(240 days)
The Nihon Kohden NKV-330 Ventilator is intended to provide ventilation and oxygen concentration for patients who are breathing spontaneously but need partial ventilation support due to respiratory failure or chronic respiratory insufficiency. It is intended for children weighing 12.5 kg or greater to adult patients. It offers noninvasive ventilation, invasive ventilation, and respiratory monitoring. The NKV-330 is intended for use in hospital-type facilities, and in-hospital transportation by qualified and trained users under the directions of a physician.
The NKV-330 is a servo-controlled ventilator that is designed to meet the gas delivery and performance requirements for pediatric through adult patients. The NKV-330 design is comprised of two major components, a Breath Delivery Unit (BDU) and a Graphic User Interface (GUI). The GUI allows clinicians to set ventilator control parameters such as PEEP and inspiratory pressure, to set alarm limits such as high inspiratory pressure alarm, to view monitored numeric values, and to view waveforms. The BDU assembly contains a blower and the electronics required to perform breath delivery. Ambient air is taken into the blower and mixed with oxygen which is flow rate controlled by a proportional valve. The mixed gas is provided to the patient. The microprocessor controls the blower and the proportional valve to deliver the pressure and oxygen concentration which are set by the user. It also provides various alarms and other design features to maximize patient safety.
The provided document is a 510(k) summary for the Nihon Kohden NKV-330 Ventilator System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that a device meets specific acceptance criteria through a clinical study or detailed performance metrics.
Therefore, the document does not contain the specific acceptance criteria, reported device performance in those terms, details of a specific study proving it, sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone algorithm performance, or ground truth details as requested.
The document primarily states that the device's technical characteristics are "substantially equivalent" to a predicate device (Philips/Respironics V60 Ventilator) and lists various non-clinical performance data and standards compliance to support this claim. It explicitly states "Clinical performance data was not required to demonstrate substantial equivalence."
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 1, 2022
Nihon Kohden OrangeMed, Inc. Jacqueline Villanueva Quality Engineering Manager 1800 E. Wilshire Avenue Santa Ana, California 92705
Re: K213521
Trade/Device Name: Nihon Kohden NKV-330 Ventilator System Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous ventilator Regulatory Class: Class II Product Code: MNT Dated: June 1, 2022 Received: June 3, 2022
Dear Jacqueline Villanueva:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brandon Blakely, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213521/S002
Device Name Nihon Kohden NKV-330 Ventilator System
Indications for Use (Describe)
The Nihon Kohden NKV-330 Ventilator is intended to provide ventilation and oxygen concentration for patients who are breathing spontaneously but need partial ventilation support due to respiratory failure or chronic respiratory . It is intended for children weighing 12.5 kg or greater to adult patients. It offers noninvasive ventilation, invasive ventilation, and respiratory monitoring. The NKV-330 is intended for use in hospital-type facilities, and inhospital transportation by qualified and trained users under the directions of a physician.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Nihon Kohden OrangeMed, Inc. The logo consists of a blue graphic element followed by the text "NIHON KOHDEN" in blue. Below this, the text "Nihon Kohden OrangeMed, Inc." is written in orange. The logo is simple and clean, with a focus on the company name.
510(k) Summary
1. Date
July 1, 2022
2. Submitter / Manufacturing Location
Nihon Kohden OrangeMed, Inc. 1800 E. Wilshire Avenue Santa Ana, CA 92705 USA
Company Contact 3.
Jacqueline Villanueva – Quality Assurance Director Email: jvillanueva@orange-med.com Phone: (949) 502-6448 x7045
Common Name of Device 4.
Assist Ventilator
Trade Name ട്.
Nihon Kohden NKV-330 Ventilator System
6. Classification Name
| Product Code: | MNT |
|---|---|
| Ventilator, Continuous, Minimal Ventilatory Support, Facility Use | |
| Regulation Number: | 21 CFR 868:5895 |
| Device Class: | II |
| Review Panel: | Anesthesiology |
7. Predicate Device
Predicate Device
| Product: | V60 Ventilator with PPV and Auto-Trak + Software Options |
|---|---|
| 510k #: | K102985 |
| Manufacturer: | Respironics California, LLC |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Nihon Kohden OrangeMed, Inc. The logo consists of the text "NIHON KOHDEN" in blue, with a stylized blue graphic to the left. Below the blue text is the text "Nihon Kohden OrangeMed, Inc." in orange. The logo is simple and clean, with a focus on the company name.
