Search Filters

Search Results

Found 88 results

510(k) Data Aggregation

    K Number
    K243956
    Date Cleared
    2025-06-10

    (169 days)

    Product Code
    Regulation Number
    868.1400
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Nihon Kohden Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nihon Kohden TG-980P/TG-980P1 CO2 Sensor Kit is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status.

    Along with other methods indicated by the physician for medical diagnosis, this device is intended as an indicator of patient carbon dioxide concentration during expiration. The device is intended for use by qualified medical personnel within a hospital, ambulatory care, transport, or any other clinical environment.

    Device Description

    The TG-980P and TG-980P1, collectively referred to as TG-980P/TG-980P1, are CO2 Sensor Kits used to measure the concentration of carbon dioxide (CO2 or CO2) during patient expiration. The TG-980P/TG-980P1 is intended for use by qualified medical personnel as an aid for determining patient ventilatory status within a hospital, ambulatory care, transport, or any other clinical environment.

    The TG-980P/TG-980P1 CO2 Sensor Kit comprises three main components: a CO2 sensor, an interface connector, and a sensor cable. It utilizes the mainstream measurement method to sample gas directly from the patient's airway and non-dispersive infrared (NDIR) absorption technology to measure CO2 concentration during patient expiration. CO2 gas absorbs infrared light at specific wavelengths, and the amount absorbed is directly related to CO2 concentration. The CO2 sensor is equipped with a light source that generates infrared light. The light passes through an attached Nihon Kohden accessory (except the Thermal Airflow Sensor) through which the expired air flows and is converted by photodetectors to voltage, which is used to calculate CO2 concentration. The calculated digital data is then transmitted and displayed on a connected patient monitor or other device.

    The TG-980P/TG-980P1 CO2 Sensor Kit can be connected via the interface connector and used with Nihon Kohden devices for which the operator's manual specifies compatibility with the TG-980P/TG-980P1.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to a hardware device (CO2 Sensor Kit) and not an AI/ML software. Therefore, many of the requested criteria regarding AI-specific studies (e.g., sample size for training set, number of experts for ground truth, MRMC study) are not applicable.

    However, I can extract the relevant acceptance criteria and details of the non-clinical performance studies conducted for the device.


    Acceptance Criteria and Device Performance (Non-AI Device)

    1. A table of acceptance criteria and the reported device performance

    The document provides the performance specifications for the CO2 Sensor Kit, rather than explicit "acceptance criteria" in the typical sense of a target for a specific study. The reported device performance is compared to the predicate device's performance.

    Performance CharacteristicAcceptance/Predicate SpecificationSubject Device (TG-980P/TG-980P1) Reported Performance
    Trade/Device NameNihon Kohden TG-970P Series CO2 Sensor KitNihon Kohden TG-980P/TG-980P1 CO2 Sensor Kit
    Regulatory ClassClass IIClass II
    Product CodeCCK (868.1400)CCK (868.1400)
    Intended UseTo be used under the control of a healthcare professional to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status; an indicator of patient carbon dioxide concentration during expiration.Equivalent, with slightly changed wording but same meaning.
    Indications for UseTo measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status; an indicator of patient carbon dioxide concentration during expiration. For use by qualified medical personnel within a hospital or clinical environment.Same as predicate, but with clarification of intended use environments to include "ambulatory care, transport, or any other clinical environment."
    Type of UsePrescription Device OnlyPrescription Device Only
    Intended PopulationAdult and children patients 7 kg or more (Predicate) / Adults, children, infants and neonates (Reference)Neonates to adults. All types of genders, races, and languages. (Same as Reference device's broader population)
    Patient ContactIntubated (Predicate) / Intubated/Non-intubated (Reference)Intubated/Non-intubated (Same as Reference device)
    ConfigurationCO2 sensor, Connector, Sensor cable, CO2 adapterCO2 sensor, Connector, Sensor Cable (CO2 adapter removed)
    Dimension (CO2 sensor)37 x 8.3 x 13.7 mm ±10%37 x 8.3 x 13.7 mm ±10%
    Weight (Sensor part)4 g ±1 g4 g ±2 g (Weight tolerance adjusted)
    Sampling methodMainstream infrared absorptionMainstream infrared absorption
    Measurement principleSingle-wave spectroscopic method (Non-dispersive infrared gas analyzing method (NDIR))Single-wave spectroscopic method (Non-dispersive infrared gas analyzing method (NDIR))
    CO2 measurement methodQuantitative methodQuantitative method
    CalibrationYESYES
    EtCO2 determinationYESYES
    No Breath detect limitCO2
    Ask a Question

    Ask a specific question about this device

    K Number
    K242750
    Device Name
    Central Station
    Date Cleared
    2025-06-05

    (266 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Nihon Kohden Digital Health Solutions, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Central Station is a network device, intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters.

    Central Station does not perform any data processing on the data received from the Nihon Kohden compatible devices. When Central Station is connected with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the Central Station can:
    • Admit and discharge patients on the Nihon Kohden network.
    • Display and manage compatible devices' real-time patient clinical data, vital signs, alarms and waveforms.
    • Review and trend data calculated by connected Nihon Kohden devices.
    • Store and transfer historical clinical data for the connected systems.
    • Print patient data.

    Central Station is intended for use in professional medical facilities by trained medical personnel.

    Device Description

    Central Station is software only product that is installed on a Commercial Off the Shelf (COTS) Computer.

    Central Station displays waveforms data and numerical data from a connected bedside monitor, vital sign telemeter, or multiple patient receiver unit on the screen.

    Central Station is a network device, intended to provide remote patient monitoring to medical personnel. Central Station displays a list of measured values and a trend graph. Numerical data and various waveforms are color-coded for each parameter. Central Station also has the function of displaying an alarm.

    Alarm indication in Central Station is displayed as a result of a judgment by the bedside monitor, vital sign telemeter, or multiple patient receiver units connected to the Central Station. Central Station itself does not have the function to perform alarm indication judgment.

    AI/ML Overview

    The provided document is a 510(k) clearance letter for the "Central Station" device. This type of document primarily focuses on establishing substantial equivalence to a legally marketed predicate device rather than detailing specific performance acceptance criteria and study results in the same way as a full clinical trial report or a detailed design validation report would.

    The document states:

    • "Central Station does not perform any data processing on the data received from the Nihon Kohden compatible devices."
    • "Alarm indication in Central Station is displayed as a result of a judgment by the bedside monitor, vital sign telemeter, or multiple patient receiver units connected to the Central Station. Central Station itself does not have the function to perform alarm indication judgment."
    • "The results of the substantial equivalence assessment, taken together with non-clinical bench testing, software verification, and validation demonstrate that the Central station does not raise concerns regarding its safety and effectiveness compared to its predicate device and operates in accordance with claimed indications for use."

    Given these statements, the "Central Station" device is essentially a display, recording, and communication hub. It does not perform diagnostic algorithms or make independent judgments that would typically necessitate the kinds of detailed performance metrics (like sensitivity, specificity, or reader agreement) that are usually established through extensive multi-reader, multi-case (MRMC) studies with expert ground truth. Its primary function is to accurately display and relay data processed by other Nihon Kohden devices.

    Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" for this specific device (Central Station) would primarily revolve around:

    • Software Verification and Validation (V&V): Ensuring the software correctly displays, records, and transmits data as designed, and that all features (admit/discharge, trend display, printing, network communication) function as intended without errors.
    • Non-Clinical Bench Testing: Confirming interoperability with compatible devices, accuracy of data display, alarm relay, and network functionality.
    • Cybersecurity Compliance: Meeting regulatory requirements for cybersecurity.

    The document explicitly states that "The software documentation was prepared following the FDA's 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices' (released June 14, 2023), specifically for an Enhanced Documentation Level." and "Verification testing was conducted at the system integration level to confirm that the device software fulfills its requirements and that safety and security risk mitigations, where applicable, were effective. Additionally, system-level testing was carried out to show that the software addresses user needs. All unit, integration, and system-level tests successfully met the test protocols."

    Based on the provided text, it is not possible to extract the specific quantitative performance metrics (like sensitivity, specificity, or effect sizes for human readers) that would be relevant for a device performing complex data processing or diagnostic functions. The document emphasizes substantial equivalence and basic functional verification, not advanced AI/diagnostic performance validation.

    However, I can infer the spirit of typical acceptance criteria and how a device like this would be proven to meet them, guided by the information provided.


