K Number
K241485

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-10-09

(138 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TruAbutment DS is a patient-specific CAD/CAM abutment, which is directly connected to endosseous dental implants and is intended to be used as an aid in prosthetic rehabilitation. It is compatible with the following systems: Astra OsseoSpeed EV (K130999, K120414), Biomet 3i Full OSSEOTITE Tapered Certain (K130949), DIO UF (II) Internal Submerged (K161987, K170608, K173975), Neoss ProActive® (K083561), Osstem TS (K161604), Camlog Screw-Line (K083496), Conelog Screw-Line (K113779), Implant Direct Legacy2 (K192221), BioHorizons Internal Implant System (K093321, K143022, K071638), MegaGen AnyRidge Internal Implant (K140091). All digitally designed abutments and/or copings for use with the TruAbutments are intended to be sent to a TruAbutment-validated milling center for manufacture.

TruBase is a titanium component that is directly connected to endosseous dental implants to provide support for patient-specific prosthetic restorations, such as copings or crowns. It is indicated for a screw-retained single tooth or cement-retained single tooth and bridge restorations. It is compatible with the following systems: Astra OsseoSpeed EV (K130999), Biomet 3i Full OSSEOTITE Tapered Certain (K130949), DIO UF(II) Internal Submerged (K161987, K170608, K173975), Neoss ProActive® (K083561), Camlog Screw-Line (K083496), Conelog Screw-Line (K113779), Implant Direct Legacy2 (K192221). All digitally designed abutments and/or copings for use with the TruAbutment are intended to be sent to a TruAbutment-validated milling center for manufacture.

Device Description

TruAbutment DS, TruBase and abutment screw are made of Titanium grade Ti-6A1-4V ELI (meets ASTM Standard F136). TruAbutment DS, TruBase are supplied with two identical screws which are used: (1) For fixing the abutment into the endosseous implant. (2) For dental laboratory use during construction of related restoration. TruAbutment DS, TruBase are provided non-sterile. Therefore, it must be sterilized before use. TruAbutment DS, TruBase are devices that can only be sold, distributed, or used upon the order of an authorized healthcare provider, generally referred to as prescription (Rx) devices.

TruAbutment DS system includes patient-specific abutments that are placed into the dental implant to provide support for the prosthetic restoration. The subject abutments are indicated for serew-retained restorations. The design and manufacturing of the patient-specific abutments take into consideration the shape of the final prosthesis based on the patient's intra-oral indications using CAD/CAM system during the manufacturing. All manufacturing processes of TruAbutment DS are conducted at the TruAbutment milling center.

TruBase is a two-piece abutment. The base component is premanufactured and is used to support a cemented CAD/CAM zirconia superstructure. The base and the zirconia superstructure together form the final abutment. CAD/CAM customized superstructure that composes the final abutment is intended to be sent to a TruAbutment-validated milling center to be designed and milled, according to the prosthetic planning and patient clinical situation. The superstructure is cemented to the TruBase in the lab. Use "RelyX Unicem 2Automix" as an adhesive extra orally to connect.

AI/ML Overview

The provided text is a 510(k) summary for the TruAbutment DS and TruBase devices. It primarily focuses on demonstrating substantial equivalence to a predicate device (TruAbutment DS, K203649) and does not detail an acceptance criteria table with reported device performance in the manner of a clinical study. The text describes non-clinical testing performed, but not a study designed to prove the device meets acceptance criteria related to a specific clinical outcome or diagnostic accuracy.

Therefore, many of the requested items (acceptance criteria table, sample size for test/training sets, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance, type of ground truth) are not applicable based on the content of this 510(k) summary, which is a premarket notification for a medical device primarily based on demonstrating substantial equivalence through engineering and mechanical testing, not clinical performance or AI algorithm validation studies.

However, I can extract the information provided regarding non-clinical testing for the devices.


Acceptance Criteria and Study for TruAbutment DS & TruBase

Based on the provided 510(k) summary, the "acceptance criteria" and "study" described are focused on non-clinical mechanical performance testing and demonstration of substantial equivalence to a predicate device, rather than a clinical study proving performance against specific clinical or diagnostic accuracy metrics with human or AI components.

Here's the relevant information extracted and presented based on the document:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table with specific quantitative acceptance criteria alongside actual reported numerical performance results for the new devices in the context of a comparative study proving their performance against such criteria. Instead, it states that "The results of the above tests have met the criteria of the standard and demonstrated substantial equivalence with the reference devices." This implies a qualitative "met standard" outcome rather than specific numerical performance data.

The tables provided describe the design limits of the devices and compare them to the predicate device, not performance data from a test:

TruAbutment DS Design Parameters (Acceptance Criteria are implied by meeting these limits)

Design ParameterSubject Device (TruAbutment DS) Design LimitPrimary Predicate Device (K203649) Design LimitReported Device Performance (Implied)
Minimum and Maximum abutment angle (°)0 ~ 250 ~ 25Met specified range
Minimum and Maximum cuff height (mm)0.5 ~ 6.00.5 ~ 6.0Met specified range
Minimum and Maximum diameter at abutment/implant interface (Ø, mm)3.3 ~ 8.03.3 ~ 8.0Met specified range
Minimum and Maximum length of the abutment (mm)6 ~ 116 ~ 11Met specified range
Minimum wall thickness at abutment/implant interface (mm)0.40.4 ~ 0.9Met specified range
Minimum and Maximum length of abutment post (length above the abutment collar / gingival height) (mm)4.0 ~ 7.04.0 ~ 7.0Met specified range

TruBase Design Parameters (Acceptance Criteria are implied by meeting these limits)

Design ParameterSubject Device (TruBase) Design LimitPrimary Predicate Device (K203649) Design LimitReported Device Performance (Implied)
Minimum and Maximum angulation (°)0 ~ 150 ~ 15Met specified range
Minimum and Maximum gingival (cuff) height (mm)0.5 ~ 5.00.5 ~ 5.0Met specified range
Minimum and Maximum diameter at abutment/implant interface (Ø, mm)5.0 ~ 8.05.0 ~ 8.0Met specified range
Minimum thickness (mm)0.40.4Met specified value
Minimum and Maximum length of abutment post (length above the abutment collar / gingival height) (mm)4.0 ~ 6.04.0 ~ 6.0Met specified range

For mechanical performance, the document states:
"Mechanical performance testing was performed according to ISO 14801. For compatible OEM implant line, worst-case constructs were subjected to static compression and compression fatigue testing. The fatigue limit data for all other implant lines demonstrated the construct strengths to be sufficient for their intended use."

This confirms that the acceptance criteria for mechanical performance were "sufficient for their intended use" as defined by ISO 14801 and worst-case testing, but quantitative results are not provided.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: The document mentions "worst-case constructs" were tested for mechanical performance, and "the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws)" was evaluated for MRI environment conditions. However, specific numerical sample sizes for these tests are not provided.
  • Data Provenance: The data comes from non-clinical laboratory testing following international standards (ISO 14801, ISO 17665-1/2, ISO 10993 series). The country of origin and retrospective/prospective nature are not applicable as it's not a clinical data study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • This is not applicable. The "ground truth" for this type of device (dental abutments) is established through adherence to engineering design specifications, material standards (ASTM F136), and performance under mechanical stress tests (ISO 14801), as well as compliance with sterilization and biocompatibility standards. It does not involve expert interpretation of images or clinical outcomes in the same way an AI diagnostic device would.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • This is not applicable as there is no human interpretation or subjective assessment of data requiring adjudication. Testing is based on objective measurements against engineering standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This is not applicable. The device is an endosseous dental implant abutment, not an AI diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This is not applicable. The device is a physical dental component, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for validating these devices is adherence to engineering specifications, material properties, and performance standards (e.g., passing specific load-bearing and fatigue tests per ISO 14801, meeting biocompatibility requirements, maintaining dimensional accuracy). "Dimensional analysis and reverse engineering" were used to confirm compatibility.

8. The sample size for the training set

  • This is not applicable. There is no "training set" as this is a physical medical device, not an AI/machine learning algorithm.

