mini guide wire is used for placement of the sheath and dilator into the radial artery. The mini guide wire is offered in two versions, either a stainless steel (spring coil) model or a polyurethane (nitinol core) plastic model.

    A guide wire inserter is also provided to assist in insertion of the mini guide wire into the cannula.

    AI/ML Overview

    The provided 510(k) summary describes the Glidesheath Slender, a catheter introducer. This document focuses on demonstrating substantial equivalence to a predicate device (Terumo Glidesheath K082644) through performance testing, material comparison, and biocompatibility evaluations, rather than reporting on a study measuring a specific clinical outcome with defined acceptance criteria for device performance.

    Therefore, many of the requested sections related to acceptance criteria, device performance, expert-established ground truth, MRMC studies, standalone performance, and training/test set details are not applicable as this is a submission for a medical device that relies on bench testing and material equivalency, not an AI/software as a medical device (SaMD).

    Here's an analysis based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria or quantifiable device performance results for clinical outcomes. Instead, it lists various performance tests conducted according to ISO standards and internal standards to verify substantial equivalence to the predicate device. The implied acceptance criterion for each test is that the Glidesheath Slender performs adequately and comparably to the predicate device, thereby ensuring safety and effectiveness.

    Table of Performance Tests and General Outcome (as inferred from the document):