    Inferred Acceptance Criteria and Device Performance for "Central Station"

    Given the device's stated function (display, record, print, and relay data without processing it for diagnosis or alarm judgment), the acceptance criteria would focus on functional correctness, data integrity, interoperability, and system reliability, rather than diagnostic accuracy metrics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred)Reported Device Performance (Inferred from document)
    Functional Performance1. Data Display Accuracy: All physiological data (waveforms, numerics, alarms) received from connected compatible devices are accurately and synchronously displayed."Display and manage compatible devices' real-time patient clinical data, vital signs, alarms and waveforms." "Numeric data and various waveforms are color-coded for each parameter." Implied: Accuracy of display confirmed via functional tests against known inputs from connected devices.
    2. Data Recording & Storage: Historical clinical data, trends, and events are accurately stored and retrievable for the specified durations/files."Store and transfer historical clinical data for the connected systems." Review history data storage (e.g., Trendgraph: 120 hours, Arrhythmia recall: 1,500 files, Event list: 10,000 files). Implied: Storage and retrieval validated against specified capacities and data integrity checks.
    3. Printing Functionality: Patient data can be accurately printed, including specified parameters like patient info, waveforms, and trends."Print patient data." Ability to print various review windows (Trend, Full Disclosure, Arrhythmia Recall, etc.). Implied: Printing validated for completeness and accuracy of generated reports.
    4. Patient Management: Functions for admitting, discharging, pausing, and transferring patients within the network operate correctly."Admit and discharge patients on the Nihon Kohden network." Support for Admit, Discharge, Pause, Transfer functions (within one CS, between CSs, manual/auto entry). Implied: Workflow and data handling for patient management validated.
    5. Alarm Relay: Alarms generated by connected bedside monitors are accurately received and displayed/mimicked by Central Station."Alarm indication in Central Station is displayed as a result of a judgment by the bedside monitor... Central Station itself does not have the function to perform alarm indication judgment." Implied: Alarms from connected devices are correctly received and presented as per design.
    Interoperability & Connectivity1. Compatible Device Connection: Successful and stable connection to all specified Nihon Kohden bedside monitors, telemetry receivers/transmitters.Compatible with NK Bedside Monitors (BSM: 1700, 3000, 6000, G9, G5, G7), Vital Signs Monitor (SVM-7200), NK Telemetry (GZ-120/130/140), Multiple Patient Receiver/Transmitters (ORG-9700/9100, ZS-940, ZM-520/521/530/531), Central Monitor (CNS-6201/6801/2101). Max 32 connections. Implied: Connectivity and data exchange verified through testing with all listed compatible devices.
    2. Network Communication: Reliable communication via NET-9/LS-NET protocol within the Nihon Kohden network."Network: Yes (NK Network)." "Communication protocol: NET-9/LS-NET communication." Implied: Network communication validated for stability, data integrity, and compliance with protocol.
    Software Quality & Reliability1. Software Functionality: All software features operate as per specifications without critical errors or crashes."Verification testing was conducted at the system integration level to confirm that the device software fulfills its requirements... All unit, integration, and system-level tests successfully met the test protocols." Implied: Comprehensive software V&V confirmed functional correctness and absence of defects.
    2. Cybersecurity: Device adequately mitigates cybersecurity risks and complies with relevant guidance."Cybersecurity information has been provided in line with the FDA's 'Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions' guidance document, dated September 27, 2023." "Interoperability-related risk management activities are included in Cybersecurity Risk Management activities." Implied: Security testing confirmed adherence to cybersecurity standards.
    Risk Management1. Safety & Effectiveness: Risks associated with interoperability, anticipated users, and foreseeable misuse are adequately addressed."The CENTRAL STATION has been designed and verified through a risk analysis that considers the risks associated with interoperability, the anticipated users, reasonably foreseeable misuse, and reasonably foreseeable combinations of events that can result in a hazardous situation." Implied: Risk analysis documented and mitigation verified, ensuring the device does not raise new safety/effectiveness concerns compared to the predicate.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: Not explicitly stated as a "sample size" in the context of patient data for diagnostic algorithms, because the device doesn't perform diagnostic processing. The "test set" for this device would be a collection of test cases covering all functional requirements, interoperability scenarios with different compatible devices, alarm conditions (relayed), data storage limits, network loads, and user interaction flows. The document mentions "system integration level" and "system-level testing" and that "all unit, integration, and system-level tests successfully met the test protocols." This implies a comprehensive set of non-clinical, bench-level tests.
    • Data Provenance: Not applicable in the sense of clinical patient data (e.g., from specific countries, retrospective/prospective studies), as the device does not process primary patient data for diagnosis. The data used for testing would be simulated, generated, or derived from compatible Nihon Kohden monitor outputs in a lab setting to verify the Central Station's display and communication functions.

    3. Number of Experts and Qualifications for Ground Truth

    • Not Applicable in the traditional sense for diagnostic AI. The ground truth for this device's performance would be the expected output based on its functional specifications and the known inputs from the connected Nihon Kohden devices. For example, if a connected monitor transmits an HR of 70 bpm, the ground truth is that the Central Station must display 70 bpm. These "ground truths" are established by engineering design specifications, not human expert consensus on clinical findings.
    • Experts Involved: Software engineers, quality assurance engineers, subject matter experts on the physiological monitoring systems, and potentially clinical users for usability and workflow testing. Their qualifications would be in device design, software development, testing, and clinical application.

    4. Adjudication Method for the Test Set

    • Not Applicable in the context of clinical interpretation adjudication (e.g., 2+1 radiologist consensus). Adjudication in this context would be internal to the software development and testing process:
      • Test Pass/Fail Criteria: Predetermined pass/fail criteria for each test case.
      • Bug/Defect Resolution: Issues found during testing are logged as bugs, investigated by engineers, and resolved, followed by retesting.
      • Verification Sign-off: Test leads or design engineers review test results and formally sign off on the successful completion of verification.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, not performed for this device. An MRMC study is relevant for AI systems that assist human readers in making diagnostic decisions (e.g., radiologists reading X-rays with AI assistance). The Central Station device does not perform any such diagnostic assistance; it merely displays data from other (likely already cleared) devices. Its function is analogous to a monitor or a remote display unit, not a diagnostic AI.

    6. Standalone (Algorithm Only) Performance

    • Not Applicable. This device does not have a "standalone algorithm" that performs diagnostic or data processing functions independently. Its function is entirely dependent on receiving data from other compatible Nihon Kohden devices.

    7. Type of Ground Truth Used

    • Functional Specifications and Truth from Connected Devices: The ground truth for this device is based on its functional design specifications (e.g., "display received data," "store data for X hours," "print Y parameters") and the verified output from the connected Nihon Kohden bedside monitors and telemetry systems. It's about data integrity and display accuracy, not clinical outcomes or pathology.

    8. Sample Size for the Training Set

    • Not Applicable. This device is described as "software only product" that "does not perform any data processing on the data received," and "Central Station itself does not have the function to perform alarm indication judgment." This strongly implies it is a rule-based or deterministic system, not a machine learning/AI system that requires a "training set" in the common sense (i.e., for learning patterns from data for prediction or classification). Therefore, there is no training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there is no training set for an AI algorithm, there is no corresponding ground truth establishment process for a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K231778
    Date Cleared
    2024-03-01

    (259 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Nihon Kohden OrangeMed, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-550 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-550 is intended for use in hospitals and hospital-type facilities, as well as, for in-hospital transportation.

    Device Description

    The Nihon Kohden NKV-550 Series Ventilator System consists of a graphic user interface (GUI) and a breath delivery unit (BDU). The GUI allows clinicians to set ventilator control parameters such as tidal volume and inspiratory pressure, to set alarm limits such as high inspiratory pressure alarm, to view monitored numeric values, to view waveform and loops, and to operate various features through the apps. The BDU contains a microprocessor that receives inputs from the electronic system and controls the pneumatic system for breath delivery to the patient. It also provides various alarms, a safety valve, and other design features to maximize patient safety.

    AI/ML Overview

    I am unable to extract specific acceptance criteria for the Nihon Kohden NKV-550 Series Ventilator System or details of a study proving it meets them. The provided text from the FDA 510(k) Summary primarily focuses on establishing substantial equivalence to predicate devices, rather than detailing a specific performance study with acceptance criteria.

