(5 days)
The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.
The Entry Needle is an accessory device which is used to gain access to the vein or artery including but not limited to the radial artery, for placement of the Mini Guide Wire.
The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery including but not limited to the radial artery.
The Glidesheath is comprised of an introducer sheath and a dilator. The Glidesheath is coated with a hydrophilic coating to reduce the frictional resistance of the sheath when inserting or removing the sheath from the patient's blood vessel. The Sheath and Dilator contain bismuth, making these devices visible under fluoroscopy. The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.
The Entry Needle is an accessory device which is used to gain access to the vein or artery including but not limited to the radial artery, for placement of the Mini Guide Wire.
The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery including but not limited to the radial artery. The Guide Inserter which is attached to the Mini Guide Wire holder is used to straighten out the wire.
Accessories to the Glidesheath are the metal entry needle and the mini guide wire. Both the metal entry needle and the mini guide wire are packaged with the Glidesheath in a pouch prior to sterilization.
The provided text describes the 510(k) summary for the Glidesheath introducer sheath, comparing it to a predicate device (K082644) to demonstrate substantial equivalence. The information focuses on the device's design, materials, specifications, and performance testing, which is largely bench testing and biocompatibility studies.
Here's an analysis of the requested information based on the provided document:
Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state "acceptance criteria" in a quantitative, pass/fail threshold manner for each performance test. Instead, it lists the performance tests that were successfully passed. The implication is that passing these tests constitutes meeting the acceptance criteria for a medical device of this type.
Table of Acceptance Criteria and Reported Device Performance:
| Performance Test Category | Specific Test | Reported Device Performance |
|---|---|---|
| Needle | Needle surface free from defects | Successfully passed (implied: surface was free from defects) |
| Needle OD | Successfully passed (implied: met specified OD) | |
| Needle length | Successfully passed (implied: met specified length) | |
| Needle ID | Successfully passed (implied: met specified ID) | |
| Needle hub conical entry angle | Successfully passed (implied: met specified angle) | |
| Bevel indicator visibility | Successfully passed (implied: indicator was visible) | |
| Bevel indicator position | Successfully passed (implied: met specified position) | |
| Needle to hub joint strength | Successfully passed (implied: met strength requirements) | |
| Gauge Luer taper | Successfully passed (implied: conformed to Luer taper standards) | |
| Liquid leakage from fitting assembly under pressure | Successfully passed (implied: no liquid leakage) | |
| Air leakage into the fitting assembly during aspiration | Successfully passed (implied: no air leakage) | |
| Separation force of fitting assembly | Successfully passed (implied: met separation force requirements) | |
| Unscrewing torque of fitting assembly | Successfully passed (implied: met unscrewing torque requirements) | |
| Ease of assembly | Successfully passed (implied: easy to assemble) | |
| Resistance to overriding | Successfully passed (implied: resisted overriding) | |
| Stress cracking | Successfully passed (implied: no stress cracking) | |
| Corrosion resistance | Successfully passed (implied: resistant to corrosion) | |
| Guide Wire | Guidewire surface free from defects | Successfully passed (implied: surface was free from defects) |
| Tip buckling test | Successfully passed (implied: tip resisted buckling) | |
| Test for resistance of guidewires to damage by flexing | Successfully passed (implied: guidewire resisted damage from flexing) | |
| Test for fracture of guidewires | Successfully passed (implied: no fracture) | |
| Test for distal tip retention | Successfully passed (implied: distal tip was retained) | |
| Guidewire OD | Successfully passed (implied: met specified OD) | |
| Guidewire length | Successfully passed (implied: met specified length) | |
| Test for corrosion resistance | Successfully passed (implied: resistant to corrosion) | |
| Dilator | Dilator surface free from defects | Successfully passed (implied: surface was free from defects) |
| Dilator tip ID | Successfully passed (implied: met specified tip ID) | |
| Dilator to hub joint strength | Successfully passed (implied: met strength requirements) | |
| Sheath | Sheath surface free from defects | Successfully passed (implied: surface was free from defects) |
| Sheath tip ID | Successfully passed (implied: met specified tip ID) | |
| Sheath to housing joint strength | Successfully passed (implied: met strength requirements) | |
| Housing to cap joint strength | Successfully passed (implied: met strength requirements) | |
| Inserter | Guidewire inserter surface free from defects | Successfully passed (implied: surface was free from defects) |