Reference Device
| Product: | NKV-550 Series Ventilator System |
|---|---|
| 510k #: | K181695 |
| Manufacturer: | Nihon Kohden OrangeMed, Inc. |
8. Description of Device
The NKV-330 is a servo-controlled ventilator that is designed to meet the gas delivery and performance requirements for pediatric through adult patients. The NKV-330 design is comprised of two major components, a Breath Delivery Unit (BDU) and a Graphic User Interface (GUI). The GUI allows clinicians to set ventilator control parameters such as PEEP and inspiratory pressure, to set alarm limits such as high inspiratory pressure alarm, to view monitored numeric values, and to view waveforms. The BDU assembly contains a blower and the electronics required to perform breath delivery. Ambient air is taken into the blower and mixed with oxygen which is flow rate controlled by a proportional valve. The mixed gas is provided to the patient. The microprocessor controls the blower and the proportional valve to deliver the pressure and oxygen concentration which are set by the user. It also provides various alarms and other design features to maximize patient safety.
Product Intended Function 9.
The Nihon Kohden NKV-330 Ventilator is an assist ventilator intended to support adult and pediatric patients who are breathing spontaneously and require a tidal volume of 100mL or more with an oxygen concentration of 21 to 100%. It is also intended for intubated patients meeting the same selection criteria as the noninvasive applications.
10. Indication for Use (Intended Medical Indication)
The Nihon Kohden NKV-330 Ventilator is intended to provide ventilation and oxygen concentration for patients who are breathing spontaneously but need partial ventilation support due to respiratory failure or chronic respiratory insufficiency. It is intended for children weighing 12.5 kg or greater to adult patients. It offers noninvasive ventilation, invasive ventilation, and respiratory monitoring. The NKV-330 is intended for use in hospital-type facilities, and in-hospital transportation by qualified and trained users under the directions of a physician.
11. Intended Patient Population
Adult and pediatric patients that are breathing spontaneously but need partial ventilation support through invasive or non-invasive ventilation.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for NIHON KOHDEN. The logo consists of a blue abstract symbol on the left, resembling a stylized wave or mountain range. To the right of the symbol is the company name, "NIHON KOHDEN," also in blue. The font is sans-serif and appears to be bolded.
Nihon Kohden OrangeMed, In
12. Intended Part of the Body or Type of Tissue Applied to or Interacted with
The device is intended for use with a breathing circuit attached to a patient interface for noninvasive ventilation or patients that are under invasive ventilation and meet the selection criteria for noninvasive application. For noninvasive ventilation, the part of the body or type of tissue applied to or interacted with would include a facial or nasal cannula, applied to the mouth and/or nose. For invasive ventilation, the part of the body or type of tissue applied to or interacted with would include an endotracheal tube inserted through the mouth or the nose or a tracheostomy tube inserted into the trachea.
13. Intended Environment of Use (Use Environment)
The device is intended for use in hospitals and hospital-type facilities, which provide respiratory care for patients requiring respiratory support.
The device may be used for intra-hospital transport within a hospital-type facility. The device is not intended for transport between hospital-type facilities, therefore is not a transport ventilator as defined by ISO 80601-2-12 Section 201.1.1.
The device is not to be used in the presence of flammable anesthetics and MRI applications.
{6}------------------------------------------------
14. Accessories
The NKV-330 is designed to be used with ventilator support accessories, such as, filters and breathing circuits. Optional accessories have been tested to be compatible with the NKV-330.
