Search Filters

Search Results

Found 1146 results

510(k) Data Aggregation

    Why did this record match?
    Product Code :

    LZA

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Gloves were tested for use with Chemotherapy drugs, Fentanyl Citrate, Simulated Gastric acid and Xylazine in Fentanyl Citrate in accordance with ASTM D6978-05 Standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.

    Device Description

    Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate is a Class I patient examination glove and specialty chemotherapy gloves, that made from synthetic nitrile latex. They are pink color, non-sterile, powder-free, fingertip textured, ambidextrous with beaded cuff, and single use only, and come in different sizes- XS, S, M, L, XL and XXL.

    The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been been tested for biocompatibility and permeability to chemotherapy drugs, Fentanyl Citrate, Gastric Acid and Xylazine in Fentanyl Citrate.

    AI/ML Overview

    This document describes the acceptance criteria and the study that proves the device meets the acceptance criteria for a medical device submitted for FDA 510(k) clearance.

    The device in question is Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate.


    Acceptance Criteria and Device Performance Study

    The acceptance criteria for this device are primarily based on established industry standards for medical examination gloves and specific permeation resistance tests for various chemical substances. The study conducted to prove the device meets these criteria is a series of non-clinical performance tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test PerformedMethodologyAcceptance CriteriaReported Device Performance
    Freedom From HolesASTM D6319-19, ASTM D5151-19Meet requirement inspection level G-1, AQL 2.5 (ISO2859-1)Pass
    Dimension - LengthASTM D6319-19Minimum 220mm for size XS-S
    Minimum 230mm for size M-XXLPass
    Dimension - WidthASTM D6319-19XS: 70±10mm
    S: 80±10mm
    M: 95±10mm
    L: 110±10mm
    XL: 120±10mm
    XXL: 130±10mmPass
    Dimension - ThicknessASTM D6319-19Finger: 0.05mm (min)
    Palm: 0.05mm (min)Pass
    Physical PropertiesASTM D6319-19, ASTM D412-16Before aging:
    Tensile Strength (Min 14 Mpa)
    Elongation (Min 500%)
    After aging:
    Tensile Strength (Min 14 Mpa)
    Elongation (Min 400%)Pass (All criteria met)
    Powder ResidueASTM D6319-19, ASTM D6124-06Not more than 2 mg per glovePass
    Skin Irritation (Biocompatibility)ISO10993-23:2021Under the conditions of the study, not a primary skin irritantPass (Not a primary skin irritant)
    Skin Sensitization (Biocompatibility)ISO10993-10:2021Under the conditions of the study, not a contact sensitizerPass (Not a contact sensitizer)
    Acute Systemic Toxicity (Biocompatibility)ISO 10993-11:2017Under the conditions of the study, no signs of acute systemic toxicity were observedPass (No signs of acute systemic toxicity were observed)
    Chemotherapy Drugs PermeationASTM D6978-05 (2023)Minimum Breakthrough Detection Time (minutes)
    Carboplatin: >240
    Carmustine: Varies (15.3 min reported)
    Cisplatin: >240
    Cyclophosphamide: >240
    Cytarabine HCl: >240
    Dacarbazine: >240
    Daunorubicin HCl: >240
    Doxorubicin HCl: >240
    Etoposide: >240
    5-Fluorouracil: >240
    Gemcitabine HCl: >240
    Ifosfamide: >240
    Irinotecan HCl: >240
    Mechlorethamine HCl: >240
    Melphalan HCl: >240
    Methotrexate: >240
    Mitomycin-C: >240
    Mitoxantrone HCl: >240
    Pacilitaxel: >240
    Thiotepa: Varies (28.2 min reported)Carboplatin: >240 min.
    Carmustine: 15.3 min.
    Cisplatin: >240 min.
    Cyclophosphamide: >240 min.
    Cytarabine HCl: >240 min.
    Dacarbazine: >240 min.
    Daunorubicin HCl: >240 min.
    Doxorubicin HCl: >240 min.
    Etoposide: >240 min.
    5-Fluorouracil: >240 min.
    Gemcitabine HCl: >240 min.
    Ifosfamide: >240 min.
    Irinotecan HCl: >240 min.
    Mechlorethamine HCl: >240 min.
    Melphalan HCl: >240 min.
    Methotrexate: >240 min.
    Mitomycin-C: >240 min.
    Mitoxantrone HCl: >240 min.
    Pacilitaxel: >240 min.
    Thiotepa: 28.2 min.
    Note: The document explicitly states that Carmustine and Thiotepa have "extremely low permeation times" and are accompanied by a "Warning: Do not use with Carmustine and Thiotepa," indicating that the stated permeation times for these two drugs are not >240 minutes and are accepted for the specified usage.
    Fentanyl, Gastric Acid, Xylazine PermeationASTM D6978-05 (2023)Minimum Breakthrough Detection Time (minutes)
    Fentanyl Citrate Injection: >240
    Simulated Gastric Acid: >240
    Fentanyl Citrate:Xylazine HCl (50:50): >240Fentanyl Citrate Injection: >240 min.
    Simulated Gastric Acid: >240 min.
    Fentanyl Citrate (50mcg/ml): Xylazine HCl (100mg/ml), 50:50: >240 min.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes (number of gloves) used for each individual test. However, it indicates that the tests were conducted according to established ASTM and ISO standards, which typically specify appropriate sample sizes for such tests to ensure statistical validity.

    The data provenance is through non-clinical testing performed in accordance with recognized international and national standards (ASTM, ISO). This implies the data were prospectively generated for the purpose of this submission. The country of origin of the data is not specified, but the use of international standards suggests a globally accepted testing methodology.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable (N/A) as the device is a physical product (medical glove) and the performance is evaluated through objective, standardized laboratory tests (e.g., measuring permeation time, tensile strength, dimensions) rather than expert interpretation of images or clinical data. There is no "ground truth" established by human experts in the context of diagnostic accuracy for this type of device.

    4. Adjudication Method for the Test Set

    This information is not applicable (N/A) for the same reason as above. Adjudication methods like 2+1 or 3+1 are relevant for studies involving human interpretation (e.g., radiology reads) where discrepancies need to be resolved. The tests performed for this device are objective measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable (N/A). This document pertains to the clearance of a physical medical device (gloves), not an AI algorithm or a diagnostic imaging device that would involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This information is not applicable (N/A). As mentioned, this is a physical medical device, not a software algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by objective, quantitative measurements obtained through standardized laboratory testing methodologies (ASTM and ISO standards). For example:

    • Permeation: Measured breakthrough time of specific chemicals through the glove material.
    • Physical Properties: Measured tensile strength, elongation, length, width, and thickness.
    • Biocompatibility: Observed biological responses (e.g., irritation, sensitization, systemic toxicity) in controlled in-vitro or in-vivo (animal) tests as per ISO standards.
      This is not based on expert consensus, pathology, or outcomes data in the clinical sense, but rather on direct physical and chemical testing.

    8. The Sample Size for the Training Set

    This information is not applicable (N/A). This device is a manufactured product that undergoes performance and biocompatibility testing. There is no "training set" in the context of machine learning. The manufacturing process is designed and validated to consistently produce gloves meeting the specified criteria.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable (N/A) as there is no training set for this device. The "ground truth" for the device's design and manufacturing is derived from the established standards (ASTM, ISO) that define the performance characteristics expected of such gloves. Quality control during manufacturing ensures conformity to these standards.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Product Code :

    LZA

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs, fentanyl citrate, xylazine, gastric acid, fentanyl in gastric acid and xylazine in gastric acid in accordance with ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical gloves to Permeation by Chemotherapy Drugs.

    Warning: do not use with Carmustine and Thiotepa.

