K Number
K221930
Manufacturer
Date Cleared
2023-03-09

(251 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CEREGLIDE™ 71 Intermediate Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. The CEREGLIDE™ 71 Intermediate Catheter is also indicated for use as a conduit for retrieval devices.

Device Description

The CEREGLIDE™ 71 Intermediate Catheter is a variable stiffness, single lumen catheter designed to be introduced over a steerable guide wire or microcatheter into the neuro vasculature. The catheter shaft is composed of a stainless steel variable pitch braid with a PTFE inner liner to facilitate movement of guide wires and other devices. The exterior of the catheter shaft is covered with polymer materials, which encapsulate the stainless steel braid construction. The catheter has a stiff proximal shaft which transitions into the flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The distal end of the catheter has a radiopaque marker band to facilitate fluoroscopic visualization and has a hydrophilic coating to provide lubricity to reduce friction during navigation in the vasculature. The proximal end of the catheter has a luer fitting located on the end of the catheter hub. An ID band is placed at the distal end of the hub over a strain relief. The catheter is packaged with a Tuohy Borst rotating hemostasis valve (RHV) with a side port and two slit introducers as accessories. The RHV with side port is used for flushing and insertion of catheters. The slit introducers are designed to introduce the catheter into the base catheter and protect the distal tip of the catheter during insertion into the RHV of the base catheter.

AI/ML Overview

The CEREGLIDE™ 71 Intermediate Catheter is a medical device for facilitating neurovascular interventions. The document provided outlines the non-clinical performance data to demonstrate its substantial equivalence to a legally marketed predicate device, the CERENOVUS Large Bore Catheter (K191237).

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the CEREGLIDE™ 71 Intermediate Catheter are implicitly defined by a series of bench tests and biocompatibility tests, with the reported performance consistently being "PASS" or "Samples met the established acceptance criteria," sometimes exceeding comparator devices.

Test CategorySpecific TestAcceptance Criteria (Implicit)Reported Device Performance
Design VerificationVisual InspectionCatheter meets visual requirements described in ISO 10555-1 Section 4.4.PASS: Samples met the established acceptance criteria
Catheter IDCatheter internal diameters meet requirements.PASS: Samples met the established acceptance criteria
Catheter ODCatheter outer diameters meet requirements.PASS: Samples met the established acceptance criteria
Catheter Working LengthWorking length of the catheter as defined in ISO 10555-1 Section 3.6.PASS: Samples met the established acceptance criteria
Catheter Tip LengthCatheter tip length meets requirements.PASS: Samples met the established acceptance criteria
System Air LeakageNo air leak into the hub subassembly.PASS: Samples met the established acceptance criteria
System Liquid LeakageCatheter joint strength meets the freedom from leakage (liquid during pressurization) requirements of ISO 10555-1:2013, section 4.7.PASS: Samples met the established acceptance criteria
Delamination of PTFE LinerPTFE appropriately adhered to the inner lumen of the Catheter with braid reinforcement.PASS: Samples met the established acceptance criteria
Kink (Distal & Proximal)Catheter remains stable and does not kink during use.PASS: Samples met the established acceptance criteria
Tip MovementCatheter meets the tip column stiffness requirement.PASS: Samples met the established acceptance criteria
Tip Linear StiffnessTip flexibility of the Catheter is acceptable, relative to other devices of similar design.PASS: Samples met the established acceptance criteria
Coating Lubricity and DurabilityLubriciousness and durability of the Catheter's hydrophilic coating are verified.PASS: Samples met the established acceptance criteria
Coating LengthCatheter hydrophilic coating length meets the design requirements.PASS: Samples met the established acceptance criteria
Peak Tensile StrengthCatheter joint strength meets the requirements of Section 4.5 of ISO 10555-1.PASS: Samples met the established acceptance criteria
Introducer IDIntroducer internal diameters meet the requirements.PASS: Samples met the established acceptance criteria
Particulate CountCoating integrity of the Catheter's outer surface meets the requirements for content of Particle Matter in alignment with USP<788> counting methods and compared to the reference predicate device.PASS: Samples met the established acceptance criteria
Burst PressureCatheter can withstand the maximum hydrostatic pressure.PASS: Samples met the established acceptance criteria
Introducer Working LengthWorking length of the introducer is confirmed.PASS: Samples met the established acceptance criteria
Introducer Separation ForceForce required to separate the introducer is confirmed.PASS: Samples met the established acceptance criteria
Torque TestNumber of revolutions to failure of the Catheter in simulated anatomy meets or exceeds established criteria.PASS: Samples exceeded comparator devices in revolutions to failure
Design ValidationIn Vitro Usability StudiesUsability parameters such as trackability, tip stability, durability, and (ancillary) device compatibility with tracking of the Catheter to target sites and delivery of a stent-retriever in the neurovascular model are acceptable.PASS: Samples met the established acceptance criteria
BiocompatibilityChemical CharacterizationDetermination of extractable species from the test article in purified water, isopropyl alcohol, and hexane is acceptable.PASS
CytotoxicityLack of potential cytotoxic effects using monolayers of L-929 mouse fibroblast cells when evaluated with 1X MEM extract.PASS
Sensitization – Guinea Pig MaximizationNo allergenic potential or sensitizing capacity.PASS
Irritation - Rabbit Intracutaneous ReactivityNo chemicals that may leach or be extracted capable of causing local irritation in the dermal tissues of rabbits.PASS
Acute Systemic Toxicity in MiceNo potential toxic effects as a result of a single-dose systemic injection in mice.PASS
Pyrogenicity – Materials Mediated Rabbit PyrogenNo febrile response in rabbits caused by a saline extract of the test article.PASS
ASTM Hemolysis Study: Direct and Extraction MethodsAcceptable hemolytic potential of the test articles and test article extract according to ASTM method (F756-17).PASS
SC5b-9 Complement Activation AssayTest article is not capable of generating the activation fragment, SC5b-9, which contributes to inflammatory immune response in humans, as measured in Normal Human Serum (NHS).PASS
ASTM Heparinized Platelet and Leukocyte CountMedical materials exposed to human whole blood do not adversely affect the platelet and leukocyte ratios in whole blood.PASS
ASTM Partial Thromboplastin Time (PTT)No coagulation abnormalities in the intrinsic coagulation pathway that would indicate an adverse effect.PASS
In Vivo ThromboresistanceAcceptable thrombogenic potential of a blood contacting medical device in comparison to a predicate device.PASS
Shelf-LifeStability TestingDevice maintains its performance and safety for 1 year when subjected to standard transportation conditioning and aging.PASS: Results of testing on the subject device all met established acceptance criteria.