9. How the ground truth for the training set was established

  • This is not applicable for the reason above.

Summary of the Study Proving Device Meets Criteria (Based on Provided Text):

The "study" conducted was primarily non-clinical laboratory testing to demonstrate the mechanical performance, sterilization efficacy, and biocompatibility of the TruAbutment DS and TruBase devices. The objective was to show substantial equivalence to an existing legally marketed device (TruAbutment DS, K203649) by proving that the new devices meet established design limits and performance standards relevant to dental implant abutments.

  • Mechanical Testing: Performed on "worst-case constructs" according to ISO 14801 for static compression and compression fatigue. The outcome was that "construct strengths [were] sufficient for their intended use."
  • Sterilization Testing: Performed per ISO 17665-1:2006, 17665-2:2009 and ANSI/AAMI ST79:2010.
  • Biocompatibility Testing: Performed per ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.
  • MRI Environment Evaluation: A non-clinical worst-case MRI review was done using scientific rationale and published literature to assess magnetically induced displacement force and torque.
  • Dimensional Analysis and Reverse Engineering: Conducted on the implant-to-abutment connection platform to assess critical design aspects and tolerances, confirming compatibility.

The overall conclusion was that the devices "met the criteria of the standard and demonstrated substantial equivalence with the reference devices," thus indicating they met their implied acceptance criteria for safety and performance as medical devices. Clinical testing was explicitly stated as "not necessary."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA part of the logo is in blue.

October 9, 2024

TruAbutment Inc. Ki Yoon Nam RA Senior Manager 17666 Fitch Irvine, California 92614

Re: K241485

Trade/Device Name: TruAbutment DS; TruBase Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: September 12, 2024 Received: September 12, 2024

Dear Ki Yoon Nam:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K241485

Device Name

TruAbutment DS; TruBase

Indications for Use (Describe)

{4}------------------------------------------------

TruAbutment DS

TruAbutment DS is a patient-specific CAD/CAM abutment, which is directly connected to endosseous dental implants and is intended to be used as an aid in prosthetic rehabilitation. It is compatible with the following systems:

· Astra OsseoSpeed EV (K130999, K120414)

  • : 3.0, 3.6, 4.2, 4.8, 5.4 mm.
  • · Biomet 3i Full OSSEOTITE Tapered Certain (K130949)
  • : 3.25, 4.0, 5.0, 6.0 mm.
  • · DIO UF (II) Internal Submerged (K161987, K170608, K173975)

: 3.3, 3.8, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0 (Narrow, Regular) mm.

  • · Neoss ProActive® (K083561)
  • : 3.25, 3.5, 4.0, 4.5, 5.0, 5.5, 6.0 (NP, SP) mm.
  • · Osstem TS (K161604)
  • : 3.0, 3.5 (3.7), 4.0 (4.2), 4.5 (4.6), 5.0 (5.1), 6.0 (6.0), 7.0 (6.8) (Mini, Regular) mm.
  • · Camlog Screw-Line (K083496)
  • : 3.3, 3.8, 4.3, 5.0, 6.0 mm.
  • · Conelog Screw-Line (K113779)
  • : 3.3, 3.8, 4.3, 5.0 mm.
  • · Implant Direct Legacy2 (K192221)
  • : 3.2 mm.
  • · BioHorizons Internal Implant System (K093321, K143022, K071638)
  • : 3.0, 3.4, 3.8, 4.6, 5.8 mm.
  • · MegaGen AnyRidge Internal Implant (K140091)
  • : 3.5, 4.0, 4.5, 5.0, 5.5 (3.5) mm.

All digitally designed abutments and/or copings for use with the TruAbutments are intended to be sent to a TruAbutment-validated milling center for manufacture.

TruBase

TruBase is a titanium component that is directly connected to endosseous dental implants to provide support for patient-specific prosthetic restorations, such as copings or crowns. It is indicated for a screw-retained single tooth or cement-retained single tooth and bridge restorations. It is compatible with the following systems:

  • · Astra OsseoSpeed EV (K130999): 3.0 mm.
  • · Biomet 3i Full OSSEOTITE Tapered Certain (K130949): 3.25, 4.0, 5.0, 6.0 mm.
  • · DIO UF(II) Internal Submerged (K161987, K170608, K173975): 3.3, 3.8, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0 (Narrow, Regular) mm.
  • · Neoss ProActive® (K083561): 3.25, 3.5, 4.0, 4.5, 5.0, 5.5, 6.0 (NP, SP) mm.
  • · Camlog Screw-Line (K083496); 3.3. 3.8. 4.3. 5.0. 6.0 mm.
  • · Conelog Screw-Line (K113779): 3.3, 3.8, 4.3, 5.0 mm.
  • · Implant Direct Legacy2 (K192221): 3.2 mm.

All digitally designed abutments and/or copings for use with the TruAbutment are intended to be sent to a TruAbutment-validated milling center for manufacture.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for a company called "Abutment ru". The logo is in blue and features the word "Abutment" in a bold, sans-serif font, stacked on top of the letters "ru", which are also in a bold, sans-serif font. To the left of the word "ru" is a graphic that resembles the threads of a screw or bolt, which may be a visual representation of the word "abutment".

510(k) Summary K241485

SubmitterOfficial Correspondent
TruAbutment Inc.TruAbutment Inc.
Rawa AbukhalafKi Yoon Nam
17666 Fitch,17666 Fitch,
Irvine, CA 92614 USAIrvine, CA 92614 USA
Email: rawa.abukhalaf@truabutment.comEmail: kiyoon.nam@truabutment.com
Phone: 1-714-956-1488Phone: 1-714-956-1488

Device Information

  • . Trade Name: TruAbutment DS, TruBase
  • Common Name: Abutment, Implant, Dental, Endosseous
  • Classification Name: Endosseous dental implant abutment
  • Product Code: NHA
  • Panel: Dental
  • Regulation Number: 21 CFR 872.3630
  • Device Class: Class II
  • Date prepared: 10/09/2024. ●

Primary Predicate / Reference Devices:

The subject device is substantially equivalent to the following predicate device:

Primary Predicate:

  • . TruAbutment DS (K203649) by TruAbutment Inc.

Reference Devices for OEM Compatibilities:

  • Astra OsseoSpeed EV (K130999, K120414) by Astra Tech, Inc. ●
  • Biomet 3i Full OSSEOTITE Tapered Certain (K130949) by Zimmer Biomet. ●
  • DIO UF (II) Internal Submerged (K161987, K170608, K173975) by DIO Implant Co., Ltd. ●
  • Neoss ProActive® (K083561) by Neoss Limited. ●
  • Camlog Screw-Line (K083496) by CAMLOG Biotechnologies GmbH. .
  • Conelog Screw-Line (K113779) by CAMLOG Biotechnologies GmbH.
  • Osstem TS Fixture System (K161604) by Osstem Implant Co., Ltd. ●
  • BioHorizons Laser-Lok 3.0 Implant System (K093321) by Biohorizons Implant Systems, Inc. .

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a logo for a company called "Abutment ru". The logo is primarily blue and features the word "Abutment" in a bold, sans-serif font, stacked above the letters "ru", which are in a smaller, cursive font. To the left of the text, there is a stylized image of a screw, also in blue, which appears to be part of the "A" in "Abutment".

  • Biohorizons Tapered Internal Implant System (K143022) by Biohorizons Implant Systems, Inc. .
  • Biohorizons Tapered Internal Implant System (K071638) by Biohorizons Implant Systems, Inc. ●
  • . Xpeed AnyRidge Internal Implant System (K140091) by MegaGen Implant Co., Ltd.
  • Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, . Legacy4 fixture-mounts (K192221) by Implant Direct LLC.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image is a logo for a company called "Abutment ru". The word "Abutment" is in a sans-serif font and is placed above the letters "ru", which are in a larger, bolder sans-serif font. To the left of the letters "ru" is a graphic of a screw. The entire logo is in a light blue color.

