K Number
K970836

Validate with FDA (Live)

Date Cleared
1997-08-26

(183 days)

Product Code
Regulation Number
864.5220
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TRUCOUNT Absoluce Count Tubes are intended for use as an accessory to TriTEST in vitro diagnostics, such as that described in K965053 (CD3/CD4/CD45), to allow computation of positive cells per known volume of blood, using flow cytometty.

The TRUCOUNT Counting Control Beads are provided in three levels, and are intended to provide a means to check the reproducibility and accuracy of volumeric pipetting of blood samples. They are an accessory to TriTEST reagens, when used with TRUCOUNT Absolute Count Tubes to obtain absolute counts of positive cells using flow cytomerry.

Indications for Use:

  • For erythrocyte lysed whole blood .
  • For use with FACS Loader ●
  • For use with IVD immunophenotyping reagents .
  • For use on flow cytometers with designated detection ranges ●
  • For use to obtain absolute counts by flow cytometry ◆
Device Description

TRUCOUNT Absolute Count Tubes are fluorescent particles supplied as a peller in a rube. The peller contains a known number of beads. A blood sample of known volume is added to the rube. Then the appropriate reagent is added to the mixture. By comparing the number of positive cell events to the number of bead events in the flow cytomerry data ourpur, posicive cells per unit volume of blood sample can be computed.

TRUCOUNT Counting Control Beads are fluorescent particles supplied as suspensions in three bead concentrations, High, Medium and Low. The fluorescence of these beads is similar to that of the Absolute Count Tube in FL1 and FL3, but is different in FL2, so that both beads may be run in the same sample, and distinguished from one another. Control beads are added to samples of normal blood prepared with TriTEST reagents and Absolute Count Tubes. The Concrol Bead count is obtained by comparison to the count of the Absolute Count Tubes. These experimental counts are compared to the expected count on the Control Bead label, to check the accuracy of pipetting.

AI/ML Overview

The provided document, K970836, is a 510(k) submission for the TRUCOUNT Absolute Count Tubes and TRUCOUNT Counting Control Beads by Becton Dickinson Immunocytometry Systems, cleared in 1997. It describes the device, its intended use, comparison to predicate devices, and performance data. However, the document does not present a typical "acceptance criteria and study that proves the device meets the acceptance criteria" in the format commonly seen for contemporary medical devices. Instead, it describes performance testing completed to support the device's claims for substantial equivalence.

Here's an analysis of the provided information within the requested framework, highlighting where information is not explicitly stated or is inferred from the context of an older 510(k) submission.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" with quantitative targets. Instead, it describes performance characteristics that were assessed and found "acceptable" to support the device's claims. The reported performance is qualitative for stability and accuracy.

Acceptance Criteria Category (Implied)Specific Criteria (Inferred/Stated)Reported Device Performance
StabilitySupport stated shelf-life.- TRUCOUNT Absolute Count Tube: Found acceptable to support a one-year shelf life. - TRUCOUNT Counting Control Beads: Found acceptable to support a six-month shelf life. - Absolute Count Tubes: Remain stable after foil pouch opened if stored closed at room temperature with blue desiccant.
Accuracy/ReproducibilitySupport ranges for acceptable performance cited in the Package Insert.- Counting Control Beads: Determined and found acceptable to support the ranges for acceptable performance cited in the Package Insert.
Functionality (with FACS Loader)Function as expected when used with FACS Loader and manually.- Absolute Count Tubes and Counting Control Beads: Seen to function as expected when used with the FACS Loader, as well as manually.

2. Sample Size Used for the Test Set and Data Provenance

The document states: "Performance of the product was established by testing at Becton Dickinson Immunocytometry Systems laboratories in San Jose, California."

  • Test Set Sample Size: Not explicitly stated. The document doesn't provide numerical sample sizes for any of the performance tests (stability, accuracy, reproducibility, or functionality).
  • Data Provenance:
    • Country of Origin: United States (San Jose, California).
    • Retrospective or Prospective: Not explicitly stated, but the description of "testing at Becton Dickinson Immunocytometry Systems laboratories" suggests prospective internal testing conducted for the submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device (TRUCOUNT Absolute Count Tubes and Counting Control Beads) is a calibration/control accessory for flow cytometry. Its performance is based on intrinsic physical and chemical properties and its ability to provide a known count of beads for volumetric measurements. Ground truth for its performance would typically be established through manufacturing controls, certified reference materials, and metrological standards, rather than expert consensus on clinical images or interpretations.

