K Number
K151812
Date Cleared
2016-01-26

(208 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Efficia CM10, CM12, CM100, CM120 and CM150 patient monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The Efficia CM10, CM12, CM100, CM120 and CM150 patient monitors are intended for monitoring, analysis, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.

Parameter (patient types): ECG (Adult, Pediatric, Neonatal); Respiration - Impedance (Adult, Pediatric, Neonatal); Respiration - RRa (Adult); Respiration - awRR (Adult, Pediatric, Neonatal); NBP (Adult, Pediatric, Neonatal); SpO2 (Adult, Pediatric, Neonatal); SpHb (Adult, Pediatric); SpCO (Adult, Pediatric); Temperature (Adult, Pediatric, Neonatal); CO2 (Adult, Pediatric, Neonatal); IBP (Adult, Pediatric, Neonatal); Cardiac Output (Adult, Pediatric); Arrhythmia (Cardiotach) (Adult, Pediatric, Neonatal); Arrhythmia (Basic) (Adult, Pediatric); Arrhythmia (Advanced) (Adult, Pediatric); ST segment Analysis (Adult);

Contraindications: Not for transport outside the healthcare facility

Device Description

The subject Philips Efficia CM10, CM120 and CM150 patient monitors are a family of physiological vital signs monitors. The monitors come with a variety of display sized (10 inch, 12 inch or 15 inch) and with a variety of measurements, some optional. The monitors primarily run on AC power with a secondary battery. The exterior material is the same as the primary predicate devices.

AI/ML Overview

This document describes the regulatory submission for the Philips Efficia CM10, CM12, CM100, CM120, and CM150 patient monitors, which are physiological vital signs monitors. The submission includes information about the device's intended use, parameters measured, and a comparison to predicate devices. The listed parameters, which include ECG, Respiration, NBP, SpO2, Temperature, IBP, Cardiac Output, Arrhythmia Analysis, and ST-Segment monitoring, also specify the patient types (adult, pediatric, neonatal) for which each parameter is indicated.

1. Table of Acceptance Criteria and Reported Device Performance
The document states that "Performance data supports that the subject devices performance to same or better level of safety and effectiveness as the predicate devices." and that the devices "use the same algorithms and have the similar performance to their predicates." Specific acceptance criteria or quantitative performance metrics (e.g., sensitivity, specificity, accuracy thresholds) are not explicitly provided in the provided text, nor are specific numerical results for the device performance against such criteria. The submission relies on the assertion that the device's performance is at least equivalent to the predicate devices.

2. Sample size used for the test set and data provenance
The provided text does not explicitly state the sample size used for any test set or the data provenance (e.g., country of origin, retrospective/prospective) for performance studies. It generally refers to "performance data" and a "device history file" without specific details.

3. Number of experts used to establish the ground truth for the test set and their qualifications
This information is not provided in the document.

4. Adjudication method for the test set
This information is not provided in the document.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, an effect size of how much human readers improve with AI vs without AI assistance
This information is not provided in the document. The device is a patient monitor, not an AI-assisted diagnostic tool for human readers in the traditional sense of an MRMC study.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
While the document states that the ECG, NBP, invasive blood pressure, arrhythmia, and ST-segment analysis algorithms are Philips proprietary or derived from established Philips algorithms, and other parameters use algorithms from other manufacturers (e.g., Masimo, Respironics), it does not explicitly detail a standalone performance study. The statement "Performance data supports that the subject devices performance to same or better level of safety and effectiveness as the predicate devices" implies some form of performance evaluation, but the methodology and whether it was standalone are not described.

7. The type of ground truth used
The type of ground truth used for performance evaluation is not specified in the provided text. It can be inferred that, for physiological parameters like ECG, NBP, SpO2, etc., ground truth would typically come from reference devices, established clinical methods, or expert interpretation of physiological waveforms.

8. The sample size for the training set
This information is not provided in the document. The algorithms are either Philips proprietary and "well established" or from other manufacturers and "unchanged," suggesting they were developed and trained prior to this specific submission, but details about the training set size are absent.

