K Number
K142580

Validate with FDA (Live)

Date Cleared
2014-12-01

(77 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Chemtrue® Multi-Panel DOA Dip Card Tests:
The Chemtrue® Multi-Panel DOA Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of Amphetamine, Cocaine, Marijuana, Morphine, Methamphetamine, Phencyclidine, Benzodiazepines, Ecstasy, Methadone and Oxycodone drugs in human urine. The test cut-off concentrations and the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMP / METd-Methamphetamine1000
MorphineMORMorphine300
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ9-THC9-COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue® Multi-Panel DOA Dip Card Tests panel can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the prefered confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates and Oxycodone. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

Chemtrue® Multi-Panel DOA Dip Card with OPI 2000 Tests:
The Chemtrue® Multi-Panel DOA Dip Card with OPI 2000 Tests are rapid lateral flow immunoassays for the qualitative detection of Amphetamine, Cocaine, Marijuana, Opiates 2000, Methamphetamine, Phencyclidine, Benzodiazepines. Barbiturates, Ecstasy, Methadone and Oxycodone drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMP /METd-Methamphetamine1000
OpiatesOPIMorphine2000
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ9-THC9-COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemrue® Multi-Panel DOA Dip Card with OP1 2000 Tests can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GCMS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates and Oxycodone. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

Device Description

The Chemtrue® DOA Tests are colloidal gold based lateral flow immunoassays for the rapid, qualitative detection of drugs of abuse in human urine. The tests are single-use, in vitro diagnostic devices, which come in the formats of dip card, cup, or cassette, as indicated by the test name.

AI/ML Overview

The provided document, K142580, describes the substantial equivalence determination for the Chemtrue® Multi-Panel DOA Dip Card Tests. It references previous 510(k) clearances (K102203, K111322, K121339, and K123080) for the individual drug analytes, stating that the test strips, sample matrix, test format, and cut-off concentrations for these analytes are identical to those previously cleared. Therefore, the performance characteristics and studies for the individual analyte tests are found in those referenced 510(k) submissions.

This document focuses on the new combination of these previously cleared test strips into multi-panel kits and confirms that verification and validation activities were conducted for this combination.

Here's an analysis of the acceptance criteria and study information provided (or inferred from the context of a 510(k) summary for a diagnostic device):


1. Table of Acceptance Criteria and Reported Device Performance

Since this 510(k) references prior clearances for the individual analytes, the general acceptance criteria for such qualitative tests involve achieving a certain level of agreement (e.g., sensitivity and specificity) with a validated reference method (like GC/MS or LC/MS) at and around the cut-off concentration. The document states that the performance characteristics (precision, specificity, interference, method comparison, stability, and lay-user study information) for each individual drug analyte were established in the predicate 510(k)s.

The "reported device performance" specifically mentioned in this document relates to the intended use and cut-off concentrations, which are considered to be "Same" as the predicate devices.

Acceptance Criteria (General for Qualitative DOA Tests)Reported Device Performance (as stated in K142580)
Qualitative detection of drugs of abuse in human urineQualitative detection of Amphetamine, Cocaine, Marijuana, Morphine, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates, Ecstasy, Methadone, and Oxycodone drugs in human urine.
Specific cut-off concentrations for each analyteChemtrue® Multi-Panel DOA Dip Card Tests:- Amphetamine: 1000 ng/mL- Cocaine: 300 ng/mL- Methamphetamine: 1000 ng/mL- Morphine: 300 ng/mL- Phencyclidine: 25 ng/mL- Marijuana: 50 ng/mL- Benzodiazepines: 300 ng/mL- Barbiturates: 300 ng/mL- Ecstasy: 500 ng/mL- Methadone: 300 ng/mL- Oxycodone: 100 ng/mLChemtrue® Multi-Panel DOA Dip Card with OPI 2000 Tests:- Amphetamine: 1000 ng/mL- Cocaine: 300 ng/mL- Methamphetamine: 1000 ng/mL- Opiates: 2000 ng/mL- Phencyclidine: 25 ng/mL- Marijuana: 50 ng/mL- Benzodiazepines: 300 ng/mL- Barbiturates: 300 ng/mL- Ecstasy: 500 ng/mL- Methadone: 300 ng/mL- Oxycodone: 100 ng/mL All cut-offs are "Same" as predicate devices.
Maintain performance characteristics (precision, specificity, interference, stability, method comparison, lay-user accuracy) as previously cleared devices.Stated that "Test strips, sample matrix, test format, and cut-off concentrations for these drugs of abuse analytes are identical to those cleared under K102203, K111322, K121339, and K123080. See K102203, K111322, K121339, and K123080 for additional precision, specificity, interference, method comparison, stability and lay-user study information."
Verification and validation of the multi-panel combination."Verification and validation activities were conducted to support combining the test strips to create the multi-panel candidate devices, including interference studies and usability testing." Specific results not detailed in this document.
Intended Use (Prescription and OTC)"Same" as predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated in this document for the overall multi-panel device. The document defers to prior 510(k)s (K102203, K111322, K121339, and K123080) for the individual analyte performance. For a typical qualitative DOA test submission, sample sizes would often involve testing a range of concentrations around the cut-off (e.g., -25%, -50%, +25%, +50% of cut-off) with multiple replicates (e.g., 20 or more) for each analyte, across multiple lots.
  • Data Provenance: Not explicitly stated in this document. Given it's a submission to the FDA, it's generally expected that studies would be conducted in a way that is acceptable to the FDA, likely involving studies in the US or under internationally recognized standards. The document does not specify if the data was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided in the K142580 document. For diagnostic tests of this nature, the "ground truth" for individual samples (e.g., whether a urine sample actually contains a drug above a certain concentration) is typically established by definitive analytical methods, not human expert consensus.

