(230 days)
The Early Vue VS30 is intended to measure, display, alarm and, record physiological information of adult, pediatric and neonatal patients in hospitals and in out-of-hospital patient care is administered by a healthcare professional (such as clinics, outpatient surgery facilities, long-term care facilities, and physician offices). It is not intended for use in mobile settings such as ambulances and aircraft. Clinical users may use the monitor during patient transport within a healthcare facility.
The intended use of measurement parameters for each patient type is shown in table below (Parameter (patient types)): Non-invasive BP--NBP (Adult, Pediatric, Neonatal); SpO2 --Masimo rainbow SET or Philips FAST (Adult, Pediatric, Neonatal); Pulse Rate -- PR -- derived from SpO2 or NBP (Adult, Pediatric, Neonatal); Temperature-Predictive, Temporal (Adult, Pediatric, Neonatal); end tidal CO2--etCO2 (Adult, Pediatric, Neonatal); Respiration Rate: airway respiration--awRR (Adult, Pediatric, Neonatal); Respiration Rate: acoustic respiration--RRa (Adult, Pediatric);
Non-invasive Total Hemoglobin--SpHb (Adult, Pediatric)
Contra-indications: Not for transport outside the healthcare facility. Not for use in home setting.
The subject device EarlyVue VS30 is a platform redesigned from the predicate device Suresign VS4 to provide the measurement of the same physiological parameters. It is a multi-parameter compact, portable monitor, offering several configurations and optional features that are intended to meet the customer's needs. The VS30 is designed with a simple, intuitive user interface for ease of operation. It is used to measure or monitor NBP, SpO2, SpHb, RRa, CO2 and optional temperature primarily in non-acute settings.
The subject VS30 also provides the wireless option. The wireless functions for VS30 are for the monitors to communicate with the hospital EMR system or to communicate with Philips IGS software over the hospital's wireless infrastructure.
Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:
Device: EarlyVue VS30 Vital Signs Monitor
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the EarlyVue VS30 are implied by its comparison to the predicate device, Philips SureSigns VS4. The document asserts that the VS30 has the "same fundamental technological characteristics" and that "performance data demonstrates the subject device Early Vue VS30 is as safe and effective as the predicate device for all the measurements." Therefore, the predicate device's specifications serve as the de facto acceptance criteria for each measurement parameter.
| Parameter | Acceptance Criteria (Predicate Device VS4 Performance) | Reported Device Performance (EarlyVue VS30) |
|---|---|---|
| Non-invasive BP (NBP) | Technique: Oscillometric, stepwise deflation. Measurement Range: Adult: Systolic 30-270mmHg, Diastolic 10-240mmHg, Mean 20-250mmHg; Pediatric: Systolic 30-180mmHg, Diastolic 10-150mmHg, Mean 20-160mmHg; Neonate: Systolic 30-130mmHg, Diastolic 10-100mmHg, Mean 20-120mmHg. Accuracy: Max Std. Dev: 8 mmHg, Max Mean Error: +/- 5 mmHg. Derived PR Accuracy: 40-100 bpm: +/- 5 bpm; 101-200 bpm: +/- 5% of reading; 201-300 bpm: +/- 10% of reading. Initial Cuff Inflation: Adult: 160 mmHg; Pediatric: 140 mmHg; Neonatal: 100 mmHg. | Technique: Oscillometric, stepwise deflation. Measurement Range: Adult: Systolic 30-270mmHg, Diastolic 10-240mmHg, Mean 20-250mmHg; Pediatric: Systolic 30-180mmHg, Diastolic 10-150mmHg, Mean 20-160mmHg; Neonate: Systolic 30-130mmHg, Diastolic 10-100mmHg, Mean 20-120mmHg. Accuracy: Max Std. Dev: 8 mmHg, Max Mean Error: +/- 5 mmHg. Derived PR Accuracy: 40-100 bpm: +/- 5 bpm; 101-200 bpm: +/- 5% of reading; 201-300 bpm: +/- 10% of reading. Initial Cuff Inflation: Adult: 160 mmHg; Pediatric: 140 mmHg; Neonatal: 100 mmHg. Reported as "Identical" |
| SpO2 (Philips SpO2) | Measurement Range: SpO2: 0-100%, Pulse Rate: 30-300 bpm. PR Accuracy: Greater of +2% or +1 bpm. Resolution: SpO2: 1%. Accuracy (70-100%): Philips reusable sensors: +/- 2% (M1191B, BL, A, AL, M1192A), +/- 3% (M1191T, M1192T, M1193T adult, M1193A, M1194A, M1195A, M1196A, T, S), +/- 4% (M1193T neonatal). Philips disposable sensors: +/- 2% (M1132A, M1133A, M1134A), +/- 3% (M1131A). | Measurement Range: SpO2: 0-100%, Pulse Rate: 30-300 bpm. PR Accuracy: Greater of +2% or +1 bpm. Resolution: SpO2: 1%. Accuracy (70-100%): Philips reusable sensors: +/- 2% (M1191B, BL, A, AL, M1192A), +/- 3% (M1191T, M1192T, M1193T adult, M1193A, M1194A, M1195A, M1196A, T, S), +/- 4% (M1193T neonatal). Philips disposable sensors: +/- 2% (M1132A, M1133A, M1134A), +/- 3% (M1131A). Reported as "Identical" |
| SpO2 (Masimo SpO2) | Measurement Range: 0-100%. Resolution: 1%. Accuracy: No motion: 60-80 +/- 3% (adult/ped/inf), 70-100 +/- 2% (adult/ped/inf), 70-100 +/- 3% (neonates). Motion: 70-100 +/- 3% (all). Low perfusion: 70-100 +/- 2% (all). | Measurement Range: 0-100%. Resolution: 1%. Accuracy: No motion: 60-80 +/- 3% (adult/ped/inf), 70-100 +/- 2% (adult/ped/inf), 70-100 +/- 3% (neonates). Motion: 70-100 +/- 3% (all). Low perfusion: 70-100 +/- 2% (all). Reported as "Identical" |
