(84 days)
The IntelliVue CL SpO2 Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters SpO2 and pulse rate wirelessly. The intended use of the IntelliVue CL SpO2 Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A, is for monitoring, recording, and alarming arterial oxygen saturation and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device.
The IntelliVue CL NBP Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters non-invasive blood pressure and pulse rate wirelessly.
The intended use of the IntelliVue CL NBP Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A is for monitoring, recording, and alarming of systolic, diastolic, and mean pressure and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device.
The IntelliVue Patient Monitors MP5, MP5T, MP2, X2 are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside of the hospital environment. The MP2, X2, and MP5 monitors are also intended for use during patient transport outside of the hospital environment.
They are not intended for home use. The monitors are only for use on one patient at a time. They are not therapeutic devices.
The IntelliVue Telemetry System Transceiver TRx4841A is indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
Intended to provide ambulatory and bedside monitoring, recording and alarming of ECG and SpO2 parameters of adult and pediatric patients in transport and hospital environments.
It is intended to be used by trained healthcare personnel. It is not intended for home use. The device is only for use on one patient at a time. It is not for use with infant or neonatal patients.
The new Philips IntelliVue CL SpO2 Pod is a small, battery powered, wrist worn pulse oximeter device for cableless monitoring of patients. It contains Philips FAST-Sp02 (Fourier Artifact Suppression Technology) to provide reliable saturation values under various artifact conditions including motion and low perfusion. It provides continuous operating mode and intermittent operating mode with configurable measurement intervals. Integrated monochrome LCD display shows measured values, measurement signal quality, battery state, and RF signal strength. It has three hardkeys for basic operation and navigation. It supports specialized Philips reusable and disposable Sp02 sensors.
The new Philips IntelliVue CL NBP Pod is a small, battery powered, non-invasive blood pressure and pulse rate measurement device for cableless monitoring of patients. It uses oscillometric method for measuring NBP. It produces numerics for systolic, diastolic and mean blood pressure values and pulse rate. Integrated monochrome LCD Display shows measured values, battery state, and RF signal strength. It has three hardkeys for basic operation and navigation. It supports specialized Philips reusable and disposable NBP cuffs.
The Philips IntelliVue MP5, MP5T, MP2, X2 are robust, portable, lightweight, compact in size and modular in design patient monitors with integrated displays and multiple physiological measurements and interfaces to dedicated external measurement devices. With the built-in battery the monitors can also function during transport situations.
With the built-in physiological measurements the monitors MP5, MP2, and X2 provide the following measurement capability: Invasive Blood Pressure, Temperature, CO2, ECG/Respiration, SpO2, and NIBP. The MP5T model provides internal SpO2 and NIBP measurements.
The monitors MP5, MP5T, MP2, X2 can be interfaced to several dedicated external devices to extend the measurement capabilities. The monitors MP5, MP2 and X2 can also be connected to another IntelliVue patient monitor (MP20 to MP90), where they act as multi-measurement modules acquiring measurements for the host monitor. The network interfaces provide the monitors MP5, MP2 and X2 with networking capability via wired or wireless network connections. The monitors MP5, MP5T, MP2, and X2 can also interface via cable or wirelessly via short range radio to the IntelliVue Telemetry Transceiver TRx4841A.
As a result of the current modification, the MP5, MP5T, MP2, X2 patient monitors can be interfaced wirelessly via short range radio to the new CL NBP and SpO2 Pods.
The Philips IntelliVue Telemetry System Transceiver TRx4841A is a small, lightweight, battery powered patient-worn device for monitoring ECG and Sp02. The TRx4841A Transceiver uses transmission through a body (finger tip) method and Philips FAST-Sp02 algorithm to measure SpO2. The TRx4841A Transceiver provides ECG using standard ECG and EASI derived 12-lead monitoring, 6lead monitoring with 2 V-leads and EASI derived 12-lead ECG technology with only 5 leads.
