K Number
K251366

Validate with FDA (Live)

Date Cleared
2025-10-09

(161 days)

Product Code
Regulation Number
882.1400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EEG-1260A Neurofax System
The EEG-1260A series Neurofax is intended to record, measure and display cerebral and extracerebral activity for EEG and Sleep Studies. These data may be used by the clinician in sleep disorders, epilepsies and other related disorders as an aid in diagnosis. The device is intended for use by medical personnel in any location within a medical facility, laboratory, clinic, or nursing home or outside of a medical facility under direct supervision of a medical professional.

JE-940A EEG Amplifier Unit
The EEG amplifier unit is intended to acquire and record the EEG and other biological signals (ECG, EMG, respiration, EOG, snoring, body position, SpO2, pulse waveform) of a patient, and transmit the measurement data to the electroencephalograph. The transmitted measurement data is displayed on the electroencephalograph screen and provides information to evaluate the functional state of the brain, brain-related diseases and disorders, and sleep disorders.

The device is intended for use by qualified medical personnel within a medical facility, or by staff in an equivalent facility under the direct supervision of qualified medical personnel.

The device is available for use on any patient as determined by qualified medical personnel.

LS-940A Photic Light
The photic stimulator is a light source for photic stimulation to confirm the responsiveness of EEG to photic stimulation during EEG tests and to test visual evoked potentials.

The device is intended for use by qualified medical personnel within a medical facility, or by staff in an equivalent facility under the direct supervision of qualified medical personnel.

The device is available for use on any patient as determined by qualified medical personnel.

Device Description

The EEG-1260A Neurofax is an electroencephalograph system specifically designed for use in healthcare facilities. This device is designed to measure and display the patient's electroencephalogram (EEG) and polysomnography (PSG) signals, providing information and analysis of brain electrical activity.

The JE-940A EEG amplifier unit is a new amplifier unit and is an input unit of the EEG-1260A. The JE-940A Amplifier unit acquires and measures EEG and other polysomnography signals (ECG, EMG, respiration, EOG, snoring, body position, SpO2, and pulse waveforms) associated with EEG/PSG testing, and transmits the acquired data to the EEG-1260A Neurofax. The JE940A operates on AC power or on battery power for mobile EEG measurements. The JE-940A offers an option to connect with the JE-944A Mini electrode junction box, which enhances the operational efficiency and mobility in EEG measurements.

The LS-940A Photic stimulator is a device which provides visual stimuli in the form of flashing light and is used to assess a patient's EEG responses to light stimulation. The parameters for flashing the light signal are controlled by the EEG-1260A Neurofax.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - EEG-1260A Neurofax System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

October 9, 2025

Nihon Kohden Corporation
℅ Charlemagne Chua
Senior Manager, Regulatory Affairs
Nihon Kohden America, LLC
15353 Barranca Parkway
Irvine, California 92618

Re: K251366
Trade/Device Name: EEG-1260A Neurofax System (EEG-1260A); JE-940A EEG Amplifier Unit (JE940A); LS-940A Photic Stimulator (LS-940A)
Regulation Number: 21 CFR 882.1400
Regulation Name: Electroencephalograph
Regulatory Class: Class II
Product Code: OLT, GWQ, OLV, GWE
Dated: September 12, 2025
Received: September 12, 2025

Dear Charlemagne Chua:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251366 - Charlemagne Chua Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251366 - Charlemagne Chua Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Patrick Antkowiak -S

for
Jay Gupta
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K251366

Device Name: EEG-1260A Neurofax System (EEG-1260A); JE-940A EEG Amplifier Unit (JE-940A); LS-940A Photic Stimulator (LS-940A)

Indications for Use (Describe)

EEG-1260A Neurfax System
The EEG-1260A series Neurofax is intended to record, measure and display cerebral and extracerebral activity for EEG and Sleep Studies. These data may be used by the clinician in sleep disorders, epilepsies and other related disorders as an aid in diagnosis. The device is intended for use by medical personnel in any location within a medical facility, laboratory, clinic, or nursing home or outside of a medical facility under direct supervision of a medical professional.