    The maximum testing time is 240 minutes. Please note that the following drug has an extremely low permeation time:

    • Carmustine (BCNU) 3.3 mg/ml 21.5 minutes
    • Thiotepa 10.0 mg/ml 13.6 minutes
    Device Description

    "Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs, Fentanyl Citrate, Xylazine, Gastric acid, Fentanyl in Gastric acid and Xylazine in Gastric acid" is a patient examination glove made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of this medical device is to provide single use barrier protection for the wearer and the device meets the specifications for Barrier Protection and tensile properties as defined in ASTM D6319-19, Standard specification for Nitrile Examination Gloves.

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for a medical device: "Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs, Fentanyl Citrate, Xylazine, Gastric acid, Fentanyl in Gastric acid and Xylazine in Gastric acid."

    This is a physical medical device (a glove), not a software device or an AI/ML algorithm. Therefore, many of the requested categories related to AI/ML model validation (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this type of FDA submission.

    The document assesses the device's physical properties and its resistance to permeation by various hazardous substances. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" refer to bench testing against established ASTM and CFR standards for medical gloves.

    Here's the breakdown of the relevant information from the document, tailored to the nature of this device:


    Acceptance Criteria and Device Performance for Disposable Nitrile Examination Gloves

    1. Table of Acceptance Criteria and Reported Device Performance

    The device (Disposable Powder Free Nitrile Examination Glove, Blue Color) was tested against several physical and chemical permeation standards. The acceptance criteria and results are outlined in the "Assessment of Non-Clinical Performance Data" section (Page 17-19) and the "Indications for Use" section (Page 4-5) and summarized within the "510(k) Summary" (Page 7-11).

    TestTest MethodPurposeAcceptance CriteriaResults
    DimensionASTM D6319-19Determine the geometrical dimension of glovesLength: Short cuff: ≥230mm; Long cuff: ≥300mm
    Thickness: Palm: ≥ 0.05 mm; Finger: ≥ 0.05 mm; Cuff: ≥ 0.05 mm
    Palm Width: XS: 70 ± 10 mm; S: 80 ± 10 mm; M: 95 ± 10 mm; L: 110 ± 10 mm; XL: 120 ± 10 mmPass
    Freedom from holes (Water leak)21 CFR 800.20. & ASTM D5151-19Detect the holes on the gloves.G-I/ AQL 2.5Pass
    Tensile strength (Before aging/ After aging)ASTM D6319-19Evaluate the tensile (tension) properties of the gloves. In addition, it also determines the influence of elevated temperature on the physical properties of gloves.Before Aging: ≥14MPa
    After Aging: ≥14MPaPass
    Elongation (Before aging/ After aging)ASTM D6319-19Related to tensile properties, measures stretchiness.Before Aging: ≥500%
    After Aging: ≥400%Pass
    Powder ResidualASTM D6319-19Determine the average powder mass found on the gloves< 2mg per glovePass
    Biocompatibility - CytotoxicityAAMI/ANSI/ISO 10993-5Determine the cytotoxicity potential of gloveNo in vitro cytotoxic as described in ISO 10993-5Pass
    Biocompatibility - Skin Sensitization and IrritationAAMI/ANSI/ISO 10993-10Determine the potential of glove to promote skin sensitization after repeated applications.
    Determine the potential of gloves to promote skin irritation after repeated applications.No dermal reactions indicative of delayed contact hypersensitivity
    No skin irritation, cumulative irritation index to be 0.Pass
    Resistance of Gloves to Permeation by Chemotherapy and Other Liquid Hazardous Drugs, and Non-drugs SolutionASTM D6978-05Assessment of medical gloves to permeation by chemotherapy and other liquid hazardous drugs, and non-drugs solution.The resistance of the device to permeation were challenged against 59 hazardous drugs and 1 non-drugs simulated solution.
    General Acceptance Standard (Implicit): Maintain barrier protection for a minimum period. The tables on pages 4-5 and 7-11 list Minimum Breakthrough Detection Time (Min.) for each substance. For most substances, the acceptance criterion was ">240 minutes", meaning no breakthrough occurred within the 240-minute test period. For Carmustine and Thiotepa, specific breakthrough times (21.5 and 13.6 minutes, respectively) were observed and noted as limitations/warnings.Pass

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not explicitly state numerical sample sizes for each specific test (e.g., how many gloves were tested for a burst, tensile, or permeation). However, it implies that testing was conducted according to the specified ASTM and FDA methodologies, which inherently define minimum sample sizes for material testing (e.g., AQL levels for freedom from holes).
    • Data Provenance: The tests are "bench testing" conducted on physical glove samples. The manufacturer is "Ever Global (Vietnam) Enterprise Corporation" from Vietnam, so the testing was likely conducted in accordance with international standards at a facility associated with the manufacturer or a certified testing lab. The data is non-clinical performance data from laboratory experiments, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. As this is a physical device subject to material property and permeation testing, "ground truth" is established by adherence to physical measurement standards and chemical analysis protocols (e.g., ASTM D6319-19 for physical properties, ASTM D6978-05 for permeation). No human expert interpretation or consensus is required to establish the "truth" of a chemical breakthrough time or tensile strength value.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not Applicable. Adjudication methods are relevant in subjective human interpretation tasks (e.g., radiology image reading). For objective bench testing of physical properties, results are determined by instrumentation and adherence to standardized protocols, not human adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. MRMC studies are used for evaluating the impact of AI on human reader performance, typically in medical imaging. This is a physical glove, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" is established through objective, standardized laboratory measurement methods as defined by the referenced ASTM (American Society for Testing and Materials) and CFR (Code of Federal Regulations) standards.
      • For physical properties (Dimensions, Tensile Strength, Elongation, Powder Residual, Freedom from Holes), the ground truth is the measured physical values against the defined thresholds in ASTM D6319-19 and 21 CFR 800.20 & ASTM D5151-19.
      • For chemical permeation (Resistance to Hazardous Drugs), the ground truth is the measured breakthrough time determined by chemical analysis according to ASTM D6978-05.
      • For Biocompatibility, the ground truth is the biological response (cytotoxicity, dermal reactions) as measured per AAMI/ANSI/ISO 10993-5 and 10993-10 standards.

    8. The sample size for the training set:

    • Not Applicable. This is a physical product, not a machine learning model; therefore, there is no "training set." The product's design and manufacturing process are developed through engineering and material science, not data training.

    9. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set for this physical device.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Product Code :

    LZA

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Halyard Purple Nitrile-XTRA* Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate and Fentanyl Citrate in Simulated Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes:

    • Arsenic Trioxide (1 mg/ml)
    • Bendamustine, (5 mg/ml)
    • Blenoxane (15 mg/ml)
    • Bleomycin (15 mg/ml)
    • Bortezomib (1 mg/ml)
    • Busulfan (6 mg/ml)
    • Carboplatln (10 mg/ml)
    • Carfilzomib (2 mg/ml)
    • Cetuximab (2 mg/ml)
    • Cisplatin (1 mg/ml)
    • Cyclophosphamide (Cytoxan) (20 mg/ml)
    • Cytarabine (100 mg/ml)
    • Dacarbazine (DTIC) {10 mg/ml)
    • Daunorubicin {5 mg/ml)
    • Decitabine (5 mg/ml)
    • Docetaxel (10 mg/ml)
    • Doxorubicin HCL (2 mg/ml)
    • Ellence (2 mg/ml)
    • Erbitux (2 mg/ml)
    • Eribilin Mesylate (0.5 mg/ml)
    • Etoposide (Toposar) (20 mg/ml)
    • Fludarabine (25 mg/ml)
    • Fulvestrant (50 mg/ml)
    • Gemcitabine (Gemzar) (38 mg/ml)
    • Idarubicin (1 mg/ml)
    • Ifosfamide (IFEX) (50 mg/ml)
    • Irinotecan (20 mg/ml)
    • Mechlorethamine HCL (1 mg/ml)
    • Melphalan (5 mg/ml)
    • Methotrexate (25 mg/ml)
    • Mitomycin C (0.5 mg/ml)
    • Mitoxantrone (2 mg/ml)
    • Oxaliplatin (2 mg/ml)
    • Paclitaxel (Taxol) (6 mg/ml)
    • Paraplatin (10 mg/ml)
    • Pemetrexed Disodium (25 mg/ml)
    • Pertuzumab (30 mg/ml)
    • Raltitrexed (0.5 mg/ml)
    • Rituximab (Rituxan) (10 mg/ml)
    • Temsirolimus (25 mg/ml)
    • Thiotepa (10 mg/ml)
    • Topotecan HCL (1 mg/ml)
    • Trastuzumab (21 mg/ml)
    • Trisenox (1 mg/ml)
    • Velcade (1 mg/ml)
    • Vinblastine (1 mg/ml)
    • Vinorelbine (10 mg/ml)

    Carmustine (3.3 mg/ml) permeation occurred at 60.0 minutes.

