AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System are indicated for treating fractures of various bones including the clavicle, pelvis, scapula, long bone (humerus, ulna, radius, femur, tibia and fibula), and small bone (metacarpals, metatarsals, phalanges).

The system is indicated for use in adult patients only. All implants are for single use only.

Device Description

The NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System is submitted as a bundled submission. Throughout the submission there is a mention of NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System that represents the range of products covered under this 510(k) submission. The NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System are subdivided into the following categories:

S. No.CategoryTypes
01Mini, Small and LargeVolar Plates, T-Shaped, Angled Plates,Reconstruction plates
02Dynamic compressionSmall, Narrow, Lengthening
03Hip PlateDynamic Hip Screw Plates
04Bone ScrewsLocking and Non-Locking Version

Generally, there are the following types of bone plates used with cortical (cortex), cancellous, and Locking Screws. These bone plates are generally designed on the basis of the bone contour and anatomy.

S. No.TypeSubtype
01Mini, Small and LargeFragmentVolar Plates, T-Shaped, Angled Plates,Reconstruction plates
02Dynamic compressionSmall, Narrow, Lengthening Narrow, Broad Plates
03Hip PlateDynamic Hip Screw Plates

NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System consists of various shape and sizes of plates featuring compression and locking holes, full-threaded-cortical, locking selftapping screws, compression and dynamic screws.

The plates and screws are fabricated from Stainless Steel and Titanium. The implants are supplied non-sterile and should be steam-sterilized prior to use.

The system contains several models based on the size of the device and application site such as fixation/reconstruction of small fragment bones, forefoot, mid-foot, and the ankle. The plate implants are available in models such as Reconstruction Plates, T- Plates, , Anatomical Plates, and Clavicle Plates.

These all are mainly divided into the following:

  • . Large Fragment Plates
  • Small Fragment Plates
  • Hip Plates (DHS Plate)

The thickness of the plates varies from 1.3mm to 6mm; and the nomber of the holes varies from 2 to 20. The Locking screw implants are in corresponding diameter ranges from 2.7mm, and 4.9 mm diameters with lengths varying as per the requirements and lengths of 10mm to 90 mm

The Non-locking cortical screw implants are in 3.5mm diameter and length range from 10mm to 80mm.

The Non-locking cancellous screws are in 6.5mm diameter and length range from 30mm to 120mm The Non-locking cannulated cancellous screws are in 4.0mm diameter and length range from 10mm to 72mm

The Lag Screw is of 12.5mm Diameter and length from 50mm to 120mm

AI/ML Overview

This document is a 510(k) summary for the Nebula Brand Locking Bone Plates and Screws Osteosynthesis Plating System and Nebula Brand of DHS Plating System. It focuses on demonstrating substantial equivalence to predicate devices, rather than proving the device meets acceptance criteria through a clinical study.

Therefore, the requested information about acceptance criteria, study design, sample sizes, expert ground truth, MRMC studies, standalone performance, and training set details cannot be extracted from this document, as it is not a clinical study report for device performance.

The non-clinical performance testing section briefly mentions:

  • Non-Clinical Performance Testing: Performance was demonstrated through test parameters outlined in ASTM F-382 (Static and Dynamic Four Point Bend Testing) and ASTM F-384 (Static and Dynamic Four Point Bend Testing). Torsional and Axial pullout testing was conducted on the screws per ASTM F-543. These are standard test methods for mechanical properties of bone plates and screws. The document states that the performance is "similar" to predicate devices, implying that the results of these tests met a level comparable to the predicate devices, which is the basis for proving substantial equivalence.

Here's a breakdown of why many of your questions cannot be answered by this document:

  1. A table of acceptance criteria and the reported device performance: Not provided. The document focuses on demonstrating similarities to predicate devices based on design, materials, and non-clinical performance testing (mechanical properties). It doesn't present specific acceptance criteria in terms of clinical outcomes or diagnostic accuracy (which would be relevant for an AI/diagnostic device, not a bone fixation system).
  2. Sample sizes used for the test set and the data provenance: Not applicable. This document refers to mechanical testing of physical devices, not a test set of patient data.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for clinical data is not relevant to this type of device submission.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/diagnostic device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable in the context of clinical performance. The "ground truth" for this device's performance would be its mechanical properties meeting established standards (ASTM).
  8. The sample size for the training set: Not applicable. There is no "training set" for this type of device.
  9. How the ground truth for the training set was established: Not applicable.

In summary: This FDA 510(k) clearance letter and its associated summary are for a bone fixation system (plates and screws). The regulatory pathway for such devices typically relies on demonstrating substantial equivalence to existing legally marketed predicate devices through material characterization, mechanical testing, and sometimes benchtop or animal studies. It does not involve clinical studies with patient data, AI, or human reader performance, which are common for diagnostic or AI/ML-based medical devices.

{0}------------------------------------------------

December 9, 2019

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.

