K Number
K063712
Manufacturer
Date Cleared
2007-03-12

(88 days)

Product Code
Regulation Number
884.5980
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Avaulta™ Support System is indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended either as mechanical support or bridging material for the fascial defect.

Device Description

The Avaulta™ Support System includes a sterile, single use, permanent implant that provides long term reinforcement to support structures in the correction of anterior or posterior vaginal wall prolapse. The central soft knit section provides compliant organ support while the strong knit arms provide improved strength for tension free fixation of the implant.

The Avaulta™ Plus Biosynthetic Support System and Avaulta™ Solo support system both utilize a nonabsorbable monofilament, polypropylene mesh to provide long-term reinforcement for support structures. The Avaulta™ Plus Biosynthetic Support System adds a porous, acellular, ultrathin sheet of crosslinked collagen attached to the polypropylene mesh which serves to establish a protective barrier between mucosal tissue and the polypropylene mesh and contains apertures uniformly sized to allow for ingrowth of host tissue and capillary vessels.

AI/ML Overview

This 510(k) premarket notification describes a surgical mesh device, the Avaulta™ Solo Support System and Avaulta™ Plus Biosynthetic Support System. The notification focuses on establishing substantial equivalence to predicate devices and provides limited information regarding detailed acceptance criteria and a specific study proving the device meets those criteria.

Based on the provided document, here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of specific acceptance criteria (e.g., tensile strength, pore size, biocompatibility metrics with defined thresholds) and corresponding reported performance metrics for the Avaulta™ Support System.

Instead, the document states:

  • "The appropriate testing to determine substantial equivalence was completed."
  • "This includes testing in accordance with Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh (March 22, 1999)."

This implies that the acceptance criteria were based on the guidance document for surgical mesh and that the device performance met the requirements outlined in that guidance, thereby demonstrating substantial equivalence to the predicate devices. The specific numerical values for these criteria are not provided in this summary.

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the document. The 510(k) summary does not detail specific sample sizes for particular tests or the provenance of any data (e.g., country of origin, retrospective/prospective). It generally refers to "appropriate testing."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not applicable and therefore not provided in the context of this device. The Avaulta™ Support System is a physical surgical mesh, not an AI or diagnostic device that requires expert-established ground truth for its performance evaluation in the way a medical imaging AI would. Its performance is evaluated through material testing and clinical equivalence to predicate devices.

4. Adjudication Method for the Test Set

This information is not applicable and therefore not provided. Adjudication methods like 2+1 or 3+1 are used in studies involving human readers and AI for diagnostic tasks, not for evaluating a physical medical device like surgical mesh.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

This information is not applicable and therefore not provided. An MRMC study is relevant for evaluating the impact of AI on human reader performance in diagnostic tasks. The Avaulta™ Support System is a surgical implant, not a diagnostic tool assisted by AI, so such a study would not be performed for this device.

6. If a Standalone Study (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

This information is not applicable and therefore not provided. This concept is relevant to AI algorithms. The Avaulta™ Support System is a physical medical device.

7. The Type of Ground Truth Used

The concept of "ground truth" in the way it's used for AI or diagnostic studies (e.g., pathology, outcomes data) is not directly applicable to the primary performance evaluation documented here for a surgical mesh.

Instead, the "truth" for this device's performance evaluation would be established through:

  • Material properties testing: Adherence to established mechanical, physical, and chemical standards (e.g., tensile strength, burst strength, pore size, biocompatibility using ISO standards).
  • Clinical equivalence: Demonstrating that the device performs as intended for tissue reinforcement in vaginal wall prolapse, comparable to the predicate devices, which have a known safety and effectiveness profile. This "ground truth" would be established through clinical experience with the predicate devices and the understanding that the new device's materials and design are substantially equivalent.

8. The Sample Size for the Training Set

This information is not applicable and therefore not provided. "Training set" is a concept used in machine learning for AI model development. This device is a physical product, not an AI algorithm.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable and therefore not provided. As explained above, the concept of a training set and its ground truth does not apply to the development and evaluation of this surgical mesh.

Summary of Device Performance and Study Information (Based on the document):

The document functions as a 510(k) summary, aiming to demonstrate substantial equivalence to legally marketed predicate devices. The "study" mentioned is the "appropriate testing to determine substantial equivalence" which "includes testing in accordance with Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh (March 22, 1999)." This implies that standard material science, biocompatibility, and potentially preclinical testing (as dictated by the guidance) were conducted, and the results demonstrated that the Avaulta™ Support System met the requirements of the guidance and was comparable to the predicate devices. Specific quantitative data, sample sizes for these tests, or detailed expert reviews (beyond regulatory review) are not typically presented in this condensed form of a 510(k) summary.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the number 2063712 in the upper left corner. To the right of the number is the fraction 1/2. Below the fraction is the word "BARD" in all capital letters.

