(269 days)
The DePuy Synthes Reconstructive Plates 'Y' Plates are indicated for pelvic and acetabular reconstructive surgery and the reduction of fractures in the distal humerus, the clavicle, and the scapula.
The DePuy Synthes Anatomical Locking Plate System indicated to treat fractures (including but not limited to) the radius, ulna, humerus, tibia, fibula, femur.
The DePuy Synthes CABP System is a plate and screw system intended to treat fractures of the proximal humerus and distal tibia. Proximal humerus fractures include two-part greater tubercle fractures and fracture dislocations, two-part surgical neck fractures and fracture dislocations, three-part fractures or fracture dislocations, fractures in osteopenic bone, and nonunions and malunions. Distal tibia fractures include acute fractures in osteopenic bone, and nonunions and malunions.
The DePuy Synthes 3.5 mm 90° Cannulated Limited Contact Angled Blade Plates are intended for the fixation of fractures and non-unions of the proximal humerus and distal tibia.
The DePuy Synthes Large Fragment DCL is intended for fixation of various long bones, such as the humerus, femur and tibia. Also for use in fixation of osteopenic bone and fixation of non-unions or malunions.
The DePuy Synthes Large Fragment Locking Compression Plate (LCP) System - T Plate is intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. Also for use in fixation of osteopenic bone and fixation of non-unions and malunions.
The DePuy Synthes Curved Reconstruction Plate is intended for fixation of fractures, osteotomies, and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, and fibula, particularly in osteopenic bone as a part of the DePuy Synthes Small Fragment DCL System.
The DePuy Synthes One-Third Tubular Plate is intended for fixation of fractures, osteotomies, and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, and fibula, particularly in osteopenic bone as a part of the DePuy Synthes Small Fragment DCL System.
The DePuy Synthes LCP Proximal Humerus Plate is indicated for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
The DePuy Synthes Pilon Plate is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal tibia.
DePuy Synthes TomoFix™ Osteotomy System is intended for open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia and lateral distal femur, treatment of bone and joint deformities, and malalignment caused by injury or disease such as osteoarthritis.
The DePuy Synthes 3.5 mm Broad LCP & 4.5 mm LCP Distal Humerus Plates are indicated for fractures of the distal humerus.
The DePuy Synthes Clavicle Hook Plate is intended for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.
The DePuy Synthes 3.5 / 4.5mm LCP® Metaphyseal Plate is intended for the fixation of various long bones, excluding the femur. It is also for use in fixation of osteopenic bone and fixation of nonunions or malunions.
The DePuy Synthes 3.5 mm LCP Distal Humerus System is indicated for intraarticular fractures of the distal humerus, commninuted supra-condylar fractures, osteotomies, and non-unions of the distal humerus.
The DePuy Synthes LCP Proximal Humerus Plate, Long is indicated for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
The DePuy Synthes Curved Broad Plates are intended for fixation of various long bones, such as the humerus, femur, and tibia. They are also for use in fixation of peri-prosthetic fractures, osteopenic bone and non-unions or malunions. The DePuy Synthes Curved Condylar Plates are intended for buttressing multifragmentary distal femur fractures, including: supracondylar, intra-articular and extra-articular condylar fractures, peri-prosthetic fractures and fractures in normal or osteopenic bone, nonunions/malunions, and osteotomies of the femur.
The DePuy Synthes Low Profile Reconstruction Plates are intended for pelvic and acetabular reconstruction and fixation of fractures of the distal humerus, clavicle, and scapula.
The DePuy Synthes 3.5 mm Spring Plate is intended for pelvic and acetabular reconstructive surgery and fracture fixation of the distal fibula.
The DePuy Synthes 3.5 mm LCP Hook Plate is indicated for fractures, osteotomies and nonunions of small bones including the ulna, radius, tibia and fibula, particularly in osteopenic bone.
The DePuy Synthes 3.5 mm LCP Periarticular Proximal Humerus Plates are indicated for fractures, fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
The DePuy Synthes 3.5 mm Locking Compression Plate (LCP) System: The DePuy Synthes 3.5 mm LCP is indicated for fixation of fractures, osteotomies and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, fibula, particularly in osteopenic bone for adult patients. These plates are also indicated for fracture fixation of diaphyseal and metaphyseal areas of long bones in pediatric patients. The DePuy Synthes 4.5 mm Locking Compression Plate (LCP) System: The DePuy Synthes 4.5 mm LCP is indicated for fixation of various long bones, such as the humerus, femur and tibia and for use in fixation of peri-prosthetic fractures, osteopenic bone and fixation of non-unions or malunions in adult patients. These plates are also indicated for fracture fixation of diaphyseal and metaphyseal areas of long bones in pediatric patients.
The DePuy Synthes 3.5mm Locking Attachment Plate is intended for use with DePuy Synthes LCP plates to augment the stabilization of fractures, including periprosthetic fractures and fractures in the presence of intramedullary implants, in the femur, tibia and humerus, particularly in osteopenic bone.
The DePuy Synthes 3.5 mm Curved Narrow and Broad LCP Plates are intended for fixation of fractures, osteotomies and non-unions of clavicle, scapula, olecranon, humerus, radius, pelvis, distal tibia and fibula, particularly in osteopenic bone for adult patients. The DePuy Synthes 4.5 mm Curved Narrow and Broad LCP Plates are intended for fixation of various long bones, such as the humerus, femur, and tibia. They are also for use in fixation of periprosthetic fractures, osteopenic bone, and non-unions or malunions in adult patients. The DePuy Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad LCP Plates are also indicated for fracture fixation of diaphyseal and metaphyseal areas of long bones in pediatric patients.
The DePuy Synthes 2.7 mm / 3.5 mm VA-LCP Anterior Clavicle Plate System is indicated for fixation of fractures, malunions, non-unions, and osteotomies of the clavicle in adults, and in both adolescents (12-18 years) and transitional adolescents (18-21 years), in which the clavicular growth plates have fused or in which the growth plates will not be crossed by the plate system.
The DePuy Synthes 3.5 mm LCP Clavicle Plate System is indicated for fixation of fractures, non-unions, and osteotomies of the clavicle in adults, and in both adolescents (12-18 years) and transitional adolescents (18-21 years), in which clavicular growth plates have fused or in which the growth plates will not be crossed by the plate system.
The DePuy Synthes Variable Angle LCP Elbow System is intended for fixation of fractures of the distal humerus, olecranon and ulna in adults and adolescents (12-21) in which the growth plates have fused or in which the growth plates will not be crossed by the plate system. Specifically, Distal humerus plates are indicated for intra-articular fractures, commninuted supracondylar fractures, osteotomies, malunions and non-unions of the distal humerus. Olecranon and Proximal ulna plates are indicated for fractures, osteotomies, malunions and non-unions of the olecranon and proximal ulna.
The DePuy Synthes Variable Angle LCP Elbow System is intended for fixation of fractures of the distal humerus, olecranon and ulna in adults and adolescents (12-21) in which the growth plates have fused. Specifically, Distal humerus plates are indicated for intra-articular fractures, comminuted supracondylar fractures. malunions and non-unions of the distal humerus. Olecranon and Proximal ulna plates are indicated for fractures, malunions and non-unions of the olecranon and proximal ulna.
The DePuy Synthes TOMOFIX Osteotomy System is intended for osteotomies, treatment of bone and joint deformities, fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia. Specifically, -The TOMOFIX Medial Proximal Tibia Plates are indicated for open- and close-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the medial proximal tibia -The TOMOFIX Lateral Proximal Tibia Plates are indicated for open- and closed-wedge osteotomies fixation of fractures. and malalignment caused by injury or disease, such as osteoarthritis, of the lateral proximal tibia -The TOMOFIX Lateral Distal Femur Plates are indicated for open- and closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the lateral distal femur -The TOMOFIX Medial Distal Femur Plates are indicated for closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the medial distal femur
The DePuy Synthes 2.4/2.7 mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long is intended for intra- and extra-articular fractures, osteotomies, nonunions and malunions of the distal radius, with or without extension into the radial diaphysis in adults and adolescents (12-21) where the growth plates have fused or will not be crossed, and the following adolescent distal radius fractures: - Intra-articular fractures exiting the epiphysis - Intra-articular fractures exiting the metaphysis - Physeal crush injuries - Any injuries which cause growth arrest to the distal radius
The 3.5 mm Reconstruction Plate is available in multiple lengths. It has a smooth underside, is notched between the screw holes and can be adjusted (contoured) in three dimensions. The screw holes are designed to provide limited bidirectional compression of the fracture via the tapered entrance of the hole, and accept 3.5 mm cortex screws. The 4.5 mm Reconstruction Plate has a smooth underside, is notched between the screw holes and can be adjusted (contoured) in three dimensions. The oval holes are designed to provide limited bidirectional compression of the fracture via the tapered entrance of the hole which allows 25° longitudinal and 7° lateral angulation of screws. It is available in multiple lengths and accepts 4.5 mm cortex and 6.5 mm cancellous screws. The 3.5 mm Curved Reconstruction Plate is available in multiple lengths. It has a smooth underside and is notched between screw holes and can be adjusted (contoured) in three dimensions. The screw holes are designed to provide limited bidirectional compression of the fracture via the tapered entrance of the hole and accept 3.5 mm cortex screws.
The Anatomical Locking Plate System utilizes a locking feature that secures the screw to the plate, enabling stable fixation to be achieved via unicortical fixation. The plates are available with and without an anatomically contoured head, in a variety of sizes. The head of the screw has a round cross section with a conical underside to fit into the plate. The underside has a tapered thread to match the design of the holes in the plate. The screws are available in thread diameters from 3 mm to 5 mm, in a variety of lengths.
The CABP is a straight plate with a blade at the head to allow for better fixation in the head of the humerus or in the distal tibia. The blade of the plate is cannulated to fit over a guide wire, allowing for the adjustment of the wire placement several times without adversely affecting the final result. There are cuts in the undersurface of the plate the surface area of the plate in contact with bone. The area of contact between the plate and the bone is decreased in an effort to reduce damage to the cortical blood supply under the plat, and resultant reduction in damage-induced porosis and remodeling of the bone near the plate undersurface. Plate undercuts also make the bending properties of the plate more uniform, which facilitate contouring. The plate has round and dynamic compression screw holes, accepts 4.5mm and 6.5 mm screws, is available in multiple blade lengths, and is manufactured from stainless steel and commercially pure titanium.