2. Sample Sizes Used for the Test Set and Data Provenance

The document states: "All testing was conducted using sampling methods as required by internal procedure." Specific sample sizes for each test are not explicitly provided.

The data provenance is non-clinical bench testing and in-vitro testing. There is no mention of human clinical data or patient data. Therefore, there is no country of origin or retrospective/prospective distinction for patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. This device's evaluation relies on objective engineering and material science tests (bench and in-vitro, and biocompatibility tests), not expert interpretation of diagnostic images or clinical outcomes. Therefore, there is no "ground truth" established by experts in this context.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods (like 2+1, 3+1) are typically used for establishing ground truth in clinical or image-based studies where expert consensus is required for ambiguous cases. As the studies described are bench and biocompatibility tests, human adjudication is not relevant.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted as this is a device clearance based on substantial equivalence using non-clinical performance data, not an AI or diagnostic imaging device requiring human reader interaction.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This pertains to an intermediate catheter, which is a physical medical device, not a software algorithm or AI.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

The "ground truth" for the non-clinical tests is established by pre-defined engineering specifications, material properties, and regulatory standards (e.g., ISO standards, ASTM methods, USP guidance). The device's performance is compared against these objective criteria. For the biocompatibility tests, the "ground truth" is typically the absence of adverse biological reactions as defined by the specific test (e.g., no cytotoxicity, no sensitization, no irritation).

8. The Sample Size for the Training Set

Not applicable. There is no "training set" as this is not a machine learning or AI-based device.

9. How the Ground Truth for the Training Set Was Established**

Not applicable. As there is no training set, there is no ground truth establishment for it.

{0}------------------------------------------------

March 9, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Cerenovus, Inc. Ariell Joiner, Ph.D. Manager, Regulatory Affairs 6303 Blue Lagoon Drive, Suite 315 Miami. Florida 33126

Re: K221930

Trade/Device Name: CEREGLIDE 71 Intermediate Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP Dated: February 2, 2023 Received: February 3, 2023

Dear Ariell Joiner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan -S

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221930

Device Name

CEREGLIDE™ 71 Intermediate Catheter

Indications for Use (Describe)

The CEREGLIDE™ 71 Intermediate Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the neurovascular system. The CEREGLIDE™ 71 Intermediate Catheter is also indicated for use as a conduit for retrieval devices.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K221930

