Search Filters

Search Results

Found 719 results

510(k) Data Aggregation

    K Number
    K251557
    Manufacturer
    Date Cleared
    2025-06-12

    (22 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K251557**
    Trade/Device Name: Bard® Mesh; Bard® Mesh Pre-Shaped
    Regulation Number: 21 CFR 878.3300
    Name:** | Mesh, Surgical, Polymeric |
    | Regulatory Class: | II |
    | Regulation Number: | 21 CFR 878.3300
    Name:** | Mesh, Surgical, Polymeric |
    | Regulatory Class: | II |
    | Regulation Number: | 21 CFR 878.3300
    The subject device indications reflect the definition of a surgical mesh given in 21 CFR 878.3300.
    Product Code: FTL | Class II Surgical Mesh 21 CFR 878.3300 Product Code: FTL | Identical |
    | **Intended

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bard® Mesh is indicated to reinforce soft tissue where weakness exists, i.e., repair of hernias and chest wall defects.

    Bard® Mesh Pre-shaped is indicated for the repair of inguinal hernia defects.

    Device Description

    The Bard® Mesh is a nonabsorbable, sterile mesh designed for the reinforcement of soft tissue where weakness exists, i.e., repair of hernias and chest wall defects. The Bard® Mesh Pre-Shaped is a nonabsorbable, pre-trimmed sterile mesh designed for reinforcement of inguinal hernias. The Bard® Mesh is offered in several sizes of rectangular flat sheets and in several pre-shaped configurations, known as Bard® Mesh Pre-Shaped. Bard® Mesh product family products consist of the standard, rectangular or square sizes which can be trimmed as needed to provide surgeons with more freedom to customize the prosthesis prior to implantation. The Bard® Mesh portfolio currently ranges from 1" x 4" to 10" x 14" rectangle mesh. The Bard® Mesh is comprised of knitted polypropylene monofilaments in a square or rectangular flat sheet with selvage edges. The Bard Mesh Pre-Shaped is comprised of knitted polypropylene monofilaments in a pre-shaped configuration with a pre-cut into a shape commonly used by surgeons to protect the spermatic cord or similar structure as part of an inguinal hernia repair. All clinical, biological, and technical characteristics of the Bard® Mesh and Bard® Mesh Pre-Shaped are identical except that Bard® Mesh Pre-Shaped is only indicated for inguinal hernias due to the pre-cut/pre-shaped nature of the product.

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for a medical device called Bard® Mesh. However, this document does not contain the information typically found in a study demonstrating how an AI/algorithm-driven device meets acceptance criteria.

    The Bard® Mesh is a surgical mesh, a physical medical device used to reinforce soft tissue, such as in hernia repair. It is a Class II device, and the clearance is based on its substantial equivalence to a preamendment predicate device (meaning it was marketed before May 28, 1976).

    The document specifically mentions:

    • "As there are no new or increased risks identified, no verification or validation activities are required for these labeling changes. Design verification testing was conducted to support the performance of the pre-shaped configuration." This indicates that for this type of physical device and the nature of the changes (labeling and a pre-shaped configuration), extensive clinical studies or complex performance evaluations are not required in the same way they would be for a novel diagnostic or AI-powered device.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, and ground truth establishment, which are typical for studies validating AI or diagnostic algorithms, is not present in this 510(k) letter. This clearance is based on the device's established technological characteristics and historical use, demonstrating its equivalence to a long-marketed product, rather than proving performance against specific acceptance criteria through a new study as would be done for an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243843
    Manufacturer
    Date Cleared
    2025-06-04

    (173 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Iceland

    Re: K243843
    Trade/Device Name: Tendon Protect (50242)
    Regulation Number: 21 CFR 878.3300
    Device Code:** OWY (Primary), FTM
    Device Classification: Class II
    Regulation Number: 21 CFR 878.3300
    Device Code:** OWY (Primary), FTM
    Device Classification: Class II
    Regulation Number: 21 CFR 878.3300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kerecis® Tendon Protect is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

    Device Description

    Kerecis® Tendon Protect is a medical device made from solid fish skin (North Atlantic Cod) sheets of thickness no less than 0.4 mm. The product is intended to provide a covering around the newly repaired tendon for the management and protection of tendon injuries in which there has been no substantial loss of tissue. The device with natural structure and proteins from fish skin is applied around the newly repaired tendon. The structure of the fish skin provides a supportive environment for new host tissue deposition as part of wound healing while still providing an initial barrier facilitating the sliding of the repaired tendon within the surrounding tissue during tendon use. The biomechanical strength for the tendon itself is provided separately by sutures used to repair the tendon tear and/or by sutures or bone anchors used to attach the tendon tissue to the bone.