    Performance Testing (Sheath)MethodImplied Acceptance Criteria / Outcome
    SurfaceISO 11070 Sec. 4.3Conforms to visual/tactile inspection; comparable to predicate.
    Corrosion resistanceISO 11070 Sec. 4.4Resistant to corrosion; comparable to predicate.
    RadiodetectabilityISO 11070 Sec. 4.5Visible under fluoroscopy; comparable to predicate.
    Dimensional verificationISO 11070 Sec. 7.2Dimensions within specified tolerances; comparable to predicate.
    Freedom from leakage (sheath)ISO 11070 Sec. 7.3No leakage; comparable to predicate.
    Freedom from leakage (hemostasis valve)ISO 11070 Sec. 7.4No leakage; comparable to predicate.
    Force at break (sheath)ISO 11070 Sec. 7.6Meets strength requirements; comparable to predicate.
    Force at break (sheath to hub)ISO 11070 Sec. 7.6Meets strength requirements; comparable to predicate.
    Sheath to dilator fitISO 11070 Sec. A.1Proper fit; comparable to predicate.
    Rollback testISO 11070 Sec. A.1Demonstrates adequate coating adhesion/stability; comparable to predicate.
    Puncture model testISO 11070 Sec. A.1Demonstrates appropriate performance upon insertion; comparable to predicate.
    Flexibility (kink angle)ISO 11070 Sec. A.1Meets flexibility requirements; comparable to predicate.
    Flexibility (radius of curvature)ISO 11070 Sec. A.1Meets flexibility requirements; comparable to predicate.
    Catheter insertion resistanceInternal StandardLow resistance comparable to predicate.
    Penetration resistanceInternal StandardAppropriate resistance; comparable to predicate.
    External surface sliding performanceInternal StandardSmooth sliding; comparable to predicate.
    Hydrophilic coating separation resistanceInternal StandardCoating remains intact; comparable to predicate.
    Hydrophilic coating particulate evaluationFDA Guidance, USP788Meets particulate limits; comparable to predicate.
    Performance Testing (Dilator)
    SurfaceISO 11070 Sec. 4.3Conforms to visual/tactile inspection; comparable to predicate.
    Corrosion resistanceISO 11070 Sec. 4.4Resistant to corrosion; comparable to predicate.
    Dimensional verificationISO 11070 Sec. 9.2Dimensions within specified tolerances; comparable to predicate.
    Conical fittingISO 11070 Sec. 9.3.2, ISO 594-1Proper conical fit; comparable to predicate.
    Strength of union (dilator to hub)ISO 11070 Sec. 9.3.3Meets strength requirements; comparable to predicate.
    Performance Testing (Stainless Steel Needle)
    SurfaceISO 11070 Sec. 4.3Conforms to visual/tactile inspection; comparable to predicate.
    Corrosion resistanceISO 11070 Sec. 4.4Resistant to corrosion; comparable to predicate.
    RadiodetectabilityISO 11070 Sec. 4.5Visible under fluoroscopy; comparable to predicate.
    Dimensional verificationISO 11070 Sec. 5.2Dimensions within specified tolerances; comparable to predicate.
    Needle pointISO 11070 Sec. 5.3Sharp and properly formed; comparable to predicate.
    Conical fittingISO 11070 Sec. 5.4.1, ISO 594-1Proper conical fit; comparable to predicate.
    Strength of union (needle tube to hub)ISO 11070 Sec. 5.4.2Meets strength requirements; comparable to predicate.
    Performance Testing (Surflo IV Needle)
    SurfaceISO 11070 Sec. 4.3Conforms to visual/tactile inspection; comparable to predicate.
    Corrosion resistanceISO 11070 Sec. 4.4Resistant to corrosion; comparable to predicate.
    RadiodetectabilityISO 11070 Sec. 4.5Visible under fluoroscopy; comparable to predicate.
    Dimensional verificationISO 11070 Sec. 5.2Dimensions within specified tolerances; comparable to predicate.
    Needle pointISO 11070 Sec. 5.3Sharp and properly formed; comparable to predicate.
    Conical fittingISO 11070 Sec. 5.4.1, ISO 594-1Proper conical fit; comparable to predicate.
    Strength of union (needle tube to hub)ISO 11070 Sec. 5.4.2Meets strength requirements; comparable to predicate.
    Performance Testing (Surflo IV Catheter)
    SurfaceISO 11070 Sec. 4.3Conforms to visual/tactile inspection; comparable to predicate.
    Corrosion resistanceISO 11070 Sec. 4.4Resistant to corrosion; comparable to predicate.
    RadiodetectabilityISO 11070 Sec. 4.5Visible under fluoroscopy; comparable to predicate.
    Catheter to needle fitISO 11070 Sec. 6.2Proper fit; comparable to predicate.
    Strength of union (catheter to hub)ISO 11070 Sec. 6.3Meets strength requirements; comparable to predicate.
    Conical fittingISO 11070 Sec. 6.4, ISO 594-1Proper conical fit; comparable to predicate.
    Dimensional verificationISO 11070 Sec. 6.5Dimensions within specified tolerances; comparable to predicate.
    Performance Testing (Plastic Guide Wire)
    SurfaceISO 11070 Sec. 4.3Conforms to visual/tactile inspection; comparable to predicate.
    RadiodetectabilityISO 11070 Sec. 4.5Visible under fluoroscopy; comparable to predicate.
    Dimensional verificationISO 11070 Sec. 8.2Dimensions within specified tolerances; comparable to predicate.
    Test for fracture of guide wireISO 11070 Sec. 8.4Resistant to fracture; comparable to predicate.
    Resistance to damage by flexingISO 11070 Sec. 8.5Resistant to damage from flexing; comparable to predicate.
    Performance Testing (Stainless Steel Guide Wire)
    SurfaceISO 11070 Sec. 4.3Conforms to visual/tactile inspection; comparable to predicate.
    Corrosion resistanceISO 11070 Sec. 4.4Resistant to corrosion; comparable to predicate.
    RadiodetectabilityISO 11070 Sec. 4.5Visible under fluoroscopy; comparable to predicate.
    Dimensional verificationISO 11070 Sec. 8.2Dimensions within specified tolerances; comparable to predicate.
    Test for fracture of guide wireISO 11070 Sec. 8.4Resistant to fracture; comparable to predicate.
    Resistance to damage by flexingISO 11070 Sec. 8.5Resistant to damage from flexing; comparable to predicate.
    Strength of union of safety wire and coilISO 11070 Sec. 8.6Meets strength requirements; comparable to predicate.
    Strength of union of core wire and coilISO 11070 Sec. 8.7Meets strength requirements; comparable to predicate.
    BiocompatibilityISO 10993-1:2009Classified as Externally Communicating Devices, Circulating Blood, Limited Contact ( Bacterial Endotoxins Test, FDA Guideline (1987)
    TiO2 Particle ReleaseISO10993-12Potential release much lower than tolerable intake (NOAEL); no safety concern.

