(266 days)
Not Found
No.
The document explicitly states that the device "does not perform any data processing on the data received" and "Central Station itself does not have the function to perform alarm indication judgment." It also does not mention AI, DNN, or ML. The device functions primarily as a display, recording, and printing system for data processed by other Nihon Kohden devices.
No
The device is a "Central Station" software that displays, records, and prints physiological data from other monitoring devices. It does not perform any data processing for therapeutic purposes, nor does it deliver any therapy. Its function is limited to monitoring and data management.
No
Explanation: The device is a 'Central Station' that displays, records, and prints physiological data from other monitors and devices. It does not perform any data processing or alarm judgment itself, rather it presents data processed and judged by connected Nihon Kohden devices. Its function is to facilitate remote monitoring and data management, not to diagnose.
Yes
The device is explicitly stated as a "software only product that is installed on a Commercial Off The Shelf (COTS) Computer." Its functions are all software-based, primarily focusing on displaying, recording, and managing data from other Nihon Kohden devices. There is no indication of integrated hardware or image processing, AI, or other functionalities that would typically categorize it as a hardware device.
No.
The device displays and manages physiological data from other medical devices, but does not perform tests on specimens derived from the human body.
N/A
Intended Use / Indications for Use
Central Station is a network device, intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters.
Central Station does not perform any data processing on the data received from the Nihon Kohden compatible devices. When Central Station is connected with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the Central Station can:
• Admit and discharge patients on the Nihon Kohden network.
• Display and manage compatible devices' real-time patient clinical data, vital signs, alarms and waveforms.
• Review and trend data calculated by connected Nihon Kohden devices.
• Store and transfer historical clinical data for the connected systems.
• Print patient data.
Central Station is intended for use in professional medical facilities by trained medical personnel.
Product codes
MSX, MHX, DRQ, MWI
Device Description
Central Station is software only product that is installed on a Commercial Off the Shelf (COTS) Computer.
Central Station displays waveforms data and numerical data from a connected bedside monitor, vital sign telemeter, or multiple patient receiver unit on the screen.
Central Station is a network device, intended to provide remote patient monitoring to medical personnel. Central Station displays a list of measured values and a trend graph. Numerical data and various waveforms are color-coded for each parameter. Central Station also has the function of displaying an alarm.
Alarm indication in Central Station is displayed as a result of a judgment by the bedside monitor, vital sign telemeter, or multiple patient receiver units connected to the Central Station. Central Station itself does not have the function to perform alarm indication judgment.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
professional medical facilities by trained medical personnel.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Study Type: Nonclinical/Bench Studies, Software Verification, Validation
Key Results: The results of the substantial equivalence assessment, taken together with non-clinical bench testing, software verification, and validation demonstrate that the Central station does not raise concerns regarding its safety and effectiveness compared to its predicate device and operates in accordance with claimed indications for use.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).
FDA 510(k) Clearance Letter - Central Station
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
June 5, 2025
Nihon Kohden Digital Health Solutions, LLC
Arthur Webb
Director Software Quality
14 Bunsen
Irvine, California 92618
Re: K242750
Trade/Device Name: Central Station
Regulation Number: 21 CFR 870.2300
Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm)
Regulatory Class: Class II
Product Code: MSX, MHX, DRQ, MWI
Dated: May 2, 2025
Received: May 2, 2025
Dear Arthur Webb:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
Page 2
K242750 - Arthur Webb Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
K242750 - Arthur Webb Page 3
Sincerely,
Jennifer W. Shih -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology, Diagnostics,
and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Indications for Use
Submission Number (if known)
K242750
Device Name
Central Station
Indications for Use (Describe)
Central Station is a network device, intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters.
Central Station does not perform any data processing on the data received from the Nihon Kohden compatible devices. When Central Station is connected with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the Central Station can:
• Admit and discharge patients on the Nihon Kohden network.
• Display and manage compatible devices' real-time patient clinical data, vital signs, alarms and waveforms.
• Review and trend data calculated by connected Nihon Kohden devices.
• Store and transfer historical clinical data for the connected systems.
• Print patient data.
Central Station is intended for use in professional medical facilities by trained medical personnel.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
Page 5
Page 1 of 9
Nihon Kohden Digital Health Solutions
Proprietary and Confidential
Traditional 510(k) – Central Station
510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance 21 CFR §807.92.
