(110 days)
The PV 2000 Synchro®2 Guidewire series of products are intended for general intravascular use, including the neuro and peripheral vasculature. The device can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.
The PV 2000 is a member of the Synchro® Guidewire family of products having a 0.014" outside diameter, being a sterile, single use/disposable product, with a shapeable tip which is used to gain intravascular access to and facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for peripheral and neuro diagnostic and interventional procedures. The guidewire can be torqued to facilitate navigation through the vasculature. A torque device, (Merit Medical (K936032)) is supplied with the wire to facilitate deployment & positioning. A guidewire introducer/shaping mandrel (B. Braun (K760389)) is also supplied and may be used to aid introduction of the guidewire into the catheter hub and/or hemostasis valve and to gently shape the guidewire's distal flexible tip, if desired, according to standard practice. Neither the guidewire introducer nor the torque device are intended to enter the body. The product is projected to be provided in a 100cm - 300cm length range, with 200cm being nominal. The Nitinol tip length is projected to be presented in a 35cm length. A traditional range of flexibility profiles shall also be provided, ranging from support (stiff) to flex (soft). The device is coated on the outer diameter with a lubricious coating over the distal portion of the device. The marker coil is platinum wire at the distal tip of the device to aid visualization under fluoroscopy. The subject device has the ability to access distal, tortuous vasculature, with steerability and torque transmission properties.
The provided text describes safety and performance tests for the PV 2000 Synchro®2 Guidewire to demonstrate substantial equivalence to predicate devices, rather than establishing acceptance criteria or a study proving that a device meets its acceptance criteria. Therefore, most of the requested information cannot be extracted directly from the given content.
However, based on the information provided, here's what can be addressed:
1. A table of acceptance criteria and the reported device performance:
The document states that "Performance testing of materials comprising the PV 2000 Synchro®2 Guidewire was conducted in accordance with ISO 11070:1998, Sterile, Single-Use Intravascular Catheter Introducers." It then lists the types of tests performed and their outcomes. While specific numerical acceptance criteria are not explicitly stated, the reported performance is that each test yielded "acceptable safety & performance outcomes" or "acceptable results."
| Test Category | Acceptance Criteria (Not explicitly stated in numerical terms) | Reported Device Performance |
|---|---|---|
| Performance Testing (ISO 11070:1998) | Assumed to be compliance with ISO standard and safety/performance benchmarks | "acceptable safety & performance outcomes" |
| Dimensional inspection | Assumed to meet design specifications | "acceptable safety & performance outcomes" |
| Fatigue assessment | Assumed to meet durability requirements | "acceptable safety & performance outcomes" |
| Tip flexibility | Assumed to meet functional requirements | "acceptable safety & performance outcomes" |
| Tip shaping | Assumed to meet functional requirements | "acceptable safety & performance outcomes" |
| Tensile strength | Assumed to meet mechanical strength requirements | "acceptable safety & performance outcomes" |
| Guidewire compatibility testing | Assumed to be compatible with other devices | "acceptable safety & performance outcomes" |
| Performance under simulated conditions | Assumed to meet functional requirements in simulated clinical use | "acceptable safety & performance outcomes" |
| Torsional strength | Assumed to meet mechanical strength requirements | "yielded acceptable results" |
| Torqueability | Assumed to meet functional requirements for steerability | "yielded acceptable results" |
| EO sterilization adoption tests | Assumed to meet sterilization effectiveness requirements | "yielded acceptable results" |
| Biocompatibility Testing (ISO 10993-1) | Assumed to be compliance with ISO standard for biological safety | "confirmed biocompatibility of the subject device" |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
This information is not provided in the given text. The document mentions "Performance testing of materials" and "tests," but does not specify sample sizes for these tests. The country of origin of data or whether it was retrospective or prospective is also not mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not provided in the given text, as the tests described are primarily engineering and material performance tests, not clinical studies requiring expert ground truth assessment.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not provided in the given text.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not provided in the given text. The device is a guidewire, a physical medical instrument, not an AI-assisted diagnostic or interventional system that would typically undergo MRMC studies with human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable as the device is a physical guidewire, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the performance and biocompatibility tests, the "ground truth" or reference for acceptance would be:
- Engineering specifications and regulatory standards: For dimensional inspection, fatigue, tensile strength, torsional strength, torqueability, tip flexibility, tip shaping, guidewire compatibility, and performance under simulated conditions.
- ISO 10993-1 and ISO 11070:1998 standards: For biocompatibility and design/material performance.
- Sterilization validation protocols: For EO sterilization adoption tests.
There is no mention of expert consensus, pathology, or outcomes data as "ground truth" for the tests described.
8. The sample size for the training set:
This information is not applicable as the device is not an AI algorithm that requires a training set.
9. How the ground truth for the training set was established:
This information is not applicable as the device is not an AI algorithm that requires a training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a date and a number. The date at the top of the image is July 7, 2026. Below the date is the number 1053268. The text is black against a white background.