9. How the ground truth for the training set was established
This information is not provided in the document. As with the test set, the establishment of ground truth for training would likely involve reference devices, clinical standards, or expert review of physiological data, but no specifics are given.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The text is centered horizontally. The background is white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 26, 2016

Philips Medical Systems Mary Kruitwagen OA and Regulatory Affairs 3000 Minuteman Road Andover, MA 01810-1099

Re: K151812

Trade/Device Name: Philips Efficia CM10 863301, Philips Efficia CM12 863303, Philips Efficia CM100 863300. Philips Efficia CM120 863302, Philips Efficia CM150 863304 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, MLD, DSJ, DSK, DXN, DRT, DQA, DSA, FLL, CCK, BZQ Dated: December 16, 2015 Received: December 21, 2015

Dear Mary Kruitwagen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

[signature]

for Pram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for UseForm Approved: OMB No. 0910-0120Expiration Date: January 31, 2017See PRA Statement below.
510(k) Number (if known)K151812
Device NamePhilips Efficia CM10 863301, Philips Efficia CM12 863303, Philips Efficia CM100 863300, Philips Efficia CM120 863302Philips Efficia CM150 863304
Indications for Use (Describe)The Efficia CM10, CM12, CM100, CM120 and CM150 patient monitors are indicated for use by health careprofessionals whenever there is a need for monitoring the physiological parameters of patients. The Efficia CM10,CM12, CM100, CM120 and CM150 patient monitors are intended for monitoring, analysis, recording and alarmingof multiple physiological parameters in healthcare environments for patient types listed below. Additionally, themonitor may be used in transport situations within a healthcare facility.
Parameter (patient types): ECG (Adult, Pediatric, Neonatal); Respiration - Impedance (Adult, Pediatric, Neonatal);Respiration - RRa (Adult); Respiration - awRR (Adult, Pediatric, Neonatal); NBP (Adult, Pediatric, Neonatal); SpO2(Adult, Pediatric, Neonatal); SpHb (Adult, Pediatric); SpCO (Adult, Pediatric); Temperature (Adult, Pediatric, Neonatal);CO2 (Adult, Pediatric, Neonatal); IBP (Adult, Pediatric, Neonatal); Cardiac Output (Adult, Pediatric); Arrhythmia(Cardiotach) (Adult, Pediatric, Neonatal); Arrhythmia (Basic) (Adult, Pediatric); Arrhythmia (Advanced) (Adult,Pediatric); ST segment Analysis (Adult);
Contraindications: Not for transport outside the healthcare facility
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

{3}------------------------------------------------

Efficia CM10, CM12, CM100, CM120 and CM150 Patient Monitors 510(k) Summary

Preparation DateSeptember 18, 2015
Submission Type:Traditional 510(k)
Regulatory InformationClassificationPhysiological Monitor, Patient Monitor
Common/ Usual NameMulti-parameter Patient monitor
Trade NamePhilips Efficia
Proprietary NameEfficia CM10Efficia CM12Efficia CM100Efficia CM120Efficia CM150
Models (with Reference numbers)CM10 (863301)CM12 (863303)CM100 (863300)CM120 (863302)CM150 (863304)
Device SponsorPhilips Medical Systems,3000 Minuteman Rd,Andover, MA, USA, 01810-1099
Device OwnerPhilips Medical Systems,3000 Minuteman Rd,Andover, MA, USA 01810-1099
Establishment Registration number:1218950
Sponsor/Manufacturer/Owner/Operator:1217116
ContactMary KruitwagenPhilips Medical Systems,3000 Minuteman Rd,Andover, MA, USA, 01810-1099Phone: (978) 659-4932Fax: 978-659-7323mary.kruitwagen@philips.com

{4}------------------------------------------------

Peter Schipelliti, Philips Medical Systems, 3000 Minuteman Rd, Andover, MA 01810-1099 USA Email: peter.schipelliti@philips.com Phone: (978) 659-4744 Fax: 978-659-7323

Device Classification:

The product codes and individual classification monographs applicable to the subject devices are listed below:

Device PanelClassificationProCodeDescription
Cardiovascular870.1025, IIMHXMonitor, Physiological, Patient (with arrhythmia detection or alarms)
Cardiovascular870.1025, IIMLDMonitor, St Segment With Alarm
Cardiovascular870.1110, IIDSJAlarm, Blood Pressure
Cardiovascular870.1110, IIDSKComputer, Blood Pressure
Cardiovascular870.1130, IIDXNSystem, Measurement, Blood Pressure, Non-Invasive
Cardiovascular870.2300, IIDRTMonitor, Cardiac (incl. Cardiotachometer & Rate Alarm)
Cardiovascular870.2700, IIDQAOximeter
Cardiovascular870.2900, IDSACable, Transducer and Electrode, incl. Patient connector
Cardiovascular890.1175, IIIKDcable, electrode
General Hospital andPersonal Use880.2910, IIFLLThermometer, Electronic, Clinical
Anesthesiology &Respiratory Therapy868.1400, IICCKAnalyzer, Gas,
Anesthesiology &Respiratory Therapy868.2375, IIBZQMonitor, Breathing Frequency

Device Description

The subject Philips Efficia CM10, CM120 and CM150 patient monitors are a family of physiological vital signs monitors. The monitors come with a variety of display sized (10 inch, 12 inch or 15 inch) and with a variety of measurements, some optional. The monitors primarily run on AC power with a secondary battery. The exterior material is the same as the primary predicate devices.