JE-940A EEG Amplifier Unit
The EEG amplifier unit is intended to acquire and record the EEG and other biological signals (ECG, EMG, respiration, EOG, snoring, body position, SpO2, pulse waveform) of a patient, and transmit the measurement data to the electroencephalograph. The transmitted measurement data is displayed on the electroencephalograph screen and provides information to evaluate the functional state of the brain, brain-related diseases and disorders, and sleep disorders.

The device is intended for use by qualified medical personnel within a medical facility, or by staff in an equivalent facility under the direct supervision of qualified medical personnel.

The device is available for use on any patient as determined by qualified medical personnel.

LS-940A Photic Light
The photic stimulator is a light source for photic stimulation to confirm the responsiveness of EEG to photic stimulation during EEG tests and to test visual evoked potentials.

The device is intended for use by qualified medical personnel within a medical facility, or by staff in an equivalent facility under the direct supervision of qualified medical personnel.

The device is available for use on any patient as determined by qualified medical personnel.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 1 of 16

510(k) SUMMARY

Administrative Information

Sponsor: Nihon Kohden Corporation
1-31-4 Nishiochiai, Shinjuku-Ku
Tokyo, Japan 161-8560
FDA Est. Registration: 9611252

Submitter: Nihon Kohden America, LLC
15353 Barranca Parkway
Irvine, CA 92618
FDA Est. Registration: 2080783

Primary Correspondent/U.S. Agent: Charlemagne Chua
Senior Manager, Regulatory Affairs
15353 Barranca Parkway
Irvine, CA 92618
Office Phone: (949) 268-0868
Email: charlemgne_chua@nihonkohden.com

Subject Device Information

Submission Type: Traditional 510(k)

Common Device Name: Non-Normalizing Quantitative Electroencephalograph Software

Regulation Medical Specialty: Neurology

Classification Panel: Neurology

Product Codes: OLT, GWQ, OLV, GWE

Premarket Review: Office of Neurological and Physical Medicine Devices (OHT5)
Neurosurgical, Neurointerventional and Neurodiagnostic Devices (DHT5A)

Regulation Number: 882.1400

Classification: Class II

Decision Type: 510(k)

Page 6

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 2 of 16

Proprietary Name: EEG-1260A Neurofax

Reason for 510(k) Submission: Modification to an Existing Device and Introduction of New Accessories

Original 510(k) / Predicate: EEG-1200A Series Neurofax (K080546)

Predicate Manufacturer: Nihon Kohden Corporation

Submission Date: October 3, 2025

Page 7

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 3 of 16

1. PRODUCT DESCRIPTION

The EEG-1260A Neurofax is an electroencephalograph system specifically designed for use in healthcare facilities. This device is designed to measure and display the patient's electroencephalogram (EEG) and polysomnography (PSG) signals, providing information and analysis of brain electrical activity.

The JE-940A EEG amplifier unit is a new amplifier unit and is an input unit of the EEG-1260A. The JE-940A Amplifier unit acquires and measures EEG and other polysomnography signals (ECG, EMG, respiration, EOG, snoring, body position, SpO2, and pulse waveforms) associated with EEG/PSG testing, and transmits the acquired data to the EEG-1260A Neurofax. The JE940A operates on AC power or on battery power for mobile EEG measurements. The JE-940A offers an option to connect with the JE-944A Mini electrode junction box, which enhances the operational efficiency and mobility in EEG measurements.

The LS-940A Photic stimulator is a device which provides visual stimuli in the form of flashing light and is used to assess a patient's EEG responses to light stimulation. The parameters for flashing the light signal are controlled by the EEG-1260A Neurofax.

2. INDICATION FOR USE / INTENDED USE

EEG-1260A

Intended Use: The EEG-1260A Electroencephalographs record electroencephalogram and related vital signals and displays the electroencephalogram and vital signals as waveforms.