| Accessory | Description | Manufacturer | Model # Tested for SystemCompatibility |
|---|---|---|---|
| Respiratory GasHumidifier | F&P MR850Respiratory Humidifier | Fisher Paykel | MR850 |
| Heated Single LimbBreathing Circuit | Fisher & Paykel Single LimbAdult Circuit Kit with Chamber | Fisher Paykel | RT219 |
| Heated Single LimbBreathing Circuit | Fisher & Paykel Single LimbAdult Circuit Kit with Chamber | Fisher Paykel | RT202 |
| Breathing CircuitBacterial Filter | F&P Inspiratory/ ExpiratoryFilter | Fisher Paykel | RT019 |
| Nasal Cannula | Teleflex Comfort Flow PlusCannula | Teleflex | 2412-13 (S)2412-12 (M)2412-11 (L) |
| NIV face mask | 6700 Series Non-vented FullFace Mask | Hans Rudolph | 113550 (L), 113551 (M),113552 (S), 113553 (XS),113554 (Petite) |
| NIV face mask | Respironics PerforMax PediatricSE Total Face Mask | Philips | 1119026 (XS) |
| NIV Face Mask Set | NPPV Cap-ONE Mask Set | Nihon KohdenOrangeMed | MSK3313P (L),MSK3314P (M),MSK3315P (S),MSK3316P (XS) |
| SPO2 Monitor | Nihon Kohden SpO2 MonitorKit | Nihon Kohden Corp. | JL-500P1 |
| SPO2 Sensor | Nihon Kohden Finger Probe | Nihon Kohden Corp. | TL-201T |
| SPO2 Sensor | Nihon Kohden Multi-site Probe | Nihon Kohden Corp. | TL-220T |
| CO2Monitor | Nihon Kohden CO2 Monitor Kit | Nihon Kohden Corp. | TG-980P |
| CO2Airway Adapter | Nihon Kohden CO2 AirwayAdapter | Nihon Kohden Corp. | YG-211T |
| On Airway FlowSensor | EnviteC SpiroQuant H | EnviteC | TF-330Z(E07-00-0001) |
| Exhalation Port | Exhalation Port | Nihon KohdenOrangeMed | EXH3301P |
{7}------------------------------------------------
15. Summary of Technical Characteristics
The technological characteristics of the Nihon Kohden NKV-330 Ventilator System are substantially equivalent to the predicate device as compared and summarized in the table below.
*Reference for predicate device information: "Respironics V60 Ventilator User Manual" (1047358 Rev C) and Philips V60 Ventilator Specification (Oct 2016).
| TechnicalCharacteristic | Nihon Kohden NKV-330Ventilator System | Philips/Respironics V60Ventilator* (PredicateDevice – K102985) | Comparison |
|---|---|---|---|
| Indication for Use | The Nihon Kohden NKV-330 Ventilator is intendedto provide ventilation andoxygen concentration forpatients who arebreathing spontaneouslybut need partial ventilationsupport due to respiratoryfailure or chronicrespiratory insufficiency. Itis intended for childrenweighing 12.5 kg or greaterto adult patients. It offersnoninvasive ventilation ,invasive ventilation , andrespiratory monitoring. TheNKV-330 is intended foruse in hospitals, hospital-type facilities, and in-hospital transportationby qualified and trainedusers under the directionsof a physician. | The Respironics V60Ventilator is an assistventilator and is intended toaugment patient breathing.It is intended forspontaneously breathingindividuals who requiremechanical ventilation:patients with respiratoryfailure, chronic respiratoryinsufficiency, or obstructivesleep apnea in a hospital orother institutional settingsunder the direction of aphysician.The ventilator is intended tosupport pediatric patientsweighing 20 kg (44 lb)or greater to adult patients.It is also intended forintubated patientsmeeting the same selectioncriteria as the noninvasiveapplications . The ventilatoris intended to be used byqualified medicalprofessionals, such asphysicians, nurses, andrespiratory therapists.The ventilator is intended tobe used only with variouscombinations ofRespironics-recommendedpatient circuits,interfaces (masks),humidifiers, and other | Substantially equivalentBoth ventilators areintended for patients whohave spontaneous breathes.Both ventilators can beused for invasiveventilation (i.e., throughintubation) andnoninvasive ventilation.The minimum patientweights in both ventilators(12.5 kg for the NKV-330and 20 kg for the V60) arewithin the samesubpopulation of"Children" under the FDAdefinition for Pediatrics. |
| TechnicalCharacteristic | Nihon Kohden NKV-330Ventilator System | Philips/Respironics V60Ventilator* (PredicateDevice - K102985) | Comparison |
| IntendedEnvironment of use | Intra-hospital transportwithin a hospital orhospital-type facility. Thedevice is not intended fortransport between hospitalsor hospital-type facilities,therefore is not a transportventilator as defined byISO 80601-2-12 Section201.1.1. | In a hospital or otherinstitutional settings underthe direction of a physician.In other section thanintended use from the V60manual, there is adescription to use theventilator during intra-hospital transport | Same |
| Anatomical Site | Patient face and airways | Patient face and airways | Same |
| Target Population | Adult and pediatric patients | Adult and pediatric patients | Same |
| WaveformComparison | Comparable waveform testresult | Comparable waveform testresult | Substantially equivalent |