    The "reported device performance" is implicitly that the device successfully met all these technical and biological safety requirements, thereby demonstrating substantial equivalence to the predicate device.

    2. Sample size(s) used for the test set and the data provenance

    The document refers to "bench tests" and "each lot of production" for LAL tests. It does not specify a distinct "test set" in the context of patient data. The tests are focused on component and assembly performance, material properties, and sterilization, typically involving a representative number of units or material samples from manufacturing lots. No patient data or clinical data is mentioned for these performance tests.

    • Test Set Sample Size: Not specified in the context of patient data; applies to manufactured units/samples for bench testing.
    • Data Provenance: Not applicable as these are bench tests and laboratory evaluations, not data from a specific country of origin or retrospective/prospective patient studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device's evaluation relies on established engineering standards, biocompatibility protocols, and manufacturing quality controls, not an expert consensus on a "ground truth" derived from patient cases.

    4. Adjudication method for the test set

    Not applicable. No expert adjudication process is described for these types of performance tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/SaMD device, and therefore, an MRMC study is not relevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an AI/SaMD device.

    7. The type of ground truth used

    The "ground truth" for this device's evaluation is defined by established engineering specifications, material composition standards, ISO/FDA recognized test methods, and biocompatibility requirements. For instance, the ground truth for "Dimensional verification" is the specified blueprint dimensions, and for "Corrosion resistance," it's the absence of corrosion under specified conditions.

    8. The sample size for the training set

    Not applicable. This is not an AI/SaMD device; there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. There is no "training set" or corresponding ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102008
    Device Name
    GLIDESHEATH
    Date Cleared
    2010-07-21

    (5 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    GLIDESHEATH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.

    The Entry Needle is an accessory device which is used to gain access to the vein or artery including but not limited to the radial artery, for placement of the Mini Guide Wire.

    The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery including but not limited to the radial artery.

    Device Description

    The Glidesheath is comprised of an introducer sheath and a dilator. The Glidesheath is coated with a hydrophilic coating to reduce the frictional resistance of the sheath when inserting or removing the sheath from the patient's blood vessel. The Sheath and Dilator contain bismuth, making these devices visible under fluoroscopy. The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.

    The Entry Needle is an accessory device which is used to gain access to the vein or artery including but not limited to the radial artery, for placement of the Mini Guide Wire.

    The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery including but not limited to the radial artery. The Guide Inserter which is attached to the Mini Guide Wire holder is used to straighten out the wire.

    Accessories to the Glidesheath are the metal entry needle and the mini guide wire. Both the metal entry needle and the mini guide wire are packaged with the Glidesheath in a pouch prior to sterilization.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Glidesheath introducer sheath, comparing it to a predicate device (K082644) to demonstrate substantial equivalence. The information focuses on the device's design, materials, specifications, and performance testing, which is largely bench testing and biocompatibility studies.

    Here's an analysis of the requested information based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" in a quantitative, pass/fail threshold manner for each performance test. Instead, it lists the performance tests that were successfully passed. The implication is that passing these tests constitutes meeting the acceptance criteria for a medical device of this type.

    Table of Acceptance Criteria and Reported Device Performance:

