K Number
K110028
Manufacturer
Date Cleared
2011-03-17

(71 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated to provide the continuous noninvasive monitoring data obtained from the Masimo Radical 7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (SpO₂) and pulse rate to multiparameter devices for the display of those devices.

Device Description

The Radical 7 Pulse CO-Oximeter and accessories (Radical 7) include the Masimo Rainbow SET technology. The Radical 7 Pulse CO-Oximeter provides nonitoring of arterial oxygen saturation (SpO3), pulse rate (PR), carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (g/dl SpHb), and/or respiration rate (RRa). Other information displayed by the Radical 7 Pulse CO-Oximeter includes: Low Signal IQ (Low SIQ), Perfusion Index (PI), Pleth Variability Index (PVI), Total Arterial Oxygen Content (SpOC), Hematocrit (SpHct), Signal Identification Quality (SIQa), Respiration Indicator (R), alarm status, alarm silence, battery life, sensor status, trends, and pleth waveform. The Radical 7 Pulse CO-Oximeter also has a touchscreen and output interfaces, which include: SatShare connection to multi-parameter monitors, Nurse Call analog output, and RS-232 serial output, and wireless radio.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the Masimo Radical 7 Pulse CO-Oximeter and Accessories, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Stated Accuracy)Reported Device Performance (Achieved Accuracy)Patient PopulationCondition
SpO2
60-80 ± 3%≤ 3% (RMS)Adults/Pediatrics/InfantsNo Motion
70-100 ± 2%≤ 2% (RMS)Adults/Pediatrics/InfantsNo Motion
70-100 ± 3%≤ 3% (RMS)Adults/Pediatrics/Infants/NeonatesMotion
70-100 ± 2%≤ 2% (RMS)Adults/Pediatrics/Infants/NeonatesLow Perfusion
PR
25-240 ± 3 bpm≤ 3 bpm (RMS)Adults/Pediatrics/Infants/NeonatesNo Motion/Low Perfusion
25-240 ± 5 bpm≤ 5 bpm (RMS)Adults/Pediatrics/Infants/NeonatesMotion
SpCO1-40 ± 3%≤ 3% (RMS)Adults/Pediatrics/Infants
SpMet1-15 ± 1%≤ 1% (RMS)Adults/Pediatrics/Infants/Neonates
SpHb8-17 +1 g/dl≤ 1 g/dl (RMS)Adults/Pediatrics
RRa4-70 + 1 breath per minute≤ 1 breath per minute (RMS)Adults

Note: The document states "The studies resulted in accuracies (rms) of equal to or less than the respective accuracies as stated above in the Radical 7 specifications." This implies that the reported device performance met or exceeded the "Acceptance Criteria" values listed in the specification table.

2. Sample Size Used for the Test Set and Data Provenance

  • SpO2, SpCO, SpMet (Adult/Pediatric/Infant): Healthy adult volunteers. The exact number of volunteers is not specified, but the range of SpO2 (60-100%), SpCO (0-40%), and SpMet (0-15%) was tested.
  • SpHb (Adult/Pediatric): Healthy adult volunteers and surgical patients with light to dark skin pigmentation. The exact number of patients is not specified, but the range of SpHb (8-17 g/dl) was tested.
  • SpO2 and SpMet (Neonatal): 16 neonatal NICU patients.
    • Age range: 7-135 days old
    • Weight range: 0.5-4.25 kg
    • Number of data samples: 79 samples collected over a range of 70-100% SaO2 and 0.5-2.5% MetHb.
  • RRa: Adults. The exact number of subjects is not specified.

Data Provenance: The studies appear to be prospective clinical studies conducted with human subjects. The country of origin for the data is not explicitly stated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of device (pulse CO-Oximeter) uses a laboratory CO-Oximeter as the ground truth reference, not human experts. Therefore, information regarding "number of experts" and "qualifications of those experts" is not applicable in this context.

4. Adjudication Method for the Test Set

Not applicable. The ground truth is established by a laboratory CO-Oximeter, which provides objective measurements, removing the need for human adjudication of results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No, an MRMC comparative effectiveness study was not done. This device is a measurement device and its performance is evaluated against a golden standard reference (laboratory CO-Oximeter), not against human interpretation or a human-AI collaborative workflow.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the studies described are standalone performance evaluations. The device's measurements (SpO2, PR, SpCO, SpMet, SpHb, RRa) are directly compared against the ground truth established by a laboratory CO-Oximeter or other appropriate reference standards, without human interpretation in the loop impacting the accuracy determination.

