K Number
K250081

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-04-11

(88 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant.
The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant.
The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant.
The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant.

Device Description

Atlantis® Abutment in Titanium refers to the following abutments that are the subject of this 510(k):

  • Atlantis® Abutment,
  • Atlantis® Crown Abutment,
  • Atlantis® Conus Abutment, and
  • Atlantis® Healing Abutment.

The Atlantis® Abutments in Titanium are patient-specific dental abutments that are intended for attachment to dental implants in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function. The design of the Atlantis® Abutments in Titanium is derived from patient dental models and is completed by Dentsply Sirona using computer-assisted design (CAD) technology according to the clinician's prescription. The final CAD design of the Atlantis® Abutment in Titanium is fabricated using computer-assisted manufacturing (CAM) to produce a customized, patient-specific device. Design and fabrication of the Atlantis® Abutment in Titanium is completed in internal Dentsply Sirona manufacturing facilities.

Alternatively, the CAD design, according to the clinician's prescription, can be performed by a laboratory or clinician in an FDA cleared abutment design software (3Shape Abutment Designer Software, K151455) within the design envelope of the Atlantis® Abutments which is codified in the validated and locked design library of the cleared software. Fabrication of the Atlantis® Abutment is then completed in internal Dentsply Sirona manufacturing facilities.

The Atlantis® Abutment in Titanium serves as a connection of the prosthetic construction and the endosseous implant. The lower part of the abutment is designed to fit with the specific implant geometry it is compatible, and the upper part design is according to the patient's specific anatomy. The Atlantis® Abutment in Titanium, including the Conus abutments and Healing abutments, are available in Titanium or Gold-shaded Titanium (titanium nitride layer applied using PVD (Physical Vapor Deposition)). The Crown Abutment is only available in Titanium.

The Atlantis® Abutment design envelope became the basis for the other more specific designs that make up the Atlantis® Abutments in Titanium. The Atlantis® Abutment is intended for use with an endosseous implant and for single tooth restoration.

The Atlantis® Crown Abutment in Titanium incorporates a design that is a combination of an abutment and an anatomically accurate crown to constitute the final finished device. It functions as a substructure that also serves as the final abutment, in a partially or completely edentulous patient.

The Atlantis® Conus abutment supports a removable prosthesis (bridges and overdentures) which is retained by friction fit to the abutment. The abutment connects to the prosthesis via caps embedded in the prosthesis.

The Atlantis® Healing Abutment is used with the compatible implants for temporary use during soft tissue healing after one-stage and two-stage surgeries. It is designed based on the planned final Atlantis® Abutment or Atlantis® Crown Abutment, using the same emergence profile as those abutments to achieve an aesthetic outcome during the soft tissue healing phase.

Unless otherwise noted in this summary, the design envelope has remained unchanged from the previously cleared systems. Parameters such as gingival height and abutment diameter are dependent on the different implant system compatibilities.

AI/ML Overview

The provided document is an FDA 510(k) clearance letter for Atlantis® Abutments in Titanium. This document indicates that the device has received clearance and is considered substantially equivalent to previously marketed devices. However, it is not a clinical study report and therefore does not contain the detailed information typically found in a study that proves a device meets acceptance criteria derived from a clinical trial.

Specifically, the document primarily focuses on non-clinical testing (e.g., MRI compatibility, biocompatibility, reprocessing, and CAD/CAM software validation) to support a labeling change and the substantial equivalence to predicate devices. It explicitly states:

  • "8. Clinical Tests Summary and Conclusion: Not applicable. There are no clinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence."

Therefore, I cannot extract the requested information (Acceptance Criteria Table, Sample Size, Data Provenance, Ground Truth Experts, Adjudication Method, MRMC Study, Standalone Performance, Ground Truth Type, Training Set Size, Training Set Ground Truth Establishment) from this document as it pertains to a clinical study proving device performance against acceptance criteria.

The information from the document that is relevant to non-clinical testing and device characteristics is provided below.


Non-Clinical Testing and Device Characteristics from the 510(k) Clearance Letter

While a clinical study proving performance against acceptance criteria is not presented, the document does describe the non-clinical tests performed to support the device's safety and effectiveness and its substantial equivalence to predicate devices.

1. A table of acceptance criteria and the reported device performance:

Since this is a 510(k) clearance notice based on substantial equivalence and non-clinical testing, there isn't a direct "acceptance criteria table" in the traditional sense of a clinical trial's performance endpoints. However, the document outlines standards and guidance documents used as criteria for the non-clinical tests, and the stated conclusion is that the device meets these requirements.

Test CategoryAcceptance Criteria (Standards/Guidance Fulfilled)Reported Device Performance (Conclusion stated)
MR Safety- ASTM F2052-21: Measurement of magnetically induced displacement force- ASTM F2213-17: Measurement of magnetically induced torque- ASTM F2119-07 (2013): Evaluation of MR image artifacts from passive implants- ASTM F2503-20: Marking medical devices for MR safety- FDA guidance: "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" (October 10, 2023)MR Conditional. Testing included in reference device (K221094) supports labeling update.The proposed Atlantis® Abutments in Titanium meet the requirements.
Biocompatibility- ISO 10993-1:2018: Biological evaluation of medical devices – Part 1- ISO 10993-5:2009: Cytotoxicity testingBiocompatibility assessment was performed and no new questions regarding biocompatibility are raised. Cytotoxicity testing was performed.
Reprocessing (Sterilization Drying Time)- ANSI/AAMI/ISO 17665-1:2006/(R)2013, Annex D- ANSI/AAMI/ISO 14937:2009/(R)2013, Annex D (Approach 3)- AAMI TIR12:2020- ANSI/AAMI/ISO TIR17665-2:2009- FDA guidance: "Reprocessing Medical Device in Health Care Settings: Validation Methods and Labeling"Reprocessing testing was performed to validate the proposed drying time for steam sterilization. The sterilization drying time validation followed the FDA guidance and relevant standards.
CAD/CAM Software Validation & Verification (Atlantis® Abutment Milling)To ensure that the same design limitations/constraints present in the 3Shape Atlantis® Abutment library match those in the internal CAD system, and that the design parameters cannot be exceeded.Software validation was performed and confirmed that the additional option of using the 3Shape design module for design of Atlantis® Abutments does not raise new questions of safety and performance as it was confirmed that the locked design library has the same design limits as the internal VAD software.
Mechanical Performance (Fatigue Testing)- ISO 14801:2016: Implants – Dynamic loading test for endosseous dental implants- FDA Guidance: Root-form Endosseous Dental Implants and Endosseous Dental Abutments – Class II Special Controls Guidance Document.The internally documented and proposed modifications did not affect the performance of the devices and no new fatigue testing was needed. The abutments continue to meet the requirements.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Non-Clinical Tests: Not explicitly stated as numerical sample sizes (e.g., 'n=X implants'). The testing refers to meeting standard requirements (e.g., ASTM, ISO guidelines), which imply specific numbers of test units, but these are not provided in this summary document.
  • Data Provenance: The testing was conducted by Dentsply Sirona (the manufacturer) and references previous FDA clearances (K numbers) and established standards. The data is part of the 510(k) submission to the FDA. The provenance of specific raw data (country of origin, retrospective/prospective) is not detailed in this summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not applicable. This document does not describe a process for establishing ground truth using human experts for the purpose of validating an AI/software device's diagnostic performance against human readers. The validation activity for the software (Atlantis® Abutment Milling) was to ensure its design limitations mirrored the internal CAD system and did not fall outside acceptable parameters.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. This is not a clinical study involving human interpretation or adjudication of findings.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. The document explicitly states: "Clinical Tests Summary and Conclusion: Not applicable. There are no clinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence." Therefore, no MRMC study was conducted or reported.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • The "software validation" for "Atlantis® Abutment Milling" is a form of standalone performance evaluation for the CAD capabilities. It confirms that the software's design outputs (e.g., abutment geometry) from the 3Shape Abutment Designer Software conform to the established design envelope and limitations of Dentsply Sirona's internal CAD system (Virtual Abutment Design - VAD). This is a technical validation of the algorithm's design capability, not a diagnostic performance study.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the MR Safety testing, the ground truth or "reference standard" is the physical behavior of the device under MR conditions, measured against established ASTM standards and FDA guidance.
  • For Biocompatibility, the ground truth is the biological response, evaluated against ISO standards.
  • For Reprocessing, the ground truth is the effectiveness of sterilization and drying, validated against AAMI/ISO standards and FDA guidance.
  • For CAD/CAM Software Validation, the "ground truth" for the 3Shape software is the "validated and locked design library" and "design limitations" of Dentsply Sirona's internal VAD software, ensuring the external software produces designs within the company's approved specifications.
  • For Mechanical Performance (Fatigue Testing), the ground truth is the structural integrity and durability under dynamic loading, evaluated against ISO standards and FDA guidance.

8. The sample size for the training set:

  • Not applicable. This is not an AI/ML device that requires a "training set" in the context of supervised learning for diagnostic tasks. The software referred to (3Shape Abutment Designer Software) is a CAD software for design, not a machine learning model for prediction or classification.

9. How the ground truth for the training set was established:

  • Not applicable, as there is no "training set" in the machine learning sense described in this document. The "ground truth" for the CAD software's validation was its adherence to the manufacturer's established design rules and limitations.

FDA 510(k) Clearance Letter - Atlantis® Abutments in Titanium

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

Dentsply Sirona
Laura Sorbin
Sr. Technical RA Manager
221 West Philadelphia Street Suite 60W
York, Pennsylvania 17401

Re: K250081
Trade/Device Name: Atlantis® Abutments in Titanium
Regulation Number: 21 CFR 872.3630
Regulation Name: Endosseous Dental Implant Abutment
Regulatory Class: Class II
Product Code: NHA
Dated: January 8, 2025
Received: January 13, 2025

Dear Laura Sorbin:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

April 11, 2025

Page 2

K250081 - Laura Sorbin Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K250081 - Laura Sorbin Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Andrew I. Steen -S

Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (6/20) Page 1 of 2

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K250081

Device Name: Atlantis® Abutments in Titanium

Indications for Use (Describe)

Atlantis Abutment & Atlantis Abutment Milling

The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant.

Atlantis Crown Abutment

The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant.

Atlantis Conus Abutment

The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant.

Atlantis Healing Abutment

The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant.