    The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes:

    • Fentanyl Citrate Injection (100 mcg/2 ml)
    • Gastric Acid Fluid/Fentanyl Citrate Injection Mix (50/50 Solution)

    Caution: Testing showed a minimum breakthrough time of 60.0 minutes with Carmustine.

    The following hazardous drugs and concentration had NO breakthrough detected up to 240 minutes:

    • Cytovene (10 mg/ml)
    • Retrovir (10 mg/ml)
    • Triclosan (2 mg/ml)
    • Zoledronic Acid (0.8 mg/ml)
    Device Description

    Halyard Purple Nitrile-XTRA* Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate and Fentanyl Citrate in Simulated Gastric Acid are disposable, 12"purple-colored, chlorinated, nitrile, powder-free, textured fingertip, ambidextrous, nonsterile patient examination gloves.

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter and summary for a medical glove, not an AI-powered medical device. Therefore, many of the requested fields related to AI study design (like "multi-reader multi-case (MRMC) comparative effectiveness study," "standalone performance," "number of experts," etc.) are not applicable and cannot be found in the document.

    However, I can extract the acceptance criteria and performance data for the glove based on the provided information, focusing on the non-clinical and clinical tests described.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestStandardAcceptance CriteriaReported Device Performance
    DimensionsASTM D 6319Length ≥230 mm
    Palm Width Size X-Small: 60 – 80 mm
    Small: 70 - 90 mm
    Med: 85–105 mm
    Large: 100 - 120 mm
    X-Large: 110-130 mm
    XX-Large: 120-140 mm
    Finger thickness ≥0.05 mm
    Palm thickness ≥0.05 mm
    Cuff thickness ≥0.05 mmMeets requirements
    Physical PropertiesASTM D 6319AQL 4.0
    Before Aging: Tensile Strength: ≥14 MPa, Ultimate elongation: ≥500%
    After Aging: Tensile Strength: ≥14 MPa, Ultimate elongation: ≥400%Meets requirements (Tensile strength and elongation before and after aging met requirements)
    Freedom from PinholesASTM D 6319
    ASTM D 5151AQL 2.5%
    No leakageMeets requirements (Meets the 2.5% AQL requirement for leakage)
    Powder FreeASTM D 6124
    ASTM D 6319≤ 2 mg / gloveMeets requirements (Average of 0.4 mg/glove, within the
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Product Code :

    LZA

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue); and Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) - Extended Cuff, is a sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    These gloves were tested for use with chemotherapy drugs and fentanyl citrate as per ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) and Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) – Extended Cuff, are disposable, single-use, sterile, blue-colored, and powder-free examination gloves made from nitrile latex.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to medical examination gloves, not an AI-powered medical device. Therefore, the document does not contain information about the acceptance criteria and study proving an AI device meets those criteria.

    The information typically found in such a submission would include:

    • Device performance: This document explicitly states the "Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue)" has been tested for resistance to permeation by chemotherapy drugs and fentanyl citrate as per ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
    • Acceptance Criteria/Performance Data: The acceptance criteria for the gloves are based on relevant ASTM standards for physical properties, barrier integrity, and biocompatibility. The chemotherapy drug resistance is presented as "Minimum Breakthrough Detection Time in Minutes."
    • Study Design (Non-Clinical): The studies are non-clinical, primarily laboratory-based physical and chemical tests compliant with recognized international and national standards (e.g., ASTM, ISO).
    • Ground Truth: For these types of devices, the "ground truth" is typically defined by quantifiable physical and chemical properties measured against established standard methods (e.g., measuring milligrams of residual powder, breakthrough time of chemicals). There are no human experts, adjudication, or training/test sets in the context of AI for this type of device.

    Given this, I cannot extract the information required for an AI-powered device from this document. The sections you asked for (sample size for test/training, expert qualifications, MRMC studies, standalone performance, etc.) are specific to AI/software as a medical device (SaMD) assessments and are not present in a submission for examination gloves.

    However, I can summarize the acceptance criteria and reported performance for the examination gloves as provided in the document:

    Acceptance Criteria and Reported Device Performance (for Examination Gloves)

    This table summarizes the non-clinical acceptance criteria and the "result" or reported performance for the Sterile Nitrile Powder Free Examination Gloves.

    Test / CharacteristicAcceptance Criteria / Standard ReferenceReported Device Performance
    Freedom from holesMeets ASTM D5151-19 (R2023): AQL 1.5%Pass
    Residual PowderMeets ASTM D6124-06 (R2022): Average less than 2 mg/glovePass
    Dimensions (Length)Meets ASTM D6319-19 (R2023): Overall Length: ≥ 230 mm (Subject Device 1); ≥ 260 mm (Subject Device 2)Pass (Conforms to ASTM D6319 width, thickness, and length requirements)
    Dimensions (Width)Meets ASTM D6319-19 (R2023): Specific ranges for XS, S, M, L, XLPass (Conforms to ASTM D6319 width, thickness, and length requirements)
    Dimensions (Thickness)Meets ASTM D6319-19 (R2023): Palm Thickness: ≥ 0.05 mm; Finger Thickness: ≥ 0.05 mmPass (Conforms to ASTM D6319 width, thickness, and length requirements)
    Tensile Strength Before AgingMeets ASTM D6319-19 (R2023): ≥ 14 MPaPass (Conforms)
    Tensile Strength After AgingMeets ASTM D6319-19 (R2023): ≥ 14 MPaPass (Conforms)
    Ultimate Elongation Before AgingMeets ASTM D6319-19 (R2023): ≥ 500 %Pass (Conforms)
    Ultimate Elongation After AgingMeets ASTM D6319-19 (R2023): ≥ 400 %Pass (Conforms)
    Primary Skin IrritationISO 10993-23: Not an irritantUnder the conditions of the study, not an irritant.
    Dermal SensitizationISO 10993-10: Not a sensitizerUnder the conditions of the study, not a sensitizer.
    Acute Systemic ToxicityISO 10993-11: No evidence of acute systemic toxicityUnder the conditions of this study, no acute systemic toxicity.
    Chemotherapy Drug PermeationASTM D6978 (No specific 'Pass/Fail' criterion listed, rather minimum breakthrough times are reported.)Listed in table below.
    Fentanyl Citrate PermeationASTM D6978 (No specific 'Pass/Fail' criterion listed, rather minimum breakthrough times are reported.)> 240 minutes

    Chemotherapy Drug and Fentanyl Citrate Permeation Results

    Device Performance (Minimum Breakthrough Detection Time in Minutes)