    Performance Test CategorySpecific TestReported Device Performance
    NeedleNeedle surface free from defectsSuccessfully passed (implied: surface was free from defects)
    Needle ODSuccessfully passed (implied: met specified OD)
    Needle lengthSuccessfully passed (implied: met specified length)
    Needle IDSuccessfully passed (implied: met specified ID)
    Needle hub conical entry angleSuccessfully passed (implied: met specified angle)
    Bevel indicator visibilitySuccessfully passed (implied: indicator was visible)
    Bevel indicator positionSuccessfully passed (implied: met specified position)
    Needle to hub joint strengthSuccessfully passed (implied: met strength requirements)
    Gauge Luer taperSuccessfully passed (implied: conformed to Luer taper standards)
    Liquid leakage from fitting assembly under pressureSuccessfully passed (implied: no liquid leakage)
    Air leakage into the fitting assembly during aspirationSuccessfully passed (implied: no air leakage)
    Separation force of fitting assemblySuccessfully passed (implied: met separation force requirements)
    Unscrewing torque of fitting assemblySuccessfully passed (implied: met unscrewing torque requirements)
    Ease of assemblySuccessfully passed (implied: easy to assemble)
    Resistance to overridingSuccessfully passed (implied: resisted overriding)
    Stress crackingSuccessfully passed (implied: no stress cracking)
    Corrosion resistanceSuccessfully passed (implied: resistant to corrosion)
    Guide WireGuidewire surface free from defectsSuccessfully passed (implied: surface was free from defects)
    Tip buckling testSuccessfully passed (implied: tip resisted buckling)
    Test for resistance of guidewires to damage by flexingSuccessfully passed (implied: guidewire resisted damage from flexing)
    Test for fracture of guidewiresSuccessfully passed (implied: no fracture)
    Test for distal tip retentionSuccessfully passed (implied: distal tip was retained)
    Guidewire ODSuccessfully passed (implied: met specified OD)
    Guidewire lengthSuccessfully passed (implied: met specified length)
    Test for corrosion resistanceSuccessfully passed (implied: resistant to corrosion)
    DilatorDilator surface free from defectsSuccessfully passed (implied: surface was free from defects)
    Dilator tip IDSuccessfully passed (implied: met specified tip ID)
    Dilator to hub joint strengthSuccessfully passed (implied: met strength requirements)
    SheathSheath surface free from defectsSuccessfully passed (implied: surface was free from defects)
    Sheath tip IDSuccessfully passed (implied: met specified tip ID)
    Sheath to housing joint strengthSuccessfully passed (implied: met strength requirements)
    Housing to cap joint strengthSuccessfully passed (implied: met strength requirements)
    InserterGuidewire inserter surface free from defectsSuccessfully passed (implied: surface was free from defects)
    SystemSystem use in modelSuccessfully passed (implied: system functioned correctly in a model)
    BiocompatibilityPhysicochemical ProfileSuccessfully passed (implied: met profile requirements)
    CytotoxicitySuccessfully passed (implied: no cytotoxic effects)
    SensitizationSuccessfully passed (implied: no sensitization)
    Acute Intracutaneous ReactivitySuccessfully passed (implied: no intracutaneous reactivity)
    Acute Systemic ToxicitySuccessfully passed (implied: no systemic toxicity)
    HemolysisSuccessfully passed (implied: no hemolysis)
    ThromobogenicitySuccessfully passed (implied: not thrombogenic)
    Complement Activation AssaySuccessfully passed (implied: no complement activation)
    Unactivated Partial Thromboplastin Time AssaySuccessfully passed (implied: appropriate coagulation response)
    In Vitro HemolysisSuccessfully passed (implied: no in vitro hemolysis)
    GenotoxicitySuccessfully passed (implied: no genotoxic effects)
    Pyrogen StudySuccessfully passed (implied: non-pyrogenic)
    Extractable Metals and Acidity/AlkalinitySuccessfully passed (implied: met requirements for extractables)
    SterilizationANSI / AAMI / ISO 11135 validationSuccessfully validated to SAL of 10^-6

    Study Details for Acceptance Criteria Proof:

    1. Sample size used for the test set and data provenance:

      • The document does not specify sample sizes for individual performance tests (e.g., how many needles were tested for joint strength).
      • It also does not indicate the data provenance (e.g., country of origin, retrospective or prospective). Given the nature of bench testing, it's typically performed in a laboratory setting by the manufacturer.
    2. Number of experts used to establish the ground truth for the test set and their qualifications:

      • This information is not provided. For mechanical and biocompatibility testing, ground truth is typically established by adhering to recognized international standards and internal quality control specifications, rather than relying on expert consensus in the way clinical studies do. The "experts" would be the engineers and scientists performing and interpreting the standardized tests.
    3. Adjudication method for the test set:

      • This information is not applicable or not provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving subjective interpretations (e.g., image reading). The listed performance tests are primarily objective, quantitative, or qualitative assessments against defined standards.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size:

      • No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for diagnostic imaging devices or devices where human interpretation is a key component, often comparing human performance with and without AI assistance. This document describes a physical medical device (introducer sheath) that does not involve diagnostic interpretation or AI.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No, a standalone algorithm performance study was not done. This device is a physical mechanical instrument and does not incorporate an AI algorithm.
    6. The type of ground truth used:

      • The ground truth for most performance tests (Needle, Guide Wire, Dilator, Sheath, Inserter, System) would be based on engineering specifications, material science standards, and functional requirements for similar medical devices.
      • For Biocompatibility, the ground truth is established by international biological evaluation standards (ISO 10993), which define acceptable thresholds for biological responses (e.g., absence of cytotoxicity, genotoxicity, etc.).
      • For Sterilization, the ground truth is based on sterilization validation standards (ANSI/AAMI/ISO 11135) to achieve a defined Sterility Assurance Level (SAL) of 10^-6.
    7. The sample size for the training set:

      • This is not applicable. The Glidesheath is a physical medical device, not an AI or machine learning model that requires a "training set."
    8. How the ground truth for the training set was established:

      • This is not applicable for the same reason as point 7.
    Ask a Question

    Ask a specific question about this device

    K Number
    K082644
    Device Name
    GLIDESHEATH
    Date Cleared
    2008-09-18

    (7 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    GLIDESHEATH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.

    The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery including but not limited to the radial artery.

    The Radifocus Obturator is also an accessory device which is used by placing it into the sheath to create an occlusion and further provide support to the wall of the indwelling sheath while it remains in place within the vein or artery, including but not limited to the radial artery, after removal of a catheter.

    Device Description

    The Glidesheath is comprised of an introducer sheath and a dilator. The Glidesheath is coated with a hydrophilic coating to reduce the frictional resistance of the sheath when inserting or removing the sheath from the patient's blood vessel. The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.

    The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery including but not limited to the radial artery. The Mini Guide Wire is inserted through a cannula placed in the patient's blood vessel. The Glidesheath is then inserted over the Mini Guide Wire and into the blood vessel. The Mini Guide Wire is then withdrawn from the vessel. The Dilator maintains the integrity of the Sheath and dilates the blood vessel while the Glidesheath is being placed into the vessel. The Dilator can be removed and an appropriate catheter can then be inserted.

    The Obturator is also an accessory device which is used by placing it into the sheath to create an occlusion and further provide support to the wall of the indwelling sheath while it remains in place within the vein or artery, including but not limited to the radial artery, after removal of a catheter. The Obturator is sold separately.

    The Sheath, Dilator and Obturator contain bismuth, making these devices visible under fluoroscopy.

    Accessories to the Glidesheath are the mini guide wire, Surflo IV catheter, Syringe, and Scalpel, and the RADIFOCUS Obturator. Depending on the product code, these accessories may or may not be contained within the kit. All of the accessories are packaged with the Glidesheath in a pouch prior to sterilization.

    AI/ML Overview

    The provided text is a 510(k) summary for the Glidesheath, a medical device. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a dedicated study with performance metrics in the way, for example, an AI/ML algorithm would be evaluated for accuracy.

    Therefore, many of the requested categories (e.g., sample size for test/training set, number of experts, adjudication method, MRMC study, standalone performance, type of ground truth) are not applicable to this type of device submission based on the provided information.

    Here's an analysis based on the information provided, highlighting what is included and what is not:


    Acceptance Criteria and Device Performance Summary

    Since this is a 510(k) submission for substantial equivalence for a medical device (introducer sheath, dilator, guide wire, obturator), the "acceptance criteria" and "device performance" are primarily focused on meeting regulatory requirements for safety and effectiveness through comparison to already marketed predicate devices, rather than quantitative performance metrics like sensitivity, specificity, or accuracy for a diagnostic or AI device.