| Design | Consists of a graphic userinterface to set ventilationand to monitor ventilationand SpO2 and CO2, breathdelivery unit, breathingcircuit, and cart;Controls oxygen deliverythrough a proportionalvalve and controls airdelivery through an internalblower | Consists of a graphic userinterface to set ventilationand to monitor ventilation,breath delivery unit,breathing circuit, and cart;Controls oxygen deliverythrough a proportionalvalve and controls airdelivery through an internalblower | Substantially equivalentThe NKV-330 has optionalSpO2 and CO2 monitors,for the users who desire touse these monitors foradded safety. The referencedevice, NKV-550 SeriesVentilator System(K181695), contains thesame optional SpO2 andCO2 monitors. |
| Chemicals Deployedto Patient | Ambient air and oxygen | Ambient air and oxygen | Same |
| Delivery method toPatient | Positive pressure | Positive pressure | Same |
| Energy Used forDevice | AC Power, a backupbattery DC Power (Ni-Mhbattery), and a main battery(Li-ion battery) | AC Power, and optionalinternal DC Power (Li-ionbattery) | Substantially equivalentThe NKV-330 containstwo batteries for addedsafety. The V60 has onebattery. |
| Control principle | Time-cycled, volume-target, pressure-controlled | Time-cycled, volume-target, pressure-controlled | Same |
| TechnicalCharacteristic | Nihon Kohden NKV-330Ventilator System | Philips/Respironics V60Ventilator* (PredicateDevice - K102985) | Comparison |
| Modes and ControlsSettings | PCV mode | PCV mode | Substantially equivalent |
| PRVC mode (VolumeTarget) | AVAPS mode (VolumeTarget) | Substantially equivalent | |
| -Max Pressure for PRVC | Max Pressure for AVAPS | ||
| -Min Pressure for PRVC | Min Pressure for AVAPS | ||
| S/T mode (Spontaneous PS/ Timed PCV) | S/T mode (Spontaneous PS/ Timed PCV) | For trigger, both the NKV-330 and the V60 use flowshape method fortriggering. The NKV-330provides conventional flowtrigger in addition. | |
| -IPAP | -IPAP | ||
| -EPAP/CPAP | -EPAP | ||
| Spont/PS mode (PS set tozero is CPAP) | CPAP mode | ||
| Inspiratory Time or I:E | Inspiratory Time (i-Time) | NKV-330 uses a peak flow% for cycling off. | |
| Pressure Release (pressurerelief during the active partof exhalation only forCPAP) | C-Flex (pressure reliefduring the activepart of exhalation only forCPAP) | V60 uses Auto-Track byflow shape method. | |
| Trigger; flow trigger oradvanced trigger (Flowshape method for trigger) | Trigger; auto-track trigger(flow shape method fortrigger) | ||
| Cycling off; peak flow % | Cycling off; Auto-Track(flow shape method) | ||
| PPV mode | PPV mode | ||
| Max E (PPV) | Max E (PPV) | NKV-330 has additional | |
| Max R (PPV) | Max R (PPV) | settings for limiting Min | |
| Max P (PPV) | Max P (PPV) | and Max inspiratory time(Ti). | |
| Max VT (PPV) | Max V (PPV) | ||
| O2 concentration (FiO2) | O2 concentration (FiO2) | ||
| Ramp up time | Ramp time | ||
| Resp Rate (frequency) | Resp Rate (frequency) | ||
| Slope (rise time) | Rise Time | ||
| Min/Max Ti | |||
| O2 Therapy ModeFlow (1-60 L/min)Circuit Obstruction Alarm(automatic) | Substantially equivalent toreference device, NKV-550 Series VentilatorSystem (K181695).**O2 therapy mode deliversonly a constant flow atFiO2 that is set by the user. |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for NIHON KOHDEN. The logo consists of a blue abstract graphic on the left, followed by the company name "NIHON KOHDEN" in blue, sans-serif font. The graphic appears to be a stylized representation of waves or a similar abstract design.
Nihon Kohden OrangeMed, Inc.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for NIHON KOHDEN. The logo consists of a blue abstract symbol on the left and the company name "NIHON KOHDEN" in blue text on the right. The symbol appears to be a stylized representation of waves or a similar abstract design.
Nihon Kohden OrangeMed, Inc.
*Reference manufacturer's user manual "Respironics V60 Ventilator User Manual" (1047358 Rev C) and Philips V60 Ventilator Specification (Oct 2016)
** The O2 Therapy operation, applicable settings, monitors, and alarms are the same between proposed device, NKV-330 Series Ventilator System, and reference device, NKV-550 Series Ventilator System. NKV-330 added a Circuit Obstruction alarm. Ventilator performance during O2 Therapy comparison testing was completed and demonstrated substantial equivalence between proposed and reference device.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Nihon Kohden OrangeMed, Inc. The logo consists of a blue abstract graphic to the left of the text "NIHON KOHDEN" in blue. Below this is the text "Nihon Kohden OrangeMed, Inc." in orange. The logo is simple and professional, likely representing a medical or technology company.