    Chemotherapy Drug and ConcentrationSubject Device Result
    Carmustine (3.3 mg/ml)38.3
    Cisplatin (1.0 mg/ml)> 240
    Cyclophosphamide (20.0 mg/ml)> 240
    Dacarbazine (10.0 mg/ml)> 240
    Doxorubicin HCl (2.0 mg/ml)> 240
    Etoposide (20.0 mg/ml)> 240
    Fluorouracil (50.0 mg/ml)> 240
    Methotrexate (25.0 mg/ml)> 240
    Mitomycin C (0.5 mg/ml)> 240
    Paclitaxel (6.0 mg/ml)> 240
    Thiotepa (10.0 mg/ml)78.6
    Vincristine Sulfate (1.0 mg/ml)> 240
    Vidaza (5-Azacytidine), 25 mg/ml> 240
    Busulfan, 6 mg/ml> 240
    Carboplatin, 10 mg/ml> 240
    Docetaxel, 10 mg/ml> 240
    Epirubicin HCl, 2 mg/ml> 240
    Gemcitabine HCl, 38 mg/ml> 240
    Ifosfamide, 50 mg/ml> 240
    Irinotecan, 20 mg/ml> 240
    Mitoxantrone HCl, 2 mg/ml> 240
    Oxaliplatin, 5 mg/ml> 240
    Vinorelbine, 10 mg/ml> 240
    Fentanyl Citrate Injection (100mcg/2ml)> 240

    Since the provided document is not for an AI device, the following points cannot be addressed:

    • Sample sized used for the test set and the data provenance: Not applicable to a physical glove.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for gloves is based on physical/chemical measurements by laboratory technicians following standard protocols.
    • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For gloves, it's objective physical/chemical measurements according to ASTM/ISO standards.
    • The sample size for the training set: Not applicable (no AI training involved).
    • How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Product Code :

    LZA

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The glove is disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves were tested for use with Chemotherapy drugs, Fentanyl and Gastric acid in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.

    Device Description

    Nitrile Powder-Free Exam Gloves 2.8, Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid (light blue, dark blue)

    AI/ML Overview

    The provided document is a 510(k) premarket notification approval letter for Nitrile Powder-Free Exam Gloves. It details the device's substantial equivalence to existing predicate devices and lists the results of permeation tests for various chemotherapy drugs, opioid fentanyl citrate, and simulated gastric acid.

    However, the document does not describe a study involving an AI/software medical device requiring ground truth establishment, expert adjudication, or MRMC comparative effectiveness studies. The "acceptance criteria" discussed in this document relate to the gloves' resistance to permeation by specific chemicals, measured in "Minimum Breakthrough Detection Time."

    Therefore, based solely on the provided text, it is not possible to answer the questions related to AI/software device performance, expert ground truth, adjudication methods, or MRMC studies.

    Here's an attempt to answer the questions based on the information available in the document, acknowledging that most questions are not applicable to this type of device:


    Acceptance Criteria and Device Performance for Nitrile Powder-Free Exam Gloves

    The study presented here is a series of permeation tests conducted to determine the resistance of Nitrile Powder-Free Exam Gloves to various hazardous substances, specifically chemotherapy drugs, opioid fentanyl citrate, and simulated gastric acid. The acceptance criteria are implicit in the detailed breakthrough detection times, aiming to demonstrate the gloves' barrier properties against these substances. The tests were performed in accordance with ASTM D6978-05 standards.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document directly states the "Minimum Breakthrough Detection Time" for each substance for both the Light Blue and Dark Blue gloves. The implicit acceptance criterion would be that the gloves demonstrate a sufficiently long breakthrough time (ideally >240 minutes) for safe handling, or to clearly identify substances for which the gloves are not suitable. The reported performance is the actual measured breakthrough time.

    Light Blue Glove Performance:

    Chemotherapy Drug / SubstanceMinimum Breakthrough Detection Time (minutes)
    Bendamustine HCl, 5 mg/ml>240 min
    Bortezomib (Velcade), 1 mg/ml>240 min
    Busulfan, 6 mg/ml>240 min
    Carboplatin, 10 mg/ml>240 min
    Carmustine, 3.3 mg/ml12.6 min.
    Carfilzomib, 2 mg/ml>240 min
    Cetuximab (Erbitux), 2 mg/ml>240 min
    Chloroquine, 50mg/ml>240 min
    Cisplatin, 1 mg/ml>240 min
    Cladribine, 1.0 mg/ml>240 min
    Cyclophosphamide, 20 mg/ml>240 min
    Cyclosporin A,100mg/ml>240 min
    Cytarabine HCl, 100 mg/ml>240 min
    Cytovene, 10 mg/ml>240 min
    Dacarbazine, 10 mg/ml>240 min
    Daunorubicin HCl, 5 mg/ml>240 min
    Decitabine, 5 mg/ml>240 min
    Docetaxel, 10 mg/ml>240 min
    Doxorubicin HCl, 2 mg/ml>240 min
    Epirubicin HCl, 2 mg/ml>240 min
    Etoposide, 20 mg/ml>240 min
    Fludarabine Phosphate, 25 mg/ml>240 min
    5-Fluorouracil, 50 mg/ml>240 min
    Gemcitabine HCl, 38 mg/ml>240 min
    Idarubicin HCl, 1 mg/ml>240 min
    Ifosfamide, 50 mg/ml>240 min
    Irinotecan HCl, 20 mg/ml>240 min
    Mechlorethamine HCl, 1 mg/ml>240 min
    Melphalan HCl, 5 mg/ml>240 min
    Methotrexate, 25 mg/ml>240 min
    Mitomycin-C, 0.5 mg/ml>240 min
    Mitoxantrone HCl, 2 mg/ml>240 min
    Oxaliplatin, 2 mg/ml>240 min
    Paclitaxel, 6 mg/ml>240 min
    Retrovir, 10 mg/ml>240 min
    ThioTEPA, 10 mg/ml27.1 min.
    Trisonex, 1 mg/ml>240 min
    Vidaza (Azacitidine), 25 mg/ml>240 min
    Vincristine Sulfate, 1 mg/ml>240 min
    Fentanyl Citrate Injection, 50mcg/ml>240 min
    Gastric Acid>240 min
    Fentanyl Citrate:Xylazine HCl, 50:50>240 min

    Note: Warnings are explicitly provided for Carmustine and ThioTEPA due to extremely low permeation times.

    Dark Blue Glove Performance:

    Chemotherapy Drug / SubstanceMinimum Breakthrough Detection Time (minutes)
    Bendamustine HCl, 5 mg/ml>240 min
    Bortezomib (Velcade), 1 mg/ml>240 min
    Busulfan, 6 mg/ml>240 min
    Carboplatin, 10 mg/ml>240 min
    Carmustine, 3.3 mg/ml11.0 min.
    Carfilzomib, 2 mg/ml>240 min
    Cetuximab (Erbitux), 2 mg/ml>240 min
    Chloroquine, 50mg/ml>240 min
    Cisplatin, 1 mg/ml>240 min
    Cladribine, 1.0 mg/ml>240 min
    Cyclophosphamide, 20 mg/ml>240 min
    Cyclosporin A,100mg/ml>240 min
    Cytarabine HCl, 100 mg/ml>240 min
    Cytovene, 10 mg/ml>240 min
    Dacarbazine, 10 mg/ml>240 min
    Daunorubicin HCl, 5 mg/ml>240 min
    Decitabine, 5 mg/ml>240 min
    Docetaxel, 10 mg/ml>240 min
    Doxorubicin HCl, 2 mg/ml>240 min
    Epirubicin HCl, 2 mg/ml>240 min
    Etoposide, 20 mg/ml>240 min
    Fludarabine Phosphate, 25 mg/ml>240 min
    5-Fluorouracil, 50 mg/ml>240 min
    Gemcitabine HCl, 38 mg/ml>240 min
    Idarubicin HCl, 1 mg/ml>240 min
    Ifosfamide, 50 mg/ml>240 min
    Irinotecan HCl, 20 mg/ml>240 min
    Mechlorethamine HCl, 1 mg/ml>240 min
    Melphalan HCl, 5 mg/ml>240 min
    Methotrexate, 25 mg/ml>240 min
    Mitomycin-C, 0.5 mg/ml>240 min
    Mitoxantrone HCl, 2 mg/ml>240 min
    Oxaliplatin, 2 mg/ml>240 min
    Paclitaxel, 6 mg/ml>240 min
    Retrovir, 10 mg/ml>240 min
    ThioTEPA, 10 mg/ml26.0 min.
    Trisonex, 1 mg/ml>240 min
    Vidaza (Azacitidine), 25 mg/ml>240 min
    Vincristine Sulfate, 1 mg/ml>240 min
    Fentanyl Citrate Injection, 50mcg/ml>240 min
    Gastric Acid>240 min
    Fentanyl Citrate:Xylazine HCl, 50:50>240 min

    Note: Warnings are explicitly provided for Carmustine and ThioTEPA due to extremely low permeation times.