Implant Manufacturer: Dentsply Sirona
Implant System: OsseoSpeed/OsseoSpeed TX
Implant Diameter (in mm): 3.0 (Yellow), 3.5, 4.0 (Aqua), 4.5, 5.0 (Lilac), 4.0 (Blue - Previous)
Abutment Platform Diameter (in mm): 2.330 (Yellow), 2.866 (Aqua), 3.865 (Lilac), 3.368 (Blue - Previous)

Implant Manufacturer: MIS Implant Technologies Inc.
Implant System: V3 (NP, SP), C1 (NP, SP, WP)
Implant Diameter (in mm): V3: 3.30 (NP), 3.90, 4.3, 5.0 (SP); C1: 3.30 (NP), 3.75, 4.2 (SP), 5.0 (WP)
Abutment Platform Diameter (in mm): V3: 2.765 (NP), 3.16 (SP); C1: 2.76 (NP), 3.16 (SP), 4.01 (WP)

Implant Manufacturer: Altatec GmbH
Implant System: Conelog Screw-Line Implant
Implant Diameter (in mm): 3.3, 3.8, 4.3, 5.0
Abutment Platform Diameter (in mm): 2.80 (3.3), 3.00 (3.8, 4.3), 3.80 (5.0)

Implant Manufacturer: Nobel Biocare
Implant System: NobelActive
Implant Diameter (in mm): 3.0
Abutment Platform Diameter (in mm): 2.461

Implant System: Branemark System (NP, RP, WP)
Implant Diameter (in mm): 3.3 (NP), 3.75, 4.0 (RP), 5.0, 5.5 (WP)

Page 5

FORM FDA 3881 (6/20) Page 2 of 2

Abutment Platform Diameter (in mm): 3.476 (NP), 4.069 (RP), 5.033 (WP)

Implant Manufacturer: BioHorizons
Implant System: Prodigy System Internal/Tapered Internal
Implant Diameter (in mm): 3.5, 4.5, 5.7
Abutment Platform Diameter (in mm): 3.434 (3.5), 4.399 (4.5), 5.644 (5.7)

Implant System: Maestro
Implant Diameter (in mm): 3.5, 4.0, 5.0
Abutment Platform Diameter (in mm): 3.708 (3.5), 4.242 (4.0), 5.258 (5.0)

Implant Manufacturer: Biomet 3i
Implant System: Certain, Certain Prevail, Certain XP
Implant Diameter (in mm): Certain: 3.25, 4, 5, 6; Certain Prevail: 3.4, 4, 5; Certain XP: 5, 6
Abutment Platform Diameter (in mm): 3.404 (3.25, 3.4), 4.012 (4), 4.942 (5), 5.951 (6)

Implant System: Osseotite Tapered and Osseotite Parallel Walled
Implant Diameter (in mm): 3.25, 3.75, 4.0, 5.0, 6.0
Abutment Platform Diameter (in mm): 4.056 (3.25, 3.75, 4.0), 4.978 (5.0), 5.985 (6.0)

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 6

510(k) SUMMARY
for
Atlantis® Abutments in Titanium (K250081)

1. Submitter Information:

Dentsply Sirona
221 West Philadelphia Street
Suite 60W
York, PA 17401, USA

Contact Person: Laura Sobrin
Telephone Number: 717-849-4434
Email: laura.sobrin@dentsplysirona.com
Date Prepared: April 11, 2025

2. Device Name:

  • Proprietary Name: Atlantis® Abutments in Titanium
  • Classification Name: Abutment, Implant, Endosseous, Root-Form
  • CFR Number: 872.3630
  • Device Class: Class II
  • Primary Product Code: NHA

3. Predicate/Reference Devices:

The primary purpose of the Bundled Traditional 510(k) premarket notification is to modify the labeling of the Atlantis® Abutments in Titanium to update the safety in an MR environment to indicate MR Conditional. This includes a total of 11 implant system compatibilities/clearances.

The predicate device, K193529, which is an Atlantis® Abutment specifically compatible with the MIS Conical Connection implant system, represents the most recently cleared implant system compatibility for the Atlantis® Abutments in Titanium.

The 510(k) clearances for the additional implant system compatibilities bundled in this submission are referred to as reference devices. Therefore, the proposed devices are modifications to the predicate device, K193529, and modification to the additional 10 implant system compatibilities/clearances, reference devices, listed below.

Page 1 of 50

Page 7

Page 2 of 50

Predicate Device Name – Implant system compatibility included in this submission for MR Conditional labeling update510(k)Product Code and Classification RegCompany Name
Atlantis® Abutment for MIS Conical Connection ImplantsK193529NHA, Abutment, Implant, Dental, Endosseous 872.3630, Endosseous dental implant abutmentDentsply Sirona Inc.
Reference Device Name – Additional implant system compatibilities bundled in this submission for MR Conditional labeling update510(k)Product Code and Classification RegCompany Name
Atlantis® Abutment for Conelog ImplantK161030NHA, 872.3630Dentsply Sirona Inc.
Atlantis® Abutment for NobelActive 3.0K151039NHA, 872.3630Dentsply Sirona Inc.
Atlantis® Abutments for Astra Osseospeed 3.0 Implant System (currently marketed as OsseoSpeed/OsseoSpeed TX)K081666NHA, 872.3630Astra Tech Inc., now Dentsply Sirona Inc.
Atlantis® Abutment for BioHorizons ImplantK073258NHA, 872.3630Atlantis Components Inc., now Dentsply Sirona Inc.
Atlantis® Abutment for Astra Implant (currently marketed as OsseoSpeed/OsseoSpeed TX)K070833NHA, 872.3630Atlantis Components Inc., now Dentsply Sirona Inc.
Atlantis® Abutment for 3i Certain MicroMini ImplantK062197NHA, 872.3630Atlantis Components Inc., now Dentsply Sirona Inc.
Atlantis® Abutment for 3i MicroMini Implant (also cleared compatibility with Branemark NP and BioHorizons NP)K062069NHA, 872.3630Atlantis Components Inc., now Dentsply Sirona Inc.
Atlantis® Abutment for Astra Implant (currently marketed as OsseoSpeed/OsseoSpeed TX)K051652NHA, 872.3630Atlantis Components Inc., now Dentsply Sirona Inc.
Atlantis® Abutment for 3i Certain ImplantK051011NHA, 872.3630Atlantis Components Inc., now Dentsply Sirona Inc.
Atlantis® Abutment and Atlantis® Abutment Screw (cleared compatibility with Biomet Osseotite, Branemark RP/WP, and BioHorizons RP/WP)K981858DZE, Implant, endosseous, root-form* 872.3640, Endosseous dental implantAtlantis Components Inc., now Dentsply Sirona Inc.

*This is an older 510(k) where the Atlantis® Abutment was cleared under product code DZE, Implant, Endosseous, Root-form

In addition to the above reference devices, the below reference devices are added to close the gap in technological characteristic differences between the predicate device and the proposed modified Atlantis® Abutment in Titanium implant system compatibilities.

Page 8

Page 3 of 50

Reference Device Name510(k)Product Code & Classification RegCompany NameReason for inclusion
OmniTaper EV Dental Implants; DS Implants with EV connection DS Implants abutments with EV ConnectionK221094DZE, 872.3640 NHA, 872.3630Dentsply Sirona Inc.Reference to MR testing on worst-case implants/abutments devices Reference to same reprocessing instructions (reference device TiDesign EV)
Atlantis® SuprastructuresK151439NHA, 872.3630Dentsply Sirona Inc.Similar Angulated Screw Access (ASA) feature
3Shape Abutment Designer SoftwareK151455PNP, 872.36303Shape A/SThe 3Shape software is part of the additional proposed workflow that allows the user to design the Atlantis Abutment within the Atlantis design limitations prior to manufacturing by Dentsply Sirona.

4. Description of Device:

Atlantis® Abutment in Titanium refers to the following abutments that are the subject of this 510(k):

  • Atlantis® Abutment,
  • Atlantis® Crown Abutment,
  • Atlantis® Conus Abutment, and
  • Atlantis® Healing Abutment.

The Atlantis® Abutments in Titanium are patient-specific dental abutments that are intended for attachment to dental implants in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function. The design of the Atlantis® Abutments in Titanium is derived from patient dental models and is completed by Dentsply Sirona using computer-assisted design (CAD) technology according to the clinician's prescription. The final CAD design of the Atlantis® Abutment in Titanium is fabricated using computer-assisted manufacturing (CAM) to produce a customized, patient-specific device. Design and fabrication of the Atlantis® Abutment in Titanium is completed in internal Dentsply Sirona manufacturing facilities.

Alternatively, the CAD design, according to the clinician's prescription, can be performed by a laboratory or clinician in an FDA cleared abutment design software (3Shape Abutment Designer Software, K151455) within the design envelope of the Atlantis® Abutments which is codified in the validated and locked design library of the cleared software. Fabrication of the Atlantis® Abutment is then completed in internal Dentsply Sirona manufacturing facilities.

The Atlantis® Abutment in Titanium serves as a connection of the prosthetic construction and the endosseous implant. The lower part of the abutment is designed to fit with the specific implant geometry it is compatible, and the upper part design is according to the patient's specific anatomy. The Atlantis® Abutment in Titanium, including the Conus abutments and Healing abutments, are available in Titanium or Gold-shaded Titanium (titanium nitride layer applied using PVD (Physical Vapor Deposition)). The Crown Abutment is only available in Titanium.

Page 3 of 50

Page 9

Page 4 of 50

The Atlantis® Abutment design envelope became the basis for the other more specific designs that make up the Atlantis® Abutments in Titanium. The Atlantis® Abutment is intended for use with an endosseous implant and for single tooth restoration.

The Atlantis® Crown Abutment in Titanium incorporates a design that is a combination of an abutment and an anatomically accurate crown to constitute the final finished device. It functions as a substructure that also serves as the final abutment, in a partially or completely edentulous patient.

The Atlantis® Conus abutment supports a removable prosthesis (bridges and overdentures) which is retained by friction fit to the abutment. The abutment connects to the prosthesis via caps embedded in the prosthesis.

The Atlantis® Healing Abutment is used with the compatible implants for temporary use during soft tissue healing after one-stage and two-stage surgeries. It is designed based on the planned final Atlantis® Abutment or Atlantis® Crown Abutment, using the same emergence profile as those abutments to achieve an aesthetic outcome during the soft tissue healing phase.

Unless otherwise noted in this summary, the design envelope has remained unchanged from the previously cleared systems. Parameters such as gingival height and abutment diameter are dependent on the different implant system compatibilities.

Table 4.1 below includes the main specifications of the Atlantis® Abutments in Titanium that are subject of this submission.

Table 4.1 Specifications of the Atlantis® Abutments in Titanium that are subject of this submission

ItemSpecification
Atlantis® Abutment in Titanium for MIS V3 and C1 implants
Abutment platform diameterØ2.765, 2.76, 3.16, 4.01 mm
Compatible Implant diameterV3: Ø3.3, 3.9, 4.3, 5.0 mm; C1: Ø3.3, 3.75, 4.2, 5.0 mm
Compatible Implant SystemMIS V3 (K163349), MIS C1 (K172505)
ConnectionInternal Conical Connection
Atlantis® Abutment in Titanium for Conelog implants
Abutment platform diameterØ2.80, 3.00, 3.80 mm
Compatible Implant diameterØ 3.3, 3.8, 4.3, 5.0 mm
Compatible Implant SystemConelog Screw-Line Implant 3.3, 3.8/4.3, 5.0 (K113779)
ConnectionInternal Conical Connection
Atlantis® Abutment in Titanium for Nobel Active 3.0 implants
Abutment platform diameterØ2.461 mm
Compatible Implant diameter3.0 mm
Compatible Implant SystemNobel Biocare Active Implant 3.0 (K102436)
ConnectionInternal Conical Connection
Atlantis® Abutment in Titanium for OsseoSpeed 3.0 implants
Abutment platform diameterØ2.330 mm
Compatible Implant DiameterØ 3.0 mm
Compatible Implant SystemOsseoSpeed 3.0 – Yellow, OsseoSpeed TX 3.0 - Yellow (K080396, K101732)
ConnectionInternal Conical Connection
Atlantis® Abutment in Titanium for BioHorizons implants
Abutment platform diameterØ3.434, 4.399, 5.644 mm
Compatible Implant Diameter3.5, 4.5, 5.7 mm
Compatible Implant SystemBioHorizons Internal 3.5, 4.5, 5.7 (K042429)
ConnectionInternal Hex, Tapered Internal
Atlantis® Abutment in Titanium for OsseoSpeed 3.5 and 4.0 mm implants
Abutment platform diameterØ2.866 mm
Compatible Implant Diameter3.5, 4.0 mm