16. Summary of Non-Clinical Performance Data
Performance of the Nihon Kohden NKV-330 Ventilator was demonstrated by the following:
- Software Verification .
- Electrical Safety & EMC Testing
- Agency Testing to Applicable Standards
- Performance of Ventilation Modes and Control Settings ●
- Device Functionality
- Power performance with AC Mains power or Reserve battery power
- . Endurance/Reliability
- Essential Performance and Worst Case VBS
- Environmental
- Cleaning & Disinfection
- Biocompatibility
- Waveform Comparison with Predicate ●
- Human Factors/Usability .
- Risk Management ●
- . Accessory Compatibility
Where applicable, the Nihon Kohden NKV-330 Ventilator has been tested in compliance with the following standards:
| Standard | Title |
|---|---|
| ANSI AAMI ES 60601-12005+AC1;A2 (R2012) | Medical Electrical Equipment - Part 1: General Requirements for Basic Safety andEssential Performance |
| IEC 60601-1Ed. 3.1 2012-08 | Medical Electrical Equipment - Part 1: General Requirements For Basic Safety AndEssential Performance |
| IEC 60601-1-2Ed. 4.0 2014-02 | Medical Electrical Equipment - Part 1-2: General Requirements for Safety - CollateralStandard: Electromagnetic Compatibility - Requirements and Tests |
| IEC 60601-1-6Ed. 3.1 2013-10 | Medical electrical equipment - Part 1-6: General requirements for safety - CollateralStandard: Usability |
| IEC 60601-1-8Ed 2.1 2012-11 | Medical Elec. Equip. - Part 1-8: General Req. for Basic Safety & Essential Perf. –Collateral Standard: General Req., Tests & Guidance for Alarm Systems in MedicalElec. Equip. & Medical Elec. Systems |
| IEC 60601-1-9Ed. 1.1 2013-06 | Medical electrical equipment - Part 1-9: General requirements for basic safety andessential performance - Collateral Standard: Requirements for environmentallyconscious design |
| IEC 60601-2-49Ed. 2.0 2011-02 | Particular requirements for the basic safety and essential performance of multifunctionpatient monitoring equipment |
| IEC 62133-2Ed. 1.0 2017-02 | Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safetyrequirements for portable sealed secondary lithium cells, and for batteries made fromthem, for use in portable applications - Part 2: Lithium systems |
| IEC 62366Ed. 1.1 2014-01 | Medical devices — Application of usability engineering to medical devices —Amendment 1 |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for NIHON KOHDEN. The logo consists of a blue square with a white design inside, followed by the text "NIHON KOHDEN" in blue. The design inside the square appears to be a stylized wave or curve.
| Standard | Title |
|---|---|
| ISO 80601-2-12Ed. 1 2011-04 | Medical Electrical Equipment - Part 2-12: Particular Requirements for Basic Safetyand Essential Performance of Critical Care Ventilators |
| ISO 80601-2-55Ed. 2 2018-02 | Medical Electrical Equipment - Part 2-55: Particular Requirements for The BasicSafety and Essential Performance of Respiratory Gas Monitor |
| ISO 80601-2-61Ed. 2 2017-12 | Medical electrical equipment - Part 2-61: Particular requirements for basic safety andessential performance of pulse oximeter equipment |
| FDA BiocompatibilityGuidance 2016-06 | Use of International Standard ISO 10993-1, “Biological Evaluation of MedicalDevices Part 1: Evaluation and testing within a risk management process”,Guidance for Industry and Food and Drug Administration Staff |
| ISO 10993-1Ed. 5 2018-08 | Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within aRisk Management Process |
| ISO 18562-1Ed. 1 2017-03 | Biocompatibility Evaluation of Breathing Gas Pathways in Healthcare Applications |
| ISO 14971Ed. 3 2019-12 | Medical devices - Application of risk management to medical devices |
17. Summary of Animal Performance Data
Not Applicable - Animal performance data was not required to demonstrate substantial equivalence.
18. Summary of Clinical Performance Data
Not Applicable - Clinical performance data was not required to demonstrate substantial equivalence.
19. Conclusion
The evaluation and testing performed demonstrates that the Nihon Kohden NKV-330 Ventilator is as safe and effective as the predicate device.
§ 868.5895 Continuous ventilator.
(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).