(338 days)
For in vitro diagnostic use to identify and enumerate percentages and absolute counts of CD3+ and CD3+CD4+ lymphocytes in blood.
For use with any flow cytometer equipped with a 488 nm laser and capable of detection in the ranges: 515-545 nm, 562-607 nm, and > 650 nm.
For use with erythrocyte-lysed peripheral whole blood.
For use with or without an isotype control.
To characterize and monitor some forms of autoimmune disease.
To characterize and monitor some forms of immunodeficiency disease, such as in HIV-infected individuals.
The BDIS TriTEST CD3 fluorescein isothiocyanate (FITC)/CD4 phycoerythrin (PE)/CD45 peridinin chlorophyll protein (PeCP) reagent is a three-color, direct immunofluorescence reagent for identifying and enumerating percentages of T lymphocytes (CD3+) and T-helper (CD3+CD4+) cells in erythrocyte-lysed whole blood (LWB). When used with TRUCOUNT Absolute Count Tubes, the product will yield absolute counts in cells/UL. The Becton Dickinson TriTEST/TRUCOUNT system for immunophenotyping consists of a flow cytometer (either from BDIS or from another manufacture), conjugated monoclonal reagent (TriTEST CD3 FITC/CD4 PE/CD45 PerCP) and TRUCOUNT Absolute Count Tubes.
The process to obtain lymphocyte subset percentages includes: 1) obtaining a whole blood sample, 2) cell-surface antigen staining with three-color monoclonal antibody reagents, 3) erythrocyte lysis, and 4) flow cytometric acquisition and analysis of list mode data. Analysis involves computing the ratio of reagent-positive events (CD3+CD4+) to the CD45 positive events, and expressing the ratio as a percentage.
To obtain absolute counts, the TriTEST reagent and whole blood are added directly to an Absolute Count Tube prior to lysis. The remaining process, until analysis, is identical to that for percentages. Analysis for absolute counts requires that an additional region, be identified and the events in this region counted. The proportion of reagent positive events to bead events (P) is computed. The absolute count is P x (beads/pellet)/(volume of blood sample).
When monoclonal antibody reagents are added to human whole blood, the fluorochrome-labeled antibodies bind specifically to antigens on the surface of leukocytes, thus identifying lymphocyte populations. The patient blood sample is added to the counting bead pellet and is treated with fluorochrome-labeled antibodies and the erythrocytes are lysed with FACS Lysing Solution. The flow cytometer is set up so that cell populations for most samples occupy approximately the same region of fluorescence space. The sample is then introduced into the flow cytometer and the stained cells and beads fluoresce when excited by a laser beam.
The three-color reagent permits identification of lymphocyte subsets using fluorescence gating instead of forward scatter gating. This three-color reagent allows direct gating on the CD45-positive population using a combination of fluorescence and side scatter parameters. By gating on the CD45-positive population, a maximum number of lymphocytes may be captured in the gate and non-lymphocyte contamination may be minimized.
Here's an analysis of the provided text, focusing on the acceptance criteria and the study that proves the device meets those criteria:
Device Name: Becton Dickinson TriTEST24 reagent CD3 FITC/CD4 PE/CD45 PerCP; TRUCOUNTTM Absolute Count Tubes
Intended Use: For in vitro diagnostic use to identify and enumerate percentages and absolute counts of CD3+ and CD3+CD4+ lymphocytes in blood.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state numerical acceptance criteria in a clear, quantitative manner. Instead, performance is generally described in terms of "equivalence," "acceptable," and "good correlation" compared to a predicate device or expected behavior.