4. Adjudication Method for the Test Set

Not applicable. As noted above, the performance assessment is based on quantifiable physical properties and functional checks, not on interpretation by human experts that would require adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done, and is not applicable for this type of device. An MRMC study is typically used for diagnostic interpretation devices where human readers' performance is being evaluated, especially with and without AI assistance. This device is an accessory that assists in counting cells, not in interpreting diagnostic images or clinical data.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Parts of the testing could be considered "standalone" in the sense that the device's intrinsic characteristics (like bead count, fluorescence, stability) are being evaluated. However, the device's ultimate "performance" is as an accessory used with other devices (flow cytometers and reagents) and by human operators. The document mentions "functions as expected when used with the FACS Loader, as well as manually," indicating an evaluation of its performance in conjunction with other components of the workflow, rather than a purely standalone algorithm.

7. The Type of Ground Truth Used

The ground truth for the device's performance is implicitly tied to:

  • Manufacturing Specifications / Certified Bead Counts: For the TRUCOUNT Absolute Count Tubes, the "known number of beads" in the tube constitutes a pre-defined ground truth established during manufacturing and rigorously controlled.
  • Expected Counts (for Control Beads): For the TRUCOUNT Counting Control Beads, the "expected count on the Control Bead label" serves as the ground truth against which experimental counts are compared to check pipetting accuracy.
  • Established Analytical Methods: Stability is assessed against established protocols for degradation (e.g., measuring physical or chemical changes over time).

This is an analytical ground truth based on device design and manufacturing controls, not clinical outcomes, pathology, or expert consensus on a diagnostic finding.

8. The Sample Size for the Training Set

Not applicable. This device is a consumable accessory for flow cytometry, not an AI/ML algorithm that requires a training set. The performance testing described is for product validation, not algorithm training.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

K970836

Attachment D Summary of Safety & Effectiveness

AUG 26 1997

Submitter Information (21 CFR 807.92(a)(1))

Submitter:Becton Dickinson Immunocytometry Systems
2350 Qume Drive
San Jose, CA 95131-1807

Anna Longwell, Esq. Contact: Director, Regulatory Affairs - Corporate (408) 954-2254

Summary Date: February 24, 1997

Name of Device and Classification (21 CFR 807-92(2)(2))

TRUCOUNT TM Absolute Count Tubes Name:

TRUCOUNTTM Counting Control Beads

Classification: Accessories to a Class II Product, Lymphocyte immunophenoryping reagent

Predicate Device (21 CFR 807,92(a)(3))

Lymphocyre immunophenotyping reagents, when used for absolute counts are subscantially equivalent to Simultest TM reagents, cleared in the Simultest IMK-Lymphocyce Kic, (K913192), plus ADCC, or, in the case of TriTEST™ reagents CD3/CD4/CD45 and CD3/CD8/CD45, to FACSCount, K933486. The absolute count application is the subject of current lymphocyte immunophenotyping reagent submissions.

Description of the Device (21 CFR 807.92(a)(4))

TRUCOUNT Absolute Count Tubes are fluorescent particles supplied as a peller in a rube. The peller contains a known number of beads. A blood sample of known volume is added to the rube. Then the appropriate reagent is added to the mixture. By comparing the number of positive cell events to the number of bead events in the flow cytomerry data ourpur, posicive cells per unit volume of blood sample can be computed.

TRUCOUNT Counting Control Beads are fluorescent particles supplied as suspensions in three bead concentrations, High, Medium and Low. The fluorescence of these beads is similar to that of the Absolute Count Tube in FL1 and FL3, but is different in FL2, so that both beads may be run in the same sample, and distinguished from one another. Control beads are added to samples of normal blood prepared with TriTEST reagents and Absolute Count Tubes. The Concrol Bead count is obtained by comparison to the count of the Absolute Count Tubes. These experimental counts are compared to the expected count on the Control Bead label, to check the accuracy of pipetting.