Page 4 of 50

Page 10

Page 5 of 50

ItemSpecification
Compatible Implant SystemOsseoSpeed 3.5/4.0 – Aqua, OsseoSpeed TX 3.5/4.0 - Aqua (K053384, K101732)
ConnectionInternal Conical Connection
Atlantis® Abutment in Titanium for Biomet 3i Certain implants
Abutment platform diameterØ3.404 mm
Compatible Implant Diameter3.25, 3.4 mm
Compatible Implant SystemBiomet 3i Certain 3.25, 4/3, Biomet 3i Certain Prevail 3/4/3 (3.4), 4/3 (K051461)
ConnectionInternal Hexagon Connection
Atlantis® Abutment in Titanium for Biomet, Nobel Biocare, and BioHorizons implants
Abutment platform diameterBiomet: Ø 4.056 mm Nobel: Ø 3.476 mm BioHorizons: 3.708 mm
Compatible Implant DiameterBiomet: 3.25, 3.75, 4.0 mm Nobel: 3.3 mm BioHorizons: 3.5 mm
Compatible Implant System3i Osseotite Tapered and Parallel Walled (K014235, K031475, K041402) Nobel Biocare Branemark System NP 3.3 (K050705) BioHorizons 3.5 (K010458)
ConnectionExternal Hexagon Connection
Atlantis® Abutment in Titanium for Osseospeed/Osseospeed TX implants
Abutment platform diameterØ 3.368, 3.865 mm
Compatible Implant DiameterØ 4.0, 4.5, 5.0 mm
Compatible Implant SystemOsseoSpeed 4.0 Blue (Previous) (K041492) OsseoSpeed/OsseoSpeed TX 4.5, 5.0 – Lilac (K053384, K101732)
ConnectionInternal Conical Connection
Atlantis® Abutment in Titanium for 3i Certain implants
Abutment platform diameterØ 4.012, 4.942, 5.951 mm
Compatible Implant DiameterØ 4.0, 5.0, 6.0 mm
Compatible Implant SystemBiomet 3i Certain 4.0, 5.0, 6.0, 5/4, Biomet 3i Certain Prevail 4/5/4, 5/6/5, 5/4, 6/5, Biomet Certain XP 4/5, 5/6 (K031475)
ConnectionInternal hexagon connection
Atlantis® Abutment in Titanium for Biomet, Nobel Biocare, and BioHorizons implants
Abutment platform diameterBiomet: 4.056, 4.978, 5.985 mm Nobel: 4.069, 5.033 mm BioHorizons: 4.242, 5.258 mm
Compatible Implant DiameterBiomet: 3.75, 4.0, 5.0, 6.0 mm Nobel: 3.75, 4.0, 5.0, 5.5 mm BioHorizons: 4.0, 5.0 mm
Compatible Implant SystemBiomet 3i Osseotite 3.75, 4.0, 5.0, 6.0, Biomet Osseotite XP 4/5, 5/6 (K983347) Branemark RP 3.75 and 4.0, Branemark WP 5.0 and 5.5, (K974828, K022562) BioHorizons External 4.0 and 5.0 (K972313)
ConnectionExternal Hexagon Connection

Page 5 of 50

Page 11

Page 6 of 50

5. Intended Use and Indications for Use:

The proposed devices and the predicate device, Atlantis® Abutment for MIS Conical Connection Implants (K193529), have the same intended use. The proposed devices are intended to be used in the upper or lower jaw for supporting tooth replacements to restore chewing function.

The Indications for Use of the proposed devices are similar to those of the predicate device. Refer to Table 5.1 for a comparison.

Table 5.2 includes additional tables that provide a comparison between the originally cleared Indications for Use and the proposed Indications for Use statement for each implant system compatibility, and a discussion on the differences.

Page 6 of 50

Page 12

Page 7 of 50

Table 5.1: Comparison of the Intended Use/Indications for Use of the Proposed Devices Atlantis® Abutments in Titanium and the Predicate Device (K193529)

ItemProposed Devices Atlantis® Abutments in TitaniumPredicate Device Atlantis® Abutment for MIS Conical Connection Implants (K193529)Discussion
Intended useDental implant abutments are intended to be used in the upper or lower jaw for supporting tooth replacements to restore chewing function.Dental implant abutments are intended to be used in the upper or lower jaw for supporting tooth replacements to restore chewing function.Same
Indications for UseAtlantis Abutment & Atlantis Abutment Milling The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant. Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries.The Atlantis® Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented, or screw retained to the abutment. The abutment screw is intended to secure the Atlantis® Abutment to the endosseous implant. The Atlantis® Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the Atlantis® Crown Abutment to the endosseous implant. The Atlantis® Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment. The abutment screw is intended to secure the Atlantis® Conus Abutment to the endosseous implant. The Atlantis® Healing Abutment can be used with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The abutment screw is intended to secure theAdded Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the 3Shape software (K151455). "The prosthesis can be cemented, or screw retained to the abutment." was removed as a crown cannot be screwed to an abutment. Restorations are always cemented to an abutment. The consolidated Indications for Use include each implant system compatibility information by implant system manufacturer.

Page 7 of 50

Page 13

Page 8 of 50

ItemProposed Devices Atlantis® Abutments in TitaniumPredicate Device Atlantis® Abutment for MIS Conical Connection Implants (K193529)Discussion
The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: Dentsply Sirona Implant System: OsseoSpeed/OsseoSpeed TX Implant Diameter (in mm): 3.0 (Yellow), 3.5, 4.0 (Aqua), 4.5, 5.0 (Lilac), 4.0 (Blue - Previous) Abutment Platform Diameter (in mm): 2.330 (Yellow), 2.866 (Aqua), 3.865 (Lilac), 3.368 (Blue - Previous) Implant Manufacturer: MIS Implant Technologies Inc. Implant System: V3 (NP, SP), C1 (NP, SP, WP) Implant Diameter (in mm): V3: 3.30 (NP), 3.90, 4.3, 5.0 (SP); C1: 3.30 (NP), 3.75, 4.2 (SP), 5.0 (WP) Abutment Platform Diameter (in mm): V3: 2.765 (NP), 3.16 (SP); C1: 2.76 (NP), 3.16 (SP), 4.01 (WP) Implant Manufacturer: Altatec GmbH Implant System: Conelog Screw-Line Implant Implant Diameter (in mm): 3.3, 3.8, 4.3, 5.0 Abutment Platform Diameter (in mm): 2.80 (3.3), 3.00 (3.8, 4.3), 3.80 (5.0) Implant Manufacturer: Nobel Biocare Implant System: NobelActive Implant Diameter (in mm): 3.0 Abutment Platform Diameter (in mm): 2.461 Implant System: Branemark System (NP, RP, WP) Implant Diameter (in mm): 3.3 (NP), 3.75, 4.0 (RP), 5.0, 5.5 (WP) Abutment Platform Diameter (in mm): 3.476 (NP), 4.069 (RP), 5.033 (WP) Implant Manufacturer: BioHorizons Implant System: Prodigy System Internal/Tapered Internal Implant Diameter (in mm): 3.5, 4.5, 5.7Atlantis® Healing Abutment to the endosseous implant. Atlantis® Abutment is compatible with MIS Conical Connection implant from MIS Implant System. Atlantis® products are compatible with the implants shown in the table below. Implant manufacturer – MIS-IMPLANT TECHNOLOGIES INC Trade Name- Abutment Platform Diameter – Implant Diameter Atlantis Abutment for MIS V3 NP - Ø2.765mm -V3: Ø3.30 mm Atlantis Abutment for MIS C1 NP - Ø2.76mm - C1: Ø3.30 mm Atlantis Abutment for MIS C1 & V3 SP - Ø3.16mm - C1: Ø3.75,4.2 mm, V3: Ø3.90,4.3,5.0 mm Atlantis Abutment for MIS C1 WP - Ø4.01mm C1: Ø5.0 mm

Page 8 of 50

Page 14

Page 9 of 50

ItemProposed Devices Atlantis® Abutments in TitaniumPredicate Device Atlantis® Abutment for MIS Conical Connection Implants (K193529)Discussion
Abutment Platform Diameter (in mm): 3.434 (3.5), 4.399 (4.5), 5.644 (5.7) Implant System: Maestro Implant Diameter (in mm): 3.5, 4.0, 5.0 Abutment Platform Diameter (in mm): 3.708 (3.5), 4.242 (4.0), 5.258 (5.0) Implant Manufacturer: Biomet 3i Implant System: Certain, Certain Prevail, Certain XP Implant Diameter (in mm): Certain: 3.25, 4, 5, 6; Certain Prevail: 3.4, 4, 5; Certain XP: 5, 6 Abutment Platform Diameter (in mm): 3.404 (3.25, 3.4), 4.012 (4), 4.942 (5), 5.951 (6) Implant System: Osseotite Tapered and Osseotite Parallel Walled Implant Diameter (in mm): 3.25, 3.75, 4.0, 5.0, 6.0 Abutment Platform Diameter (in mm): 4.056 (3.25, 3.75, 4.0), 4.978 (5.0), 5.985 (6.0)

Page 9 of 50

Page 15

Page 10 of 50

Table 5.2: Comparison between the originally cleared Indications for Use and the proposed Indications for Use statement for each Atlantis Abutments in Titanium

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
K193529The ATLANTIS® Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the ATLANTIS® Abutment to the endosseous implant. The ATLANTIS® Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS® Crown Abutment to the endosseous implant. The ATLANTIS® Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment. The abutment screw is intended to secure the ATLANTIS® Conus Abutment to the endosseous implant. The ATLANTIS® Healing Abutment can be used with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The abutment screw is intended to secure the ATLANTIS® Healing Abutment to the endosseous implant. ATLANTIS® Abutment is compatible with MIS Conical Connection implant from MIS Implant System.Atlantis Abutment & Atlantis Abutment Milling The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant. Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage orAdded Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the 3Shape software (K151455). Removed incorrect statement around "The prosthesis [crown] can be cemented or screw retained to the abutment". This was removed because a crown cannot be screwed to an abutment. Crowns are always cemented to an abutment.

Page 10 of 50

Page 16

Page 11 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
ATLANTIS® products are compatible with the implants shown in the table below. Implant manufacturer - MIS-IMPLANT TECHNOLOGIES INC Trade Name- Abutment Platform Diameter-Implant Diameter Atlantis Abutment for MIS V3 NP - ∅2.765mm -V3: ∅3.30 mm Atlantis Abutment for MIS Cl NP - ∅2.76mm - Cl: ∅3.30 mm Atlantis Abutment for MIS C1 & V3 SP - ∅3. l 6mm - Cl: ∅3.75,4.2 mm, V3: ∅3.90,4.3,5.0 mm Atlantis Abutment for MIS Cl WP - ∅4.01 mm - Cl: ∅5.0 mmtwo-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: MIS Implant Technologies Inc. Implant System: V3 (NP, SP), C1 (NP, SP, WP) Implant Diameter (in mm): V3: 3.30 (NP), 3.90, 4.3, 5.0 (SP); C1: 3.30 (NP), 3.75, 4.2 (SP), 5.0 (WP) Abutment Platform Diameter (in mm): V3: 2.765 (NP), 3.16 (SP); C1: 2.76 (NP), 3.16 (SP), 4.01 (WP)
K161030The ATLANTIS™ Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the ATLANTIS Abutment to the endosseous implant. The ATLANTIS™ Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS Crown Abutment to the endosseous implant. The ATLANTIS™ Conus Abutment is intended forAtlantis Abutment & Atlantis Abutment Milling The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant. Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended forAdded Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the 3Shape software (K151455). Removed incorrect statement around "The prosthesis [crown] can be cemented or screw retained to the abutment". This was removed because a crown cannot be screwed to an abutment. Crowns are always cemented to an abutment. Added Healing Abutment indications.