General Description

TruAbutment DS, TruBase and abutment screw are made of Titanium grade Ti-6A1-4V ELI (meets ASTM Standard F136). TruAbutment DS, TruBase are supplied with two identical screws which are used:

  • (1) For fixing the abutment into the endosseous implant.
  • (2) For dental laboratory use during construction of related restoration.

TruAbutment DS, TruBase are provided non-sterile. Therefore, it must be sterilized before use.

TruAbutment DS, TruBase are devices that can only be sold, distributed, or used upon the order of an authorized healthcare provider, generally referred to as prescription (Rx) devices.

Mechanical resistance of the implant-abutment connection is essential to ensure the correct long-term functional performance of the complete dental restoration. Dimensional compatibility and mechanical performance of bases and screws together with the underlying implant are of primary importance. These concepts are the basis upon which the system design characteristics and functional performance are established.

The following table shows the subject device abutment platform sizes for each of the OEM implant lines and sizes.

TruAbutment PlatformDiameter / CompatibleImplant SystemTruAbutment DSEngagingTruAbutmentDSNon-EngagingTruBaseEngagingTruBaseNon-Engaging
Astra OsseoSpeed EV
3.0X000
3.6X0XX
4.2X0XX
4.8X0XX
5.4X0XX
Biomet 3i FullOSSEOTITE TaperedCertain
3.4X000
4.1X000
5.0X000
6.0X000
DIO UF (II) Internal
NarrowX000
RegularX000
Neoss ProActive®
NPX000
SPX000
Camlog Screw-Line
3.30000
3.80000

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows a logo for a company called "Abutment ru". The logo is primarily blue and features the word "Abutment" in a stylized font above the letters "ru". To the left of the letters "ru" is a graphic that resembles the side view of a screw or bolt, also in blue. The overall design is simple and modern, with a focus on the company's name and a visual element related to engineering or construction.

4.30000
5.00000
6.00000
Conelog Screw-Line
3.30000
3.80000
4.30000
5.00000
Implant Direct Legacy2
3.20000
Osstem TS
MiniX0XX
RegularX0XX
BioHorizons Internal
3.0X0XX
3.5X0XX
4.5X0XX
5.7X0XX
MegaGen Xpeed
AnyRidge Internal
3.5X0XX

X: Not included in this submission

O: Included in this submission

TruAbutment DS

TruAbutment DS system includes patient-specific abutments that are placed into the dental implant to provide support for the prosthetic restoration. The subject abutments are indicated for serewretained restorations.

The design and manufacturing of the patient-specific abutments take into consideration the shape of the final prosthesis based on the patient's intra-oral indications using CAD/CAM system during the manufacturing. All manufacturing processes of TruAbutment DS are conducted at the TruAbutment milling center.

Design Limitation for TruAbutment DS

Design ParameterDesign Limit
Minimum and Maximum abutment angle (°)0 ~ 25
Minimum and Maximum cuff height (mm)0.5 ~ 6.0
Minimum and Maximum diameter at abutment/implant interface (Ø, mm)3.3 ~ 8.0
Minimum thickness (mm)0.4
Minimum and Maximum length of abutment post(length above the abutment collar / gingival height) (mm)4.0 ~ 7.0

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows a logo for a company called "Abutment ru". The logo is primarily blue and features the word "Abutment" in a bold, sans-serif font, stacked above the letters "ru", which are in a smaller, cursive font. The "A" in "Abutment" is stylized to resemble a screw or bolt, possibly indicating the company's focus or industry.

TruBase

TruBase is a two-piece abutment. The base component is premanufactured and is used to support a cemented CAD/CAM zirconia superstructure. The base and the zirconia superstructure together form the final abutment.

CAD/CAM customized superstructure that composes the final abutment is intended to be sent to a TruAbutment-validated milling center to be designed and milled, according to the prosthetic planning and patient clinical situation. The superstructure is cemented to the TruBase in the lab. Use "RelyX Unicem 2Automix" as an adhesive extra orally to connect.

Raw material blanks

  • InCoris Zi (ZrO2) by Sirona Dental Systems GmbH, L size blanks, cleared under K123664.

Cement

  • . RelyX Unicem 2Automix by 3M ESPE, cleared under K100756.
    Design Limitation for Zirconia superstructure:
Design ParameterDesign Limit
Minimum and Maximum angulation (°)0~15
Minimum and Maximum cuff height (mm)0.5 ~ 5.0
Minimum and Maximum diameter at abutment/implant interface (Ø, mm)5.0 ~ 8.0
Minimum thickness (mm)0.4
Minimum and Maximum length of abutment post(length above the abutment collar / gingival height) (mm)4.0 ~ 6.0

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image is a logo for a company called "Abutment ru". The logo is primarily blue and features the word "Abutment" in a bold, sans-serif font stacked above the letters "ru", which are stylized to appear as if they are connected. To the left of the text, there is a graphic of a screw, also in blue, which adds a visual element related to construction or engineering.

Indication for Use

TruAbutment DS

TruAbutment DS is a patient-specific CAD/CAM abutment, which is directly connected to endosseous dental implants and is intended to be used as an aid in prosthetic rehabilitation. It is compatible with the following systems:

  • · Astra OsseoSpeed EV (K130999, K120414) : 3.0, 3.6, 4.2, 4.8, 5.4 mm.
  • · Biomet 3i Full OSSEOTITE Tapered Certain (K130949) : 3.25, 4.0, 5.0, 6.0 mm.
  • · DIO UF (II) Internal Submerged (K161987, K170608, K173975) : 3.3, 3.8, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0 (Narrow, Regular) mm.
  • · Neoss ProActive® (K083561) : 3.25, 3.5, 4.0, 4.5, 5.0, 5.5, 6.0 (NP, SP) mm.
  • · Osstem TS (K161604) : 3.0, 3.5 (3.7), 4.0 (4.2), 4.5 (4.6), 5.0 (5.1), 6.0 (6.0), 7.0 (6.8) (Mini, Regular) mm.
  • · Camlog Screw-Line (K083496) : 3.3, 3.8, 4.3, 5.0, 6.0 mm.
  • · Conelog Screw-Line (K113779) : 3.3, 3.8, 4.3, 5.0 mm.
  • · Implant Direct Legacy2 (K192221) : 3.2 mm.
  • · BioHorizons Internal Implant System (K093321, K143022, K071638) : 3.0. 3.4. 3.8. 4.6. 5.8 mm.
  • · MegaGen AnyRidge Internal Implant (K140091) : 3.5, 4.0, 4.5, 5.0, 5.5 (3.5) mm.

All digitally designed abutments and/or copings for use with the TruAbutments are intended to be sent to a TruAbutment-validated milling center for manufacture.

TruBase

TruBase is a titanium component that is directly connected to endosseous dental implants to provide support for patient-specific prosthetic restorations, such as copings or crowns. It is indicated for a screw-retained single tooth or cement-retained single tooth and bridge restorations. It is compatible with the following systems:

  • . Astra OsseoSpeed EV (K130999): 3.0 mm.
  • Biomet 3i Full OSSEOTITE Tapered Certain (K130949): 3.25.4.0, 5.0, 6.0 mm.
  • DIO UF(II) Internal Submerged (K161987, K170608, K173975): 3.3, 3.8, 4.0, 4.5, 5.0, 5.5, 6.0, . 6.5, 7.0 (Narrow, Regular) mm.
  • . Neoss ProActive® (K083561): 3.25, 3.5, 4.0, 4.5, 5.0, 5.5, 6.0 (NP, SP) mm.
  • Camlog Screw-Line (K083496): 3.3, 3.8, 4.3, 5.0, 6.0 mm.
  • Conelog Screw-Line (K113779): 3.3, 3.8, 4.3, 5.0 mm.
  • Implant Direct Legacy2 (K192221): 3.2 mm. ●

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows a logo for a company called "Abutment ru". The logo is blue and white. The word "Abutment" is written in white letters on a blue background. The letters "ru" are written in blue letters on a white background. The logo has a jagged edge on the left side.