| Pulse Rate (Masimo SpO2) | Range: 25-240 bpm. Resolution: 1 bpm. Accuracy: +/- 5 bpm motion or low perfusion (worst case). | Range: 25-240 bpm. Resolution: 1 bpm. Accuracy: +/- 5 bpm motion or low perfusion (worst case). Reported as "Identical" |
| SpHb (Masimo) | Range: 0-25 g/dL. Accuracy: 8-17 g/dL (arterial/venous) adults/pediatrics: +/- 1 g/dL. Resolution: 0.1 g/dL. | Range: 0-25 g/dL. Accuracy: 8-17 g/dL (arterial/venous) adults/pediatrics: +/- 1 g/dL. Resolution: 0.1 g/dL. Reported as "Identical" |
| Respiration Rate (RRa - Masimo) | Range: 0-70 bpm. Accuracy: 4-70 bpm: +/- 1 bpm (adults >30kg and adults/pediatrics >10kg). Resolution: 1 bpm. | Range: 0-70 bpm. Accuracy: 4-70 bpm: +/- 1 bpm (adults >30kg and adults/pediatrics >10kg). Resolution: 1 bpm. Reported as "Identical" |
| Temperature (Predictive) | Measurement Range: 26.7–43.3°C (80–110°F). Accuracy (monitoring mode): +/- 0.1°C (+/- 0.2°F). Measurement Time: Oral: 4-6s, Adult axillary: 12-15s, Pediatric axillary: 10-13s, Rectal: 10-13s. Resolution: 0.1°C. Measurement Sites: Oral, rectal, axillary. | Measurement Range: 26.7–43.3°C (80–110°F). Accuracy (monitoring mode): +/- 0.1°C (+/- 0.2°F). Measurement Time: Oral: 4-6s, Adult axillary: 12-15s, Pediatric axillary: 10-13s, Rectal: 10-13s. Resolution: 0.1°C (+/- 0.1°F). Measurement Sites: Oral, rectal, axillary. Reported as "Identical" |
| Temperature (Temporal) | Range: 16°C - 43°C. Clinical Accuracy: +/- 0.2°F or 0.1°C per ASTM E1112. Arterial heat balance range: 34.5°C - 43°C (Normal: 35.9°C to 37.5°C). Resolution: 0.1°C or 0.1°F. Response Time: <=1 second. | Range: 16°C - 43°C. Clinical Accuracy: +/- 0.2°F or 0.1°C per ASTM E1112. Arterial heat balance range: 34.5°C - 43°C (Normal: 35.9°C to 37.5°C). Resolution: 0.1°C or 0.1°F. Response Time: <=1 second. Reported as "Identical" |
| end tidal CO2 (etCO2) | Measurement Range: 0-150 mmHg. Sampling Rate: 20 samples/second. Flow Rate: 50ml/min, +15ml/min, -7.5ml/min. Warm-up Time: 30 seconds (typical). Calibration Interval: Initial: 1200 hrs or 1 year; Subsequent: 1 year or 4000 hrs. Accuracy: 0-38 mmHg: +/- 2 mmHg; 39-150 mmHg: +/-(5% of reading + 0.08% for every 1 mmHg above 38 mmHg). Response Time: ~3.9 seconds. | Measurement Range: 0-150 mmHg. Sampling Rate: 20 samples/second. Flow Rate: Same. Warm-up Time: 30 seconds (typical). Calibration Interval: Initial: 1200 hrs or 1 year; Subsequent: 1 year or 4000 hrs. Accuracy: 0-38 mmHg: +/- 2 mmHg; 39-150 mmHg: +/-(5% of reading + 0.08% for every 1 mmHg above 38 mmHg). Response Time: ~3.9 seconds. Reported as "Identical" |
| Respiration Rate (awRR - CO2) | Range: 0-150 respirations/minute. Accuracy: +/- 1 rpm (0-70 rpm); +/- 2 rpm (71-120 rpm); +/- 3 rpm (121-150 rpm). | Range: 0-150 respirations/minute. Accuracy: +/- 1 rpm (0-70 rpm); +/- 2 rpm (71-120 rpm); +/- 3 rpm (121-150 rpm). Reported as "Identical" |
| Battery operation time | 3 hours | 6 hours (Improved) |
| User Interface | 8.4" resistive touchscreen, fixed key switches. | 10.1" LCD touchscreen, soft keys. Verified and validated, passed verification and validation. |
| Monitor-to-monitor communication | No (for predicate) | Yes (Non-medical application, allows viewing saved patient records remotely, viewing last saved record for up to 100 patients, viewing trends. Fully verified and passed tests.) |
| RFID scanner | No (for predicate) | Yes (Non-medical application, entry of patient ID and user ID. Fully verified and passed tests.) |
| Operating Temperature | 10°C to 40°C monitor (16°C to 33°C with tympanic, 0°C to 50°C Masimo, 16°C to 40°C Exergen, 10°C to 40°C WelchAllyn) | 10°C to 40°C monitor (10°C to 40°C WelchAllyn, 16°C to 40°C Exergen) Reported as "Same" |
| Software Operating System | Win CE | Linux (recoded software). Implied acceptance through comprehensive software verification and validation. |
| Hardware Components | VS4 hardware | New CPU, PCAs. Implied acceptance through comprehensive hardware verification and validation. |
| Wireless Module | VS4 wireless module | Different from VS4, but from same supplier, has FCC certificate and ID. Tested to and passed ANSI C63.27-2017. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes (number of patients, number of measurements) used for performance testing (e.g., NBP accuracy, SpO2 accuracy). It generally refers to "verification and validation" and "performance data."
For the SpO2 and SpHb measurements using Masimo technology, the document references the Masimo RRa respiration rate K110028 and the Oridion MicroMediCO2 module K123690. The accuracy claims for these integrated modules would presumably be based on studies performed during their original clearance, but specific details about those test sets are not in this document.
The document does not specify the country of origin of the data or whether the studies were retrospective or prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not provide information on the number of experts or their qualifications for establishing ground truth for the test set. For physiological measurements like NBP, SpO2, and temperature, ground truth is typically established using reference methods (e.g., arterial catheter for NBP, co-oximetry for SpO2, calibrated thermometer for temperature) rather than expert consensus on interpretive tasks.
