K Number
K142185

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-11-19

(103 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EXPEDIUM Verse System is intended to provide immobilization of spinal segments in sketally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine.

The EXPEDIUM Verse System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used in a posterior percutaneous approach with MIS instrumentation, the EXPEDIUM Verse System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM Verse System metallic implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. The EXPEDIUM Verse system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Device Description

The EXPEDIUM® Verse System is designed to provide intraoperative polyaxial to monoaxial conversion. It facilitates easier rod capture and provides a powerful and precise reduction mechanism. Verse is a reduced profile thoracolumbar implant for use for with wide range of patient statures. EXPEDIUM Verse is a self-contained, efficient, and intuitive instrument system that is compatible with EXPEDIUM 5.5 rods, hooks and mono screws to enhance versatility.

AI/ML Overview

The provided document is a 510(k) premarket notification for the DePuy Synthes Spine EXPEDIUM® Verse Spine System. It describes the device, its intended use, and comparability to predicate devices, but does not contain information related to software or AI algorithms, nor does it provide a study that directly assesses the performance of an AI device against specific acceptance criteria.

Therefore, I cannot extract the requested information regarding acceptance criteria, device performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth for AI-related studies.

The document discusses mechanical testing for the physical medical device (spinal implant system), not an AI or software device. The "Non-clinical Test Summary" and "Conclusion" sections refer to these mechanical tests.

To answer your request, I would need a document detailing the performance of an AI-based device, including its acceptance criteria and a study demonstrating its performance.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other, representing the department's focus on people and their well-being.

November 19, 2014

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

DePuy Synthes Spine, a Johnson & Johnson Company Catherine Kilshaw, M.S. Senior Regulatory Affairs Associate 325 Paramount Drive Raynham, Massachusetts 02767

Re: K142185

Trade/Device Name: DePuy Synthes Spine EXPEDIUM® Verse Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNI, MNH, KWP, KWQ Dated: August 21, 2014 Received: August 22, 2014

Dear Ms. Kilshaw:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142185

Device Name

DePuy Synthes Spine EXPEDIUM® Verse Spine System

Indications for Use (Describe)

The EXPEDIUM Verse System is intended to provide immobilization of spinal segments in sketally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine.

The EXPEDIUM Verse System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used in a posterior percutaneous approach with MIS instrumentation, the EXPEDIUM Verse System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM Verse System metallic implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. The EXPEDIUM Verse system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."


{3}------------------------------------------------

510(k) Summary

Date Prepared: November 18, 2014

Submitter:DePuy Synthes Spine325 Paramount DriveRaynham, MA 02767
Contact Person:Catherine KilshawSr. Regulatory Affairs AssociateTelephone: (508) 880-8412Fax: (508) 828-3797
Trade Name:DePuy Synthes Spine EXPEDIUM® Verse Spine System
Device Class:III
Product Code:NKB OSH KWP KWQ MNH MNI
Common Name:Pedicle Screw Spine System
Classification Name:Pedicle Screw Spinal System
Regulation Number:21 CFR 888.3070
PrimaryPredicate Device:K111136 EXPEDIUM® Spine System, VIPER® Spine System, VIPER® 2Spine System
Additional PredicateDevices:K033901 Merlin Spine System (later re-named EXPEDIUM Spine System)K070387 EXPEDIUM Spine System
Reference Devices:K091994 MOUNTAINEER LAMINOPLASTY SYSTEMK124004 NAVIGATED CD HORIZON SOLERA SCREWDRIVERS, TAPS, ILIAC TAPS,LEGACY TAPS
Device Description:The EXPEDIUM® Verse System is designed to provide intraoperative polyaxial tomonoaxial conversion. It facilitates easier rod capture and provides a powerfuland precise reduction mechanism. Verse is a reduced profile thoracolumbar

{4}------------------------------------------------

implant for use for with wide range of patient statures. EXPEDIUM Verse is a self-contained, efficient, and intuitive instrument system that is compatible with EXPEDIUM 5.5 rods, hooks and mono screws to enhance versatility.

Indications: The EXPEDIUM Verse System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The EXPEDIUM Verse System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used in a posterior percutaneous approach with MIS instrumentation, the EXPEDIUM Verse System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM Verse System metallic implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The EXPEDIUM Verse system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Materials:Titanium alloy Ti64AIV, Cobalt chromium alloy CoCr
Comparison to
Predicate Device:The substantial equivalence of the subject device to the predicate identifiedabove is based upon the similarity of intended use, design (fundamentalscientific technology) materials, performance, sterility and biocompatibility.
Non-clinical Test
Summary:Mechanical testing (using the ASTM 1717 Standard Test Methods for SpinalImplant Constructs in a Vertebrectomy Model standard) was provided in order

{5}------------------------------------------------

to provide data to support a substantial equivalence determination. These testswere performed to characterize the properties and functionality of the screw, aswell as to allow comparison with established acceptance criteria.
The following mechanical tests were conducted on the screw:
●Static Compression
Dynamic Compression●
●Static Torsion
Clinical Test
Summary:N/A
Conclusion:Based upon the predicate comparison, the intended use, similar technologicalcharacteristics and the results of the static compression testing, dynamiccompression testing and static torsion testing, the proposed device issubstantially equivalent to the predicate device.

N/A