All digitally designed abutments and/or copings for use with the TruAbutments are intended to be sent to a TruAbutment-validated milling center for manufacture.

ImplantSystemImplantBodyDiameter(mm)ImplantLength(mm)Model No.ImplantPlatformDiameter(mm)Type of Implant-AbutmentConnection
AstraOsseoSpeedEV(K130999)(K120414)3.08263013.0Internal Splineconnection"Index System"
926302
1126303
1326304
1526305
3.6626310
826311
9263123.6
1126313
1326314
1526315
1726316
4.2626320
8263214.2
926322
1126323
1326324
1526325
1726326
4.86263404.8
826341
926342
1126343
1326344
1526345
1726346
5.46263605.4
826361
926362
1126363
1326364
1526365
3.258.5IENT32853.4
ImplantSystemImplantBodyDiameter(mm)ImplantLength(mm)Model No.ImplantPlatformDiameter(mm)Type of Implant-AbutmentConnection
Biomet 3i FullOSSEOTITETaperedCertain(K130949)4.010IFNT32104.1Image: Internal Hex Connection
11.5IFNT3211
13IFNT3213
15IFNT3215
8.5IFNT485
10IFNT410
11.5IFNT411
13IFNT413
15IFNT415
5.08.5IFNT5855.0Image: Internal Hex Connection
10IFNT510
11.5IFNT511
13IFNT513
15IFNT515
6.08.5IFNT685
10IFNT610
11.5IFNT611
6.0
13IFNT613
15IFNT615
DIO UF (II)InternalSubmergedImplant(K161987)(K170608)(K173975)3.38.5UF(II)N 3308SNarrow(K161987)Image: Internal Hex Connection
10UF(II)N 3310S
11.5UF(II)N 3311S
13UF(II)N 3313S
15UF(II)N 3315S
3.88.5UF(II) 3808S
10UF(II) 3810S
11.5UF(II) 3811S
13UF(II) 3813S
15UF(II) 3815SRegular(K170608)
16UF(II) 3816S
4.07UF(II) 4007S
8.5UF(II) 4008S
10UF(II) 4010S
11.5UF(II) 4011S
13UF(II) 4013S
15UF(II) 4015S
ImplantSystemImplantBodyDiameter(mm)ImplantLength(mm)Model No.ImplantPlatformDiameter(mm)Type of Implant-AbutmentConnection
4.516UF(II) 4016S
7UF(II) 4507S
8.5UF(II) 4508S
10UF(II) 4510S
11.5UF(II) 4511S
13UF(II) 4513S
15UF(II) 4515S
16UF(II) 4516S
5.07UF(II) 5007S
8.5UF(II) 5008S
10UF(II) 5010S
11.5UF(II) 5011S
13UF(II) 5013S
15UF(II) 5015S
16UF(II) 5016S
5.57UF(II) 5507S
8.5UF(II) 5508S
10UF(II) 5510S
11.5UF(II) 5511S
13UF(II) 5513S
15UF(II) 5515S
16UF(II) 5516S
6.07UF(II) 6007S
8.5UF(II) 6008S
10UF(II) 6010S
11.5UF(II) 6011S
13UF(II) 6013S
6.57UF(II) 6507SWide(K173975)
8.5UF(II) 6508S
10UF(II) 6510S
11.5UF(II) 6511S
13UF(II) 6513S
7.07UF(II) 7007S
8.5UF(II) 7008S
10UF(II) 7010S
ImplantSystemImplantBodyDiameter(mm)ImplantLength(mm)Model No.ImplantPlatformDiameter(mm)Type of Implant-AbutmentConnection
NeossProActive®(K083561)3.2513UF(II) 7013S3.5(NP)Image: Internal Hex Connection
921176
1121177
1321178
1521179
3.5721181
921182
1121183
1321184
1521185
1721186
4.07211874.0(SP)
921188
1121189
1321190
1521191
1721192
4.5721193
921194
1121195
1321196
1521197
1721198
5.0721199
921200
1121201
1321202
1521203
1721205
5.5921206
1121207
1321208
1521211
6.0921212
1121213
1321221
ImplantSystemImplantBodyDiameter(mm)ImplantLength(mm)Model No.ImplantPlatformDiameter(mm)Type of Implant-AbutmentConnection
3.01521222Image: Internal Hex connection
8.5TS3M3008S2.1(Mini)
10TS3M3010S
11.5TS3M3011S
13TS3M3013S
3.5 (3.7)8.5TS3M3508S
10TS3M3510S
11.5TS3M3511S
13TS3M3513S
4.0 (4.2)7TS3S4007S
8.5TS3S4008S
10TS3S4010S
11.5TS3S4011S
13TS3S4013S
OSSTEM TS(K161604)4.5 (4.6)7TS3S4507S
8.5TS3S4508S
10TS3S4510S
11.5TS3S4511S
13TS3S4513S
5.0 (5.1)7TS3S5007S
8.5TS3S5008S
10TS3S5010S
11.5TS3S5011S
13TS3S5013S
6.0 (6.0)7TS3S6007S
8.5TS3S6008S
10TS3S6010S
11.5TS3S6011S
13TS3S6013S
7.0 (6.8)7TS3S7007S
8.5TS3S7008S
10TS3S7010S
11.5TS3S7011S
13TS3S7013S
3.311K1046.33113.3
13K1046.3313
ImplantSystemImplantBodyDiameter(mm)ImplantLength(mm)Model No.ImplantPlatformDiameter(mm)Type of Implant-AbutmentConnection
CamlogScrew-Line(K083496)3.816K1046.33163.8Tube-in-tubeconnection
3.89K1046.3809
3.811K1046.3811
3.813K1046.3813
4.316K1046.38164.3
4.39K1046.4309
4.311K1046.4311
4.313K1046.4313
CamlogScrew-Line(K083496)4.316K1046.43165.0
5.09K1046.5009
5.011K1046.5011
5.013K1046.5013
6.016K1046.50166.0
6.09K1046.6009
6.011K1046.6011
6.013K1046.6013
6.016K1046.60166.0Conical abutmentconnectoin
ConelogScrew-Line(K113779)3.39C1065.3309
3.311C1065.3311
3.313C1065.3313
3.316C1065.3316
3.87C1065.38073.8
3.89C1065.3809
3.811C1065.3811
3.813C1065.3813
ConelogScrew-Line(K113779)3.816C1065.38164.3
4.37C1065.4307
4.39C1065.4309
4.311C1065.4311
5.04.313C1065.43135.0
5.016C1065.4316
5.07C1065.5007
5.09C1065.5009
5.05.011C1065.50115.0
5.05.013C1065.5013
5.05.016C1065.5016
ImplantSystemImplantBodyDiameter(mm)ImplantLength(mm)Model No.ImplantPlatformDiameter(mm)Type of Implant-AbutmentConnection
ImplantDirectLegacy2(K192221)3.288232083.0Image: Internal Hex Connection
10823210
11.5823211
13823213
16823216
BioHorizonsLaser-Lok(K093321)3.010.5TLX30103.0Image: Internal Hex Connection
12TLX3012
15TLX3015
3.49TLX34093.0
10.5TLX3410
12TLX3412
15TLX3415
3.418TLX34183.0
3.89TLX38093.5
10.5TLX3810
12TLX3812
15TLX3815
3.818TLX38183.5
BioHorizonsTaperedInternal(K071638)(K143022)4.67.5TLX46074.5
9TLX4609
10.5TLX4610
12TLX4612
15TLX4615
18TLX4618
7.5TLX5807
9TLX5809
5.810.5TLX58105.7
9TLX5809
10.5TLX5810
12TLX5812
15TLX5815
18TLX5818
7FANIHX3507C
8FANIHX3508C
ImplantSystemImplantBodyDiameter(mm)ImplantLength(mm)Model No.ImplantPlatformDiameter(mm)Type of Implant-AbutmentConnection
MegaGenXpeedAnyRidge®(K140091)3.510FANIHX3510C3.5Image: Internal Hex Connection
11.5FANIHX3511C
13FANIHX3513C
15FANIHX3515C
18FANIHX3518C
4.07FANIHX4007C
8FANIHX4008C
10FANIHX4010C
11.5FANIHX4011C
13FANIHX4013C
15FANIHX4015C
4.57FANIHX4507C
8FANIHX4508C
10FANIHX4510C
11.5FANIHX4511C
13FANIHX4513C
15FANIHX4515C
18FANIHX4518C
5.07FANIHX5007C
8FANIHX5008C
10FANIHX5010C
11.5FANIHX5011C
13FANIHX5013C
15FANIHX5015C
18FANIHX5018C
5.57FANIHX5507C
8FANIHX5508C
10FANIHX5510C
11.5FANIHX5511C
13FANIHX5513C
15FANIHX5515C
18FANIHX5518C