4. Adjudication Method for the Test Set
The document does not describe any adjudication method used for the test set, as this is typically applicable to subjective interpretations by multiple experts (e.g., radiology reads), not direct physiological measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
This information is not applicable to this submission. The EarlyVue VS30 is a vital signs monitor that directly measures physiological parameters. It does not involve "human readers" interpreting medical images or other data with or without "AI assistance." The device's function is to provide direct measurements, not to aid in human diagnostic interpretations.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, the primary function of the EarlyVue VS30 is as a standalone device that performs physiological measurements without a human-in-the-loop for the measurement process itself. The "performance data" referred to throughout the document would be standalone performance of the device's measurement functions.
7. The Type of Ground Truth Used
The ground truth for the physiological measurements (NBP, SpO2, Temperature, CO2) would typically be established using:
- Reference Devices/Methods:
- NBP: Invasive arterial blood pressure measurements or comparison to a validated reference sphygmomanometer.
- SpO2: Co-oximetry for arterial oxygen saturation.
- Temperature: Calibrated reference thermometers (e.g., mercury-in-glass, highly accurate electronic thermometers).
- CO2: Calibrated gas analyzers.
The document states that the technologies used are "identical" to the predicate device, which in turn relied on established measurement technologies (Philips picoNBP, Masimo Rainbow SET, Oridion MicroMediCO2 module). These underlying technologies would have their own validation against established ground truth methods.
8. The Sample Size for the Training Set
The document does not provide information on the sample size for a training set. This is because the EarlyVue VS30 is not described as an AI/ML-based device that learns from a training set in the typical sense (e.g., for image classification or prediction). Its functions are based on established physiological measurement principles and algorithms. The software was "recoded" compared to the predicate, implying engineering-based development and verification, rather than a data-driven training process.
9. How the Ground Truth for the Training Set Was Established
Since no explicit training set is mentioned for an AI/ML context, this question is not directly applicable. The device's "ground truth" refers to its accuracy against established medical standards for physiological measurements. The document outlines a comprehensive verification and validation process (software, hardware, environmental, reliability, safety, performance, applicable consensus standards: IEC 60601 series, ISO 80601 series, etc.) to ensure the device performs as intended and is as safe and effective as the predicate.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and consists of the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG ADMINISTRATION".
October 29, 2019
Philips Medizin Systeme Boeblingen GmbH Greg Li Regulatory Affairs Specialist Hewlett-Packard Strasse 2 Boeblingen, 71034 DE
Re: K190624
Trade/Device Name: EarlyVue VS30 Regulation Number: 21 CFR 870.1100 Regulation Name: Blood pressure alarm Regulatory Class: Class II Product Code: DSJ, DSK, DXN, DOA, DSA, FLL Dated: September 23, 2019 Received: September 24, 2019
Dear Greg Li:
This letter corrects our substantially equivalent letter of October 27, 2019.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Stephen C. Browning -S5
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K190624
Device Name EarlyVue VS30
Indications for Use (Describe)
The Early Vue VS30 is intended to measure, display, alarm and, record physiological information of adult, pediatric and neonatal patients in hospitals and in out-of-hospital patient care is administered by a healthcare professional (such as clinics, outpatient surgery facilities, long-term care facilities, and physician offices). It is not intended for use in mobile settings such as ambulances and aircraft. Clinical users may use the monitor during patient transport within a healthcare facility.
The intended use of measurement parameters for each patient type is shown in table below (Parameter (patient types)): Non-invasive BP--NBP (Adult, Pediatric, Neonatal); SpO2 --Masimo rainbow SET or Philips FAST (Adult, Pediatric, Neonatal); Pulse Rate -- PR -- derived from SpO2 or NBP (Adult, Pediatric, Neonatal); Temperature-Predictive, Temporal (Adult, Pediatric, Neonatal); end tidal CO2--etCO2 (Adult, Pediatric, Neonatal); Respiration Rate: airway respiration--awRR (Adult, Pediatric, Neonatal); Respiration Rate: acoustic respiration--RRa (Adult, Pediatric);
Non-invasive Total Hemoglobin--SpHb (Adult, Pediatric)
Contra-indications: Not for transport outside the healthcare facility. Not for use in home setting.
| Type of Use (Select one or both, as applicable) |
|---|
| 区 Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Philips Healthcare. The word "PHILIPS" is written in large, bold, blue letters. Below the word "PHILIPS" is the text "Philips Healthcare" written in a smaller, black font.
510K Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807-92(c).
-
- The submitter and contact person of this pre-market notification is:
The submitter/manufacturer:
- The submitter and contact person of this pre-market notification is:
Philips Medizin Systeme Böblingen GmbH Hewlett-Packard Strasse 2 71034 Böblingen Germany
Contact person:
Greg Li Philips Healthcare 3000 Minuteman Road Andover, MA 01810 United States
Tel: 978-659-4227 Fax: 978-659-7323 Email: Greg.li@philips.com
-
- Date this summary is prepared: This summary is prepared on October 24, 2019.
-
- Subject device
Name of Device: EarlyVue VS30 Vital Signs Monitor Trade Name: EarlyVue VS30 Common or Usual name: multi-parameter patient monitor Regulatory Class: II Classification names and Product Code:
| Device Panel | Classification | Product Code | Description |
|---|---|---|---|
| Cardiovascular | 870.1100, II | DSJ | Alarm, Blood Pressure |
| 870.1110, II | DSK | Computer, Blood Pressure | |
| 870.1130, II | DXN | System, Measurement, BloodPressure, Non-Invasive |
{4}------------------------------------------------
| Device Panel | Classification | Product Code | Description |
|---|---|---|---|
| 870.2700, II | DQA | Oximeter | |
| 870.2900, II | DSA | Cable, transducer and electrode,patient connector | |
| GeneralHospital andPersonal Use | 880.2910, II | FLL | Thermometer, Electronic, Clinical |
| Anesthesiology& RespiratoryTherapy | 868.1400, II | CCK | Analyzer, Gas |
4. Predicate Device:
Philips SureSigns VS4 Vital Signs Monitor cleared under K163649, K151761, K133961.