SUBMITTER:
Nihon Kohden Digital Health Solutions, LLC
14 Bunsen, Irvine, CA 92618
Phone: (949) 474-9207
Preparation date: May 02, 2025
Primary Contact Person:
Arthur Webb
Director Software Quality
14 Bunsen, Irvine, CA 92618
Office Phone: (949)-445-8984
Mobile Phone: (949)-577-3849
Email: art_webb@nihonkohden.com
Secondary Contact Person:
Tejal Digambar Warange
Software Regulatory Affairs Specialist
14 Bunsen, Irvine, CA 92618
Mobile Phone: (857)-869-8214
Email: tejal_warange@nihonkohden.com
Device:
Device Name: Central Station
Device Common Name: System, Network and Communication, Physiological Monitor
Proprietary Name: Central Station
Marketing Names: Central Station
Regulation Medical Specialty: Cardiovascular
Classification Panel: Cardiovascular
Regulation Number: 870.2300
Device Class: II
Submission Type: 510(k)
Product Code: MSX
Subsequent Product Codes: DRQ, MHX, MWI
Premarket Review: Office of Cardiovascular Devices (OHT2), Cardiac Electrophysiology, Diagnostics, and Monitoring Devices (DHT2A)
Submission Basis: New Device
Previous 510(k) submissions: None
Submission Date: May 02, 2025
Predicate device:
Device Trade Name: CNS – 2101 Central Monitor
510(k) Number: K223567
Page 6
Page 2 of 9
Traditional 510(k) – Central Station
K242750
DEVICE DESCRIPTION
Central Station is software only product that is installed on a Commercial Off the Shelf (COTS) Computer.
Central Station displays waveforms data and numerical data from a connected bedside monitor, vital sign telemeter, or multiple patient receiver unit on the screen.
Central Station is a network device, intended to provide remote patient monitoring to medical personnel. Central Station displays a list of measured values and a trend graph. Numerical data and various waveforms are color-coded for each parameter. Central Station also has the function of displaying an alarm.
Alarm indication in Central Station is displayed as a result of a judgment by the bedside monitor, vital sign telemeter, or multiple patient receiver units connected to the Central Station. Central Station itself does not have the function to perform alarm indication judgment.
INDICATIONS FOR USE/INTENDED USE
Central Station is a network device, intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters.
Central Station does not perform any data processing on the data received from the Nihon Kohden compatible devices. When Central Station is connected with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the Central Station can:
• Admit and discharge patients on the Nihon Kohden network.
• Display and manage compatible devices' real-time patient clinical data, vital signs, alarms and waveforms.
• Review and trend data calculated by connected Nihon Kohden devices.
• Store and transfer historical clinical data for the connected systems.
• Print patient data.
Central Station is intended for use in professional medical facilities by trained medical personnel.
SUBMISSION SCOPE
Nihon Kohden Digital Health Solutions (NKDHS) is requesting market clearance for the Central Station. The Central Station communicates with other NK devices on the NK network (NET-9).
Comparison of Technological Characteristics with the Predicate Device
Central Station has the same intended use/ indication for use, and technological Characteristics, as the predicate CNS–2101 Central Monitor. Both the predicate as well as subject device intended to display, record, and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters.
Page 7
Page 3 of 9
Traditional 510(k) – Central Station
K242750
COMPARISON TABLE
The table below contains a detailed comparison of the Central Station to the Predicate Device (K223567).
Characteristics | Central Station (Proposed Device) | CNS-2101 Central Monitor (Predicate Device: K223567) Info on predicate | Comparison |
---|---|---|---|
Regulation items | |||
Classification Panel | Cardiovascular | Cardiovascular | Same |
Device Name | System, Network And Communication, Physiological Monitors | System, Network And Communication, Physiological Monitors | Same |
Regulatory Class | Class II (performance standards) | Class II (performance standards) | Same |
Regulatory Number | 21CFR 870.2300- Cardiac monitor (including cardiotachometer and rate alarm) | 21CFR 870.2300- Cardiac monitor (including cardiotachometer and rate alarm) | Same |
Product Code | Primary Product Codes: MSX - System, network and communication, physiological monitors |
Subsequent Product Codes:
MHX- Monitor, physiological, patient (with arrhythmia detection or alarms)
DRQ- Amplifier and signal conditioner, transducer signal
MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) | Primary Product Codes: MSX - System, network and communication, physiological monitors
Subsequent Product Codes:
MHX- Monitor, physiological, patient (with arrhythmia detection or alarms)
DRQ- Amplifier and signal conditioner, transducer signal
MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) | Both devices share the same MSX primary code and performance specifications. The subsequent (secondary) product codes are related to the type of the devices the CNS can connect to. The NK systems were cleared with the additional secondary codes. |
Page 8
Page 4 of 9
Traditional 510(k) – Central Station
K242750
Characteristics | Central Station (Proposed Device) | CNS-2101 Central Monitor (Predicate Device: K223567) Info on predicate | Comparison |
---|---|---|---|
Indications for Use | Central Station is a network device, intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters. |
Central Station does not perform any data processing on the data received from the Nihon Kohden compatible devices. When Central Station is connected with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the Central Station can:
• Admit and discharge patients on the Nihon Kohden network.