| System | System use in model | Successfully passed (implied: system functioned correctly in a model) |
| Biocompatibility | Physicochemical Profile | Successfully passed (implied: met profile requirements) |
| Cytotoxicity | Successfully passed (implied: no cytotoxic effects) | |
| Sensitization | Successfully passed (implied: no sensitization) | |
| Acute Intracutaneous Reactivity | Successfully passed (implied: no intracutaneous reactivity) | |
| Acute Systemic Toxicity | Successfully passed (implied: no systemic toxicity) | |
| Hemolysis | Successfully passed (implied: no hemolysis) | |
| Thromobogenicity | Successfully passed (implied: not thrombogenic) | |
| Complement Activation Assay | Successfully passed (implied: no complement activation) | |
| Unactivated Partial Thromboplastin Time Assay | Successfully passed (implied: appropriate coagulation response) | |
| In Vitro Hemolysis | Successfully passed (implied: no in vitro hemolysis) | |
| Genotoxicity | Successfully passed (implied: no genotoxic effects) | |
| Pyrogen Study | Successfully passed (implied: non-pyrogenic) | |
| Extractable Metals and Acidity/Alkalinity | Successfully passed (implied: met requirements for extractables) | |
| Sterilization | ANSI / AAMI / ISO 11135 validation | Successfully validated to SAL of 10^-6 |
Study Details for Acceptance Criteria Proof:
-
Sample size used for the test set and data provenance:
- The document does not specify sample sizes for individual performance tests (e.g., how many needles were tested for joint strength).
- It also does not indicate the data provenance (e.g., country of origin, retrospective or prospective). Given the nature of bench testing, it's typically performed in a laboratory setting by the manufacturer.
-
Number of experts used to establish the ground truth for the test set and their qualifications:
- This information is not provided. For mechanical and biocompatibility testing, ground truth is typically established by adhering to recognized international standards and internal quality control specifications, rather than relying on expert consensus in the way clinical studies do. The "experts" would be the engineers and scientists performing and interpreting the standardized tests.
-
Adjudication method for the test set:
- This information is not applicable or not provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving subjective interpretations (e.g., image reading). The listed performance tests are primarily objective, quantitative, or qualitative assessments against defined standards.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size:
- No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for diagnostic imaging devices or devices where human interpretation is a key component, often comparing human performance with and without AI assistance. This document describes a physical medical device (introducer sheath) that does not involve diagnostic interpretation or AI.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No, a standalone algorithm performance study was not done. This device is a physical mechanical instrument and does not incorporate an AI algorithm.
-
The type of ground truth used:
- The ground truth for most performance tests (Needle, Guide Wire, Dilator, Sheath, Inserter, System) would be based on engineering specifications, material science standards, and functional requirements for similar medical devices.
- For Biocompatibility, the ground truth is established by international biological evaluation standards (ISO 10993), which define acceptable thresholds for biological responses (e.g., absence of cytotoxicity, genotoxicity, etc.).
- For Sterilization, the ground truth is based on sterilization validation standards (ANSI/AAMI/ISO 11135) to achieve a defined Sterility Assurance Level (SAL) of 10^-6.
-
The sample size for the training set:
- This is not applicable. The Glidesheath is a physical medical device, not an AI or machine learning model that requires a "training set."
-
How the ground truth for the training set was established:
- This is not applicable for the same reason as point 7.
{0}------------------------------------------------
510(k) Summary
FDA CDRH DMC
| A. Device Name | JUL 21 2010 | |
|---|---|---|
| Proprietary Name | Glidesheath | Received |
| Classification Name | Catheter Introducer (as per 870.1340) | |
| Common Name | Introducer Sheath | |
| Product Code | DYB |
B. Intended Use
The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.
The Entry Needle is an accessory device which is used to gain access to the vein or artery including but not limited to the radial artery, for placement of the Mini Guide Wire.
The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery including but not limited to the radial artery.
C. Device Description
The Glidesheath is comprised of an introducer sheath and a dilator. The Glidesheath is coated with a hydrophilic coating to reduce the frictional resistance of the sheath when inserting or removing the sheath from the patient's blood vessel. The Sheath and Dilator contain bismuth, making these devices visible under fluoroscopy. The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.