| Performance Characteristic | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| Accuracy (Absolute Counts CD3+) | Substantial equivalence to predicate device (FACSCount system) | "Accuracy data demonstrated the TriTEST/TRUCOUNT product's equivalence to FACSCount." |
| Accuracy (Absolute Counts CD3+CD4+) | Substantial equivalence to predicate device (FACSCount system) | "Accuracy data demonstrated the TriTEST/TRUCOUNT product's equivalence to FACSCount." |
| Use with or without Isotype Control | Performance should be acceptable and comparable in both scenarios | "Data indicated that the reagent may be used with or without an isotype control." |
| Reference Range Studies | Ability to establish site-specific reference ranges | "The reference range for CD3+ and CD3+CD4+ lymphocytes was a function of sex and site... Each site must determine its own reference range." (Implies the device provides robust data for this purpose). |
| Stability (Age of blood pre-staining) | Acceptable performance over a defined time period post-draw | "Staining the samples within 24 hours of draw and analyzing samples within 24 hours of staining or alternatively, staining the samples within 48 hours of draw and analyzing them within 6 hours is recommended." (Implies stability within these parameters). |
| Stability (Age of stain post-staining) | Acceptable performance over a defined time period post-staining | "Staining the samples within 24 hours of draw and analyzing samples within 24 hours of staining or alternatively, staining the samples within 48 hours of draw and analyzing them within 6 hours is recommended." (Implies stability within these parameters). |
| Within-specimen Reproducibility | Acceptable consistency within replicates from the same specimen | "Results demonstrated acceptable within-sample reproducibility." |
| Linearity | Proportional response over a relevant concentration range | "Results indicate a linear response over this range [16,700 to 200 lymphocytes/UL and from 31,000 to 2,500 WBC/UL]." |
| Cross Reactivity / Specificity | Maintenance of established specificity | "Conjugation and product formulation have not changed their specificity [as reported in literature]." |
| Cross-Platform Reproducibility (Percentages) | Acceptable performance on non-BD devices | "For percentage results, the reagent may be used with flow cytometers not made by Becton Dickinson." |
| Cross-Platform Reproducibility (Absolute Counts) | Good correlation with BD devices, with transparency about potential bias | "There was a small (<20%) non-zero bias, but good correlation between results on a Becton Dickinson flow cytometer versus a Coulter cytometer was observed. Therefore, users will be advised that they must validate performance characteristics for absolute counts." (Implicit acceptance despite bias, with user guidance). |
2. Sample Sizes Used for the Test Set and Data Provenance
- Accuracy Study (Absolute Counts CD3+): 197 donors (152 abnormal, 45 normal).
- Accuracy Study (Absolute Counts CD3+CD4+): Number of donors not explicitly stated, but "data for CD3+CD4+ was grouped according to range of number of cells."
- Within-specimen reproducibility (BDIS lab): 1 high, 1 medium, and 1 low specimen (10 replicates each).
- Within-specimen reproducibility (Clinical sites): n=94 for CD3+ donors, n=80 for CD3+CD4+ donors (3 aliquots from each donor).
- Linearity: Blood samples from 3 normal donors.
- Stability: Whole blood samples were tested at 0, 24, 48, and 72 hours post-draw, and stained samples at 0 and 24 hours from staining. (Number of donors not specified).
- Reference Range: Not specified, but suggests multiple sites.
Data Provenance:
The studies were performed at:
- Cleveland Clinic
- Johns Hopkins Hospital
- Institute of Tropical Medicine
- University of North Carolina
- Becton Dickinson Immunocytometry Systems laboratories in San Jose, California
Indicates a multi-center study involving both a commercial lab and academic/clinical institutions within the USA (Cleveland Clinic, Johns Hopkins, UNC). "Institute of Tropical Medicine" might suggest international data, but the context generally points to the US for the other named sites. The data appears to be prospective or a mix of retrospective and prospective, generated specifically for these performance studies.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not provide this information. The predicate device comparison implies the "ground truth" for accuracy was the performance of the BDIS FACSCount system, which is another device rather than expert consensus on patient samples. For other analyses like reproducibility or linearity, the "ground truth" is inherent to the measurements and expected biological/analytical behavior rather than expert interpretation of a diagnostic outcome.
4. Adjudication Method for the Test Set
Not applicable. The "ground truth" is derived from direct measurements and comparison to a predicate device, not through expert review needing adjudication of subjective interpretations.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not performed. The device is for objective enumeration of cell populations, not for image interpretation or diagnosis that would typically involve multiple human readers. The comparison is between the new device and a predicate device (FACSCount system) for quantitative measurements.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, the studies evaluate the standalone performance of the TriTEST/TRUCOUNT system. While human operators are involved in sample preparation and operating the flow cytometer, the evaluation focuses on the accuracy, reproducibility, and linearity of the device's measurements of cell populations, independent of human diagnostic interpretation. The flow cytometer, reagents, and counting tubes constitute the "algorithm" or system being tested.
7. The Type of Ground Truth Used
The primary ground truth for accuracy was the performance of the predicate device, the BDIS FACSCount system.
For other studies:
- Reproducibility: The consistency of repeated measurements on the same sample.
- Linearity: The expected proportional response to serial dilutions, likely confirmed instrumentally or against established cell count methodologies.
- Stability: Measured values over time compared to baseline.
8. The Sample Size for the Training Set
The document does not explicitly mention a "training set" in the context of machine learning or algorithm development. The studies described are validation studies of an established methodology (flow cytometry with specific reagents and beads) against a predicate device. Therefore, a separate, distinct "training set" as understood in AI/ML is not applicable here. The development of the formulation of the reagent itself would have preceded these validation studies.
9. How the Ground Truth for the Training Set Was Established
As noted above, a "training set" in the context of an AI/ML algorithm is not described. The device is a diagnostic assay kit (reagents and counting tubes) used with a flow cytometer. The "ground truth" for development of such a kit would trace back to established immunological principles and cell biology, validated through prior scientific research and predicate device performance.