    The document states:

    • "Performance of the Nihon Kohden NKV-550 Series Ventilator System was demonstrated by the following testing performed in compliance with Design Controls: Software Verification, Device Functionality Testing, Human Factors/Usability Testing, Bench Test Comparison with Reference Device, Risk Management."
    • "Summary of Clinical Performance Data: Not Applicable – Clinical performance data was not required to demonstrate substantial equivalence."
    • "Performance: Met ISO 80601-2-12 requirements on essential performance of critical care ventilator"

    This indicates that the device's performance was evaluated against the general requirements of the ISO 80601-2-12 standard for critical care ventilators and through internal design control processes like software verification and functionality testing, but a standalone study with explicit acceptance criteria and performance metrics for the device itself (beyond meeting the ISO standard) is not detailed in this summary.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor details about sample sizes, ground truth establishment, expert involvement, adjudication methods, or MRMC studies, as this information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223567
    Date Cleared
    2023-07-31

    (244 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Nihon Kohden Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CNS-2101 central monitor is a networked multi-patient monitoring system, that is intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters. The CNS-2101 does not perform any data processing on the Nihon Kohden compatible devices. When the CNS-2101 is to connect with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the CNS-2101 can:

    • · Admit and discharge patients on the Nihon Kohden network.
    • · Display and manage compatible devices' real-time patient clinical data.
    • Mimic the alarms of connected devices when a measured parameter falls outside a preset limits or when an arrhythmia is detected.
    • · Review and trend data calculated by connected Nihon Kohden devices.
    • · Store and transfer historical clinical data for the connected systems.
    • · Print patient data.

    The CNS-2101 is intended for use in professional medical facilities by trained medical personnel.

    Device Description

    The CNS-2101 central monitor is a central monitoring device designed to support medical personnel to provide medical care to multiple patients at the same time. It acquires vital sign data from multiple monitoring devices such as bedside monitors and displays the acquired data such as ECG and pulse rate on the screen as well as informing alarms.

    The CNS-2101 can communicate with other devices through a network connection. The CNS-2101 can acquire vital sign data directly from multiple monitoring devices (e.g., bedside monitors) connected to Nihon Kohen Monitoring device network or using multiple patient receivers and transmitters, or by a combination of both methods. The parameters to monitor on the central monitor can be changed as necessary by selecting a monitoring device such as a bedside monitor or transmitter and changing the parameter settings for that device.

    The CNS-2101 is designed to be installed in a location outside the patient environment such as a nurse's station for central monitoring.

    AI/ML Overview

    The provided text is a 510(k) summary for the Nihon Kohden CNS-2101 Central Monitor. This document focuses on demonstrating substantial equivalence to a predicate device through comparison of features and adherence to recognized standards. It does not present specific acceptance criteria with numeric performance values or detailed performance study results in the way typically found for AI/ML-driven devices with diagnostic or prognostic claims.

    Therefore, for the information requested:

    1. A table of acceptance criteria and the reported device performance: This information is not explicitly provided in the document. The document states that the device was tested according to international and FDA-recognized consensus standards to verify and validate its functionality and technical characteristics. However, specific performance metrics (e.g., sensitivity, specificity, accuracy) are not listed for the CNS-2101.

    2. Sample size used for the test set and the data provenance: This information is not provided. The testing described is verification and validation (V&V) against design specifications and consensus standards, not clinical performance testing with patient data in the context of diagnostic accuracy.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not applicable and therefore not provided. "Ground truth" in the context of diagnostic accuracy is not relevant to the described V&V testing of a central monitoring system.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable and therefore not provided.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable and therefore not provided. The CNS-2101 is a central monitoring system that displays data from other devices; it does not utilize AI to assist human readers in interpretation or diagnosis.

    6. If a standalone (i.e., algorithm-only without human-in-the-loop performance) was done: Not applicable and therefore not provided. The CNS-2101 is a networked multi-patient monitoring system, not a standalone diagnostic algorithm. It displays data from already cleared bedside monitors and telemetry systems. It "does not perform any data processing on the data received from the Nihon Kohden compatible devices."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable and therefore not provided, as the testing relates to the functional performance and safety of the system rather than diagnostic accuracy against a specific condition.

    8. The sample size for the training set: Not applicable and therefore not provided. The document describes a central monitoring system, not an AI/ML device that requires training data.

    9. How the ground truth for the training set was established: Not applicable and therefore not provided.

    Summary of available information regarding performance:

    • Acceptance Criteria: Not explicitly defined with performance metrics like accuracy, sensitivity, or specificity. Instead, the acceptance criteria are implied by adherence to recognized consensus standards for medical electrical equipment, software lifecycle, risk management, usability, electromagnetic compatibility, and alarm systems.
    • Study Proving Device Meets Acceptance Criteria: Non-clinical verification and validation testing was conducted on the CNS-2101 based on the following standards and guidance:
      • ISO 14971:2019 Medical devices - Application of risk management to medical devices
      • IEC 62304:2006/A1:2015 Medical device software - Software life cycle processes
      • ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment—Part 1: General requirements for basic safety and essential performance
      • IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
      • IEC 60601-1-8: 2012 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
      • IEC 60601-2-27:2011 Medical electrical equipment--Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
      • ANSI/AAMI/IEC 62366-1:2015 Medical devices -Part 1: Application of usability engineering to medical devices.
      • Various FDA guidance documents related to software validation, cybersecurity, interoperable medical devices, off-the-shelf software, EMC, and human factors.
    • Reported Device Performance: The document states that "All function in the CNS-2101 have been validated by the FDA-recognized consensus standards, therefore, the validity of the results can be ensured." It also notes that "Tests conducted within the software development cycle demonstrate that communication between the CNS-2101 and the Nihon Kohden 510(k)-cleared monitoring devices is robust and stable enough to exchange accurate physiological data in real time."

    In essence, the document confirms that the CNS-2101 complies with relevant safety and performance standards for its intended function as a central monitoring system, without requiring a clinical validity study involving comparison to a "ground truth" or expert adjudication for diagnostic accuracy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222644
    Date Cleared
    2023-05-31

    (272 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Nihon Kohden OrangeMed, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nihon Kohden NKV-440 Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-440 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-440 is intended for use in hospitals and hospital-type facilities, as well as, for in-hospital transportation.

    Device Description

    The NKV-440 is a servo-controlled ventilator that is designed to meet the gas delivery and performance requirements for neonate through adult patients. The NKV-440 design is comprised of two major components, a Breath Delivery Unit (BDU) and a Graphic User Interface (GUI). The GUI allows clinicians to set ventilator control parameters such as PEEP and inspiratory pressure, to set alarm limits such as high inspiratory pressure alarm, to view monitored numeric values, and to view waveforms. The BDU assembly contains a blower and the electronics required to perform breath delivery. Ambient air is taken into the blower and mixed with oxygen which is flow rate controlled by a proportional valve. The mixed gas is provided to the patient. The microprocessor controls the blower and the proportional valve to deliver the pressure and oxygen concentration which are set by the user. It also provides various alarms and other design features to maximize patient safety.

    AI/ML Overview

    The provided document is an FDA 510(k) premarket notification for the Nihon Kohden NKV-440 Ventilator System. It details the device's indications for use, technological characteristics, and comparison to a predicate device (Nihon Kohden NKV-550 Ventilator System).

    The document does not include information about AI/ML algorithm performance, acceptance criteria for such algorithms, or studies involving human readers or ground truth established by experts for image analysis. Therefore, I cannot address most of the specific points requested regarding acceptance criteria and the study proving the device meets them, as it relates to AI/ML device performance.

    This 510(k) pertains to a physical medical device (a ventilator), and its substantial equivalence is demonstrated through engineering performance testing against established standards and comparison of technical characteristics with a legally marketed predicate device.

    However, I can extract the general acceptance criteria for this type of device based on the information provided:

    Summary of Acceptance Criteria and Device Performance (Based on the provided document for a Ventilator):

    Since this is a ventilator and not an AI/ML diagnostic tool, the "acceptance criteria" discussed are related to performance against established engineering and safety standards, and functional equivalence to a predicate device, rather than diagnostic accuracy metrics.

