(419 days)
The EXPEDIUM and VIPER/VIPER2 Spine Systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.
The EXPEDIUM and VIPER/VIPER2 Spine System metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used in a posterior percutaneous approach with MIS instrumentation, the VIPER/VIPER2 System metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM and VIPER/VIPER2 System metallic implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. The EXPEDIUM and VIPER/VIPER2 systems are intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The EXPEDIUM and VIPER/VIPER2 PEEK rods are only indicated for fusion procedures for spinal stenosis with instability (no greater than Grade I spondylolisthesis) from LI-SI in skeletally mature patients.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM and VIPER/VIPER2 System metallic implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The EXPEDIUM and VIPER/VIPER2 systems are intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The subject EXPEDIUM® Spine System and VIPER®/VIPER®2 Systems are pedicle screw scoliosis systems for the treatment of pediatric patients consisting of a wide range of components in a variety of geometries and sizes.
The provided document does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria in terms of diagnostic accuracy or algorithmic performance.
Instead, this is a 510(k) summary for the "EXPEDIUM® Spine System, VIPER® System, VIPER®2 System," which are pedicle screw spinal systems. The document focuses on demonstrating substantial equivalence to previously cleared predicate devices, an approval pathway primarily concerned with the safety and effectiveness of a new device compared to an existing one, rather than novel performance benchmarks.
Here's a breakdown of what the document does state regarding performance:
-
A table of acceptance criteria and the reported device performance: This information is not present. The document does not define specific performance metrics or acceptance criteria for the device's function or a study comparing it to those criteria.
-
Sample size used for the test set and the data provenance: Not applicable. No test set or data provenance for performance evaluation is mentioned.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No ground truth establishment is described.
-
Adjudication method for the test set: Not applicable. No test set adjudication is described.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a surgical implant (pedicle screw system), not an AI-assisted diagnostic tool, so an MRMC study is irrelevant in this context.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithmic or AI device.
-
The type of ground truth used: Not applicable. No ground truth is mentioned.
-
The sample size for the training set: Not applicable. No training set is mentioned.
-
How the ground truth for the training set was established: Not applicable. No training set ground truth is mentioned.
Key statements from the document regarding performance:
- "No new testing was performed. Published clinical results and engineering analysis supported the expansion of indications." (Page 2, section "Performance Data")
- "The substantial equivalence justification demonstrates that the device is as effective, and performs as well as the predicate device." (Page 3, Conclusion)
In summary, this 510(k) submission relies on demonstrating substantial equivalence to predicate devices, drawing on existing clinical results and engineering analyses, rather than presenting a performance study with defined acceptance criteria for the device itself.
{0}------------------------------------------------
K111136 - page 1 of 3
510(K) SUMMARY
JUN 1 4 2012
| Contact: | Kevin G. StevensDePuy Spine, Inc.325 Paramount DriveRaynham, MA 02767508.828.3288 |
|---|---|
| Device Trade Name: | EXPEDIUM® Spine System, VIPER® System, VIPER®2 System |
| Manufacturer: | DePuy Spine, Inc.325 Paramount DriveRaynham, MA 02767 |
| Common Name: | Pedicle Screw System |
| Classification: | 21 CFR §888.3050; Spinal interlaminar fixation orthosis21 CFR §888.3060; Spinal intervertebral body fixation orthosis21 CFR §888.3070; Pedicle screw spinal system |
| Class: | III |
| Product Code: | MNI, MNH, NKB, KWQ, KWP, OSH |
. Indications For Use:
The EXPEDIUM and VIPER/VIPER2 Spine Systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.
The EXPEDIUM and VIPER/VIPER2 Spine System metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures
{1}------------------------------------------------
(i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used in a posterior percutaneous approach with MIS instrumentation, the VIPER/VIPER2 System metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM and VIPER/VIPER2 System metallic implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. The EXPEDIUM and VIPER/VIPER2 systems are intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Device Description:
The subject EXPEDIUM® Spine System and VIPER®/VIPER®2 Systems are pedicle screw scoliosis systems for the treatment of pediatric patients consisting of a wide range of components in a variety of geometries and sizes.
Predicate Device(s):
The EXPEDIUM® Spine System and VIPER®/VIPER®2 System were shown to be substantially equivalent to previously cleared predicate devices.
Performance Data:
No new testing was performed. Published clinical results and engineering analysis supported the expansion of indications.
Materials:
The subject EXPEDIUM Spine System components have been previously cleared in commercially pure titanium or titanium alloy conforming to ASTM F-67, ASTM F-136 or ASTM F-1472 specifications, stainless steel conforming to ASTM F-138, ASTM F-1314, or F-2229 specifications, cobalt-nickelchromium- molybdenum alloy wire conforming to ASTM F-562 specifications, as well as longitudinal rods in cobalt-chromium-molybdenum alloy conforming to ASTM F-1537.
The subject VIPER/VIPER2 System components have been previously cleared in titanium alloy conforming to ASTM F-136 specifications, stainless steel conforming to ASTM F-138 or ASTM F-1314
{2}------------------------------------------------
KII 1136 page 3 of 3
specifications, as well as longitudinal rods in cobalt-chromium-molybdenum alloy conforming to ASTM F-1537.
Conclusion:
The substantial equivalence justification demonstrates that the device is as effective, and performs as well as the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is depicted in black, and the text is also in black, set against a white background. The overall design is simple and recognizable, representing the department's focus on health and human welfare.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 27, 2013
DePuy Spine, Inc. % Mr. Kevin G. Stevens 325 Paramount Drive Raynham, Massachusetts 02767
Re: K111136
Trade/Device Name: EXPEDIUM Spine System, VIPER System, VIPER2 System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNI, MNH, KWP, KWQ Dated: June 08, 2012 Received: June 11, 2012
Dear Mr. Stevens:
This letter corrects our substantially equivalent letter of June 14, 2012.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Kevin G. Stevens
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/8 description: The image shows the name "Erin Dreith" in a stylized font. The first name, "Erin," is written in a bold, sans-serif font. The last name, "Dreith," is written in a more decorative font, with the letters overlapping and outlined.
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
EXPEDIUM Spine System and VIPER Systems 510(k)
Indications for Use 4.
510(k) Number (if known): K | | | 36
Device Name: EXPEDIUM® Spine System, VIPER® System, VIPER®2 System
The EXPEDIUM and VIPER/VIPER2 Spine Systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.
The EXPEDIUM and VIPER/VIPER2 Spine System metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvaures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used in a posterior percutaneous approach with MIS instrumentation, the VIPER/VIPER2 System metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
The EXPEDIUM and VIPER/VIPER2 PEEK rods are only indicated for fusion procedures for spinal stenosis with instability (no greater than Grade I spondylolisthesis) from LI-SI in skeletally mature patients.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM and VIPER/VIPER2 System metallic implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The EXPEDIUM and VIPER/VIPER2 systems are intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Prescription Use V AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
KITI 36 510(k) Number_
N/A