The Entry Needle is an accessory device which is used to gain access to the vein or artery including but not limited to the radial artery, for placement of the Mini Guide Wire.
The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery including but not limited to the radial artery. The Guide Inserter which is attached to the Mini Guide Wire holder is used to straighten out the wire.
{1}------------------------------------------------
Accessories to the Glidesheath are the metal entry needle and the mini guide wire. Both the metal entry needle and the mini guide wire are packaged with the Glidesheath in a pouch prior to sterilization.
D. Principle of Operation / Technology
The Glidesheath and its accessories are operated manually or by a manual process.
Using the Guide Inserter to straighten the Mini Guide Wire, the Mini Guide Wire is inserted through a cannula placed in the patient's blood vessel. The Glidesheath is then inserted over the Mini Guide Wire and into the blood vessel. The Mini Guide Wire is then withdrawn from the vessel. The Dilator maintains the integrity of the Sheath and dilates the blood vessel while the Glidesheath is being placed into the vessel. The Dilator can be removed and an appropriate catheter can then be inserted.
E. Design / Materials
| Current Glidesheath | Predicate Glidesheath (K082644) | |
|---|---|---|
| Sheath | ||
| Tubing | Ethylene-Tetrafluoroethylene (ETFE) copolymer containingbismuth trioxide | Ethylene-Tetrafluoroethylene (ETFE) copolymer containingbismuth trioxide |
| HydrophilicCoating | Dimethyl acrylamide-glycidyl methacrylatecopolymer | Dimethyl acrylamide-glycidylmethacrylate copolymer |
| Housing | Polypropylene | Polypropylene |
| Valve | Silicone Rubber | Silicone Rubber |
| Cap | Polypropylene | Polypropylene |
| Caulking Pin | Stainless Steel | Stainless Steel |
| Sheath Support | Styrene elastomer | Styrene elastomer |
| Side Tube | Polybutadiene | Polybutadiene |
| Three-way stopcock | ||
| Holder | Polycarbonate | Polycarbonate |
| Cock | Polyethylene | Polyethylene |
| Luer Cap | Polypropylene | Polypropylene |
| Dilator | ||
| Tube | Polypropylene containingbismuth subcarbonate | Polypropylene containing bismuthsubcarbonate |
| Hub | Polypropylene | Polypropylene |
{2}------------------------------------------------
| Current Glidesheath | Predicate Glidesheath (K082644) | ||
|---|---|---|---|
| Caulking Pin | Stainless Steel | Stainless Steel | |
| Mini Guide Wire | Plastic Option | Metallic Option | |
| Spring | Palladium | Polyurethane/tungsten with silicon coating | Stainless Steel |
| Core | Nickel-Titanium alloy | Nickel-Titanium alloy | Stainless Steel |
| Inserter | Polyethylene | Polyethylene | |
| Needle Hub | Polycarbonate | Polycarbonate | |
| Cannula | Stainless Steel | Stainless Steel | |
| Cap | Polypropylene | Polycarbonate |
Both devices have similar parts which function in the same manner. Differences in materials between the Glidesheath device covered in this submission and the predicate device, cleared under K082644, raise no new issues of safety or effectiveness.
F. Specifications
| Current Glidesheath | Predicate Glidesheath(K082644) | |
|---|---|---|
| Sheath Size | 5 & 6 French | 4, 5 & 6 French |
| Sheath Length | 10 cm | 10 - 25 cm |
| Sheath Hydrophilic Coating | 10 cm(entire length of shaft) | 10 - 25cm(entire length of shaft) |
| Dilator Length | 15.5 cm | 15.5 - 30.5 cm |
| Guide Wire OD | 0.021" | 0.021" - 0.038" |
| Guide Wire Length | 45 cm | 10 - 180 cm |
| Entry Needle: Size | 21G | 20 G - 21 G |
| Entry Needle Length | 1.5" | 1.5" |
| IV Catheter | Not Available | 16 G - 22 G |
| IV Catheter Length | Not Available | 1" - 2.5" |
| Scalpel | Not Available | Available |
| Syringe | Not Available | Available |
| Obturator | Not Available | Available |
All specifications of the Glidesheath included in this submission are within the specification ranges of the predicate device. The comparison of specifications raises no new issues of safety or effectiveness.