Page 11 of 50

Page 17

Page 12 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment. The abutment screw is intended to secure the ATLANTIS Conus Abutment to the endosseous implant. ATLANTIS® products are compatible with the implants shown in the table below. Implant manufacturer - MIS-IMPLANT TECHNOLOGIES INC Trade Name- Abutment Platform Diameter-Implant Diameter Atlantis Abutment for MIS V3 NP - ∅2.765mm -V3: ∅3.30 mm Atlantis Abutment for MIS C1 NP - ∅2.76mm - C1: ∅3.30 mm Atlantis Abutment for MIS C1 & V3 SP - ∅3. l 6mm - C1: ∅3.75,4.2 mm, V3: ∅3.90,4.3,5.0 mm Atlantis Abutment for MIS C1 WP - ∅4.01 mm - C1: ∅5.0 mmuse with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: MIS Implant Technologies Inc. Implant System: V3 (NP, SP), C1 (NP, SP, WP) Implant Diameter (in mm): V3: 3.30 (NP), 3.90, 4.3, 5.0 (SP); C1: 3.30 (NP), 3.75, 4.2 (SP), 5.0 (WP) Abutment Platform Diameter (in mm): V3: 2.765 (NP), 3.16 (SP); C1: 2.76 (NP), 3.16 (SP), 4.01 (WP)
K161030The ATLANTIS™ Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the ATLANTIS Abutment to the endosseous implant. The ATLANTIS™ Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS Crown Abutment to the endosseous implant. The ATLANTIS™ Conus Abutment is intended forAtlantis Abutment & Atlantis Abutment Milling The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant. Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended forAdded Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the 3Shape software (K151455). Removed incorrect statement around "The prosthesis [crown] can be cemented or screw retained to the abutment". This was removed because a crown cannot be screwed to an abutment. Crowns are always cemented to an abutment. Added Healing Abutment indications.

Page 12 of 50

Page 18

Page 13 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment. The abutment screw is intended to secure the ATLANTIS Conus Abutment to the endosseous implant. ATLANTIS™ products are compatible with the implants shown in the table below. Implant Manufacturer - Trade Name - Implant Diameter - Abutment Platform Diameter Altatec GmbH CONELOG SCREW-LINE Implant ∅3.3, 3.8, 4.3, 5.0 mm ∅3.3, 3.8, 4.3, 5.0 mmuse with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: Altatec GmbH Implant System: Conelog Screw-Line Implant Implant Diameter (in mm): 3.3, 3.8, 4.3, 5.0 Abutment Platform Diameter (in mm): 2.80 (3.3), 3.00 (3.8, 4.3), 3.80 (5.0)Correction was made to the abutment platform diameter since the original indications repeated the implant diameter measurements instead of specifically referring to the abutment platform diameter measurements.
K151039The ATLANTIS™ Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the ATLANTIS™ Abutment to the endosseous implant. The ATLANTIS™ Crown Abutment is intended forThe Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant. Atlantis Crown AbutmentAdded Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the 3Shape software (K151455). Removed incorrect statement around "The prosthesis [crown] can be

Page 13 of 50

Page 19

Page 14 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS™ Crown Abutment to the endosseous implant. The ATLANTIS™ Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment retained by friction fit to the abutment. The abutment screw is intended to secure the ATLANTIS™ Conus Abutment to the endosseous implant. ATLANTIS™ Abutment for NobelActive 3.0 is compatible with the NobelActive 3.0 implant.The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: Nobel Biocare Implant System: NobelActive Implant Diameter (in mm): 3.0 Abutment Platform Diameter (in mm): 2.461cemented or screw retained to the abutment". This was removed because a crown cannot be screwed to an abutment. Crowns are always cemented to an abutment. Added the abutment platform diameter information to be consistent with the more recent Atlantis Abutment clearances. Added Healing Abutment indications. Although this specific implant system compatibility does not have a Conus Abutment (Conus Abutment was cleared but now discontinued), the Atlantis Conus Abutment is still included in the general indications as all Atlantis Abutments in Titanium share the same indications and Instructions for Use. The user will refer to the compatibility chart to confirm which abutments are available with which implant system.
K081666The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the maxillary lateral incisors and mandibular lateral and central incisors. TheThe Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandibleAdded Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the

Page 14 of 50

Page 20

Page 15 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. This device is compatible with the Astra Tech OsseoSpeed 3.0 mm Implant. Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional. Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the Implant.or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant. Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: Dentsply Sirona Implant System: OsseoSpeed/OsseoSpeed TX Implant Diameter (in mm): 3.0 (Yellow) Abutment Platform Diameter (in mm): 2.330 (Yellow)3Shape software (K151455). Removed the statement around "The prosthesis [crown] can be cemented retained to the abutment". This was removed as this is redundant since crowns are always cemented to an abutment. Added the abutment platform diameter information to be consistent with the more recent Atlantis Abutment clearances. Added Healing Abutment, Conus Abutment, and Crown Abutment indications. This abutment is not offered as a Conus Abutment. However, all Atlantis Abutments in Titanium share the same indications and Instructions for Use. The user will refer to the compatibility chart to confirm which abutments are available with which implant system. Limitations for use statement ("limited to maxillary lateral incisors and mandibular lateral and central incisors") specific to small diameters such as the Atlantis Abutment for OsseoSpeed/OsseoSpeed TX 3.0 are now partially captured in the Contraindications section of the Instructions for Use. In addition, this limitation for use is fully captured in the implant system Indications for Use (K080396, K101732). The implant system indications and placement

Page 15 of 50

Page 21

Page 16 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
limitations dictate where the abutment will be placed. Therefore, the limitations for use are not repeated in the indications. The special note was removed altogether for the following reasons: - The language regarding surgical technique (early loading protocol) is more appropriate for the implant system indications and has therefore been removed from the abutment indications. - The highly angled abutment limitations were moved to the Contraindications section of the Instructions for Use.
K073258The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. This device is compatible with the following manufacturers' implant systems: BioHorizons The Prodigy System™ Dental Implants 3.5mm, 4.0mm, 5.0mm and 6.0mm Implants. Please note: This device may be used in an early load situation,but is dependent on the specific implant system and protocol used by the dental professional. Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant. Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended for use with an endosseous implant to support aAdded Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the 3Shape software (K151455). Removed the statement around "The prosthesis [crown] can be cemented retained to the abutment". This was removed as this is redundant since crowns are always cemented to an abutment. Added the abutment platform diameter information to be consistent with the more recent Atlantis Abutment clearances. Added Healing Abutment, Conus Abutment, and Crown Abutment

Page 16 of 50

Page 22

Page 17 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: BioHorizons Implant System: Prodigy System Internal/Tapered Internal Implant Diameter (in mm): 3.5, 4.5, 5.7 Abutment Platform Diameter (in mm): 3.434 (3.5), 4.399 (4.5), 5.644 (5.7)indications. The dental implant system diameter was updated to be consistent with the implant manufacturer platform diameter information. The implant body diameters are 3.5 mm, 4.0 mm, 5.0 mm and 6.0 mm, which correspond to the implant manufacturer's designation as follows: 3.5 = 3.5; 4.0 = 4.5; and 5.0/6.0 = 5.7. The special note was removed altogether for the following reasons: - The language regarding surgical technique (early loading protocol) is more appropriate for the implant system indications and has therefore been removed from the abutment indications. - The highly angled abutment limitations were moved to the Contraindications section of the Instructions for Use.
K070833The Atlantis Abutment is intended for use as an accessory to an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. The Atlantis Abutment for Astra Implant is compatible with Astra's OsseoSpeed™ 3.5(S) mm, 4.0(S) mm Internal Hex Implants. Please note: This device may be used in an early loadThe Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant. Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the finalAdded Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the 3Shape software (K151455). Removed the statement around "The prosthesis [crown] can be cemented retained to the abutment". This was removed as this is redundant since crowns are always cemented to an abutment.

Page 17 of 50

Page 23

Page 18 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
situation, but is dependent on the specific implant system and protocol used by the dental professional. Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: BioHorizons Implant System: Prodigy System Internal/Tapered Internal Implant Diameter (in mm): 3.5, 4.5, 5.7 Abutment Platform Diameter (in mm): 3.434 (3.5), 4.399 (4.5), 5.644 (5.7)Added the abutment platform diameter information to be consistent with the more recent Atlantis Abutment clearances. Added Healing Abutment, Conus Abutment, and Crown Abutment indications. The dental implant system diameter was updated to be consistent with the implant manufacturer platform diameter information. The implant body diameters are 3.5 mm, 4.0 mm, 5.0 mm and 6.0 mm, which correspond to the implant manufacturer's designation as follows: 3.5 = 3.5; 4.0 = 4.5; and 5.0/6.0 = 5.7. The special note was removed altogether for the following reasons: - The language regarding surgical technique (early loading protocol) is more appropriate for the implant system indications and has therefore been removed from the abutment indications. - The highly angled abutment limitations were moved to the Contraindications section of the Instructions for Use.
K070833The Atlantis Abutment is intended for use as an accessory to an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. The Atlantis Abutment for Astra Implant is compatible with Astra's OsseoSpeed™ 3.5(S) mm, 4.0(S) mm Internal Hex Implants. Please note: This device may be used in an early loadThe Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant. Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the finalAdded Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the 3Shape software (K151455). Removed the statement around "The prosthesis [crown] can be cemented retained to the abutment". This was removed as this is redundant since crowns are always cemented to an abutment.

Page 18 of 50

Page 24

Page 19 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: Dentsply Sirona Implant System: OsseoSpeed/OsseoSpeed TX Implant Diameter (in mm): 3.5, 4.0 (Aqua) Abutment Platform Diameter (in mm): 2.866 (Aqua)Added the abutment platform diameter information to be consistent with the more recent Atlantis Abutment clearances. Added Healing Abutment, Conus Abutment, and Crown Abutment indications. The special note was removed altogether for the following reasons: - The language regarding surgical technique (early loading protocol) is more appropriate for the implant system indications and has therefore been removed from the abutment indications. - The highly angled abutment limitations were moved to the Contraindications section of the Instructions for Use.
K062197The Atlantis Abutment is intended for use as an accessory to an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant.Added Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the 3Shape software (K151455). Removed the statement around "The

Page 19 of 50

Page 25

Page 20 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
situation, but is dependent on the specific implant system and protocol used by the dental professional. Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis Is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: Dentsply Sirona Implant System: OsseoSpeed/OsseoSpeed TX Implant Diameter (in mm): 3.5, 4.0 (Aqua) Abutment Platform Diameter (in mm): 2.866 (Aqua)prosthesis [crown] can be cemented retained to the abutment". This was removed as this is redundant since crowns are always cemented to an abutment. Added the abutment platform diameter information to be consistent with the more recent Atlantis Abutment clearances. Added Healing Abutment, Conus Abutment, and Crown Abutment indications. The special note was removed altogether for the following reasons: - The language regarding surgical technique (early loading protocol) is more appropriate for the implant system indications and has therefore been removed from the abutment indications. - The highly angled abutment limitations were moved to the Contraindications section of the Instructions for Use.

