K Number
K173942

Validate with FDA (Live)

Date Cleared
2018-03-05

(69 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs using ASTM D6978-05 and will be labeled with a statement of compliance and a summary of the testing results.

Device Description

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous, and come in different sizes - Extra Small, Small, Medium, Large and Extra Large. Gloves meet the specification of ASTM D6319-10 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978-05, as guide lined by the FDA Medical Glove Guidance Manual.

AI/ML Overview

The provided text describes testing and acceptance criteria for Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs (K173942), focusing on their resistance to chemotherapy drug permeation. This is a medical device, not an AI/ML-based diagnostic or prognostic tool, therefore, many of the typical questions regarding AI model evaluation (like sample size for training set, number of experts for ground truth, MRMC studies, etc.) are not applicable to this submission.

The study proves the device's substantial equivalence to a predicate device (K172525) and its performance according to established industry standards.

Here's an analysis of the acceptance criteria and the study performance for the chemotherapy drug permeation aspect, as that's where specific performance data is provided:


Acceptance Criteria and Reported Device Performance

The key performance criterion for these gloves regarding chemotherapy drugs is the "Minimum Breakthrough Detection Time" as per ASTM D6978-05. While specific "acceptance criteria" (e.g., "must be >X minutes") are not explicitly stated as numerical thresholds for all drugs, the implicit acceptance criterion is to demonstrate a breakthrough time for each tested chemotherapy drug, and for certain drugs, specifically those known to have low permeation times (Carmustine and Thiotepa), to explicitly state their observed values, even if low. For the majority of drugs, the goal is often to demonstrate a breakthrough time exceeding a common testing duration (e.g., >240 minutes, which is 4 hours).

Table of Acceptance Criteria (Implicit) and Reported Device Performance:

Tested Chemotherapy DrugImplicit Acceptance Criteria (based on common medical glove standards and the data presented)Reported Device Performance (Minimum Breakthrough Detection Time in minutes)
Bleomycin SulfateDemonstrate permeation data; ideally >240 min>240
BusulfanDemonstrate permeation data; ideally >240 min>240
CarboplatinDemonstrate permeation data; ideally >240 min>240
CarmustineDemonstrate permeation data; acknowledge low time if applicable12.4
CisplatinDemonstrate permeation data; ideally >240 min>240
ChloroquineDemonstrate permeation data; ideally >240 min>240
CyclophosphamideDemonstrate permeation data; ideally >240 min>240
CyclosporinDemonstrate permeation data; ideally >240 min>240
Cytarabine HCIDemonstrate permeation data; ideally >240 min>240
DacarbazineDemonstrate permeation data; ideally >240 min>240
Daunorubicin HCIDemonstrate permeation data; ideally >240 min>240
DocetaxelDemonstrate permeation data; ideally >240 min>240
Doxorubicin HydrochlorideDemonstrate permeation data; ideally >240 min>240
Epirubicin (Ellence)Demonstrate permeation data; ideally >240 min>240
EtoposideDemonstrate permeation data; ideally >240 min>240
FludarabineDemonstrate permeation data; ideally >240 min>240
FluorouracilDemonstrate permeation data; ideally >240 min>240
GemcitabineDemonstrate permeation data; ideally >240 min>240
IdarubicinDemonstrate permeation data; ideally >240 min>240
IfosfamideDemonstrate permeation data; ideally >240 min>240
IrinotecanDemonstrate permeation data; ideally >240 min>240
Mechlorethamine HCIDemonstrate permeation data; ideally >240 min>240
MelphalanDemonstrate permeation data; ideally >240 min>240
MethotrexateDemonstrate permeation data; ideally >240 min>240
MitomycinDemonstrate permeation data; ideally >240 min>240
MitoxantroneDemonstrate permeation data; ideally >240 min>240
OxaliplatinDemonstrate permeation data; ideally >240 min>240
PaclitaxelDemonstrate permeation data; ideally >240 min>240
ParaplatinDemonstrate permeation data; ideally >240 min>240
RetrovirDemonstrate permeation data; ideally >240 min>240
RituximabDemonstrate permeation data; ideally >240 min>240
ThiotepaDemonstrate permeation data; acknowledge low time if applicable24.4
TopotecaneDemonstrate permeation data; ideally >240 min>240
TrisonexDemonstrate permeation data; ideally >240 min>240
VelcadeDemonstrate permeation data; ideally >240 min>240
Vincristine SulfateDemonstrate permeation data; ideally >240 min>240