TruAbutment DS and TruBase are compatible with the following OEM devices:

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image features a logo for "Abutment ru" in a stylized, modern design. The word "ABUTMENT" is prominently displayed in a bold, sans-serif font, with the "A" slightly larger than the other letters. To the left of the word, there is a graphic resembling a screw or threaded cylinder, suggesting a connection to engineering or construction. The letters "ru" are placed to the right of "ABUTMENT" in a smaller, italicized font, possibly indicating a website domain or regional affiliation.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows a logo for a company called "Abutment ru". The word "Abutment" is in a larger, bold, sans-serif font, with the "A" in blue and the rest of the word in white. The letters "ru" are in a smaller, cursive font, also in blue, and are positioned to the right and slightly below the word "Abutment". The logo has a tooth-like design incorporated into the "A" of "Abutment", suggesting a dental or medical focus.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image features a logo for "Abutment ru" in a stylized, modern design. The word "ABUTMENT" is prominently displayed in a bold, sans-serif font, with the "A" slightly larger than the other letters. To the left of the word, there is a graphic resembling a screw or threaded cylinder, suggesting a connection to engineering or construction. The letters "ru" are placed to the right of "ABUTMENT" in a smaller, italicized font, possibly indicating a website domain or regional affiliation.

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows a logo for a company called "Abutment ru". The word "Abutment" is in a larger, bold, sans-serif font and is colored in a bright blue. To the left of the word "ru" is a graphic of a screw, also in blue, which is likely meant to represent a dental implant abutment.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows a logo for a company called "Abutment ru". The word "Abutment" is in a sans-serif font, with the "A" being larger than the other letters. The letters "ru" are in a larger, bolder font and are placed to the right of the word "Abutment". To the left of the word "Abutment" is a graphic of a screw.

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows a logo for a company called "Abutment ru". The word "Abutment" is in a sans-serif font, with the "A" being larger than the other letters. The letters "ru" are in a larger, bolder font and are placed to the right of the word "Abutment". To the left of the word "Abutment" is a graphic of a screw.

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows a logo for a company called "Abutment ru". The word "Abutment" is in a bold, sans-serif font, with the letters "ABUTMENT" stacked on top of each other. The letters "ru" are in a smaller, sans-serif font and are located to the right of the word "Abutment". The entire logo is in a bright blue color. The bottom of the logo has a design that looks like a screw.

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image shows a logo for a company called "Abutment ru". The logo is in a bright blue color and features the word "Abutment" in a bold, sans-serif font. Below the word "Abutment" is the letters "ru" in a larger, stylized font. The logo has a modern and professional look.

Summary of Technological Characteristic

The subject base and blank devices are identical to the K203649 base and blank devices besides the compatible implant bodies. They are substantially equivalent in intended use, the material and connection interfaces to the implants identical for each diameter and connection type. A comparison demonstrating substantial equivalence follows at the end of this section.