The TRx4841A Transceiver sends patient digitized ECG/SpO2 signals via cable or wirelessly (via SRR) to the IntelliVue Patient Monitors MP5, MP5T, MP2, X2 or wirelessly (via WMTS 1.4 GHz) to the Philips IntelliVue Information Center, which displays ECG waveforms, heart rate, and SpO2 values, detects and analyzes the ECG for cardiac arrhythmias and displays alarm conditions and INOP's. Displays, settings, recordings, and alarms are controlled from the IntelliVue Information Center or the Patient Monitors.
As a result of the current modification, the transceiver TRx4841A can be interfaced wirelessly via short range radio to the new CL NBP and SpO2 Pods. The transceiver TRx4841A sends patient digitalized SpO2/NBP signals from the connected CL NBP and SpO2 Pods wirelessly (via WMTS 1.4 GHz) to the Philips IntelliVue Information Center, which displays the physiological measurement values and indicates alarms.
Here is an analysis of the provided text regarding the acceptance criteria and supporting studies for the Philips IntelliVue CL SpO2 Pod and CL NBP Pod:
Device: Philips IntelliVue CL SpO2 Pod and IntelliVue CL NBP Pod
1. Table of Acceptance Criteria and Reported Device Performance
| Parameter | Acceptance Criteria (Standard) | Reported Device Performance (Summary) |
|---|---|---|
| IntelliVue CL SpO2 Pod | ||
| SpO2 Accuracy | ISO 9919 (recognized consensus standard) | Met performance requirements of ISO 9919 and recommendations of FDA draft guidance. |
| Pulse Rate Accuracy | Within specified measurement range | Validated within specifications, substantial equivalent to predicate device. |
| Averaging of Numerics (SpO2, Pulse Rate, Perfusion Index) | Not explicitly stated but implied to be within specifications. | Verification of averaging for calculated numerics. |
| IntelliVue CL NBP Pod | ||
| Systolic/Diastolic Blood Pressure Accuracy | ANSI/AAMI SP10 (recognized consensus standard) | Met performance requirements of ANSI/AAMI SP10. |
| Pulse Rate Accuracy | Within specified measurement range | Validated within specified measurement range. |
| Overall NBP performance compared to predicate devices | Not explicitly stated as a separate acceptance criterion, but implied. | Produced same results as appropriate predicate devices. |
2. Sample Sizes Used for the Test Set and Data Provenance
- IntelliVue CL SpO2 Pod: A "clinical desaturation study" was performed. The specific number of subjects/samples is not provided in the given text.
- IntelliVue CL NBP Pod: A "clinical study" using the auscultatory reference method was performed. The specific number of subjects/samples is not provided in the given text.
- Data Provenance: The text does not specify the country of origin. Both studies are described as "clinical studies" and "clinical performance validation," which implies prospective data collection, but this is not explicitly stated.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
- IntelliVue CL SpO2 Pod: The text refers to a "clinical desaturation study" to determine SpO2 accuracy. The ground truth (reference SpO2 values) for such studies is typically established using a co-oximeter connected to arterial blood samples analyzed by laboratory personnel. However, the exact number and qualifications of experts involved in establishing this reference are not specified in the provided text.
- IntelliVue CL NBP Pod: The ground truth for blood pressure accuracy was established using the "auscultatory reference method" according to ANSI/AAMI SP10. This method typically involves trained observers (e.g., medical professionals) using a stethoscope and sphygmomanometer. However, the number and specific qualifications of these experts are not provided in the text.
4. Adjudication Method for the Test Set
The text does not provide any information regarding an adjudication method for either the SpO2 or NBP clinical studies.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A multi-reader multi-case (MRMC) comparative effectiveness study was not performed for these devices. The studies described are validation studies against established reference methods and standards, focusing on the device's standalone accuracy rather than human reader improvement with AI assistance.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, standalone performance studies were conducted. The clinical studies for both the CL SpO2 Pod and CL NBP Pod evaluate the accuracy of the device's measurements (SpO2, pulse rate, blood pressure) against a reference standard. This is a measure of the device's algorithm-only performance.
7. Type of Ground Truth Used
- IntelliVue CL SpO2 Pod: The clinical desaturation study (to determine SpO2 accuracy) implies the use of a co-oximeter as the gold standard for arterial oxygen saturation.