    Acceptance Criteria (General for Ventilators)Reported Device Performance (Nihon Kohden NKV-440 Ventilator System)
    Intended Use Equivalence: Provide continuous ventilation for adult, pediatric, and neonatal patients requiring invasive or noninvasive respiratory support, including mandatory and spontaneous ventilation modes and respiratory monitoring, for use in hospitals and hospital-type facilities, as well as in-hospital transport.Met: "The Nihon Kohden NKV-440 Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-440 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-440 is intended for use in hospitals and hospital-type facilities, as well as, for in-hospital transportation." (Identical to predicate)
    Clinical Conditions Equivalence: Use in hospitals, hospital-type facilities, and in-hospital transportation for patients needing ventilation therapy.Met: "Hospitals, hospital-type facilities and in-hospital transportation for patients who need ventilation therapy." (Identical to predicate)
    Anatomical Site Equivalence: Patient airways.Met: "Patient airways." (Identical to predicate)
    Target Population Equivalence: Adult, pediatric, and neonatal patients.Met: "Adult, pediatric and neonatal patients." (Identical to predicate)
    User Equivalence: Trained clinicians, not lay users.Met: "Trained clinicians, not lay users." (Identical to predicate)
    Compliance with ISO 80601-2-12 (Essential Performance of Critical Care Ventilator):Met: "Met ISO 80601-2-12 requirements on essential performance of critical care ventilator." (Identical to predicate)
    Waveform Comparison: Comparable waveform test result to predicate device.Met: "Comparable waveform test result."
    Chemicals Deployed to Patient Equivalence: Air and oxygen.Met: "Air and oxygen." (Identical to predicate)
    Delivery Method to Patient Equivalence: Positive pressure.Met: "Positive pressure." (Identical to predicate)
    Energy Used for Device Equivalence: AC Power and internal DC Power (Li-ion battery).Met: "AC Power and internal DC Power (Li-ion battery)." (Identical to predicate, noting "Exactly same battery packs are used in both NKV-440 and NKV-550.")
    Control Principle Equivalence: Time-cycled, volume-constant, pressure-controlled.Met: "time-cycled, volume-constant, pressure-controlled." (Identical to predicate)
    Breathing Modes Equivalence: All specified invasive and non-invasive modes (A/CMV-PC, A/CMV-VC, A/CMV-PRVC, SIMV-PC-PS, SIMV-VC-PS, SIMV-PRVC-PS, SPONT-CPAP, SPONT-PS, SPONT-VS, APRV, CPAP, O2 Therapy).Met: All listed modes are identical to the predicate device.
    Compliance with Relevant Standards: (e.g., ANSI AAMI ES 60601-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62133, IEC 62304, IEC 62366-1, ISO 10993-1, ISO 14971, ISO 18562-1, ISO 80601-2-12, ISO 80601-2-55, ISO 80601-2-61).Met: The device "has been tested in compliance with the following standards" (a comprehensive list is provided).
    Other Non-Clinical Performance: Performance of Ventilation Modes and Control Settings, Device/Software Functionality, Power performance, Essential Performance and Worst Case VBS, Environmental, Product Endurance/Reliability, Biocompatibility, Cleaning & Disinfection, EMC and Electrical Safety, Human Factors/Usability, Risk Management, Compatibility with 3rd Party Devices.Demonstrated: "Performance of the Nihon Kohden NKV-440 Ventilator was demonstrated by the following..." (followed by the list above).

    Regarding the points specific to AI/ML device studies, the document states "Not Applicable" for clinical and animal performance data, and does not mention any AI/ML components inherent to the device or its assessment.

    Therefore, the following points cannot be answered from the provided text:

    1. Sample sizes used for the test set and the data provenance: Not applicable, as this is not an AI/ML clinical study. Performance is demonstrated through engineering tests and comparison to a predicate device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.

    In conclusion, the supplied document describes the regulatory clearance for a conventional medical ventilator by demonstrating substantial equivalence to a predicate device through non-clinical performance data and adherence to recognized standards. It does not involve AI/ML technology or human reader studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220989
    Date Cleared
    2022-07-29

    (116 days)

    Product Code
    Regulation Number
    870.1025
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Nihon Kohden Digital Health Solutions, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Next Generation NetKonnect is interface the Nihon Kohden monitoring network with networked client PCs to annunciate and display patient monitoring information to healthcare providers. The device is intended for use in near real-time monitoring of routing patient status and alarm events. Next Generation NetKonnect supplements the primary patient monitoring system by providing a forwarding for annunciating and displaying patient alarm events and event related information including vital signs values and waveforms. The system is intended for use on any patients as determined by qualified medical personnel within a hospital or clinical environment.

    Device Description

    NGNK is a software product that runs on Microsoft Windows architecture. NGNK will communicate with Nihon Kohden devices by a network connection and through the Gateway. The NGNK is intended to interface the Nihon Kohden monitoring network to annunciate and display patient monitoring information to healthcare providers. The device is intended for use in near real-time monitoring of patient status and alarm events. NGNK supplements the primary patient monitoring system by providing remote monitoring capability, including vital signs, alarms and waveforms.

    AI/ML Overview

    The provided document describes the FDA 510(k) clearance for the Next Generation NetKonnect device. This device is a software product designed to interface with the Nihon Kohden monitoring network to annunciate and display patient monitoring information to healthcare providers. It is an update to a previously cleared device, NetKonnect Remote Network Extension (K112637).

    The document focuses on demonstrating substantial equivalence to the predicate device, rather than proving performance against specific quantitative acceptance criteria for a novel device. The primary changes are updates to the User Interface and support for current 64-bit Windows architectures.

    Here's an breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria in terms of performance metrics (e.g., sensitivity, specificity, accuracy) for physiological measurements or alarm detection. This is because the device primarily functions as a display and annunciation system for data already processed by the primary patient monitoring system, which itself has established performance.

    Instead, the "acceptance criteria" appear to be implicit in demonstrating that the modified device performs its intended functions (displaying vital signs, waveforms, and alarm events) reliably and safely, consistent with its predicate. The "reported device performance" is summarized in a qualitative manner through the non-clinical testing.

    Feature/RequirementAcceptance Criteria (Implied)Reported Device Performance
    Functional EquivalenceThe device must perform the same core functions as the predicate: interface with the monitoring network, annunciate/display patient monitoring info (vital signs, waveforms, alarm events), and act as a supplementary system.The "Description of Communication requirements" and "Performance Specification" sections indicate that the subject device (NGNK) "SAME" as the predicate for receiving and forwarding waveform/numerical data using Nihon Kohden's proprietary “NET9” protocol. The primary function of displaying patient history information (Trend Data, Hemodynamics lists, Arrhythmia recall, ST recall, Full disclosure, ECG 12 lead analysis) is also stated as "SAME".
    Safety and EffectivenessThe modifications (UI update, OS support) must not introduce new questions of safety or effectiveness. The device must be as safe and effective as the predicate."The required testing of the Next Generation NetKonnect was performed in accordance with the requirements of the design control guidelines and established quality assurance processes to demonstrate substantial equivalence... The software testing demonstrated that the software version meets it design requirements." Conclusion: "Based on the design and results of testing, it can be concluded that the subject device - Next Generation NetKonnect - is as safe, as effective, and performs as well as or better than the predicate."
    Alarm NotificationAlarm notification functionality should be clear and not replace the primary alarm system. (Note: NGNK differs slightly but is deemed substantially equivalent).Predicate: "Alarm notification is a secondary notification system and does not replace the primary alarm notification at the bedside monitor." Subject device: "SAME." However, the subject device "NGNK does not forward alarm notification. This difference does not affect the device from performing in a manner in which it is intended to operate from the predicate device “NetKonnect”". This implies that while the mechanism might be different, it still serves the intended supplementary purpose without compromising the primary system.
    Operating System SupportThe device should run on a compatible and current operating system.Predicate: "Micorsoft windows that support Win32 service processes." Subject device: "Substantially equivalent: NGNK now supports current standard 64 bit Windows archetures." This is an improvement to support modern IT environments.
    User Interface (UI) DesignThe UI should be updated to be consistent with current technology and user experience while maintaining clarity and functionality similar to connected FDA cleared devices.Predicate: "User interface is similar to the connected FDA cleared devices Nihon Kohden bedside monitoring and primary central monitoring systems." Subject device: "Substantially equivalent: Updates in the user interface to be consistent with current user interfaces on connected FDA cleared devices Nihon Kohden bedside monitoring and primary central monitoring systems." The primary objective of NGNK was stated as "to update the User Interface due to advancements in technologies and match current on-market products for a consistent experience."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states: "No clinical testing was required or performed since substantial equivalence of the device was supported by the non-clinical testing."

    Therefore, there is no clinical test set, sample size, or specific data provenance (country, retrospective/prospective) to report for the performance of the device on patient data. The evaluation was primarily internal engineering and software testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Given that no clinical testing was performed and the evaluation relied on non-clinical (software) testing, there were no external experts or "ground truth" derived from patient data for an independent performance evaluation. The "ground truth" for software testing usually refers to specified design requirements and expected outputs, which are internally verified by the manufacturer's engineering team.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    There was no adjudication method used for a clinical test set, as no clinical testing was performed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was performed. The device is not an AI-assisted diagnostic tool that would typically involve human readers and their performance improvement. It is a display and annunciation system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is software that operates "standalone" in the sense that it processes and displays data from patient monitors. However, its purpose is to complement human monitoring in a clinical environment. The software testing mentioned is effectively a standalone performance evaluation against its design requirements and functional specifications. The specific results of this software testing (e.g., bug reports, pass/fail rates for specific test cases) are not detailed in this summary.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical (software) testing, the "ground truth" would be the design specifications and functional requirements of the software. This involves verifying that the software correctly receives, processes, and displays the data as intended, and that the user interface functions as designed, without introducing errors or safety concerns. This type of ground truth is established by the manufacturer during the design and development process.