4. Adjudication Method for the Test Set

  • Not applicable in the typical sense for a qualitative diagnostic device where chemical analysis provides the "ground truth." The reference method itself (e.g., GC/MS, LC/MS) is considered the gold standard.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • No, an MRMC study was not done, and this is not an AI/Reader-based device. The Chemtrue® Multi-Panel DOA Dip Card Tests are rapid lateral flow immunoassays designed for direct visual interpretation. There is no AI component or human reader assistance mechanism to evaluate in an MRMC study. However, the document mentions "lay-user study information" in the predicate devices, which would assess how well a typical non-professional user can interpret the results, which is a form of human performance assessment.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • N/A. This is a physical, in vitro diagnostic test, not an algorithm, so the concept of "algorithm only" performance (standalone) does not apply. The device's performance is its direct ability to detect analytes in a given sample.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

  • The implicit ground truth that would have been used for the predicate devices (and thus for these tests, by extension) is definitive analytical methods, specifically Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS). The Indications for Use section explicitly states these are "the preferred confirmatory methods" to obtain a "confirmed assay result" after a preliminary result from the dip card.

8. The Sample Size for the Training Set

  • Not explicitly stated for the individual analytes in this document, as it refers to previous 510(k)s. For this type of qualitative immunoassay, there isn't a "training set" in the machine learning sense. Instead, the "training" (or development and optimization) of the assay's chemical components (antibodies, conjugates) would involve numerous experiments with various concentrations of analytes and interferents to establish optimal performance around the cut-off.

9. How the Ground Truth for the Training Set Was Established

  • As mentioned above, there isn't a "training set" in the AI/ML context here. The development and validation of the chemical components and detection mechanisms for an immunoassay rely on precisely prepared samples with known concentrations of the target analytes, often verified by highly accurate analytical chemistry methods like GC/MS or LC/MS.

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 1, 2014

CHEMTRON BIOTECH, INC. JANE ZHANG QA/RA DIRECTOR 9245 BROWN DEER ROAD, SUITE B SAN DIEGO CA 92121

Re: K142580

Trade/Device Name: Chemtrue® Multi Panel DOA Dip Card Tests Chemtrue® Multi Panel DOA Dip Card with OPI 2000 Tests Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: II Product Code: DKZ, LCM, DIO, LDJ, LAF, DJG, DIS, JXM, DJC, DJR, DNK Dated: November 3, 2014 Received: November 4, 2014

Dear Ms. Jane Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142580

Device Name

Chemtrue® Multi-Panel DOA Dip Card Tests Chemtrue® Multi-Panel DOA Dip Card with OPI 2000 Tests

Indications for Use (Describe)

Chemtrue® Multi-Panel DOA Dip Card Tests:

The Chemtrue® Multi-Panel DOA Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of Amphetamine, Cocaine, Marijuana, Morphine, Methamphetamine, Phencyclidine, Benzodiazepines, Ecstasy, Methadone and Oxycodone drugs in human urine. The test cut-off concentrations and the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMP / METd-Methamphetamine1000
MorphineMORMorphine300
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ9-THC9-COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue® Multi-Panel DOA Dip Card Tests panel can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the prefered confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates and Oxycodone. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

Chemtrue® Multi-Panel DOA Dip Card with OPI 2000 Tests:

The Chemtrue® Multi-Panel DOA Dip Card with OPI 2000 Tests are rapid lateral flow immunoassays for the qualitative detection of Amphetamine, Cocaine, Marijuana, Opiates 2000, Methamphetamine, Phencyclidine, Benzodiazepines. Barbiturates, Ecstasy, Methadone and Oxycodone drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMP /METd-Methamphetamine1000
OpiatesOPIMorphine2000
PhencyclidinePCPPhencyclidine25

FORM FDA 3881 (8/14)

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

MarijuanaTHC11-nor-Δ9-THC9-COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemrue® Multi-Panel DOA Dip Card with OP1 2000 Tests can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GCMS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates and Oxycodone. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

AS REQUIRED BY 21 CFR 807.92(c)

A. SUBMITTER: Chemtron Biotech, Inc. 9245 Brown Deer Road, Suite B, San Diego, CA

  1. TEL: 858-450-0044; FAX: 858-450-0046

Contact Person: Jane Zhang, Director of QA/RA

Official FDA Correspondent 9245 Brown Deer Road, Suite B San Diego, CA 92121 Office: (858) 450-0044; FAX: (858) 450-0046 Email: jane@uschemtronbio.com

Date Prepared: November 20, 2014

B. DEVICE

Trade or Proprietary Name: Chemtrue® Multi-Panel DOA Dip Card Tests Chemtrue® Multi-Panel DOA Dip Card with OPI 2000 Tests

Common Name: Multi-Drug Urine Test Panel

Regulatory Class: Class II

This 510(k) has been submitted for clearance of new drugs of abuse dip card devices that combines drugs of abuse test strips from previously FDA cleared Chemtrue Single/Multi-Panel Drug Screen Dip Card/Cassette Tests (K102203, K111322, K121339, and K123080). These tests have been previously cleared for prescription use (K102203 and K111322) and over-the-counter use (K121339 and K123080) for the following drugs of abuse analytes in the table below. The new devices are intended for over-the-counter and prescription use:

Drug of AbuseProduct CodePanelRegulation Section
AmphetamineDKZToxicology 9121CFR 862.3100, Amphetamine Test System
CocaineDIOToxicology 9121 CFR 862.3250, Cocaine and metabolites Test System
MethamphetamineLAFToxicology 9121 CFR 862.3610, Methamphetamine Test System
OpiatesDJGToxicology 9121 CFR 862.3650, Opiate Test System
PhencyclidineLCMToxicology 91Unclassified, Enzyme immunoassay Phencyclidine
MarijuanaLDJToxicology 9121 CFR 862.3870, Cannabinoids Test System
BenzodiazepinesJXMToxicology 9121 CFR 862.3170, Benzodiazepines Test System
BarbituratesDISToxicology 9121 CFR 862.3150, Barbiturates Test System
Ecstasy (MDMA)DJCToxicology 9121 CFR 862.3610, Methamphetamine Test System
MethadoneDJRToxicology 9121 CFR 862.3620, Methadone Test System
OxycodoneDJGToxicology 9121 CFR 862.3650, Opiate Test System
MorphineDNKToxicology 9121 CFR 862.3640, Morphine Test System

{5}------------------------------------------------

C. PREDICATE DEVICES

510(k): K102203, K111322, K121339 and K123080. Chemtrue Single/Multi-Panel Drug Screen Dip Card/Cassette Tests

D. INDICATIONS FOR USE:

Device Name: Chemtrue® Multi-Panel DOA Dip Card Tests

Chemtrue® Multi-Panel DOA Dip Card with OPI 2000 Tests

Chemtrue® Multi-Panel DOA Dip Card Tests:

The Chemtrue® Multi-Panel DOA Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of Amphetamine, Cocaine, Morphine, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates, Ecstasy, Methadone and Oxycodone drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration(ng/mL)
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMP /METd-Methamphetamine1000
MorphineMORMorphine300
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ⁹-THC9-COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue® Multi-Panel DOA Dip Card Tests panel can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates and Oxycodone. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