Page 20 of 50

Page 26

Page 21 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
K062197The Atlantis Abutment for 3i Certain MicroMini Implant is compatible with 3i Osseotite NT® Certain and Parallel Walled Implants both with 3.25 mm diameters. Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional. Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant fixture.The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant.Added Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the 3Shape software (K151455). Removed the statement around "The prosthesis [crown] can be cemented retained to the abutment". This was removed as this is redundant since crowns are always cemented to an abutment. Added the abutment platform diameter information to be consistent with the more recent Atlantis Abutment clearances. Biomet 3i implant platform of 3.25 corresponds to the 3.4 implant body size. Therefore, implants with designation 3.4, 3.25, or 3/4 correspond to the same implant size and therefore are compatible with the same Atlantis Abutment. The 3i Certain MicroMini Implant naming convention is no longer used by the implant manufacturer. Reference to the implant system has been updated to indicate 3i Certain and Certain Prevail implant system. Added Healing Abutment, Conus Abutment, and Crown Abutment indications. The special note was removed altogether for the following reasons: - The language regarding surgical technique (early loading protocol) is more appropriate for the implant system indications and has therefore been removed from the abutment indications. - The highly angled abutment limitations were moved to the

Page 21 of 50

Page 27

Page 22 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: Biomet 3i Implant System: Certain, Certain Prevail Implant Diameter (in mm): 3.25, 3.4 Abutment Platform Diameter (in mm): 3.404Contraindications section of the Instructions for Use.

Page 22 of 50

Page 28

Page 23 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
K062069The Atlantis Abutment is intended for use as an accessory to an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. The Atlantis Abutment for 3i MicroMini Implant is compatible with the 3i: Osseotite NT® Certain and Parallel Walled Implants both with 3.25mm diameter; Nobel Biocare: Mk Ill 3.3m, Threaded IlL Series 3.3mm, IIL Immediate Impression 3.25mm, Cylindrical BL Series 325mm; BioHorizons: Maestro™ 3.5mm; Lifecore: Self-tapping 3.75mm, Threaded 3.3mm, Cylinder 3.4mm, Titanium 3.3mm; Sterugold: Implamed Hex Screw 3.3mm, 3.75mm, 4.0mm and 5.0mm, Hex Cylinder 3.3mm; Innova: Endopore® External 3.5mm and Entegram T External 3.25mm. Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional. Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant. Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis HealingThis Atlantis Abutment is not available as "Atlantis Abutment Milling". Since the Indications for Use are generic for all proposed Atlantis Abutments in Titanium subject of this submission, the Atlantis Abutment Milling is part of the indications. The user will use the compatibility chart to confirm which abutments are available for design in 3Shape (Atlantis Abutment Milling). Removed the statement around "The prosthesis [crown] can be cemented retained to the abutment". This was removed as this is redundant since crowns are always cemented to an abutment. Added the abutment platform diameter information to be consistent with the more recent Atlantis Abutment clearances. Many of the implant systems originally cleared are discontinued. The compatibilities that remain active and included in the proposed Indications for Use are: • BioHorizons Maestro (which was listed in the K062069 indications as "Maestro™ 3.5mm") • Osseotite Tapered and Osseotite Parallel Walled (which was listed in the K062069 indications as "Osseotite NT Certain and Parallel Walled Implants both

Page 23 of 50

Page 29

Page 24 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
K062069The Atlantis Abutment is intended for use as an accessory to an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. The Atlantis Abutment for 3i MicroMini Implant is compatible with the 3i: Osseotite NT® Certain and Parallel Walled Implants both with 3.25mm diameter; Nobel Biocare: Mk Ill 3.3m, Threaded IlL Series 3.3mm, IIL Immediate Impression 3.25mm, Cylindrical BL Series 325mm; BioHorizons: Maestro™ 3.5mm; Lifecore: Self-tapping 3.75mm, Threaded 3.3mm, Cylinder 3.4mm, Titanium 3.3mm; Sterugold: Implamed Hex Screw 3.3mm, 3.75mm, 4.0mm and 5.0mm, Hex Cylinder 3.3mm; Innova: Endopore® External 3.5mm and Entegram T External 3.25mm. Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional. Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.Abutment to the endosseous implant. Implant Manufacturer: BioHorizons Implant System: Maestro Implant Diameter (in mm): 3.5 Abutment Platform Diameter (in mm): 3.708 Implant Manufacturer: Biomet 3i Implant System: Osseotite Tapered and Osseotite Parallel Walled Implant Diameter (in mm): 3.25 Abutment Platform Diameter (in mm): 4.056 Implant Manufacturer: Nobel Biocare Implant System: Branemark System (NP) Implant Diameter (in mm): 3.3 Abutment Platform Diameter (in mm): 3.476with 3.25mm diameter") • Branemark (which was listed in the indications as "MK III 3.3mm") The special note was removed altogether for the following reasons: - The language regarding surgical technique (early loading protocol) is more appropriate for the implant system indications and has therefore been removed from the abutment indications. - The highly angled abutment limitations were moved to the Contraindications section of the Instructions for Use. Added Healing Abutment, Conus Abutment, and Crown Abutment indications. This abutment is not offered as a Healing Abutment. All Atlantis Abutments in Titanium share the same indications and Instructions for Use. The user will refer to the compatibility chart to confirm which abutments are available with which implant system.
K051652The Atlantis Abutment is intended for use as an accessory to an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant.Added Atlantis Abutment Milling to the Atlantis Abutment indications. Software validation is included in the submission to demonstrate that the same design limitations exist between the internal CAD software and the 3Shape software (K151455). Removed the statement around "The prosthesis [crown] can be cemented

Page 24 of 50

Page 30

Page 25 of 50

510(k)Predicate/Reference Device Indications for UseProposed Device Indications for UseDiscussion
Atlantis Crown Abutment The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant. Atlantis Conus Abutment The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment- retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant. Atlantis Healing Abutment The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant. Implant Manufacturer: BioHorizons Implant System: Maestro Implant Diameter (in mm): 3.5 Abutment Platform Diameter (in mm): 3.708 Implant Manufacturer: Biomet 3i Implant System: Osseotite Tapered and Osseotite Parallel Walled Implant Diameter (in mm): 3.25 Abutment Platform Diameter (in mm): 4.056 Implant Manufacturer: Nobel Biocare Implant System: Branemark System (NP) Implant Diameter (in mm): 3.3 Abutment Platform Diameter (in mm): 3.476retained to the abutment". This was removed as this is redundant since crowns are always cemented to an abutment. Added Healing Abutment, Conus Abutment, and Crown Abutment indications. This abutment is not offered as a Healing Abutment. All Atlantis Abutments in Titanium share the same indications and Instructions for Use. The user will refer to the compatibility chart to confirm which abutments are available with which implant system.

Page 25 of 50

Page 31

Page 26 of 50

6. Technological Comparison:

The primary purpose of the Bundled Traditional 510(k) premarket notification is to modify the labeling of the Atlantis® Abutments in Titanium to update the safety in an MR environment to indicate MR Conditional. The Abutments will be changed from MR Not Evaluated to MR Conditional. The labeling change reflects the requirements in the FDA guidance document, "Testing and Labeling Devices for Safety in the Magnetic Resonance (MR) Environment" and ensures that the clinician has the necessary information to communicate to the patient regarding the MR Safety of the proposed devices.

In addition to the MR Conditional labeling changes, internally documented changes and proposed modifications are included in the submission.

Angulated Screw Access (ASA)

Dentsply Sirona added the Angulated Screw Access (ASA) feature to some Atlantis® Abutments, as reviewed in reference device K151439, Atlantis® Suprastructures. The angulated screw access allows the clinician to access the screw access channel from the lingual side, which facilitates the installation procedure of the abutment.

Healing, Conus and Crown Abutment offering

Dentsply Sirona expanded the abutment offering to include Healing Abutments, Conus Abutments, and Crown Abutments that were not originally cleared with some Atlantis® Abutments. This offering is already cleared for the Atlantis® Abutment for MIS Conical Connection Implants in predicate device, K193529. The additional abutments (Healing, Conus, and Crown) are within the design envelope of the corresponding cleared Atlantis® Abutment implant system compatibilities and have the same mating geometry to the implant, which have been tested per ISO 14801 in respective original clearances.

Screw and abutment diameter modifications

Minor modification was made to some of the abutment screws since the original clearance of the Atlantis® Abutments in Titanium. These changes were made to standardize dimensions across the various custom abutment screws to simplify the procurement and quality control process. Upon review, the changes to the screws did not cause an interference with the internal geometry of the abutment and the modified screws are considered as safe and effective as the predicate device and reference device screws.

Limited to two implant system compatibilities (reference devices K073258 and K051652), the abutment diameter was slightly modified. Compatibility analysis was performed and the results indicate that the modifications do not result in a new worst-case for fatigue testing. Fatigue testing submitted in the original submissions remains valid.

Additional Compatibility

Added compatibility of existing Atlantis® Abutments in Titanium (K081666, K070833, K051652) with the Osseospeed TX implant system (K101732). The OsseoSpeed TX implant system has the same internal geometry as the OsseoSpeed Implant System (K080396, K053384, K041492) and therefore the Atlantis® Abutments in Titanium for OsseoSpeed/OsseoSpeed TX are compatible with both implant systems.

Atlantis Abutment Milling (Design in 3Shape)

Some Atlantis® Abutments (predicate device, K193529, and reference devices K151039,

Page 25 of 50

Page 32

Page 26 of 50

K161030, K081666, K073258, K070833, K062197, K051652, K051011) can be designed by the user via 3Shape (3Shape Abutment Designer Software, K151455). After the abutment is designed by the lab, the abutment is manufactured by Dentsply Sirona. This proposed workflow is also referred to as Atlantis® Abutment Milling in this submission. The design limitations that are locked in the 3Shape abutment library are the same design limitations as those in the internal DS CAD system called Virtual Abutment Design (VAD). To ensure that the same design limitations/constraints present in the 3Shape Atlantis® Abutment library match those in the internal CAD system, and that the design parameters cannot be exceeded, validation was performed. The software validation confirms that the additional option of using the 3Shape design module for design of Atlantis® Abutments does not raise new questions of safety and performance as it was confirmed that the locked design library for Atlantis® Abutments in 3Shape has the same design limitations as those built into the internal VAD software.

An overall comparison of the technological characteristics of the proposed modified Atlantis® Abutments in Titanium to the predicate device (K193529) and reference devices K161030, K151039, K081666, K073258, K070833, K062197, K062069, K051652, K051011, K981858) is provided in the Comparison Tables below.