    The "study" that proves the device meets the "acceptance criteria" is essentially the comparison to predicate devices and adherence to relevant standards/tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    CategoryAcceptance Criteria (Implied from 510(k) goals)Reported Device Performance (as demonstrated by comparison/testing)
    Intended UseSubstantially equivalent to predicate devices (Glidesheath K062858, Cordis Avanti K962746)Used to facilitate catheter placement into veins/arteries, including radial artery. Mini Guide Wire and Obturator as accessories.
    Design / MaterialsNo new issues of safety and effectiveness compared to predicate devicesDifferences in materials between modified and predicate device raise no new safety/effectiveness issues. Includes hydrophilic coating.
    Biological Safety (Biocompatibility)Biocompatible for prolonged contact (24 hrs to 30 days) with circulating blood.Blood contacting materials tested per ISO 10993-1, results demonstrate biocompatibility.
    SterilizationSterilized to a Sterility Assurance Level (SAL) of 10^-6.Validated according to EN ISO 11135-1, device sterilized to SAL of 10^-6.
    Risk ManagementConducted according to EN ISO 14971.A risk/hazard analysis was conducted according to EN ISO 14971.
    Manufacturing ControlsIncludes visual, functional, dimensional, and sterility tests.Manufacturing controls include visual, functional, dimensional and sterility tests.
    Overall Safety and Effectiveness (SE)Substantially equivalent to predicate devices."Differences between the devices do not raise any significant issues of safety or effectiveness."
    Physical SpecificationsMeet specified ranges for sheath sizes, lengths, guide wire OD, and Surflo IV catheter.Ranges: Sheath 4, 5, 6 Fr; Sheath Length 10-25 cm; Dilator Length 15.5-30.5 cm; Guide Wire OD 0.021"-0.038", 10-180cm; Surflo IV Catheter 16G-22G, 1"-2.5" length.

    2. Sample Size used for the test set and the data provenance

    • Not Applicable. This is a 510(k) submission for a physical medical device, not a data-driven device (like an AI/ML algorithm) that typically relies on a "test set" in the conventional sense of machine learning. The "testing" involves biocompatibility tests, sterilization validations, and performance characteristic determinations based on engineering principles and comparisons to predicate devices, not data sets of patient information.
    • The provenance of data is not specified in terms of country of origin or retrospective/prospective, as typical for clinical data studies. The testing is laboratory and bench-top oriented.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. Ground truth in the context of diagnostic performance metrics is not established for this type of device. The verification and validation activities are based on engineering standards, material science, and regulatory compliance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. Adjudication methods are relevant for subjective evaluations, typically in clinical studies or expert consensus for ground truth. This is not the type of evaluation described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. An MRMC study is relevant for diagnostic imaging devices, particularly with AI assistance. This device is an introducer sheath for facilitating catheter placement, not an imaging or diagnostic interpretation device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an algorithm-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable in the sense of clinical ground truth for diagnostic accuracy. For physical device testing, "ground truth" equates to established engineering specifications, validated test methods (e.g., ISO standards for biocompatibility and sterilization), and objective measurements.

    8. The sample size for the training set

    • Not Applicable. This device does not involve a "training set" like an AI/ML algorithm.

    9. How the ground truth for the training set was established

    • Not Applicable. This device does not involve a "training set" or "ground truth" in the context of data-driven learning.
    Ask a Question

    Ask a specific question about this device

    K Number
    K062858
    Date Cleared
    2006-10-20

    (25 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    MODIFICATION TO GLIDESHEATH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery. The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery. The RADIFOCUS Obturator is also an accessory device which is used by placing it into the sheath to create an occlusion and further provide support to the wall of the indwelling sheath while it remains in place within the vein or artery after removal of a catheter.