Indication for Use: The EEG-1260A Neurofax is intended to record, measure and display cerebral and extracerebral activity for EEG and Sleep Studies. These data may be used by the clinician in sleep disorders, epilepsies and other related disorders as an aid in diagnosis. The device is intended for use by medical personnel in any location within a medical facility, laboratory, clinic, or nursing home or outside of a medical facility under direct supervision of a medical professional.

JE-940A

Intended Use: The JE-940A amplifier unit is intended to acquire and record the EEG and other biological signals (ECG, EMG, respiration, EOG, snoring, body position, SpO2, pulse waveform) of a patient, and transmit the measurement data to the electroencephalograph.

Indication for Use: The EEG amplifier unit is intended to acquire and record the EEG and other biological signals (ECG, EMG, respiration, EOG, snoring, body position, SpO2, pulse waveform) of a patient, and transmit the measurement data to the electroencephalograph. The transmitted measurement data is displayed on the electroencephalograph screen and provides information to evaluate the functional state of the brain, brain-related diseases and disorders, and sleep disorders.

The device is intended for use by qualified medical personnel within a medical facility, or by staff in an equivalent facility under the direct supervision of qualified medical personnel.

Page 8

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 4 of 16

The device is available for use on any patient as determined by qualified medical personnel.

LS-940A

Intended Use: LS-940A is intended to evoke response photic stimulator.

Indication for Use: The photic stimulator is a light source for photic stimulation to confirm the responsiveness of EEG to photic stimulation during EEG tests and to test visual evoked potentials.

The device is intended for use by qualified medical personnel within a medical facility, or by staff in an equivalent facility under the direct supervision of qualified medical personnel.

The device is available for use on any patient as determined by qualified medical personnel.

3. SUBMISSION SCOPE

Nihon Kohden (NK) is requesting market clearance for a modified Nihon Kohden device, the EEG-1260A Neurofax and newly introduced accessories. The new accessories of the EEG1260A Neurofax include the JE-940A EEG Amplifier Unit and LS-940A Photic Stimulator.

4. COMPARISON WITH THE PREDICATE DEVICES

4.1 Subject Device Information

Table 1 provides the regulation and classification for the subject device and accessories.

Table 1 Regulatory Information on EEG-1260A, JE-940A, and LS-940A

DeviceRegulationProduct CodeDevice ClassificationClassification
EEG1260A§882.1400OLT, GWQ, OLV, GWENon-Normalizing Quantitative Electroencephalograph SoftwareClass II
JE-940A§882.1400GWQ, OLV, OLTFull-Montage Standard ElectroencephalographClass II
LS-940A§882.1890GWEStimulator, Photic, Evoked ResponseClass II

Page 9

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 5 of 16

4.2 Predicate Device Information

Table 2 lists the basic information about the predicate devices including 510(k) number, device name, 510(k) holder, and clearance date. Regulatory information used for comparison is provided in Table 3

Table 2 Predicate Device General Information

Device510(k) NumberDevice Name510(k) HolderClearance Date
EEG-1260AK080546EEG-1200A Series NeurofaxNihon Kohden Corp.August 29, 2008
JE-940AK080546JE-921A Electrode Junction BoxNihon Kohden Corp.August 29, 2008
LS-940AK101691Lifelines Photic StimulatorLifelines Ltd.August 10, 2010

Table 3 Predicate Device Regulatory Information

DeviceRegulationProduct CodeDevice ClassificationClassification
EEG-1200A§882.1400OLT, GWQ, OLVNon-Normalizing Quantitative Electroencephalograph SoftwareClass II
JE-921A§882.1400GWQ, OLVFull-Montage Standard ElectroencephalographClass II
Lifelines Photic Stimulator§882.1890GWEStimulator, Photic, Evoked ResponseClas II