- IntelliVue CL NBP Pod: The clinical study used the auscultatory reference method according to ANSI/AAMI SP10 for blood pressure.
8. Sample Size for the Training Set
The provided text does not mention or describe a training set for these devices. The descriptions focus on performance validation studies (test sets) for their accuracy against established medical standards. This suggests a traditional medical device development approach where algorithms are designed based on known physiological principles and validated against clinical data, rather than being "trained" in the machine learning sense on a distinct training dataset.
9. How the Ground Truth for the Training Set Was Established
Since no training set is described, information on how its ground truth was established is not applicable.
{0}------------------------------------------------
KIOL600 PI/7
AUG 3 1 2010
510 (k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
- The submitter of this premarket notification is: Markus Stacha Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-2840, Fax: ++49 7031 463-2442 e-mail: markus.stacha@philips.com
This summary was prepared on June 02, 2010.
- The names of the devices are:
New devices: Trade name: Philips IntelliVue CL SpO2 Pod Common name: Telemetry Transceiver
Trade name: Philips IntelliVue CL NBP Pod Common name: Telemetry Transceiver
Modified devices:
Trade names: Philips IntelliVue Patient Monitors MP5, MP5T, MP2, X2
Common names: Multiparameter Patient Monitors
Trade name: Philips IntelliVue Telemetry System Transceiver TRx4841A
Common name: Telemetry Transceiver
Classification names are as follows:
| Device Panel | Classification | ProCode | Description |
|---|---|---|---|
| CardiovascularDevices | §870.1025, II | DSI | Detector and Alarm, Arrhythmia |
| §870.1100, II | DSJ | Alarm, Blood-Pressure | |
| §870.1110, II | DSK | Computer, Blood-Pressure | |
| §870.1120, II | DXQ | Cuff, Blood-Pressure | |
| §870.1130, II | DXN | System, Measurement, Blood-Pressure, Non-Invasive | |
| §870.1435, II | DXG | Computer, Diagnostic, Pre-Programmed, Single-Function | |
| §870.2030, II | DRT | Monitor, Cardiac (incl.Cardiotachometer & Rate Alarm) | |
| §870.2060, II | DRQ | Amplifier and SignalConditioner, Transducer Signal | |
| §870.2700, II | DQA | Oximeter | |
| §870.2900, I | DSA | Cable, Transducer andElectrode, incl. PatientConnector | |
| §870.2910, II | DRG | Transmitters and Receivers,Physiological Signal,Radiofrequency |
{1}------------------------------------------------
| Device Panel | Classification | ProCode | Description |
|---|---|---|---|
| - | MSX | System, Network andCommunication, PhysiologicalMonitors |
-
- The new Philips IntelliVue CL SpO2 Pod and IntelliVue CL NBP Pod are substantially equivalent to the previously cleared Nihon Kohden Transmitter ZS-940PA series marketed pursuant to K043517, Philips IntelliVue Telemetry System Transceiver TRx4841A marketed pursuant to K081793, K041741, K040357, and Philips Multi-Measurement Server M3000A/M3001A marketed pursuant to K001333, K013427, K020531, K030973, K033715, K062268, K090360.
The new reusable NBP cuffs 989803163171, 989803163191, 989803163211 and Extension Air Hose 989803163131 for the CL NBP Pod are substantially equivalent to the Philips reusable cuffs M1573A, M1574A, M1575A marketed pursuant to K001333, Philips Extension Air Hose M1598B marketed pursuant to K031187.
- The new Philips IntelliVue CL SpO2 Pod and IntelliVue CL NBP Pod are substantially equivalent to the previously cleared Nihon Kohden Transmitter ZS-940PA series marketed pursuant to K043517, Philips IntelliVue Telemetry System Transceiver TRx4841A marketed pursuant to K081793, K041741, K040357, and Philips Multi-Measurement Server M3000A/M3001A marketed pursuant to K001333, K013427, K020531, K030973, K033715, K062268, K090360.