    The fish skin is fully integrated into the surrounding tissue over time, with corresponding new host tissue deposition. The device is biocompatible, non-crosslinked, bioresorbable, strong, and pliable. Its tensile strength supports fixation by sutures customary for tendon protection surgical procedures. The device is intended for one-time use and supplied sterile in labelled Tyvek® pouch packaging.

    The subject device is available in following sizes:

    • 3x5 cm size
    • 6x9 cm size

    Kerecis® Tendon Protect is a standalone, single use medical device made from fish skin that is technologically similar to the predicate device Tendon Wrap Tendon Protector (K053655) (Substantial Equivalence) and identical to the Kerecis® Reconstruct™ (K202430).

    AI/ML Overview

    The provided text is a 510(k) clearance letter and summary for a medical device called "Tendon Protect." It details the device's characteristics, indications for use, and a summary of the non-clinical performance testing conducted to demonstrate its safety and effectiveness and substantial equivalence to predicate devices. However, it does not contain the specific detailed information typically found in acceptance criteria tables or a comprehensive study report, particularly regarding a multi-reader multi-case (MRMC) study, standalone algorithm performance, or the detailed methodology for ground truth establishment that would be associated with AI/ML-based medical devices.

    The document describes a physical medical device (surgical mesh made from fish skin) rather than an AI/ML-based diagnostic or assistive software. Therefore, many of the questions related to AI/ML specific criteria (such as sample size for training set, number of experts for ground truth, adjudication methods, MRMC studies, and standalone algorithm performance) are not applicable to this particular device submission.

    Here's a breakdown of the applicable information found in the document:

    1. A table of acceptance criteria and the reported device performance:

    The document mentions that the device "met all the pre-defined acceptance criteria" for various tests, but it does not provide a table detailing the specific numerical or quantitative acceptance criteria and the corresponding reported performance values. It only states that the tests were "conducted and passed successfully according to the test specifications."

    Test CategoryTest NameAcceptance CriteriaReported Device Performance
    BiocompatibilityCytotoxicity (ISO 10993-5)(Not specified, but implied "pass")Passed
    Sensitization (10993-10)(Not specified, but implied "pass")Passed
    Irritation or intracutaneous reactivity (ISO 10993-10)(Not specified, but implied "pass")Passed
    Material mediated pyrogenicity (ISO 10993-11)(Not specified, but implied "pass")Passed
    Implantation Effects (ISO 10993-6)(Not specified, but implied "pass")Passed
    Genotoxicity (ISO 10993-3)(Not specified, but implied "pass")Passed
    Acute systemic toxicity (ISO 10993-11)(Not specified, but implied "pass")Passed
    Subacute/subchronic toxicity (ISO 10993-11)(Not specified, but implied "pass")Passed
    Bench TestingTendon Summative Usability Evaluation (ISO 62366-2)Pre-defined acceptance criteria for safety and effectivenessMet all pre-defined acceptance criteria
    Ball burst testing(Not specified, but implied "met acceptance criteria")Met acceptance criteria
    Suture retention testing(Not specified, but implied "met acceptance criteria")Met acceptance criteria
    Tensile strength testing(Not specified, but implied "met acceptance criteria")Met acceptance criteria
    Animal Testing (GLP Chicken Study)Histopathology evaluation(Not specified, but implied "met acceptance criteria")Acceptance criteria met; exhibited similar overall performance to predicate
    Macroscopic evaluations(Not specified, but implied "met acceptance criteria")Acceptance criteria met; exhibited similar overall performance to predicate
    Tendon biomechanical function assessment(Not specified, but implied "met acceptance criteria")Acceptance criteria met; exhibited similar overall performance to predicate
    Overall animal health(Not specified, but implied "met acceptance criteria")Acceptance criteria met; exhibited similar overall performance to predicate