TruAbutment DS

AttributesProposed DevicePrimary Predicate Device
Trade NameTruAbutment DSTruAbutment DS (K203649)
Indications for UseTruAbutment DS is a patient-specific CAD/CAM abutment,directly connected to endosseous dental implants and isintended for use as an aid in prosthetic rehabilitation. It iscompatible with the following systems:Astra OsseoSpeed EV (K130999, K120414): 3.0, 3.6, 4.2, 4.8, 5.4 mm.Biomet 3i Full OSSEOTITE Tapered Certain (K130949): 3.25, 4.0, 5.0, 6.0 mm.DIO UF(II) Internal Submerged (K161987, K170608,K173975): 3.3, 3.8, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0 (Narrow, Regular)Neoss ProActive® (K083561): 3.25, 3.5, 4.0, 4.5, 5.0, 5.5, 6.0 (NP, SP) mm.Osstem TS (K161604): 3.0, 3.5 (3.7), 4.0 (4.2), 4.5 (4.6), 5.0 (5.1), 6.0 (6.0), 7.0 (6.8)(Mini, Regular) mm.Camlog Screw-Line (K083496): 3.3, 3.8, 4.3, 5.0, 6.0 mm.The TruAbutment DS is patient specific. CAD/CAM abutment,directly connected to endosseous dental implants and is intendedfor use as an aid in prosthetic rehabilitation. It is compatible withthe following systems:MIS C1 Conical Connection Implant (K172505, K112162): 3.3 (NP), 3.75, 4.2, 5.0 (SP, WP) mm.Neodent Implant System - GM Helix (K163194, K180536): 3.5, 3.75, 4.0, 4.3, 5.0 (3.0), 6.0 (3.0) mm.Nobel Biocare Groovy Implants (K050258): 3.5, 4.3, 5.0, 6.0 (NP, RP, WP, 6.0) mm.Straumann BLX Implant (K173961, K181703, K191256): 3.5, 3.75, 4.0, 4.5, 5.0, 5.5, 6.5 (RB, WB) mm.Straumann Tissue Level Implant (K111357): 3.3 (NNC) mm.All digitally designed abutments and/or copings for use with theTruAbutment DS abutments are intended to be sent to aTruAbutment-validated milling center for manufacture.
AttributesProposed DevicePrimary Predicate Device
Trade NameTruAbutment DSTruAbutment DS (K203649)
Conelog Screw-Line (K113779): 3.3, 3.8, 4.3, 5.0 mm.Implant Direct Legacy2 (K192221): 3.2 mm.BioHorizons Internal (K093321, K071638, K143022): 3.0, 3.4, 3.8, 4.6, 5.8 mm.MegaGen AnyRidge Internal Implant (K140091): 3.5, 4.0, 4.5, 5.0, 5.5 (3.5) mm.All digitally designed abutments and/or copings for use withthe TruAbutment DS abutments are intended to be sent to aTruAbutment-validated milling center for manufacture.TruAbutment DS is compatible with the following devices:Astra OsseoSpeed EV (K130999, K120414)Implant Body Diameter (mm): 3.0 /Implant Platform (mm):3.0 / Internal Spline connection.Implant Body Diameter (mm): 3.6 /Implant Platform (mm):3.6 / Internal Spline connection.Implant Body Diameter (mm): 4.2 /Implant Platform (mm):4.2 / Internal Spline connection.Implant Body Diameter (mm): 4.8 /Implant Platform (mm):4.8 / Internal Spline connection.Implant Body Diameter (mm): 5.4 /Implant Platform (mm):5.4/ Internal Spline connection.Biomet 3i Full OSSEOTITE Tapered Certain(K130949)
AttributesProposed DevicePrimary Predicate Device
Trade NameTruAbutment DSTruAbutment DS (K203649)
Implant Body Diameter (mm): 3.25 / Implant Platform (mm): 3.4 / Internal Hex connection.
Implant Body Diameter (mm): 4.0 / Implant Platform (mm): 4.1 / Internal Hex connection.
Implant Body Diameter (mm): 5.0 / Implant Platform (mm): 5.0 / Internal Hex connection.
Implant Body Diameter (mm): 6.0 / Implant Platform (mm): 6.0 / Internal Hex connection.
DIO UF(II) Internal Submerged (K161987, K170608, K173975)
Implant Body Diameter (mm): 3.3 /Implant Platform (mm): Narrow / Internal Hex connection.
Implant Body Diameter (mm): 3.8 /Implant Platform (mm): Regular / Internal Hex connection.
Implant Body Diameter (mm): 4.0 /Implant Platform (mm): Regular / Internal Hex connection.
Implant Body Diameter (mm): 4.5 /Implant Platform (mm): Regular / Internal Hex connection.
Implant Body Diameter (mm): 5.0 /Implant Platform (mm): Regular / Internal Hex connection.
Implant Body Diameter (mm): 5.5 /Implant Platform (mm): Regular / Internal Hex connection.
Implant Body Diameter (mm): 6.0 /Implant Platform (mm): Wide / Internal Hex connection.
Implant Body Diameter (mm): 6.5 /Implant Platform (mm): Wide / Internal Hex connection.
Implant Body Diameter (mm): 7.0 /Implant Platform (mm):
AttributesProposed DevicePrimary Predicate Device
Trade NameTruAbutment DSTruAbutment DS (K203649)
Wide / Internal Hex connection.Neoss ProActive® (K083561)Implant Body Diameter (mm): 3.25 / Implant Platform (mm): 3.5 / Internal Hex connection.Implant Body Diameter (mm): 3.5 /Implant Platform (mm): 4.0 / Internal Hex connection.Implant Body Diameter (mm): 4.0 /Implant Platform (mm): 4.0 / Internal Hex connection.Implant Body Diameter (mm): 4.5 /Implant Platform (mm): 4.0 / Internal Hex connection.Implant Body Diameter (mm): 5.0 /Implant Platform (mm): 4.0 / Internal Hex connection.Implant Body Diameter (mm): 5.5 /Implant Platform (mm): 4.0 / Internal Hex connection.Implant Body Diameter (mm): 6.0 /Implant Platform (mm) 4.0 / Internal Hex connection.Osstem TS (K161604)Implant Body Diameter (mm): 3.0 / Implant Platform (mm): 2.1 / Internal Hex connection.Implant Body Diameter (mm): 3.5 / Implant Platform (mm): 2.1 / Internal Hex connection.Implant Body Diameter (mm): 4.0 / Implant Platform (mm): 2.5/ Internal Hex connection.Implant Body Diameter (mm): 4.5 / Implant Platform (mm): 2.5 / Internal Hex connection.Implant Body Diameter (mm): 5.0/ Implant Platform (mm):
AttributesProposed DevicePrimary Predicate Device
Trade NameTruAbutment DSTruAbutment DS (K203649)
2.5 / Internal Hex connection.Implant Body Diameter (mm): 6.0 / Implant Platform (mm):2.5 / Internal Hex connection.Implant Body Diameter (mm): 7.0 / Implant Platform (mm):2.5 / Internal Hex connection.Camlog Screw-Line (K083496)Implant Body Diameter (mm): 3.3 / Implant Platform (mm):3.3 / Tube-in-tube connection.Implant Body Diameter (mm): 3.8 /Implant Platform (mm):3.8 / Tube-in-tube connection.Implant Body Diameter (mm): 4.3 /Implant Platform (mm):4.3 / Tube-in-tube connection.Implant Body Diameter (mm): 5.0 /Implant Platform (mm):5.0 / Tube-in-tube connection.Implant Body Diameter (mm): 6.0 /Implant Platform (mm):6.0/ Tube-in-tube connection.Conelog Screw-Line (K113779)Implant Body Diameter (mm): 3.3 /Implant Platform (mm):3.3 / Conical abutment connection.Implant Body Diameter (mm): 3.8 /Implant Platform (mm):3.8 / Conical abutment connection.Implant Body Diameter (mm): 4.3 /Implant Platform (mm):4.3 / Conical abutment connection.Implant Body Diameter (mm): 5.0 /Implant Platform (mm):5.0 / Conical abutment connection.
AttributesProposed DevicePrimary Predicate Device
Trade NameTruAbutment DSTruAbutment DS (K203649)
Implant Direct Legacy2 (K192221)Implant Body Diameter (mm): 3.2 /Implant Platform(mm): 3.0 / Internal Hex connection.
BioHorizons Internal Implant System (K093321, K071638,K143022)Implant Body Diameter (mm): 3.0/ Implant Platform (mm):3.0 / Internal Hex connection.Implant Body Diameter (mm): 3.4/ Implant Platform (mm):3.0 / Internal Hex connection.Implant Body Diameter (mm): 3.8/ Implant Platform (mm):3.5 / Internal Hex connection.Implant Body Diameter (mm): 4.6/ Implant Platform (mm):4.5 / Internal Hex connection.Implant Body Diameter (mm): 5.8/ Implant Platform (mm):5.7 / Internal Hex connection.
MegaGen AnyRidge Internal Implant System (K140091)Implant Body Diameter (mm): 3.5/ Implant Platform (mm):3.5 / Internal Hex connection.Implant Body Diameter (mm): 4.0/ Implant Platform (mm):3.5 / Internal Hex connection.Implant Body Diameter (mm): 4.5/ Implant Platform (mm):3.5 / Internal Hex connection.Implant Body Diameter (mm): 5.0/ Implant Platform (mm):3.5 / Internal Hex connection.Implant Body Diameter (mm): 5.5/ Implant Platform (mm):3.5 / Internal Hex connection.
AttributesProposed DevicePrimary Predicate Device
Trade NameTruAbutment DSTruAbutment DS (K203649)
CAD Design LimitsMinimum and Maximum abutment angle (°): 025Minimum and Maximum gingival (cuff) height (mm): 0.56.0Minimum and Maximum diameter at abutment/implantinterface (Ø,mm): 3.38.0Minimum and Maximum length of abutment (mm): 6.011.0Minimum wall thickness at the abutment/implant interface(mm): 0.4Minimum and Maximum length of abutment post (lengthabove the abutment collar/gingival height) (mm): 4.0~7.0Minimum and Maximum abutment angle (°): 025Minimum and Maximum Gingival (Cuff) Height (mm): 0.56.0Minimum and Maximum diameter at abutment/implant interface(Ø,mm): 3.38.0Minimum and Maximum length of abutment (mm): 6.011.0Minimum wall thickness at the abutment/implant interface (mm):0.40.9Minimum and Maximum length of abutment post (length above theabutment collar/gingival height) (mm): 4.07.0
ConnectionInternal ConnectionsInternal Connections
SterilityPackaged non-sterilePackaged non-sterile
MaterialTi-6Al-4V ELI (ASTM F136)Ti-6Al-4V ELI (ASTM F136)
Abutment SeatSits on TaperSits on Taper
Anatomical SiteOral CavityOral Cavity
ConstructionMachinedMachined
Type of RetentionScrew-retained to the implant. The prosthesis can be cement-retained to the abutment.Screw-retained to the implant. The prosthesis can be cement-retained tothe abutment.

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image is a logo for a company called "Abutment ru". The word "Abutment" is in a bold, sans-serif font and is placed above the letters "ru", which are in a cursive font. To the left of the letters "ru" is a graphic of a screw. The entire logo is in a light blue color.

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image contains a logo with the word "ABUTMENT" in a bold, sans-serif font, stacked above the letters "ru" in a cursive font. To the left of the word "ABUTMENT" is a stylized image of a screw. The entire logo is in a light blue color. The logo appears to be for a company or product related to dental implants or abutments.

{23}------------------------------------------------

Image /page/23/Picture/0 description: The image shows a logo for a company called "Abutment ru". The word "Abutment" is in a sans-serif font and is stacked on top of the letters "ru", which are in a cursive font. To the left of the word "ru" is a graphic of a screw.

{24}------------------------------------------------

Image /page/24/Picture/0 description: The image shows a logo for a company called "Abutment ru". The word "Abutment" is in a sans-serif font and is stacked on top of the letters "ru", which are in a cursive font. To the left of the word "ru" is a graphic of a screw.

{25}------------------------------------------------

Image /page/25/Picture/0 description: The image shows a logo for a company called "Abutment Tru". The word "Abutment" is in a sans-serif font and is stacked on top of the word "Tru", which is in a cursive font. To the left of the word "Tru" is a graphic of a screw.