Page 10

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 6 of 16

4.3 Comparison Table

Table 4 Comparison Table of the EEG-1260A Neurofax and its Predicate

DescriptionEEG-1260A Subject DeviceEEG-1200A (K080546) Predicate DeviceComment
General
Manufacturer (Applicant)Nihon Kohden CorporationNihon Kohden CorporationNo change
Classification panelNeurologyNeurologyNo change
Device classClass IIClass IINo change
Regulatory number21 CFR Part 882.140021 CFR Part 882.1400No change
Product codeOLT, GWQ, OLVOLT, GWQ, OLVNo change
Indications for useThe EEG-1260A series Neurofax is intended to record, measure and display cerebral and extracerebral activity for EEG and Sleep Studies. These data may be used by the clinician in sleep disorders, epilepsies and other related disorders as an aid in diagnosis. The device is intended for use by medical personnel in any location within a medical facility, laboratory, clinic, or nursing home or outside of a medical facility under direct supervision of a medical professional.The EEG-1200A Series Neurofax is intended to record, measure and display cerebral and extracerebral activity for EEG and Sleep Studies. These data may be used by the clinician in Sleep Disorder, Epilepsies and other related disorders as an aid in diagnosis. The device is intended for use by medical personnel in any location within a medical facility, laboratory, clinic or nursing home or outside of a medical facility under direct supervision of a medical professional.No change* *Note: after K080546 was FDA cleared, EEG-1200A was renamed as EEG1260A.
Intended useThe EEG-1260A Electroencephalographs record electroencephalogram and related vital signals and displays the electroencephalogram and vital signals as waveforms.Not SpecifiedNo change
Intended PopulationPatient age not specifiedPatient age not specifiedNo change
Cybersecurity
CybersecurityComply with FDA guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (2023)Not embeddedChange

Page 11

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 7 of 16

DescriptionEEG-1260A Subject DeviceEEG-1200A (K080546) Predicate DeviceComment
Cybersecurity - Protection dataEncryption function complies with FIPS 140-2 Level.1Not embeddedChange
Compatible Input Unit
EEG amplifierJE-940A (New)JE-921A JE-120A (K113117) WEE-1200 (K171124) AE-120A (K183529)Change The predicate device is available with the same optional compatible amplifier(s) as with the subject device through the 510(k) clearance K113117, K171124, K183529.
Function (Software)
PE-210AK, Switch Box ProgramAvailableNot Available at the time of K080546 clearance.Change
Identification and characterization of areas of burstsuppression pattern in the EEG.Available (Optional Software QP160AK)Not available at the time of K080546 clearance.Change
Seizure detectionAvailable (Optional Software QP160AK)Not available at the time of K080546 clearance.Change
Digital video softwareAvailable (Optional Software QP110AK)Available (Optional Software QP110AK)No Change

Page 12

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 8 of 16

Table 5 Comparison Table of the JE-940A EEG Amplifier Unit and its Predicate

DescriptionJE-940A Subject Accessory DeviceJE-921A (K080546) Predicate DeviceComments
General
ManufacturerNihon Kohden CorporationNihon Kohden CorporationNo Change
Product codeGWQ, OLVGWQ, OLVNo Change
Indications for useThe EEG amplifier unit is intended to acquire and record the EEG and other biological signals (ECG, EMG, respiration, EOG, snoring, body position, SpO2, pulse waveform) of a patient, and transmit the measurement data to the electroencephalograph. The transmitted measurement data is displayed on the electroencephalograph screen and provides information to evaluate the functional state of the brain, brain-related diseases and disorders, and sleep disorders. The device is intended for use by qualified medical personnel within a medical facility, or by staff in an equivalent facility under the direct supervision of qualified medical personnel. The device is available for use on any patient as determined by qualified medical personnel.Not clearly defined at the time of K080546 clearance.Change
Intended useThe JE-940A amplifier unit is intended to acquire and record the EEG and other biological signals (ECG, EMG, respiration, EOG, snoring, body position, SpO2, pulse waveform) of a patient, and transmit the measurement data to the electroencephalograph. JE-944A is intended to relay EEG and other vital signals to the JE-940A.Not clearly defined at the time of K080546 clearance.Change