The DePuy Synthes 3.5 mm 90° Cannulated LC-ABP provides stable fracture fixation and rotational control for fractures of the proximal humerus and distal tibia. The plates feature a low profile limited contact dynamic compression plate (LC-DCP®) design and dynamic compression unit (DCU) screw holes. The blade portion of the plate is cannulated to accept a 2.0 mm guide wire. The plates are available in various sizes to accommodate varying patient anatomy.
The Large Fragment DCL system consists of limited-contact profile plates in broad and narrow sizes, which include combination dynamic compression/locking screw holes. The plates accept 4.5 mm cortex, 6.5 mm cancellous, 4.5 mm cannulated, 7.0 mm cannulated, and 5.0 mm locking screws. This device is manufactured in either stainless steel or titanium.
The DePuy Synthes Locking Compression Plate (LCP) System - T Plate is a buttress plate and screw system. The primary feature of the plate is round holes combined with locking screw holes. The locking screws form a locked, fixed angle construct with the plate, while the standard screws facilitate reduction and create compression between the plate and bone. The plates accept 4.5 mm cortex, 6.5 mm cancellous, 4.5 mm cannulated, 7.0 mm cannulated, 7.3 mm cannulated, 4.0 mm and 5.0 mm locking screws.
The DePuy Synthes Curved Reconstruction Plate line extension is a pre-curved version of the currently marketed Straight Locking Reconstruction Plate. Both of these plates will be included as part of the DePuy Synthes Small Fragment DCL System. Both the curved and straight plates have the same intended use and there is no change in safety or efficacy.
The DePuy Synthes One-Third Tubular DCL Plate line extension is a threaded version of the currently marketed Synthes Third Tubular Plate. The threaded plates will accept locking screws and therefore can be included as part of the Small Fragment DCL System. The new plates have the same intended use as other plates in the system and there is no change in safety or efficacy.
The DePuy Synthes LCP Proximal Humerus Plates are designed to match the anatomy of the proximal humerus. These plates can be applied to either the right or left humerus. The proximal portion of each plate has threaded holes that accept 3.5 mm or 2.7 mm screws. The distal portion of the plate has combination holes that allow the option of using 3.5 mm locking or cortex screws, or 4.0 mm cancellous screws to accomplish plate fixation. These plates will be offered as an addition to the DePuy Synthes Small Fragment LCP (formerly DCL) System.
The DePuy Synthes Pilon Plate is a metal plate that offers screw to plate locking designed for various fracture modes of the distal end of the tibia.
The TomoFix™ Osteotomy System consists of five different titanium plates with locking and combination holes. There are two plates (left and right) for the lateral distal femur, 2 plates (left and right) for the lateral proximal tibia, and 1 plate for the medial proximal tibia.
The DePuy Synthes 3.5 mm Broad LCP & 4.5 mm Broad LCP Distal Humerus Plates are contoured to match the anatomy of the distal humerus with a limited contact low profile design. The plate has dynamic compression holes combined with locking holes which accept 3.5 & 4.5 mm cortex, 3.5 & 4.5 mm self-tapping cortex, 3.5 mm shaft, 3.5,4.0, & 5.0 mm locking, and 4.0 mm cancellous screws. The plates are available in various lengths.
The DePuy Synthes Clavicle Hook Plate is an anatomically designed L-shaped plate featuring a distal joint bridging hook with a rounded end. The shaft of the plate contains either dynamic compression screw holes or LCP plate holes. The plates accept 3.5 mm cortex, 4.0 mm cancellous and 3.5 mm locking screws (LCP Clavicle Hook Plate only).
The DePuy Synthes 3.5 / 4.5 mm LCP® Metaphyseal Plates are contourable to match the anatomy, have a limited contact design, and are tapered at the plate head. The 3.5 / 4.5 mm metaphyseal plates feature combination dynamic compression / locking screws holes (combi-holes). The plates in this system accept 3.5 / 4.5 mm cortex, 3.5, 4.0, & 5.0 mm locking, and 4.0 mm or 6.5 mm cancellous screws. The plate also has a 2.0 mm hole for preliminary fixation with k-wires.
The DePuy Synthes 3.5 mm LCP Distal Humerus System consists of medial and posterolateral distal humerus plates of various lengths and 2.7 mm locking screws. The plates are pre-contoured to match the anatomy of the distal humerus with a limited contact low profile design. The plate features locking compression holes and conical locking holes which accept 2.4, 3.5, & 4.0 mm cortex screws, 2.4, 2.7 & 3.5 mm locking screws, and 4.0 mm cancellous screws. The System will be available in Stainless Steel and Titanium.
The DePuy Synthes LCP Proximal Humerus Plates, long are pre-contoured to match the anatomy of the proximal humerus with a limited contact low profile design. The plate features locking holes and combination locking and compression holes which accept 2.7, 3.5, & 4.0 mm cortex screws, 3.5 mm locking screws, and 4.0 mm cancellous screws. The System will be available in Stainless Steel and Titanium. These plates will be offered as an addition to the DePuy Synthes Small Fragment LCP (formerly DCL) System.
The DePuy Synthes LCP® Curved Plates have a slight curve to better match the anatomy of the bone. The plates have a limited contact profile design and includes combination dynamic compression/locking screw holes.
The DePuy Synthes 3.5 mm Low Profile Reconstruction Plates consist of straight, curved, and J-shaped plates. The shape and profile of these plates are designed to minimize soft tissue irritation and minimize the need for intraoperative contouring. However, when intraoperative contouring is required for a precise fit with the bone, the plate design allows for simple and uniform bending.
The DePuy Synthes 3.5mm Spring Plate is a variation of the DePuy Synthes one-third tubular plate with collar which utilizes two sharp spikes at the bottom surface and a pre-bent convex shape to aid in the reduction of small bone fragments while conforming to bony uneven surfaces. The plate incorporates a 1 - 10 hole design in lengths ranging from 19.5mm - 132mm and accepts either 3.5mm cortex or pelvic screws. In addition the plates are manufactured from Stainless Steel and Commercially Pure Titanium and provided STERILE and NON STERILE.
The DePuy Synthes 3.5 mm LCP Hook Plate is a low-profile, precontoured bone fixation plate intended for the treatment of fractures, osteotomies, and non-unions in small bones. The plates feature a preformed dual hook on one end and are available in stainless steel and titanium.
The DePuy Synthes 3.5 mm LCP Periarticular Proximal Humerus Plates are low-profile, precontoured bone fixation plates intended for the treatment of fractures of the humerus. The plates are offered sterile and non-sterile and are available in stainless steel and titanium.
The DePuy Synthes 3.5 mm and 4.5 mm LCP Plates with Expanded Indications consist on.5 mm LCP plates, 4.5 rom Narrow LCP Plates, 4.5 mm Broad LCP plates and 4.5 mm Curved Broad LCP Plates for fracture fixation in adults and pediatric patients. These plates accept locking, cortex and cancellous screws.
The DePuy Synthes 3.5mm Locking Attachment Plates are low-profile, pre-contoured plates which are intended to be used with existing Synthes to stabilize fractures. The plates are available in stainless steel and titanium.
The DePuy Synthes Curved Narrow and Broad LCP Plates are available in stainless steel and titanium, and consist of limited-contact profile plates in 3.5 mm and 4.5mm narrow and broad sizes. The plates feature Dynamic Compression Plate (DCP) holes combined with locking screw holes. The 3.5mm plates accept 3.5mm cortex and locking screws and 4.0 mm cancellous screws, and the 4.5mm plates accept 4.5 mm cortex screws, 4.0mm and 5.0mm locking screws, 4.5 mm cannulated screws, 5.0 mm periprosthetic screws, and 6.5 mm cancellous screws.
The DePuy Synthes 2.7 mm / 3.5 mm VA-LCP Anterior Clavicle Plate System consists of plates of various lengths and variable angle screws that provide the flexibility to lock screws in trajectories that can diverge from the central axis of the plate hole. The system features medial and lateral plates that are pre-contoured to match the anatomy of the clavicle and a limited contact, low profile design. The plate with lateral extension features Variable Angle (VA) Locking holes and Dynamic Compression Plate (DCP) holes. The medial plate features Dynamic Compression Plate (DCP) holes combined with VA locking screw holes.
The DePuy Synthes 3.5 mm LCP Clavicle Plate System consists of metallic plates and screws that offer screw to plate locking designed for various fracture modes of the clavicle.
The DePuy Synthes Variable Angle LCP Elbow System contains plates intended to treat fractures of the distal humerus and proximal ulna. A variety of plate configurations are included in the system to allow for fixation of multiple fracture patterns. Specifically, the system includes several plate configurations for fixation of the distal humerus which are intended to be used in a two-plate construct where plates are positioned medially and laterally. Additionally, the system includes plates for fixation of the olecranon and proximal ulna. In its entirety, the following plate types are included in the system: - -Medial Distal Humerus Plate - -Lateral Distal Humerus Plate - -Olecranon Plate - -Proximal Olecranon Plate Extra-articular Proximal Ulna Plate The system accepts existing conical and locking screws as well as new metaphyseal screws, and allows for both dynamic compression and locking through Combi holes. The plates are universally designed for both left and right use and will be offered in both stainless steel and titanium.
The DePuy Synthes Variable Angle LCP Elbow System contains posterolateral and medial plates intended to treat fractures of the distal humerus. The plates are used together in a two-plate, 90° construct and accept existing screws. New 2.7mm Metaphyseal Screws are also compatible with the System.
The DePuy Synthes TOMOFIX Osteotomy System consists of five different titanium plate families with locking and combination locking/compression holes. This system features plates designed to provide stable fixation of osteotomies of the distal femur and proximal tibia. The DePuy Synthes TOMOFIX Medial Distal Femur Plates are part of the DePuy Synthes TOMOFIX Osteotomy System used for closed wedge femoral osteotomies. The subject plates are anatomically contoured to fit the medial distal femur, are available in right and left versions, as well as sterile and non-sterile.