The new disposable NBP cuffs 989803163181, 989803163201, 989803163221 for the CL NBP Pod are substantially equivalent to the Philips disposable cuffs M4574B, M4575B, M4577B marketed pursuant to K071885, Hewlett-Packard/Philips disposable cuffs M1876A, M1877A, M1878A marketed pursuant to K941811, K925910, К923682, К923343, К922058, К910490, К903771, К903523, К900032, K896030, K882609.
The modified reusable Sp02 sensor Mobile CL RSp02-1A Reusable SpO2 Sensor for IntelliVue CL SpO2 Pod is substantially equivalent to the Philips reusable SpO2 sensor M1192AN marketed pursuant to K033715 and Nellcor disposable Sp02 sensor MAX-A marketed pursuant K081937.
The modified disposable SpO2 sensor Mobile CL DSpO2-1A Single Patient Sp02 Sensor for IntelliVue CL Sp02 Pod is substantially equivalent to the Philips disposable SpO2 sensor M1134A marketed pursuant to K091572 Philips reusable SpO2 sensor M1192AN marketed pursuant to K033715, and Nellcor disposable SpO2 sensor MAX-A marketed pursuant to K081937.
The modified Philips IntelliVue Patient Monitors MP5, MP5T, MP2, X2 are substantially equivalent to previously cleared IntelliVue Patient Monitors MP5, MP5, MP2, X2 marketed pursuant to K062392, K063725, K071426, K072070, K081793, K082633, K083228, K083517, K091395, K100939.
The modified Philips IntelliVue Telemetry System Transceiver TRx4841A is substantially equivalent to previously cleared IntelliVue Telemetry System Transceiver TRx4841A marketed pursuant to K081793, K041741, K040357.
4. Description of the devices
The new Philips IntelliVue CL SpO2 Pod is a small, battery powered, wrist worn pulse oximeter device for cableless monitoring of patients. It contains Philips FAST-Sp02 (Fourier Artifact Suppression Technology) to provide reliable saturation
{2}------------------------------------------------
values under various artifact conditions including motion and low perfusion. It provides continuous operating mode and intermittent operating mode with configurable measurement intervals. Integrated monochrome LCD display shows measured values, measurement signal quality, battery state, and RF signal strength. It has three hardkeys for basic operation and navigation. It supports specialized Philips reusable and disposable Sp02 sensors.
The new Philips IntelliVue CL NBP Pod is a small, battery powered, non-invasive blood pressure and pulse rate measurement device for cableless monitoring of patients. It uses oscillometric method for measuring NBP. It produces numerics for systolic, diastolic and mean blood pressure values and pulse rate. Integrated monochrome LCD Display shows measured values, battery state, and RF signal strength. It has three hardkeys for basic operation and navigation. It supports specialized Philips reusable and disposable NBP cuffs.
The Philips IntelliVue MP5, MP5T, MP2, X2 are robust, portable, lightweight, compact in size and modular in design patient monitors with integrated displays and multiple physiological measurements and interfaces to dedicated external measurement devices. With the built-in battery the monitors can also function during transport situations.
With the built-in physiological measurements the monitors MP5, MP2, and X2 provide the following measurement capability: Invasive Blood Pressure, Temperature, CO2, ECG/Respiration, SpO2, and NIBP. The MP5T model provides internal SpO2 and NIBP measurements.
The monitors MP5, MP5T, MP2, X2 can be interfaced to several dedicated external devices to extend the measurement capabilities. The monitors MP5, MP2 and X2 can also be connected to another IntelliVue patient monitor (MP20 to MP90), where they act as multi-measurement modules acquiring measurements for the host monitor. The network interfaces provide the monitors MP5, MP2 and X2 with networking capability via wired or wireless network connections. The monitors MP5, MP5T, MP2, and X2 can also interface via cable or wirelessly via short range radio to the IntelliVue Telemetry Transceiver TRx4841A.
As a result of the current modification, the MP5, MP5T, MP2, X2 patient monitors can be interfaced wirelessly via short range radio to the new CL NBP and SpO2 Pods.
The Philips IntelliVue Telemetry System Transceiver TRx4841A is a small, lightweight, battery powered patient-worn device for monitoring ECG and Sp02. The TRx4841A Transceiver uses transmission through a body (finger tip) method and Philips FAST-Sp02 algorithm to measure SpO2. The TRx4841A Transceiver provides ECG using standard ECG and EASI derived 12-lead monitoring, 6lead monitoring with 2 V-leads and EASI derived 12-lead ECG technology with only 5 leads.