Fundamental Scientific Technology:

The subject Efficia CM10, CM120 and CM150 patient monitors include the following parameters (some optional): ECG, Respiration, NBP, SpO, Temperature, Invasive Blood Pressure, Cardiac Output, Arrhythmia Analysis, ST-Segment monitoring, and CO2 .monitoring.

  • . The ECG algorithm is a Philips algorithm which is well established in the marketplace and is used in the primary predicate devices (K123900).
  • Respiration rate can be measured through impedance technology, from capnography or via acoustic respiration. The Impedance respiration algorithm is the same algorithm as used in the primary predicate devices (K123900).
  • Respiration from capnography, shown as awRR, is either from Microstream from the Oridion algorithm or main stream CO2.
    • The Sidestream method is the same as in the primary predicate (K123900). The main stream CO2 including O respiration is from Respironics.
    • The algorithm of the respirations from the mainstream CO₂ is from Respironics M2105A sensor (K042601 O and K050762) and is used unchanged in the subject devices.
    • The acoustic respirations (awRR) measurement is achieved via the Masimo rainbow Set using their O algorithm unchanged and previously cleared (K100428).

{5}------------------------------------------------

  • The NBP algorithm is proprietary Philips algorithm and is the same algorithm as in the primary predicate device (K123900).
  • The subject devices offer two different SpO3 algorithms. The default SpO3 algorithm is a Philips proprietary algorithm and is unchanged from the primary predicate device (K123900). Optionally, the user can select the Masimo Set or Masimo rainbow Set. This option uses Masimo's algorithm unchanged from its predicate (K100428). SpHb and SpCO are part of Masimo's algorithm and are implemented unchanged from the Masimo predicate.
  • . The temperature measurement is similar algorithm of primary predicate (K123900).
  • The invasive blood pressure algorithm is a Philips algorithm which is unchanged from that in the primary predicate device (K123900).
  • . The arrhythmia algorithm and the ST-Segment analysis algorithm used the Philips ST/AR algorithm and incorporate a subset of the STAR algorithm as cleared by K101521.

Performance data supports that the subject devices performance to same or better level of safety and effectiveness as the predicate devices. Clinical user preference and usability studies support that the customer requirements are met. The subject devices are in compliance with the following standards:

IEC 60601-1IEC 60601-2-27ISO 80601-2-55
IEC 60601-1-2IEC 80601-2-30ISO 80601-2-56
IEC 60601-1-6IEC 60601-2-34ISO 80601-2-61
IEC 62366 included IEC 62304IEC 60601-2-49ANSI/AAMI EC-13
IEC 60601-1-8

The predicates for the subject devices are:

K12390003/19/2013Philips SureSigns VM Series Patient Monitors
K05076204/22/2005Philips MP20, MP30, MP40, MP50, MP60, MP70 MP90 IntellliVue Patient Monitor
K04260111/19/2004Respironics CAPNOSTAT 5 CO2 sensor
K10042807/09/2010Masimo Rainbow SET® Radical 7R CO-Oximeter and Accessories
K10152107/25/2010Philips ST/AR ST and Arrhythmia Software, Release K.O
Primary 510(k) Predicate

Philips SureSigns VM Series Patient Monitors

K123900 03/19/2013

Indications for Use:

Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The intended uses are for monitoring, analysis, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.

Patient Types
ParameterAdultPediatricNeonatal
ECG
Respiration - Impedance
Respiration - RRa

{6}------------------------------------------------

Patient Types
ParameterAdultPediatricNeonatal
Respiration - awRR
NBP
SpO2
SpHb
SpCO
Temperature
CO2
IBP
Cardiac Output
Arrhythmia (Cardiotach)
Arrhythmia (Basic)
Arrhythmia (Advanced)
ST segment Analysis

Contraindications: Not for transport outside the healthcare facility

Conclusion

The subject Efficia CM10, CM120 and CM150 patient monitors utilize the same fundamental scientific technology as the predicate devices on the market. The energy source is electrical with secondary battery back like the primary predicate. The subject Efficia CM10, CM120 and CM150 patient monitors use the same algorithms and have the similar performance to their predicates. They have the same or better level of safety and essential performance as compared to their predicate devices when reviewing the data of the device history file. It is our assessment that the subject devices demonstrate that they are substantially equivalent to the identified predicates and have the same level of safety and effectiveness.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.