• Display and manage compatible devices' real-time patient clinical data, vital signs, alarms and waveforms.
• Review and trend data calculated by connected Nihon Kohden devices.
• Store and transfer historical clinical data for the connected systems.
• Print patient data.
Central Station is intended for use in professional medical facilities by trained medical personnel. | The CNS-2101 central monitor is a networked multipatient monitoring system, that is intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters.
The CNS-2101 does not perform any data processing on the data received from the Nihon Kohden compatible devices. When the CNS-2101 is to connect with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the CNS-2101 can:
• Admit and discharge patients on the Nihon Kohden network.
• Display and manage compatible devices' real-time patient clinical data.
• Mimic the alarms of connected devices when a measured parameter falls outside a preset limits or when an arrhythmia is detected.
• Review and trend data calculated by connected Nihon Kohden devices.
• Store and transfer historical clinical data for the connected systems.
• Print patient data.
The CNS-2101 is intended for use in professional medical facilities by trained medical personnel. | Same Intended use but the indications for uses has some minor editorial |
| Hardware | | | |
| Display | 1,920× 1,080 resolution 23.8-inch color TFT type LCD | 1,920× 1,080 resolution 23.8-inch color TFT type LCD | Same |
| Operation System | Microsoft Windows 10 IoT (Embedded) | Microsoft Windows 10 IoT (Embedded) | Same. |
| Storage media | 512GB (Solid State Drive) | 480 GB SSD (Solid State Drive) | Substantial Equivalence. |
Page 9
Page 5 of 9
Traditional 510(k) – Central Station
K242750
Characteristics | Central Station (Proposed Device) | CNS-2101 Central Monitor (Predicate Device: K223567) Info on predicate | Comparison |
---|---|---|---|
Indications for Use | Central Station is a network device, intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters. |
Central Station does not perform any data processing on the data received from the Nihon Kohden compatible devices. When Central Station is connected with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the Central Station can:
• Admit and discharge patients on the Nihon Kohden network.
• Display and manage compatible devices' real-time patient clinical data, vital signs, alarms and waveforms.
• Review and trend data calculated by connected Nihon Kohden devices.
• Store and transfer historical clinical data for the connected systems.
• Print patient data.
Central Station is intended for use in professional medical facilities by trained medical personnel. | The CNS-2101 central monitor is a networked multipatient monitoring system, that is intended to display, record and print monitored physiological data from Nihon Kohden bedside monitors, telemetry receiver and/or transmitters.
The CNS-2101 does not perform any data processing on the data received from the Nihon Kohden compatible devices. When the CNS-2101 is to connect with the Nihon Kohden bedside monitors and telemetry receivers/transmitters the CNS-2101 can:
• Admit and discharge patients on the Nihon Kohden network.
• Display and manage compatible devices' real-time patient clinical data.
• Mimic the alarms of connected devices when a measured parameter falls outside a preset limits or when an arrhythmia is detected.
• Review and trend data calculated by connected Nihon Kohden devices.
• Store and transfer historical clinical data for the connected systems.
• Print patient data.