    8. The sample size for the training set

    The document does not mention any training set. This is not an AI/machine learning device that typically requires training data. It is a communication and display software.

    9. How the ground truth for the training set was established

    As there is no training set, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220976
    Date Cleared
    2022-07-21

    (108 days)

    Product Code
    Regulation Number
    868.2775
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Nihon Kohden Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Smart Cable NMT Module and Accessories are indicated for monitoring the relaxation of the patient when neuromuscular blockades are administered. The Smart Cable NMT Module and Accessories are intended to be used as a system that requires Nihon Kohden compatible electrodes and bedside monitoring systems. The Smart Cable NMT Module and Accessories are intended for use by medical personnel in clinical settings and are available by prescription only.

    The Life Scope® BSM-6000 Series Bedside Monitoring Systems are intended to monitor, display, and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signals produced by the electrocardiogram to generate visible and/or audible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, blood oxygen saturation (SpO2), non- invasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, BIS, cardiac output (CO), oxygen concentration (FiO2), carbon dioxide concentration (CO2), EtCO2, respiratory rate and inspired and expired anesthetic agents and anesthetic gases including N2O, halothane, isoflurane, enflurane, sevoflurane and desflurane. The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits. The device may also be used to condition and transmit physiological signals via radio frequency. The device can interface to external equipment to display numerical and waveform data and alarms from the external devices. Supported external devices include CO2 monitors, BIS monitors, Anesthetic agents/gases detection system, Anesthesia machine, Ventilators, CCO monitors, TOF monitors, CCO/SvO2 Monitors, EEG monitoring device, tcPO2/tcPCO2 monitors and external devices which output analog voltage signal. The device will be available for use by medical personnel on patients within a medical facility on all patient populations. The system requires a BSM-6000 core unit with a compatible input unit: AY Series or Life Scope® PT BSM-1700 Series.

    The Life Scope® CSM-1901 Bedside Monitoring Systems are intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signal produced by the heart and monitor the electrocardiogram to generate visible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, blood oxygen saturation (SpO2), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, BIS, cardiac output (CO), oxygen concentration (FiO2), carbon dioxide concentration (CO2), EtCO2, respiratory rate, inspired and expired anesthetic agents and anesthetic gases including N2O, halothane, isoflurane, enflurane and desflurane. The device also displays patient data from external devices such as ventilators, TOF monitors, and EEG measuring unit. The device may generate and audible and/or visual alarm when a measured rate falls outside preset limits. The device will be available for use by trained medical facility on all patient populations, including adult, neonate, infant, child, and adolescent subgroups. The system requires a CSM-1901 core unit with a compatible input unit: AY Series or Life Scope® PT BSM-1700 Series.

    The Life Scope® PT BSM-1700 Series Beside Monitor and Accessories are intended acquire and transfer electrical impulses from the patient to the main unit of the device. The BSM-1700 Series has both an input and transport/standalone mode. The BSM-1700 Series input unit monitors physiological data and may generate an audible and/or visual alarm when a measured rate falls outside preset limits or when disconnected from the core unit of the input unit can be removed from one core unit and connected to another device's core unit. The Life Scope® PT BSM-1700 Series can be used in transport mode where data is transferred from one device by using with or with or without WLAN technology. In standalone mode, the device does not require a core unit. The BSM-1700 Series can acquire the following parameter signals: Electrocardiogram (ECG), Impedance respiration (Imp Resp), Non-invasive blood pressure (NBP), Arterial oxygen saturation (SpO2), Carbon dioxide concentration (CO2), Invasive blood pressure (IBP), Temperature (Temp), Cardiac Output (CO), TOF and Bispectral Index (BIS). When the BSM-1700 Series is used in transport or standalone mode, the following can be analyzed and displayed: Electrocardiogram (ECG), Impedance respiration (Imp Resp), Non-invasive blood pressure (NIBP), Arterial oxygen saturation (SpO2), Carbon dioxide concentration (CO2), Invasive blood pressure (IBP), Temperature (Temp), Cardiac Output (CO), and Bispectral Index (BIS). The Data Acquisition Unit (DAU) is an optional accessory. The DAU is used to communicate between the compatible parent core unit and the input unit using connection cables. The keys on the DAU allow operation of the bedside monitor remotely. The DAU is only compatible with the Life Scope® PT BSM-1700 Series and the AY Series Input Units.

    AY Series and Accessories are intended to acquire and transfer electrical impulses from the patient to the core unit of the device. The input unit can acquire the following parameter signals: Electrocardiogram (ECG), Impedance respiration (Imp Resp), Non-invasive blood pressure (NIBP), Arterial oxygen saturation (SpO2), Carbon dioxide concentration (CO2), Invasive blood pressure (IBP), Temperature (Temp), Cardiac Output (CO), TOF and Bispectral Index (BIS). AA Series smart expansion unit adds additional MULTI sockets to an AY Series input unit and can only be used with compatible monitoring systems. The Data Acquisition Unit (DAU) is an optional accessory. The DAU is used to communicate between the compatible parent core unit and the input unit using connection cables. The keys on the DAU allow operation of the bedside monitor remotely. The DAU is only compatible with the Life Scope® PT BSM-1700 Series Input Units.

    Device Description

    The Nihon Kohden Smart Cable NMT Module (NMT Module) and Accessories is an optional accessory for the Nihon Kohden bedside monitoring systems. The Smart Cable NMT Module and Accessories TOF (Train of Four) are based on EMG technology. With this system, the user can apply electrical stimulation on the ulnar nerve to detect the muscle's action potential. The reaction to the electrical impulse can be visualized on the connected monitoring system. The Smart Cable NMT Module and Accessories can assist medical persomel to quantitatively determine the level of muscle relaxation. This information can be used to determine the dose of muscle relaxants and regional anesthetics when performing anesthesia in a clinical setting. It is intended for use by medical personnel in the operating room, or intensive care unit. NMT Module is a system comprised of NMT Module. Main Cable, Holder, and EMG Electrode. The NMT module is connected to an electrode via Main Cable. The electrode is a single-use electrode array, and each array includes two stimulating electrodes, two recording electrodes, and one ground electrode. TheNMT module can transmit an electrical stimulation pulse to the patient and can receive EMG signals via the electrode array. Thecaptured data from the disposable electrode is sent to the monitoring system via the Smart Cable interface connector. The various stimulation settings are a lso sent to the monitoring system to display. The AF-201P NMT Module is used to control the electrical stimulation and to measure theresponse. The operational setting is controlled via buttons on the module or a touch screen.

    The Life Scope® BSM-6000 Series Bedside Monitoring Systems are intended to monitor, display and record physiological data to provide cardial signs monitoring within a medical facility. The device is intended to produce a visualrecord of the electrical signals produced by the heart and monitor the electrocardiogram to generate visible and/or audible alarms when an arrhythmia exists. The device is a lso intended to monitor heart rate, pulse rate, blood oxygen saturation (SpO2), non- invasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, BIS, Cardiac Output (CO), oxygen concentration (FiO2), carbon dioxide concentration (CO2), EtCO2, respiratory rate and inspired and expired a nesthetic a gents and a nesthetic gases including CO2, O2, N2O, Halothane, Isoflurane, Sevoflurane and Desflurane. Anesthetic agents and gases are detected using the cleared AG-920R A Anesthetic Agent Detection System. The device can interface to external equipment to display numerical and waveform data and a larms from the external devices. Supported external devices include AG-920RA Anestheic Agent Detection System, Ventilators, CO2 Monitors, BIS Monitors, BIS Monitors, CCO/SvO2 Monitors and continuous NIBP Monitors. The devicemay generate anaudible and or visual alarm when a measured rate falls outside preset limits. This device may also be used to condition and transmit physiological signa ls via radio frequency. The system requires a BSM 6000 core unit with a compatible input unit: AY Series or Life Scope® PT BSM-1700 Series.

    The Life Scope® CSM-1901 Bedside Monitoring Systems which continuously monitors physiological information of a patient and is used in an operation room, a recovery room, general wards, ICU, CCU. HCU. NICU and an emergency room. These systems are placed near the patient and is intended to display patient's vital signs. These systems can also be connected to other external patient monitoring devices. In a ddition, these systems can communicate patient's data to a central monitoring station via network to monitor multiple patients. The input unit is common to NK parent devices that require both a core unit and input unit. NK manufactures the input units with three (3) SpO2 options. The Life Scope® CSM-1 901 Bedside Monitoring Systems have interchangeable input units that contains the MULTI socket ports. For larger monitoring systems is data a cquisition unit is required to transmit data from the input unit to the core unit. The bedside monitoring systems require both a coreunit and an input unit interprets the electrical impulses from the patient's body and transfers this data into the core unit calculates the electrical impulses. Each monitor has a color display and is intended for one patient. The intended populations are all patient populations under the care of health professionals.