{3}------------------------------------------------
G. Performance
The Glidesheath successfully passed all of the following performance tests:
Needle
- Needle surface free from defects .
- ◆ Needle OD
- Needle length .
- . Needle ID
- . Needle hub conical entry angle
- Bevel indicator visibility .
- . Bevel indicator position
- . Needle to hub joint strength
- Gauge Luer taper .
- Liquid leakage from fitting assembly under pressure .
- . Air leakage into the fitting assembly during aspiration
- . Separation force of fitting assembly
- Unscrewing torque of fitting assembly ●
- Ease of assembly .
- Resistance to overriding .
- Stress cracking .
- . Corrosion resistance
Guide Wire
- Guidewire surface free from defects
- Tip buckling test .
- Test for resistance of guidewires to damage by flexing .
- Test for fracture of guidewires
- . Test for distal tip retention
- Guidewire OD .
- Guidewire length .
- Test for corrosion resistance .
Dilator
- . Dilator surface free from defects
- . Dilator tip ID
- . Dilator to hub joint strength
Sheath
- Sheath surface free from defects 0
- . Sheath tip ID
- . Sheath to housing joint strength
{4}------------------------------------------------
- Housing to cap joint strength .
Inserter
-
Guidewire inserter surface free from defects .
System -
System use in model ●
H. Biocompatibility and Sterilization
The Glidesheath is classified as Externally Communicating Devices, Circulating Blood, Limited Contact (≤ 24h). Results of the testing demonstrate that the blood contacting materials are biocompatible.
Blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing". The Glidesheath successfully passed all of the following biocompatibility tests:
- Physicochemical Profile .
- Cytotoxicity
- Sensitization .
- � Acute Intracutaneous Reactivity
- Acute Systemic Toxicity .
- Hemolysis .
- Thromobogenicity t
- Complement Activation Assay .
- Unactivated Partial Thromboplastin Time Assay .
- In Vitro Hemolysis .
- Genotoxicity .
- Pyrogen Study t
- Extractable Metals and Acidity/Alkalinity (needle assembly) .
Sterilization conditions have been validated according to ANSI / AAMI / ISO 11135. Sterilization of Health Care Products-Part I: Requirements for development, validation and routine control of a sterilization process for medical devices to provide a Sterility Assurance Level (SAL) of 10-6.
I. Substantial Equivalence
The performance of the Glidesheath in this submission is substantially equivalent to the performance of the predicate device. The equivalence was shown through
{5}------------------------------------------------
comparison of component materials and specifications, performance and biocompatibility testing and sterilization validation.
The Glidesheath is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the predicate device the Glidesheath, cleared under K082644. Differences between the devices do not raise any significant issues of safety or effectiveness.
J. Submitter Information
| Prepared By: | Mr. Daniel R. PlonskiRegulatory Affairs Specialist |
|---|---|
| Prepared For: | Terumo Medical Corporation950 Elkton Blvd.Elkton, MD 21921Phone: (410) 392-7213Fax: (410) 398-7395Email: daniel.plonski@terumomedical.com |
| Date Prepared: | July 2, 2010 |
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Contro! Room -WO66-G609 Silver Spring, MD 20993-0002
Terumo Medical Corporation c/o Mark Job Reviewer Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313
Re: K102008 ·
Trade/Device Name: Glidesheath Regulation Number: 21 CFR 870.1250 Regulation Name: Catheter, Introducer Regulatory Class: Class II (Two) Product Code: DYB Dated: July 15, 2010 Received: July 16, 2010
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
UNI 2 1 2010
{7}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
R. Vahner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Indications for Use
Comments of the comments of the comments of the comments of the comments of the contribution of the contribution of the contribution of the contribution of the contribution o
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Glidesheath™
Indications For Use:
UUU. 21 2010
The Glidesheath is used to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.
The Entry Needle is an accessory device which is used to gain access to the vein or artery including but not limited to the radial artery, for placement of the Mini Guide Wire.
The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery including but not limited to the radial artery.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
bma R. hichner
Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_K102008
§ 870.1340 Catheter introducer.
(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).