page 1 of 2
Summary of Safety and Effectiveness
| Prepared November 22, 2005Revised March 2, 2006 | |||||
|---|---|---|---|---|---|
| GeneralProvisions | Submitter of 510(k)Premarket Notification: | Boston Scientific - Precision Vascular2405 West Orton CircleWest Valley City, UT 84119Phone: 801.974.1700Fax: 801.974.1740 | |||
| Contact Person: | Rick GaykowskiVP, Regulatory/Clinical AffairsQuality Systems & Operations | ||||
| Device Trade Name:Device Generic Name: | PV 2000 Synchro®2 GuidewireGuidewire | ||||
| The predicate devices are listed in the table below. | |||||
| PredicateDevices | Device | Manufacturer | 510(k) Number/Date | Pro Code | |
| Synchro™ .014" | Precision Vascular | K032146, 12 August 2003 | DQX | ||
| Synchro™ .010"Synchro™ .016"Transend | Precision VascularPrecision VascularTarget Therapeutics | K032146, 12 August 2003K050964, 13 June 2005K971254, 01 July 1997* | DQXDQXDQX | ||
| * This information is assumed based upon our best knowledge. | |||||
| Classification | Class II, 21 CFR 870.1330, Wire, Guide, Catheter 74DQX | ||||
| PerformanceStandards | Performance standards have not been established by FDA under section 514 of theFederal Food, Drug and Cosmetic Act. | ||||
| Intended Use | The PV 2000 Synchro®2 Guidewire series of products is intended for peripheral andneurovascular use. The device can be used to selectively introduce and positioncatheters and other interventional devices within the peripheral and neurovasculature.This device should be used only by physicians trained in percutaneous, intravasculartechniques and procedures. | ||||
| DeviceDescription | The PV 2000 is a member of the Synchro® Guidewire family of products having a0.014" outside diameter, being a sterile, single use/disposable product, with ashapeable tip which is used to gain intravascular access to and facilitate the positioningand exchange of interventional devices in small diameter, tortuous vasculature forperipheral and neuro diagnostic and interventional procedures. The guidewire can betorqued to facilitate navigation through the vasculature. A torque device, (Merit Medical(K936032)) is supplied with the wire to facilitate deployment & positioning. A guidewire |
Page 11 of 64
{1}------------------------------------------------
K05 3268
page 2042
introducer/shaping mandrel (B. Braun (K760389)) is also supplied and may be used to aid introduction of the guidewire into the catheter hub and/or hemostasis valve and to gently shape the guidewire's distal flexible tip, if desired, according to standard practice. Neither the guidewire introducer nor the torque device are intended to enter the body. The product is projected to be provided in a 100cm - 300cm length range, with 200cm being nominal. The Nitinol tip length is projected to be presented in a 35cm length. A traditional range of flexibility profiles shall also be provided, ranging from support (stiff) to flex (soft). The device is coated on the outer diameter with a lubricious coating over the distal portion of the device. The marker coil is platinum wire at the distal tip of the device to aid visualization under fluoroscopy. The subject device has the ability to access distal, tortuous vasculature, with steerability and torque transmission properties.
Technological Characteristics Technological similarities between the subject PV 2000 Synchro®2 Guidewire and the PVS 1300 Synchro™ predicate remain identical. This is also true for competitive predicate device features including the basal design and dimensions, generic materials & construction, and hydrophilic coating. There are no new questions raised regarding safety or efficacy of the PVS 1300 Synchro® 0.014" Guidewire.
Biocompatibility of the PV 2000 Synchro®2 Guidewire materials have been verified in accordance with ISO 10993-1, Biological Evaluation of Medical Devices - Part 1. Materials test results confirmed biocompatibility of the subject device when tested as an external communicating, blood contact, short duration (<24 hours) device.
Safety and Performance Tests
Performance testing of materials comprising the PV 2000 Synchro®2 Guidewire was conducted in accordance with ISO 11070:1998, Sterile, Single-Use Intravascular Catheter Introducers. Ventication testing for the subject device included dimensional inspection, fatique assessment, tip flexibility, tip shaping, tensile strength, guidewire compatibility testing and performance under simulated conditions. Subject product testing has yielded acceptable safety & performance outcomes.
In addition, torsional strength, torqueability, and EO sterilization adoption tests also The results of these tests, in conjunction with the vielded acceptable results. substantial equivalence claims as outlined in the premarket notification, effectively demonstrate the PV 2000 Synchro®2 Guidewires' substantial equivalence to the cited predicate devices.
Summary of Substantial Equivalence
Based on the indications for use, technological characteristics, and safety and performance testing, the subject PV 2000 Synchro@2 Guidewire meets the minimum requirements that are considered adequate for its intended use and is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available quidewires/cited predicates.
Page 12 of 64
{2}------------------------------------------------
Food and Drug Administration
Image /page/2/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird-like figure.
9200 Corporate Boulevard Rockville MD 20850 MAR 1 3 2006
Boston Scientific - Precision Vascular c/o Mr. Rick Gaykowski V.P. Regulatory/Clinical Affairs Quality Systems and Operations 2405 West Orton Circle West Valley City, UT 84119
Re: K053268
PVS 2000 Synchro®2 Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Wire, Guide, Catheter Regulatory Class: Class II Product Code: DQX Dated: February 15, 2006 Received: February 16, 2006
Dear Mr. Gaykowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality
{3}------------------------------------------------
Page 2 - Mr. Rick Gaykowski
systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation 335this (Q) 70g visions 531-542 of the Act); 21 CFR 1000-1050. Control provisions (Soctions 551 - 12 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 This letter with anow you're ough maning of substantial equivalence of your device to a legally premarket notification: "The PDF maing sification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at notiance at (240) 276-0120. Also, please note the regulation entitled, Connate the Office of Central of Centralion" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small general intermational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blymmman for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K053268
Device Name: PV 2000 Synchro®2 Guidewire
Indications For Use: The PV 2000 Synchro®2 Guidewire series of products are intended for general intravascular use, including the neuro and peripheral vasculature. The device can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.
Prescription Use ﮯ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
ivision Sigh-O ivision of Clardiova
Page 1 of
510(k) Number K053268
§ 870.1330 Catheter guide wire.
(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.