I. SubmitterCerenovus, Inc.6303 Blue Lagoon DriveSuites 215 & 315Miami, FL 33126 USA
Contact Person: Ariell JoinerTel: (908) 249-0182Email: aioiner@its.ini.com

II. Date Prepared

March 8, 2023

III. Device Information

Table 1. Device Information
Device Proprietary NameCEREGLIDE™ 71 Intermediate Catheter
Common or Usual nameCatheter, Percutaneous, Neurovasculature
Classification Name21 CFR 870.1250 – Catheter, Percutaneous
Regulatory ClassificationII
FDA Product CodeQJP

IV. Predicate Device Information

V. Device

The primary predicate device is listed below in Table 2 and the reference device is listed in Table 3.

Table 2. Primary Predicate Device
510(k) NumberDate ClearedNameManufacturer
K191237November 8.2019CERENOVUS LargeBore CatheterMedos InternationalSARL*
*510(k) was previously held by Codman & Shurtleff, Inc. The 510(k) was transferred to
Medos International SARL, which is the current holder of the 510(k).
Table 3. Reference Device
510(k) NumberDate ClearedNameManufacturer
K150366March 27, 2015SOFIA PLUS/DistalAccess CatheterMicroVention, Inc.

The CEREGLIDE™ 71 Intermediate Catheter is a variable stiffness, single lumen Description catheter designed to be introduced over a steerable guide wire or microcatheter into the neuro vasculature. The catheter shaft is composed of a stainless steel variable pitch braid with a PTFE inner liner to facilitate movement of guide wires and other devices. The exterior of the catheter shaft is covered with polymer materials, which encapsulate the stainless steel braid construction. The catheter has a stiff proximal shaft which transitions into the flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The distal end of the catheter has a radiopaque marker band to facilitate fluoroscopic visualization and has a hydrophilic coating to provide lubricity to reduce friction during navigation in the vasculature. The proximal end of the catheter has a luer fitting located on the end of the catheter hub. An ID band is placed at the distal end of the hub over a strain relief.

{4}------------------------------------------------

V. DeviceDescription,continuedThe catheter is packaged with a Tuohy Borst rotating hemostasis valve (RHV) with aside port and two slit introducers as accessories. The RHV with side port is used forflushing and insertion of catheters. The slit introducers are designed to introduce thecatheter into the base catheter and protect the distal tip of the catheter during insertioninto the RHV of the base catheter.
VI. Indicationsfor UseThe CEREGLIDE™ 71 Intermediate Catheter is indicated for use in facilitating theinsertion and guidance of appropriately sized interventional devices into a selectedblood vessel in the neurovascular system. The CEREGLIDE™ 71 IntermediateCatheter is also indicated for use as a conduit for retrieval devices.
VII. PredicateComparisonA comparison of the similarities and differences of product features between theCEREGLIDE™ 71 Intermediate Catheter and the predicate device is presented inTable 4.
Table 4. Subject and Predicate Device Comparison Summary
DescriptionSubject Device:CEREGLIDETM 71 Intermediate CatheterPredicate Device:CERENOVUS Large Bore Catheter(K191237)
Product CodeQJPDQY
Regulatory NameCatheter, PercutaneousSame
ClassificationClass II - 21 CFR 870.1250Same
Basic DesignVariable stiffness single lumen catheterSame
Indications ForUseThe CEREGLIDETM 71 Intermediate Catheter is indicatedfor use in facilitating the insertion and guidance ofappropriately sized interventional devices into a selectedblood vessel in the neurovascular system. TheCEREGLIDETM 71 Intermediate Catheter is alsoindicated for use as a conduit for retrieval devices.The CERENOVUS Large Bore Catheter is indicated for use infacilitating the insertion and guidance of appropriatelysized interventional devices into a selected blood vessel in theneurovascular system. The CERENOVUS Large BoreCatheter is also indicated for use as a conduit for retrievaldevices.
Dimensions:
Length115 - 137cm125 - 135cm
Inner Diameter(ID)0.071"Same
Distal OuterDiameter (OD)0.082"(0.0825" max)0.081"
Proximal OD0.0837" max0.0825"
Catheter CoatingHydrophilicSame
Coating Length55 cm30 cm
Materials:
Marker BandMetal Platinum (90%) / Iridium (10%)Same
BraidStainless SteelSame
LinerPTFE LinerSame
HubPolyamideSame
Strain ReliefSameSame
Outer JacketPebax and UrethanePebax, Urethane, Nylon
Accessories Included:
Hemostasis ValveIncluded: Tuohy Borst Hemostasis Valve with Side PortExtension TubingSame
IntroducerSheathIntroducer (2)Peel-Away Sheath Introducer (2)
SterilizationMethodEthylene OxideSame
SterilityAssurance Level(SAL)10-6Same
PackagingPolyethylene Hoop and Mounting Card, Tyvek® Pouch,CartonSame
Shelf Life1 year1 year