    Device Description

    The Glidesheath is comprised of an introducer sheath and a dilator.

    The Glidesheath is used to facilitate placement of a catheter through the skin into a vein or artery. A Mini Guide Wire (with Inserter) may be included with the Glidesheath. The mini guide wire is available in a stainless steel or a nickel-titanium alloy polyurethane coated configuration. The Inserter does not contact blood and is used strictly for guiding the Guide Wire into a cannula or Introducer.

    The A kit also contains a Surflo IV catheter, 2.5ml syringe, and scalpel for use in priming the system and gaining initial access to the vessel. Once access is obtained, the Mini Guide Wire is inserted through the cannula which was placed in the patient's blood vessel. The Glidesheath is then inserted over the Mini Guide Wire and into the blood vessel. The Mini Guide Wire is then withdrawn from the vessel. The Ditator maintains the integrity of the Sheath and dilates the blood vessel while the Glidesheath is being placed into the vessel. The Dilator can be removed and an appropriate catheter can then be inserted. The RADIFOCUS Obturator is an accessory device which creates an occlusion when inserted into the Sheath. The Obturator also provides support to the indwelling Sheath after the catheter is removed.

    The Sheath, Dilator and Obturator contain bismuth, making these devices visible under fluoroscopy.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Glidesheath, a medical device, and its substantial equivalence to a predicate device. It indicates performance testing was conducted to demonstrate this equivalence. However, the document does not describe acceptance criteria, the specific study design to prove acceptance criteria, or the detailed results of such a study in terms of quantitative performance metrics. Instead, it lists the types of verification tests performed and states that "None of the data raises any new issues of safety and effectiveness."

    Therefore, based solely on the provided text, I cannot complete the requested tables and information as it is not present in the document. The general nature of the performance section, which focuses on demonstrating substantial equivalence to a predicate device rather than meeting specific, quantified acceptance criteria for a novel functionality, means this level of detail is absent.

    Here's what can be extracted and what cannot:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Not specified in document"None of the data raises any new issues of safety and effectiveness." (This is a general statement, not specific performance data against criteria.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: Not specified.
    • Data Provenance: Not specified (since this is a device modification, likely internal laboratory testing, but not explicitly stated).
    • Retrospective or Prospective: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Number of Experts: Not applicable. This is a physical medical device, not an AI or diagnostic imaging device that requires expert ground truth establishment in the traditional sense. The "ground truth" would be the engineering specifications and performance of the device itself.
    • Qualifications of Experts: Not applicable.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not applicable. This is a physical medical device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, this is not an AI or diagnostic imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Standalone Performance: Not applicable. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Type of Ground Truth: For the "verification tests" performed (Leakage/clogging, Tensile strength of connections, etc.), the "ground truth" would be established engineering standards, material specifications, and physical measurement against those standards. It's an internal verification process, not clinical "ground truth" in the diagnostic sense.

    8. The sample size for the training set:

    • Sample Size for Training Set: Not applicable. This is a physical medical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established:

    • Ground truth for training set: Not applicable.

    Summary of available information:

    The document focuses on demonstrating substantial equivalence of a modified Glidesheath device to a predicate device (Glidesheath K033681). This is done through a series of "verification tests" on the physical properties of the device components. The tests listed are:

    • Leakage/clogging
    • Tensile strength of connections
    • Separation force of dilator and sheath
    • Internal sliding resistance
    • External sliding resistance
    • Penetration force

    The conclusion from these tests and a risk analysis is that "None of the data raises any new issues of safety and effectiveness," thus supporting substantial equivalence. The document does not provide specific numerical acceptance criteria or quantitative results for these tests.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033681
    Device Name
    GLIDESHEATH
    Manufacturer
    Date Cleared
    2003-12-11

    (17 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    GLIDESHEATH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery. The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery. The RADIFOCUS Obturator is also an accessory device which is used by placing it into the sheath to create an occlusion and further provide support to the wall of the indwelling sheath while it remains in place within the vein or artery after removal of a catheter.

    Device Description

    The Glidesheath is comprised of an introducer sheath and a dilator.