Page 13

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 9 of 16

DescriptionJE-940A Subject Accessory DeviceJE-921A (K080546) Predicate DeviceComments
Prescription / Over-the-counter usePrescriptionPrescriptionNo Change
Intended populationPatient age not specifiedPatient age not specifiedNo Change
System Configuration
Compatible Junction BoxJE-922A Mini Electrode Junction Box JE-944A Mini Electrode Junction BoxJE-922A Mini Electrode Junction BoxChange
Connection method (Data transfer)Wired communication: USB or LAN (Wired LAN: 100 BASE-TX)Wired communication: USB or LANNo Change
Connection securityEncryption method: AES-256 Encryption key: 256 bit (32 bytes)Not embeddedChange
Acquisition Features
ParametersEEG, ECG, EMG, respiration, EOG, snoring, body position, SpO2, pulse waveformEEG, ECG, EMG, respiration, EOG, snoring, body position, SpO2, pulse waveform, CO2Change
Total number of inputs6349Change
- EEG input jacks2525
- Multi-purpose input jacks164
- Bipolar input jacks16 (8 pairs) (All bipolar input jacks can also be used as multipurpose jacks)14 (7 pairs) (8 jacks (4 pairs) of bipolar input jacks can also be used as multipurpose jacks)
- DC input jacks44
- Body position channel1Not available
- SpO2 connector11
- CO2 connectorNot available1
Sampling frequencySelect from the following: 4000 Hz, 2000 Hz, 1000 Hz, 500 Hz, 200 HzSelect from the following: 1000 Hz, 500 Hz, 200 Hz, 100Hz
Input impedance200 MΩ100 MΩ
Internal noise levelLess than 1.5 μVp-p (0.5Hz to 60Hz)Less than 1.5 μVp-p (0.5Hz to 60Hz)
CMRR120 dB or greater (EEG input, multi-purpose input) 100 dB or greater (bipolar input)105 dB or greater (EEG input/extra input, 60Hz) 100 dB or greater (bipolar input, 60 Hz)
Offset tolerance±1000 mV or greater±600 mV
Input range±1.6 mV, ±3.2 mV, or ±12.8 mV (selectable)±3.2 mV
A/D conversion24 bits16 bits
Anti-aliasing filterSelect from the following:

Page 14

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 10 of 16

DescriptionJE-940A Subject Accessory DeviceJE-921A (K080546) Predicate DeviceComments
1,200 Hz, 600 Hz, 300 Hz, 120 Hz, 70 Hz, 60 Hz300 Hz
Time constant2 seconds (0.08 Hz) or 10 seconds (0.016 Hz) (selectable)2 seconds
EEGComply with IEC 80601-2-26:2019Comply with IEC 60601-2-26:2002Change
Polysomnography Signals (Measurement Parameters other than EEG)
ECGLinearity and dynamic range: Amplitude of displayed waveform within 10 % of 50 μV of the input signal (Input signal of triangular wave with amplitudes of 0.5 mV, 1 mV, 2 mV, and 6 mV p-v, 6.25Hz) Input impedance: 10 MΩ or greater (5 mV sinusoidal signal with offset voltages of 300 mV and -300 mV, 10Hz) Common mode rejection: At least 60 dB at the supply mains frequency (Sine wave, 50/60 Hz) At least 45 dB at twice the supply mains frequency (Sine wave, 100/120 Hz) System noise: Not exceed 50 μV p-v over any 10-second periodLinearity and dynamic range: Amplitude of displayed waveform within 10 % of 50 μV of the input signal (Input signal of triangular wave with amplitudes of 0.5 mV, 1 mV, 2 mV, and 6 mV p-v, 6.25Hz) Input impedance: 10 MΩ or greater (5 mV sinusoidal signal with offset voltages of 300 mV and -300 mV, 10Hz) Common mode rejection: At least 60 dB at the supply mains frequency (Sine wave, 50/60 Hz) At least 45 dB at twice the supply mains frequency (Sine wave and 100/120 Hz) System noise: - Not exceed 50 μV p-v over any 10-second periodNo change
EMGAmplitude of the displayed waveform is between 45μVp-p and 55μVp-p when a voltage of 10Hz and 50μVp-p is applied.Amplitude of the displayed waveform is between 45μVp-p and 55μVp-p when a voltage of 10Hz and 50μVp-p is applied.No change
Respiration (air flow)Available (waveform)Available (waveform)No change
EOGAmplitude of the displayed waveform is between 45μVp-p and 55μVp-p when a voltage of 10Hz and 50μVp-p is applied.Amplitude of the displayed waveform is between 45μVp-p and 55μVp-p when a voltage of 10Hz and 50μVp-p is applied.No change
SnoringAvailable (waveform)Available (waveform)No change
Body positionAvailableAvailableNo change
SpO2 declared range and pulse rate (PR)SpO2 Declared range - 70 to 100%SpO2SpO2 Declared range - 70 to 100%SpO2Change