Reference Device: Philips Efficia CM patient monitors with 510(k) K151812
5. Product Description
The subject device EarlyVue VS30 is a platform redesigned from the predicate device Suresign VS4 to provide the measurement of the same physiological parameters. It is a multi-parameter compact, portable monitor, offering several configurations and optional features that are intended to meet the customer's needs. The VS30 is designed with a simple, intuitive user interface for ease of operation. It is used to measure or monitor NBP, SpO2, SpHb, RRa, CO2 and optional temperature primarily in non-acute settings.
The subject VS30 also provides the wireless option. The wireless functions for VS30 are for the monitors to communicate with the hospital EMR system or to communicate with Philips IGS software over the hospital's wireless infrastructure.
The differences in the technological characteristics are as follows:
- a) The subject VS30 does not provide Tympanic temperature measure whilst the predicate VS4 does have this option
- b) The outside appearance of the VS30 monitor looks different. User interface is changed comparing to the predicate VS4.
- c) The VS30 has different hardware comparing to VS4, including new Central Processing Unit (CPU), Printed Circuit Board Assemblies (PCAs).
{5}------------------------------------------------
- d) The VS30 software operating system is Linux while VS4 uses Win CE. The software is recoded.
- e) The wireless module used on VS30 is different from VS4. They are both from the same supplier. The VS30 wireless module has FCC certificate and FCC identification.
- f) The subject VS30 has a new feature called monitor-to-monitor communication.
- g) VS30 added RFID scanner for entry of patient ID and user ID.
-
- The subject device EarlyVue VS30 has the same Intended Use as the legally marketed predicate device, SureSigns VS4 (the predicate VS4 has Tympanic temperature measurement but the subject VS30 does not have this measurement).
The indications for use for the subject EarlyVue VS30 is as following:
The Early Vue VS30 is intended to measure, display, alarm and, record physiological information of adult, pediatric and neonatal patients in hospitals and in out-of-hospital patient care settings in which care is administered by a healthcare professional (such as clinics, outpatient surgery facilities, long-term care facilities, and physician offices). It is not intended for use in mobile settings such as ambulances and aircraft. Clinical users may use the monitor during patient transport within a healthcare facility.
The intended use of measurement parameters for each patient type is shown in table below:
| Patient Types | |||
|---|---|---|---|
| Parameter | Adult | Pediatric | Neonatal |
| Non-invasive BP (NBP) | √ | √ | √ |
| SpO2 (Masimo rainbow SET orPhilips FAST) | √ | √ | √ |
| Pulse Rate (PR) derived from SpO2or NBP | √ | √ | √ |
| Temperature (Predictive, Temporal) | √ | √ | √ |
| end tidal CO2 (etCO2) | √ | √ | √ |
| Respiration Rate: airway respiration(awRR) | √ | √ | √ |
| Respiration Rate: acousticrespiration (RRa) | √ | √ | |
| Non-invasive Total Hemoglobin(SpHb) | √ | √ |
{6}------------------------------------------------
Contra-indications: Not for transport outside the healthcare facility. Not for use in home setting.
{7}------------------------------------------------
-
- The subject device EarlyVue VS30 has the same fundamental technological characteristics as the legally marketed predicate device SureSigns VS4. Through the pre-defined verification and validation, the performance data demonstrates the subject device Early Vue VS30 is as safe and effective as the predicate device for all the measurements.
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarlyVue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
|---|---|---|---|
| Model/modelnumber/Referencenumbers | SureSigns VS4/863283 | EarlyVue VS30/863380 | New Model is assigned to VS30 |
{8}------------------------------------------------
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarlyVue VS30 Vital Signs MonitorSW A.00.00 | Comparison | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Indications foruse | The SureSigns VS4 vital signs monitoris for use by health care professionalswhenever there is a need formonitoring the physiologicalparameters of patients. The SureSignsVS4 is for monitoring, recording andalarming of multiple physiologicalparameters in healthcare environmentsfor patient types listed below.Additionally, the monitor may be usedin transport situations within ahealthcare facility.Parameter Adult Pediatric Neonatal NBP ✓ ✓ ✓ SpO2 ✓ ✓ ✓ Temperature ✓ ✓ ✓ CO2 ✓ ✓ ✓ RRa ✓ ✓ SpHb ✓ ✓ Contraindications: Not for transportoutside the healthcare facility | The EarlyVue VS30 is intended to measure,display, alarm and, record physiologicalinformation of adult, pediatric and neonatalpatients in hospitals and in out-of-hospital patientcare settings in which care is administered by ahealthcare professional (such as clinics,outpatient surgery facilities, long-term carefacilities, and physician offices). It is notintended for use in mobile settings such asambulances and aircraft. Clinical users may usethe monitor during patient transport within ahealthcare facility.The intended use of measurement parameters foreach patient type is shown in table below:Parameter Adult Pediatric Neonatal Non-invasive BP (NBP) ✓ ✓ ✓ SpO2 (Masimo rainbowSET or Philips FAST) ✓ ✓ ✓ Pulse Rate (PR)derived from SpO2 orNBP ✓ ✓ ✓ Temperature(Predictive, Temporal) ✓ ✓ ✓ end tidal CO2 (etCO2) ✓ ✓ ✓ Respiration Rate:airway respiration(awRR) ✓ ✓ ✓ Respiration Rate:acoustic respiration(RRa) ✓ ✓ Non-invasive TotalHemoglobin (SpHb) ✓ ✓ Contra-indications: Not for transport outside thehealthcare facility. Not for use in home setting. | --both predicate VS4 and subject VS30are intended to measure/ monitor/alar/record patient physiological information;--VS4 identified clinical setting ashealthcare environments, VS30 clarifiesthis with details hospitals and in out-of-hospital patient care settings in whichcare is administered by a healthcareprofessional (such as clinics, outpatientsurgery facilities, long-term carefacilities, and physician offices).Technically the clinical setting is notchanged.--patient types for VS4 and VS30 areidentical--parameters are clarified in VS30comparing to VS4. VS30 will not provideTympanic temperature but VS4 has.Respiration Rate: airway RR (awRR) andPulse Rate (PR) derived from SpO2 orNBP are not new. VS4 had these but notlisted in the indications table.So the parameters are identical except theVS30 will not provide Tympanictemperature.Summary Indications for use andintended use of VS30 and VS4 are thesame. |