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

  • A. SUBMITTER INFORMATION:
    Submitter's Name:

Address:

C. R. Bard, Inc. Bard Urological Division 13183 Harland Drive Covington, GA 30014

Contact Person: John C. Knorpp Contact Person's Telephone Number: 678-342-4920 Contact Person's Fax: 770-788-5513

  • B. DEVICE NAME:
Trade Name(s):Avaulta ™ Solo Support SystemAvaulta ™ Plus Biosynthetic Support System
Common/Usual Name:Surgical Mesh
Classification Names:79 - Mesh, Surgical, Polymeric (OTP, PAI)
CFR Reference:21 CFR 878.3300
Classification Panel:General and Plastic Surgery

C. PREDICATE DEVICE NAME:

Bard CollaMend™ Implant - K052322 Trade Names: UGYTEX® Dual Knit Mesh – K051503

  • D. Device Description:
    The Avaulta™ Support System includes a sterile, single use, permanent implant that provides long term reinforcement to support structures in the correction of anterior or posterior vaginal wall prolapse. The central soft knit section provides compliant organ support while the strong knit arms provide improved strength for tension free fixation of the implant.

The Avaulta™ Plus Biosynthetic Support System and Avaulta™ Solo support system both utilize a nonabsorbable monofilament, polypropylene mesh to provide long-term reinforcement for support structures. The Avaulta™ Plus Biosynthetic Support System adds a porous, acellular, ultrathin sheet of crosslinked collagen attached to the polypropylene mesh which serves to establish a protective barrier between mucosal tissue and

{1}------------------------------------------------

K0637/D
2/D

the polypropylene mesh and contains apertures uniformly sized to allow for ingrowth of host tissue and capillary vessels.

  • E. Intended Use:
    The Avaulta™ Support System is indicated for tissue reinforcement and longlasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended either as mechanical support or bridging material for the fascial defect.

F. TECHNOLOGICAL CHARACTERISTICS SUMMARY:

The subject Avaulta™ Support System has the same intended use, general design and fundamental scientific technology as the predicate devices.

  • G. Performance Data Summary:
    The appropriate testing to determine substantial equivalence was completed. This includes testing in accordance with Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh (March 22, 1999).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The overall design is simple and professional, reflecting the department's role in public health and welfare.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SEP 2 8 2012

John C. Knorpp, RAC Manager, Regulatory Affairs Bard Urological Division 13183 Harland Drive COVINGTON GA 30014

Re: K063712 Trade/Device Name: Avaulta Solo™ Support System Avaulta Plus™ Biosynthetic Support System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTP, PAI Dated: December 8, 2006 Received: December 14, 2006

Dear Mr. Knorpp:

This letter corrects our substantially equivalent letter of March 12, 2007.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 -

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely yours,

Benjamin R. Krebs

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

C.R. Bard, Inc., Bard Urological Division Avaulta™ Support System Premarket Notification [510(k)]

KO 63712

1.3 Indications for Use Statement

510(k) Number (if known): __

Device Name: _ Avaulta™ Support System

Indications for Use:

The Avaulta™ Support System is indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended either as mechanical support or bridging material for the fascial defect.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

Pelo R

(Division Sign-Off) Division of General, Restorative. and Neurological Devices

11063112 510(k) Number

(Recommended Format 11/13/2003)

§ 884.5980 Surgical mesh for transvaginal pelvic organ prolapse repair.

(a)
Identification. Surgical mesh for transvaginal pelvic organ prolapse repair is a prescription device intended to reinforce soft tissue in the pelvic floor. This device is a porous implant that is made of synthetic material, non-synthetic material, or a combination of synthetic and non-synthetic materials. This device does not include surgical mesh for other intended uses (§ 878.3300 of this chapter).(b)
Classification. Class III (premarket approval).(c)
Date premarket application approval or notice of completion of a product development protocol is required. A premarket application approval or notice of completion of a product development protocol for a device is required to be filed with the Food and Drug Administration on or before July 5, 2018, for any surgical mesh described in paragraph (a) of this section that was in commercial distribution before May 28, 1976, or that has, on or before July 5, 2018, been found substantially equivalent to a surgical mesh described in paragraph (a) of this section that was in commercial distribution before May 28, 1976. Any other surgical mesh for transvaginal pelvic organ prolapse repair shall have an approved premarket application or declared completed product development protocol in effect before being placed in commercial distribution.