The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long will be available in stainless steel and titanium. The head of the plate will remain the same as the existing 2.4mm VA-LCP Two-Column Volar Distal Radius Plate (K102694) and will therefore allow the use of the existing guide block for these plates. The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long will come in 3 different lengths (7 shaft holes, 10 shaft holes and 13 shaft holes). To accommodate different patient anatomy, the plate with 7 shaft holes will be available with three different plate head widths (Standard, Narrow and Wide). The 10 shaft hole and the 13 shaft hole plate will only be released with the Standard plate head width and additionally they feature a curvature in the shaft. All plates will feature left- and right-specific designs and will be offered in sterile packed only. The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plates, Extra-Long are designed to accept existing 2.4mm Variable Angle Locking Screws, 2.4mm Cortex Screws, 2.4mm Locking Screws, 1.8mm VA Buttress Pins and 1.8mm Locking Buttress Pins in the plate head holes. In the plate shaft, existing 2.4mm Cortex Screws, 2.7mm Cortex Screws, 2.4mm Variable Angle Locking Screws, 2.7mm Variable Angle Locking Screws, 2.4mm Locking Screws, and 2.7mm Locking Screws can be inserted.
The provided text describes several DePuy Synthes Trauma Orthopedic Plates and Screws that have received FDA 510(k) clearance (K180310) for "MR Conditional" labeling. The primary study presented for proving the device meets acceptance criteria is a series of non-clinical tests to assess the device's compatibility with the Magnetic Resonance (MR) environment.
Here's an organized breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The acceptance criteria for "MR Conditional" labeling are based on established ASTM (American Society for Testing and Materials) standards. The reported device performance is that the devices meet these standards.
| Acceptance Criteria (based on ASTM Standards) | Reported Device Performance |
|---|---|
| Magnetically Induced Displacement Force: (ASTM F2052-14) | The non-clinical data demonstrate the subject devices raise no new questions of safety or efficacy. |
| Magnetically Induced Torque: (ASTM F2213-06) | The non-clinical data demonstrate the subject devices raise no new questions of safety or efficacy. |
| Radio Frequency (RF) Heating: (ASTM F2182-11a) | The non-clinical data demonstrate the subject devices raise no new questions of safety or efficacy. |
| Image Artifacts: (ASTM F2119-07) | The non-clinical data demonstrate the subject devices raise no new questions of safety or efficacy. |
Explanation: For each device covered under this 510(k) submission, the document explicitly states: "Non-clinical testing is provided to support the conditional safety of the [Device Name] in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy." This indicates that the devices met the requirements of these ASTM standards, thus satisfying the acceptance criteria for MR Conditional labeling.
2. Sample Size Used for the Test Set and Data Provenance:
The document primarily refers to non-clinical testing. This type of testing typically involves physical device samples rather than human or animal subjects for evaluating MR compatibility. The document does not specify a numerical sample size for the devices tested for MR compatibility. The provenance is implied to be through laboratory testing under controlled conditions as per ASTM standards. There is no mention of country of origin of data in this context.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
This information is not provided in the document. For non-clinical testing of MR compatibility, the "ground truth" is typically established by the specifications of the ASTM standards themselves, interpreted and performed by testing engineers or technicians with expertise in medical device testing and MR safety. No specific number or qualifications of such experts are mentioned.
4. Adjudication Method for the Test Set:
This information is not provided in the document. As the testing is non-clinical and based on quantitative ASTM standards, a human adjudication method (like 2+1, 3+1) is typically not applicable. The results would be objectively evaluated against the pass/fail criteria defined by the ASTM standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
No, a MRMC comparative effectiveness study was not done. The submission focuses on the "MR Conditional" status of the orthopedic plates and screws, which relates to their physical properties in an MR environment, not their diagnostic performance or how they impact human reader interpretation. Therefore, there is no mention of an effect size for human readers with or without AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:
Not applicable. The device is a physical orthopedic implant (plates and screws), not an algorithm or an AI-powered diagnostic tool. Therefore, there is no "standalone performance" of an algorithm.
7. The Type of Ground Truth Used:
The ground truth for the non-clinical MR compatibility testing is based on established testing standards (ASTM F2052-14, ASTM F2213-06, ASTM F2182-11a, and ASTM F2119-07). These standards define the methods and acceptance criteria for classifying a device as "MR Conditional."
8. The Sample Size for the Training Set:
Not applicable. As the tests are non-clinical, there is no "training set" in the context of machine learning or AI models. The testing involves physical samples under controlled conditions.
9. How the Ground Truth for the Training Set Was Established:
Not applicable. No training set was used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Synthes (USA) Products LLC/DePuy Orthopaedics Inc Jeffrey Krawiec Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380
November 1, 2018
Re: K180310
Trade/Device Name: DePuy Synthes Trauma Orthopedic Plates and Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC, KTT, KTW, LXT Dated: September 28, 2018 Received: October 1, 2018
Dear Jeffrey Krawiec:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products, and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed Jesse by Jesse Muir -S Date: 2018.11.01 Muir -S 18:42:03 -04'00' For: Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes Reconstructive Plates 'Y' Plates - MR Conditional
Indications for Use (Describe)
The DePuy Synthes Reconstructive Plates 'Y' Plates are indicated for pelvic and acetabular reconstructive surgery and the reduction of fractures in the distal humerus, the clavicle, and the scapula.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes Anatomical Locking Plate System - MR Conditional
Indications for Use (Describe)
The DePuy Synthes Anatomical Locking Plate System is indicated to treat fractures (including but not limited to) the radius, ulna, humerus, tibia, fibula, femur.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{4}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes Cannulated Angle Blade Plate (CABP) System - MR Conditional
Indications for Use (Describe)
The DePuy Synthes CABP System is a plate and screw system intended to treat fractures of the proximal humerus and distal tibia.
Proximal humerus fractures include two-part greater tubercle fracture dislocations. two-part surgical neck fractures and fracture dislocations, three-part fracture dislocations, fractures in osteopenic bone, and nonunions and malunions.
Distal tibia fractures include acute fractures in osteopenic bone, and nonunions and malunions.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 3.5 mm 90° Cannulated Limited Contact-Angled Blade Plates – MR Conditional
Indications for Use (Describe)
The DePuy Synthes 3.5 mm 90° Cannulated Limited Contact Angled Blade Plates are intended for the fixation of fractures and non-unions of the proximal humerus and distal tibia.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes Large Fragment Dynamic Compression Locking (DCL) System - MR Conditional
Indications for Use (Describe)
The DePuy Synthes Large Fragment DCL is intended for fixation of various lones, such as the humerus, femur and tibia. Also for use in fixation of osteopenic bone and fixation of non-unions or malunions.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{7}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes Large Fragment Locking Compression Plate (LCP) System - T Plate - MR Conditional
Indications for Use (Describe)
The DePuy Synthes Large Fragment Locking Compression Plate (LCP) System - T Plate is intended to buttress metaphyseal fractures of the proximal humerus, medial tibia. Also for use in fixation of osteopenic bone and fixation of non-unions and malunions.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{8}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes Curved Reconstruction Plate - MR Conditional
Indications for Use (Describe)
The DePuy Synthes Curved Reconstruction Plate is intended for fixation of fractures, and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, particularly in osteopenic bone as a part of the DePuy Synthes Small Fragment DCL System.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{9}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes One-Third Tubular Dynamic Compression Locking (DCL) Plate - MR Conditional
Indications for Use (Describe)
The DePuy Synthes One-Third Tubular Plate is intended for fixation of fractures, and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, particularly in osteopenic bone as a part of the DePuy Synthes Small Fragment DCL System.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{10}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes LCP Proximal Humerus Plate - MR Conditional
Indications for Use (Describe)
The DePuy Synthes LCP Proximal Humerus Plate is indicated for fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{11}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes Pilon Plate - MR Conditional
Indications for Use (Describe)
The DePuy Synthes Pilon Plate is intended for fixation of complex intra- and extra-articular fractures of the distal tibia.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{12}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes TomoFix Osteotomy System - MR Conditional
Indications for Use (Describe)
DePuy Synthes TomoFix Osteotomy System is intended for open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia and lateral distal femur, treatment of bone and joint deformities, and malalignment caused by injury or disease such as osteoarthritis.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{13}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 3.5 mm Broad LCP & 4.5 mm LCP Distal Humerus Plate – MR Conditional
Indications for Use (Describe)
The DePuy Synthes 3.5 mm Broad LCP & 4.5 mm LCP Distal Humerus Plates are indicated for fractures of the distal humerus.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{14}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes Clavicle Hook Plate - MR Conditional
Indications for Use (Describe)
The DePuy Synthes Clavicle Hook Plate is intended for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{15}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 3.5 / 4.5 mm LCP® Metaphyseal Plates - MR Conditional
Indications for Use (Describe)
The DePuy Synthes 3.5 / 4.5 LCP Metaphyseal Plate is intended for fixation of various long bones, excluding the femur. It is also for use in fixation of osteopenic bone and fixation of nonunions or malunions.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{16}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 3.5 mm LCP Distal Humerus System - MR Conditional
Indications for Use (Describe)
The DePuy Synthes 3.5 mm LCP Distal Humerus System is indicated for intraarticular fractures of the distal humerus, commninuted supra-condylar fractures, osteotomies, and non-unions of the distal humerus.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{17}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes LCP® Proximal Humerus Plates, Long - MR Conditional
Indications for Use (Describe)
The DePuy Synthes LCP Proximal Humerus Plate, Long is indicated for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{18}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes LCP® Curved Plates - MR Conditional
Indications for Use (Describe)
The DePuy Synthes Curved Broad Plates are intended for fixation of various long bones, such as the humerus, femur, and tibia. They are also for use in fixation of peri-prosthetic fractures, osteopenic bone and non-unions or malunions.