    2. Sample Size and Data Provenance for the Test Set:
    The document does not explicitly state the sample size (number of gloves tested for each substance) used for these permeation tests. It only lists the test results. The data provenance is stated as being tested "in accordance with ASTM D6978-05 standards," which implies a standardized laboratory testing setting. There is no information regarding the country of origin of the data or whether it was retrospective or prospective.

    3. Number of Experts and Qualifications for Ground Truth:
    This section is not applicable to this type of device and study. The "ground truth" here is the physical measurement of chemical permeation, which is determined by laboratory instrumentation and protocols defined by the ASTM standard, not by human expert interpretation or consensus.

    4. Adjudication Method for the Test Set:
    This section is not applicable. Adjudication methods like "2+1" or "3+1" are used in studies involving human readers/experts evaluating data (e.g., medical images) to establish a consensus ground truth. In this study, the "ground truth" is a direct physical measurement.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
    This section is not applicable. MRMC studies are performed for AI/software devices where human readers interpret medical data, often comparing human performance with and without AI assistance. This document describes chemical permeation tests of gloves, not a diagnostic or prognostic medical device involving human interpretation.

    6. Standalone (Algorithm Only) Performance:
    This section is not applicable. This refers to the performance of an AI algorithm without human intervention. The device in question is a physical glove, not an algorithm.

    7. Type of Ground Truth Used:
    The "ground truth" for this study is a physical measurement of chemical permeation through the glove material, determined by laboratory testing protocols as defined by the ASTM D6978-05 standards. It is not derived from expert consensus, pathology, or outcomes data.

    8. Sample Size for the Training Set:
    This section is not applicable. This study does not involve a training set as it's not an AI/machine learning model development context.

    9. How the Ground Truth for the Training Set was Established:
    This section is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Product Code :

    LZA

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on examiner's hand to prevent contamination between patient and examiner. In addition, the glove was evaluated using ASTM D6978 for permeation of chemotherapy drugs and fentanyl citrate. The glove is single use and non-sterile.

    The following chemotherapy drugs have been tested on the glove:

    Chemotherapy DrugsConcentrationBreakthrough Detection Time
    Carmustine*3.3 mg/mL23.2 min
    Cisplatin1.0 mg/mL>240 min
    Cyclophosphamide20.0 mg/mL>240 min
    Dacarbazine10.0 mg/mL>240 min
    Doxorubicin HCL2.0 mg/mL>240 min
    Etoposide20.0 mg/mL>240 min
    Fluorouracil50.0 mg/mL>240 min
    Methotrexate25.0 mg/mL>240 min
    Paclitaxel6.0 mg/mL>240 min
    ThioTepa*10.0 mg/mL13.5 min
    Mitomycin C0.5 mg/mL>240 min
    Vincristine Sulfate1.0 mg/mL>240 min
    FentanylConcentrationBreakthrough Detection Time
    Fentanyl Citrate100.0 mcg/2mL>240 min

    *Please note that the following drugs have low permeation time.

    1. Carmustine
    2. ThioTepa

    Warning: Do not use with Carmustine & ThioTepa

    Device Description

    The proposed Powder-Free Polychloroprene Examination Glove (Green Color) Tested for Use with Chemotherapy Drugs and Fentanyl is a disposable device intended for over-the-counter use and is provided powder-free and non-sterile. These patient examination gloves are formulated using Polychloroprene. The gloves are not made with natural rubber latex. General specifications of the glove are as below:

    1. Overall Length: 240 mm minimum
    2. Width: 95 ± 5 mm minimum (for medium glove)
    3. Palm Thickness: 0.05 mm minimum
    4. Finger Thickness: 0.05 mm minimum
    5. Cuff Thickness: 0.05 mm minimum
    6. Tensile Strength
      • a. Before Aging: 14 MPa minimum
      • b. After Aging: 14 MPa minimum
    7. Ultimate Elongation
      • a. Before Aging: 500 % minimum
      • b. After Aging: 400 % minimum
    8. Pinhole AQL: 2.5
    AI/ML Overview

    The provided FDA 510(k) clearance letter and associated documentation pertain to a Powder-Free Polychloroprene Examination Glove. This is a physical medical device, not an AI/software device. Therefore, the questions related to AI/software performance criteria (such as multi-reader multi-case studies, human-in-the-loop performance, training/test set ground truth, and expert adjudication) are not applicable to this documentation.

    The acceptance criteria and performance data provided in the document relate to the physical and chemical properties of the gloves, and their resistance to permeation by chemotherapy drugs and fentanyl.

    Here's an analysis of the acceptance criteria and study that proves the device meets them, based on the provided text, focusing on the relevant physical device aspects:

    Acceptance Criteria and Reported Device Performance

    1. Table of Acceptance Criteria and the Reported Device Performance

    CharacteristicStandard / RequirementDevice Performance (Proposed Device K250342)
    DimensionsASTM D6977-19
    Length (XS)≥ 220 mm≥ 251 mm
    Length (S)≥ 220 mm≥ 250 mm
    Length (M)≥ 230 mm≥ 251 mm
    Length (L)≥ 230 mm≥ 248 mm
    Length (XL)≥ 230 mm≥ 250 mm
    Length (XXL)≥ 230 mm≥ 250 mm
    Width (XS)70 ± 10 mm79 - 80 mm
    Width (S)80 ± 10 mm86 - 88 mm
    Width (M)95 ± 10 mm98 - 100 mm
    Width (L)110 ± 10 mm111 - 113 mm
    Width (XL)120 ± 10 mm115 - 118 mm
    Width (XXL)130 ± 10 mm127 - 129 mm
    Thickness (Finger)≥ 0.05 mm≥ 0.11 mm
    Thickness (Palm)≥ 0.05 mm≥ 0.08 mm
    Thickness (Cuff)≥ 0.05 mm≥ 0.06 mm
    Freedom from HolesASTM D6977-19, 21 CFR 800.20 (AQL 2.5)Passed (for all sizes XS-XXL)
    Physical PropertiesASTM D6977-19 (ASTM D412-16, D573)
    Before Aging (Tensile)≥ 14 MPa16.83 - 33.4 MPa (depending on size)
    Before Aging (Elongation)≥ 500 %521 - 596 % (depending on size)
    After Aging (Tensile)≥ 14 MPa23.8 - 40.2 MPa (depending on size)
    After Aging (Elongation)≥ 400 %428 - 462 % (depending on size)
    Residual PowderASTM D6977-19, ASTM D6124-06 (≤ 2.0 mg/glove)0.22 - 0.32 mg/glove (depending on size)
    Biocompatibility
    Skin IrritationISO 10993-10 (Score ≤ 0.4)Not considered an irritant
    Skin SensitizationISO 10993-10 (Score 240 min
    Cyclophosphamide->240 min
    Dacarbazine->240 min
    Doxorubicin HCL->240 min
    Etoposide->240 min
    Fluorouracil->240 min
    Methotrexate->240 min
    Paclitaxel->240 min
    ThioTepa-13.5 min*
    Mitomycin C->240 min
    Vincristine Sulfate->240 min
    Fentanyl PermeationASTM D6978-05 (Minimum Breakthrough Detection Time)
    Fentanyl Citrate->240 min

    *Note: The document explicitly warns that Carmustine and ThioTepa have low permeation times and should not be used with the glove. This is a crucial labeling instruction based on the test results.

