Search Results
Found 3 results
510(k) Data Aggregation
(110 days)
BOSTON SCIENTIFIC - PRECISION VASCULAR
The PV 2000 Synchro®2 Guidewire series of products are intended for general intravascular use, including the neuro and peripheral vasculature. The device can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.
The PV 2000 is a member of the Synchro® Guidewire family of products having a 0.014" outside diameter, being a sterile, single use/disposable product, with a shapeable tip which is used to gain intravascular access to and facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for peripheral and neuro diagnostic and interventional procedures. The guidewire can be torqued to facilitate navigation through the vasculature. A torque device, (Merit Medical (K936032)) is supplied with the wire to facilitate deployment & positioning. A guidewire introducer/shaping mandrel (B. Braun (K760389)) is also supplied and may be used to aid introduction of the guidewire into the catheter hub and/or hemostasis valve and to gently shape the guidewire's distal flexible tip, if desired, according to standard practice. Neither the guidewire introducer nor the torque device are intended to enter the body. The product is projected to be provided in a 100cm - 300cm length range, with 200cm being nominal. The Nitinol tip length is projected to be presented in a 35cm length. A traditional range of flexibility profiles shall also be provided, ranging from support (stiff) to flex (soft). The device is coated on the outer diameter with a lubricious coating over the distal portion of the device. The marker coil is platinum wire at the distal tip of the device to aid visualization under fluoroscopy. The subject device has the ability to access distal, tortuous vasculature, with steerability and torque transmission properties.
The provided text describes safety and performance tests for the PV 2000 Synchro®2 Guidewire to demonstrate substantial equivalence to predicate devices, rather than establishing acceptance criteria or a study proving that a device meets its acceptance criteria. Therefore, most of the requested information cannot be extracted directly from the given content.
However, based on the information provided, here's what can be addressed:
1. A table of acceptance criteria and the reported device performance:
The document states that "Performance testing of materials comprising the PV 2000 Synchro®2 Guidewire was conducted in accordance with ISO 11070:1998, Sterile, Single-Use Intravascular Catheter Introducers." It then lists the types of tests performed and their outcomes. While specific numerical acceptance criteria are not explicitly stated, the reported performance is that each test yielded "acceptable safety & performance outcomes" or "acceptable results."
Test Category | Acceptance Criteria (Not explicitly stated in numerical terms) | Reported Device Performance |
---|---|---|
Performance Testing (ISO 11070:1998) | Assumed to be compliance with ISO standard and safety/performance benchmarks | "acceptable safety & performance outcomes" |
Dimensional inspection | Assumed to meet design specifications | "acceptable safety & performance outcomes" |
Fatigue assessment | Assumed to meet durability requirements | "acceptable safety & performance outcomes" |
Tip flexibility | Assumed to meet functional requirements | "acceptable safety & performance outcomes" |
Tip shaping | Assumed to meet functional requirements | "acceptable safety & performance outcomes" |
Tensile strength | Assumed to meet mechanical strength requirements | "acceptable safety & performance outcomes" |
Guidewire compatibility testing | Assumed to be compatible with other devices | "acceptable safety & performance outcomes" |
Performance under simulated conditions | Assumed to meet functional requirements in simulated clinical use | "acceptable safety & performance outcomes" |
Torsional strength | Assumed to meet mechanical strength requirements | "yielded acceptable results" |
Torqueability | Assumed to meet functional requirements for steerability | "yielded acceptable results" |
EO sterilization adoption tests | Assumed to meet sterilization effectiveness requirements | "yielded acceptable results" |
Biocompatibility Testing (ISO 10993-1) | Assumed to be compliance with ISO standard for biological safety | "confirmed biocompatibility of the subject device" |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
This information is not provided in the given text. The document mentions "Performance testing of materials" and "tests," but does not specify sample sizes for these tests. The country of origin of data or whether it was retrospective or prospective is also not mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not provided in the given text, as the tests described are primarily engineering and material performance tests, not clinical studies requiring expert ground truth assessment.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not provided in the given text.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not provided in the given text. The device is a guidewire, a physical medical instrument, not an AI-assisted diagnostic or interventional system that would typically undergo MRMC studies with human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable as the device is a physical guidewire, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the performance and biocompatibility tests, the "ground truth" or reference for acceptance would be:
- Engineering specifications and regulatory standards: For dimensional inspection, fatigue, tensile strength, torsional strength, torqueability, tip flexibility, tip shaping, guidewire compatibility, and performance under simulated conditions.
