K Number
K172525

Validate with FDA (Live)

Date Cleared
2017-11-13

(84 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with chemotherapy drugs using ASTM D6978-05 and will be labeled with a statement of compliance and a summary of the testing results.

Device Description

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous, and come in different sizes - Extra Small, Small, Medium, Large and Extra Large.

Gloves meet the specification of ASTM D6319-10 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978-05, as guide lined by the FDA Medical Glove Guidance Manual.

AI/ML Overview

This document describes the premarket notification for "Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs" (K172525). The primary purpose of the submission is to demonstrate substantial equivalence to a predicate device (K163146) based on technological characteristics and performance data, particularly regarding resistance to chemotherapy drug permeation.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are primarily defined by adherence to ASTM standards and comparative performance against the predicate device.

Acceptance Criteria/StandardReported Device Performance
ASTM D6319-10 (Standard Specification for Nitrile Examination Gloves for Medical Application):Device "meets the specification of ASTM D6319-10" and demonstrates "the standard specifications and performance are appropriate for the requirements of powder free nitrile examination gloves and specialty chemotherapy gloves."
* Physical Properties (Before Aging):
* Tensile Strength15 Mpa min
* Ultimate Elongation500% min
* Physical Properties (After Aging):
* Tensile Strength14 Mpa min
* Ultimate Elongation400% min
* Freedom from Holes (ASTM D5151-06)In accordance with ASTM D5151-06, following ASTM D6319 AQL 2.5/Inspection Level G-I
* Powder ContentMax. 0.52 mg per glove
ASTM D6978-05 (Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs):Gloves "have been tested for use with chemotherapy drugs using ASTM D6978-05" and "will be labeled with a statement of compliance and a summary of the testing results."
* Chemotherapy Drug Permeation (Minimum Breakthrough Detection Time in minutes)
* Carmustine (3.3 mg/ml)12.4 minutes
* Cisplatin (1.0 mg/ml)No breakthrough for up to 240 minutes
* Cyclophosphamide (20 mg/ml)No breakthrough for up to 240 minutes
* Doxorubicin HCl (2.0 mg/ml)No breakthrough for up to 240 minutes
* Etoposide (20.0 mg/ml)No breakthrough for up to 240 minutes
* Fluorouracil (50.0 mg/ml)No breakthrough for up to 240 minutes
* Mitoxantrone (2.0 mg/ml)No breakthrough for up to 240 minutes
* Paclitaxel (6.0 mg/ml)No breakthrough for up to 240 minutes
* Thiotepa (10.0 mg/ml)24.4 minutes
* Dacarbazine (10.0 mg/ml)No breakthrough for up to 240 minutes
* Ifosfamide (50.0 mg/ml)No breakthrough for up to 240 minutes
* Vincristine Sulfate (1.0 mg/ml)No breakthrough for up to 240 minutes
Biocompatibility Testing"Under the conditions of the study, the proposed device is not a sensitizer, or an irritant." These studies were performed on the proposed device.

2. Sample Size and Data Provenance for Test Set

  • Sample Size: The document does not explicitly state the sample size used for each specific test (e.g., number of gloves tested for permeation, holes, or physical properties). However, it implies that the testing was conducted "in accordance with ASTM D5151-06, following ASTM D6319 AQL 2.5/Inspection Level G-I" for freedom from holes, which dictates sample size based on lot size and AQL. For other tests like chemotherapy drug permeation (ASTM D6978-05), the standard itself would define the testing methodology and implied sample size per drug.
  • Data Provenance: The manufacturer, Central Medicare Sdn. Bhd., is located in Teluk Intan, Perak, Malaysia. The tests were likely conducted in Malaysia or by a testing facility in adherence to the specified ASTM standards. The data is retrospective in the sense that it represents characterization of the manufactured gloves to support the 510(k) submission.

3. Number of Experts and Qualifications for Ground Truth

This device is not an AI/imaging device requiring expert interpretation for ground truth. The "ground truth" for glove performance is established through standardized physical and chemical tests (e.g., ASTM D6319-10, ASTM D6978-05). The "experts" are the qualified laboratory personnel performing these tests according to the standard operating procedures of the testing facilities. Their qualifications typically involve training and certification in laboratory testing methodologies relevant to material science and medical device performance standards.

4. Adjudication Method for Test Set

Not applicable for this type of device and testing. Test results are objective measurements from standardized protocols rather than subjective interpretations requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is a physical medical device (examination glove), not an imaging or diagnostic AI system.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

Not applicable. This is a physical medical device. The performance evaluated is the physical and chemical properties of the glove itself.