Continued on next page

{5}------------------------------------------------

VIII. Non-Clinical Performance Data

Performance Testing - Bench

Appropriate testing was identified based on design, risk analyses and the intended use of the CEREGLIDE™ 71 Intermediate Catheter to demonstrate that it is substantially equivalent to the legally marketed predicate device. The following performance data are being provided in support of the substantial equivalence determination. All testing was conducted using sampling methods as required by internal procedure. The bench testing included the following tests:

Table 5. Performance Testing Summary
TestTest SummaryResult
Design Verification
VisualInspectionConfirm that the Catheter meets the visual requirementdescribed in ISO 10555-1 Section 4.4.PASS:Samples met the establishedacceptance criteria
Catheter IDVerify that the Catheter internal diameters meet therequirements.PASS:Samples met the establishedacceptance criteria
Catheter ODVerify that the Catheter outer diameters meet the requirements.PASS:Samples met the establishedacceptance criteria
CatheterWorkingLengthConfirm the working length of the catheter as defined in ISO10555-1 Section 3.6.PASS:Samples met the establishedacceptance criteria
Catheter TipLengthVerify the catheter tip length of the Catheter.PASS:Samples met the establishedacceptance criteria
System AirLeakageVerify that there is no air leak into the hub subassembly.PASS:Samples met the establishedacceptance criteria
System LiquidLeakageVerify that the catheter joint strength meets the freedom fromleakage (liquid during pressurization) requirements of ISO10555-1:2013, section 4.7.PASS:Samples met the establishedacceptance criteria
Delamination ofPTFE LinerVerify that the PTFE has appropriately adhered to the innerlumen of the Catheter with braid reinforcement.PASS:Samples met the establishedacceptance criteria
Kink (Distal &Proximal)Confirm that the Catheter meets the requirement for thecatheter to remain stable and not kink during use.PASS:Samples met the establishedacceptance criteria
Tip MovementConfirm that the Catheter meets the tip column stiffnessrequirement.PASS:Samples met the establishedacceptance criteria
Tip LinearStiffnessTest the tip flexibility of the Catheter, relative to other devicesof similar design.PASS:Samples met the establishedacceptance criteria
CoatingLubricity andDurabilityVerify the lubriciousness and durability of the Catheter'shydrophilic coating.PASS:Samples met the establishedacceptance criteria
Coating LengthVerify that the Catheter hydrophilic coating length meets thedesign requirements.PASS:Samples met the establishedacceptance criteria
Peak TensileStrengthVerify that the Catheter joint strength meets the requirements ofSection 4.5 of ISO 10555-1.PASS:Samples met the establishedacceptance criteria
Introducer IDVerify that the introducer internal diameters meet therequirements.PASS:Samples met the establishedacceptance criteria
ParticulateCountVerify that the coating integrity of the Catheter's outer surfacemeets the requirements for content of Particle Matter inalignment with USP<788> counting methods and compared tothe reference predicate device.PASS:Samples met the establishedacceptance criteria
Table 5. Performance Testing Summary, continued
TestTest SummaryResult
Design Verification, continued
Burst PressureConfirm the maximum hydrostatic pressure the Cathetercan withstand using a Crescent Hydraulic Burst-leakTester.PASS:Samples met the establishedacceptance criteria
Introducer WorkingLengthConfirm the working length of the introducer.PASS:Samples met the establishedacceptance criteria
IntroducerSeparation ForceConfirm the force required to separate the introducer.PASS:Samples met the establishedacceptance criteria
Torque TestTo determine the number of revolutions to failure of theCatheter in simulated anatomy.PASS:Samples exceeded comparatordevices in revolutions to failure
Design Validation
In Vitro UsabilityStudiesThe in-vitro studies were conducted to evaluate usabilityparameters such as trackability, tip stability, durability, and(ancillary) device compatibility with tracking of theCatheter to target sites and delivery of a stent-retriever inthe neurovascular model.PASS:Samples met the establishedacceptance criteria

{6}------------------------------------------------

VIII. Non-Clinical Performance Data, continued

Performance Testing - Animal

No animal studies were required as appropriate verification and validation of the device design were achieved based on the similarities of the proposed device and the predicate device, and from results of bench testing.