- ISO 10993-1 and ISO 11070:1998 standards: For biocompatibility and design/material performance.
- Sterilization validation protocols: For EO sterilization adoption tests.
There is no mention of expert consensus, pathology, or outcomes data as "ground truth" for the tests described.
8. The sample size for the training set:
This information is not applicable as the device is not an AI algorithm that requires a training set.
9. How the ground truth for the training set was established:
This information is not applicable as the device is not an AI algorithm that requires a training set.
Ask a specific question about this device
(56 days)
BOSTON SCIENTIFIC - PRECISION VASCULAR
The PV 1900 Synchro® 0.016" Neuro Guidewire series of products is intended for peripheral and neurovascular use. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.
The PV 1900 is a member of the Synchro® Guidewire family of products having a 0.016" outside diameter, being a sterile, single use/disposable product, with a shapeable tip which is used to gain intravascular access to and facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for peripheral and neuro diagnostic and interventional procedures. The guidewire can be torqued to facilitate navigation through the vasculature. A torque device, (Boston Scientific (K903606)) is supplied with the wire to facilitate deployment & positioning. A guidewire introducer (B. Braun (K760389)) is also supplied and may be used to aid introduction of the guidewire into the catheter hub and/or hemostasis valve and to gently shape the guidewire's distal flexible tip, if desired, according to standard practice. Neither the guidewire introducer nor the torque device are intended to enter the body. The product is projected to be provided in a 100cm - 300cm length range with 140-180cm being nominal. The Nitinol tip length is projected to be presented in a 25cm - 35cm range. A traditional range of flexibility profiles shall also be provided ranging from support (stiff) to flex (soft). The device is coated on the outer diameter with a lubricious coating over the distal portion of the device. The marker coil is platinum wire at the distal tip of the device to aid visualization under fluoroscopy. The subject device has the ability to access distal, tortuous vasculature, with steerability and torque transmission properties.
This is a 510(k) premarket notification for a medical device (a guidewire), not an AI/ML device. Therefore, the typical acceptance criteria and study design for AI/ML performance evaluation are not applicable.
However, I can extract the information relevant to the device's performance given the context of this document.
Summary of Device Performance Information Provided:
Since this is a filing for a traditional medical device (guidewire) and not an AI/ML diagnostic or prognostic system, the requested information (acceptance criteria, sample size, ground truth, experts, MRMC, standalone performance) for AI/ML devices is not directly applicable in the sense of predictive or classification performance metrics.
Instead, the "acceptance criteria" for this device are implicitly tied to demonstrating substantial equivalence to predicate guidewires by meeting various physical and mechanical performance characteristics. The "study" refers to the engineering and biological testing performed on the device.
Here's an attempt to map the requested AI/ML-centric questions to the provided information for this traditional medical device:
-
Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly present acceptance criteria in a quantifiable table with corresponding performance values for each criterion as one would for an AI/ML device (e.g., sensitivity >X%, specificity >Y%). Instead, it states that "Performance testing of materials comprising the PV 1900 Synchro® 0.016" Guidewire was conducted in accordance with ISO 11070:1998, Sterile, Single-Use Intravascular Catheter Introducers." and "Subject product testing has yielded acceptable safety & performance outcomes."
The performance aspects tested and the general outcome are listed.
Acceptance Criteria (Implicit) Reported Device Performance Compliance with ISO 10993-1 for Biocompatibility (for external communicating, blood contact, short duration
Ask a specific question about this device
(11 days)
BOSTON SCIENTIFIC - PRECISION VASCULAR
The "PivotTM" Steerable Microcatheter is intended to be used to access tortuous vasculature for sub-selective controlled infusion or delivery of diagnostic, embolic, and therapeutic agents into the distal, peripheral, coronary, and neurovasculature, and for guide wire exchange/support during diagnostic or interventional procedures.