The manufacturer clearly states the two drugs with "extremely low permeation times" (Carmustine and Thiotepa), indicating transparency and adherence to relevant standards (ASTM D6978-05).


Study Details:

This submission is for a medical device (gloves), not an AI/ML diagnostic. Therefore, the following points typical for AI/ML device evaluations are not applicable (N/A):

  1. Sample Size used for the Test Set and the Data Provenance:

    • Test Set Sample Size: The document does not specify a "sample size" in terms of number of gloves tested for each drug, but rather the performance against a list of chemicals. ASTM D6978-05 (Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs) outlines the methodology for such testing. Typically, this would involve a statistically significant number of specimens to ensure reproducibility and reliability, but the exact number isn't in this summary.
    • Data Provenance: The testing was done on the "Proposed Device" manufactured by Central Medicare Sdn. Bhd., located in Teluk Intan, Perak, Malaysia. The data is prospective, a result of new testing on the device.
  2. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts: N/A for this type of device. The "ground truth" is established by physical/chemical testing as per recognized ASTM standards.

  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: N/A. Physical/chemical testing does not involve human adjudication in this context.

  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A. This is not an AI/ML device.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: N/A. This is a physical product, not an algorithm.

  6. The type of ground truth used:

    • The ground truth for chemotherapy drug permeation is established through laboratory testing using standardized methods (ASTM D6978-05), which measures the time it takes for a chemical to permeate through a material. This is an objective, quantitative measurement, not based on expert consensus, pathology, or outcomes data in the traditional medical sense.
    • Other ground truths for the gloves' properties (tensile strength, elongation, freedom from holes, powder content) are established via their respective ASTM standards (e.g., ASTM D6319-10, ASTM D5151-06, ASTM D6124-06).
  7. The sample size for the training set: N/A. This is not an AI/ML device.

  8. How the ground truth for the training set was established: N/A. This is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The first logo on the left is the Department of Health & Human Services - USA logo. The second logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.

March 5, 2018

Central Medicare Sdn. Bhd. Arivalagan Subramaniam QA/RA Manager PT 2609-2620, BT 8, Jalan Changkat Jong Teluk Intan, 36000 My

Re: K173942

Trade/Device Name: Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZC, LZA Dated: December 27, 2017 Received: January 2, 2018

Dear Arivalagan Subramaniam:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K173942

Device Name

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs

Indications for Use (Describe)

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with chemotherapy drugs using ASTM D6978-05 and will be labeled with a statement of compliance and a summary of the testing results.

Chemotherapy Drug Permeation

The following chemicals have been tested with these gloves:

Tested Chemotherapy DrugConcentrate(mg/ml)Minimum Breakthrough Detection Time (minutes)
Bleomycin Sulfate15>240
Busulfan6>240
Carboplatin10>240
Carmustine3.312.4
Cisplatin1>240
Chloroquine50>240
Cyclophosphamide20>240
Cyclosporin100>240
Cytarabine HCI100>240
Dacarbazine10>240
Daunorubicin HCI5>240
Docetaxel10>240
Doxorubicin Hydrochloride2>240
Epirubicin (Ellence)2>240
Etoposide20>240
Fludarabine25>240
Fluorouracil50>240
Gemcitabine38>240
Idarubicin1>240
Ifosfamide50>240
Irinotecan20>240
Mechlorethamine HCI1>240
Melphalan5>240
Methotrexate25>240
Mitomycin0.5>240
Mitoxantrone2>240
Oxaliplatin5>240
Paclitaxel6>240
Paraplatin10>240
Retrovir10>240
Rituximab10>240
Thiotepa1024.4
Topotecane1>240