The CNS-2101 is intended for use in professional medical facilities by trained medical personnel. | Same Intended use but the indications for uses has some minor editorial |
| Hardware | | | |
| Display | 1,920× 1,080 resolution 23.8-inch color TFT type LCD | 1,920× 1,080 resolution 23.8-inch color TFT type LCD | Same |
| Operation System | Microsoft Windows 10 IoT (Embedded) | Microsoft Windows 10 IoT (Embedded) | Same. |
| Storage media | 512GB (Solid State Drive) | 480 GB SSD (Solid State Drive) | Substantial Equivalence. |
| Audio | Built-in speakers | Built-in speakers | Same |
| Operation Methods | | | |
| Touchscreen | Yes | Yes | Same |
| Keyboard | Yes | Yes | Same |
| Mouse | Yes | Yes | Same |
| Installation method | Stand / Wall mount | Stand / Wall mount | Same |
| Power source | AC Power Supply | AC Power Supply | Same |
| Power interruption support | No | Yes- Internal battery (3-minute backup) | The CNS-2101 has an internal battery. Central Station supports an external Uninterruptible power supply battery. The difference does not raise questions of safety and effectiveness of the subject device. |
| Network | Yes | Yes (NK Network) | Same |
| Compatible monitoring systems | NK Bedside Monitors (MHX): BSM: 1700, 3000, 6000, G9, G5, G7
Vital Signs Monitor (MWI): SVM-7200
NK Telemetry (MHX/DRG): GZ-120/130/140
Multiple Patient Receiver and Transmitters (DRG): ORG-9700/9100
(DRT): ZS-940, ZM-520/521/530/531
Central Monitor (MHX): CNS6201/6801/2101 | NK Bedside Monitors (MHX): BSM: 1700, 3000, 6000, G9, G5, G7
Vital Signs Monitor (MWI): SVM-7200
NK Telemetry (MHX/DRG): GZ-120/130/140
Multiple Patient Receiver and Transmitters (DRG): ORG-9700/9100
(DRT): ZS-940, ZM-520/521/530/531
Central Monitor (MHX): CNS- 6201 and 6801 | Same The CNS communicates with other monitoring stems within the manufacturers product portfolio. |
| Extended Display | Yes | Yes | Same |
| Max connections are supported for one Central Station | Up to 32 | Up to 32 | Same |
| Communication protocol (and compatible monitors) | NET-9/LS-NET communication | NET-9/LS-NET communication | Same - Both devices use a communication protocol to communicate between devices on the network. |
| Output to EMR | Yes | Yes | Same |
| Serial Data Output capable | Yes | Yes | Same |
| Display | | | |
| Wave Display | Yes | Yes | Same |
| Numeric Display | Yes | Yes | Same |
Page 10
Page 6 of 9
Traditional 510(k) – Central Station
K242750
Characteristics | Central Station (Proposed Device) | CNS-2101 Central Monitor (Predicate Device: K223567) Info on predicate | Comparison |
---|---|---|---|
Patient Admit/Discharge | |||
Patient Admit/Discharge/Transfer | Following functions are supported | ||
• Admit, Discharge, Pause, and Transfer functions | |||
• Patient transfer within one central station | |||
• Patient transfer between central stations | |||
• Entering patient information (Manual, Auto) | Following functions are supported | ||
• Admit, Discharge, Pause, and Transfer functions | |||
• Patient transfer within one central station | |||
• Patient transfer between central stations | |||
• Entering patient information (Manual, Auto) | Same | ||
Review window | |||
Review windows | • Trend window | ||
• Tabular Trend window | |||
• Full Disclosure window | |||
• Expanded Waveform window | |||
• Arrhythmia Recall window | |||
• ST Recall window | |||
• Event List window | |||
• Alarm Events window | |||
• ECG 12 Lead Analysis window | |||
• Hemodynamics List window | |||
• SpO2 Trend window | • Trend window | ||
• Tabular Trend window | |||
• Full Disclosure window | |||
• Expanded Waveform window | |||
• Arrhythmia Recall window | |||
• ST Recall window | |||
• Event List window | |||
• Alarm Events window | |||
• ECG 12 Lead Analysis window | |||
• Hemodynamics List window | |||
• SpO2 Trend window | Same - Both devices can connect and display monitored parameters. The viewable data points are similar on each of the monitoring system. | ||
Review history | • Trendgraph data: 120 hours | ||
• Tabular trend data: 120 hours | |||
• Full disclosure and expanded waveform data: 120 hours | |||
• Arrhythmia recall data: 1,500 files | |||
• Event list data: 10,000 files | |||
• Alarm events data: 120 files | |||
• ECG 12 lead analysis data: 200 files | |||
• ST recall data: 120 files | |||
• Hemodynamics list data: 256 files | |||
• SpO2 trendgraph data: 120 hours | • Trendgraph data: 120 hours | ||
• Tabular trend data: 120 hours | |||
• Full disclosure and expanded waveform data: 120 hours | |||
• Arrhythmia recall data: 1,500 files | |||
• Event list data: 10,000 files | |||
• Alarm events data: 120 files | |||
• ECG 12 lead analysis data: 200 files | |||
• ST recall data: 120 files | |||
• Hemodynamics list data: 256 files | |||
• SpO2 trendgraph data: 120 hours | Same - Both devices provide review history data on the central monitor system each of the system can store data. The differences do not affect their substantial equivalence. | ||