The "PivotTM" Steerable Microcatheter is a 1.9F/2.4F (nominal distal/proximal) tubular device, ~150 cm in length, with a lumen to be used for delivery of contrast, drugs, or embolics. The lumen is constructed from a polymeric material and has an inside diameter of 0.017". The device is coated on the outer diameter with a lubricious coating over the distal segment of the device. Two radiopaque markers are positioned at the distal tip of the device to aid visualization under fluoroscopy. The proximal end of the device has a standard luer adapter for attachment of accessories and can be used to flush the lumen. The subject device has the ability to access distal, tortuous vasculature over a guide wire, deliver embolics and agents, and has the ability to be steered like a guide wire as needed.
The "Pivot™" Steerable Microcatheter was evaluated against several performance criteria, demonstrating its safety and effectiveness.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Conformity to ISO 10555-1 | Performance testing was conducted in accordance with ISO 10555-1, "Sterile, Single-Use Intravascular Catheters -- Part 1." |
Dimensional Inspection | Verification testing included dimensional inspection. The results were acceptable. |
Hub Integrity | Verification testing included hub integrity. The results were acceptable. |
Flow Rate Measurements | Verification testing included flow rate measurements. The results were acceptable. |
Burst Strength | Verification testing included burst strength. The results were acceptable. |
Tensile Strength | Verification testing included tensile strength. The results were acceptable. |
Guidewire Compatibility | Verification testing included guidewire compatibility. The results were acceptable. |
Performance under Simulated Conditions | Verification testing included performance under simulated conditions. The results were acceptable. |
Torsional Strength | Torsional strength tests were conducted. The results were acceptable. |
Torqueability | Torqueability tests were conducted. The results were acceptable. |
Corrosion Resistance | Corrosion resistance tests were conducted. The results were acceptable. |
Substantial Equivalence | The "Pivot™" Steerable Microcatheter effectively demonstrates substantial equivalence to predicate devices based on indications for use, technological characteristics (basal tubular design, dimensions, polymeric materials and construction, hydrophilic coating), and safety and performance testing. No new questions regarding safety and efficacy were raised. |
2. Sample Size Used for the Test Set and Data Provenance
The provided document does not specify a separate "test set" in the context of clinical trials or data for algorithm performance. The performance evaluation was based on bench testing and simulated conditions, which typically do not involve human subjects or a distinct "test set" in the way an AI/ML device would. Therefore, sample sizes for a test set and data provenance (e.g., country of origin, retrospective/prospective) are not applicable in this context.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Given that the evaluation was based on bench testing and simulated conditions of a physical device, there were no "experts" establishing ground truth in the context of interpreting medical images or clinical data. The acceptance criteria were based on engineering and performance standards (ISO 10555-1 and specific mechanical tests).
4. Adjudication Method for the Test Set
As the evaluation was based on bench testing and simulated conditions, an adjudication method for a test set (e.g., 2+1, 3+1) is not applicable. The assessment was likely based on standardized testing protocols and objective measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The device is a physical microcatheter, and the evaluation was based on its physical and mechanical performance, not on improving human reader performance with AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This question is not applicable. The "Pivot™" Steerable Microcatheter is a physical medical device, not an algorithm or AI model. Therefore, "standalone algorithm performance" is not relevant to its evaluation.
7. The Type of Ground Truth Used
The "ground truth" for the device's performance was established through objective measurements against predefined engineering standards and specifications. This includes:
- Compliance with ISO 10555-1.
- Results from physical tests such as dimensional inspection, hub integrity, flow rate, burst strength, tensile strength, guidewire compatibility, performance under simulated conditions, torsional strength, torqueability, and corrosion resistance.
8. The Sample Size for the Training Set
This question is not applicable. The "Pivot™" Steerable Microcatheter is a physical device, not an AI/ML model, so there is no "training set" in the context of machine learning.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable, as there is no training set for a physical microcatheter.
Ask a specific question about this device
Page 1 of 1