{3}------------------------------------------------

Trisonex1>240
Velcade1>240
Vincristine Sulfate1>240

*Please note that the following drugs have extremely low permeation times:

Carmustine: 12.4 minutes and Thiotepa: 24.4 minutes

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

510(k) Number: K173942

Date Prepared: 5th March 2018

1.0 Submitter:

Name:Arivalagan SubramaniamQA/RA Manager
Address:Central Medicare Sdn. Bhd.PT 2609-2620, Batu 8,Jalan Changkat Jong, 36000 Teluk Intan,Perak, Malaysia
Contact:Tel: +605-629 0000 Fax: +605-629 0001
510(k) Submission Prepared by:Muhammad Abdul RahmanRegulatory Affairs OfficerCentral Medicare Sdn. Bhd.

2.0 Name of the device:

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs

Common Names:Chemotherapy Gloves, Exam Gloves
Device Class:Class I
Classification Names:Patient Examination Gloves Specialty (21 CFR 880.6250 product code LZC, LZA)

3.0 Identification of the Legally Marketed Devices that equivalency is claimed:

Trade Name:Blue Non Sterile Powder Free NitrileExamination Gloves Tested for Use withChemotherapy Drugs
510(k) Number:K173942
Predicate Device510(k) Number:K172525

4.0 Description of The Device:

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous, and come in different sizes - Extra Small, Small, Medium, Large and Extra Large.

Gloves meet the specification of ASTM D6319-10 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978-05, as guide lined by the FDA Medical Glove Guidance Manual.

{5}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

5.0 Indications for Use

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs using ASTM D6978-05 and will be labeled with a statement of compliance and a summary of the testing results.

Chemotherapy Drug Permeation

The following chemicals have been tested with these gloves:

Tested Chemotherapy DrugConcentrate (mg/ml)Minimum BreakthroughDetection Time (minutes)
Bleomycin Sulfate15>240
Busulfan6>240
Carboplatin10>240
Carmustine3.312.4
Cisplatin1>240
Chloroquine50>240
Cyclophosphamide20>240
Cyclosporin100>240
Cytarabine HCI100>240
Dacarbazine10>240
Daunorubicin HCI5>240
Docetaxel10>240
Doxorubicin Hydrochloride2>240
Epirubicin (Ellence)2>240
Etoposide20>240
Fludarabine25>240
Fluorouracil50>240
Gemcitabine38>240
Idarubicin1>240
Ifosfamide50>240
Irinotecan20>240
Mechlorethamine HCI1>240
Melphalan5>240
Methotrexate25>240
Mitomycin0.5>240
Mitoxantrone2>240
Oxaliplatin5>240
Paclitaxel6>240
Paraplatin10>240
Retrovir10>240
Rituximab10>240
Thiotepa1024.4
Topotecane1>240
Trisonex1>240
Velcade1>240
Vincristine Sulfate1>240

{6}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

*Please note that the following drugs have extremely low permeation times: Carmustine: 12.4 minutes and Thiotepa: 24.4 minutes

6.0 Summary of the Technological Characteristics of the Device:

The proposed and predicate devices all share the same technological characteristics and design by meeting ASTM D6319-10 and following the FDA's Medical Glove Guidance Manual.

The purpose of this submission is to gain premarket clearance for the proposed device as a specialty chemotherapy glove and patient examination glove by proving substantial equivalence to Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs (K172525) as a predicate device. The proposed device will also be known as Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs.

Biocompatibility studies were performed on the proposed device. Under the conditions of the study, the proposed device is not a sensitizer, or an irritant.

The proposed and predicate devices are made of nitrile synthetic latex. The proposed device have similar technological characteristics compared to the predicate devices, as both devices meet ASTM D6319-10 and follow FDA's Medical Glove Guidance Manual. The technological characteristics and design for the proposed device have been assessed and confirmed that the standard specifications and performance are appropriate for the requirements of powder free nitrile examination gloves and specialty chemotherapy aloves.

