K Number
K181139

Validate with FDA (Live)

Date Cleared
2018-07-06

(67 days)

Product Code
Regulation Number
888.3070
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OLYMPIC Posterior Spinal Fixation System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine (T1-S2/Ilium): degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), spinal tumor, and failed previous fusion (pseudarthrosis). When used as an adjunct to fusion, the OLYMPIC Posterior Spinal Fixation System is intended to be used with autograft/allograft.

In addition, the OLYMPIC Posterior Spinal Fixation System is intended for treatment of severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, having implants attached to the lumbosacral spine and/or ilium with removal of the implants after the attainment of a solid fusion. Levels of pedicle and non-pedicle fixation for these patients are L3-sacrum/ilium.

When used for posterior non-cervical pedicle and non-pediatric patients, the OLYMPIC Posterior Spinal Fixation System is intended to be used with autograft. Pediatric pediatric pedicle and non-pedicle fixation is limited to a posterior approach.

Device Description

The OLYMPIC PSFS is a top loading thoracolumbar, sacral, and iliac fixation system implanted from the posterior approach and designed to provide fixation during the fusion process. The system is composed of preassembled uniplanar screws, hooks, rods, crosslinks, and rod connectors. The system is supported by a comprehensive set of instruments to install the implants within the system.

AI/ML Overview

The provided text is a 510(k) summary for the OLYMPIC Posterior Spinal Fixation System. This document is a premarket notification for a medical device to demonstrate that it is substantially equivalent to a legally marketed predicate device.

Based on the content of the provided document, here's an analysis regarding acceptance criteria and supporting studies:

  1. A table of acceptance criteria and the reported device performance

The document does not provide a table of acceptance criteria or specific numerical performance metrics for the device. Instead, it relies on demonstrating substantial equivalence to predicate devices through a non-clinical test.

Acceptance CriteriaReported Device Performance
Not explicitly defined as numerical acceptance criteria. The underlying acceptance is likely to meet or exceed the performance of the predicate devices.The results of the Dynamic Compression Bending – ASTM F1717 evaluations indicate that the OLYMPIC PSFS is equivalent to predicate devices.
  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document describes a non-clinical test (Dynamic Compression Bending – ASTM F1717) but does not specify a sample size for this test. It also does not mention data provenance as it's a mechanical test, not data from patients.

  1. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable. The study conducted was a non-clinical mechanical test, not a study requiring expert interpretation or ground truth establishment based on human observation or diagnosis.

  1. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable. As explained above, the study was a mechanical test, not one involving human interpretation or adjudication.

  1. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a spinal fixation system (implants), not an AI-powered diagnostic tool. Therefore, the concept of human readers improving with AI assistance is not relevant to this submission.

  1. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone (algorithm only) performance study was not done. This device is a physical medical implant, not an algorithm or AI system.

  1. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This is not applicable. As it was a non-clinical mechanical test, the "ground truth" would be established by the physical properties and mechanical behavior of the materials and device in controlled laboratory settings, against established standards (ASTM F1717) and comparative predicate device performance.

  1. The sample size for the training set

This is not applicable. There was no training set as this is a physical medical device and not an AI/ML or software device that undergoes a training phase.

  1. How the ground truth for the training set was established

This is not applicable, as there was no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Astura Medical, LLC Mr. Parker Kelch Quality Manager 3168 Lionshead Ave, Suite 100 Carlsbad, California 92010

Re: K181139

Trade/Device Name: OLYMPIC Posterior Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw Systems Regulatory Class: Class II Product Code: NKB, KWP Dated: April 27, 2018 Received: April 30, 2018

Dear Mr. Kelch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'neill -S
for MNM

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181139

Device Name OLYMPIC Posterior Spinal Fixation System

Indications for Use (Describe)

The OLYMPIC Posterior Spinal Fixation System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine (T1-S2/Ilium): degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), spinal tumor, and failed previous fusion (pseudarthrosis). When used as an adjunct to fusion, the OLYMPIC Posterior Spinal Fixation System is intended to be used with autograft/allograft.

In addition, the OLYMPIC Posterior Spinal Fixation System is intended for treatment of severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, having implants attached to the lumbosacral spine and/or ilium with removal of the implants after the attainment of a solid fusion. Levels of pedicle and non-pedicle fixation for these patients are L3-sacrum/ilium.