{3}------------------------------------------------
: .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The TRx4841A Transceiver sends patient digitized ECG/SpO2 signals via cable or wirelessly (via SRR) to the IntelliVue Patient Monitors MP5, MP5T, MP2, X2 or wirelessly (via WMTS 1.4 GHz) to the Philips IntelliVue Information Center, which displays ECG waveforms, heart rate, and SpO2 values, detects and analyzes the ECG for cardiac arrhythmias and displays alarm conditions and INOP's. Displays, settings, recordings, and alarms are controlled from the IntelliVue Information Center or the Patient Monitors.
As a result of the current modification, the transceiver TRx4841A can be interfaced wirelessly via short range radio to the new CL NBP and SpO2 Pods. The transceiver TRx4841A sends patient digitalized SpO2/NBP signals from the connected CL NBP and SpO2 Pods wirelessly (via WMTS 1.4 GHz) to the Philips IntelliVue Information Center, which displays the physiological measurement values and indicates alarms.
5. Intended Use/ Indications for Use
The IntelliVue CL SpO2 Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters SpO2 and pulse rate wirelessly. The intended use of the IntelliVue CL SpO2 Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A, is for monitoring, recording, and alarming arterial oxygen saturation and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device.
This is the same intended use as that of the predicate Nihon Kohden Transmitter ZS-940PA series, except for the compatible monitoring devices and number of physiological parameters: whereas the Transmitter ZS-940PA provides SpO2, pulse rate, noninvasive blood pressure, ECG, and respiration, the CL SpO2 Pod only provides SpO2 and pulse rate. The new CL SpO2 Pod is compatible with Philips monitoring devices whereas the Nihon Kohden Transmitter ZS-940PA is compatible with Nihon Kohden monitoring devices.
The IntelliVue CL NBP Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters non-invasive blood pressure and pulse rate wirelessly.
The intended use of the IntelliVue CL NBP Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A is for monitoring, recording, and alarming of systolic, diastolic, and mean pressure and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device.
This is the same intended use as that of the predicate Nihon Kohden Transmitter ZS-940PA series, except for the compatible monitoring devices and number of physiological parameters:
{4}------------------------------------------------
whereas the Transmitter ZS-940PA provides non-invasive blood pressure, pulse rate, SpO2, ECG, and respiration, the CL NBP Pod only provides non-invasive blood pressure and pulse rate. The new CL SpO2 Pod is compatible with Philips monitoring devices whereas the Nihon Kohden Transmitter ZS-940PA is compatible with Nihon Kohden monitoring devices.
The modified Philips IntelliVue Patient Monitors MP5, MP5T, MP2, X2 have the same intended use as the legally marketed monitors MP5, MP5T, MP2, X2.
The IntelliVue Patient Monitors MP5, MP5T, MP2, X2 are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside of the hospital environment. The MP2, X2, and MP5 monitors are also intended for use during patient transport outside of the hospital environment. They are not intended for home use. The monitors are only for use on one patient at a time. They are not therapeutic devices.
The modified IntelliVue Telemetry System Transceiver TRx4841A has the same intended use as the legally marketed Transceiver TRx4841A.
The IntelliVue Telemetry System Transceiver TRx4841A is indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended to provide ambulatory and bedside monitoring, recording and alarming of ECG and SpO2 parameters of adult and pediatric patients in transport and hospital environments. It is intended to be used by trained healthcare personnel. It is not intended for home use. The device is only for use on one patient at a time. It is not for use with infant or neonatal patients.
-
6.Technological Characteristics
The new IntelliVue CL SpO2 Pod and CL NBP Pod and their medical accessories have the same or substantial equivalent technological characteristics, such as design, material, energy source, portability, user interface, radio technology, measurement principle, as the legally marketed predicate devices The new IntelliVue CL SpO2 Pod and CL NBP Pod and their medical accessories use partly different patient skin contacting materials, than the predicates. The only relevant characteristic for the intended applications is the biocompatibility of these materials. The biocompatibility of the different materials, hence the equivalence to the predicate devices, has been proved by successful testing to the applicable biocompatibility standards of the IEC 10993-x standard family.