    The AY Series Input unit is used with the monitoring systems platforms, when connected to the core unit of the parent device, the inputunit collects electrical impulses, and the core units calculates and displays on the core unit's screen.

    The BSM-1700 Series Bedside Monitor is a multifunctional device used as an input unit, transport/standalone monitor.
    · Input unit for other monitoring systems platforms, when connected to the core unit of the parent device, the input unit collects electrical impulses, and the core units calculates and displays on the core unit's screen.
    · When the patient needs to be transported, the BSM-1700 Series can be removed from the core unit, transport mode can be enabled and can be used with or without WLAN. When WLAN is enabled real time data viewing on the Nihon Kohden network or if it is disabled the BSM-1700 Series will display monitoring data and store the review data.
    • When the device is removed from the coreunit it functions as a standalone or independent monitoring system.
    The BSM-1700 Series has a display monitor that is disabled when connect to a core unit. Each monitor has a cobr display and is intended for one patient. When used as an inputunit with the coreunit, the system monitors a dvanced parameters. The intended populations are all patient populations under the care of health professionals. In all modes, the BSM-1700 Series uses the Smart Cable technology that is used to connect to other accessories used to collect electrical impulses. The BSM-1700 Series interprets the electrical impulses from the patient's body. When connected to a core unit, advanced calculations can be achieved. The device may generate a naudible and/or visual a larm when a measured rate falls outside preset lim its.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and related study information, based on the provided text:

    Important Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device, not necessarily a detailed clinical study report proving performance against novel acceptance criteria. In this specific case, the submission emphasizes that there are no new acceptance criteria because the device (Life Scope PT BSM-1700 Series Bedside Monitor) is being updated to include a "standalone mode," but its underlying technology and performance characteristics are unchanged from its previously cleared version (K213316). Therefore, the "study that proves the device meets the acceptance criteria" largely relies on the prior clearance and non-clinical testing for the updated mode.


    1. Table of Acceptance Criteria and Reported Device Performance

    Since the document explicitly states there are no new technological changes or safety/performance claims, the acceptance criteria are effectively the performance specifications of the predicate device (K213316). The reported device performance is stated to be "the same" as the predicate.

    Characteristic/Parameter (Acceptance Criteria)Reported Device Performance (Subject Device: Life Scope BSM-1700 Series Bedside Monitor)
    Intended Use/Indications for UseExpanded to include standalone mode functionality for parameters as listed below. (Note: This is a labeling update, not a change in underlying performance for already cleared functions.)
    Compatible Core Unit OptionsSame as predicate (Life Scope® BSM-6000, CSM-1901, G5, G7 Series)
    BSM-1700 Touchscreen Display (Input Unit)Disabled
    Measuring Parameters (Core unit + Input)ECG, Imp Resp, NIBP, SpO2, CO2, IBP, Temp, CO, TOF, BIS
    Touchscreen Display (Standalone/Transport)Enabled
    Resolution (Standalone/Transport)640 x 480
    Number of Traces (Standalone/Transport)9, 12 (when 12 leads ECG)
    Waveform Display (Standalone/Transport)ECG, Imp Resp, NIBP, SpO2, CO2, IBP, Temp, CO, BIS
    Numeric Data Display (Standalone/Transport)Heart Rate, Pulse Rate, VPC rate, ST level, Respiration Rate, NIBP (sys/dia/mean), Temp (2), SpO2, EtCO2, FiCO2, Cardiac Output, BIS, IBP (3) (sys/dia/mean)
    Alarm DisplayAlarm sound, highlighted alarm display, alarm lamp
    Alarm Suspend/SilenceYes
    Number of ECG Electrodes3, 6 or 10
    Defibrillation Discharge ProtectionYes
    Electrosurgery Interface filterYes
    Pacing detectionYes
    Heart Rate Counting MethodAverage, instantaneous (Beat to beat)
    Heart Rate Counting Range0, 15 to 300 bpm
    Heart Rate Counting Accuracy± 2 bpm
    ST Level Measuring Range-2.5 to +2.5 mV
    VPC Counting Range0 to 99 per min
    Arrhythmia AlarmYes
    Arrhythmia Recall (Standalone Mode)32768 items, 72 hours
    Respiration Rate Display0 to 150 bpm
    SpO2 Declared Range, Accuracy (NK type)70 to 100% (with sensor), 80 to 100% ± 2 %SpO2, 70 to 80% ± 3 %SpO2
    NIBP Measuring MethodOscillometric
    NIBP Measuring Range (Adult/Pediatric)10 - 280 mmHg
    NIBP Measuring Accuracy± 3 mmHg
    IBP Measuring Range-50 to 300 mmHg
    IBP Measuring Accuracy±1mmHg±1digit (-50 to 100 mmHg), ±1%±1 digit (100 to 300 mmHg)
    Temperature Measuring Range0 to 45°C
    Temperature Accuracy±0.1 °C (25 to 45°C), ± 0.2 °C (0 to 25°C)
    CO2 Measuring RangeDepends on CO2 unit (e.g., 0 to 100mmHg for TG-900P/920P)
    Cardiac Output Measuring Range0.5 to 20 L/min
    Cardiac Output Accuracy±5%
    Full Disclosure Storage (Standalone Mode)72 hours, 5 waves
    Trend Display Time (Standalone Mode)1, 2, 4, 8, 24, 72 hours
    Battery Operation Time5 hours
    General SafetyMeets IEC 60601-1:2005+Amd.1:2012, ANSI/AAMI/ES 60601-1:2005 (R2012)
    EMCMeets IEC60601-1-2:2014

    Study Proving Device Meets Acceptance Criteria

    The document states:

    • "The device performance and software have not changed from the original submission K213316." (Sections 2.4, 3.4, 4.4, 5.4)
    • "For this submission the BSM-1700 Bedside Monitor has been updated to include the standalone mode. There are no specification changes, technological changes or safety and Performance claims. The indications for use and labeling have been updated to include the standalone mode." (Section 6)
    • "The Life Scope BSM-1700 Series Bedside Monitor was subjected to tests to electromagnetic, environmental, safety, and performance testing procedures. These tests verified the operation of the device. The software validation tested the operation of the software function of the device, the results confirmed that the device performed within specifications." (Section 6.3 Safety & Performance Tests)

    Given this, the "study" is primarily an affirmation that the previously cleared performance for K213316 remains valid, coupled with non-clinical verification and validation testing of the additional "standalone mode" functionality to ensure it operates within the established specifications without altering the core performance characteristics.

    1. Sample size used for the test set and the data provenance:

      • The document does not specify a sample size for a clinical test set in this 510(k) summary. This type of submission relies heavily on non-clinical (bench) testing to confirm performance when significant changes are not made.
      • The data provenance is implied to be from Nihon Kohden's internal testing. The submission is focused on demonstrating substantial equivalence rather than presenting a de novo clinical study with patient data. As such, the data would be laboratory-generated from testing the device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • This information is not provided because no clinical study requiring expert ground truth establishment for a test set is detailed. The "ground truth" for non-clinical performance would be derived from established engineering and metrological standards and test procedures.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This is not applicable since the submission focuses on non-clinical performance and engineering validation rather than human reader studies.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This device is a physiological monitor, not an AI-powered diagnostic tool, and the submission does not mention AI assistance.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • The term "standalone" in the provided document refers to a mode of operation for the BSM-1700 device, where it functions independently without being connected to a core unit. It does not refer to a standalone performance study of an algorithm.
      • However, the performance tests described in Section 6.3 were likely conducted in a "standalone" fashion for the device's functional parameters, meaning the device itself was tested to ensure its outputs (e.g., heart rate, SpO2 accuracy) met specifications. The submission highlights that "The device performance and software have not changed from the original submission K213316," implying that the standalone performance requirements were met by the previous device and confirmed for this update.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the non-clinical testing, the ground truth would be based on calibrated reference standards and established measurement methodologies for physiological parameters (e.g., known electrical signals for ECG, certified NIBP simulators, calibrated temperature baths, etc.).
      • No clinical ground truth (like pathology or outcomes data) is described in this submission, as it's not a clinical study.
    7. The sample size for the training set:

      • This information is not provided. Medical devices like this (physiological monitors) typically undergo engineering development, verification, and validation testing rather than a "training set" in the machine learning sense. Any underlying algorithms (e.g., for signal processing, arrhythmia detection) would have been developed and validated during earlier stages or for previous clearances.
    8. How the ground truth for the training set was established:

      • Not applicable for this type of device and submission. If any algorithms involved "training," the ground truth would have been established using rigorously collected physiological data and expert annotation/labeling, but this detail is not present in the 510(k) summary.
    Ask a Question

    Ask a specific question about this device

    K Number
    K213521
    Date Cleared
    2022-07-01

    (240 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Nihon Kohden OrangeMed, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nihon Kohden NKV-330 Ventilator is intended to provide ventilation and oxygen concentration for patients who are breathing spontaneously but need partial ventilation support due to respiratory failure or chronic respiratory insufficiency. It is intended for children weighing 12.5 kg or greater to adult patients. It offers noninvasive ventilation, invasive ventilation, and respiratory monitoring. The NKV-330 is intended for use in hospital-type facilities, and in-hospital transportation by qualified and trained users under the directions of a physician.