{9}------------------------------------------------
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarlyVue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
|---|---|---|---|
| Physicalspecifications:HeightLengthDepthWeight | Width: 26.0 cm (10.2 in)Height: 22.0 cm (8.6 in)Depth: 14.5 cm (5.7 in)Weight: 3.6 Kg (8 lbs) includingbattery, excluding temperature,excluding optional recorder, optionalconnectivity package or mountinghardware. | Width: 33 cmHeight: 25 cmDepth: 18 cmWeight: <8 kg when fully optioned and withbatteryinstalled | Similar dimensionsSimilar. The weight in VS30 isfor fully optioned and the numberfor VS4 does not include options.VS30 has bigger screen but stillsimilar to VS4 |
| Device keys | Screen type: 8.4" 5-Wire resistivetouch screen display with LEDbacklightThe following fixed key switches arelocated on the front panel with theassociated labelsAlarm silence "Silence" NBP start/stop "NBP" NBP interval "Interval" Main "Main" Print "Print | 10.1" (25.7 cm) LCD touch screen with LEDbacklightThe control keys are soft keys. | VS30 changes fixed keys to softkeys.The new user interface has beenverified and validated. The VS30passed the verification andvalidation. |
| Number ofwaveforms | Two: one for etCO2 and the other onefor SpO2 or RRa, or SpO2 with RRaoverlapped | Two: one for etCO2 and the other one for SpO2or RRa, or SpO2 with RRa overlapped | Identical |
| PhysiologicalMeasurements:varies byconfigurationor option | Temperature (has Tympanictemperature option), Non-invasiveBlood Pressure, SpO2, Pulse Rate,SpHb, RRa, CO2 | Temperature (no Tympanic temperature option),Non-invasive Blood Pressure, SpO2, Pulse Rate,SpHb, RRa, CO2 | Same except that the subjectVS30 does not have Tympanictemperature but the predicate VS4has. |
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarly Vue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
| NBPSpecifications | Philips picoNBP – K051366Technique: Oscillometric using stepwise deflation pressureIn Adult Mode, the NBP measurement range is:Systolic 30-270mmHg Diastolic 10-240mmHg Mean 20-250mmHg In Pediatric Mode, the NBP measurement range is:Systolic 30-180mmHg Diastolic 10-150mmHg Mean 20-160mmHg In Neonate Mode, the NIBP measurement has the following ranges for its values:Systolic 30-130mmHg Diastolic 10-100mmHg Mean 20-120mmHg accuracy over the listed ranges:Maximum Standard Deviation: 8 mmHgMaximum Mean Error: +/- 5 mmHgA Pulse Rate is derived from the NBP measurement with accuracy:40 – 100 bpm: +/- 5 bpm | Philips picoNBP – K051366Technique: Oscillometric using stepwise deflation pressureIn Adult Mode, the NBP measurement range is:Systolic 30-270mmHg Diastolic 10-240mmHg Mean 20-250mmHg In Pediatric Mode, the NBP measurement range is:Systolic 30-180mmHg Diastolic 10-150mmHg Mean 20-160mmHg In Neonate Mode, the NIBP measurement has the following ranges for its values:Systolic 30-130mmHg Diastolic 10-100mmHg Mean 20-120mmHg accuracy over the listed ranges:Maximum Standard Deviation: 8 mmHgMaximum Mean Error: +/- 5 mmHgA Pulse Rate is derived from the NBP measurement with accuracy:40 – 100 bpm: +/- 5 bpm | Identical |
| Predicate Device | Subject Device | Comparison | |
| Comparative | SureSigns VS4 Vital Signs Monitor | EarlyVue VS30 Vital Signs Monitor | |
| Characteristic | SW A.07 | SW A.00.00 | |
| Latest 510(k): K163649 | |||
| 101 - 200 bpm: +/- 5% of reading | 101 - 200 bpm: +/- 5% of reading | ||
| 201 - 300 bpm: +/- 10% of reading | 201 - 300 bpm: +/- 10% of reading | ||
| (average over NIBP measurement | (average over NIBP measurement cycle)Initial | ||
| cycle) | cuff inflation | ||
| Initial cuff inflation | - Adult: 160 mmHg | ||
| - Adult: 160 mmHg | - Pediatric: 140 mmHg | ||
| - Pediatric: 140 mmHg | - Neonatal: 100 mmHg | ||
| - Neonatal: 100 mmHg | Subsequent cuff inflation: 30 mmHg above the | ||
| Subsequent cuff inflation: 30 mmHg | last measured Systolic value | ||
| above the last measured Systolic value | -one programmable interval with up to five | ||
| -one programmable interval with up to | periods. Each period can be composed of a | ||
| five periods. Each period can be | duration and interval. The options for each are: | ||
| composed of a duration and interval. | Duration: Off, 2, 5, 15, 30, 60, 90, 120, 150, 180, | ||
| The options for each are: | 210, 240, 270, 300 Minutes | ||
| Duration: Off, 2, 5, 15, 30, 60, 90, 120, | Interval: Off, 1, 2, 3, 5, 10, 15, 30, 60, 90, 120 | ||
| 150, 180, 210, 240, 270, 300 Minutes | Minutes | ||
| Interval: Off, 1, 2, 3, 5, 10, 15, 30, 60, | |||
| 90, 120 Minutes | |||
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarly Vue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
| SpO2: PhilipsSpO2 | Measurement range- SpO2: 0 - 100%- Pulse rate: 30 - 300 bpm | Measurement range- SpO2: 0 - 100%- Pulse rate: 30 - 300 bpm | Identical |
| Pulse rate accuracy: Greater of + 2%or + 1 bmpResolution of SpO2 measurement: 1% | Pulse rate accuracy: Greater of + 2% or + 1 bmpResolution of SpO2 measurement: 1% | ||
| Wavelength range: 500 - 1000 nm forall specified sensors | Wavelength range: 500 - 1000 nm for allspecified sensors | ||
| Maximum optical output power: ≤15mW for all specified sensors | Maximum optical output power: ≤15mW for allspecified sensors | ||
| Variable pitch tone for SpO2 | Variable pitch tone for SpO2 | ||
| Pleth wave displayed | Pleth wave displayed | ||
| configurable alarm delay times:High/Low SpO2 Alarm delay time,seconds: 0-30Increments of change: 1Desat Alarm: delay time, seconds: 0-30Increments of change: 1Default setting is : 0 seconds | configurable alarm delay times:High/ Low SpO2 Alarm delay time, seconds: 0-30Increments of change: 1Desat Alarm: delay time, seconds: 0-30Increments of change: 1Default setting is : 0 seconds | ||