    2. Sample Size Used for the Test Set and Data Provenance

    • The document does not explicitly state the specific sample sizes used for each physical, chemical, and permeation test. However, tests are typically performed on a statistically relevant number of samples to ensure representativeness and robust results according to the specified ASTM and ISO standards.
    • Data Provenance: The tests appear to have been conducted by Encompass Industries Sdn. Bhd. or a contracted lab on their behalf. The company is located in Kertih Bio-Polymer Park, Kemaman, Terengganu, Malaysia. The data is prospective in the sense that the tests were specifically performed on the proposed device for the purpose of this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • This question is not applicable as the document describes a physical medical device (examination glove) and its material/permeation performance, not a software/AI device that requires expert human interpretation of data for ground truth establishment. The "ground truth" for these tests is based on objective, standardized laboratory measurements (e.g., precise measurements of length, thickness, tensile strength; chemical analysis for powder residue; standardized biological assays for biocompatibility; and analytical chemistry for drug permeation breakthrough times).

    4. Adjudication Method for the Test Set

    • This question is not applicable for the same reason as point 3. Testing of physical device characteristics and chemical permeation is typically performed according to established laboratory protocols with quantitative, measurable endpoints, not through a process of expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This question is not applicable as the document describes a physical medical device (examination glove), not an AI-assisted diagnostic or therapeutic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • This question is not applicable as the document describes a physical medical device (examination glove), not an AI algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" for the performance of the examination gloves is based on objective, standardized laboratory measurements and adherence to established international and national standards (e.g., ASTM D6977-19, ASTM D6319-19, ASTM D5151-19, ASTM D412-16, ASTM D573, ASTM D6124-06, ASTM D6978-05, ISO 10993-10, ISO 10993-11, 21 CFR 800.20).
      • Examples include:
        • Physical measurements: Direct measurement of glove dimensions (length, width, thickness).
        • Mechanical properties: Tensile strength and elongation determined by standard material testing methods.
        • Integrity: Freedom from holes determined by water leak testing (ASTM D5151-19).
        • Chemical analysis: Measurement of residual powder.
        • Biocompatibility: In-vivo (skin irritation, sensitization, acute systemic toxicity) tests using animal models, following ISO standards.
        • Permeation: Breakthrough detection time of specific chemotherapy drugs and fentanyl using analytical methods (ASTM D6978-05).

    8. The Sample Size for the Training Set

    • This question is not applicable as the document describes a physical medical device, which does not involve a "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    • This question is not applicable for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250496
    Date Cleared
    2025-07-07

    (137 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    LZA

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder Free Polyurethane Examination Gloves (Green, Blue, Pink, White) is a disposable device intended for medical purpose that is worn on the examiners hand or finger to prevent contamination between patient and examiner.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for Powder Free Polyurethane Examination Gloves. It is not a document describing an AI/ML-driven medical device or software. Therefore, the information requested in your prompt (acceptance criteria, study details for AI/ML performance, ground truth establishment, etc.) cannot be extracted from this letter.

    This letter confirms that the gloves are "substantially equivalent" to legally marketed predicate devices, meaning they meet the safety and effectiveness standards for their intended use as barrier protection. The "study" here refers to the documentation and evidence the manufacturer provided to demonstrate this substantial equivalence based on established performance characteristics for medical gloves (e.g., barrier integrity, tensile strength, biocompatibility), not an AI/ML efficacy study.

    To answer your prompt, I would need a document related to the clearance of an AI/ML medical device, which typically includes detailed performance studies and acceptance criteria related to its algorithmic accuracy and clinical utility.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Product Code :

    LZA

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Halyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Pairs or Sterile Singles, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes:

    • Azacitidine (25 mg/ml)
    • Bendamustine HCl (5 mg/ml)
    • Bleomycin Sulfate (15 mg/ml)
    • Bortezomib (1 mg/ml)
    • Busulfan (6 mg/ml)
    • Capecitabine (26 mg/ml)
    • Carboplatin (10 mg/ml)
    • Carfilzomib (2 mg/ml)
    • Cetuximab (2 mg/ml)
    • Cisplatin (1 mg/ml)
    • Cladribine (1 mg/ml)
    • Cyclophosphamide (20 mg/ml)
    • Cytarabine HCl (100 mg/ml)
    • Dacarbazine (10 mg/ml)
    • Dactinomycin (0.5 mg/ml)
    • Daunorubicin HCl (5 mg/ml)
    • Decitabine (5 mg/ml)
    • Docetaxel (10 mg/ml)
    • Doxorubicin HCl (2 mg/ml)
    • Epirubicin HCl (2 mg/ml)
    • Etoposide (20 mg/ml)
    • Fludarabine Phosphate (25 mg/ml)
    • Fluorouracil (50 mg/ml)
    • Fulvestrant (50 mg/ml)
    • Gemcitabine HCl (38 mg/ml)
    • Idarubicin HCl (1 mg/ml)
    • Ifosfamide (50 mg/ml)
    • Irinotecan HCl (20 mg/ml)
    • Leuprolide Acetate (5 mg/ml)
    • Mechlorethamine HCl (1 mg/ml)
    • Melphalan HCl (5 mg/ml)
    • Methotrexate (25 mg/ml)
    • Mitomycin C (0.5 mg/ml)
    • Mitoxantrone HCl (2 mg/ml)
    • Oxaliplatin (5 mg/ml)
    • Paclitaxel (6 mg/ml)
    • Pemetrexed (25 mg/ml)
    • Raltitrexed (0.5 mg/ml)
    • Rituximab (10 mg/ml)
    • Temsirolimus (25 mg/ml)
    • Topotecan HCl (1 mg/ml)
    • Trisenox (Arsenic Trioxide) (1 mg/ml)
    • Vinblastine Sulfate (1 mg/ml)
    • Vincristine (1 mg/ml)
    • Vinorelbine Tartrate (10 mg/ml)

    The following chemotherapy drugs and concentration showed breakthrough detected in less than 100 minutes:

    • Carmustine (3.3 mg/ml) No breakthrough up to 44.5 minutes.
    • Thiotepa (10 mg/ml) No breakthrough up to 99.1 minutes.
    • Warning- Not for use with Carmustine and ThioTEPA

    The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes:

    • Fentanyl Citrate Injection (100 mcg/2 ml)
    • Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution

    The following hazardous drugs and concentration had NO breakthrough detected up to 240 minutes:

    • Chloroquine (50 mg/ml)
    • Cyclosporin A (100 mg/ml)
    • Cytovene (10 mg/ml)
    • Retrovir (10 mg/ml)
    • Triclosan (2 mg/ml)
    • Zoledronic Acid (0.8 mg/ml)
    Device Description

    The Halyard Purple Nitrile* Powder‐Free Exam Gloves, Sterile Pairs or Sterile Singles, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid are disposable, 9.5" purple‐colored, chlorinated, nitrile, powder‐free, textured fingertip, ambidextrous, sterile patient examination gloves.

    AI/ML Overview

    The provided text is a 510(k) clearance letter and summary for medical examination gloves, not an AI/ML device. Therefore, it does not contain information related to AI/ML specific acceptance criteria or study details such as:

    • Sample size used for the test set and data provenance (e.g., country of origin, retrospective/prospective).
    • Number of experts used to establish ground truth and their qualifications.
    • Adjudication method for the test set.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study.
    • Standalone (algorithm-only) performance.
    • Type of ground truth used (expert consensus, pathology, outcomes data).
    • Sample size for the training set.
    • How ground truth for the training set was established.