7. Type of Ground Truth Used

The ground truth used is based on:

  • Performance Standards: Adherence to established industry standards such as ASTM D6319-10 (for general glove properties) and ASTM D6978-05 (for chemotherapy drug permeation).
  • Objective Laboratory Measurements: Direct measurements of tensile strength, elongation, freedom from holes, powder content, and breakthrough times for chemotherapy drugs. These are quantifiable results obtained through controlled experimental procedures.
  • Biocompatibility Studies: In vitro or in vivo tests to assess sensitization and irritation, which also yield objective results based on established protocols.

8. Sample Size for Training Set

Not applicable. This is not an AI/machine learning device that requires a training set. The "device" is a physical product.

9. How Ground Truth for Training Set Was Established

Not applicable, as no training set is relevant for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The word "Administration" is on the second line.

November 13, 2017

Central Medicare Sdn. Bhd. Arivalagan Subramaniam QA/RA Manager PT 2609 - 2620, BT 8, Jalan Changkat Jong Teluk Intan, 36000 MALAYSIA

Re: K172525

Trade/Device Name: Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZC, LZA Dated: August 15, 2017 Received: August 21, 2017

Dear Arivalagan Subramaniam:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K172525

Device Name

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs

Indications for Use (Describe)

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with chemotherapy drugs using ASTM D6978-05 and will be labeled with a statement of compliance and a summary of the testing results.

Chemotherapy Drug Permeation

The following chemicals have been tested with these gloves:

Tested Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time (minutes)
* Carmustine (3.3 mg/ml)Min minutes before breakthrough = 12.4
Cisplatin (1.0 mg/ml)No breakthrough for up to 240 minutes
Cyclophosphamide (20 mg/ml)No breakthrough for up to 240 minutes
Doxorubicin HCl (2.0 mg/ml)No breakthrough for up to 240 minutes
Etoposide (20.0 mg/ml)No breakthrough for up to 240 minutes
Fluorouracil (50.0 mg/ml)No breakthrough for up to 240 minutes
Mitoxantrone (2.0 mg/ml)No breakthrough for up to 240 minutes
Paclitaxel (6.0 mg/ml)No breakthrough for up to 240 minutes
* Thiotepa (10.0 mg/ml)Min minutes before breakthrough = 24.4
Dacarbazine (10.0 mg/ml)No breakthrough for up to 240 minutes
Ifosfamide (50.0 mg/ml)No breakthrough for up to 240 minutes
Vincristine Sulfate (1.0 mg/ml)No breakthrough for up to 240 minutes
  • Please note that the following drugs have extremely low permeation times:

Carmustine: 12.4 minutes and Thiotepa: 24.4 minutes

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 TelukIntan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

510(k) Number: K172525

Date Prepared: 13th October 2017

1.0 Submitter:

Name:Arivalagan SubramaniamQA/RA Manager
Address:Central Medicare Sdn. Bhd.PT 2609-2620, Batu 8,Jalan Changkat Jong, 36000 Teluk Intan,Perak, Malaysia
Contact:Tel: +605-629 0000 Fax: +605-629 0001
510(k) Submission Prepared by:Muhammad Abdul RahmanRegulatory Affairs OfficerCentral Medicare Sdn. Bhd.

2.0 Name of the device:

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs

Common Names:Chemotherapy Gloves, Exam Gloves
Device Class:Class I
Classification Names:Patient Examination Gloves Specialty (21 CFR 880.6250 product code LZC, LZA)

3.0 Identification of the Legally Marketed Devices that equivalency is claimed:

Trade Name:Blue Non Sterile Powder Free Nitrile
Examination Gloves Tested for Use with
Chemotherapy Drugs

510k Number: K172525

4.0 Description of the Device:

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous, and come in different sizes - Extra Small, Small, Medium, Large and Extra Large.

{4}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 TelukIntan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

Gloves meet the specification of ASTM D6319-10 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978-05, as guide lined by the FDA Medical Glove Guidance Manual.

5.0 Intended Use of the Device:

Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with chemotherapy drugs using ASTM D6978-05 and will be labeled with a statement of compliance and a summary of the testing results.

Chemotherapy Drug Permeation

The following chemicals have been tested with these gloves:

Tested Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time (minutes)
*Carmustine (3.3 mg/ml)Min minutes before breakthrough = 12.4
Cisplatin (1.0 mg/ml)No breakthrough for up to 240 minutes
Cyclophosphamide (20 mg/ml)No breakthrough for up to 240 minutes
Doxorubicin HCl (2.0 mg/ml)No breakthrough for up to 240 minutes
Etoposide (20.0 mg/ml)No breakthrough for up to 240 minutes
Fluorouracil (50.0 mh/ml)No breakthrough for up to 240 minutes
Mitoxantrone (2.0 mg/ml)No breakthrough for up to 240 minutes
Paclitaxel (6.0 mg/ml)No breakthrough for up to 240 minutes
*Thiotepa (10.0 mg/ml)Min minutes before breakthrough = 24.4
Dacarbazine (10.0 mg/ml)No breakthrough for up to 240 minutes
Ifosfamide (50.0 mg/ml)No breakthrough for up to 240 minutes
Vincristine Sulfate (1.0 mg/ml)No breakthrough for up to 240 minutes
  • Please note that the following drugs have extremely low permeation times: Carmustine: 12.4 minutes and Thiotepa: 24.4 minutes

6.0 Summarv of the Technological Characteristics of the Device:

The proposed and predicate devices all share the same technological characteristics and design by meeting ASTM D6319-10 and following the FDA's Medical Glove Guidance Manual.

