K Number
K050630
Date Cleared
2005-03-22

(11 days)

Product Code
Regulation Number
870.1210
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The "PivotTM" Steerable Microcatheter is intended to be used to access tortuous vasculature for sub-selective controlled infusion or delivery of diagnostic, embolic, and therapeutic agents into the distal, peripheral, coronary, and neurovasculature, and for guide wire exchange/support during diagnostic or interventional procedures.

Device Description

The "PivotTM" Steerable Microcatheter is a 1.9F/2.4F (nominal distal/proximal) tubular device, ~150 cm in length, with a lumen to be used for delivery of contrast, drugs, or embolics. The lumen is constructed from a polymeric material and has an inside diameter of 0.017". The device is coated on the outer diameter with a lubricious coating over the distal segment of the device. Two radiopaque markers are positioned at the distal tip of the device to aid visualization under fluoroscopy. The proximal end of the device has a standard luer adapter for attachment of accessories and can be used to flush the lumen. The subject device has the ability to access distal, tortuous vasculature over a guide wire, deliver embolics and agents, and has the ability to be steered like a guide wire as needed.

AI/ML Overview

The "Pivot™" Steerable Microcatheter was evaluated against several performance criteria, demonstrating its safety and effectiveness.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Conformity to ISO 10555-1Performance testing was conducted in accordance with ISO 10555-1, "Sterile, Single-Use Intravascular Catheters -- Part 1."
Dimensional InspectionVerification testing included dimensional inspection. The results were acceptable.
Hub IntegrityVerification testing included hub integrity. The results were acceptable.
Flow Rate MeasurementsVerification testing included flow rate measurements. The results were acceptable.
Burst StrengthVerification testing included burst strength. The results were acceptable.
Tensile StrengthVerification testing included tensile strength. The results were acceptable.
Guidewire CompatibilityVerification testing included guidewire compatibility. The results were acceptable.
Performance under Simulated ConditionsVerification testing included performance under simulated conditions. The results were acceptable.
Torsional StrengthTorsional strength tests were conducted. The results were acceptable.
TorqueabilityTorqueability tests were conducted. The results were acceptable.
Corrosion ResistanceCorrosion resistance tests were conducted. The results were acceptable.
Substantial EquivalenceThe "Pivot™" Steerable Microcatheter effectively demonstrates substantial equivalence to predicate devices based on indications for use, technological characteristics (basal tubular design, dimensions, polymeric materials and construction, hydrophilic coating), and safety and performance testing. No new questions regarding safety and efficacy were raised.

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not specify a separate "test set" in the context of clinical trials or data for algorithm performance. The performance evaluation was based on bench testing and simulated conditions, which typically do not involve human subjects or a distinct "test set" in the way an AI/ML device would. Therefore, sample sizes for a test set and data provenance (e.g., country of origin, retrospective/prospective) are not applicable in this context.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Given that the evaluation was based on bench testing and simulated conditions of a physical device, there were no "experts" establishing ground truth in the context of interpreting medical images or clinical data. The acceptance criteria were based on engineering and performance standards (ISO 10555-1 and specific mechanical tests).

4. Adjudication Method for the Test Set

As the evaluation was based on bench testing and simulated conditions, an adjudication method for a test set (e.g., 2+1, 3+1) is not applicable. The assessment was likely based on standardized testing protocols and objective measurements.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The device is a physical microcatheter, and the evaluation was based on its physical and mechanical performance, not on improving human reader performance with AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable. The "Pivot™" Steerable Microcatheter is a physical medical device, not an algorithm or AI model. Therefore, "standalone algorithm performance" is not relevant to its evaluation.

7. The Type of Ground Truth Used

The "ground truth" for the device's performance was established through objective measurements against predefined engineering standards and specifications. This includes:

  • Compliance with ISO 10555-1.
  • Results from physical tests such as dimensional inspection, hub integrity, flow rate, burst strength, tensile strength, guidewire compatibility, performance under simulated conditions, torsional strength, torqueability, and corrosion resistance.

8. The Sample Size for the Training Set

This question is not applicable. The "Pivot™" Steerable Microcatheter is a physical device, not an AI/ML model, so there is no "training set" in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable, as there is no training set for a physical microcatheter.