Table 6.1 Comparison of the technological characteristics between proposed modified Atlantis® Abutment in Titanium for MIS V3 and C1 implants and predicate device, Atlantis® Abutment in Titanium for MIS V3 and C1 implants (K193529)

ItemModified Atlantis® Abutment in Titanium for MIS implants V3 and C1Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Prosthesis attachmentScrew-retained Cement-retained Friction FitScrew-retained Cement-retained Friction FitSame
AbutmentsAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentSame
RestorationSingle or Multi-unitSingle or Multi-unitSame
Abutment platform diameterØ2.765, 2.76, 3.16, 4.01 mmØ2.765, 2.76, 3.16, 4.01 mmSame
Compatible Implant diameterV3: Ø3.3, 3.9, 4.3, 5.0 mm C1: Ø3.3, 3.75, 4.2, 5.0 mmV3: Ø3.3, 3.9, 4.3, 5.0 mm C1: Ø3.3, 3.75, 4.2, 5.0 mmSame
Compatible Implant SystemMIS V3 (K163349), MIS C1 (K172505)MIS V3 (K163349), MIS C1 (K172505)Same
ConnectionInternal Conical ConnectionInternal Conical ConnectionSame
Abutment angulationStraight, Up to 30°Straight, Up to 30°Same
Maximum abutment15 mm15 mmSame

Page 26 of 50

Page 33

Page 27 of 50

ItemModified Atlantis® Abutment in Titanium for MIS implants V3 and C1Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
height
Maximum abutment width from implant axis6 mm6.5 mmDifferent, however in K193529, the tested abutment maximum width from the implant axis was below 6.0 mm.
Minimum abutment width from implant axis1.3525 mm1.65 mmDifferent, however in K193529, the worst-case abutment that was tested was within the minimum abutment width from the implant axis.
Minimum abutment post height4 mm4 mmSame
Material: AbutmentTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumSame
Material: ScrewTitanium alloyTitanium alloySame
Screw typeStraight screw Angulated screw – access hole on the lingual sideStraight screwDifferent due to additional ASA option.
Screw access angulationStraight or angled up to 20°StraightDifferent due to additional ASA option.
Screw drive featureHex HexalobularHexDifferent due to the hexalobular screw drive feature for Atlantis ASA screwdriver.
MR SafetyMR ConditionalMR Not EvaluatedDifferent. Testing included in reference device (K221094) supports labeling update.

Page 27 of 50

Page 34

Page 28 of 50

Table 6.2: Comparison of the technological characteristics between proposed modified Atlantis® Abutment in Titanium for Conelog implants, predicate device, Atlantis® Abutment in Titanium (K193529), and reference device Atlantis® Abutment in Titanium for Conelog implants (K161030)

ItemModified Atlantis® Abutment in Titanium for Conelog implantsReference device Atlantis® Abutment in Titanium for Conelog implants (K161030)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Prosthesis attachmentScrew-retained Cement-retained Friction FitScrew-retained Cement-retained Friction FitScrew-retained Cement-retained Friction FitSame
AbutmentsAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentSame offering as the predicate device. The same design envelope applies to all Atlantis Healing Abutments in Titanium.
RestorationSingle or Multi-unitSingle or Multi-unitSingle or Multi-unitSame
Abutment platform diameterØ2.80, 3.00, 3.80 mmØ2.80, 3.00, 3.80 mmØ2.765, 2.76, 3.16, 4.01 mmSame as reference device
Compatible Implant DiameterØ 3.3, 3.8, 4.3, 5.0 mmØ 3.3, 3.8, 4.3, 5.0 mmV3: Ø3.3, 3.9, 4.3, 5.0 mm C1: Ø3.3, 3.75, 4.2, 5.0 mmSame as reference device
Compatible Implant SystemConelog Screw-Line Implant 3.3 Conelog Screw-Line Implant 3.8/4.3 Conelog Screw-Line Implant 5.0 (K113779)Conelog implant system (K113779)MIS V3 (K163349), MIS C1 (K172505)Same. No change. Table now specifies the individual compatible Conelog implants instead of the system as a whole.
ConnectionInternal Conical ConnectionInternal Conical ConnectionInternal Conical ConnectionSame as reference device
Abutment angulationStraight, Up to 30°Straight, Up to 30°Straight, Up to 30°Same
Maximum abutment height15 mm15 mm15 mmSame
Maximum abutment width from implant axis6 mm6.5 mm6.5 mmDifferent, however in K161030, the tested abutment maximum width from the implant axis was below 6.0 mm.
Minimum abutment width from implant axis1.3525 mm1.65 mm1.65 mmDifferent, however in K161030, the worst-case abutment that was tested was within the minimum abutment width from

Page 28 of 50

Page 35

Page 29 of 50

ItemModified Atlantis® Abutment in Titanium for Conelog implantsReference device Atlantis® Abutment in Titanium for Conelog implants (K161030)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
the implant axis.
Minimum abutment post height4 mm4 mm4 mmSame
Material: AbutmentTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumSame
Material: ScrewTitanium alloyTitanium alloyTitanium alloySame
Screw typeStraight screwStraight screwStraight screwSame
Screw access angulationStraightStraightStraightSame
Screw drive featureHexHexHexSame
MR SafetyMR ConditionalMR Not EvaluatedMR Not EvaluatedDifferent. Testing included in reference device (K221094) supports labeling update.

Table 6.3: Comparison of the technological characteristics between proposed modified Atlantis® Abutment in Titanium for Nobel Active 3.0 implants, predicate device, Atlantis® Abutment in Titanium (K193529), and reference device Atlantis® Abutment in Titanium for Nobel Active 3.0 implants (K151039)

ItemModified Atlantis® Abutment in Titanium for Nobel Active 3.0 implantsReference device Atlantis® Abutment in Titanium for Nobel Active 3.0 implants (K151039)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Prosthesis attachmentScrew-retained Cement-retained Note: Atlantis Conus Abutment with friction fit is not currently marketed for this implantScrew-retained Cement-retained Friction FitScrew-retained Cement-retained Friction FitSame
AbutmentsAtlantis Abutment Atlantis Crown Abutment Atlantis Healing Abutment Note: Atlantis Conus AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentOffering is covered by predicate device. The additional offering is within the same design envelope as the reference device.

Page 29 of 50

Page 36

Page 30 of 50

ItemModified Atlantis® Abutment in Titanium for Nobel Active 3.0 implantsReference device Atlantis® Abutment in Titanium for Nobel Active 3.0 implants (K151039)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
is not currently marketed for this implantthe implant axis.
Minimum abutment post height4 mm4 mm4 mmSame
Material: AbutmentTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumSame
Material: ScrewTitanium alloyTitanium alloyTitanium alloySame
Screw typeStraight screwStraight screwStraight screwSame
Screw access angulationStraightStraightStraightSame
Screw drive featureHexHexHexSame
MR SafetyMR ConditionalMR Not EvaluatedMR Not EvaluatedDifferent. Testing included in reference device (K221094) supports labeling update.

Table 6.3: Comparison of the technological characteristics between proposed modified Atlantis® Abutment in Titanium for Nobel Active 3.0 implants, predicate device, Atlantis® Abutment in Titanium (K193529), and reference device Atlantis® Abutment in Titanium for Nobel Active 3.0 implants (K151039)

ItemModified Atlantis® Abutment in Titanium for Nobel Active 3.0 implantsReference device Atlantis® Abutment in Titanium for Nobel Active 3.0 implants (K151039)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Prosthesis attachmentScrew-retained Cement-retained Note: Atlantis Conus Abutment with friction fit is not currently marketed for this implantScrew-retained Cement-retained Friction FitScrew-retained Cement-retained Friction FitSame
AbutmentsAtlantis Abutment Atlantis Crown Abutment Atlantis Healing Abutment Note: Atlantis Conus AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentOffering is covered by predicate device. The additional offering is within the same design envelope as the reference device.

Page 30 of 50

Page 37

Page 31 of 50

ItemModified Atlantis® Abutment in Titanium for Nobel Active 3.0 implantsReference device Atlantis® Abutment in Titanium for Nobel Active 3.0 implants (K151039)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
is not currently marketed for this implant
RestorationSingle or Multi-unitSingle or Multi-unitSingle or Multi-unitSame
Abutment platform diameterØ2.461 mmØ2.461 mmØ2.765, 2.76, 3.16, 4.01 mmSame as reference device
Compatible Implant Diameter3.0 mm3.0 mmV3: Ø3.3, 3.9, 4.3, 5.0 mm C1: Ø3.3, 3.75, 4.2, 5.0 mmSame as reference device
Compatible Implant SystemNobel Biocare Active Implant 3.0 (K102436)Nobel Biocare Active Implant 3.0 (K102436)MIS V3 (K163349), MIS C1 (K172505)Same as reference device
ConnectionInternal Conical ConnectionInternal Conical ConnectionInternal Conical ConnectionSame as reference device
Abutment angulationStraight, Up to 30°Straight, Up to 30°Straight, Up to 30°Same
Maximum abutment height15 mm15 mm15 mmSame
Maximum abutment width from implant axis6 mm6.5 mm6.5 mmDifferent, however in K151039, the tested abutment maximum width from the implant axis was below 6.0 mm.
Minimum abutment width from implant axis1.3525 mm1.65 mm1.65 mmDifferent, however in K151039, the worst-case abutment that was tested was within the minimum abutment width from the implant axis.
Minimum abutment post height4 mm4 mm4 mmSame
Material: AbutmentTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumSame
Material: ScrewTitanium alloyTitanium alloyTitanium alloySame
Screw typeStraight screw Angulated screw – access hole onStraight screwStraight screwDifferent due to additional ASA option.

Page 31 of 50

Page 38

Page 32 of 50

ItemModified Atlantis® Abutment in Titanium for Nobel Active 3.0 implantsReference device Atlantis® Abutment in Titanium for Nobel Active 3.0 implants (K151039)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
the lingual side
Screw access angulationStraight or angled up to 30°StraightStraightDifferent due to additional ASA option. The ASA screw access channel ranges from 0 to 30°.
Screw drive featureUnigrip HexalobularUnigripHexDifferent due to the hexalobular screw drive feature for Atlantis ASA screwdriver.
MR SafetyMR ConditionalMR Not EvaluatedMR Not EvaluatedDifferent. Testing included in reference device (K221094) supports labeling update.

Table 6.4: Comparison of the technological characteristics between proposed modified Atlantis® Abutment in Titanium for OsseoSpeed 3.0 implants, predicate device, Atlantis® Abutment in Titanium (K193529), and reference device Atlantis® Abutment in Titanium for AstraTech OsseoSpeed 3.0 implants (K081666)

ItemModified Atlantis® Abutment in Titanium for OsseoSpeed 3.0 implantsReference device Atlantis® Abutment in Titanium for OsseoSpeed 3.0 implants (K081666)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Prosthesis attachmentScrew-retained Cement-retainedScrew-retained Cement-retainedScrew-retained Cement-retained Friction FitSame as reference device
AbutmentsAtlantis Abutment Atlantis Crown Abutment Atlantis Healing AbutmentAtlantis AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentOffering is covered by predicate device. The additional offering is within the same design envelope as the reference device.
RestorationSingle or Multi-unitSingle or Multi-unitSingle or Multi-unitSame
Abutment platform diameterØ2.330 mmØ2.330 mmØ2.765, 2.76, 3.16, 4.01 mmSame as reference device
Compatible Implant DiameterØ 3.0 mmØ 3.0 mmV3: Ø3.3, 3.9, 4.3, 5.0 mm C1: Ø3.3, 3.75, 4.2, 5.0 mmSame as reference device

