(164 days)
Not Found
No
The description focuses on the physical components and function of a blood pressure cuff, with no mention of AI or ML. The "Not Found" entries for AI/ML mentions and training/test sets further support this.
No.
The device is a blood pressure cuff, which is an accessory used for measurement and does not provide a therapeutic effect.
No
The device is a cuff used for non-invasive blood pressure measurement, which is a physiological measurement, not a diagnostic one in itself. The Intended Use states it's for "non-invasive measurement of infant, pediatric and adult human blood pressure," implying it provides data for diagnosis rather than making a diagnosis.
No
The device description explicitly states it comprises physical components like tubing, an inelastic sleeve, an inflatable bladder, and a hook and loop fastener, which are hardware.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the device is for the "non-invasive measurement of infant, pediatric and adult human blood pressure." This is a measurement taken directly from the patient's body, not from a sample of bodily fluid or tissue.
- Device Description: The description details a cuff that is wrapped around a patient's limb and connected to a blood pressure measuring system. This is a physical interface with the patient, not a device that analyzes a biological sample.
- Lack of IVD Characteristics: IVD devices are designed to examine specimens derived from the human body (like blood, urine, tissue) to provide information about a physiological state, health, or disease. This device does not perform such analysis.
Therefore, the Philips Multi-Patient and Single-Patient cuffs are considered non-invasive medical devices used for physiological measurement, not in vitro diagnostics.
N/A
Intended Use / Indications for Use
Philips Multi-Patient and Single-Patient cuffs are to be used with identified devices intended for use by, or under the supervision of, a licensed physician or other healthcare provider for the non-invasive measurement of infant, pediatric and adult human blood pressure.
Product codes (comma separated list FDA assigned to the subject device)
DXQ, OED
Device Description
The device comprises tubing attached to an inelastic sleeve with an integral inflatable bladder that is wrapped around a patient's limb. It is secured by a hook and loop fastener. Device tubing is attached to an NIBP measuring system.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
patient's limb
Indicated Patient Age Range
infant, pediatric and adult
Intended User / Care Setting
licensed physician or other healthcare provider
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Device testing was performed according to the Validation Traceability Matrix. These activities were conducted to establish the performance and reliability characteristics of the modified devices. All tests passed.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1120 Blood pressure cuff.
(a)
Identification. A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.(b)
Classification. Class II (performance standards).
0
71888 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
Philips series of Multi-Patient Cuffs and Single-Patient Cuffs
Submitter's Name and Address
Submitter's Name: | Philips Medical Systems |
---|---|
Division: | Medical Consumables and Sensors |
Address: | 3000 Minuteman Road |
City, State, and Zip: | Andover, MA, 01810 |
Contact Person
Name: | Peter Schipelliti |
---|---|
Title: | Regulatory Affairs Specialist |
Telephone: | (978) 687-1501 |
Facsimile: | (978) 659-7712 |
E-mail: | Peter.schipelliti@philips.com |
Date of Summary
Date:
December 11, 2007
Manufacturing Facility Address
Manufacturer: | Philips Medical Systems |
---|---|
Address: | 3000 Minuteman Road |
City, State, and ZIP: | Andover, MA, 01810 |
Establishment Registration Number
Est. Registration Number: 1218950
New Device Details
Proprietary of Trade Names
Proprietary or Trade Name:
Philips series of Multi-Patient Cuffs and Single-Patient Cuffs
New Device Common Name
Common Name:
Blood Pressure Cuff
Philips Medical Systems APPENDIX E Philips series of Multi-Patient Cuffs and Single-Patient Cuffs 510(k)
December 11, 2007 Page 1 of 3
2007
1
New Device Class
Device Class: | Class II |
---|---|
New Device Product Code | |
Device Procode: | DXQ |
New Device CFR | |
Device CFR: | 870.1120 |
New Device Classification Panel | |
Classification Panel: | Cardiovascular |
New Device Classification Name | |
Classification Name: | Cuff, Blood Pressure |
Reason for submission · | |
Reason for submission | Modified Devices |
Predicates
| Predicate #1 | (K001333) Reusable Non-invasive Blood Pressure Comfort
Cuff |
|--------------|----------------------------------------------------------------|
| Predicate #2 | (K901252) Disposable Non-invasive Blood Pressure Cuff |
| Predicate #3 | (K884421) Reusable Non-invasive Blood Pressure Cuff |
Comparison of Technological Characteristics
The modified devices have the same technological characteristics as the legally marketed predicate devices. Safety characteristics are unchanged by these modifications
Summary of Intended Uses
Philips Multi-Patient and Single-Patient cuffs are to be used with identified devices intended for use by, or under the supervision of, a licensed physician or other healthcare provider for the non-invasive measurement of infant, pediatric and adult human blood pressure.