Page 15

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 11 of 16

DescriptionJE-940A Subject Accessory DeviceJE-921A (K080546) Predicate DeviceComments
Pulse Rate Display Range - 30 to 300 beats/min Alarm - None Comply with ISO 80601-2-61:2017Alarm - None Compliance to ISO 80601-2-61: 2017Compliance to ISO 80601-2-61: 2017
Pulse waveformAvailableAvailableNo Change
CO2Not AvailableOptionalChange
Operating Mode
RecordingThe operation to start or stop data recording is available on both the device panel and from the electroencephalograph screen.The operation to start or stop data recording is available only from the electroencephalograph screen.Change
Data backupAvailableNot availableChange
Battery operationAvailable Use of four AA batteries (alkaline primary batteries or NiMH secondary batteries).Not availableChange
Lead checkAvailableNot availableChange
Calibration data, log information (ISO 15189)AvailableNot availableChange
Environmental
Operation conditionsTemperature 10 to 35˚C (50 to 95˚F) Humidity 30 to 85% (noncondensing) Atmospheric pressure 70 to 106 kPaTemperature 10 to 35˚C (50 to 95˚F) Humidity 30 to 80% (noncondensing) Atmospheric pressure 70 to 106 kPaChange
Storage conditionsTemperature −20 to +65˚C (−4 to +140˚F) Humidity 10 to 95% Atmospheric pressure 70 to 106 kPaTemperature −20 to +60˚C (−4 to +140˚F) Humidity 10 to 80% Atmospheric pressure 70 to 106 kPaNo Change

Page 16

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 12 of 16

Table 6 Comparison Table of the LS-940A Photic Stimulator and its Predicate

DescriptionLS-940A Subject Accessory DeviceLifelines Photic Stimulator (K101691) Predicate Accessory DeviceComparison
Regulatory Information
ManufacturerNihon Kohden CorporationLifeLines LtdChange
Product codeGWEGWENo Change
Indication for useThe photic stimulator is a light source for photic stimulation to confirm the responsiveness of EEG to photic stimulation during EEG tests and to test visual evoked potentials. The device is intended for use by qualified medical personnel within a medical facility, or by staff in an equivalent facility under the direct supervision of qualified medical personnel. The device is available for use on any patient as determined by qualified medical personnel.The Lifelines Photic Stimulator is indicated for photic activation of the EEG during an EEG study and in the generation of visual evoked potentials.Change
Intended useLS-940A is intended to evoke response photic stimulator.Not SpecifiedNo change
Intended PopulationPatient age not specifiedPatient age not specifiedNo change
Specification
Light sourceLEDLEDNo change
Illuminance Flash rate- Typ. 26,000 lx / Single Manual flash (0.5 Hz, 1 to 33 Hz) - Typ. 26,000 lx (flash rate) 0.5 Hz, 1 to 33 Hz) - Typ. 13,800 lx (flash rate 40, 50, 60 Hz) (30 cm away from the lightemitting surface)- 700 lm typical, 900 lm max - 13,000 lux at one 1 foot 1 to 60 Hz or single manual flashChange
Frequency- NORMAL 0.5, 1 – 33 (1Hz step), 40, 50, 60Hz - RANDOM 0.5, 1 – 33 (1Hz step)The frequency is constant.