    Instead, the document focuses on the physical and chemical properties of the gloves and their resistance to various chemicals, using established ASTM and ISO standards for medical gloves.

    However, I can extract the acceptance criteria and reported device performance for these medical gloves based on the provided document.


    Analysis of Acceptance Criteria and Device Performance for Medical Gloves

    The document details the non-clinical testing performed on the Halyard Purple Nitrile* Powder-Free Exam Gloves to demonstrate their safety and effectiveness. The acceptance criteria and the results of these tests are clearly outlined.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestStandardAcceptance CriteriaReported Device Performance
    DimensionsASTM D 6319Length ≥230 mm
    Palm Width Size Small: 70 - 90 mm
    Med: 85–105 mm
    Large: 100 - 120 mm
    Finger thickness ≥0.05 mm
    Palm thickness ≥0.05 mm
    Cuff thickness ≥0.05 mmMeets requirements
    Physical Properties (Tensile Strength & Elongation)ASTM D 6319AQL 4.0
    Before Aging:
    Tensile Strength: ≥14 MPa
    Ultimate elongation: ≥500%
    After Aging:
    Tensile Strength: ≥14 MPa
    Ultimate elongation: ≥400%Meets requirements. The physical dimensions and properties (tensile strength before and after aging, and elongation before and after aging) met the requirements of the standard.
    Freedom from PinholesASTM D 6319, ASTM D 5151AQL 2.5%
    No leakageMeets requirements. Testing of the subject device shows it meets the 2.5% AQL requirement in the standards for leakage.
    SterilityANSI/AAMI/ISO 1113710⁻⁶10⁻⁶
    Powder FreeASTM D 6124, ASTM D 6319≤ 2 mg / gloveMeets requirements. Residual powder on the subject device is an average of 0.4 mg/glove, which is within the powder-free limit of
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Product Code :

    LZA

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    These gloves were tested for use with chemotherapy drugs, fentanyl citrate and gastric acid as per ASTM D6978-05 (2023) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:

    The following drugs and concentration had NO breakthrough detected up to 240 minutes:

    • Bendamustine HCl (5 mg/mL)
    • Bleomycin Sulfate (15 mg/mL)
    • Busulfan (6 mg/mL)
    • Carboplatin (10 mg/mL)
    • Carfilzomib (2 mg/mL)
    • Cetuximab (Erbitux) (2 mg/mL)
    • Chloroquine (50 mg/mL)
    • Cisplatin (1 mg/mL)
    • Cladribine (1 mg/mL)
    • Cyclophosphamide (20 mg/mL)
    • Cyclosporin A (100 mg/mL)
    • Cytarabine HCl (100 mg/mL)
    • Cytovene (10 mg/mL)
    • Dacarbazine (10 mg/mL)
    • Daunorubicin HCl (5 mg/mL)
    • Decitabine (5 mg/mL)
    • Docetaxel (10 mg/mL)
    • Doxorubicin HCl (2 mg/mL)
    • Epirubicin HCl (2 mg/mL)
    • Etoposide (20 mg/mL)
    • Fludarabine Phosphate (25 mg/mL)
    • Fluorouracil (50 mg/mL)
    • Fulvestrant (50 mg/mL)
    • Gemcitabine HCl (38 mg/mL)
    • Idarubicin HCl (1 mg/mL)
    • Ifosfamide (50 mg/mL)
    • Irinotecan HCl (20 mg/mL)
    • Mechlorethamine HCl (1 mg/mL)
    • Melphalan HCl (5 mg/mL)
    • MESNA (100 mg/mL)
    • Methotrexate (25 mg/mL)
    • Mitomycin C (0.5 mg/mL)
    • Mitoxantrone HCl (2 mg/mL)
    • Oxaliplatin (2 mg/mL)
    • Paclitaxel (6 mg/mL)
    • Pemetrexed (25 mg/mL)
    • Propofol (10 mg/mL)
    • Raltitrexed (0.5 mg/mL)
    • Retrovir (10 mg/mL)
    • Rituximab (10 mg/mL)
    • Temsirolimus (25 mg/mL)
    • Topotecan HCl (1 mg/mL)
    • Triclosan (1 mg/mL)
    • Trisenox (Arsenic Trioxide) (1 mg/mL)
    • Velcade (Bortezomib) (1 mg/mL)
    • Vidaza (Azacitidine) (25 mg/mL)
    • Vinblastine Sulfate (1 mg/mL)
    • Vincristine Sulfate (1 mg/mL)
    • Vinorelbine Tartrate (10 mg/mL)
    • Zoledronic Acid (0.8 mg/mL)

    The following chemotherapy drugs and concentration showed breakthrough detected in less than 60 minutes:

    • Carmustine (3.3 mg/mL), breakthrough detected at 55.5 minutes
    • Thiotepa (10 mg/mL), breakthrough detected at 50.8 minutes

    Warning: Not recommended for use with Carmustine and Thiotepa

    The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes:

    • Fentanyl Citrate Injection (100mcg/2mL)
    • Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution
    Device Description

    The subject device, Nitrile Powder Free Examination Gloves with Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid is a single use, disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This product demonstrated reduced potential for sensitizing users to chemical additives, supported by a negative skin sensitization test (Modified Draize-95 Test) and tested for use with chemotherapy drugs, fentanyl citrate and gastric acid.

    The gloves are made of nitrile rubber, powder free, ambidextrous with beaded cuff. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that facilitates the user in donning the gloves without using lubricant and donning powder on the glove surface. These gloves are offered in six sizes (XS, S, M, L, XL, XXL), and supplied in non-sterile state.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary concern a medical device, "Nitrile Powder Free Examination Gloves with Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid." This document details the product's attributes, testing, and comparison to a predicate device for regulatory clearance.

    However, the questions you've asked (acceptance criteria, study details, sample sizes, expert qualifications, adjudication, MRMC studies, standalone performance, ground truth, and training set details) are highly specific to the performance of an AI/ML medical device rather than a physical product like examination gloves. The provided document describes the physical and chemical performance of the gloves (e.g., resistance to permeation by chemotherapy drugs, physical properties, biocompatibility, and skin sensitization), not the performance of an artificial intelligence or machine learning algorithm.

    Therefore, many of your questions are not applicable to the information contained within this 510(k) clearance for medical gloves. I will answer the applicable questions based on the provided text, and explicitly state when a question is not relevant to this type of device.


    Acceptance Criteria and Device Performance for Nitrile Examination Gloves

    The "acceptance criteria" for these gloves are primarily physical, chemical, and biological performance standards, as demonstrated through various ASTM and ISO tests. The "reported device performance" indicates whether the gloves met these pre-defined standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Requirement)Reported Device Performance (Result)
    Physical Properties (ASTM D6319-19):
    - Tensile strength (Before Aging): min 14 MPaPass
    - Ultimate elongation (Before Aging): min 500%Pass
    - Tensile strength (After Aging): min 14 MPaPass
    - Ultimate elongation (After Aging): min 400%Pass
    Dimensions (ASTM D6319-19):
    - Length: XS: min 220mm, S: min 220mm, M: min 230mm, L: min 230mm, XL: min 230mm, XXL: min 230mmPass
    - Palm Width: XS: 70 ± 10mm, S: 80 ± 10mm, M: 95 ± 10mm, L: 110 ± 10mm, XL: 120 ± 10mm, XXL: 130 ± 10mmPass
    - Thickness (Finger): min 0.05mmPass
    - Thickness (Palm): min 0.05mmPass
    Watertight Test (Freedom from Holes - ASTM D5151-19):Pass Inspection Level G1, AQL 2.5
    Residual Powder Content (ASTM D6124-06):Residual powder
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Product Code :

    LZA

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 (2023), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

    Tested Chemotherapy DrugsConcentrationMinimum Breakthrough Detection Time in minutes
    *Carmustine3.3mg/ml22.5
    Cisplatin1 mg/ml>240
    Cyclophosphamide20 mg/ml>240
    Dacarbazine10 mg/ml>240
    Doxorubicin, HCl2 mg/ml>240
    Etoposide20 mg/ml>240
    Fluorouracil50 mg/ml>240
    Methotrexate25 mg/ml>240
    Mitomycin0.5 mg/ml>240
    Oxaliplatin5 mg/ml>240
    Paclitaxel6 mg/ml>240
    *Thiotepa10 mg/ml34.5
    Vincristine1 mg/ml>240
    Tested Opioid DrugsConcentrationMinimum Breakthrough Detection Time in minutes
    Fentanyl Citrate Injection100mcg/2ml>240

    Warning: Do not use with Carmustine and Thiotepa.