The modification of the IntelliVue Patient Monitors MP5, MP5T, MP2, X2 and the IntelliVue Telemetry System Transceiver TRx4841A is solely limited to the software changes necessary due to the
{5}------------------------------------------------
wireless communication with the introduced IntelliVue CL SpO2 Pod and CL NBP Pod. This software modification does not affect the technological characteristics of the modified patient monitors MP5, MP5T, MP2, X2 and the transceiver TRx4841A.
7.Brief discussion of non-clinical performance tests
The determination of the substantial equivalence of the performance of the new IntelliVue CL SpO2 Pod and CL NBP Pod and their measurement accessories was based, in addition to the clinical performance validation, on bench performance testing. For the CL Sp02 Pod the accuracy of the pulse rate in the specified measurement range has been validated. In addition, the verification of averaging of the calculated numerics (Sp02, pulse rate, perfusion index) has been verified. The bench performance testing has demonstrated that the new CL SpO2 Pod has performed within specifications, which have been substantial equivalent to those of the predicate device.
For the CL NBP Pod the accuracy of systolic/ diastolic and mean pressure, such as accuracy of pulse rate in the specified measurement range has been validated. The comparative testing has demonstrated that the new CL NBP Pod and its cuffs have produced the same results as the appropriate predicate devices.
8.Brief discussion of clinical performance tests
The determination of substantial equivalence of the performance of the new IntelliVue CL SpO2 Pod and CL NBP Pod and their measurement accessories was based, in addition to the bench performance testing, on clinical performance validation. For the CL SpO2 Pod and its sensors the clinical desaturation study to determine the Sp02 accuracy in the specified measurement range, according to the recognized consensus standard ISO 9919, has been performed. The clinical desaturation study has demonstrated that the new CL SpO2 Pod and its sensors have worked within the specified accuracy and have met the performance requirements of the ISO 9919 standard and the recommendations of the draft guidance document 'Pulse Oximeters - Premarket Notification Submissions [510(k)s]' .
For the CL NBP Pod and its cuffs the clinical study to determine the systolic and diastolic accuracy of blood pressure in the specified measurement range, using the auscultatory reference method according to the recognized consensus standard ANSI/AAMI SP10, has been performed. The clinical study has demonstrated that the new CL NBP Pod and its cuffs have worked within the specified accuracy and have met the performance requirements of the ANSI/AAMI SP10 standard.
9.Conclusion
ﻟﺔ ، ﺳﻴﺮ
Verification and validation testing activities were conducted to establish the performance, safety, functionality, and reliability characteristics of the new and modified devices.
Page 19
{6}------------------------------------------------
Declarations of Conformity to the above listed recognized consensus standards are provided in the Annexes A.01 to A.06.
9.2 Summary Reports
The new IntelliVue CL SpO2 Pod and CL NBP Pod, the modified IntelliVue Patient Monitors MP5, MP5T, MP2, X2 and the modified IntelliVue Telemetry System Transceiver TRx4841A have been subject to a series of verification and validation (V&V) testing, as required by the applicable standards, relevant Sp02 and NIBP guidance documents, hazard analysis, and additional R&D V&V requirements, in order to , demonstrate their safe and effective use in accordance with specifications under simulated and clinical use conditions.
The V&V activities have confirmed safe, reliable, secure, and effective function within specifications of the new IntelliVue CL SpO2 Pod and CL NBP Pod, the modified Monitors MP5, MP5T, MP2, X2 and Transceiver TRx4841A, in the intended environment of use.
In addition to the testing required in the applicable recognized consensus standards and particular guidance documents, the following testing has been performed:
- · Environmental testing
- . Mechanical testing on the new IntelliVue CL SpO2 Pod and CL NBP Pod.
- Temperature and humidity testing on the new IntelliVue CL SpO2 Pod and CL NBP Pod.