Page 17

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 13 of 16

5. PERFORMANCE

5.1 Summary of Non-Clinical Performance Testing

Nihon Kohden conducted non-clinical bench performance testing as part of the design verification and validation activities. The EEG-1260A, JE-940A, and LS-940A were evaluated for reprocessing & shelf-life, biocompatibility, software & firmware, cybersecurity & interoperability, electromagnetic compatibility & electrical, mechanical, wireless & thermal safety, system connection, and usability performance. The same test methods and specifications were used as for the predicate device, establishing equivalency with the predicate devices. The testing was conducted using EEG-1260A, JE-940A, and LS-940A as a system. These tests were performed according to the international and FDA-recognized consensus standards listed in Table 7.

Table 7 Applied Standards and Guidance List for the EEG-1260A, JE-940A, and LS-940A

Recognition NumberDescription
5-125ISO 14971:2019 Medical devices - Application of risk management to medical devices
13-79IEC 62304:2006+A1:2015 Medical device software -Software life cycle processes
13-122IEC 81001-5-1: 2021 Health software and health IT systems safety, effectiveness and security - Part 5-1: Security - Activities in the product life cycle
19-4IEC 60601-1:2005+A1:2012+A2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
19-36IEC 60601-1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
19-50IEC TS 60601-4-2:2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical system
5-129IEC 62366-1:2015+A1:2020 Medical devices - Part 1: Application of usability engineering to medical devices
Not RecognizedIEC 80601-2-26:2019 Medical electrical equipment - Part 2-26: Particular requirements for the basic safety and essential performance of electroencephalographs
1-139ISO 80601-2-61:2017 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
2-258ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
10-130ANSI Z80.36:2021 Ophthalmic - Light Hazard Protection for Ophthalmic Instruments

Page 18

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 14 of 16

Recognition NumberDescription
Not recognizedISO 15004-2:2007 Ophthalmic instruments - Fundamental requirements and test methods Part 2: Light hazard protection
12-249IEC 62471:2006 Photobiological safety of lamps and lamp systems
Guidance"General Principles of Software Validation" Document issued on: January 11, 2002 "Content of Premarket Submissions for Device Software Functions" Document Issued on June 14, 2023. "Multiple Function Device Products: Policy and Considerations", Document issued on June 14, 2023. Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, Document Issued on September 27, 2023. "Postmarket Management of Cybersecurity in Medical Devices" Document Issued on December 28, 2016 "Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices" Document issued on September 6, 2017 "Off-The-Shelf Software Use in Medical Devices" Document Issued on September 27, 2023. "Electromagnetic Compatibility (EMC) of Medical Devices" Document issued on June 6, 2022. "Applying Human Factors and Usability Engineering to Medical Devices" Document issued on: February 3, 2016 Draft guidance "Content of Human Factors Information in Medical Device Marketing Submissions" December, 2022.

6. SUBSTANTIAL EQUIVALENCE DISCUSSION

The comparison of the EEG-1260A Neurofax to its predicate device (EEG-1200A Series Neurofax K080546), the JE-940A EEG Amplifier unit to its predicate device (JE-921A Electrode Junction Box K080546), and the LS-940A Photic Stimulator to its predicate device (Lifelines Photic Stimulator K101691) and the discussion of the key aspects are presented in the following sections.

Discussion for Intended Use

In comparing the Intended Use between the EEG-1260A Neurofax and its predicate device, the Intended Use are the same.

In comparing the Intended Use between the JE-940A EEG Amplifier Unit and its predicate device, the Intended Use are the same.

Page 19

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 15 of 16

In comparing the Intended Use between the LS-940A Photic Stimulator and its predicate device, the Intended Use for the predicate device and LS-940A Photic Stimulator are used as a light source for photic stimulation,

6.1 Indications for Use

In comparing the Indications for Use between the EEG-1260A Neurofax and its predicate device, there were no changes in the Indications for Use.