The DePuy Synthes Curved Condylar Plates are intended for buttressing multifragmentary distal femur fractures. including: supracondylar, intra-articular condylar fractures, peri-prosthetic fractures and fractures in normal or osteopenic bone, nonunions/malunions, and osteotomies of the femur.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ---------------------------------------------- | --------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{19}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes Low Profile Reconstruction Plates - MR Conditional
Indications for Use (Describe)
The DePuy Synthes Low Profile Reconstruction Plates are intended for pelvic and acetabular reconstruction and fixation of fractures of the distal humerus, clavicle, and scapula.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{20}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 3.5 mm Spring Plate - MR Conditional
Indications for Use (Describe)
The DePuy Synthes 3.5 mm Spring Plate is intended for pelvic and acetabular reconstructive surgery and fracture fixation of the distal fibula.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{21}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 3.5 mm LCP Hook Plate - MR Conditional
Indications for Use (Describe)
The DePuy Synthes 3.5 mm LCP Hook Plate is indicated for fractures and non-unions of small bones including the ulna, radius, tibia and fibula, particularly in osteopenic bone.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{22}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 3.5mm LCP Periarticular Proximal Humerus Plates - MR Conditional
Indications for Use (Describe)
The DePuy Synthes 3.5 mm LCP Periarticular Proximal Humerus Plates are indicated for fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{23}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 3.5 mm and 4.5 mm Locking Compression Plate (LCP) System with Expanded Indications - MR Conditional
Indications for Use (Describe) The DePuy Synthes 3.5 mm Locking Compression Plate (LCP) System:
The DePuy Synthes 3.5 mm LCP is indicated for fixation of fractures, osteotomies and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, fibularly in osteopenic bone for adult patients,
These plates are also indicated for fracture fixation of diaphyseal areas of long bones in pediatric patients.
The DePuy Synthes 4.5 mm Locking Compression Plate (LCP) System:
The DePuy Synthes 4.5 mm LCP is indicated for fixation of various long bones, such as the humerus, femur and tibia and for use in fixation of peri-prosthetic fractures, osteopenic bone and fixation of non-unions in adult patients.
These plates are also indicated for fracture fixation of diaphyseal areas of long bones in pediatric patients.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{24}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 3.5 mm Locking Attachment Plates - MR Conditional
Indications for Use (Describe)
The DePuy Synthes 3.5mm Locking Attachment Plate is intended for use with DePuy Synthes LCP plates to augment the stabilization of fractures, including periprosthetic fractures in the presence of intramedullary implants, in the femur, tibia and humerus, particularly in osteopenic bone.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | ☑ |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{25}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad Locking Compression Plates (LCP) - MR Conditional
Indications for Use (Describe)
The DePuy Synthes 3.5 mm Curved Narrow and Broad LCP Plates are intended for fixation of fractures and non-unions of clavicle, scapula, olecranon, humerus, radius, pelvis, distal tibia and fibularly in osteopenic bone for adult patients.
The DePuy Synthes 4.5 mm Curved Narrow and Broad LCP Plates are intended for fixation of various long bones, such as the humerus, femur, and tibia. They are also for use in fixation of periprosthetic fractures, osteopenic bone, and nonunions or malunions in adult patients.
The DePuy Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad LCP Plates are also indicated for fracture fixation of diaphyseal and metaphyseal areas of long bones in pediatric patients.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{26}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 2.7 mm / 3.5 mm VA-LCP Anterior Clavicle Plate System - MR Conditional
Indications for Use (Describe)
The DePuy Synthes 2.7 mm / 3.5 mm VA-LCP Anterior Clavicle Plate System is indicated for fixatures, malunions, non-unions, and osteotomies of the clavicle in adults, and in both adolescents (12-18 years) and transitional adolescents (18-21 years), in which the clavicular growth plates have fused or in which the growth plates will not be crossed by the plate system.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{27}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 3.5 mm LCP Clavicle Plate System - MR Conditional
Indications for Use (Describe)
The DePuy Synthes 3.5 mm LCP Clavicle Plate System is indicated for fixation of fractures, non-unions, and osteotomies of the clavicle in adults, and in both adolescents (12-18 years) and transitional adolescents (18-21 years), in which clavicular growth plates have fused or in which the growth plates will not be crossed by the plate system.
| Type of Use (Select one or both, as applicable) |
|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{28}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes Variable Angle LCP Elbow System - MR Conditional
Indications for Use (Describe)
The DePuy Synthes Variable Angle LCP Elbow System is intended for fixation of fractures of the distal humerus, olecranon and ulna in adults and adolescents (12-21) in which the growth plates have fused or in which the growth plates will not be crossed by the plate system. Specifically,
-
Distal humerus plates are indicated for intra-articular fractures, comminuted supracondylar fractures, malunions and non-unions of the distal humerus.
-
Olecranon and Proximal ulna plates are indicated for fractures, malunions and non-unions of the olecranon and proximal ulna.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{29}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes Variable Angle LCP Elbow System (Medial and Posterolateral Distal Humerus Plates) – MR Conditional
Indications for Use (Describe)
The DePuy Synthes Variable Angle LCP Elbow System is intended for fixation of fractures of the distal humerus, olecranon and ulna in adults and adolescents (12-2 1) in which the growth plates have fused. Specifically,
· Distal humerus plates are indicated for intra-articular fractures, comminuted supracondylar fractures. malunions and non-unions of the distal humerus.
· Olecranon and Proximal ulna plates are indicated for fractures, malunions and non-unions of the olecranon and proximal ulna.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{30}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes TOMOFIX Osteotomy System - MR Conditional
Indications for Use (Describe)
The DePuy Synthes TOMOFIX Osteotomy System is intended for osteotomies, treatment of bone and joint deformities, fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia.
Specifically,
-The TOMOFIX Medial Proximal Tibia Plates are indicated for open- and close-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the medial proximal tibia
-The TOMOFIX Lateral Proximal Tibia Plates are indicated for open- and closed-wedge osteotomies fixation of fractures. and malalignment caused by injury or disease, such as osteoarthritis, of the lateral proximal tibia
-The TOMOFIX Lateral Distal Femur Plates are indicated for open- and closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the lateral distal femur
-The TOMOFIX Medial Distal Femur Plates are indicated for closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the medial distal femur
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{31}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180310
Device Name
DePuy Synthes 2.4/2.7mm VA LCP Two-Column Volar Distal Radius Plate, Extra-Long - MR Conditional
Indications for Use (Describe)
The DePuy Synthes 2.4/2.7 mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long is intended for intra-and extra-articular fractures, osteotomies, nonunions of the distal radius, with or without extension into the radial diaphysis in adults and adolescents (12-21) where the growth plates have fused or will not be crossed, and the following adolescent distal radius fractures:
-
Intra-articular fractures exiting the epiphysis
-
Intra-articular fractures exiting the metaphysis
-
Physeal crush injuries
-
Any injuries which cause growth arrest to the distal radius
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{32}------------------------------------------------
Image /page/32/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a golden circular design on the left, followed by the company name "DePuy Synthes" in blue. To the right of the company name, the phrase "People inspired" is written in a cursive font. The text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is located beneath the company name in a smaller font size.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes Reconstructive Plates 'Y' Plates – MR Conditional
Classification Name(s): Appliance, Fixation, Nail/Blade/Plate Combination, Single Component
Regulatory Class: Class II, §888.3030
Product Code(s): KTW
1.3. Predicate Device
K792291 Synthes Reconstructive Plates 'Y' Plates
1.4. Device Description
The 3.5 mm Reconstruction Plate is available in multiple lengths. It has a smooth underside, is notched between the screw holes and can be adjusted (contoured) in three dimensions. The screw holes are designed to provide limited bidirectional compression of the fracture via the tapered entrance of the hole, and accept 3.5 mm cortex screws
The 4.5 mm Reconstruction Plate has a smooth underside, is notched between the screw holes and can be adjusted (contoured) in three dimensions. The oval holes are designed to provide limited bidirectional compression of the fracture via the tapered entrance of the hole which allows 25° longitudinal and 7° lateral angulation of screws. It is available in multiple lengths and accepts 4.5 mm cortex and 6.5 mm cancellous screws.
{33}------------------------------------------------
Image /page/33/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a symbol that looks like two intersecting circles, followed by the company name in blue, and the tagline "People inspired" in a light cursive font. Below the company name, there is a smaller text that reads "PART OF THE Johnson & Johnson FAMILY OF COMPANIES".
The 3.5 mm Curved Reconstruction Plate is available in multiple lengths. It has a smooth underside and is notched between screw holes and can be adjusted (contoured) in three dimensions. The screw holes are designed to provide limited bidirectional compression of the fracture via the tapered entrance of the hole and accept 3.5 mm cortex screws.
1.5. Intended Use
The DePuy Synthes Reconstructive Plates 'Y' Plates are indicated for pelvic and acetabular reconstructive surgery and the reduction of fractures in the distal humerus, the clavicle, and the scapula.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes Reconstructive Plates 'Y' Plates. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Reconstructive Plates 'Y' Plates in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{34}------------------------------------------------
Image /page/34/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol on the left, followed by the company name in blue text. To the right of the company name is the phrase "People inspired" in a light brown script font. Below the company name is the phrase "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" in smaller, gray text.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes Anatomical Locking Plate System – MR Conditional
Classification Name(s): Screw, Fixation, Bone
Regulatory Class: Class II, §888.3040
Product Code(s): HWC, HRS
1.3. Predicate Device
K961413 Synthes Anatomical Locking Plate System
1.4. Device Description
The Anatomical Locking Plate System utilizes a locking feature that secures the screw to the plate, enabling stable fixation to be achieved via unicortical fixation. The plates are available with and without an anatomically contoured head, in a variety of sizes. The head of the screw has a round cross section with a conical underside to fit into the plate. The underside has a tapered thread to match the design of the holes in the plate. The screws are available in thread diameters from 3 mm to 5 mm, in a variety of lengths.
1.5. Intended Use
The DePuy Synthes Anatomical Locking Plate System indicated to treat fractures (including but not limited to) the radius, ulna, humerus, tibia, fibula, femur.