All specified test requirements have been met. Please refer to Sub-Sec 9.2.1 'Environmental Testing' for details.
- . Radio Frequency (RF) wireless communication testing on the new CL SpO2 Pod and CL NBP Pod with the modified monitors MP5/MP5T, MP2/X2 and the modified transceiver TRx4841A, based on the FDA guidance document 'Radio-Frequency Wireless Technology in Medical Devices' .
All specified test requirements have been met.
The following specific items have been subject to the RF wireless communication tests:
- " Performance of wireless functions For details please refer to Sub-Sec. 9.2.2 'Radio Frequency (RF) Testing' and Table 16.9.1 'Summary of Tests on the IntelliVue CL SpO2 Pod/ NBP Pod, as required
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo of the Department of Health & Human Services. The logo consists of three parallel lines on the left side and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right side. The text is in a bold, sans-serif font.
Image /page/7/Picture/1 description: The image contains the text 'Public Health Service'. The text is displayed in a simple, sans-serif font. The words are arranged on a single line, with 'Public' and 'Health' appearing to be slightly larger than 'Service'.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Philips Medizin Systeme Boeblingen GmbH c/o Mr. Markus Stacha Sr. Regulatory Affairs Engineer Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany
AUG 31 2010
Re: K101600
Device Name: Philips IntelliVue CL SpO2 Pod and IntelliVue CL NBP Pod Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II (Two) Product Codes: DQA, DSI, DSJ, DSK, DXQ, DXN, DXG, DRT, DRQ, DSA, DRG, MSX, MHX Dated: June 2, 2010 Received: June 8, 2010
Dear Mr. Stacha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Mr. Markus Stacha
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please . go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
M. A. Killeen
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Indications for Use
510(k) Number (if known): KIO 1600
Device Names:
- Philips IntelliVue CL SpO2 Pod and IntelliVue CL NBP Pod ・
- Philips IntelliVue Patient Monitors MP5, MP5T, MP2, X2 -
- -Philips IntelliVue Telemetry System Transceiver TRx4841A
Indications for Use:
IntelliVue CL Sp02 Pod:
The IntelliVue CL SpO2 Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters SpO2 and pulse rate wirelessly. The intended use of the IntelliVue CL SpO2 Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A, is for monitoring, recording, and alarming arterial oxygen saturation and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device.
IntelliVue CL NBP Pod:
。 . ·····
The IntelliVue CL NBP Pod is indicated for use by health care professionals whenever there is a need for monitoring the physiological patient parameters non-invasive blood pressure and pulse rate wirelessly.
The intended use of the IntelliVue CL NBP Pod when used together with IntelliVue Patient Monitors MP5, MP5T, MP2, X2, or with the IntelliVue Telemetry System Transceiver TRx4841A is for monitoring, recording, and alarming of systolic, diastolic, and mean pressure and pulse rate of adult and pediatric patients inside hospitals. The device is intended for use by health care professionals. It is not intended for home use. The device is not a therapeutic device.
continued on next page
| Prescription(Part 21 CFR 801 Subpart D) | Yes |
|---|---|
| AND/OR | |
| Over-The-Counter(21 CFR 807 Subpart C) | No |
| Use | Use |
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE
IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M.A. Willieme
f Cardiovascular Devices
510(k) Number
Page 11
{10}------------------------------------------------
Indications for Use (continued) :
Philips IntelliVue Patient Monitors MP5, MP5T, MP2, X2:
The IntelliVue Patient Momitors MP5, MP5T, MP2, X2 are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside of the hospital environment. The MP2, X2, and MP5 monitors are also intended for use during patient transport outside of the hospital environment.
They are not intended for home use. The monitors are only for use on one patient at a time. They are not therapeutic devices.
Philips IntelliVue Telemetry System Transceiver TRx4841A:
The IntelliVue Telemetry System Transceiver TRx4841A is indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
Intended to provide ambulatory and bedside monitoring, recording and alarming of ECG and SpO2 parameters of adult and pediatric patients in transport and hospital environments.
It is intended to be used by trained healthcare personnel. It is not intended for home use. The device is only for use on one patient at a time. It is not for use with infant or neonatal patients.
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).