In comparing the Indications for Use between the JE-940A EEG Amplifier Unit and its predicate device, the Indications for Use are the same.

In comparing the Indications for Use between the LS-940A Photic Stimulator and its predicate device, the Indications for Use were modified to include more details to better support the intended use of the LS-940A as a light source for photic stimulation to confirm the responsiveness of EEG to photic stimulation during EEG tests and to test visual evoked potentials. Based on the Intended Use, the predicate device and LS-940A Photic Stimulator are regulated under the same regulation number and FDA product code.

6.2 Discussion for Technological Characteristics

The EEG-1260A Neurofax and its predicate device have the same technological characteristics.

The JE-940A EEG Amplifier Unit and its predicate device have the same technological characteristics.

The LS-940A Photic Stimulator and its predicate device have the same technological characteristics.

6.3 Discussion for Design Characteristics

There are slight differences between the EEG-1260A Neurofax System and the predicate as referenced. The differences are based on design changes to accommodate new design improvement and to ensure compliance with current applicable standards. The changes have been evaluated to ensure there is no impact on the safety and effectiveness of the EEG-1260A Neurofax System.

There are slight differences between the JE-940A EEG Amplifier Unit and its predicate device. The differences are based on design changes to accommodate new design features and to ensure compliance with current applicable standards. The changes have been evaluated to ensure there is no impact on the safety and effectiveness of the JE-940A EEG Amplifier Unit.

There are slight differences between LS-940A Photic Stimulator and the predicate accessory device. The differences are based on design changes to accommodate the subject device's intended use and to ensure compliance with current applicable standards. The changes have been evaluated to ensure there is no impact on the safety and effectiveness of the LS-940A Photic Stimulator.

Page 20

Traditional 510(k) – EEG-1260A Neurofax System
510(K) Summary

K251366-S003: 510(k) Summary
03 OCT 2025
Page 16 of 16

6.4 Discussion for Compatible Devices

The EEG-1260A Neurofax System and predicate device require the device to be connected to Nihon Kohden amplifiers. The subject device is available with the same amplifiers as with the predicate device. Currently, there are more Nihon Kohden amplifiers that have received 510(k) clearance since the predicate device has cleared by FDA, the differences do not impact the safety and effectiveness of the EEG-1260A Neurofax System when compared to the predicate device.

When compared to the predicate, the JE-940A EEG Amplifier Unit may require the use of a Mini Electrode Junction box (JE-944A). The predicate may require the Mini Electrode Junction box (JE-922A) to relay the signal. Both JE-922A and JE-944A connect to electrode leads attached to patients and relays EEG and other vital signals to EEG Amplifier Unit. Both JE-922A and JE944A are compatible with EEG-1260A Neurofax System.

6.5 Discussion for Performance Specifications

The EEG-1260A Neurofax System has been compared and evaluated using the same performance methods and testing criteria as the predicate.

The JE-940A EEG Amplifier Unit has been compared and evaluated using the same performance methods and testing criteria as the predicate accessory device.

When compared to the predicate accessory device, the LS-940A Photic Stimulator has been evaluated to the most current versions of the Applicable Standards. Any new standard has been considered to ensure safety and efficacy of the subject device.

7. CONCLUSION

The results of the substantial equivalence assessment, taken together with non-clinical bench testing, electrical safety, and electromagnetic compatibility, software verification, and validation demonstrate that the EEG-1260A Neurofax, JE-940A EEG Amplifier Unit, and LS-940A Photic Stimulator do not raise concerns regarding their safety and effectiveness compared to their predicate device and operates in accordance with claimed indications for use. EEG-1260A Neurofax, JE-940A EEG Amplifier Unit, and LS-940A Photic Stimulator are substantially equivalent to the predicate devices.

§ 882.1400 Electroencephalograph.

(a)
Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.(b)
Classification. Class II (performance standards).