The purpose of this submission is to demonstrate that the proposed device is substantially equivalent, to a legally marketed predicate device, a specialty chemotherapy glove, Hebei Hongsen Plastics Technology Co, Ltd's POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs (K163146). The proposed device will be known as Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs.

{5}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 TelukIntan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

Biocompatibility studies were performed on the proposed device. Under the conditions of the study, the proposed device is not a sensitizer, or an irritant.

The proposed and predicate devices are made of nitrile synthetic latex. The proposed device have similar technological characteristics compared to the predicate devices, as both devices meet ASTM D6319-10 and follow FDA's Medical Glove Guidance Manual.

The technological characteristics and design for the proposed device have been assessed and confirmed that the standard specifications and performance are appropriate for the requirements of powder free nitrile examination gloves and specialty chemotherapy gloves.

The following tables are summaries of the technological characteristics, biocompatibility and testing for use with chemotherapy drugs of the proposed and predicate devices.

Proposed DevicePredicate DeviceRemark
SponsorCentral Medicare Sdn. Bhd.(as both the sponsor and manufacturer)Hebei Hongsen PlasticsTechnology Co, Ltd
Trade NameBlue Non Sterile Powder Free Nitrile ExaminationGloves Tested for Use with Chemotherapy DrugsPOWDER FREE Blue Nitrile GLOVES,Tested for Use with Chemotherapy DrugsSimilar
510k NumberK172525K163146
Product CodeLZA, LZCLZA, LZCSame
Regulation Number21 CFR 880.625021 CFR 880.6250Same
ClassISame
Indications for UseBlue Non Sterile Powder Free Nitrile ExaminationGloves Tested for Use with Chemotherapy Drugsis a disposable device intended for medicalpurposes that is worn on the examiner's hands toprevent contamination between patient andexaminer.The POWDER FREE Blue NitrileGLOVES, Tested for Use withChemotherapyDrugs is a disposabledevice intended for medical purposes thatis worn on the examiner's hands toprevent contamination between patient andexaminer.Same
Powder or PowderFreePowder FreePowder FreeSame
Design FeatureAmbidextrousAmbidextrousSame
ColorBlueBlueSimilar
LabelingInformationSingle-use indication, powder free, device name,glove size, quantity, Nitrile Examination Gloves,Non SterileSingle-use indication, powder free, devicename, glove size, quantity, NitrileExamination Gloves, Non SterileSame
ChemotherapyDrugPermeation ClaimFluorouracil, Etoposide (Toposar),Cyclophosphamid (Cytoxan), Carmustine (BCNU),Thiotepa, Paclitaxel (Taxol), DoxorubicinHydrochloride, Dacarbazine (DTIC), Cisplatin,lfosfamide, Mitoxantrone, Vincristine SulfateFluorouracil, Etoposide (Toposar),Cyclophosphamid (Cytoxan), Carmustine(BCNU), Thiotepa, Paclitaxel (Taxol),Doxorubicin Hydrochloride, Dacarbazine(DTIC), Cisplatin, Carboplatin, Docetaxel,lfosfamide, Irinotecan, MechlorethamineHCL, Methotrexate, Mitomycin C,Mitoxantrone, Vincristine SulfateSimilar

General Comparison Table:

{6}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

De vice Dimensions Table:

Proposed Device
SizeTolerance
XSSMLXL
Blue Non Sterile PowderFree Nitrile ExaminationGloves Tested for Usewith Chemotherapy DrugsLengthminimum 230
Width (mm)708095110120$\pm0.10$
Thickness (mm)
Cuff0.06$\pm0.03$
Palm0.08$\pm0.03$
Finger0.10$\pm0.03$
Predicate Device
SizeTolerance
XSSMLXL
POWDER FREE BlueNitrile GLOVES, Testedfor Use withChemotherapy Drugs(163146)Lengthminimum 230
Width (mm)708095110120$\pm0.10$
Thickness (mm)
Cuff0.06$\pm0.03$
Palm0.08$\pm0.03$
Finger0.10$\pm0.03$

Performance Comparison Table:

Proposed DevicePredicate Device
ItemBlue Non Sterile PowderFree Nitrile ExaminationGloves Tested for Use withChemotherapy DrugsPOWDER FREE Blue NitrileGLOVES, Tested for Usewith Chemotherapy DrugsRemark
ColorBlueBlueSimilar
PhysicalPropertiesBeforeTensile Strength15 Mpa min15 Mpa minSame
AgingUltimate Elongation500% min500% minSame
AfterAgingTensile Strength14 Mpa min14 Mpa minSame
Ultimate Elongation400% min400% minSame
Freedom from HolesIn accordance with ASTMD5151-06, following ASTMD6319 AQL 2.5/InspectionLevel G-IBe free from holes whentested in accordance withASTM D5151 AQL 1.5Similar
Powder ContentMax. 0.52 mg per gloveMax. 0.32 mg per gloveSimilar

{7}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

Che motherapy Permeation Comparison Claim:

Chemotherapy Drug Testing - ASTM D6978-05
Proposed DevicePredicate Device
Tested ChemotherapyDrug andConcentrationMinimumBreakthroughDetection Time(minutes)Tested ChemotherapyDrug andConcentrationAverage BreakthroughDetection Time(minutes)Remark
Fluorouracil (50.0 mh/ml)No breakthrough for up to240 minutesFluorouracil (50.0 mh/ml)No breakthrough for up to240 minutesSame
Etoposide (20.0 mg/ml)No breakthrough for up to240 minutesEtoposide (20.0 mg/ml)No breakthrough for up to240 minutesSame
Cyclophosphamide (20mg/ml)No breakthrough for up to240 minutesCyclophosphamide (20mg/ml)No breakthrough for up to240 minutesSame
Carmustine (3.3 mg/ml)Mn minutes beforebreakthrough = 12.4Carmustine (3.3 mg/ml)Mn minutes beforebreakthrough = 45.0Similar
Thiotepa (10.0 mg/ml)Mn minutes beforebreakthrough = 24.4Thiotepa (10.0 mg/ml)Mn minutes beforebreakthrough = 30.0Similar
Paclitaxel (6.0 mg/ml)No breakthrough for up to240 minutesPaclitaxel (6.0 mg/ml)No breakthrough for up to240 minutesSame
Doxorubicin HCl (2.0mg/ml)No breakthrough for up to240 minutesDoxorubicin HCl (2.0mg/ml)No breakthrough for up to240 minutesSame
Dacarbazine (10.0mg/ml)No breakthrough for up to240 minutesDacarbazine (10.0mg/ml)No breakthrough for up to240 minutesSame
Cisplatin (1.0 mg/ml)No breakthrough for up to240 minutesCisplatin (1.0 mg/ml)No breakthrough for up to240 minutesSame
Carboplatin (10.0 mg/ml)Not TestedCarboplatin (10.0 mg/ml)No breakthrough for up to240 minutesWill not beclaimed
Docetaxel (10.0 mg/ml)Not TestedDocetaxel (10.0 mg/ml)No breakthrough for up to240 minutesWill not beclaimed
lfosfamide (50.0 mg/ml)No breakthrough for up to240 minutesIfosfamide (50.0 mg/ml)No breakthrough for up to240 minutesSame
Irinotecan (20.0 mg/ml)Not TestedIrinotecan (20.0 mg/ml)No breakthrough for up to240 minutesWill not beclaimed
Mechlorethamine HCL(1.0 mg/ml)Not TestedMechlorethamine HCL(1.0 mg/ml)No breakthrough for up to240 minutesWill not beclaimed
Methotrexate (25.0mg/ml)Not TestedMethotrexate (25.0mg/ml)No breakthrough for up to240 minutesWill not beclaimed
Mitomycin C (0.5 mg/ml)Not TestedMitomycin C (0.5 mg/ml)No breakthrough for up to240 minutesWill not beclaimed
Mitoxantrone (2.0 mg/ml)No breakthrough for up to240 minutesMitoxantrone (2.0 mg/ml)No breakthrough for up to240 minutesSame
Vincristine Sulfate (1.0mg/ml)No breakthrough for up to240 minutesVincristine Sulfate (1.0mg/ml)No breakthrough for up to240 minutesSame

{8}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 TelukIntan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

510(k) Summary

7.0 Non-Clinical Performance Data

The proposed device and its predicate devices share intended use, are made of the same material, are within the same minimum specifications of thickness and length by meeting ASTM D6319-10, similar permeation rates for chemotherapy drugs as per ASTM D6978-05, similar labeling, physical properties, freedom from powder, biocompatibility and water tightness.

The above test results demonstrated that the proposed device complies with the following standards: ASTM D6319.10

8.0 Clinical Performance Data

NA

9.0 Conclusion

Based on intended uses, technological characteristics and non-clinical performance data, the proposed device is substantially equivalent to the predicate device POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs (K163146).

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.