{35}------------------------------------------------
Image /page/35/Picture/1 description: The image features the DePuy Synthes logo, which includes a golden emblem resembling interconnected circles. Next to it, the company name "DePuy Synthes" is written in a bold, dark blue font. Underneath the company name, there is a smaller text indicating that it is "PART OF THE Johnson & Johnson FAMILY OF COMPANIES", and to the right of the company name, the phrase "People inspired" is written in an elegant, cursive font.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes Anatomical Locking Plate System. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Anatomical Locking Plate System in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{36}------------------------------------------------
Image /page/36/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular graphic on the left, followed by the company name in blue text. To the right of the company name is the phrase "People inspired" in a light brown script font. Below the company name is the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" in a smaller font size.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes Cannulated Angle Blade Plate (CABP) System – MR Conditional
Classification Name(s): Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Regulatory Class: Class II, §888.3030
Product Code(s): KTT
1.3. Predicate Device
K974537 Synthes Cannulated Angle Blade Plate (CABP) System
1.4. Device Description
The CABP is a straight plate with a blade at the head to allow for better fixation in the head of the humerus or in the distal tibia. The blade of the plate is cannulated to fit over a guide wire, allowing for the adjustment of the wire placement several times without adversely affecting the final result. There are cuts in the undersurface of the plate the surface area of the plate in contact with bone. The area of contact between the plate and the bone is decreased in an effort to reduce damage to the cortical blood supply under the plat, and resultant reduction in damage-induced porosis and remodeling of the bone near the plate undersurface. Plate undercuts also make the bending properties of the plate more uniform, which facilitate contouring. The plate has round and dynamic compression screw
{37}------------------------------------------------
Image /page/37/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol on the left, followed by the company name in blue text. To the right of the company name is the phrase "People inspired" in a cursive font. Below the company name is the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" in a smaller font.
holes, accepts 4.5mm and 6.5 mm screws, is available in multiple blade lengths, and is manufactured from stainless steel and commercially pure titanium.
1.5. Intended Use
The DePuy Synthes CABP System is a plate and screw system intended to treat fractures of the proximal humerus and distal tibia.
Proximal humerus fractures include two-part greater tubercle fractures and fracture dislocations, two-part surgical neck fractures and fracture dislocations, three-part fractures or fracture dislocations, fractures in osteopenic bone, and nonunions and malunions.
Distal tibia fractures include acute fractures in osteopenic bone, and nonunions and malunions.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes Cannulated Angle Blade Plate (CABP) System. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Cannulated Angle Blade Plate (CABP) System in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{38}------------------------------------------------
Image /page/38/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol on the left, followed by the company name in blue text. To the right of the company name is the phrase "People inspired" in a light brown script font. Below the company name is the phrase "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" in smaller, gray text.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 3.5 mm 90° Cannulated Limited Contact-Angled Blade Plates – MR Conditional
Classification Name(s): Appliance, Fixation, Nail/Blade/Plate Combination, Single Component
Regulatory Class: Class II, §888.3030
Product Code(s): KTW
1.3. Predicate Device
K992837 Synthes 3.5 mm 90° Cannulated Limited Contact-Angled Blade Plates
1.4. Device Description
The DePuy Synthes 3.5 mm 90° Cannulated LC-ABP provides stable fracture fixation and rotational control for fractures of the proximal humerus and distal tibia. The plates feature a low profile limited contact dynamic compression plate (LC-DCP®) design and dynamic compression unit (DCU) screw holes. The blade portion of the plate is cannulated to accept a 2.0 mm guide wire. The plates are available in various sizes to accommodate varying patient anatomy.
{39}------------------------------------------------
Image /page/39/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a geometric symbol resembling interconnected circles in a gold color, followed by the company name "DePuy Synthes" in a bold, dark blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light brown font.
1.5. Intended Use
The DePuy Synthes 3.5 mm 90° Cannulated Limited Contact Angled Blade Plates are intended for the fixation of fractures and non-unions of the proximal humerus and distal tibia.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes 3.5 mm 90° Cannulated Limited Contact-Angled Blade Plates. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 3.5 mm 90° Cannulated Limited Contact-Angled Blade Plates in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{40}------------------------------------------------
Image /page/40/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a geometric symbol resembling interconnected circles in a light brown color, followed by the company name "DePuy Synthes" in a dark blue, sans-serif font. To the right of the company name, the phrase "People inspired" is written in a cursive, light brown font, adding a personal touch to the brand's image.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes Large Fragment Dynamic Compression Locking (DCL) System - MR Conditional
Classification Name(s): Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Regulatory Class: Class II, §888.3030
Product Code(s): KTT
1.3. Predicate Device
K000682 Synthes Large Fragment Dynamic Compression Locking (DCL) System
1.4. Device Description
The Large Fragment DCL system consists of limited-contact profile plates in broad and narrow sizes, which include combination dynamic compression/locking screw holes. The plates accept 4.5 mm cortex, 6.5 mm cancellous, 4.5 mm cannulated, 7.0 mm cannulated, and 5.0 mm locking screws. This device is manufactured in either stainless steel or titanium.
{41}------------------------------------------------
Image /page/41/Picture/1 description: The image shows the logo for DePuy Synthes, which is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol with overlapping shapes, followed by the text "DePuy Synthes" in blue. To the right of the company name, the words "People inspired" are written in a cursive font.
1.5. Intended Use
The DePuy Synthes Large Fragment DCL is intended for fixation of various long bones, such as the humerus, femur and tibia. Also for use in fixation of osteopenic bone and fixation of non-unions or malunions.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes Large Fragment Dynamic Compression Locking (DCL) System. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Large Fragment Dynamic Compression Locking (DCL) System in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{42}------------------------------------------------
Image /page/42/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic element on the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is present in a smaller font size.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes Large Fragment Locking Compression Plate (LCP) System -T Plate - MR Conditional
Classification Name(s): Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Regulatory Class: Class II, §888.3030
Product Code(s): KTT
1.3. Predicate Device
K010766 Synthes (USA) Large Fragment Locking Compression Plate (LCP) System - T Plate
1.4. Device Description
The DePuy Synthes Locking Compression Plate (LCP) System - T Plate is a buttress plate and screw system. The primary feature of the plate is round holes combined with locking screw holes. The locking screws form a locked, fixed angle construct with the plate, while the standard screws facilitate reduction and create compression between the plate and bone.
The plates accept 4.5 mm cortex, 6.5 mm cancellous, 4.5 mm cannulated, 7.0 mm cannulated, 7.3 mm cannulated, 4.0 mm and 5.0 mm locking screws
{43}------------------------------------------------
Image /page/43/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a gold-colored symbol that looks like three overlapping circles, followed by the company name in blue, and the tagline "People inspired" in a light brown script font. Below the company name, in smaller font, it says "PART OF THE Johnson & Johnson FAMILY OF COMPANIES".
1.5. Intended Use
The DePuy Synthes Large Fragment Locking Compression Plate (LCP) System - T Plate is intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. Also for use in fixation of osteopenic bone and fixation of non-unions and malunions.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes (USA) Large Fragment Locking Compression Plate (LCP) System - T Plate. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes (USA) Large Fragment Locking Compression Plate (LCP) System - T Plate in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{44}------------------------------------------------
Image /page/44/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a symbol that looks like two overlapping circles, followed by the company name in blue text. To the right of the company name, the words "People inspired" are written in a cursive font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes Curved Reconstruction Plate – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K011334 Synthes Curved Reconstruction Plate
1.4. Device Description
The DePuy Synthes Curved Reconstruction Plate line extension is a pre-curved version of the currently marketed Straight Locking Reconstruction Plate. Both of these plates will be included as part of the DePuy Synthes Small Fragment DCL System. Both the curved and straight plates have the same intended use and there is no change in safety or efficacy.
1.5. Intended Use
The DePuy Synthes Curved Reconstruction Plate is intended for fixation of fractures, osteotomies, and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, and fibula, particularly in osteopenic bone as a part of the DePuy Synthes Small Fragment DCL System.
{45}------------------------------------------------
Image /page/45/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol to the left of the company name, which is written in blue. To the right of the company name, the phrase "People inspired" is written in a cursive font.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes Curved Reconstruction Plate. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Curved Reconstruction Plate in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{46}------------------------------------------------
Image /page/46/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic element on the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is present in a smaller font size.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes One-Third Tubular Dynamic Compression Locking (DCL) Plate – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K011335 Synthes One-third DCL Plate
1.4. Device Description
The DePuy Synthes One-Third Tubular DCL Plate line extension is a threaded version of the currently marketed Synthes Third Tubular Plate. The threaded plates will accept locking screws and therefore can be included as part of the Small Fragment DCL System. The new plates have the same intended use as other plates in the system and there is no change in safety or efficacy.
{47}------------------------------------------------
Image /page/47/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular graphic with overlapping shapes, followed by the company name "DePuy Synthes" in blue, and the tagline "People inspired" in a cursive font. Below the company name, there is a smaller text indicating that DePuy Synthes is part of the Johnson & Johnson family of companies.
1.5. Intended Use
The DePuy Synthes One-Third Tubular Plate is intended for fixation of fractures, osteotomies, and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, and fibula, particularly in osteopenic bone as a part of the DePuy Synthes Small Fragment DCL System.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes One-third DCL Plate. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Onethird DCL Plate in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{48}------------------------------------------------
Image /page/48/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic to the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes LCP Proximal Humerus Plate – MR Conditional
Classification Name(s): Appliance, Fixation, Nail/Blade/Plate Combination, Single Component
Regulatory Class: Class II, §888.3030
Product Code(s): KTW
1.3. Predicate Device
K011815 Synthes LCP Proximal Humerus Plate
1.4. Device Description
The DePuy Synthes LCP Proximal Humerus Plates are designed to match the anatomy of the proximal humerus. These plates can be applied to either the right or left humerus. The proximal portion of each plate has threaded holes that accept 3.5 mm or 2.7 mm screws. The distal portion of the plate has combination holes that allow the option of using 3.5 mm locking or cortex screws, or 4.0 mm cancellous screws to accomplish plate fixation. These plates will be offered as an addition to the DePuy Synthes Small Fragment LCP (formerly DCL) System.
{49}------------------------------------------------
Image /page/49/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular graphic to the left of the company name, which is written in blue. To the right of the company name is the phrase "People inspired" in a cursive font. Below the company name is the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES".
1.5. Intended Use
The DePuy Synthes LCP Proximal Humerus Plate is indicated for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes LCP Proximal Humerus Plate. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes LCP Proximal Humerus Plate in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{50}------------------------------------------------
Image /page/50/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic element on the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is present in a smaller font size.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes Pilon Plate – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K020602 Synthes Pilon Plate
1.4. Device Description
The DePuy Synthes Pilon Plate is a metal plate that offers screw to plate locking designed for various fracture modes of the distal end of the tibia.
