K Number
K043388

Validate with FDA (Live)

Date Cleared
2005-08-05

(239 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OrthADAPT® Bioimplant (Surgical Mesh) is intended to be used for implantation to reinforce soft tissue including but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, hernias, suture-line reinforcement, and reconstructive procedures. The device is also intended for the reinforcement of soft tissues repaired by sutures or suture anchors during tendon repair surgery including reinforcement of rotator cuff, patellar. Achilles, biceps, quadriceps, or other tendons.

OrthADAPT® is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps or other tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide biomechanical strength for the tendon repair.

OrthADAPT® is intended for one-time use only.

Device Description

The OrthADAPT Bioimplant is a decellularized equine pericardium. The OrthADAPT Bioimplant has been crosslinked and exposed to a liquid chemical sterillant. The product has passed the USP sterility test and satisfies FDA requirements for LAL endotoxin limit for a medical device. The product must be rinsed prior to use.

AI/ML Overview

The provided text is a 510(k) summary for the OrthADAPT® Bioimplant, a surgical mesh. It describes the device, its intended use, and its substantial equivalence to predicate devices based on biocompatibility and performance testing. However, it does not contain the detailed acceptance criteria for a study, nor does it describe a study design that would directly assess and prove the device meets specific performance criteria.

The 510(k) summary states: "The results of biocompatibility and performance testing demonstrate that OrthADAPT is comparable to the predicate devices." This implies that some testing was done to establish substantial equivalence, but the specifics of that testing, including acceptance criteria, sample sizes, ground truth establishment, or any AI-related components, are not provided in this document.

Therefore, I cannot fulfill your request for:

  1. A table of acceptance criteria and the reported device performance: The document does not list specific acceptance criteria or quantitative performance metrics for the OrthADAPT® Bioimplant.
  2. Sample sizes used for the test set and the data provenance: No sample sizes for a test set are mentioned.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not present.
  4. Adjudication method: Not applicable as no such study is described.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: The document does not mention any MRMC studies, nor does it involve AI assistance for human readers.
  6. Standalone (i.e. algorithm only without human-in-the-loop performance) study: The device is a physical surgical mesh, not an algorithm. Therefore, a standalone algorithm performance study is not relevant here.
  7. The type of ground truth used: Not specified for any performance testing described.
  8. The sample size for the training set: Not applicable as this is not an AI/algorithm-based device.
  9. How the ground truth for the training set was established: Not applicable.

Summary:

The provided document describes a medical device (surgical mesh) and its regulatory clearance based on substantial equivalence. It does not provide the information typically associated with the performance evaluation of a diagnostic or AI-driven device, such as specific acceptance criteria and detailed study results. The "performance testing" mentioned is in the context of demonstrating comparability to existing predicate devices, rather than meeting predefined, quantitative acceptance criteria for a novel performance claim beyond what is typical for a surgical mesh.

{0}------------------------------------------------

K043388

Synovis Orthopedic and Woundcare, Inc.

OrthADAPT® F

510(k) Summary

Submitter Information 1

A.Company Name:Synovis Orthopedic and Woundcare, Inc.
B.Company Address:6 Jenner, Suite 150Irvine, CA 92618
C.Company Phone:(949) 502-3240
D.Company Facsimile:(949) 502-3241
E.Contact Person:Amy BouclyManager, Regulatory Affairs/Quality Assurance
F.Date:06/25/12

Device Identification 2

A.Device Trade Name:OrthADAPT® Bioimplant
B.Common Name:Surgical Mesh
C.Classification Name(s):Surgical Mesh
D.Classification Regulation:21 CFR 878.3300
E.Device Class:Class II
F.Device Code(s):OWY, FTM
G.Advisory Panel:General and Plastic Surgery

Identification of Predicate Devices 3

The OrthADAPT Bioimplant is substantially equivalent to the Fortaflex Surgical Mesh manufactured by Organogenesis, Inc. (K020049), and Depuy Restore Orthobiologic Soft Tissue Implant, manufactured by Depuy Orthopedic, Inc. (K001738).

Page 1 of 2

{1}------------------------------------------------

Synovis Orthopedic and Woundcare, Inc.

K043388 page
2/2
OrthADAPTO® FX, PX, MX

র্ব Device Description

The OrthADAPT Bioimplant is a decellularized equine pericardium. The OrthADAPT Bioimplant has been crosslinked and exposed to a liquid chemical sterillant. The product has passed the USP sterility test and satisfies FDA requirements for LAL endotoxin limit for a medical device. The product must be rinsed prior to use.

5 Indications for Use

The OrthADAPT® Bioimplant (Surgical Mesh) is intended to be used for implantation to reinforce soft tissue including but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, hernias, suture-line reinforcement, and reconstructive procedures. The device is also intended for the reinforcement of soft tissues repaired by sutures or suture anchors during tendon repair including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons.

OrthADAPT® is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps or other tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide biomechanical strength for the tendon repair.

OrthADAPT® is intended for one-time use only.

Substantial Equivalence ਉ

The OrthADAPT is comparable to the predicate devices in terms of intended use, technology and performance: It is similar to the predicate devices in that it has the same intended use of soft tissue reinforcement, is a collagen mesh and is similar in design and materials. The results of biocompatibility and performance testing demonstrate that OrthADAPT is comparable to the predicate devices.

Page 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in blue, with three overlapping profiles suggesting community and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure, emphasizing the department's role and national scope.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Pegasus Biologics, Incorporated % Mr. Muir S. Meinhold Director. Regulatory Affairs. Quality Assurance 10 Pasteur, Suite 150 Irvine, California 92618

AUG 2 7 2012

Re: K043388

Trade/Device Name: Pegasus Biologics OrthADAPT™ FX,PX.MX Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTM. OXB. OXE. OWY Dated: June 17, 2005 Received: June 20, 2005

Dear Mr. Meinhold:

This letter corrects our substantially equivalent letter of August 5, 2005.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm,

Sincerely yours.

Mark Melkerson Director

Division of Surgical. Orthopedics and Restorative Devices Office of Device Evaluation Center for Device and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K043388

Device Name:

OrthADAPT® FX, PX, MX

Indications For Use:

The OrthADAPT® Bioimplant (Surgical Mesh) is intended to be used for implantation to reinforce soft tissue including but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, hernias, suture-line reinforcement, and reconstructive procedures. The device is also intended for the reinforcement of soft tissues repaired by sutures or suture anchors during tendon repair surgery including reinforcement of rotator cuff, patellar. Achilles, biceps, quadriceps, or other tendons.

OrthADAPT® is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps or other tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide biomechanical strength for the tendon repair.

OrthADAPT® is intended for one-time use only.

Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Keane for MM
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K043388

Page 1 of 1

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.