{26}------------------------------------------------

Image /page/26/Picture/0 description: The image is a logo for a company called "Abutment ru". The word "Abutment" is in a bold, sans-serif font, with the letters in white and the background in blue. The letters "ru" are in a smaller, sans-serif font, with the letters in white and the background in blue. To the left of the letters "ru" is a graphic of a screw, also in blue and white.

{27}------------------------------------------------

Image /page/27/Picture/0 description: The image shows a logo for a company called "Abutment ru". The word "Abutment" is in a sans-serif font and is placed above the letters "ru", which are in a larger, bolder font. To the left of the letters "ru" is a jagged line that resembles a tooth. The entire logo is in a light blue color.

TruBase

AttributesProposed DevicePrimary Predicate Device
Trade NameTruBaseTruBase (K203649)
Indications for UseTruBase is a titanium component that is directly connected toendosseous dental implants to provide support for patient-specific prosthetic restorations, such as copings or crowns. Itis indicated for a screw-retained single tooth or cement-retained single tooth and bridge restorations. It is compatiblewith the following systems:Astra OsseoSpeed EV (K130999): 3.0 mm.Biomet 3i Full OSSEOTITE Tapered Certain (K130949): 3.25, 4.0, 5.0, 6.0 mm.DIO UF(II) Internal Submerged (K161987, K170608,K173975): 3.3, 3.8, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0 (Narrow, Regular) mm.Neoss ProActive® (K083561): 3.25, 3.5, 4.0, 4.5, 5.0, 5.5, 6.0 (NP, SP) mm.Camlog Screw-Line (K083496): 3.3, 3.8, 4.3, 5.0, 6.0 mm.Conelog Screw-Line (K113779):3.3, 3.8, 4.3, 5.0 mm.Implant Direct Legacy2 (K192221): 3.0 mm.All digitally designed zirconia superstructures for use with theTruBase are intended to be sent to a TruAbutment-validatedmilling center for manufacture.TruBase is a titanium component that is directly connected toendosseous dental implants to provide support for patient-specific prosthetic restorations, such as copings or crowns. Itis indicated for a screw-retained single tooth or cement-retained single tooth and bridge restorations. It is compatiblewith the following systems:MIS C1 Conical Connection Implant (K172505, K112162): 3.3 (NP) 3.75, 4.2, 5.0 (SP, WP) mm.Neodent Implant System - GM Helix (K163194, K180536): 3.5, 3.75, 4.0, 4.3, 5.0 (3.0) 6.0(3.0) mm.Nobel Biocare Groovy Implants (K050258): 3.5. 4.3, 5.0, 6.0 (NP, RP, WP, 6.0) mm.Straumann BLX Implant (K173961, K181703, K191256): 3.5, 3.75, 4.0, 4.5, 5.0, 5.5, 6.5 (RB, WB) mm.Straumann Tissue Level Implant (K111357): 3.3 (NNC) mm.All digitally designed zirconia superstructures for use with theTruBase are intended to be sent to a TruAbutment-validatedmilling center for manufacture.
AttributesProposed DevicePrimary Predicate Device
Trade NameTruBaseTruBase (K203649)
TruBase is compatible with the following devices:
Astra OsseoSpeed EV(K130999)Implant Body Diameter (mm): 3.0 / Implant Platform(mm): 3.0 / Internal Spline connection.
Biomet 3i Full OSSEOTITE Tapered Certain (K130949)Implant Body Diameter (mm): 3.25 / Implant Platform(mm): 3.4 / Internal Hex connection.Implant Body Diameter (mm): 4.0 / Implant Platform(mm): 4.1 / Internal Hex connection.Implant Body Diameter (mm): 5.0 / Implant Platform(mm): 5.0 / Internal Hex connection.Implant Body Diameter (mm): 6.0 / Implant Platform
(mm): 6.0 / Internal Hex connection.DIO UF(II) Internal Submerged (K161987, K170608,K173975)Implant Body Diameter (mm): 3.3 / Implant Platform(mm): Narrow / Internal Hex connection.Implant Body Diameter (mm): 3.8 / Implant Platform(mm): Regular/ Internal Hex connection.
Implant Body Diameter (mm): 4.0 / Implant Platform(mm): Regular / Internal Hex connection.Implant Body Diameter (mm): 4.5 / Implant Platform(mm): Regular / Internal Hex connection.Implant Body Diameter (mm): 5.0 / Implant Platform(mm): Regular / Internal Hex connection.Implant Body Diameter (mm): 5.5 / Implant Platform(mm): Regular / Internal Hex connection.Implant Body Diameter (mm): 6.0 / Implant Platform(mm): Wide / Internal Hex connection.
AttributesProposed DevicePrimary Predicate Device
Trade NameTruBaseTruBase (K203649)
Implant Body Diameter (mm): 6.5 / Implant Platform(mm): Wide / Internal Hex connection.Implant Body Diameter (mm): 7.0 / Implant Platform(mm): Wide / Internal Hex connection.Neoss ProActive® (K083561)Implant Body Diameter (mm): 3.25 / Implant Platform(mm): 3.5 / Internal Hex connection.Implant Body Diameter (mm): 3.5 / Implant Platform(mm): 4.0 / Internal Hex connection.Implant Body Diameter (mm): 4.0 / Implant Platform(mm): 4.0 / Internal Hex connection.Implant Body Diameter (mm): 4.5 / Implant Platform(mm): 4.0 / Internal Hex connection.Implant Body Diameter (mm): 5.0 / Implant Platform(mm): 4.0 / Internal Hex connection.Implant Body Diameter (mm): 6.0 / Implant Platform(mm) 4.0 / Internal Hex connection.Camlog Screw-Line (K083496)Implant Body Diameter (mm): 3.3 / Implant Platform(mm): 3.3 / Tube-in-tube connection.Implant Body Diameter (mm): 3.8 / Implant Platform(mm): 3.8 / Tube-in-tube connection.Implant Body Diameter (mm): 4.3 / Implant Platform(mm): 4.3 / Tube-in-tube connection.Implant Body Diameter (mm): 5.0 / Implant Platform(mm): 5.0 / Tube-in-tube connection.Implant Body Diameter (mm): 6.0 / Implant Platform(mm): 6.0/ Tube-in-tube connection.
AttributesProposed DevicePrimary Predicate Device
Trade NameTruBaseTruBase (K203649)
Conelog Screw-Line (K113779)Implant Body Diameter (mm): 3.3 / Implant Platform(mm): 3.3 / Conical abutment connection.Implant Body Diameter (mm): 3.8 / Implant Platform(mm): 3.8 / Conical abutment connection.Implant Body Diameter (mm): 4.3 / Implant Platform(mm): 4.3 / Conical abutment connection.Implant Body Diameter (mm): 5.0 / Implant Platform(mm): 5.0 / Conical abutment connection.Implant Direct Legacy2 (K192221)Implant Body Diameter (mm): 3.2 / Implant Platform(mm): 3.0 / Internal Hex connection.
CAD Design LimitsMaximum angulation (°): 015Maximum cuff height (mm): 0.55.0Minimum and Maximum diameter at the abutment/implant (Ø,mm): 5.0~ 8.0Minimum thickness (mm): 0.4Minimum post height length above the abutmentcollar/gingival height (mm): 4~6Maximum angulation (°): 015Maximum cuff height (mm) 0.55.0Minimum and Maximum diameter at the abutment/implant (Ø,mm): 5.0~ 8.0Minimum thickness (mm): 0.4Minimum post height length above the abutmentcollar/gingival height (mm): 4~6
Abutment Base andScrew MaterialTi-6Al-4V ELI (ASTM F136)Ti-6Al-4V ELI (ASTM F136)
Implant-to-AbutmentConnection (s)Screw-retained to the implant. The prosthesis can be cement-retained to the abutment.Screw-retained to the implant. The prosthesis can be cement-retained to the abutment.
Type of RetentionScrew-retained.Screw-retained.
Material ofSuperstructureInCoris Zi (K123664)InCoris Zi (K123664)
AttributesProposed DevicePrimary Predicate Device
Trade NameTruBaseTruBase (K203649)
Material of CementRelyX Unicem 2Automix by 3M ESPE (K100756)RelyX Unicem 2Automix by 3M ESPE (K100756)
ManufacturingprocessesTruAbutment-validated milling centerTruAbutment-validated milling center
End-User SterilizationMoist steam sterilizationMoist steam sterilization

{28}------------------------------------------------

Image /page/28/Picture/0 description: The image shows a logo for a company called "Abutment". The word "Abutment" is written in a bold, sans-serif font, with the letters "A" and "B" being larger than the other letters. To the right of the word "Abutment" are the letters "ru" in a smaller, cursive font. The entire logo is in a bright blue color. The left side of the "A" has a jagged edge, resembling the teeth of a gear.