2
Description of Device
The device comprises tubing attached to an inelastic sleeve with an integral inflatable bladder that is wrapped around a patient's limb. It is secured by a hook and loop fastener. Device tubing is attached to an NIBP measuring system.
Device Testing
Device testing was performed according to the Validation Traceability Matrix. These activities were conducted to establish the performance and reliability characteristics of the modified devices. All tests passed.
Voluntary Standards Used in Determination of Substantial Equivalence:
The cuffs were tested according to:
- ANSI/AAMI SP10:2002 +A1:2003 "Manual, electronic, or automated . sphygmomanometers"
- EN 1060-1:1995 +A1:2002; "Non-invasive sphygmomanometers Part 1: . General Requirements"
- EN 1060-3:1997 +A1:2005; "Non-invasive sphygmomanometers Part 3: . Supplementary requirements for electro-mechanical blood pressure measuring systems"
- IEC 60601-1:1988 +A1:1991 +A2:1995; "Medical Electrical Equipment → . Part 1: General Requirements for Safety"
- . IEC 60601-2-30:1999; "Medical electrical equipment Part 2-30: Particular requirements for the safety, including essential performance, of automatic cycling non-invasive blood pressure monitoring equipment"
- ISO 10993-1:2003; "Biological evaluation of medical devices Part 1: . Evaluation and Testing"
All tests passed and therefore it is concluded that the product is safe and effective.
Conclusion
Based on similarity in technology, characteristics and the same intended use as the predicates, these devices are substantially equivalent.
3
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the staff and two snakes intertwined around it. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
DEC 2 0 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Philips Medical Systems c/o Mr. Peter Schipelliti Regulatory Affairs Specialist 3000 Minuteman Road Andover, MA 01810-1099
Re: K071885
Trade/Device Name: Multi-Patient Cuffs, Models M4552B, M4553B, M4554B, M4555B, M4556B, M4557B, M4558B, M4559B, M4562B, M4563B, M4564B, M4565B, M4566B, M4567B, M4568B, and M4569B (16 Models); and Single-Patient Cuffs, Models M4572B, M4573B, M4573B, M4574B, M4575B, M4576B, M4577B, M4578B, M4579B, M4582B, M4583B, M4584B, M4585B, M4586B, M4587B, M4588B, and, M4589B (16 Models) Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II (two) Product Code: OED Dated: November 21, 2007 Received: November 23, 2007
Dear Mr. Schipelliti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
4
Page 2 - Mr. Peter Schipelliti
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
B. Zimmerman for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
PHILIPS
510(k) Number_K071885
Device Name: Philips series of Multi-Patient Cuffs and Single-Patient Cuffs
Indications For Use:
Philips Multi-Patient and Single-Patient cuffs are to be used with identified devices intended for use by, or under the supervision of, a licensed physician or other healthcare provider for the non-invasive measurement of infant, pediatric and adult human blood pressure.
Prescription Use (21 CFR 801 Subpart D) | X |
---|---|
AND/OR | |
Over-The-Counter Use (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE) |
---|
-------------------------------------------------------- |
Bhimmerma | |
---|---|
(Division Sign-Off) | |
Division of Cardiovascular Devices | |
510(k) Number | K071885 |
Philips Medical Systems Philips Series of Multi-Patient Cuffs and Single-Patient Cuffs 510(k)
December 11, 2007 Page xv of xviii