Alarm | |||
Setting Alarm Priorities | Bedside Monitor: No | ||
Telemetry system and transmitter: Yes | Bedside Monitor: No | ||
Telemetry system and transmitter: Yes | Same | ||
Printing and Recording |
Page 11
Page 7 of 9
Traditional 510(k) – Central Station
K242750
Characteristics | Central Station (Proposed Device) | CNS-2101 Central Monitor (Predicate Device: K223567) Info on predicate | Comparison |
---|---|---|---|
Patient information (Hospital Name, Bed name, Patient name) | |||
Recording start time or file creation date and time | |||
Recording type (Multiple Waveform, Expanded Waveform, Trend, Tabular Trend, Full Disclosure, Arrhythmia Recall, ST Recall, Event List, Alarm Event, ECG 12 Lead Analysis, Hemodynamics List) and Review window parameters | |||
• Trend window | |||
• Tabular Trend window | |||
• Full Disclosure window | |||
• Expanded Waveform window | |||
• Arrhythmia Recall window | |||
• ST Recall window | |||
• Event List window | |||
• Alarm Events window | |||
• ECG 12 Lead Analysis window | |||
• Hemodynamics List window | |||
• SpO2 Trend window | Patient information (Hospital Name, Bed name, Patient name) | ||
Recording start time or file creation date and time | |||
Recording type (Multiple Waveform, Expanded Waveform, Trend, Tabular Trend, Full Disclosure, Arrhythmia Recall, ST Recall, Event List, Alarm Event, ECG 12 Lead Analysis, Hemodynamics List) and Review window parameters | |||
• Trend window | |||
• Tabular Trend window | |||
• Full Disclosure window | |||
• Expanded Waveform window | |||
• Arrhythmia Recall window | |||
• ST Recall window | |||
• Event List window | |||
• Alarm Events window | |||
• ECG 12 Lead Analysis window | |||
• Hemodynamics List window | |||
• SpO2 Trend window | Same - Both devices are capable of printing to an external network printer. | ||
Recording | No | Patient information (Hospital Name, Bed name, Patient name) | |
Recording start time or file creation date and time | |||
Recording type (Multiple Waveform, Expanded Waveform, Trend, Tabular Trend, Full Disclosure, Arrhythmia Recall, ST Recall, Event List, Alarm Event, ECG 12 Lead Analysis, Hemodynamics List) | |||
Recording duration, Recording speed (25 or 50 mm/s), Numeric data, Waveforms (up to 16 channels), Sensitivity, Lead | |||
Waveform Printing | CNS-2101 is capable of printing on a thermal printer when the parameters fall outside the limits. | ||
ECG Features | |||
ECG Functions Display | Yes | Yes | Same |
HR Display | Yes | Yes | |
ST Alarm Page settings | Upper Limit: OFF, −1.99 to 2.00 mV | ||
Lower Limit: OFF, −2.00 to 1.99 mV | Upper Limit: OFF, −1.99 to 2.00 mV | ||
Lower Limit: OFF, −2.00 to 1.99 mV | |||
ST segment analysis Display | Yes | Yes |
Page 12
Page 8 of 9
Traditional 510(k) – Central Station
K242750
Characteristics | Central Station (Proposed Device) | CNS-2101 Central Monitor (Predicate Device: K223567) Info on predicate | Comparison |
---|---|---|---|
Arrhythmia detection Display | Yes | Yes | |
Arrhythmia analysis Setting | Yes | Yes | |
QTc/QRSd display | Spot check and continuous When connected to BSM-1700, G5 and G7 | Spot check and continuous When connected to BSM-1700, G5 and G7 | |
QRS detection Display | Yes | Yes | |
Pace pulse rejection (Display) | Yes | Yes |
Summary of Nonclinical/Bench Studies
Software
The software documentation was prepared following the FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (released June 14, 2023), specifically for an Enhanced Documentation Level. The rationale for choosing this documentation level is detailed in the Documentation Level document.
Verification testing was conducted at the system integration level to confirm that the device software fulfills its requirements and that safety and
Page 13
Page 9 of 9
Traditional 510(k) – Central Station
K242750
security risk mitigations, where applicable, were effective. Additionally, system-level testing was carried out to show that the software addresses user needs. All unit, integration, and system-level tests successfully met the test protocols.
Cybersecurity information has been provided in line with the FDA's "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" guidance document, dated September 27, 2023.
The CENTRAL STATION has been designed and verified through a risk analysis that considers the risks associated with interoperability, the anticipated users, reasonably foreseeable misuse, and reasonably foreseeable combinations of events that can result in a hazardous situation. Our Interoperability-related risk management activities are included in Cybersecurity Risk Management activities.
Conclusion:
The results of the substantial equivalence assessment, taken together with non-clinical bench testing, software verification, and validation demonstrate that the Central station does not raise concerns regarding its safety and effectiveness compared to its predicate device and operates in accordance with claimed indications for use.