When used for posterior non-cervical pedicle and non-pediatric patients, the OLYMPIC Posterior Spinal Fixation System is intended to be used with autograft. Pediatric pediatric pedicle and non-pedicle fixation is limited to a posterior approach.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
AAUHILILE AIL I ARREBIER BLAN IN LINEARR

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: OLYMPIC Posterior Spinal Fixation System

Date PreparedApril 27, 2018
Submitted ByAstura Medical3186 Lionshead Ave, Suite 100Carlsbad, CA 92010Phone: 760-814-8047
ContactParker Kelch3186 Lionshead Ave, Suite 100Carlsbad, CA 92010Phone: 760-814-8047 x413Email: quality@asturamedical.com
Trade NameOLYMPIC Posterior Spinal Fixation System
Common NamePosterior spinal fixation system
Classification NameThoracolumbosacral pedicle screw systemSpinal interlaminal fixation orthosis
ClassII
Product CodeNKBKWP
CFR Section21 CFR section 888.307021 CFR section 888.3050
Device PanelOrthopedic
Primary Predicate DeviceOLYMPIC PSFS (K153446)
Secondary Predicate Device(s)Revere®/CREO™ Stabilization System - Globus Medical (K143633 / K124058 / K113395)Xia Spinal System - Stryker (K142381 /K113666 / K071373 / K060748)Expedium/Viper - DePuy Spine (K111136 / K131802)Universal Spinal System / Matrix - Synthes (K111358 / K100952) MossMiami SS - DePuy Spine (K000536, K955348)Synergy VLS - open - Cross Medical (K940631 / K950099)PWB (now Synergy) - Cross Medical (K920116)CD HORIZON LEGACY – Medtronic (K090390 / K162494 / K162379)
Device DescriptionThe OLYMPIC PSFS is a top loading thoracolumbar, sacral, and iliacfixation system implanted from the posterior approach and designed toprovide fixation during the fusion process. The system is composed ofpreassembled uniplanar screws, hooks, rods, crosslinks, and rodconnectors. The system is supported by a comprehensive set ofinstruments to install the implants within the system.
MaterialsTi-6Al-4V ELI (ASTM F136)
CP Titanium Grade 4 (ASTM F67)
Nitinol #1 (ASTM F2063)
SS 17-4 (ASTM A564)
SS 420 (ASTM A240)
SubstantialEquivalence Claimedto Predicate DevicesThe OLYMPIC PSFS is substantially equivalent to the predicate devices interms of intended use, design, materials used, mechanical safety andperformances.
Indications for UseThe OLYMPIC Posterior Spinal Fixation System is intended to provideimmobilization and stabilization of spinal segments in skeletally maturepatients as an adjunct to fusion in the treatment of the following acute andchronic instabilities or deformities of thoracic, lumbar, and sacral spine(T1-S2/Ilium): degenerative spondylolisthesis with objective evidence ofneurological impairment, fracture, dislocation, deformities or curvatures(i.e. scoliosis, kyphosis, and/or lordosis), spinal tumor, and failed previousfusion (pseudarthrosis). When used as an adjunct to fusion, theOLYMPIC Posterior Spinal Fixation System is intended to be used withautograft/allograft.
In addition, the OLYMPIC Posterior Spinal Fixation System is intendedfor treatment of severe spondylolisthesis (grades 3 and 4) of the L5-S1vertebra in skeletally mature patients receiving fusion by autogenous bonegraft, having implants attached to the lumbosacral spine and/or ilium withremoval of the implants after the attainment of a solid fusion. Levels ofpedicle and non-pedicle fixation for these patients are L3-sacrum/ilium.
When used for posterior non-cervical pedicle and non-pedicle fixation inpediatric patients, the OLYMPIC Posterior Spinal Fixation System isintended to be used with autograft and/or allograft. Pediatric pedicle andnon-pedicle fixation is limited to a posterior approach.
Non-clinical TestSummaryThe system was analyzed through Dynamic Compression Bending – ASTMF1717.The results of these evaluations indicate that the OLYMPIC PSFS isequivalent to predicate devices.
Clinical TestSummaryNo clinical studies were performed.
Conclusions: Non-Clinical and ClinicalAstura Medical considers OLYMPIC PSFS to be equivalent to the predicatedevices listed above. This conclusion is based upon the devices' similaritiesin principles of operation, technology, materials and indications for use

In accordance with 21 CFR 807.92 of the Federal Code of Regulations

{4}------------------------------------------------

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.