Page 32 of 50

Page 39

Page 33 of 50

ItemModified Atlantis® Abutment in Titanium for OsseoSpeed 3.0 implantsReference device Atlantis® Abutment in Titanium for OsseoSpeed 3.0 implants (K081666)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Compatible Implant SystemOsseoSpeed 3.0 - Yellow OsseoSpeed TX 3.0 - Yellow Note: OsseoSpeed and OsseoSpeed TX have the same connection. (K080396, K101732)Osseospeed 3.0 implant (K080396)MIS V3 (K163349), MIS C1 (K172505)The connection of the Dentsply Sirona OsseoSpeed and OsseoSpeed TX implants are identical and therefore the Atlantis Abutments in Titanium cleared in K081666 are compatible with both Dentsply Sirona OsseoSpeed (K080396) and OsseoSpeed TX (K101732) implant systems.
ConnectionInternal Conical ConnectionInternal Conical ConnectionInternal Conical ConnectionSame as reference device
Abutment angulationStraight, Up to 30°Straight, Up to 30°Straight, Up to 30°Same
Maximum abutment height15 mm15 mm15 mmSame
Maximum abutment width from implant axis6 mm5.5 mm6.5 mmDifferent but still within the 6 mm maximum limit.
Minimum abutment width from implant axis1.3525 mmNot specified1.65 mmAdded minimum abutment width from implant axis to be consistent with more recently cleared Atlantis Abutments. All Atlantis Abutments share the same design envelope.
Minimum abutment post height4 mmNot specified4 mmAdded minimum abutment post height to be consistent with more recently cleared Atlantis Abutments. All Atlantis Abutments share the same design envelope.
Material: AbutmentTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumSame
Material: ScrewTitanium alloyTitanium alloyTitanium alloySame

Page 33 of 50

Page 40

Page 34 of 50

ItemModified Atlantis® Abutment in Titanium for OsseoSpeed 3.0 implantsReference device Atlantis® Abutment in Titanium for OsseoSpeed 3.0 implants (K081666)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Screw typeStraight screw Angulated screw – access hole on the lingual sideStraight screwStraight screwDifferent due to additional ASA option.
Screw access angulationStraight or angled up to 30°StraightStraightDifferent due to additional ASA option. The ASA screw access channel ranges from 0 to 30°.
Screw drive featureUnigrip HexalobularUnigripHexDifferent due to the hexalobular screw drive feature for Atlantis ASA screwdriver.
MR SafetyMR ConditionalMR Not EvaluatedMR Not EvaluatedDifferent. Testing included in reference device (K221094) supports labeling update.

Table 6.4: Comparison of the technological characteristics between proposed modified Atlantis® Abutment in Titanium for OsseoSpeed 3.0 implants, predicate device, Atlantis® Abutment in Titanium (K193529), and reference device Atlantis® Abutment in Titanium for AstraTech OsseoSpeed 3.0 implants (K081666)

ItemModified Atlantis® Abutment in Titanium for OsseoSpeed 3.0 implantsReference device Atlantis® Abutment in Titanium for OsseoSpeed 3.0 implants (K081666)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Prosthesis attachmentScrew-retained Cement-retainedScrew-retained Cement-retainedScrew-retained Cement-retained Friction FitSame as reference device
AbutmentsAtlantis Abutment Atlantis Crown Abutment Atlantis Healing AbutmentAtlantis AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentOffering is covered by predicate device. The additional offering is within the same design envelope as the reference device.
RestorationSingle or Multi-unitSingle or Multi-unitSingle or Multi-unitSame
Abutment platform diameterØ2.330 mmØ2.330 mmØ2.765, 2.76, 3.16, 4.01 mmSame as reference device
Compatible Implant DiameterØ 3.0 mmØ 3.0 mmV3: Ø3.3, 3.9, 4.3, 5.0 mm C1: Ø3.3, 3.75, 4.2, 5.0 mmSame as reference device

Page 34 of 50

Page 41

Page 35 of 50

ItemModified Atlantis® Abutment in Titanium for BioHorizons implantsReference device Atlantis® Abutment in Titanium for BioHorizons implants (K073258)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Prosthesis attachmentScrew-retained Cement-retained Friction-fitScrew-retained Cement-retainedScrew-retained Cement-retained Friction FitSame as predicate device. Added Friction-fit with the introduction of the Conus Abutment.
AbutmentsAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentAtlantis AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentSame as the predicate device. The additional offering is within the same design envelope as the reference device.
RestorationSingle or Multi-unitSingle or Multi-unitSingle or Multi-unitSame
Abutment platform diameterØ3.434, 4.399, 5.644 mmØ3.536, 4.514, 5.720 mmØ2.765, 2.76, 3.16, 4.01 mmSimilar. Minor changes in abutment diameter. Verification and validation activities in design controls confirm continued compatibility and applicability of existing fatigue

Page 35 of 50

Page 42

Page 36 of 50

ItemModified Atlantis® Abutment in Titanium for BioHorizons implantsReference device Atlantis® Abutment in Titanium for BioHorizons implants (K073258)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
testing.
Compatible Implant Diameter3.5, 4.5, 5.7 mm3.5, 4.5, 5.7 mmV3: Ø3.3, 3.9, 4.3, 5.0 mm C1: Ø3.3, 3.75, 4.2, 5.0 mmSame as reference device
Compatible Implant SystemBioHorizons Internal 3.5 BioHorizons Internal 4.5 BioHorizons Internal 5.7 (K042429)BioHorizons Prodigy System (K042429)MIS V3 (K163349), MIS C1 (K172505)Same as reference device Compatible with the same implant system. Branding name change only.
ConnectionInternal Hex, Tapered InternalInternal Hex, Tapered InternalInternal Conical ConnectionSame as reference device
Abutment angulationStraight, Up to 30°Straight, Up to 30°Straight, Up to 30°Same
Maximum abutment height15 mm15 mm15 mmSame
Maximum abutment width from implant axis6 mm5.5 mm6.5 mmDifferent but still within the 6 mm maximum limit.
Minimum abutment width from implant axis1.3525 mmNot specified1.65 mmAdded minimum abutment width from implant axis to be consistent with more recently cleared Atlantis Abutments. All Atlantis Abutments share the same design envelope.
Minimum abutment post height4 mmNot specified4 mmAdded minimum abutment post height to be consistent with more recently cleared Atlantis Abutments. All Atlantis Abutments share the same design envelope.
Material: AbutmentTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumSame
Material: ScrewTitanium alloyTitanium alloyTitanium alloySame

Page 36 of 50

Page 43

Page 37 of 50

ItemModified Atlantis® Abutment in Titanium for BioHorizons implantsReference device Atlantis® Abutment in Titanium for BioHorizons implants (K073258)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Screw typeStraight screw Angulated screw – access hole on the lingual sideStraight screwStraight screwDifferent due to additional ASA option.
Screw access angulationStraight or angled up to 20°StraightStraightDifferent due to additional ASA option. The ASA screw access channel ranges from 0 to 20°.
Screw drive featureHexalobular HexHexHexDifferent due to the hexalobular screw drive feature for Atlantis ASA screwdriver.
MR SafetyMR ConditionalMR Not EvaluatedMR Not EvaluatedDifferent. Testing included in reference device (K221094) supports labeling update.

Table 6.6: Comparison of the technological characteristics between proposed modified Atlantis® Abutment in Titanium for OsseoSpeed 3.5 and 4.0 mm implants, predicate device, Atlantis® Abutment in Titanium (K193529), and reference device Atlantis® Abutment in Titanium for OsseoSpeed 3.5 and 4.0 mm implants (K070833)

ItemModified Atlantis® Abutment in Titanium for OsseoSpeed implants 3.5 and 4.0Reference device Atlantis® Abutment in Titanium for OsseoSpeed implants 3.5 and 4.0 (K070833)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Prosthesis attachmentScrew-retained Cement-retained Friction-fitScrew-retained Cement-retainedScrew-retained Cement-retained Friction FitSame as predicate device. Added Friction-fit with the introduction of the Conus Abutment.
AbutmentsAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentAtlantis AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentSame as the predicate device. The additional offering is within the same design envelope as the reference device.
RestorationSingle or Multi-unitSingle or Multi-unitSingle or Multi-unitSame
Abutment platform diameterØ2.866 mmØ2.866 mmØ2.765, 2.76, 3.16, 4.01 mmSame as reference device

Page 37 of 50

Page 44

Page 38 of 50

ItemModified Atlantis® Abutment in Titanium for OsseoSpeed implants 3.5 and 4.0Reference device Atlantis® Abutment in Titanium for OsseoSpeed implants 3.5 and 4.0 (K070833)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Compatible Implant Diameter3.5, 4.0 mm3.5, 4.0 mmV3: Ø3.3, 3.9, 4.3, 5.0 mm C1: Ø3.3, 3.75, 4.2, 5.0 mmSame as reference device
Compatible Implant SystemOsseoSpeed 3.5/4.0 – Aqua OsseoSpeed TX 3.5/4.0 - Aqua (K053384, K101732) Note: Microthread OsseoSpeed and Osseospeed/OsseoSpeed TX have the same connection.OsseoSpeed implants (formerly AstraTech implants) (K053384, K024111)MIS V3 (K163349), MIS C1 (K172505)The connection of the Dentsply Sirona OsseoSpeed and OsseoSpeed TX implants are identical and therefore the Atlantis Abutments in Titanium cleared in K070833 are compatible with both Dentsply Sirona OsseoSpeed (K053384) and OsseoSpeed TX (K101732) implant systems. Microthread implants branding changed to OsseoSpeed. Implants cleared in K024111 are discontinued.
ConnectionInternal Conical ConnectionInternal Conical ConnectionInternal Conical ConnectionSame as reference device
Abutment angulationStraight, Up to 30°Straight, Up to 30°Straight, Up to 30°Same
Maximum abutment height15 mm15 mm15 mmSame
Maximum abutment width from implant axis6 mm5.5 mm6.5 mmDifferent but still within the 6 mm maximum limit.
Minimum abutment width from implant axis1.3525 mmNot specified1.65 mmAdded minimum abutment width from implant axis to be consistent with more recently cleared Atlantis Abutments. All Atlantis Abutments share the same design envelope.
Minimum abutment4 mmNot specified4 mmAdded minimum abutment post height to be consistent with

Page 38 of 50

Page 45

Page 39 of 50

ItemModified Atlantis® Abutment in Titanium for OsseoSpeed implants 3.5 and 4.0Reference device Atlantis® Abutment in Titanium for OsseoSpeed implants 3.5 and 4.0 (K070833)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
post heightmore recently cleared Atlantis Abutments. All Atlantis Abutments share the same design envelope.
Material: AbutmentTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumSame
Material: ScrewTitanium alloyTitanium alloyTitanium alloySame
Screw typeStraight screw Angulated screw – access hole on the lingual sideStraight screwStraight screwDifferent due to additional ASA option.
Screw access angulationStraight or angled up to 30°StraightStraightDifferent due to additional ASA option. The ASA screw access channel ranges from 0 to 30°.
Screw drive featureHexalobular HexHexHexDifferent due to the hexalobular screw drive feature for Atlantis ASA screwdriver.
MR SafetyMR ConditionalMR Not EvaluatedMR Not EvaluatedDifferent. Testing included in reference device (K221094) supports labeling update.

Table 6.7: Comparison of the technological characteristics between proposed modified Atlantis® Abutment in Titanium for Biomet 3i Certain implants, predicate device, Atlantis® Abutment in Titanium (K193529), and reference device Atlantis® Abutment in Titanium for Biomet 3i Certain implants (K062197)

ItemModified Atlantis® Abutment in Titanium for Biomet 3i Certain implantsReference device Atlantis® Abutment in Titanium for Biomet 3i Certain implants (K062197)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Prosthesis attachmentScrew-retained Cement-retained Friction-fitScrew-retained Cement-retainedScrew-retained Cement-retained Friction FitSame as predicate device. Added Friction-fit with the introduction of the Conus Abutment.