{29}------------------------------------------------

Image /page/29/Picture/0 description: The image shows a logo for a company called "Abutment ru". The word "Abutment" is in a larger, bold, white font with a blue background. The letters "ru" are in a smaller, italicized, white font with a blue background. To the left of the text is a graphic of a screw.

{30}------------------------------------------------

Image /page/30/Picture/0 description: The image features a logo for "Abutment ru". The word "Abutment" is in a sans-serif font, with the "A" being larger than the rest of the letters. The letters "ru" are stylized in a cursive-like font and are positioned to the right and slightly below the word "Abutment". A graphic resembling the side view of a screw is placed to the left of the word "Abutment". The entire logo is in a bright blue color.

{31}------------------------------------------------

Image /page/31/Picture/0 description: The image shows a logo with the word "Abutment" in a large, bold font, stacked on top of the letters "ru" in a smaller, italicized font. The entire logo is in a bright blue color. To the left of the word "ru" is a stylized image of a screw.

{32}------------------------------------------------

Image /page/32/Picture/0 description: The image shows a logo for a company called "Abutment ru". The logo is primarily blue and features the word "Abutment" in a bold, sans-serif font, stacked above the letters "ru", which are also in blue. To the left of the "ru" is a stylized graphic that resembles a screw or a gear, also in blue, adding a visual element related to engineering or construction.

Substantial Equivalence Discussion

The subject device (TruAbutment DS) is substantially equivalent in indications and design principles to the primary predicate device listed above. The provided tables compare the Indications for Use Statements and the technological characteristics of the subject device, and the primary predicate device.

The Indications for Use Statement (IFUS) for the subject device (TruAbutment DS) is substantially equivalent in intended use to the primary predicate device (K203649). All are intended for use with endosseous dental implants in the maxilla and mandible to provide functional and esthetic rehabilitation of the edentulous maxilla and mandible.

Slight differences in the language of the subject device (TruAbutment DS) and primary predicate (K203649) Indications for Use statements do not affect the intended use as an endosseous dental implant abutment for support of a prosthesis to restore chewing function.

The minor differences between the subject device (TruAbutment DS) and the primary predicate device (K203649) are related to the compatible OEM implant lines. None of these minor differences impact substantial equivalences because all IFUS express equivalence intended to be used to facilitate dental prosthetic restorations, and the indications are expressed equivalently using different specific wording.

Design ParameterSubject Device(TruAbutment DS)Design LimitPrimary PredicateDevice (K203649)Design Limit
Minimum and Maximum abutment angle (°)0 ~ 250 ~ 25
Minimum and Maximum cuff height (mm)0.5 ~ 6.00.5 ~ 6.0
Minimum and Maximum diameter atabutment/implant interface (Ø, mm)3.3 ~ 8.03.3 ~ 8.0
Minimum and Maximum length of the abutment(mm)6~116 ~ 11
Minimum wall thickness at abutment/implantinterface (mm)0.40.4 ~ 0.9
Minimum and Maximum length of abutmentpost(length above the abutment collar / gingivalheight) (mm)4.0 ~ 7.04.0 ~ 7.0

The subject device and the predicate device indications and design parameters are identical.

The following subject device (TruBase) is substantially equivalent in indications and design principles to the primary predicate device (K203649). The subject device (TruBase) and the primary predicate device (K203649) have internal implant interface connections and are made of

{33}------------------------------------------------

Image /page/33/Picture/0 description: The image shows a logo for a company called "Abutment ru". The logo is primarily blue and features the word "Abutment" in a bold, sans-serif font, stacked above the letters "ru", which are also in a similar font but larger in size. To the left of the "ru" is a stylized image of a screw or bolt, also in blue, which adds a visual element related to construction or engineering.

Design ParameterSubject Device(TruBase)Design LimitPrimary Predicate Device(K203649) Design Limit
Minimum and Maximum angulation (°)0 ~ 150 ~ 15
Minimum and Maximum gingival (cuff)height (mm)0.5 ~ 5.00.5 ~ 5.0
Minimum and Maximum diameter atabutment/implant interface (Ø, mm)5.0 ~ 8.05.0 ~ 8.0
Minimum thickness (mm)0.40.4
Minimum and Maximum length of abutmentpost(length above the abutment collar / gingivalheight) (mm)4.0 ~ 6.04.0 ~ 6.0

Ti-6Al-4V ELI (abutments and abutment screws).

The subject devices (TruAbutment DS, TruBase) are to be sterilized by the end-user, the same as primary predicate devices (K203649). Sterilization for the subject devices (TruAbutment DS, TruBase) was performed according to ISO 17665-1 and ISO 17665-2. This sterilization validation method is the same as the primary predicate devices (K203649).

Mechanical performance testing was performed according to ISO 14801. For compatible OEM implant line, worst-case constructs were subjected to static compression and compression fatigue testing. The fatigue limit data for all other implant lines demonstrated the construct strengths to be sufficient for their intended use.

MR Environment Condition

Non-clinical worst-case MRI review was performed to evaluate the metallic DS, TruBase in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Maggetically Induced Displ acement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testin g and Evaluation 49.2 (2019): 783-795), based on the entire system including all variatio ns (all compatible implant bodies, dental abutments, and fixation screws) and material co mposition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including ma gnetically induced displacement force and torque.

{34}------------------------------------------------

Image /page/34/Picture/0 description: The image is a logo for a company called "Abutment ru". The logo is primarily blue and features the word "Abutment" in a bold, sans-serif font, with the "A" slightly larger than the other letters. Below "Abutment" is "ru" in a similar font. To the left of the text is a stylized graphic that resembles a screw or a gear, also in blue, which adds a visual element related to engineering or construction.

Non-clinical Testing

The subject device was tested to evaluate its substantial equivalence according to the following standards.

  • . Fatigue Test according to ISO 14801:2016
    The tests below were performed for the reference device (K152559, K200817) and leveraged for the subject device:

  • End-User Steam Sterilization Test according to ISO 17665-1:2006, 17665-2:2009 and ANSI/AAMI ST79:2010.

  • Biocompatibility tests according to ISO 10993-1:2009, ISO 10993-5:2009, and ISO ● 10993-10:2010.

Non-clinical test data was used to evaluate the proposed device's substantial equivalence compared to the predicate device. The results of the above tests have met the criteria of the standard and demonstrated substantial equivalence with the reference devices.

Non-clinical testing was conducted in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments," and it consisted of testing finished assembled implant/abutment systems of the worstcase scenario through fatigue testing.

Dimensional analysis and reverse engineering of the implant-to-abutment connection platform were performed, using the OEM implant body, abutment, and abutment screw. The reverse engineering included an assessment of maximum and minimum dimensions of critical design aspects and tolerances of the OEM implant body, abutment screws, along with cross-sectional images of the submission device and compatible implant body. The testing demonstrated implant to abutment compatibility and has established substantial equivalency of the proposed device with reference devices.

Clinical testing was not necessary to establish substantial equivalency of the proposed device.

Conclusion

The TruAbutment DS, TruBase constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its reference devices. TruAbutment DS, TruBase, and predicate devices substantially equivalent.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)