{0}------------------------------------------------
K9605053
Attachment C
Summary of Safety and Effectiveness
NOV 21 1997
Submitter Information (21 CFR 807.92(2)(1))
Becton Dickinson Immunocytometry Systems Submitter: 2350 Quine Drive San Jose, CA 95131-1807
- Anna Longwell, Esq. Contact: Director, Regulatory Affairs - Corporate (408) 954-2254
Summary date: November 18, 1997
Name of Device and Classification (21 CFR 807.92(2)(2))
Becton Dickinson TriTEST24 reagent CD3 FITC/CD4 PE/CD45 PerCP; Name: TRUCOUNTTM Absolute Count Tubes
Classificacion: Class II
Predicate Device (21 CFR 807.92(=1(3))
The BDIS TrillEST™ CD3/CD4/CD45 reagent with TRUCOUNT Absolute Count Tubes is substantially equivalent to the BDIS FACSCount system cleared in K933486.
Description of the Device (21 CFR 807.92(a)(4))
The BDIS TrilleST CD3 flucescein isothiogranate (FITC)/CD4 phycoerythrin (PE)/CD45 peridinin chlorophyll protein (PeCP) reagent is a three-color, direct immunofluqescence reagent for identifying and enumerating percentages of T lymphocytes (CD3+) and T-helper (CD3+CD4+) cells in aythrocyte-lysed whole blood (LWB). When used with TRUCOUNT Absolute Count Tubes, the product will yield absolute counts in cells/UL. The Becton Dickinson TrillEST/TRUCOUNT system for immunophenotyping consists of a flow cytomere (cither from BDIS or from another manufacture), conjugated monoclonal reagent (TriTEST CD3 FITC/CD4 PE/CD45 PerCP) and TRUCOUNT Absolute Count Tubes.
The process to obtain lymphocyte subset percentages includes: 1) obtaining a whole blood sample, 2) cell-surface antigen staining with three-color monoclonal antibody reagents, 3) cythrocyte lysis, and 4) flow cytometric acquisition and analysis of list mode data. Analysis involves computing the ratio of reagent-positive events (CD3+CD4+) to the CD45 positive events, and expressing the ratio as a percentage.
{1}------------------------------------------------
Summary of Safety and Effectiveness
To obtain absolute counts, the TriTEST reagent and whole blood are added directly to an Absolute Count Tube prior to lysis. The remaining process, until analysis, is identical to that for percentages. Analysis for absolute counts requires that an additional region, be identified and the events in this region counted. The proportion of reagent positive wents to bead events (P) is computed. The absolute count is I' x (beads/peller)/(volume of blood sample).
When monoclonal antibody reagents are added to human whole blood, the flucrochrome-labeled antibodies bind specifically to antigens on the surface of leucocytes, thus identifying lymphogye populations. The patient blood sample is added to the counting bead pellet and is treated with fluctochrome-labeled antibodies and the cythrocytes are lysed with FACS® Lysing Solution. The flow cytomerer is set up so that cell populations for most samples occupy approximately the same region of fluorescence space. The sample is then introduced into the flow cyrometer and the stained cells and beads fluoresce when excited by a laser beam.
The three-color reagent permits identification of lymphocyte subsets using fluorescence gating instead of forward scatter gating. This three-color reagent allows direct gating on the CD45-positive population using a combination of flucescence and side scatter parameters. By gating on the CD45positive population, a maximum number of lymphocytes may be captured in the gate and nonlymphocyte contamination may be minimized.
Intended Use (21 CFR 807.92(a)(5))
For in vitro diagnostic use to identify and enumerate percentages and absolute counts of CD3+ and CD3+CD4+ lymphocytes in blood.
Indications for Use
- For use with any flow cytometer equipped with a 488 nm laser and capable of detection ♥ in the ranges: 515-545 nm, 562-607 nm, and > 650 nm
- For use with erythrocyte-lysed peripheral whole blood .
- For use with or without an isotype control .
- To characterize and monitor some forms of autoimmune disease .
- To characterize and monitor some forms of immunodeficiency disease, such as in HIV-infected . individuals
Clinical Urility
The determination of CD3+CD4+ lymphocytes has been found useful in monitoring some forms of immunodeficiency and autoimmune disease.
{2}------------------------------------------------
Summary of Safety and Effectiveness
Comparison to Predicate Device (21 CFR 807.92(a)(6))
The three-color reagent is substantially equivalent to the predicate BDIS FACSCount System in that they share the same intended uses and similar methodologies. Results demonstrate that the products yield essentially equivalent performance characteristics. Both tests are antibody-dependent, in vitro diagnostic assays to identify T lymphocytes (CD3+) and helper/induce (CD3+CD4+) Tlymphocyte subsets. The products differ in the steps used to determine analysis gates to identify the lymphocyte population.