    2. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated for most tests (e.g., biocompatibility bench tests). For the Tendon GLP Chicken Study, it is a chronic model evaluating early, mid-term, and late time points, comparing the subject device to a control and a negative control. The specific number of animals or tendons tested is not provided.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be pre-clinical (bench and animal testing), not human clinical data, and are likely prospective given they are part of a regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable as the device is a physical surgical mesh, not a diagnostic or AI-based device requiring expert interpretation of images or signals for ground truth. Ground truth for the animal study would be based on objective scientific measurements and observations (e.g., histology, biomechanics).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This is not applicable for the reasons stated above.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable as the device is a physical surgical mesh, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable as the device is a physical surgical mesh, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the animal study:

    • Histopathology evaluation
    • Macroscopic evaluations
    • Tendon biomechanical function assessment
    • Overall animal health

    For bench testing, ground truth would be established by the physical and mechanical properties of the material itself, measured against predefined specifications.

    8. The sample size for the training set:

    This is not applicable as the device is a physical surgical mesh, not an AI/ML system requiring a training set.

    9. How the ground truth for the training set was established:

    This is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250598
    Date Cleared
    2025-06-03

    (95 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    **
    Trade/Device Name: Endoform Reconstructive Template - PLGA
    Regulation Number: 21 CFR 878.3300
    Reconstructive Template - PLGA |
    | Common name | Surgical Mesh |
    | Regulation Number | 21 CFR 878.3300
    Reconstructive Template - PLGA |
    | Common name | Surgical Mesh |
    | Regulation Number | 21 CFR 878.3300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endoform™ Reconstructive Template – PLGA is intended for use as a surgical mesh to reinforce and/or repair soft tissue where weakness exists. Indications for use include the repair of hernias and/or abdominal wall defects that require the use of reinforcing material to obtain the desired surgical outcome.

    Device Description

    Endoform™ Reconstructive Template – PLGA is a surgical mesh comprised of multiple layers of ovine-derived extracellular matrix reinforced with poly(lactic-co-Glycolic) acid (PLGA). The device design includes a range of shapes, sizes, and thicknesses in surface areas up to 400cm², to give a range of strengths as required for a particular implant procedure. Devices are terminally sterilized by ethylene oxide (EO) sterilization.

    AI/ML Overview

    This FDA 510(k) clearance letter and supporting summary for the Endoform™ Reconstructive Template – PLGA device does not involve artificial intelligence (AI) or machine learning (ML). Therefore, many of the requested elements for AI/ML device validation are not applicable.

    The device is a traditional medical device (surgical mesh) and its substantial equivalence is demonstrated through non-clinical performance data and comparison to a predicate device. There is no mention of an algorithm, AI assistance, or human readers.

    However, I can extract the information relevant to its acceptance criteria and the study that proves it meets those criteria, based on the provided document.


    Acceptance Criteria and Study Proof for Endoform™ Reconstructive Template – PLGA

    Since this is a traditional medical device (surgical mesh) and not an AI/ML device, the concept of "acceptance criteria" is related to meeting performance specifications and demonstrating biocompatibility and safety comparable to a predicate device, rather than diagnostic accuracy metrics. The "study" refers to non-clinical testing.

    Here's a breakdown based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes the types of tests conducted and the conclusion that the device meets specifications, rather than providing specific numerical acceptance criteria and results in a table format common for AI/ML.

    For this device, the "acceptance criteria" are implied by the performance characteristics demonstrated to be "substantially equivalent" to the predicate device and meeting recognized standards.

    Acceptance Criteria CategoryDescription (Implied/Direct)Reported Device Performance (Summary)
    Material CompositionConsistency with ovine-derived extracellular matrix with PLGA reinforcement; comparable to predicate's ovine-derived ECM with PGA.Comprised of multiple layers of ovine-derived extracellular matrix reinforced with poly(lactic-co-Glycolic) acid (PLGA). "Maintains the same fundamental technological characteristics as the predicate device with respect to material types."
    Physical DimensionsRange of shapes, sizes, and thicknesses for surface areas up to 400cm².Designs include a range of shapes, sizes, and thicknesses in surface areas up to 400cm². "The subject device is offered in the same range of shapes, sizes, and thicknesses as the predicate device."
    Sterilization Method & SALEthylene Oxide sterilization with a Sterility Assurance Level (SAL) of 10⁻⁶."Devices are terminally sterilized by ethylene oxide (EO) sterilization." SAL of 10⁻⁶ achieved.
    Endotoxin ContentEndotoxin content less than 20 EU/device.Endotoxin Content:
    Ask a Question

    Ask a specific question about this device

    Page 1 of 72