{1}------------------------------------------------

Summary of Safety and Effectiveness

Intended Use (21 CFR 807.92(a)(5))

TRUCOUNT Absoluce Count Tubes are intended for use as an accessory to TriTEST in vitro diagnostics, such as that described in K965053 (CD3/CD4/CD45), to allow computation of positive cells per known volume of blood, using flow cytometty.

The TRUCOUNT Counting Control Beads are provided in three levels, and are intended to provide a means to check the reproducibility and accuracy of volumeric pipetting of blood samples. They are an accessory to TriTEST reagens, when used with TRUCOUNT Absolute Count Tubes to obtain absolute counts of positive cells using flow cytomerry.

Indications for Use

  • For erythrocyte lysed whole blood .
  • For use with FACS Loader ●
  • For use with IVD immunophenotyping reagents .
  • For use on flow cytometers with designated detection ranges ●

Clinical Utility

The determination of absolute counts of subtypes of lymphocytes has been found useful in monitoring some forms of immunodeficiency and autoimmune disease.

Comparison to Predicate Device (21 CFR 807-92(a)(6))

Use of the TRUCOUNT Absolute Count Tubes with each of the TriTEST reagents was compared to either FACSCount, K933486 (for CD3/CD4/CD45 and CD3/CD45), or to the appropriate vial of IMK-Lymphocyte, K913192, plus ADCC. Substantial equivalence of the beads when used with other flow cytomerry reagents is included in the respective reagent submission (e.g., TriTEST CD3/CD4/CD45, K965053). Data for these comparisons is not included in this submission.

Use of the TRUCOUNT Counting Control Beads does not change the safety and effectiveness of the above assays.

Performance Data (21 CFR 807.92(b)(2))

Performance of the product was established by testing at Becton Dickinson Immunocytomerry Systems laboratories in San Jose, California.

  • Stability of the Absolure Count Tube and Counting Control Bead was assessed and found to be . acceptable to support a one year and six month shelf life, respectively.
  • Accuracy and reproducibility of the Counting Control Beads was determined and found to be . acceptable to support the ranges for acceptable performance cited in the Package Insert.
  • Absolute Count Tubes were found to remain stable after the foil pouch is opened, as long as the . pouch is stored closed at room temperature, and the desiccant remains blue.
  • The Absolute Count Tubes and Counting Control Beads were seen to function as expected when . used with the FACS Loader, as well as manually.

2

{2}------------------------------------------------

Summary of Safety and Effectiveness

Performance Dara-Conclusions (21 CFR 807,92(b)(3))

The results of the studies demonstrate that:

  • Expiration dating claim for the TRUCOUNT Absolute Count Tube Product is one year. .
  • Expiration dating claim for the TRUCOUNT Counting Control Beads is six months. .
  • TRUCOUNT Absolute Count Tube and TRUCOUNT Counting Control Bead Product may be . used with the FACS Loader.

The product does not change the subsrantial equivalence of the reagents it is used with, as demonstrated in the respective reagent submissions.

: :

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Anna Longwell, Esq. Director, Regulatory Affairs-Corporate Becton Dickinson Immunocytometry Systems 2350 Qume Drive San Jose, California 95131-1807 AUG 26 1997 Re : K970836/S2 Becton Dickinson TruCount™ Absolute Count Tubes and Trade Name: TruCount™ Counting Control Beads Requlatory Class: II Product Code: GKZ Dated: June 5, 1997 Received: June 10, 1997

Dear Ms. Longwell:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name: Incuquant for bout to Count Tubes program Counting Control Sends

Indications For Use:

  • For use with FACS Loader �
  • For use with in vitro diagnostic immunophenotyping reagents �
  • For use in erythrocyte lysed whole blood �
  • For use to obtain absolute counts by flow cytometry ◆

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
--------------------------------------------------------
Division Sign
Division of Clinical Chemistry Devices
510(k) Number
Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use
(Optional Format 1-2-96)

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”