Performance Data (21 CFR 807.92(b)(2))
Performance of the product was established by testing at Cleveland Clinic, Johns Hopkins Hospitul, Institute of Tropical Medicine, University of North Carolina, and at Becton Dickinson Immunocytometry Systems laboratories in San Jose, California.
Several studies were performed:
- Accuracy was determined by comparison to the FACSCount system. Absolute counts for CD3+ t were obtained from 197 (152 abnormal and 45 normal) dones. Absolute counts for CD3+CD4+ were obtained from dones; data for CD3+CD4+ was grouped according to range of number of cells. Accuray data demonstrated the TriTEST/TRUCOUNT product's equivalence to FACSCount. Additionally, a subser of the accuracy study data was analyzed for percent CD3+ and percent CD4+ first using a control to set matkers and then using only the stained sample to set quadrant markers. Data indicated that the reagent may be used with or without an isorype control.
- Reference range studies were performed. The reference range for CD3+ and CD3+CD4+ . lymphocytes was a function of sex and site. For CD3+ lymphocytes, within the male population, it was also a function of age. Each site must determine its own reference range.
- A stability study was conducted to assess the time effect relating to age of blood (time-from-. draw) and the time effect relating to the age of the stain (time-from-sample preparation), as well as the combined effect of both. Stability was decermined for both percentages and absolute counts. Stability was decemined for whole blood samples at 0, 24, 48 and 72 hours post draw. Additionally, stability was measured for stained samples at 0 and 24 hours from time of staining. A combination of the two, time from draw plus time from sample preparation was srudied. Results indicated that either 1) staining the samples within 24 hours of daw and analyzing samples within 24 hours of staining or alternatively, 2) staining the samples with 48 hours of draw and analyzing them within 6 hours is recommended.
{3}------------------------------------------------
Summary of Safety and Effectiveness
- Within-soccimen reproducibility was performed at BDIS; 10 replicates from 1 high, 1 medium, and 1 low (with respect to CD3+ and CD3+CD4+ lymphocytes) were assessed. Withinspecimen reproducibility was also performed at 3 clincial sites; 3 aliquots from each donor (n-94 for CD3+ and n=80 for CD3+CD4+) were assessed. Results demonstrated acceptable withinsample reproducibility.
- Linearity was determined using blood samples from 3 normal dones diluted to 5 . concentrations, ranging from 16,700 to 200 lymphoorcs/JLL and from 31,000 to 2,500 WBC/UL. Results indicate a linear response over this range.
- Cross reactivity of these clones is reported in the literature. Conjugation and product . formulation have not changed their specificity.
- Results from a cross platform reproducibility study indicated 1) for percentage results the reagent . may be used with flow cytometers not made by Becton Dickinson 2) for determining absolute counts usingTriTEST reagent with TRUCOUNT Absolute Count Tubes, there was a small (<20%) non-zoo bias, but good correlation between results on a Becton Dickinson flow cytomere versus a Coulte cytomere was observed. Therefore, users will be advised that they must validate performance characteristics for absolute counts.
Performance Data - Conclusions (21 CFR 807.92(b)(3))
The results of the clinical studies demonstrate that the device is as safe and effective as the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Anna Longwell, Esq. Director, Regulatory Affairs - Corporate Becton Dickinson Immunocytometry Systems 2350 Qume Drive San Jose, California 95131-1807
NOV 2 1 1997
Re : K965053 Becton Dickinson TriTEST™ Reagent CD3 FITC/CD4 Trade Name: PE/CD45 PerCP; TRUCOUNT™ Absolute Count Tubes Regulatory Class: II Product Code: GKZ Dated: Auqust 28, 1997 September 2, 1997 - - - - - -Received:
Dear Ms. Longwell:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) , it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
{5}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) Additionally, for questions on the promotion and 594-4588. advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Attachment A TriTEST CD3/CD4/CD45 with TRUCOUNT Control Beads & Absolute Count Tubes K965053
Indications for Use Statment
- For use with any flow cytometer equipped with a 488 nm laser and capable of . detection in the ranges: 515-545 nm, 562-607 nm, and > 650 nm
- For use with erythrocyte-lysed peripheral whole blood .
- For use with or without an isotype control .
- To characterize and monitor some forms of autoimmune diseasc .
- To characterize and monitor some forms of immunodeficiency disease, such as in . HIV-infected individuals
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Artu E. Makeri
Division Sign-Off) Division of Clinical Laboratory De 510(k) Number
Prescription Use (Per 21-CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
§ 864.5220 Automated differential cell counter.
(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”