{0}------------------------------------------------

K050630

Summary of Safety and Effectiveness

Prepared 8 March 2005

GeneralProvisionsSubmitter of 510(k)Premarket Notification:Boston Scientific - Precision Vascular2405 West Orton CircleWest Valley City, UT 84119Phone: 801.974.1700Fax: 801.974.1740
Contact Person:Rick GaykowskiVice President, Regulatory/ClinicalAffairs & Quality Systems
Device Trade Name:"PivotTM" Steerable Microcatheter
Device Generic Name:Infusion Catheter
The predicate devices are listed below.
PredicateDevicesDeviceManufacturer510(k) Number/DatePro Code
SDS 0.018"PVS, Inc.K020733, 21 March 2002KRA
SDS 0.010"PVS, Inc.K032494, 05 Sep 2003KRA
Pivot "SDS"PVS, Inc.K033726, 16 Jan 2004KRA
ClassificationClass II, 21 CFR 870.1210, Continuous Flush Catheter, KRA
PerformanceStandardsPerformance standards have not been established by FDA under section 514 of theFederal Food, Drug and Cosmetic Act.
Intended UseThe "PivotTM" Steerable Microcatheter is intended to be used to access tortuousvasculature for sub-selective controlled infusion or delivery of diagnostic, embolic, andtherapeutic agents into the distal, peripheral, coronary, and neurovasculature, and forguide wire exchange/support during diagnostic or interventional procedures.
DeviceDescriptionThe "PivotTM" Steerable Microcatheter is a 1.9F/2.4F (nominal distal/proximal) tubulardevice, ~150 cm in length, with a lumen to be used for delivery of contrast, drugs, orembolics. The lumen is constructed from a polymeric material and has an insidediameter of 0.017". The device is coated on the outer diameter with a lubriciouscoating over the distal segment of the device. Two radiopaque markers are positionedat the distal tip of the device to aid visualization under fluoroscopy. The proximal end ofthe device has a standard luer adapter for attachment of accessories and can be usedto flush the lumen. The subject device has the ability to access distal, tortuousvasculature over a guide wire, deliver embolics and agents, and has the ability to besteered like a guide wire as needed.

{1}------------------------------------------------

TechnologicalCharacteristicsTechnological similarities between the "Pivot™" Steerable Microcatheter and predicate devices include the basal tubular design and dimensions, polymeric materials and construction, and hydrophilic coating. In instances where the technological characteristics may differ slightly, it has been demonstrated that there are no new questions raised regarding safety and efficacy of the "Pivot™" Steerable Microcatheter.
Safety andPerformanceTestsPerformance testing of the "Pivot™" Steerable Microcatheter was conducted in accordance with ISO 10555-1, Sterile, Single-Use Intravascular Catheters -- Part 1. Verification testing for the subject device included dimensional inspection, hub integrity, flow rate measurements, burst strength, tensile strength, guidewire compatibility testing and performance under simulated conditions. Subject product testing has yielded acceptable results.In addition, torsional strength, torqueability, and corrosion resistance tests have also yielded acceptable results. The results of these tests, in conjunction with the substantial equivalence claims as outlined in the premarket notification, effectively demonstrate the "Pivot™" Steerable Microcatheter is substantial equivalent to the cited predicate devices.
Summary ofSubstantialEquivalenceBased on the indications for use, technological characteristics, and safety and performance testing, the subject "Pivot™" Steerable Microcatheter meets the minimum requirements that are considered adequate for its intended use and is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available catheters/cited predicates.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three horizontal lines curving upwards, resembling a human figure with outstretched arms.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 2 2005

Mr. Rick Gaykowski Mice President, Regulatory/Clinical Affairs & Quality Systems Boston Scientific - Precision Vascular 2405 West Orton Circle West Valley City, UT 84119

K050630 Re:

Trade/Device Name: Pivor™ Steerable Microcatheter Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous flush catheter Regulatory Class: II Product Code: KRA Dated: March 10, 2005 Received: March 11, 2005

Dear Mr. Gaykowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for assessment to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 – Mr. Rick Gaykowski

Fage 2 - "htt read as a proposion (21 CFR Part 820); and if applicable, the electronic
forth in the quality systems in and inco (Sections 531-542 of the Act); 21 CFR 1000-105 forth in the quality systems (QS) regulation (2) CFR Pat 620); and 11 CFF
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. 101 - 12 - 12 - 12 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -This letter will allow you to begin marketing your device of your device to a legally
premarket notification. The FDA finding of subscification for your device to a legally remarket notification. The FDA finding of substantial on your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
s and the results of the support (240) 276.0120 . Also, please note the r If you desire specific advice for your device on our labeling nease not the regulation entitled,
contact the Office of Compliance at (240) 276-0120. Also, please not 97). contact the Office of Compliance at (240) 270-01-20. Proxet 807.97). You may obtain
"Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obta "Misbranding by reference to premarket notification "(21 cf 1 cr from the Division of Small
other general information on your responsibilities under the Act from the Division other general information on your responsibilities under (800) 638-204) or 2001)
Manufacturers, International and Consumer Assistance at its todalfred wordindex.html. Manufacturers, International and Consumer Assistance at les to restor in over index.html.
(301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/i

Sincerely yours,

Durna R. Kirchner

Bram D. Zuckerman, M.D.

Director

Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

510(k) Number (if known): ¥650630_

Device Name: "Pivot™" Steerable Microcatheter

Indications for Use:

The "Pivot™" Steerable Microcatheter is intended to be used to access tortuous vasculature for sub-The "Pivot™" Steerable Microcather is intended to activer and therapeutic agents into the distal,
selective controlled infusion of diagnosic, emblic, and therapeupport during selective controlled infusion or delivery of diagnostic, endonic, and therebels agence of the counter of the county, or interventional procedures.

:

. . .

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseOROver-the-Counter Use
(Per 21 CFR 801.109)
Donna P. Vechinel (Division Sign-Off) Division of Cardiovascular Devices
510(k) NumberK050630

§ 870.1210 Continuous flush catheter.

(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).