Chemtrue® Multi-Panel DOA Dip Card with OPI 2000 Tests:

The Chemtrue® Multi-Panel DOA Dip Card with OPI 2000 Tests are rapid lateral flow immunoassays for the qualitative detection of Amphetamine, Cocaine, Marijuana, Opiates

{6}------------------------------------------------

  1. Methamphetamine, Phencyclidine, Benzodiazepines, Ecstasy, Methadone and Oxycodone drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMP/METd-Methamphetamine1000
OpiatesOPIMorphine2000
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor- $\Delta^{9}$ -THC9-COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue® Multi-Panel DOA Dip Card with OPI 2000 Tests can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates and Oxycodone. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

E. DEVICE DESCRIPTION

The Chemtrue® DOA Tests are colloidal gold based lateral flow immunoassays for the rapid, qualitative detection of drugs of abuse in human urine. The tests are single-use, in vitro diagnostic devices, which come in the formats of dip card, cup, or cassette, as indicated by the test name.

F. SUBSTANTIAL EQUIVALENCE INFORMATION

Comparison with the predicate devices is outlined below:

ItemCandidate DevicesPredicates
Indication(s) for useFor qualitative detection of drugs of abuse in human urineSame
Specimen TypeHuman urineSame

{7}------------------------------------------------

Methodology /TechnologicalLateral flow, competitive binding immunoassay based onthe principle of antigen and antibody immunochemistry.Same
Characteristics
ResultsQualitativeSame
Cut OffAmphetamine 1000 ng/mLCocaine 300 ng/mLMethamphetamine 1000 ng/mLMorphine 300 ng/mLOpiates 2000 ng/mLPCP 25 ng/mLTHC 50 ng/mLBenzodiazepines 300 ng/mLBarbiturates 300 ng/mLEcstasy (MDMA) 500 ng/mLMethadone 300 ng/mLOxycodone 100 ng/mLSame
ConfigurationsDip CardSame
Intended UsePrescription and OTC UseSame

G. TEST PRINCIPLE

The Chemtrue® DOA Tests are rapid lateral flow immunoassays in which chemically modified drugs (drug-protein conjugates) compete with drugs that may be present in urine. On each test strip, a drug-protein conjugate is striped on the test band of the membrane - known as the test region (T) and the anti-drug antibody-colloidal gold conjugate pads are placed at the forward end of the membrane. If target drugs are present in the urine specimen below its cut-off concentration, the solution of the colored antibody-colloidal gold conjugates moves along with the sample solution by capillary action across the membrane to the immobilized drug-protein conjugate zone on the test band region. The colored antibody-gold conjugates then complexes with the drug-protein conjugates to form visible lines. Therefore, the formation of the visible precipitant in the test band indicates a negative result. If the target drug level exceeds its cut-off concentration, the drug/metabolite antigen competes with drug-protein conjugates on the test band region for the limited antibody on the colored drug antibody-colloidal gold conjugate pad. The drug will saturate the limited antibody binding sites and the colored antibody-colloidal gold conjugate cannot bind to the drug-protein conjugate at the test region of the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result.

A band should form in the control region (C) of the devices regardless of the presence of drug in the sample to indicate that the test has been performed properly.

Monoclonal anti-drug antibodies are used on the AMP/COC/MET/MOR/OPI/ PCP/THC/ BAR/MDMA/MTD/OXY Test devices which are derived from mouse. The polyclonal antidrug antibodies are used on BZO Test devices which are derived from sheep/mouse.

{8}------------------------------------------------

H. PERFORMANCE CHARACTERISTICS

All the drugs of abuse analytes of the candidate devices (AMP, COC, MAMP, OPI, MOR, PCP, THC, BZO, BAR, MDMA, MTD, and OXY) were previously cleared under K102203, K111322, K121339, and K123080. Test strips, sample matrix, test format, and cut-off concentrations for these drugs of abuse analytes are identical to those cleared under K102203, K111322, K121339, and K123080. See K102203, K111322, K121339, and K123080 for additional precision, specificity, interference, method comparison, stability and lay-user study information.

Verification and validation activities were conducted to support combining the test strips to create the multi-panel candidate devices, including interference studies and usability testing.

I. CONCLUSION:

Based on the test principle and performance characteristics of the candidate device, it is concluded that the candidate devices are substantially equivalent to the predicate device.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).