    The Glidesheath is used to facilitate placement of a catheter through the skin into a vein or artery. A Mini Guide Wire (with Inserter) may be included with the Glidesheath. The Inserter docs not contact blood and is used strictly for guiding the Guide Wire into a cannula or Introducer.

    The Mini Guide Wire is inserted through a cannula placed in the patient's blood vessel. The Glidesheath is then inserted over the Mini Guide Wire and into the blood vessel. The Mini Guide Wire is then withdrawn from the vessel. The Dilator maintains the integrity of the Sheath and dilates the blood vessel while the Glidesheath is being placed into the vessel. The Dilator can be removed and an appropriate catheter can then be inserted. The RADIFOCUS Obturator is an accessory device which creates an occlusion when inserted into the Sheath. The Obturator also provides support to the indwelling Sheath after the catheter is removed.

    The Sheath, Dilator and Obturator contain bismuth, making these devices visible under fluoroscopy.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Glidesheath device, focusing on acceptance criteria and the study proving its compliance:

    The provided text describes a 510(k) premarket notification for a new device, the Glidesheath, and its substantial equivalence to a previously cleared device, the RADIFOCUS® Introducer II (K954234). Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are not focused on a strict performance threshold for the Glidesheath itself, but rather on demonstrating that the Glidesheath performs as well as or is substantially equivalent to the predicate device.

    Here's the breakdown of the information requested, based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria Category (Demonstrates Substantial Equivalence)Reported Device Performance (Glidesheath vs. RADIFOCUS® Introducer II)
    Leakage/cloggingNo new issues of safety and effectiveness raised.
    Tensile strength of connectionsNo new issues of safety and effectiveness raised.
    Separation force of dilator and sheathNo new issues of safety and effectiveness raised.
    Internal sliding resistanceNo new issues of safety and effectiveness raised.
    External sliding resistanceNo new issues of safety and effectiveness raised.
    Penetration forceNo new issues of safety and effectiveness raised.
    Overall Substantial Equivalence (Summary of Performance)Performance of the Glidesheath is substantially equivalent to the performance of the RADIFOCUS® Introducer II.

    Note: The document explicitly states, "None of the data raises any new issues of safety and effectiveness." This implies that the Glidesheath's performance in these categories met the acceptable levels established by the predicate device and did not introduce new risks or deficiencies. The "acceptance criteria" for each test would have been defined by the performance characteristics of the predicate device, K954234.


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the sample sizes used for the verification tests (Leakage/clogging, Tensile strength, etc.).
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. Given these are physical/mechanical property tests, they would be conducted in a laboratory setting.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    This is not applicable in this context. The "ground truth" for these performance tests is typically established through direct measurement against pre-defined engineering and material specifications, and comparison to the predicate device's performance. It does not involve human experts establishing a subjective "ground truth" in the way a diagnostic imaging study would.


    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods like "2+1" or "3+1" are relevant for subjective assessments, often in clinical studies or image interpretation. For engineering verification tests, results are typically objective measurements.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating diagnostic accuracy, especially of AI/CAD systems, and involves human readers interpreting cases. The Glidesheath is a physical medical device (introducer sheath, dilator, guide wire) and its performance is assessed via engineering and material properties, not diagnostic interpretation.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The Glidesheath is a physical device, not an algorithm or AI system.


    7. The Type of Ground Truth Used

    The "ground truth" for the verification tests was based on objective measurements of physical and mechanical properties (e.g., force, resistance, leak rates) compared against the performance characteristics of the predicate device, RADIFOCUS® Introducer II. There is no mention of expert consensus, pathology, or outcomes data being used as ground truth for these specific tests.


    8. The Sample Size for the Training Set

    This is not applicable. There is no "training set" as this is not an AI/machine learning device. The "study" here refers to verification testing of a physical device.


    9. How the Ground Truth for the Training Set was Established

    This is not applicable, as there is no training set for this type of device.


    Ask a Question

    Ask a specific question about this device

    Page 1 of 1