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarlyVue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
| SpO2 Sensors | SpO2 accuracy: With SpO2 valuesbetween 70 - 100% range oxygenationPhilips reusable sensors $\pm$ 2%: M1191B, M1191BL, M1191A, M1191AL, M1192A $\pm$ 3%: M1191T, M1192T, M1193T with adult use, M1193A, M1194A, M1195A, M1196A, M1196T, M1196S $\pm$ 4% - M1193T with Neonatal use Philips disposable sensors $\pm$ 2% - M1132A, M1133A, M1134A $\pm$ 3% - M1131A | SpO2 accuracy: With SpO2 values between 70 –100% range oxygenationPhilips reusable sensors $\pm$ 2%: M1191B, M1191BL, M1191A, M1191AL, M1192A $\pm$ 3%: M1191T, M1192T, M1193T with adult use, M1193A, M1194A, M1195A, M1196A, M1196T, M1196S $\pm$ 4% - M1193T with Neonatal use Philips disposable sensors $\pm$ 2% - M1132A, M1133A, M1134A $\pm$ 3% - M1131A | Identical |
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarlyVue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
|---|---|---|---|
| PredicativeTempMeasurementrange:(optional) | Measurement Range: 26.7 – 43.3°C(80 – 110°F)Accuracy in monitoring mode: $\pm$ 0.1 °C( $\pm$ 0.2°F)Measurement Time Oral: 4-6 seconds Adult axillary: 12-15 sec Pediatric axillary: 10-13 sec.(17 year and younger) Rectal: 10-13 seconds Resolution: 0.1 °CMeasurement sites: Oral, rectal andaxillary | Measurement Range: 26.7 – 43.3°C (80 – 110°F)Accuracy in monitoring mode: $\pm$ 0.1 °C ( $\pm$ 0.2°F)Measurement Time Oral: 4-6 seconds Adult axillary: 12-15 sec Pediatric axillary: 10-13 sec. (17 yearand younger) Rectal: 10-13 seconds Resolution: 0.1 °C ( $\pm$ 0.1 °F)Measurement sites: Oral, rectal and axillary | Identical |
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarlyVue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
| Tympanic Tempmeasurement(optional) | The ear measurement algorithm hasthese correction factors for these modebuilt in depending on setting:Ear = NoneCore = Ear mode + 1.04 °COral = Ear mode + 0.60 °CRectal = Ear mode + 1.16 °CThe overall range of the temperaturemeasurement is 33.0 °C to 42.0 °C(91.4 °F to 107.6 °F)Accuracy at overall temperature rangeis + 0.2 °C (+ 0.4 °F)The displayed resolution of thetympanic temperature measurement is0.1 °C or 0.1 °FApproximate measurement time isequal to or better than 2 seconds.Pulse timer beeps once at each of theseintervals: 15, 30, 45 and 60 seconds. | The VS30 will not provide this option | This option is not provided by thesubject VS30 |
| Masimo SpO2,SpHb, RRameasurement | yes | yes | Identical |
| Range ofMasimo SpO2measurement | 0-100% | 0-100% | Identical |
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarly Vue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
| Resolution ofSpO2measurement | 1% | 1% | Identical |
| Accuracy ofMasimo SpO2 | No motion:60-80 +/- 3%, for adults/pediatrics/infants 70-100 +/-2%, for adults/pediatrics/infants, 70-100 +/-3%, for neonates Motion: 70-100 +/-3%, for adults/pediatrics/infants/ neonates Low perfusion: 70-100 +/-2%, for adults/pediatrics/infants/ neonates | No motion:60-80 +/- 3%, for adults/pediatrics/infants 70-100 +/-2%, for adults/pediatrics/infants, 70-100 +/-3%, for neonates Motion: 70-100 +/-3%, for adults/pediatrics/infants/ neonates Low perfusion: 70-100 +/-2%, for adults/pediatrics/infants/ neonates | Identical |
| Pleth waveform | A pleth waveform is derived from measurement | A pleth waveform is derived from measurement | Identical |
| Pulse ratederived fromthe MasimoSpO2measurement | Yes | Yes | Identical |
| Pulse rate range | 25-240 bpm | 25-240 bpm | Identical |
| Pulse rateresolution | 1 bpm | 1 bpm | Identical |
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarlyVue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
| Pulse rateaccuracy | +/-5 bpm motion or low perfusionworst case | +/-5 bpm motion or low perfusion worst case | Identical |
| Range of SpHbmeasurement | 0-25 g/dL | 0-25 g/dL | Identical |
| SpHb accuracy | 8-17 g/dL (arterial or venous)adults/pediatrics: +/- 1g/dL | 8-17 g/dL (arterial or venous) adults/pediatrics:+/- 1g/dL | Identical |
| SpHbresolution | 0.1 g/dL | 0.1 g/dL | Identical |
| Range ofrespiratory rate(RRa) | 0-70 breaths per minute (bpm) | 0-70 breaths per minute (bpm) | Identical |
| RRa accuracy | In the range of 4-70 bpm: +/-1 bpm,adults (>30kg) | In the range of 4-70 bpm: +/-1 bpm, adults(>30kg) | Identical |
| Technologyused for RRameasurementand calculation | Rainbow SET technology (K110028) | Rainbow SET technology (K110028) | IdenticalPlease refer to section 7.2 forrespiration rate comparisonbetween the VS30 and MasimoRainbow SET RRa |
| RRa resolution | 1 breath per minute (bpm) | 1 breath per minute (bpm) | Identical |
| Firmware | Rainbow SET Technology, MXboard/circuitry from Masimo | Rainbow SET Technology, MX board/circuitryfrom Masimo | Identical |
| Wavelengthrange | 653 to 905 nm for all specified sensors | 500 nm to 1400 for all specified sensors | Identical |
| Maximumoptical outputpower | <=25mW for all specified sensors | <=25mW for all specified sensors | Identical |
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarly Vue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
| TemporalTemperaturemeasurement | yes | yes | Identical |
| Can displayboth °C and oF | yes | yes | Identical |
| Range oftemperaturemeasurement | 16 °C - 43 °C | 16 °C - 43 °C | Identical |
| Clinicalaccuracy | +/- 0.2 oF or 0.1 °C per ASTM E1112 | +/- 0.2 oF or 0.1 °C per ASTM E1112 | Identical |
| Arterial heatbalance rangefor bodytemperature | 34.5 °C -43 °CNormal range is defined as 35.9℃ to37.5℃, with mean of 37℃. | 34.5 °C -43 °CNormal range is defined as 35.9℃ to 37.5℃,with mean of 37°C. | Identical |
| Resolution | Both on probe and on monitor 0.1 °Cor 0.1 oF | Both on probe and on monitor 0.1 ℃ or 0.1 oF | Identical |
| Response time | <=1 second | <=1 second | Identical |
| CO2Measurement | yes | yes | Identical |
| Measurementrange | 0 - 150 mmHg | Same | Identical |
| Sampling rate | Waveform sampling, 20 samples persecond | Same | Identical |
| Flow rate | 50 ml/min, + 15 ml/min, - 7.5 ml/min | Same | Identical |