    Device Description

    The NKV-330 is a servo-controlled ventilator that is designed to meet the gas delivery and performance requirements for pediatric through adult patients. The NKV-330 design is comprised of two major components, a Breath Delivery Unit (BDU) and a Graphic User Interface (GUI). The GUI allows clinicians to set ventilator control parameters such as PEEP and inspiratory pressure, to set alarm limits such as high inspiratory pressure alarm, to view monitored numeric values, and to view waveforms. The BDU assembly contains a blower and the electronics required to perform breath delivery. Ambient air is taken into the blower and mixed with oxygen which is flow rate controlled by a proportional valve. The mixed gas is provided to the patient. The microprocessor controls the blower and the proportional valve to deliver the pressure and oxygen concentration which are set by the user. It also provides various alarms and other design features to maximize patient safety.

    AI/ML Overview

    The provided document is a 510(k) summary for the Nihon Kohden NKV-330 Ventilator System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that a device meets specific acceptance criteria through a clinical study or detailed performance metrics.

    Therefore, the document does not contain the specific acceptance criteria, reported device performance in those terms, details of a specific study proving it, sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone algorithm performance, or ground truth details as requested.

    The document primarily states that the device's technical characteristics are "substantially equivalent" to a predicate device (Philips/Respironics V60 Ventilator) and lists various non-clinical performance data and standards compliance to support this claim. It explicitly states "Clinical performance data was not required to demonstrate substantial equivalence."

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    Nihon Kohden Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Smart Cable NMT Module and Accessories are indicated for monitoring the relaxation of the patient when neuromuscular blockades are administered.

    The Smart Cable NMT Module and Accessories are intended to be used as a system that requires Nihon Kohden compatible electrodes and bedside monitoring systems. The Smart Cable NMT Module and Accessories are intended for use by medical personnel in clinical settings and are available by prescription only.

    The Life Scope® BSM-6000 Series Bedside Monitoring Systems are intended to monitor, display, and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signals produced by the heart and monitor the electrocardiogram to generate visible and/or audible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, blood oxygen saturation (SpO2), non- invasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, BIS, cardiac output (CO), oxygen concentration (FiO2), carbon dioxide concentration (CO2), EtCO2, respiratory rate and inspired anesthetic agents and anesthetic gases including N2O, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane. The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits. The device may also be used to condition and transmit physiological signals via radio frequency. The device can interface to external equipment to display numerical and waveform data and alarms from the external devices. Supported external devices include CO2 monitors, BIS monitors, Anesthetic agents/gases detection system, Anesthesia machine, Ventilators, CCO monitors. TOF monitors, CCO/SvO2 Monitors, EEG monitoring device, tcPO2/tcPCO2 monitors, and external devices which output analog voltage signal. The device will be available for use by medical personnel on patients within a medical facility on all patient populations. The system requires a BSM-6000 core unit with a compatible input unit: AY Series or Life Scope® PT BSM-1700 Series.

    The Life Scope® CSM-1901 Bedside Monitoring Systems are intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signal produced by the heart and monitor the electrocardiogram to generate visible alarms when an arthythmia exists. The device is also intended to monitor heart rate, blood oxygen saturation (SpO2), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, BIS, cardiac output (CO), oxygen concentration (FiO2), carbon dioxide concentration (CO2), EtCO2, respiratory rate, inspired and expired anesthetic agents and anesthetic gases including N20, halothane, isoflurane, enflurane and desflurane. The device also displays patient data from external devices such as ventilators. TOF monitors, and EEG measuring unit. The device may generate and audible and/or visual alarm when a measured rate falls outside preset limits. The device will be available for use by trained medical facility on all patient populations, including adult, neonate, infant, child, and adolescent subgroups. The system requires a CSM-1901 core unit with a compatible input unit: AY Series or Life Scope® PT BSM-1700 Series.

    The Life Scope® PT BSM-1700 Series and Accessories are intended acquire and transfer electrical impulses from the patient to the main unit of the device. The BSM-1700 Series input unit monitors physiological data and may generate an audible and/or visual alarm when a measured rate falls outside preset limits when discomected from the core unit of the device. The input unit can be removed from one core unit and connected to another devices core unit. The Life Scope® PT BSM-1700 Series can be used in transport mode where data is transferred from one device to another device by using with or without WLAN technology. The input unit can acquire the following parameter signals: Electrocardiogram (ECG), Impedance respiration (Imp Resp), Non-invasive blood pressure (NIBP), Arterial oxygen saturation (SpO2), Carbon dioxide concentration (CO2), Invasive blood pressure (Temp), Cardiac Output (CO), TOF and Bispectral Index (BIS). The Data Acquisition Unit (DAU) is an optional accessory. The DAU is used to communicate between the compatible parent core unit and the input unit using connection cables. The Reys on the DAU allow operation of the bedside monitor remotely. The DAU is only compatible with the Life Scope® PT BSM-1700 Series and the AY Series Input Units.

    AY Series and Accessories are intended to acquire and transfer electrical impulses from the patient to the core unit of the device. The input unit can acquire the following parameter signals: Electrocardiogram (ECG), Impedance respiration (Imp Resp), Non-invasive blood pressure (NIBP), Arterial oxygen saturation (SpO2), Carbon dioxide concentration (CO2), Invasive blood pressure (IBP), Temperature (Temp), Cardiac Output (CO), TOF and Bispectral Index (BIS). AA Series smart expansion unit adds additional MULTI sockets to an AY Series input unit and can only be used with compatible monitoring systems. The Data Acquisition Unit (DAU) is an optional accessory. The DAU is used to communicate between the compatible parent core unit and the input unit using connection cables. The keys on the DAU allow operation of the bedside monitor remotely. The DAU is only compatible with the Life Scope® PT BSM-1700 Series Input Units.

    Device Description

    The Nihon Kohden Smart Cable NMT Module (NMT Module) and Accessories is an optional accessory for the Nihon Kohden bedside monitoring systems. The Smart Cable NMT Module and Accessories TOF (Train of Four) are based on EMG technology. With this system, the user can apply electrical stimulation on the ulnar nerve to detect the muscle's action potential. The reaction to the electrical impulse can be visualized on the connected monitoring system. The Smart Cable NMT Module and Accessories can assist medical personnel to quantitatively determine the level of muscle relaxation. This information can be used to determine the dose of muscle relaxants and regional anesthetics when performing anesthesia in a clinical setting. It is intended for use by medical personnel in the operating room, recovery room, or intensive care unit. NMT Module is a system comprised of NMT Module, Main Cable, Holder, and EMG Electrode. The NMT module is connected to an electrode via Main Cable. The electrode is a single-use electrode array and each array includes two stimulating electrodes, two recording electrodes, and one ground electrode The NMT module can transmit an electrical stimulation pulse to the patient and can receive EMG signals via the electrode array. The captured data from the disposable electrode is sent to the monitoring system via the Smart Cable interface connector. The various stimulation settings are also sent to the monitoring system to display. The AF-201P NMT Module is used to control the electrical stimulation and to measure the response. The operational setting is controlled via buttons on the module or a touch screen.

    The Life Scope BSM-6000 Series Bedside Monitoring Systems are intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signals produced by the heart and monitor the electrocardiogram to generate visible and/or audible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, pulse rate, blood oxygen saturation (SpO2), non- invasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, Cardiac Output (CO), oxygen concentration (FiO2), CO2 and EtCO2, respiratory rate, BIS and inspired and expired anesthetic agents and gases including CO2, O2, N2O, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane. Anesthetic agents and gases are detected using the cleared AG-920RA Anesthetic Agent Detection System. The device can interface to external equipment to display numerical and waveform data and alarms from the external devices. Supported external devices include AG-920RA Anesthetic Agent Detection System, Ventilators, CO2 Monitors, TOF Monitors, BIS Monitors, CCO/SvO2 Monitors and continuous NIBP Monitors. The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits. This device may also be used to condition and transmit physiological signals via radiofrequency. The system requires a BSM 6000 core unit with a compatible input unit: AY Series or Life Scope® PT BSM-1700 Series.

