(89 days)
OrthADAPT® PR is intended to be used for implantation to reinforce soft tissue, including but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, hernias. suture-line reinforcement and other reconstructive procedures.
The device is also intended for the reinforcement of soft tissues repaired by sutures or suture anchors during tendon repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons.
OrthADAPT® PR is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps or other tendons. Sutures, used to repair the tear, and sutures or bone anchors. used to attach the tissue to the bone, provide biomechanical strength for the tendon repair.
OrthADAPT® PR is comprised of collagen matrix reinforced by a woven polymer to provide permanent durability. The collagen matrix, which is derived from the same equine pericardial tissue used in the fabrication of its predicate device, the OrthADAPT® Bioimplant, has been decellularized and crosslinked and the entire device has been exposed to a liquid chemical sterilant. The product passes USP sterility testing and satisfies FDA requirements for LAL endotoxin limit for a medical device. The product must be rinsed prior to use following the procedures described in the Instructions for Use.
The provided text describes the OrthADAPT® PR Bioimplant, a surgical mesh device, and its substantial equivalence to predicate devices, rather than a study with acceptance criteria in the typical sense of diagnostic performance metrics (e.g., sensitivity, specificity). This documentation is a 510(k) summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device, not on proving device efficacy through a comparative performance study with predefined acceptance criteria for AI or diagnostic performance.
Therefore, many of the requested fields cannot be filled as they pertain to a different type of study (e.g., an AI/diagnostic performance study).
Here's the information available from the text, framed within the requested table and addressing the other points where possible:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (for substantial equivalence) | Reported Device Performance (OrthADAPT® PR Bioimplant) |
|---|---|
| Intended Use: Equivalent to predicate devices. | Intended use for soft tissue reinforcement is comparable to predicate devices. Specifically: defects of abdominal/thoracic wall, muscle flap reinforcement, hernias, suture-line reinforcement, and other reconstructive procedures. Also for reinforcement of soft tissues repaired by sutures/suture anchors during tendon repair (rotator cuff, patellar, Achilles, biceps, quadriceps, etc.). |
| Technology: Similar to predicate devices. | Comprised of collagen matrix reinforced by woven polymer. Collagen matrix derived from equine pericardial tissue (same as predicate OrthADAPT® Bioimplant). Decellularized, crosslinked, exposed to liquid chemical sterilant. |
| Performance (Biocompatibility): Equivalent to predicate devices. | Biocompatibility test results indicate the device's biocompatibility profile is equivalent to the predicate devices. |
| Performance (Biomechanical): Equivalent to predicate devices and satisfies mechanical requirements for intended use. | Biomechanical test results indicate the device is equivalent to the predicate devices and satisfies mechanical performance requirements for its intended use. |
| Functionality & Tissue Response: Confirmed through in vivo study. | An animal implant study was performed to confirm the functionality and tissue response characteristics. |
| Sterility: Passes USP sterility testing. | The product passes USP sterility testing. |
| LAL Endotoxin Limit: Satisfies FDA requirements. | The product satisfies FDA requirements for LAL endotoxin limit for a medical device. |
The "study that proves the device meets the acceptance criteria" in this context refers to the collective body of biocompatibility testing, biomechanical bench testing, and an animal implant study, all performed to demonstrate substantial equivalence to predicate devices.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size: Not explicitly stated. The document mentions "biocompatibility testing," "biomechanical bench testing," and "an animal implant study." The sample size for each of these tests is not quantified in the provided text.
- Data Provenance: Not explicitly stated. The document indicates the testing was conducted to support a submission to the FDA (U.S.). It does not specify the country of origin for the animal study or other testing data, nor whether the studies were retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. This device is a surgical mesh, not an AI or diagnostic device that relies on expert interpretation of results for "ground truth" establishment in a diagnostic accuracy study. The "ground truth" here is established through objective physical and biological tests (e.g., tensile strength for biomechanical performance, histological assessment in animal studies for tissue response).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. This is not a study requiring adjudication of expert opinions for a diagnostic outcome.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not a diagnostic device or an AI-assisted interpretation tool. It is a surgical implant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for the performance evaluation of the OrthADAPT® PR Bioimplant is based on:
- Biocompatibility Standards: Meeting established biological safety standards for medical devices.
- Biomechanical Measurements: Objective quantitative measurements of mechanical properties (e.g., tensile strength, tear resistance) against performance requirements deemed suitable for its intended use and comparable to predicate devices.
- In Vivo Animal Study Observations: Histological and functional assessments of tissue response, integration, and functionality in an animal model, conforming to expectations for surgical mesh.
- Sterility and Endotoxin Testing: Objective laboratory tests (USP sterility testing, LAL endotoxin limit).
8. The sample size for the training set
- Not applicable. This is not an AI/machine learning device.
9. How the ground truth for the training set was established
- Not applicable. This is not an AI/machine learning device.
{0}------------------------------------------------
K090288 page 1/2
Synovis Orthopadic and Woundcare, Inc.