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarly Vue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
| Warm-up timefor CO2measurement | 30 seconds (typical) | 30 seconds (typical) | Identical |
| Calibrationinterval | Initial Calibration after 1,200 hours orafter one year whichever comes first,then once per year or every 4000hours, whichever comes first | Initial Calibration after 1,200 hours or after oneyear whichever comes first, then once per year orevery 4000 hours, whichever comes first | Identical |
| Accuracy ofetCO2measurement | 0 – 38 mmHg: ±2 mmHg39 - 150 mmHg: ±(5% of reading +0.08% for every 1 mmHg above 38mmHg) | 0 – 38 mmHg: ±2 mmHg39 - 150 mmHg: ±(5% of reading + 0.08% forevery 1 mmHg above 38 mmHg) | Identical |
| Range ofRespiration rate | Respiration rate can be displayed witha range of 0-150 respirations/ minute | Respiration rate can be displayed with a range of0-150 respirations/ minute | Identical |
| accuracy ofRespirationRate | ±1 rpm in the range of 0- 70 rpm±2 rpm in the range of 71 - 120 rpm±3 rpm in the range of 121 - 150 rpm | ±1 rpm in the range of 0- 70 rpm±2 rpm in the range of 71 - 120 rpm±3 rpm in the range of 121 - 150 rpm | Identical |
| Technologyused forrespiration ratecalculation | Oridion MicroMediCO2 module(K123690) | Oridion MicroMediCO2 module (K123690) | IdenticalPlease refer to section 7.1 forrespiration rate comparisonbetween the VS30 and theOridion MicroMediCO2. |
| CO2 responsetime | CO2 Response Time (Delay time +Rise time + System time) isapproximately 3.9 seconds.Long lines add an additional 3 secondsto the overall response time. | CO2 Response Time (Delay time + Rise time +System time) is approximately 3.9 seconds.Long lines add an additional 3 seconds to theoverall response time. | Identical |
| Battery poweroperation | 3 hour | 6 hour | Battery can operate with longertime |
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07Latest 510(k): K163649 | Subject DeviceEarlyVue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
| Bar Codescanner | Yes – the bar code reader allows forentry of patient identificationinformation and user ID | Yes – the bar code reader allows for entry ofpatient identification information and user ID | Identical |
| RFID scanner | No | Yes-- allows for entry of patient identificationinformation and user ID | This application is a non-medicalapplication.The subject VS30 has an optionof RFID scanner for entry ofpatient identification informationand user ID. This is fully verifiedand verification passed all tests. |
| Quick Capture | Yes | Yes | Identical |
| Quick Check | Yes | Yes | Identical |
| Quick Alerts | Yes | Yes | Identical |
| Connectionwith PhilipsIGS software | Yes | Yes | Identical |
| Get EWS fromIGS | Yes | Yes | Identical |
| On-deviceEWS | Yes | Yes | Identical |
| Deviceoperatingtemperature | 10°C to 40°C monitor16°C to 33°C monitor with tympanic10°C to 40°C for WelchAllyn0°C to 50°C for Masimo16°C to 40°C for Exergen | 10°C to 40°C monitor10°C to 40°C for WelchAllyn16°C to 40°C for Exergen | Same |
| ComparativeCharacteristic | Predicate DeviceSureSigns VS4 Vital Signs MonitorSW A.07 | Subject DeviceEarlyVue VS30 Vital Signs MonitorSW A.00.00 | Comparison |
| Storagetemperature | Latest 510(k): K163649-20°C to 50°C for the monitor-20°C to 40°C for the monitor plusaccessories-20°C to 50°C for WelchAllyn-40°C to 70°C for Masimo-20°C to 50°C for Exergen | 20°C to 50°C for the monitor-20°C to 40°C for the monitor plus accessories | Same |
| Operation andstorageHumidity | Up to 90% Relative Humidity non-condensingUp to 80% Relative Humidity non-condensing with recorder and paper15-90% for Welch Allyn predictivetemperature (non-condensing)10-90% Masimo SpO2 | Composite test range 10% to 90% RH (non-condensing) for monitorComposite test range 10% to 80% RH formonitor with print recorder and paper | Same |
| Monitor tomonitor sharing | No | Yes---view saved patient records remotely on up to15 monitors---view the last saved record for up to 100patients---view patient data trends in graphical or tabularformat, up to 35 records. | This is a non-medical application.This is a new feature . The view isonly to view records. The user cannot use one monitor tomonitor/measure patient who isconnected to other monitors. Thisis a hospital work flowimprovement feature but not amedical application.This feature is fully verified andthe verification passed the tests. |
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
{18}------------------------------------------------
{19}------------------------------------------------
{20}------------------------------------------------
{21}------------------------------------------------
7.1 Respiration rate comparison between Early VS30 and Oridion MicroMediCO2
| Parameter | Oridion MicroMediCO2 | EarlyVue VS30 | Comparison |
|---|---|---|---|
| ----------- | ---------------------- | --------------- | ------------ |
{22}------------------------------------------------
| K123690 | K190624 | ||
|---|---|---|---|
| CO2Respiration raterange | 0-150 bpm | 0 breaths/min to 150 breaths/min | Identical |
| CO2Respiration rateaccuracy | 0-70 bpm: $\u00b1$ 1 bpm71-120 bpm: $\u00b1$ 2 bpm121-150 bpm: $\u00b1$ 3 bpm | $\u00b1$ 1 rpm in the range of 0 rpm-70 rpm$\u00b1$ 2 rpm in the range of 71 rpm−120 rpm$\u00b1$ 3 rpm in the range of 121 rpm-150 rpm | Identical |
7.2 RRa respiration rate comparison between the VS30 and the Masimo RRa
| Parameter | Masimo RRa respiration rateK110028 | Early Vue VS30K190624 | Comparison |
|---|---|---|---|
| RRaMeasurementRange | 4 – 70 breaths per minute | 4 – 70 breaths per minute | Identical |
| RRa Accuracy | ± 1 breath per minute in the range 4 –70 breaths per minute,Adults (> 30 kg) and Adults/Pediatrics(> 10 kg) | ± 1 breath per minute in the range 4 – 70 breathsper minute,Adults (> 30 kg) and Adults/Pediatrics (> 10 kg) | Identical |
{23}------------------------------------------------
-
- Verification and testing activities establish the performance, functionality, and reliability characteristics of the subject device VS30. Testing involved software, hardware, environmental and reliability, safety and performance.