1.5. Intended Use
The DePuy Synthes Pilon Plate is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal tibia.
{51}------------------------------------------------
Image /page/51/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo includes a symbol that looks like two overlapping circles, the company name in blue, and the tagline "People inspired" in a light brown script font. The text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font below the company name.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes Pilon Plate. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Pilon Plate in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{52}------------------------------------------------
Image /page/52/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic element on the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is present in a smaller font size.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes TomoFix™ Osteotomy System – MR Conditional
Classification Name(s): Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Regulatory Class: Class II, §888.3030
Product Code(s): KTT
1.3. Predicate Device
K023941 Synthes TomoFixTM Osteotomy System
1.4. Device Description
The TomoFix™ Osteotomy System consists of five different titanium plates with locking and combination holes. There are two plates (left and right) for the lateral distal femur, 2 plates (left and right) for the lateral proximal tibia, and 1 plate for the medial proximal tibia.
1.5. Intended Use
The DePuy Synthes TomoFix™ Osteotomy System is intended for open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia and lateral distal femur, treatment of bone and joint deformities, and malalignment caused by injury or disease such as osteoarthritis.
{53}------------------------------------------------
Image /page/53/Picture/1 description: The image features the logo of DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo includes a circular graphic element on the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-brown font.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes TomoFix™ Osteotomy System. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes TomoFix™ Osteotomy System in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{54}------------------------------------------------
Image /page/54/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic to the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 3.5 mm Broad LCP & 4.5 mm LCP Distal Humerus Plate – MR Conditional
Classification Name(s): Plate, Fixation, Bone.
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K031178 3.5 mm Broad LCP & 4.5 mm LCP Distal Humerus Plate
1.4. Device Description
The DePuy Synthes 3.5 mm Broad LCP & 4.5 mm Broad LCP Distal Humerus Plates are contoured to match the anatomy of the distal humerus with a limited contact low profile design. The plate has dynamic compression holes combined with locking holes which accept 3.5 & 4.5 mm cortex, 3.5 & 4.5 mm self-tapping cortex, 3.5 mm shaft, 3.5,4.0, & 5.0 mm locking, and 4.0 mm cancellous screws. The plates are available in various lengths.
1.5. Intended Use
The DePuy Synthes 3.5 mm Broad LCP & 4.5 mm LCP Distal Humerus Plates are indicated for fractures of the distal humerus.
{55}------------------------------------------------
Image /page/55/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol to the left of the company name, which is written in blue. To the right of the company name, the phrase "People inspired" is written in a cursive font.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes 3.5 mm Broad LCP & 4.5 mm LCP Distal Humerus Plate. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 3.5 mm Broad LCP & 4.5 mm LCP Distal Humerus Plate in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{56}------------------------------------------------
Image /page/56/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a stylized symbol to the left of the company name, "DePuy Synthes", which is written in a dark blue font. To the right of the company name is the phrase "People inspired" in a cursive font. Below the company name is the phrase "PART OF THE johnson-johnson FAMILY OF COMPANIES" in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes Clavicle Hook Plate – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K031677 Synthes (USA) Clavicle Hook Plate
1.4. Device Description
The DePuy Synthes Clavicle Hook Plate is an anatomically designed L-shaped plate featuring a distal joint bridging hook with a rounded end. The shaft of the plate contains either dynamic compression screw holes or LCP plate holes. The plates accept 3.5 mm cortex, 4.0 mm cancellous and 3.5 mm locking screws (LCP Clavicle Hook Plate only).
1.5. Intended Use
The DePuy Synthes Clavicle Hook Plate is intended for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.
{57}------------------------------------------------
Image /page/57/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a symbol that looks like two overlapping circles, followed by the words "DePuy Synthes" in blue. To the right of the company name, the words "People inspired" are written in a cursive font.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes Clavicle Hook Plate. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Clavicle Hook Plate in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{58}------------------------------------------------
Image /page/58/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol on the left, followed by the company name in blue text. To the right of the company name is the phrase "People inspired" in a light brown script font. Below the company name is the phrase "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" in smaller, gray text.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 3.5 / 4.5 mm LCP® Metaphyseal Plates – MR Conditional
Classification Name(s): Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite
Regulatory Class: Class II, §888.3030
Product Code(s): LXT
1.3. Predicate Device
K033805 Synthes (USA) 3.5 / 4.5 mm LCP® Metaphyseal Plates
1.4. Device Description
The DePuy Synthes 3.5 / 4.5 mm LCP® Metaphyseal Plates are contourable to match the anatomy, have a limited contact design, and are tapered at the plate head. The 3.5 / 4.5 mm metaphyseal plates feature combination dynamic compression / locking screws holes (combi-holes). The plates in this system accept 3.5 / 4.5 mm cortex, 3.5, 4.0, & 5.0 mm locking, and 4.0 mm or 6.5 mm cancellous screws. The plate also has a 2.0 mm hole for preliminary fixation with k-wires.
{59}------------------------------------------------
Image /page/59/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol with intersecting shapes, followed by the company name "DePuy Synthes" in blue text. Below the company name, there is a smaller text that reads "PART OF THE Johnson & Johnson FAMILY OF COMPANIES", and to the right of the company name, there is the phrase "People inspired" in a cursive font.
1.5. Intended Use
The DePuy Synthes 3.5 / 4.5mm LCP® Metaphyseal Plate is intended for the fixation of various long bones, excluding the femur. It is also for use in fixation of osteopenic bone and fixation of nonunions or malunions.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes 3.5 / 4.5 mm LCP® Metaphyseal Plates. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 3.5 / 4.5 mm LCP® Metaphyseal Plates in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{60}------------------------------------------------
Image /page/60/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic to the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 3.5 mm LCP Distal Humerus System – MR Conditional
Classification Name(s): Single/Multiple component metallic bone fixation appliances and accessories. And Smooth/threaded metallic bone fixation fastener
Regulatory Class: Class II, §888.3030 and §888.3040
Product Code(s): KTT
1.3. Predicate Device
K033995 3.5 mm LCP Distal Humerus System
1.4. Device Description
The DePuy Synthes 3.5 mm LCP Distal Humerus System consists of medial and posterolateral distal humerus plates of various lengths and 2.7 mm locking screws. The plates are pre-contoured to match the anatomy of the distal humerus with a limited contact low profile design. The plate features locking compression holes and conical locking holes which accept 2.4, 3.5, & 4.0 mm cortex screws, 2.4, 2.7 & 3.5 mm locking screws, and 4.0 mm cancellous screws. The System will be available in Stainless Steel and Titanium.
{61}------------------------------------------------
Image /page/61/Picture/1 description: The image shows the logo for DePuy Synthes, which is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol with intersecting shapes, followed by the text "DePuy Synthes" in a bold, blue font. To the right of the company name, the words "People inspired" are written in a cursive font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font.
1.5. Intended Use
The DePuy Synthes 3.5 mm LCP Distal Humerus System is indicated for intraarticular fractures of the distal humerus, commninuted supra-condylar fractures, osteotomies, and non-unions of the distal humerus.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes 3.5 mm LCP Distal Humerus System. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 3.5 mm LCP Distal Humerus System in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{62}------------------------------------------------
Image /page/62/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic element on the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is present in a smaller font size.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes LCP® Proximal Humerus Plates, Long – MR Conditional
Classification Name(s): Single/Multiple component metallic bone fixation appliances and accessories.
Regulatory Class: Class II, §888.3030
Product Code(s): HRS, HWC
1.3. Predicate Device
K041860 Synthes (USA) LCP® Proximal Humerus Plates, Long
1.4. Device Description
The DePuy Synthes LCP Proximal Humerus Plates, long are pre-contoured to match the anatomy of the proximal humerus with a limited contact low profile design. The plate features locking holes and combination locking and compression holes which accept 2.7, 3.5, & 4.0 mm cortex screws, 3.5 mm locking screws, and 4.0 mm cancellous screws. The System will be available in Stainless Steel and Titanium. These plates will be offered as an addition to the DePuy Synthes Small Fragment LCP (formerly DCL) System.
{63}------------------------------------------------
Image /page/63/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular graphic to the left of the company name, which is written in blue. To the right of the company name is the phrase "People inspired" in a cursive font.
1.5. Intended Use
The DePuy Synthes LCP Proximal Humerus Plate, Long is indicated for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes LCP® Proximal Humerus Plates, Long. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes LCP® Proximal Humerus Plates, Long in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{64}------------------------------------------------
Image /page/64/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic element on the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is present in a smaller font size.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes LCP® Curved Plates – MR Conditional
Classification Name(s): Single/Multiple component metallic bone fixation appliances and accessories.
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K041911 Synthes LCP® Curved Plates
1.4. Device Description
The DePuy Synthes LCP® Curved Plates have a slight curve to better match the anatomy of the bone. The plates have a limited contact profile design and includes combination dynamic compression/locking screw holes.
{65}------------------------------------------------
Image /page/65/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a symbol of interconnected circles in a gold color, followed by the company name "DePuy Synthes" in a bold, blue font. Underneath the company name, there is a tagline "People inspired" in a cursive, light-colored font, and below the company name is the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES".
1.5. Intended Use
The DePuy Synthes Curved Broad Plates are intended for fixation of various long bones, such as the humerus, femur, and tibia. They are also for use in fixation of peri-prosthetic fractures, osteopenic bone and non-unions or malunions.
The DePuy Synthes Curved Condylar Plates are intended for buttressing multifragmentary distal femur fractures, including: supracondylar, intra-articular and extra-articular condylar fractures, peri-prosthetic fractures and fractures in normal or osteopenic bone, nonunions/malunions, and osteotomies of the femur.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes LCP® Curved Plates. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes LCP® Curved Plates in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{66}------------------------------------------------
Image /page/66/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol, the words "DePuy Synthes" in blue, and the words "People inspired" in a light brown script font. Below the company name, it states "PART OF THE Johnson-Johnson FAMILY OF COMPANIES".