7. The Type of Ground Truth Used

The primary ground truth used is measurements from a laboratory CO-Oximeter.

  • For SpO2, SpCO, SpMet, and SpHb, the device was tested "against a laboratory CO-Oximeter."
  • For RRa, the specific ground truth method is not detailed beyond "based on vibratory signals from respiratory sounds," but it would involve a highly accurate reference for respiratory rate.

8. The Sample Size for the Training Set

The document does not provide details on a separate "training set" or its sample size. The description focuses on clinical studies performed for performance validation, which typically serve as the test set for regulatory submission. It's possible that internal development and calibration involved various datasets, but these are not explicitly detailed as a separate "training set" in this 510(k) summary.

9. How the Ground Truth for the Training Set Was Established

As no explicit "training set" details are provided in this summary, the method for establishing its ground truth is also not mentioned. However, it can be inferred that any internal development or calibration data would likely have been established using similar reference standards (e.g., laboratory CO-Oximeters) as used for the validation studies.

{0}------------------------------------------------

510(k) Summarv

MAR 1 7 2011

Submitted by:Masimo Corporation40 Parker, Irvine, CA 92618Phone: (949) 297-7000; FAX: (949) 297-7592
ContactShelly Harris, Manager of Regulatory Affairs
Date Summary PreparedJanuary 3, 2010.
Establishment Registration No.2031172
Trade NameMasimo Radical 7 Pulse CO-Oximeter and Accessories
Common NamePulse Oximeter and Sensor
Regulation Number/Name/ Class21 CFR 870.2700/ Oximeter/ Class II
Product CodeDQA, BZQ, DPZ, JKS
Substantially Equivalent DeviceMasimo Rainbow SET® Radical 7R Pulse CO-Oximeters and Accessories510(k) Number - K100428

Description of the Device

The Radical 7 Pulse CO-Oximeter and accessories (Radical 7) include the Masimo Rainbow SET technology. The Radical 7 Pulse CO-Oximeter provides nonitoring of arterial oxygen saturation (SpO3), pulse rate (PR), carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (g/dl SpHb), and/or respiration rate (RRa). Other information displayed by the Radical 7 Pulse CO-Oximeter includes: Low Signal IQ (Low SIQ), Perfusion Index (PI), Pleth Variability Index (PVI), Total Arterial Oxygen Content (SpOC), Hematocrit (SpHct), Signal Identification Quality (SIQa), Respiration Indicator (R), alarm status, alarm silence, battery life, sensor status, trends, and pleth waveform. The Radical 7 Pulse CO-Oximeter also has a touchscreen and output interfaces, which include: SatShare connection to multi-parameter monitors, Nurse Call analog output, and RS-232 serial output, and wireless radio.

Intended Use/Indications for Use

The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masino Radical 7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Radical 7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate to multi-parameter devices for the display of those devices.

Principles of Operation

SpO2 and Pulse Rate: Pulse oximetry is governed by the principles that blood contents differ in their absorption of red and infrared light and blood pulsation.

SpCO, SpMet, and SpHb: The Radical 7 includes the Masimo Rainbow SET technology board, which calculates functional oxygen saturation (SpO₂), fractional concentration of carboxyhemoglobin (SpCO), fractional concentration of methemoglobin (SpMet), total hemoglobin concentration (SpHb) and pulse rate.

Respiratory or Respiration Rate (RRa) General Description: The Masimo Rainbow SET technology also provides respiratory or respiration rate measurements, based on vibratory signals from respiratory sounds.

{1}------------------------------------------------

510(k) Summary

Comparison to Predicate Device

The Radical 7 in this filing is the same as the predicate in the K100428 filing, with the addition of wireless radio for remote monitoring, a Bluetooth radio module, and update to the user interface controls. Below are the specifications for the Radical 7.