Nebula Surgical Private Limited Rajat Kumar Khant Director Nebula, Narmada Park-3,Amin Marg, Vidyakunj Soc.Main Road Opp.Kings Heigt Rajkot, 360005 In

Re: K191793

Trade/Device Name: Nebula Brand Locking Bone Plates and Screws Osteosynthesis Plating System, Nebula Brand of DHS Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: September 10, 2019 Received: September 10, 2019

Dear Rajat Khant:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191793

Device Name

Nebula Brand Locking Bone Plates and Screws Osteosynthesis Plating System, Nebula Brand of DHS Plating System

Indications for Use (Describe)

NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System are provided non-sterile.

NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System are indicated for treating fractures of various bones including the clavicle, pelvis, scapula, long bone (humerus, ulna, radius, femur, tibia and fibula), and small bone (metacarpals, metatarsals, phalanges).

The system is indicated for use in adult patients only. All implants are for single use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 201.2 Subpart D)
Over-The-Counter Use (21 CFR 201.2 Subpart C)

Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This Premarket Notification 510(k) Summary is prepared as required by Section 807.92

Submitter Information

510(k) Submitter:Date Prepared:NEBULA SURGICAL PVT. LTD.06-06-2019
Head Office Address:"Nebula", Narmada Park-3, Vidyakunj Society Main Road, Opp. Kings HeightAppt., Amin Marg, Rajkot - 360005, Gujarat (INDIA)
Factory address :G/1921/3, Gate no. 02, G.I.D.C. - Metoda, Dist. Rajkot - 360021, Gujarat (INDIA)
Contact:Rajat Kumar KhantDirector (Sales and Marketing)Email: executiveshrikant@gmail.comPhone: +91-9825132632

Submission Information

Proprietary Name:NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System
Common or Usual Name:Orthopaedic Bone Plates and Screws
Classification Name:21 CFR 888.3030: Plates Fixation, Bone
21 CFR 888.3040: Screws, Fixation, Bone
Product Code:HRS, HWC
Device Class:Class II
Classification Panel:Orthopaedic

Predicate Devices (with corresponding subject Device)

The subject NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System is submitted as a bundled submission. The following table lists predicate device 510(k) numbers used to demonstrate substantial equivalence. The following subject devices have been aligned with their respective predicate devices:

S NoSubject DevicePredicate Device
Small Fragment
01Distal Radius Volar Locking Plate,Left & RightSynthes (USA) VOLAR DISTAL RADIUS PLATE(K953644)
02Small Locking PlateSynthes (USA) 3.5 mm and 4.5 mm Locking Compression Plate(K082807
03Distal Tibial Medial Locking Plate,Left & RightSynthes (USA) Medial Distal Tibia Plate (K001945)
04Reconstruction Locking Plate3.5mmSynthes 4.5 mm LCP Straight Reconstruction Plate (K051986)
05Clavicle Medial Locking Plate, Left& RightSynthes 3.5mm LCP Clavicle Plate System (K111540)
06Distal Humerus D/L @ SupportLCP, Left & RightSynthes 3.5 MM LCPDISTAL HUMERUS SYSTEM (K033995)
07Distal Humerus Medial LCP Plate,Left & RightSynthes 3.5 MM LCP DISTAL HUMERUS SYSTEM (K033995)
08Distal Fibula VA LCP, Left and RightSynthes 2.7mm / 3.5mm LCP Distal Fibula Plate (K073460)
09Anterolateral Distal Tibia LockingPlate, Left & RightSynthes (USA) 2.7mm / 3.5mm LCP Anterolateral Distal Tibia Plate(K092812)
10Medial Proximal Tibia LockingPlate, Left & RightSynthes (USA) 3.5 / 4.5mm LCP® Medial Proximal Tibia Plate (K032269)
11Posterior Medial Proximal TibialLocking PlateSynthes 3.5mm LCP Proximal Tibia Plates(K030597)
12Anatomical Proximal Tibia LockingPlate, Left and RightSYNTHES ANATOMICAL LOCKING PLATE SYSTEM (K961413)
Large Fragment
13Narrow and Broad Locking Plates Ø4.5mmSynthes 3.5 and 4.5mm Locking Compression Plate System (K082807)
14Broad Locking Plate Ø 4.5mmSynthes 3.5 and 4.5mm Locking Compression Plate System (K082807)
15Distal Femur Locking Plate, Right &LeftSynthes LCP Distal FemurPlates (K062564)
16Lateral Tibia Locking Plate, Right &LeftSynthes LCP Proximal Tibia Plate (K011978)
17Philos Locking PlateSynthes (USA) LCP® Proximal Humerus Plates, Long (K041860)
18Narrow LC DCP Plate, 3.5mmSynthes 3.5 and 4.5mm Locking Compression Plate System (K082807)
19D.H.S. Plate 135 Degree withDynamic Hip Screw (Non- Locking)Long Barrel and Short Barrel D.H.S.Plate 130 Degree with Dynamic HipScrew (Non- Locking) Long Barreland Short BarrelSynthes Limited Contact- Dynamic Hip Screw Implant (LC-DHS) forK923613SSynthes Titanium Limited Contact-Dynamic Hip Screw Implant(Ti LC- DHS) for K953607
Locking and Non-Locking Screw
20D.H.S. Lag ScrewSynthes Limited Contact- Dynamic Hip Screw Implant (LC-DHS)" forK923613Synthes Titanium Limited Contact-Dynamic Hip Screw Implant(Ti LC-DHS) for K953607
21Locking Screw, Ø 2.7mm - SelfTappingSynthes Small Fragment Dynamic CompressionLocking (DCL) System(K000684)
22Locking Screw, Ø 3.5mm - SelfTappingSynthes Small Fragment Dynamic Compression Locking (DCL) System(K000684)
23Cortical Screw Ø 3.5mm 20 TPI(Self Tapping)Synthes Cortical Screws K112583
24Cannulated Cancellous Screw Ø 4.0mm, 16mm Thread, 32mm Threadand Full ThreadSynthes (USA) 6.5 mm Cancellous Screws (K061621)
25Cortical Screw Ø 4.5mm,Self-TappingSynthes Cortical Screws K112583
26Locking Screw, Ø4.9 mmSelf-TappingSynthes Large Fragment Dynamic Compression Locking (DCL) System(K000682)
27Cancellous Screw, Ø6.5mm, 16mmThread, 32mm Thread and FullThreadSynthes (USA) 6.5 mm Cancellous Screws (K061621)