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes Low Profile Reconstruction Plates – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K042377 Synthes Low Profile Reconstruction Plates
1.4. Device Description
The DePuy Synthes 3.5 mm Low Profile Reconstruction Plates consist of straight, curved, and J-shaped plates. The shape and profile of these plates are designed to minimize soft tissue irritation and minimize the need for intraoperative contouring. However, when intraoperative contouring is required for a precise fit with the bone, the plate design allows for simple and uniform bending.
1.5. Intended Use
The DePuy Synthes Low Profile Reconstruction Plates are intended for pelvic and acetabular reconstruction and fixation of fractures of the distal humerus, clavicle, and scapula.
{67}------------------------------------------------
Image /page/67/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol to the left of the company name, which is written in blue. To the right of the company name, the phrase "People inspired" is written in a cursive font.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes Low Profile Reconstruction Plates. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Low Profile Reconstruction Plates in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{68}------------------------------------------------
Image /page/68/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular graphic to the left of the company name, which is written in a dark blue font. To the right of the company name is the phrase "People inspired" in a cursive font. Below the company name is the phrase "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 3.5 mm Spring Plate – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K061973 Synthes 3.5 mm Spring Plate
1.4. Device Description
The DePuy Synthes 3.5mm Spring Plate is a variation of the DePuy Synthes one-third tubular plate with collar which utilizes two sharp spikes at the bottom surface and a pre-bent convex shape to aid in the reduction of small bone fragments while conforming to bony uneven surfaces. The plate incorporates a 1 - 10 hole design in lengths ranging from 19.5mm - 132mm and accepts either 3.5mm cortex or pelvic screws. In addition the plates are manufactured from Stainless Steel and Commercially Pure Titanium and provided STERILE and NON STERILE.
1.5. Intended Use
The DePuy Synthes 3.5 mm Spring Plate is intended for pelvic and acetabular reconstructive surgery and fracture fixation of the distal fibula.
{69}------------------------------------------------
Image /page/69/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo includes a symbol of interlocking circles in gold, followed by the company name in blue, and the tagline "People inspired" in a light tan script font. Below the company name, in smaller letters, it states "PART OF THE Johnson-Johnson FAMILY OF COMPANIES."
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes 3.5mm Spring Plate. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 3.5mm Spring Plate in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{70}------------------------------------------------
Image /page/70/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic to the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 3.5 mm LCP Hook Plate – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K082072 Synthes (USA) 3.5mm LCP Hook Plate
1.4. Device Description
The DePuy Synthes 3.5 mm LCP Hook Plate is a low-profile, precontoured bone fixation plate intended for the treatment of fractures, osteotomies, and non-unions in small bones. The plates feature a preformed dual hook on one end and are available in stainless steel and titanium.
1.5. Intended Use
The DePuy Synthes 3.5 mm LCP Hook Plate is indicated for fractures, osteotomies and nonunions of small bones including the ulna, radius, tibia and fibula, particularly in osteopenic bone.
{71}------------------------------------------------
Image /page/71/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a stylized symbol resembling interconnected circles in a gold color, followed by the company name "DePuy Synthes" in a dark blue, sans-serif font. To the right of the company name, the phrase "People inspired" is written in a cursive, light gold font, adding a personal touch to the brand identity.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes 3.5 mm LCP Hook Plate. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 3.5 mm LCP Hook Plate in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{72}------------------------------------------------
Image /page/72/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol, the words "DePuy Synthes" in blue, and the words "People inspired" in a light brown script font. Below the company name, it states "PART OF THE Johnson-Johnson FAMILY OF COMPANIES".
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 3.5 mm LCP Periarticular Proximal Humerus Plates – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K082625 Synthes (USA) 3.5 mm LCP Periarticular Proximal Humerus Plates
1.4. Device Description
The DePuy Synthes 3.5 mm LCP Periarticular Proximal Humerus Plates are low-profile, precontoured bone fixation plates intended for the treatment of fractures of the humerus. The plates are offered sterile and non-sterile and are available in stainless steel and titanium.
1.5. Intended Use
The DePuy Synthes 3.5 mm LCP Periarticular Proximal Humerus Plates are indicated for fractures, fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
{73}------------------------------------------------
Image /page/73/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol, the company name in blue, and the tagline "People inspired" in a light brown script font. The text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is located below the company name in a smaller font.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes 3.5 mm LCP Periarticular Proximal Humerus Plates. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 3.5 mm LCP Periarticular Proximal Humerus Plates in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{74}------------------------------------------------
Image /page/74/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic to the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 3.5 mm and 4.5 mm Locking Compression Plate (LCP) System with Expanded Indications - MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS, HWC
1.3. Predicate Device
K082807 Synthes 3.5 mm and 4.5 mm Locking Compression Plate (LCP) System with Expanded Indications
1.4. Device Description
The DePuy Synthes 3.5 mm and 4.5 mm LCP Plates with Expanded Indications consist on.5 mm LCP plates, 4.5 rom Narrow LCP Plates, 4.5 mm Broad LCP plates and 4.5 mm Curved Broad LCP Plates for fracture fixation in adults and pediatric patients. These plates accept locking, cortex and cancellous screws.
{75}------------------------------------------------
Image /page/75/Picture/1 description: The image shows the logo for DePuy Synthes, which is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol to the left, followed by the text "DePuy Synthes" in blue. To the right of the company name, the words "People inspired" are written in a cursive font.
1.5. Intended Use
The DePuy Synthes 3.5 mm Locking Compression Plate (LCP) System:
The DePuy Synthes 3.5 mm LCP is indicated for fixation of fractures, osteotomies and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, fibula, particularly in osteopenic bone for adult patients.
These plates are also indicated for fracture fixation of diaphyseal and metaphyseal areas of long bones in pediatric patients.
The DePuy Synthes 4.5 mm Locking Compression Plate (LCP) System:
The DePuy Synthes 4.5 mm LCP is indicated for fixation of various long bones, such as the humerus, femur and tibia and for use in fixation of peri-prosthetic fractures, osteopenic bone and fixation of non-unions or malunions in adult patients.
These plates are also indicated for fracture fixation of diaphyseal and metaphyseal areas of long bones in pediatric patients.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes 3.5 mm and 4.5 mm Locking Compression Plate (LCP) System with Expanded Indications. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 3.5 mm and 4.5 mm Locking Compression Plate (LCP) System with Expanded Indications in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{76}------------------------------------------------
Image /page/76/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic element on the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is present in a smaller font size.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 3.5mm Locking Attachment Plates – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K083573 Synthes (USA) 3.5mm Locking Attachment Plates
1.4. Device Description
The DePuy Synthes 3.5mm Locking Attachment Plates are low-profile, pre-contoured plates which are intended to be used with existing Synthes to stabilize fractures. The plates are available in stainless steel and titanium.
1.5. Indications for Use
The DePuy Synthes 3.5mm Locking Attachment Plate is intended for use with DePuy Synthes LCP plates to augment the stabilization of fractures, including periprosthetic fractures and fractures in the presence of intramedullary implants, in the femur, tibia and humerus, particularly in osteopenic bone.
{77}------------------------------------------------
Image /page/77/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a symbol that looks like two overlapping circles, followed by the words "DePuy Synthes" in blue. To the right of the company name, the words "People inspired" are written in a cursive font.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes 3.5mm Locking Attachment Plates. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 3.5mm Locking Attachment Plates in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{78}------------------------------------------------
Image /page/78/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic to the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad Locking Compression Plates (LCP) – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS, HWC
1.3. Predicate Device
K092609 Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad Locking Compression Plates (LCP)
1.4. Device Description
The DePuy Synthes Curved Narrow and Broad LCP Plates are available in stainless steel and titanium, and consist of limited-contact profile plates in 3.5 mm and 4.5mm narrow and broad sizes. The plates feature Dynamic Compression Plate (DCP) holes combined with locking screw holes. The 3.5mm plates accept 3.5mm cortex and locking screws and 4.0 mm cancellous screws, and the 4.5mm plates accept 4.5 mm cortex screws, 4.0mm and 5.0mm locking screws, 4.5 mm cannulated screws, 5.0 mm periprosthetic screws, and 6.5 mm cancellous screws.
{79}------------------------------------------------
Image /page/79/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo includes a symbol of two overlapping circles, the company name in blue, and the tagline "People inspired" in a light brown script font. The text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font below the company name.
1.5. Indications for Us
The DePuy Synthes 3.5 mm Curved Narrow and Broad LCP Plates are intended for fixation of fractures, osteotomies and non-unions of clavicle, scapula, olecranon, humerus, radius, pelvis, distal tibia and fibula, particularly in osteopenic bone for adult patients.
The DePuy Synthes 4.5 mm Curved Narrow and Broad LCP Plates are intended for fixation of various long bones, such as the humerus, femur, and tibia. They are also for use in fixation of periprosthetic fractures, osteopenic bone, and non-unions or malunions in adult patients.
The DePuy Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad LCP Plates are also indicated for fracture fixation of diaphyseal and metaphyseal areas of long bones in pediatric patients.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad Locking Compression Plates (LCP). The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad Locking Compression Plates (LCP) in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{80}------------------------------------------------
Image /page/80/Picture/1 description: The image shows the logo for DePuy Synthes, which is part of the Johnson & Johnson family of companies. The logo consists of a circular graphic on the left, followed by the text "DePuy Synthes" in a dark blue font. To the right of the company name, the words "People inspired" are written in a cursive font. Below the company name, the text "PART OF THE Johnson & Johnson FAMILY OF COMPANIES" is written in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 2.7 mm / 3.5 mm VA-LCP Anterior Clavicle Plate System – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS, HWC
1.3. Predicate Device
K101536 2.7 mm / 3.5 mm VA-LCP Anterior Clavicle Plate System
1.4. Device Description
The DePuy Synthes 2.7 mm / 3.5 mm VA-LCP Anterior Clavicle Plate System consists of plates of various lengths and variable angle screws that provide the flexibility to lock screws in trajectories that can diverge from the central axis of the plate hole.
The system features medial and lateral plates that are pre-contoured to match the anatomy of the clavicle and a limited contact, low profile design. The plate with lateral extension features Variable Angle (VA) Locking holes and Dynamic Compression Plate (DCP) holes. The medial plate features Dynamic Compression Plate (DCP) holes combined with VA locking screw holes.