FEATURESSPECIFICATIONS
Display RangeSpO2: 0-100%; PR: 25-240 bpm; SpCO: 0-99%; SpMet: 0-99.9%; SpHb: 0-25 g/dL;RRa: 0-70 breaths per minute; SpOC: 0-35 ml/dl, SpHct: 0-75%; PI: 0.02-20%; PVI: 0-100%
Accuracy: SpO2No Motion (adults/pediatrics/infants): 60-80 ± 3%; 70-100 ± 2%, ± 3%(neonates)Motion (adults/pediatrics/infants/neonates): 70-100 ± 3%Low Perfusion (adults/pediatrics/infants/neonates): 70-100 ± 2%
Accuracy: PRNo Motion/ Low Perfusion(adults/pediatrics/ infants/neonates): 25-240 ± 3 bpmMotion (adults/pediatrics/infants/neonates): 25-240 ± 5 bpm
Accuracy: SpCO1-40 + 3%, adults/pediatrics/infants
Accuracy: SpMet1-15 ± 1%, adults/pediatrics/infants/neonates
Accuracy: SpHb8-17 +1 g/dl (arterial or venous), adults/pediatrics
Accuracy: RRa4-70 + 1 breath per minute, adults
ResolutionSpO2: 1%; PR: 1 bpm; SpCO: 1%; SpMet: 0.1%; SpHb: 0.1 g/dl; RRa: 1 breath per minute
MeasurementsLow Signal IQ, PI, SpOC, SpHct, PVI, SIQa, RI
PowerAC Input Range: 100-240 VAC, 47-63 Hz; Rechargeable batteries
EnvironmentalOp temp: 32 to 122°F; Storage temp: -40 to 158°F; RH: 10-95% non-condensing
SpO2 AveragingAveraging: 2, 4, 6, 8, 10, 12 and 16 seconds; FastSat
Mode/ SensitivitySensitivity: APOD, Normal, Maximum
AlarmVolume (pulse/tone), out of limit (high/low), sensor condition, system failure, low battery
Display/IndicatorsSpO2, PR, SPCO, SpMet, SpHb, SpHbv, RRa, SpOC, SpHct, PI, PVI, Pleth waveform, SIQSIQa, RI, sensitivity, sensor status/time/messages, alarm status, battery status
Output InterfaceTrends, Satshare to Multiparameter monitors (SpO2 only), RS-232 (PC/printer), Vuelink,Spacelabs Flexport, RadNet, PSN, wireless radio
ComplianceEMC: EN 60601-1-2, Class B; Electrical Safety: IEC 60601-1/UL 60601-1, Class 1 (ACPower), internally powered, patient cable-Type BF applied part, Satshare cable-type CFapplied part; 802.11 a/b/g
Operation ModeContinuous

Clinical Summary

Clinical Studies: SpO2, SpCO and SpMet accuracy was determined by testing healthy adult volunteers in the range of 60-100% SpO2, 0-40% SpCO, and 0-15% SpMet against a laboratory CO-Oximeter. SpHb accuracy has been validated on healthy adult volunteers and on surgical patients with light to dark skin pigmentation in the range of 8-17 g/dl SpHb against a laboratory CO-oximeter. The variation in these studies equals one standard deviation which encompasses 68% of the population. SpO₂ and SpMet accuracy was determined on 16 neonatal NICU patients ranging in age from 7-135 days old and weighing between 0.5-4.25 kg. Seventy-nine (79) data samples were collected over a range of 70-100% SaO2 and 0.5-2.5% MetHb with a resultant accuracy of 2 9% SpO2 and 0.9% SpMet.

Clinical Results: No device-related adverse events. The clinical studies were performed in accordance with ISO 9919:2005. The studies resulted in accuracies (rms) of equal to or less than the respective accuracies as stated above in the Radical 7 specifications.

Non-Clinical Summary

The Radical 7 complies with the voluntary standards as detailed in this submission. Laboratory testing for rne radical 7 compleo with the voluntary overify that the Radical 7 met all design specifications and was circularitially equivalent to the predicate device. This device was not tested for biocompatibility because like the predicate device, it does not contain any patient contacting materials.

{2}------------------------------------------------

510(k) Summary

Conclusions

·

·

. .

!

The information in this 510(k) submission demonstrates that the Masimo Radical 7 Pulse CO-Oximeter is substantially equivalent to the predicate device, with respect to safety, effectiveness, and performance.

.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the left side of the logo.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002

Ms. Shelly Harris Manager of Regulatory Affairs Masimo Corporation 40 Parker Irvine, California 92618

MAR 1 7 2011

Re: K110028

Trade/Device Name: Radical 7 Pulse CO-Oximeter and Accessories Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA, BZQ, DPZ, JKS Dated: February 14, 2011 Received: February 15, 2011

Dear Ms. Harris:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Ms. Harris

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

Radical 7 Pulse CO-Oximeter and Accessories Device Name:

Indications For Use:

The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated to provide the continuous noninvasive monitoring data obtained from the Masimo Radical 7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (SpQ₂) and pulse rate to multiparameter devices for the display of those devices.

Prescription Use (Per 21 CFR 801.109 Subpart D) AND/OR

Over-The-Counter Use (Per 21 CFR 801.109 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesional Devices Division of Anostrios of Antson of Antection Controllers

510(k) Number: K1100281

Special 510(k), Radical 7 01/03/2011, Page 4-1

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).