    Note:
    Please note that the following drugs have low permeation times:
    (1) Carmustine - 22.5 minutes
    (2) Thiotepa - 34.5 minutes.

    Device Description

    The Nitrile Powder Free Examination Glove with Tremella Fuciformis Extract, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate - Blue are Class 1, Polymer Patient Examination Gloves and Chemotherapy Gloves.

    The glove is made of Nitrile Butadiene Rubber. They are ambidextrous with beaded cuffs, fingertip texture and come in different sizes - Small, Medium, Large and Extra Large. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs and fentanyl citrate as per ASTM D6978-05(2023). The gloves are single use and provided non-sterile.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to Nitrile Powder Free Examination Gloves, not a software-based medical device or AI system. Therefore, the questions regarding ground truth, expert adjudication, MRMC studies, standalone performance, training sets, and data provenance are not applicable to this document.

    However, based on the non-clinical performance testing detailed in the submission, we can address the acceptance criteria and demonstrated performance for the physical device.

    Acceptance Criteria and Reported Device Performance for Nitrile Powder Free Examination Gloves K243694

    The acceptance criteria for this medical device are based on various physical, chemical, and biological tests, primarily adhering to ASTM and ISO standards for examination gloves and their specific claims (chemotherapy and fentanyl resistance).

    1. Table of Acceptance Criteria and the Reported Device Performance

    Test MethodAcceptance CriteriaReported Device Performance
    Physical Dimensions (ASTM D6319)Length: S: ≥220mm, M/L/XL: ≥230mm
    Width: S: 80±10 mm, M: 95±10 mm, L: 110±10 mm, XL: 120±10 mm
    Thickness: Palm: ≥0.05 mm, Finger: ≥0.05 mmLength: S: ≥220mm (Pass), M/L/XL: ≥230mm (Pass)
    Width: S: 86-88/Pass, M: 98-99/Pass, L: 105-110/Pass, XL: 117-119/Pass
    Thickness: Palm: S-XL: 0.06mm/Pass, Finger: S-L: 0.08mm/Pass, XL: 0.09mm/Pass
    Watertightness Test for Detection of Holes (ASTM D5151, ASTM D6319)Meet AQL 2.5 with G1Meet AQL 2.5 with G1 (Pass)
    Powder Content (ASTM D6124, ASTM D6319)Meet the requirement of ASTM D 6124-06 (≤2.0mg/glove)S: 0.63mg/glove, M: 1.02mg/glove, L: 0.28mg/glove, XL: 0.38mg/glove (Pass)
    Physical Properties (Tensile Strength & Elongation) (ASTM D412, ASTM D6319)Tensile Strength: before aging ≥14 MPa, after aging ≥14 MPa
    Ultimate Elongation: before aging ≥500 %, after aging ≥400%Tensile Strength: Before aging: Pass, After aging: Pass
    Ultimate Elongation: Before aging: Pass, After aging: Pass
    Chemotherapy drug and Fentanyl Citrate claim (ASTM D6978)The subject device was tested for use with chemotherapy drugs and Fentanyl Citrate. Implicit acceptance is that breakthrough times meet or exceed safe working periods, and any low permeation drugs are adequately warned.No breakthrough detected up to 240 minutes for: Cisplatin, Cyclophosphamide, Dacarbazine, Doxorubicin HCl, Etoposide, Fluorouracil, Methotrexate, Mitomycin, Oxaliplatin, Paclitaxel, Vincristine, Fentanyl Citrate Injection.
    Low permeation times: Carmustine (22.5 minutes), ThioTEPA (34.5 minutes) - Warning issued for these two.
    Cytotoxicity (ISO 10993-5)Non-cytotoxicUnder conditions of the study, device extract is cytotoxic (This seems to be a clerical error in the document as it proceeds to state 'Non-cytotoxic' in subsequent summary sections of the same document for the same test, and a device would not be cleared if cytotoxic. Given the context of clearance, it likely passed or the initial statement is a misphrasing where the high concentrations showed cytotoxicity but the diluted, in-use concentrations did not, aligning with the "Under the condition of this test, exhibited cytotoxicity reactivity from 100.0% extract concentration to 25.0% extract concentrations and exhibited no cytotoxicity reactivity from 12.5% extract concentration to 3.125% extract concentrations" within the Summary of Technological Characteristic section. The overall conclusion for biocompatibility in Table 1 Part 4 of 8 "Similar" for this test, and the clearance, suggests it was deemed acceptable).
    Irritation (ISO 10993-23)Non-irritatingUnder the conditions of the study, device is not an irritant. / Pass
    Sensitization (ISO 10993-10)Non-sensitizingUnder the conditions of the study, device is not a sensitizer. / Pass
    Acute Systemic Toxicity (ISO 10993-11)Non-acute systemic toxicityUnder the conditions of the study, device did not show acute systemic toxicity in vivo. / Pass
    Bioburden Determination (ISO 11737-1:2018)≤ 200 CFU/device80-119 CFU/glove (Pass)
    Material-Mediated Pyrogenicity (USP )Non-pyrogenicUnder the conditions of the study, non-pyrogenic (Pass)

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific "sample size" for each individual test reported (e.g., how many gloves were tested for watertightness, how many for a specific drug permeation). However, these tests are typically performed on a statistically representative sample of manufactured gloves as per the respective ASTM/ISO standards.

    • Data Provenance: The testing was conducted as part of the 510(k) submission process for a device manufactured by Mah Sing Healthcare Sdn. Bhd. in Malaysia. The data would therefore be prospective for the purpose of demonstrating the device's adherence to the standards for this specific submission. The data is generated from laboratory testing of the physical product.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This relates to physical/chemical/biological testing of a medical glove, not the establishment of ground truth for an AI or imaging diagnostic system. The "ground truth" here is the objective measurement against established physical/chemical/biological standards.

    4. Adjudication method for the test set:

    Not applicable. Adjudication methods (like 2+1, 3+1) are relevant to human review of diagnostic medical data, often for establishing a consensus "ground truth" when a definitive objective truth is unavailable (e.g., interpreting medical images). For physical device testing, the results are typically objectively measured and do not require expert adjudication in this manner.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    No. MRMC studies are used to evaluate the impact of a new diagnostic aid (like AI) on physician performance. This is for a physical medical device (examination gloves), not an AI system or diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This question pertains to AI algorithm performance.

    7. The type of ground truth used:

    Objective Experimental Measurement and Compliance with Standardized Protocols. For each test (e.g., tensile strength, chemical permeation, cytotoxicity), the "ground truth" is the quantitative measurement obtained through standardized laboratory procedures (e.g., ASTM D6319 for physical properties, ASTM D6978 for chemical permeation, ISO 10993 series for biocompatibility) and compared against the defined acceptance criteria within those standards.

    8. The sample size for the training set:

    Not applicable. This question pertains to AI/machine learning models. Examination gloves do not have a "training set."

    9. How the ground truth for the training set was established:

    Not applicable. Again, this relates to AI/machine learning.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 115