{81}------------------------------------------------
Image /page/81/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo includes a symbol of two overlapping circles, the company name in blue, and the tagline "People inspired" in a light brown script font. The text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font below the company name.
The DePuy Synthes 2.7 mm / 3.5 mm VA-LCP Anterior Clavicle Plate System is indicated for fixation of fractures, malunions, non-unions, and osteotomies of the clavicle in adults, and in both adolescents (12-18 years) and transitional adolescents (18-21 years), in which the clavicular growth plates have fused or in which the growth plates will not be crossed by the plate system.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes 2.7 mm / 3.5 mm VA-LCP Anterior Clavicle Plate System. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 2.7 mm / 3.5 mm VA-LCP Anterior Clavicle Plate System in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{82}------------------------------------------------
Image /page/82/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic to the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 3.5 mm LCP Clavicle Plate System – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K111540 Synthes 3.5 mm LCP Clavicle Plate System
1.4. Device Description
The DePuy Synthes 3.5 mm LCP Clavicle Plate System consists of metallic plates and screws that offer screw to plate locking designed for various fracture modes of the clavicle.
1.5. Indications for Use
The DePuy Synthes 3.5 mm LCP Clavicle Plate System is indicated for fixation of fractures, malunions, non-unions, and osteotomies of the clavicle in adults, and in both adolescents (12-18 years) and transitional adolescents (18-21 years), in which the clavicular growth plates have fused or in which the growth plates will not be crossed by the plate system.
{83}------------------------------------------------
Image /page/83/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular symbol, the company name in blue, and the phrase "People inspired" in a light brown script font. The text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is located below the company name in a smaller font.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes 3.5 mm LCP Clavicle Plate System. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 3.5 mm LCP Clavicle Plate System in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{84}------------------------------------------------
Image /page/84/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular graphic to the left of the company name, which is written in a dark blue font. To the right of the company name is the phrase "People inspired" in a cursive font. Below the company name is the phrase "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes Variable Angle LCP Elbow System – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS, HWC
1.3. Predicate Device
K120070 Synthes Variable Angle LCP Elbow System
1.4. Device Description
The DePuy Synthes Variable Angle LCP Elbow System contains plates intended to treat fractures of the distal humerus and proximal ulna. A variety of plate configurations are included in the system to allow for fixation of multiple fracture patterns. Specifically, the system includes several plate configurations for fixation of the distal humerus which are intended to be used in a two-plate construct where plates are positioned medially and laterally. Additionally, the system includes plates for fixation of the olecranon and proximal ulna. In its entirety, the following plate types are included in the system:
- -Medial Distal Humerus Plate
- -Lateral Distal Humerus Plate
- -Olecranon Plate
- -Proximal Olecranon Plate
{85}------------------------------------------------
Image /page/85/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a stylized graphic element resembling interconnected circles in a gold color, followed by the company name "DePuy Synthes" in a dark blue, sans-serif font. To the right of the company name, the phrase "People inspired" is written in a cursive, light beige font, adding a softer touch to the overall design.
Extra-articular Proximal Ulna Plate
The system accepts existing conical and locking screws as well as new metaphyseal screws, and allows for both dynamic compression and locking through Combi holes. The plates are universally designed for both left and right use and will be offered in both stainless steel and titanium.
1.5. Indications for Use
The DePuy Synthes Variable Angle LCP Elbow System is intended for fixation of fractures of the distal humerus, olecranon and ulna in adults and adolescents (12-21) in which the growth plates have fused or in which the growth plates will not be crossed by the plate system. Specifically,
- -Distal humerus plates are indicated for intra-articular fractures, commninuted supracondylar fractures, osteotomies, malunions and non-unions of the distal humerus.
- -Olecranon and Proximal ulna plates are indicated for fractures, osteotomies, malunions and non-unions of the olecranon and proximal ulna.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes Variable Angle LCP Elbow System. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Variable Angle LCP Elbow System in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{86}------------------------------------------------
Image /page/86/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic to the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font.
Date Prepared: October 31, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes Variable Angle LCP Elbow System (Medial and Posterolateral Distal Humerus Plates) – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS, HWC
1.3. Predicate Device
K120717 Synthes Variable Angle LCP Elbow System (Medial and Posterolateral Distal Humerus Plates)
1.4. Device Description
The DePuy Synthes Variable Angle LCP Elbow System contains posterolateral and medial plates intended to treat fractures of the distal humerus. The plates are used together in a two-plate, 90° construct and accept existing screws. New 2.7mm Metaphyseal Screws are also compatible with the System.
{87}------------------------------------------------
Image /page/87/Picture/1 description: The image shows the logo for DePuy Synthes, which is part of the Johnson & Johnson family of companies. The logo consists of a gold-colored symbol on the left, followed by the company name in blue. To the right of the company name, the words "People inspired" are written in a cursive font.
The DePuy Synthes Variable Angle LCP Elbow System is intended for fixation of fractures of the distal humerus, olecranon and ulna in adults and adolescents (12-21) in which the growth plates have fused or in which the growth plates will not be crossed by the plate system. Specifically,
- -Distal humerus plates are indicated for intra-articular fractures, comminuted supracondylar fractures, osteotomies, malunions and non-unions of the distal humerus.
- Olecranon and Proximal ulna plates are indicated for fractures, osteotomies, malunions and non-unions of the olecranon and proximal ulna.
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes Variable Angle LCP Elbow System (Medial and Posterolateral Distal Humerus Plates). The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Variable Angle LCP Elbow System (Medial and Posterolateral Distal Humerus Plates) in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{88}------------------------------------------------
Image /page/88/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular graphic to the left of the company name, which is written in a dark blue font. To the right of the company name is the phrase "People inspired" in a cursive font. Below the company name is the phrase "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes TOMOFIX Osteotomy System – MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K141796 DePuy Synthes TOMOFIX Osteotomy System
1.4. Device Description
The DePuy Synthes TOMOFIX Osteotomy System consists of five different titanium plate families with locking and combination locking/compression holes. This system features plates designed to provide stable fixation of osteotomies of the distal femur and proximal tibia.
The DePuy Synthes TOMOFIX Medial Distal Femur Plates are part of the DePuy Synthes TOMOFIX Osteotomy System used for closed wedge femoral osteotomies. The subject plates are anatomically contoured to fit the medial distal femur, are available in right and left versions, as well as sterile and non-sterile.
{89}------------------------------------------------
Image /page/89/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic element on the left, followed by the company name "DePuy Synthes" in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font, adding a personal touch to the brand identity.
The DePuy Synthes TOMOFIX Osteotomy System is intended for osteotomies, treatment of bone and joint deformities, fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia.
Specifically,
- The TOMOFIX Medial Proximal Tibia Plates are indicated for open- and closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the medial proximal tibia
- -The TOMOFIX Lateral Proximal Tibia Plates are indicated for open- and closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the lateral proximal tibia
- -The TOMOFIX Lateral Distal Femur Plates are indicated for open- and closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the lateral distal femur
- -The TOMOFIX Medial Distal Femur Plates are indicated for closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the medial distal femur
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes TOMOFIX Osteotomy System. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes TOMOFIX Osteotomy System in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{90}------------------------------------------------
Image /page/90/Picture/1 description: The image displays the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a circular graphic to the left, followed by the company name in a bold, blue font. To the right of the company name, the phrase "People inspired" is written in a cursive, light-colored font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font.
Date Prepared: October 29, 2018
1.1 Submitter
Primary Contact:
Jeffrey Krawiec Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5434 Jkrawiec@its.jnj.com
Alternate Contact:
Thomas Shea Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5679 tshea@its.jnj.com
1.2. Device
Name of Device: DePuy Synthes 2.4/2.7mm VA LCP Two-Column Volar Distal Radius Plate, Extra-Long - MR Conditional
Classification Name(s): Plate, Fixation, Bone
Regulatory Class: Class II, §888.3030
Product Code(s): HRS
1.3. Predicate Device
K163046 DePuy Synthes 2.4/2.7mm VA LCP Two-Column Volar Distal Radius Plate, Extra-Long
1.4. Device Description
The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long will be available in stainless steel and titanium. The head of the plate will remain the same as the existing 2.4mm VA-LCP Two-Column Volar Distal Radius Plate (K102694) and will therefore allow the use of the existing guide block for these plates. The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long will come in 3 different lengths (7 shaft holes, 10 shaft holes and 13 shaft holes). To accommodate different patient anatomy, the plate with 7 shaft holes will be available with three different plate head widths (Standard, Narrow and Wide). The 10 shaft hole and the 13 shaft hole plate will only be released with the Standard plate head width and additionally they feature
{91}------------------------------------------------
Image /page/91/Picture/1 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a symbol that looks like two overlapping circles, followed by the company name in blue text. To the right of the company name, the words "People inspired" are written in a cursive font. Below the company name, the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is written in a smaller font.
a curvature in the shaft. All plates will feature left- and right-specific designs and will be offered in sterile packed only.
The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plates, Extra-Long are designed to accept existing 2.4mm Variable Angle Locking Screws, 2.4mm Cortex Screws, 2.4mm Locking Screws, 1.8mm VA Buttress Pins and 1.8mm Locking Buttress Pins in the plate head holes. In the plate shaft, existing 2.4mm Cortex Screws, 2.7mm Cortex Screws, 2.4mm Variable Angle Locking Screws, 2.7mm Variable Angle Locking Screws, 2.4mm Locking Screws, and 2.7mm Locking Screws can be inserted.
1.5. Intended Use
The DePuy Synthes 2.4/2.7 mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long is intended for intra- and extra-articular fractures, osteotomies, nonunions and malunions of the distal radius, with or without extension into the radial diaphysis in adults and adolescents (12-21) where the growth plates have fused or will not be crossed, and the following adolescent distal radius fractures:
- Intra-articular fractures exiting the epiphysis
- Intra-articular fractures exiting the metaphysis
- Physeal crush injuries
- Any injuries which cause growth arrest to the distal radius
1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for DePuy Synthes 2.4/2.7mm VA LCP Two-Column Volar Distal Radius Plate, Extra-Long. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 2.4/2.7mm VA LCP Two-Column Volar Distal Radius Plate, Extra-Long in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
N/A