The following tables are summaries of the technological characteristics, biocompatibility and testing for use with chemotherapy drugs of the proposed and predicate devices.

Proposed DevicePredicate DeviceRemark
SponsorCentral Medicare Sdn. Bhd.(as both the sponsor and manufacturer)Central Medicare Sdn. Bhd.(as both the sponsor and manufacturer)Same
Trade NameBlue Non Sterile Powder Free Nitrile ExaminationGloves Tested for Use with Chemotherapy DrugsBlue Non Sterile Powder Free Nitrile ExaminationGloves Tested for Use with Chemotherapy DrugsSame
510k NumberK173942K172525Same
Product CodeLZA, LZCLZA, LZCSame
Regulation Number21 CFR 880.625021 CFR 880.6250Same
ClassSame
Indications for UseBlue Non Sterile Powder Free Nitrile ExaminationGloves Tested for Use with Chemotherapy Drugsis a disposable device intended for medicalpurposes that is worn on the examiner's hands toprevent contamination between patient andexaminer.Blue Non Sterile Powder Free Nitrile ExaminationGloves Tested for Use with Chemotherapy Drugsis a disposable device intended for medicalpurposes that is worn on the examiner's hands toprevent contamination between patient andexaminer.Same
Powder or PowderFreePowder FreePowder FreeSame
Design FeatureAmbidextrousAmbidextrousSame
ColorBlueBlueSame
LabelingInformationSingle-use indication, powder free, device name,glove size, quantity, Nitrile Examination Gloves,Non SterileSingle-use indication, powder free, device name,glove size, quantity, Nitrile Examination Gloves,Non SterileSame

General Comparison Table:

{7}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

Chemotherapy Drug Permeation Claim:

Proposed DevicePredicate DeviceRemark
SponsorCentral Medicare Sdn. Bhd.(as both the sponsor and manufacturer)Central Medicare Sdn. Bhd.(as both the sponsor and manufacturer)Same
Chemotherapy DrugPermeation ClaimBleomycin Sulfate, Busulfan, Carboplatin,Carmustine, Cisplatin, Chloroquine,Cyclophosphamide, Cyclosporin, Cytarabine HCI,Dacarbazine, Daunorubicin HCI, Docetaxel,Doxorubicin Hydrochloride, Epirubicin (Ellence),Etoposide, Fludarabine, Fluorouracil,Gemcitabine, Idarubicin, Ifosfamide, Irinotecan,Mechlorethamine HCI, Melphalan, Methotrexate,Mitomycin, Mitoxantrone, Oxaliplatin, Paclitaxel,Paraplatin, Retrovir, Rituximab, Thiotepa,Topotecane, Trisonex, Velcade and VincristineSulfateFluorouracil, Etoposide (Toposar),Cyclophosphamid (Cytoxan), Carmustine(BCNU), Thiotepa, Paclitaxel (Taxol), DoxorubicinHydrochloride, Dacarbazine (DTIC), Cisplatin,Ifosfamide, Mitoxantrone, Vincristine SulfateProposedDevice Tested24 More Drugs

Device Dimensions Table:

Proposed Device
Size
Blue Non Sterile PowderFree Nitrile ExaminationGloves Tested for Usewith Chemotherapy Drugs(K173942)Length (mm)minimum 230
Width (mm)708095110120±10
Thickness (mm)
Cuff0.06±0.03
Palm0.08±0.03
Finger0.10±0.03
Predicate Device
Size
Blue Non Sterile PowderFree Nitrile ExaminationGloves Tested for Usewith Chemotherapy Drugs(K172525)Length (mm)minimum 230
Width (mm)708095110120±10
Thickness (mm)
Cuff0.06±0.03
Palm0.08±0.03
Finger0.10±0.03

{8}------------------------------------------------

central medicare sdn. Bhd. (660896 T)

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

Performance Comparison Table:

ItemProposed DevicePredicate DeviceRemark
Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs (K173942)Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs (K172525)
ColorBlueBlueSame
PhysicalPropertiesBeforeAgingTensile Strength15 Mpa min15 Mpa minSame
AgingUltimate Elongation500% min500% minSame
AfterAgingTensile Strength14 Mpa min14 Mpa minSame
AgingUltimate Elongation400% min400% minSame
Freedom from HolesIn accordance with ASTM D5151-06, following ASTM D6319 AQL 2.5/Inspection Level G-IIn accordance with ASTM D5151-06, following ASTM D6319 AQL 2.5/Inspection Level G-ISame
Powder ContentAvg. 0.44 mg per gloveAvg. 0.52 mg per gloveSame

{9}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

Chemotherapy Permeation Comparison Claim:

Chemotherapy Drug Testing - ASTM D6978-05
Tested Chemotherapy DrugConcentrate (mg/ml)Proposed Device Minimum Breakthrough Detection Time (minutes)Predicate DeviceRemark
Bleomycin Sulfate15>240Not TestedAdditional
Busulfan6>240Not TestedAdditional
Carboplatin10>240Not TestedAdditional
Carmustine3.312.412.4Same
Cisplatin1>240>240Same
Chloroquine50>240Not TestedAdditional
Cyclophosphamide20>240>240Same
Cyclosporin100>240Not TestedAdditional
Cytarabine HCI100>240Not TestedAdditional
Dacarbazine10>240>240Same
Daunorubicin HCI5>240Not TestedAdditional
Docetaxel10>240Not TestedAdditional
Doxorubicin Hydrochloride2>240>240Same
Epirubicin (Ellence)2>240Not TestedAdditional
Etoposide20>240>240Same
Fludarabine25>240Not TestedAdditional
Fluorouracil50>240>240Same
Gemcitabine38>240Not TestedAdditional
Idarubicin1>240Not TestedAdditional
Ifosfamide50>240>240Same
Irinotecan20>240Not TestedAdditional
Mechlorethamine HCI1>240Not TestedAdditional
Melphalan5>240Not TestedAdditional
Methotrexate25>240Not TestedAdditional
Mitomycin0.5>240Not TestedAdditional
Mitoxantrone2>240>240Same
Oxaliplatin5>240Not TestedAdditional
Paclitaxel6>240>240Same
Paraplatin10>240Not TestedAdditional
Retrovir10>240Not TestedAdditional
Rituximab10>240Not TestedAdditional
Thiotepa1024.424.4Same
Topotecane1>240Not TestedAdditional
Trisonex1>240Not TestedAdditional
Velcade1>240Not TestedAdditional
Vincristine Sulfate1>240>240Same

{10}------------------------------------------------

PT 2609 - 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

7.0 Based on Assessment of Non-Clinical Performance Data

The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above, and all mentioned devices follow ASTM D6319-10 reguirements and FDA's Medical Glove Guidance Manual.

The proposed device and it's predicate devices share intended use, are made of the same material, are within the same minimum specifications of thickness and length by meeting ASTM D6319-10, similar permeation rates for chemotherapy drugs as per ASTM D6978-05, similar labeling, physical properties, freedom from powder, biocompatibility and water tightness and by following the same quidance – FDA's Medical Glove Guidance Manual. Therefore, the proposed device is substantially equivalent to the predicate.

The test results demonstrated that the proposed device complies with the following standards:

ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

Guidance for Industry and FDA Staff Medical Glove Guidance Manual (Document issued on January 22, 2008)

The following standards were followed to carry out the testing:

ASTM D5151-06 (Reapproved 2011), Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6124-06 (Reaffirmation 2011), Standard Test Method for Residual Powder on Medical Gloves.

ISO 2859-1:1999, "Sampling Procedures for Inspection by Attributes – Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection.

ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

ASTM D412-16 Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers - Tension

ASTM D7160-16 Standard Practice for Determination of Expiration Dating for Medical Gloves

ASTM D573 - 04 Standard Test Method for Rubber—Deterioration in an Air Oven ASTM D6978-05 (Reapproved 2013), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs

8.0 Based on Assessment of Clinical Performance Data

N/A

9.0 Conclusion

Based on intended uses, technological characteristics and non-clinical performance data, the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs (K172525).

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.