Page 39 of 50

Page 46

Page 40 of 50

ItemModified Atlantis® Abutment in Titanium for Biomet 3i Certain implantsReference device Atlantis® Abutment in Titanium for Biomet 3i Certain implants (K062197)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
AbutmentsAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentAtlantis AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentSame as the predicate device. The additional offering is within the same design envelope as the reference device.
RestorationSingle or Multi-unitSingle or Multi-unitSingle or Multi-unitSame
Abutment platform diameterØ3.404 mmØ3.404 mmØ2.765, 2.76, 3.16, 4.01 mmSame as reference device
Compatible Implant Diameter3.25, 3.4 mm3.25, 3.4 mmV3: Ø3.3, 3.9, 4.3, 5.0 mm C1: Ø3.3, 3.75, 4.2, 5.0 mmSame as reference device
Compatible Implant SystemBiomet 3i Certain 3.25 Biomet 3i Certain 4/3 Biomet 3i Certain Prevail 3/4/3 (3.4), 4/3 (K051461)Biomet 3i Certain implants (K041235)MIS V3 (K163349), MIS C1 (K172505)Same. Compatible with the same implant system. Branding name change only. Note: K062197 referenced the incorrect implant system, an external implant system (K041235). The correct implant system is K051461.
ConnectionInternal Hexagon ConnectionInternal Hexagon ConnectionInternal Conical ConnectionSame as reference device
Abutment angulationStraight, Up to 30°Straight, Up to 30°Straight, Up to 30°Same
Maximum abutment height15 mm15 mm15 mmSame
Maximum abutment width from implant axis6 mm5.5 mm6.5 mmDifferent but still within the 6 mm maximum limit.
Minimum abutment width from implant axis1.3525 mmNot specified1.65 mmAdded minimum abutment width from implant axis to be consistent with more recently cleared Atlantis Abutments. All Atlantis Abutments share the

Page 40 of 50

Page 47

Page 41 of 50

ItemModified Atlantis® Abutment in Titanium for Biomet 3i Certain implantsReference device Atlantis® Abutment in Titanium for Biomet 3i Certain implants (K062197)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
same design envelope.
Minimum abutment post height4 mmNot specified4 mmAdded minimum abutment post height to be consistent with more recently cleared Atlantis Abutments. All Atlantis Abutments share the same design envelope.
Material: AbutmentTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumTitanium alloy, Gold Shaded TitaniumSame
Material: ScrewTitanium alloyTitanium alloyTitanium alloySame
Screw typeStraight screwStraight screwStraight screwSame
Screw access angulationStraightStraightStraightSame
Screw drive featureHexHexHexSame
MR SafetyMR ConditionalMR Not EvaluatedMR Not EvaluatedDifferent. Testing included in reference device (K221094) supports labeling update.

Page 41 of 50

Page 48

Page 42 of 50

Table 6.9: Comparison of the technological characteristics between proposed modified Atlantis® Abutment in Titanium for Osseospeed/Osseospeed TX implants, predicate device, Atlantis® Abutment in Titanium (K193529), and reference device Atlantis® Abutment in Titanium for Osseospeed implants (K051652)

ItemModified Atlantis® Abutment in Titanium for OsseoSpeed implantsReference device Atlantis® Abutment in Titanium for OsseoSpeed implants (K051652)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Prosthesis attachmentScrew-retained Cement-retained Friction-fitScrew-retained Cement-retainedScrew-retained Cement-retained Friction FitSame as predicate device. Added Friction-fit with the introduction of the Conus Abutment.
AbutmentsAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentAtlantis AbutmentAtlantis Abutment Atlantis Crown Abutment Atlantis Conus Abutment Atlantis Healing AbutmentSame as the predicate device. The additional offering is within the same design envelope as the reference device.
RestorationSingle or Multi-unitSingle or Multi-unitSingle or Multi-unitSame
Abutment platform diameterØ 3.368, 3.865 mmØ 3.463 for Ø 4.0 mm For the larger size, the submission did not specify the abutment platform diameter but the compatible implant diameter of Ø 4.5, 5.0Ø2.765, 2.76, 3.16, 4.01 mmMinor changes in abutment diameter. Verification and validation activities in design controls confirm continued compatibility and applicability of existing fatigue testing.
Compatible Implant DiameterØ 4.0, 4.5, 5.0 mmØ 4.0, 4.5, 5.0 mmV3: Ø3.3, 3.9, 4.3, 5.0 mm C1: Ø3.3, 3.75, 4.2, 5.0 mmSame as reference device
Compatible Implant SystemOsseoSpeed 4.0 Blue (Previous) (K041492) OsseoSpeed/OsseoSpeed TX 4.5, 5.0 - Lilac Note: Microthread OsseoSpeed and Osseospeed/OsseoSpeed TX have the same connection. (K053384, K101732)Microthread OsseoSpeed 4.0, 4.5, 5.0 (Compatible implant system is not specified in submission but referred to as Astra's Microthread ST 4.0, 4.5, and 5.0 mm, which corresponds to K041492, K053384)MIS V3 (K163349), MIS C1 (K172505)The connection of the Dentsply Sirona OsseoSpeed and OsseoSpeed TX implants are identical and therefore the Atlantis Abutments in Titanium cleared in K051652 are compatible with both Dentsply Sirona OsseoSpeed (K041492, K053384) and OsseoSpeed TX (K101732) implant systems.
ConnectionInternal Conical ConnectionInternal Conical ConnectionInternal Conical ConnectionSame as reference device
Abutment angulationStraight, Up to 30°Straight, Up to 30°Straight, Up to 30°Same
Maximum abutment height15 mm15 mm15 mmSame

Page 42 of 50

Page 49

Page 43 of 50

ItemModified Atlantis® Abutment in Titanium for OsseoSpeed implantsReference device Atlantis® Abutment in Titanium for OsseoSpeed implants (K051652)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
Maximum abutment width from implant axis6 mm5.5 mm6.5 mmDifferent but still within the 6 mm maximum limit.
Minimum abutment width from implant axis1.3525 mmNot specified1.65 mmAdded minimum abutment width from implant axis to be consistent with more recently cleared Atlantis Abutments. All Atlantis Abutments share the same design envelope.
Minimum abutment post height4 mmNot specified4 mmAdded minimum abutment post height to be consistent with more recently cleared Atlantis Abutments. All Atlantis Abutments share the same design envelope.
Material: AbutmentTitanium alloy, Gold Shaded TitaniumTitanium alloyTitanium alloy, Gold Shaded TitaniumSame as predicate device. Gold-shaded titanium abutments (TiN abutments) have been reviewed by the Agency during the predicate device submission as well as in earlier submissions. This corresponds to Titanium with a titanium nitride layer applied using PVD (Physical Vapor Deposition) to create a gold shade for a more esthetic outcome.
Material: ScrewTitanium alloyTitanium alloyTitanium alloySame
Screw typeStraight screw Angulated screwStraight screwStraight screwDifferent due to additional ASA option.
Screw accessStraight or angled up to 30°StraightStraightDifferent due to additional ASA option.

Page 43 of 50

Page 50

Page 44 of 50

ItemModified Atlantis® Abutment in Titanium for OsseoSpeed implantsReference device Atlantis® Abutment in Titanium for OsseoSpeed implants (K051652)Predicate device, Atlantis® Abutment in Titanium for MIS implants V3 and C1 (K193529)Discussion
angulationThe ASA screw access channel ranges from 0 to 30°.
Screw drive featureHexalobular HexHexHexDifferent due to the hexalobular screw drive feature for Atlantis ASA screwdriver.
MR SafetyMR ConditionalMR Not EvaluatedMR Not EvaluatedDifferent. Testing included in reference device (K221094) supports labeling update

7. Non-Clinical Tests Summary and Conclusion:

To support the labeling change from MR Not Evaluated or no MR information on the labeling to MR Conditional labeling, testing included in reference device, OmniTaper EV Dental Implants, DS Implants with EV connection (K221094) is included by reference. The proposed Atlantis® Abutments in Titanium meet the requirements of:

  • ASTM F2052-21, Standard test method for measurement of magnetically induced displacement force on medical devices in the magnetic resonance environment
  • ASTM F2213-17, Standard test method for measurement of magnetically induced torque on medical devices in magnetic resonance environment
  • ASTM F2119-07 (2013), Standard test method for evaluation of MR image artifacts from passive implants,
  • ASTM F2503-20, Standard practice for marking medical devices and other items for safety in the magnetic resonance environment,
  • FDA guidance document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" dated October 10, 2023

Biocompatibility assessment was performed to evaluate the biocompatibility profile of the Atlantis® Abutments in Titanium according to the following standards:

  • ISO 10993-1:2018, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
  • Cytotoxicity testing was performed according to ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

Reprocessing testing was performed to validate the proposed drying time for steam sterilization. The sterilization drying time validation followed the FDA guidance document titled "Reprocessing Medical Device in Health Care Settings: Validation Methods and Labeling".

  • Sterilization and drying time validation was performed according to:
    • ANSI/AAMI/ISO 17665-1:2006/(R)2013, Annex D, Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
    • ANSI/AAMI/ISO 14937:2009/(R)2013, Annex D (Approach 3), Sterilization of health care products – General requirements for characterization of a sterilizing agent and the development, validation, and routing control of a sterilization process
    • AAMI TIR12:2020, Designing, testing and labeling medical devices intended for processing by health care facilities: A guide for manufacturers
    • ANSI/AAMI/ISO TIR17665-2:2009, Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO ANSI/AAMI/ISO 17665-1.

CAD/CAM Software Validation & Verification Testing for Atlantis® Abutment Milling:
Software validation was performed to show that the Atlantis® abutment design limitations were implemented within the 510(k) cleared abutment design software (3Shape Abutment Designer Software, K151455).

The internally documented and proposed modifications to the Atlantis® Abutments in Titanium and screws did not affect the performance of the devices and therefore no new fatigue testing is

Page 49 of 50

Page 51

Page 45 of 50

included in this submission. The abutments continue to meet the requirements of ISO 14801:2016, Dentistry – Implants – Dynamic loading test for endosseous dental implants and FDA Guidance, Root-form Endosseous Dental Implants and Endosseous Dental Abutments – Class II Special Controls Guidance Document.

The testing described above confirmed that the proposed Atlantis® Abutments in Titanium are substantially equivalent to the predicate device (K193529) and reference devices (K161030, K151039, K081666, K073258, K070833, K062197, K062069, K051652, K051011, K981858) and do not raise new questions regarding safety and effectiveness.

8. Clinical Tests Summary and Conclusion:

Not applicable. There are no clinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence.

9. Conclusion Regarding Substantial Equivalence:

The proposed Atlantis® Abutments in Titanium have the same intended use and similar indications as compared to the predicate device Atlantis® Conical Connection Abutment for MIS Implant (K193529). The abutments labeling is modified to conform to FDA guidance document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" and testing performed on the worst-case Dentsply Sirona implant products validates the labeling change to indicate MR Conditional.

In addition to the MR labeling change which is the main purpose of this submission, additional internally documented changes and proposed changes were reviewed against the original clearance of the predicate (K193529) and reference devices (K161030, K151039, K081666, K073258, K070833, K062197, K062069, K051652, K051011, K981858) and the changes do not raise new questions about safety and effectiveness. Test data is included in this premarket notification to verify and validate the changes to the proposed devices and the results support a conclusion of substantial equivalence.

Page 50 of 50

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)