Performance Testing - Clinical

Clinical studies were not required as appropriate verification and validation of the device design were achieved based on the similarities of the proposed device and the predicate device, and from results of bench testing.

Sterilization

The CEREGLIDE™ 71 Intermediate Catheter, as packaged with included accessories, is sterilized using a validated 100% Ethylene Oxide sterilization process to ensure sterility assurance level (SAL) of 106 in accordance with ISO 11135. The CEREGLIDE™ 71 Intermediate Catheter and all accessories meet EO residuals per EN ISO 10993-7 for a limited contact delivery system - externally communicating. The CEREGLIDE™ 71 Intermediate Catheter and all accessories are for single use only.

Shelf-Life

The CEREGLIDE™ 71 Intermediate Catheter will have a shelf life of one year based on the successful completion of stability testing. Shelf life testing was performed using standard test methods and acceptance criteria. Prior to aging, all samples were exposed to standard transportation conditioning. Results of testing on the subject device all met established acceptance criteria.

Continued on next page

{7}------------------------------------------------

VIII. Non-Clinical Performance Data, continued

Biocompatibility Testing

A biological safety evaluation was conducted on the CEREGLIDE™ 71 Intermediate Catheter in accordance with the FDA guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", (Issued June 16, 2016) based on the changes to the predicate device, CERENOVUS Large Bore Catheter (K191237). Biocompatibility testing was conducted in accordance with International Standard ISO 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation of Testing Within a Risk Management Process (2018).

The following testing was performed:

Table 6. Biocompatibility Test Summary for CEREGLIDE™ 71 Intermediate
Catheter and Introducer
TestTest SummaryResults
ChemicalCharacterizationDetermination of extractable species from the test article wasperformed in purified water, isopropyl alcohol, and hexane.PASS
CytotoxicityCytotoxicity Study Using the ISO Elution Method (compliantwith ISO 10993-5:2009/(R)2014 and ISO 10993-12:2012).1X MEM extract (6 cm²/ mL at 37°C for 72 hours) was evaluatedfor potential cytotoxic effects using monolayers of L-929 mousefibroblast cells.PASS
Sensitization –Guinea PigMaximizationEvaluation of the allergenic potential or sensitizing capacity of atest article by screening of contact allergens in guinea pigs andextrapolating the results to humans.PASS
Irritation - RabbitIntracutaneousReactivityDetermination of any chemicals that may leach or be extractedfrom the test article capable of causing local irritation in thedermal tissues of rabbits.PASS
Acute SystemicToxicity in MiceScreening of test article extracts for potential toxic effects as aresult of a single-dose systemic injection in mice.PASS
Pyrogenicity –Materials MediatedRabbit PyrogenDetermination if a saline extract of the test article causes afebrile response in rabbits.PASS
ASTM HemolysisStudy: Direct andExtraction MethodsEvaluation of the hemolytic potential of the test articles and testarticle extract according to ASTM method (F756-17).PASS
SC5b-9ComplementActivation AssayMeasure of complement activation in Normal Human Serum(NHS) indicates whether a test article is capable of generating theactivation fragment, SC5b-9, which in turn contributes to theinflammatory immune response in humans.PASS
ASTM HeparinizedPlatelet andLeukocyte CountDetermination if medical materials exposed to human wholeblood would adversely affect the platelet and leukocyte ratios inwhole blood.PASS
ASTM PartialThromboplastinTime (PTT)Screening for detection of coagulation abnormalities in theintrinsic coagulation pathway. The PTT indirectly measures theformation of thrombin by its action on fibrinogen, forming thefibrin clot through a method compliant with ISO 10993-4.PASS
In VivoThromboresistanceEvaluate the thrombogenic potential of a blood contactingmedical device in comparison to a predicate device.PASS

Continued on next page

{8}------------------------------------------------

IX. Conclusion

Based upon the intended use, design, materials, function, and side-by-side in-vitro and in-vivo testing, it is concluded that the subject device, CEREGLIDE™ 71 Intermediate Catheter is substantially equivalent to the predicate device, CERENOVUS Large Bore Catheter (K191237). The differences in materials and design do not raise new questions regarding the safety and effectiveness of the device. The device, as designed, manufactured, packaged and sterilized, is substantially equivalent to the primary predicate and reference devices currently marketed under the Federal Food, Drug and Cosmetic Act.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).