OrthADAPT® PR B
510(K) SUMMARY
Submitter Information 1
| A. | Company Name: | Synovis Orthopedic and Woundcare, Inc. |
|---|---|---|
| B. | Company Address: | 6 Jenner, Suite 150Irvine, CA 92618 |
| C. | Company Phone: | (949) 502-3240 |
| D. | Company Facsimile: | (949) 502-3241 |
| E. | Contact Person: | Amy BouclyManager, Regulatory Affairs/Quality Assurance |
| F. | Date: | 06/25/12 |
Device Identification 2
| A. | Device Trade Name: | OrthADAPT® PR Bioimplant |
|---|---|---|
| B. | Common Name: | Surgical Mesh |
| C. | Classification Name(s): | Mesh, Surgical |
| D. | Classification Regulation: | 878.3300 |
| E. | Device Class: | Class II |
| F. | Device Code(s): | FTL, FTM, OXA, OXD, OWY |
| G. | Advisory Panel: | General and Plastic Surgery |
Identification of Predicate Devices 3
The OrthADAPT® PR Bioimplant is substantially equivalent to the following surgical mesh devices, which are cleared for commercial distribution in the United States:
- · OrthADAPT® Bioimplant, Pegasus Biologics, Inc., K043388, K071065
- · Avaulta Plus® BioSynthetic Support System, C.R. Bard, Inc., K063712
- · AMS Elevate® Prolapse Repair System, American Medical Systems (AMS) K080185
- · PeriPatch™ Sheet, PM Devices, K040835
- · SeriScaffold Surgical Mesh, Serica Technologies, Inc., K080442
- · Surgicraft Surgical Mesh System, Surgicraft, K072370
{1}------------------------------------------------
K090288 Page 2/2
Synovis Orthopedic and Woundcare, Inc.
OrthADAPT® PR Bloimplant K080288
4 Device Description
OrthADAPT® PR is comprised of collagen matrix reinforced by a woven polymer to provide permanent durability. The collagen matrix, which is derived from the same equine pericardial tissue used in the fabrication of its predicate device, the OrthADAPT® Bioimplant, has been decellularized and crosslinked and the entire device has been exposed to a liquid chemical sterilant. The product passes USP sterility testing and satisfies FDA requirements for LAL endotoxin limit for a medical device. The product must be rinsed prior to use following the procedures described in the Instructions for Use.
Statement of Intended Use 5
OrthADAPT® PR is intended to be used for implantation to reinforce soft tissue, including but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, hernias, sutureline reinforcement and other reconstructive procedures.
The device is also intended for the reinforcement of soft tissues repaired by sutures or suture anchors during tendon repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons.
OrthADAPT® PR is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps or other tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide biomechanical strength for the tendon repair.
Biocompatibility and Performance Data 6
Biocompatibility testing, biomechanical bench testing, and in vivo performance testing have been conducted to evaluate the biological safety and biomechanical performance characteristics of OrthADAPT® PR, Biocompatibility test results indicate that the device biocompatibility profile is equivalent to the predicate devices. Biomechanical test results indicate that the device is equivalent to the predicate devices and satisfies mechanical performance requirements for its intended use. An animal implant study was performed to confirm the functionality and tissue response characteristics of the OrthADAPT® PR.
Comparison with Predicate Devices 7
The OrthADAPT® PR is comparable to the predicate devices in terms of intended use, technology and performance: It is similar to the predicate devices in that it has the same intended use of soft tissue reinforcement, is comprised of similar materials (collagen and polymeric materials) and is similar in design. The results of testing demonstrate that the performance of OrthADAPT® PR is comparable to the predicate devices.
Page 2 of 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol in blue, with its wings spread and body facing to the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. The text is in all caps and appears to be in a sans-serif font, also in blue.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Pegasus Biologics. Incorporated % Ms. Amy Boucly Director. Regulatory Affairs 6 Jenner Street, Suite 150 Irvine, California 92618
Re: K090288
Trade/Device Name: OrthADAPT® PR Bioimplant Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class 2 Product Code: FTL, FTM, OXA, OXD, OWX Dated: February 3, 2009 Received: February 5, 2009
AUG 2 9 2012
Dear Ms. Boucly:
This letter corrects our substantially equivalent letter of May 5, 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2- Ms. Amy Boucly
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Synovis Orthopedic and Woundcare, Inc.
INDICATIONS FOR USE
510(k) Number: K090288
Device Name: OrthADAPT® PR
Indications for Use: OrthADAPT® PR is intended to be used for implantation to reinforce soft tissue, including but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, hernias. suture-line reinforcement and other reconstructive procedures.
The device is also intended for the reinforcement of soft tissues repaired by sutures or suture anchors during tendon repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons.
OrthADAPT® PR is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps or other tendons. Sutures, used to repair the tear, and sutures or bone anchors. used to attach the tissue to the bone, provide biomechanical strength for the tendon repair.
Prescription Use X (21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daiel Kune fur MM
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K090288
§ 878.3300 Surgical mesh.
(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.