{4}------------------------------------------------

510(k) Summary

{5}------------------------------------------------

Device Description

The NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System is submitted as a bundled submission. Throughout the submission there is a mention of NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System that represents the range of products covered under this 510(k) submission. The NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System are subdivided into the following categories:

S. No.CategoryTypes
01Mini, Small and LargeVolar Plates, T-Shaped, Angled Plates,Reconstruction plates
02Dynamic compressionSmall, Narrow, Lengthening
03Hip PlateDynamic Hip Screw Plates
04Bone ScrewsLocking and Non-Locking Version

Generally, there are the following types of bone plates used with cortical (cortex), cancellous, and Locking Screws. These bone plates are generally designed on the basis of the bone contour and anatomy.

S. No.TypeSubtype
01Mini, Small and LargeFragmentVolar Plates, T-Shaped, Angled Plates,Reconstruction plates
02Dynamic compressionSmall, Narrow, Lengthening Narrow, Broad Plates
03Hip PlateDynamic Hip Screw Plates

NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System consists of various shape and sizes of plates featuring compression and locking holes, full-threaded-cortical, locking selftapping screws, compression and dynamic screws.

The plates and screws are fabricated from Stainless Steel and Titanium. The implants are supplied non-sterile and should be steam-sterilized prior to use.

The system contains several models based on the size of the device and application site such as fixation/reconstruction of small fragment bones, forefoot, mid-foot, and the ankle. The plate implants are available in models such as Reconstruction Plates, T- Plates, , Anatomical Plates, and Clavicle Plates.

These all are mainly divided into the following:

  • . Large Fragment Plates
  • Small Fragment Plates
  • Hip Plates (DHS Plate)

{6}------------------------------------------------

510(k) Summarv

The thickness of the plates varies from 1.3mm to 6mm; and the nomber of the holes varies from 2 to 20. The Locking screw implants are in corresponding diameter ranges from 2.7mm, and 4.9 mm diameters with lengths varying as per the requirements and lengths of 10mm to 90 mm

The Non-locking cortical screw implants are in 3.5mm diameter and length range from 10mm to 80mm.

The Non-locking cancellous screws are in 6.5mm diameter and length range from 30mm to 120mm The Non-locking cannulated cancellous screws are in 4.0mm diameter and length range from 10mm to 72mm

The Lag Screw is of 12.5mm Diameter and length from 50mm to 120mm

Indications for Use

NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System are provided non-sterile.

NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System are indicated for treating fractures of various bones including the clavicle, pelvis, scapula, long bone (humerus, ulna, radius, femur, tibia and fibula), and small bone (metacarpals, metatarsals, phalanges).

The system is indicated for use in adult patients only. All implants are for single use only.

Comparison of Technological Characteristics

A comparison between the NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System and predicate devices has been performed which has resulted in demonstration of similarities in dimensional and performance criteria. The subject and predicate devices are similar in geometry and dimensions, intended and indicated uses, materials, sterilization methods, and performance.

Non-Clinical Performance Testing

NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System are of similar design and pattern as well as similar information was not necessary to demonstrate substantial equivalence. Performance of the NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System has been demonstrated through the test parameters outlined in ASTM F-382 and ASTM F-384 to conduct Static and Dynamic Four Point Bend Testing. Torsional and Axial pullout testing was conducted on the screws per ASTM F-543.

Conclusion

From the data presented, the NEBULA BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, NEBULA BRAND of DHS Plating System are substantially equivalent in performance for the indications and predicate devices identified.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.