    The Life Scope® CSM-1901 Bedside Monitoring Systems are systems which continuously monitors physiological information of a patient and is used in an operation room, a recovery room, general wards, ICU, CCU, HCU, NICU and an emergency room. These systems are placed near the patient and is intended to display patient's vital signs. These systems can also be connected to other external patient monitoring devices. In addition these systems can communicate patient's data to a central monitoring station via network to monitor multiple patients. The input unit is common to NK parent devices that require both a core unit and input unit. NK manufactures the input units with three (3) SpO2 options. The Life Scope® CSM-1901 Bedside Monitoring Systems have interchangeable input units that contains the MULTI socket ports. For larger monitoring systems is data acquisition unit is required to transmit data from the input unit to the core unit. The bedside monitoring systems require both a core unit and an input unit The input unit interprets the electrical impulses from the patient's body and transfers this data into the core unit. The core unit calculates the electrical impulses. Each monitor has a color display and is intended for one patient. The intended populations are all patient populations under the care of health professionals.

    AI/ML Overview

    The provided text is a 510(k) summary for the Nihon Kohden Life Scope PT BSM-1700 Series, Data Acquisition Unit, LIFE SCOPE BSM 6000 SERIES BEDSIDE MONITORING SYSTEM, and Nihon Kohden CSM-1901 BEDSIDE MONITORING SYSTEM. It describes software modifications to enable compatibility with the Smart Cable NMT Module and Accessories, and updates to the stated Indications for Use.

    The document does not contain information about an AI/algorithm-only study, a multi-reader multi-case (MRMC) comparative effectiveness study, or details on ground truth establishment for a training set. The device in question is a physiological monitor, and the software modifications are specifically related to integrating with a neuromuscular transmission (NMT) module and updating intended use statements, not for an AI/CADe (Computer-Aided Diagnosis/Detection) algorithm that would typically require such studies.

    Therefore, many of the requested criteria related to AI/algorithm performance and clinical validation studies are not applicable to the information provided in this document.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the software modifications do not change the safety, performance of the predicate devices. It refers to integration testing, not a de novo performance study with specific quantifiable acceptance criteria for a new AI algorithm.

    Acceptance Criteria (Implied)Reported Device Performance
    Device functions as intended with new software.Confirmed operation of the device when the Smart Cable NMT software was integrated into the system. Results confirmed the device performed within specifications.
    Software modifications do not raise different questions of safety and effectiveness.Software input units modifications and labeling modification do not raise different questions of safety and effectiveness.
    Compliance with relevant electrical safety and EMC standards.Complies with ANSI AAMI ES60601-1 / IEC 60601-1 for safety and IEC 60601-1-2 for EMC.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in terms of patient data. The testing mentioned is "software unit testing, integration level testing, and system-level testing" and a "system test... based on the software requirements specification." This suggests internal product testing rather than a separate clinical test set of patient data.
    • Data Provenance: Not applicable, as no external clinical data test set is described. The focus is on the integration and functional testing of the software within the device system.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The ground truth, in this context, would be the proper functioning and integration of the software and hardware, verified through engineering and system testing, not through expert reading of medical images or data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This is typically used in clinical studies for establishing ground truth for AI algorithms, which is not the subject of this 510(k).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. The document explicitly states: "No clinical tests have been submitted, referenced or relied on in this premarket notification submission for a determination of substantial equivalence." This device is not an AI-assisted diagnostic or therapeutic tool; it's a physiological monitor with updated software for compatibility.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI algorithm requiring standalone performance evaluation. The "software modification" refers to updates enabling the NMT module to work with the monitoring system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the software and system functionality: The ground truth is the predefined technical specifications and expected behavior of the device and its integrated components. This is verified through "software unit testing, integration level testing, and system-level testing."

    8. The sample size for the training set:

    • Not applicable. There's no machine learning model or AI algorithm described that would require a "training set."

    9. How the ground truth for the training set was established:

    • Not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    Nihon Kohden Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nihon Kohden Life Scope® G5 and Nihon Kohden Life Scope® G7 Bedside Monitoring System are intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signal produced by the heart and monitor the electrocardiogram to generate visible and/or audible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, pulse rate, blood oxygen saturation (SpO2), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, BIS, cardiac output (CO), oxygen concentration (02), carbon dioxide concentration (CO2), EtCO2, respiratory rate, inspired and expired anesthetic agents and anesthetic gases including N20, halothane, isoflurane, enflurane, sevoflurane, and desflurane. The device also displays patient data from external devices such as ventilators, TOF modules, CCO/SvO2 monitors, and EEG measuring units.

    The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits.

    The device will be available for use by trained medical personnel within a medical facility on all patient populations, including adult, neonate, infant, child, and adolescent subgroups.

    A-fib detection, ST measurement and QTc/QRSd monitoring are intended for adult patients only. Arrhythmia detection function is intended for child, adolescent, and adult patients.

    Device Description

    The Nihon Kohden Life Scope® G5 and Life Scope® G7 Bedside Monitoring System are an LCD touchscreen bedside monitoring system. These bedside monitors are installed near the patient and are intended to display the patient's vital signs such as ECG (basic and 12 lead), NIBP, temperature, SpO2, respiration, and CO2 and generate alarms from the bedside monitor. Additional parameters can be measured such as arrhythmia detection, ST elevation, and Train of Four (TOF) measuring parameters. Apnea and arrhythmia can also be monitored. The configuration of the bedside monitor can be adapted by the health care professionals to meet the clinical setting requirements.

    The bedside monitoring systems require both a core unit and an input unit. The input unit interprets the electrical impulses from the patient's body and transfers this data into the core unit. The core unit calculates the electrical impulses. Each monitor has a color display and is intended for one patient. The intended populations are all patient populations under the care of health professionals.

    The bedside monitor is designed so the operator can directly touch the screen from the operator position. Other optional accessories can also be used with the bedside monitor to add other parameters, allowing it to be used in a wide range of sites, such as operating rooms and intensive care units (ICU). The bedside monitor can also be connected to a network to communicate with central monitors and other Nihon Kohden devices.

    The Life Scope® G5 Bedside Monitoring System consists of two models, those models are offered in two sizes:

    • CSM-1501 bedside monitoring with core unit (CU) model CU-151R: 12.1-inch . display
    • CSM-1502 bedside monitoring with core unit (CU) model CU-152R: 15.6-inch . displav

    The Life Scope G7 Bedside Monitoring System consists of two models, those models are offered in two sizes:

    • CSM-1701 bedside monitoring with core unit (CU) model CU-171R: 15.6-inch . displav
    • CSM-1702 bedside monitoring with core unit (CU) model CU-172R: 19.0-inch . display

    The Life Scope G7 Bedside Monitoring System consists of an input unit and a data acquisition unit with either the CU-171R or the CU-172R core unit.

    AI/ML Overview

    Here is an analysis of the acceptance criteria and supporting study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission primarily focuses on establishing substantial equivalence to a predicate device rather than defining specific performance-based acceptance criteria for novel claims. The acceptance criteria are implicitly derived from the predicate device's specifications and compliance with recognized standards. Therefore, the "reported device performance" is essentially a statement of "identical" or "below" (meaning lower capacity/resolution but deemed not to affect safety/effectiveness) compared to the predicate.

    CharacteristicAcceptance Criteria (from Predicate)Subject Device (Life Scope G5/G7) PerformanceComparison
    GeneralIdentical to CSM-1901 (K201949)Identical to CSM-1901 (K201949) in Classification Panel, Regulation Number, Classification Name, Regulatory Class, Product Code, Patient Population, Setting, End-User, Biocompatibility, Shelf Life, Patient Contact, Single-Use.Identical
    Indications for UseMonitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. Monitor ECG to generate alarms when arrhythmia exists. Monitor heart rate, pulse rate, SpO2, NIBP, IBP, body temperature, BIS, CO, O2, CO2, EtCO2, respiratory rate, inspired/expired anesthetic gases. Display patient data from external devices. Generates audible/visual alarms when a measured rate falls outside preset limits. For trained medical personnel, all patient populations (adult, neonate, infant, child, adolescent). A-fib detection, ST measurement, QTc/QRSd monitoring for adult; Arrhythmia detection for child, adolescent, adult.Identical to Predicate K201949. Patient population for each ECG analysis functionality clarified.Identical
    ECG
    QTc/QRSd MonitoringYes (Spot-Check)Yes (Spot-Check) and Continuous monitoring (Automatically started when ECG monitoring begins).Below (Predicate had only spot-check, subject device adds continuous, but deemed not to affect safety/effectiveness)
    Arrhythmia Recall Files20,000 files16,384 filesBelow (16,384
    Ask a Question

    Ask a specific question about this device

    Page 1 of 9