The VS30 is tested against applicable FDA consensus IEC/ISO standards. The following standards were tested and the EarlyVue VS30 passed the tests of all these standards:
- Verification and testing activities establish the performance, functionality, and reliability characteristics of the subject device VS30. Testing involved software, hardware, environmental and reliability, safety and performance.
| Standard and Year | Description | FDA recognition number |
|---|---|---|
| IEC 60601-1 ed 3.1:2012Issue Date: 2012-08-30 | Medical electrical equipment. Part 1: General requirementsfor basic safety and essential performance | 19-4 |
| IEC 60601-1-2 ed 4.0:2014Issue Date: 2014-02-28 | Medical electrical equipment. Part 1-2: General requirementsfor basic safety and essential performance. Collateralstandard: Electromagnetic compatibility. Requirements andtests | 19-8 |
| IEC 60601-1-6 ed 3.1:2013Issue Date: 2013-10-31 | Medical electrical equipment. Part 1-6: General requirementsfor basic safety and essential performance. Collateralstandard: Usability | 5-89 |
| IEC 62366-1 ed 1.0:2015Issue Date: 2015-02-28 | Medical devices. Part 1: Application of usability engineeringto medical devices | 5-114 |
| IEC 62304 ed 1.1:2015Issue Date: 2015-06-29 | Medical device software. Software life cycle processes | 13-79 |
| IEC 60601-1-8 ed 2.1:2012Issue Date: 2012-11-30 | Medical electrical equipment. Part 1-8: General requirementsfor basic safety and essential performance. CollateralStandard: General requirements, tests and guidance for alarmsystems in medical electrical equipment and medicalelectrical systems | 5-76 |
| IEC 80601-2-30 ed 1.1:2013Issue Date: 2013-07-30 | Medical electrical equipment. Part 2-30: Particularrequirements for the basic safety and essential performance ofautomated non-invasive sphygmomanometers | 3-152 |
| IEC 60601-2-49 ed 2.0:2011Issue Date: 2011-02-28 | Medical electrical equipment. Part 2-49: Particularrequirements for the basic safety and essential performance ofmultifunction patient monitoring equipment | Notrecognized |
| ISO 80601-2-55:2011Issue Date: 2011-12-15 | Medical electrical equipment, part 2-55 — Particularrequirements for the basic safety and essential performance ofrespiratory gas monitor | 1-96 |
| ISO 80601-2-56:2017Issue Date: 2017-03 | Medical electrical equipment. Part 2-56: Particularrequirements for basic safety and essential performance ofclinical thermometers for body temperature measurement | 6-403 |
| ISO 80601-2-61:2011Issue Date: 2011-03-31 | Medical electrical equipment. Part 2-61: Particularrequirements for basic safety and essential performance ofpulse oximeter equipment | 1-85 |
| ISO 15223-1:2017Issue Date: 2016-11 | Medical devices - Symbols to be used with medical devicelabels, labelling and information to be supplied - Part 1:General requirements | 5-117 |
{24}------------------------------------------------
In summary, the subject device VS30 has the same intended use as the predicate device VS4. The VS30 does not add any new measurement comparing to the predicate device VS4 and the measurement technologies used are identical.
Any new or different features have been verified and validated, such as the monitor to monitor sharing, the new user interface including the function keys. The result supports a determination of substantial equivalence of VS30 to the predict device VS4.
The VS30 has been tested to all applicable IEC/ISO safety, EMC and performance standards and passed all the tests.
All the VS30 measurement functions were verified through verification test/integration test.
The VS30 with the wireless module was tested to and passed the ANSI C63.27-2017 standard and comply with the FDA guidance "Radio Frequency Wireless Technology in Medical Devices Guidance".
The VS30 software has been verified according to the design control process which was designed according to the FDA quality management system regulation.
All the VS30 electric interfaces has been tested and verified and comply with the FDA guidance: Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices.
In conclusion, the subject VS30 successfully passed all verifications, testing and validations. The results demonstrate that the Philips Early Vue VS30 Vital Signs Monitor meets all safety, effectiveness and performance requirements. The subject Early Vue VS30 is as safe, as effective and performs as well as the predicate device SureSigns VS4. The EarlyVue VS30 is substantially equivalent to the predicate device VS4.
§ 870.1100 Blood pressure alarm.
(a)
Identification. A blood pressure alarm is a device that accepts the signal from a blood pressure transducer amplifier, processes the signal, and emits an alarm when the blood pressure falls outside a pre-set upper or lower limit.(b)
Classification. Class II (performance standards).