K Number
K121339

Validate with FDA (Live)

Date Cleared
2012-05-17

(14 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Chemtron Biotech, Inc.'s Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of up to six of the following drugs in a variety of combinations in human urine. The designed cutoff concentrations and the calibrators used for these drugs are as follows:

Analyte Abbreviation Calibrator Cutoff Concentration
Amphetamine AMP d-Amphetamine 1000 ng/mL
Cocaine COC Benzoylecgonine 300 ng/mL
Marijuana THC 11-nor-Δ9-THC9-COOH 50 ng/mL
Methamphetamine MET d-Methamphetamine 1000 ng/mL
Opiates OPI/MOR Morphine 300 ng/mL
Phencyclidine PCP Phencyclidine 25 ng/mL

The Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are intended for the qualitative detection of drugs of abuse for health care professional, in vitro diagnostic and Over-the-Counter (OTC) use. These assays provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods.

Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

Device Description

The Drugs of Abuse (DOA) Screen Panels are one-step lateral flow immunoassays in which chemically labeled drugs (drug-protein conjugates) compete for limited antibody binding sites with drugs that may be present in urine. The test device consists of up to six test strips placed into separate panels of a plastic holder. On each test strip, a drug-protein conjugate is striped on the test band of the membrane - known as the test region (T) and the drug antibody-colloidal gold conjugate pads are placed at one end of the membrane (opposite in morphine). In the absence of drugs in the urine, the solution of the colored antibody-colloidal gold conjugates move along with the sample solution upward chromatographically by capillary action across the immobilized drug-protein conjugate zones on the test band region. The colored antibody-gold conjugates then complexes with the drug-protein conjugates to form visible lines. Therefore, the formation of the visible precipitant in the test band occurs when the test urine is negative for the drug. If any drug is present in the urine, the drug/metabolite antigen competes with drug-protein conjugates on the test band region for the limited antibody on the colored drug antibody-collooidal gold conjugate pad. When a sufficient amount of drug is present in the urine, the drug will saturate the limited antibody binding sites and the colored antibody -colloidal gold conjugate cannot bind to the drug-protein conjugate at the test region of the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result.

A control band with a different antibody reaction is added to the membrane strip at the control region (C) to indicate that the test has performed properly. This control line is manufactured as a builtin internal control of the test device and should always appear regardless of the presence of drug or metabolite. If the control line does not appear the test cassette should be discarded. The presence of this colored band in the control region also serves 1) as verification that adequate specimen volume is added (flooding, if too much urine is added, or no flow, due to insufficient urine volume), 2) the test device IS properly functioning, and 3) as reagent control.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Acceptance Criteria and Device Performance

The core acceptance criteria for this device are its accuracy of detection for various drugs of abuse when compared to a confirmatory method (GC/MS). The study aimed to demonstrate that lay-users could perform the testing and interpret the results correctly with a high level of agreement.

Table 1: Acceptance Criteria and Reported Device Performance (Summary from provided text)

AnalyteCutoff ConcentrationAcceptance Criteria (Implied)Reported Performance (Agreement with GC/MS)
Amphetamine1000 ng/mLHigh agreement (e.g., >95%)98.9% (Dip Card), 100% (Cassette)
Cocaine300 ng/mLHigh agreement (e.g., >95%)100% (Dip Card), 100% (Cassette)
Marijuana (THC)50 ng/mLHigh agreement (e.g., >95%)100% (Dip Card), 100% (Cassette)
Methamphetamine1000 ng/mLHigh agreement (e.g., >95%)100% (Dip Card), 100% (Cassette)
Opiates (Morphine)300 ng/mLHigh agreement (e.g., >95%)100% (Dip Card), 98.9% (Cassette)
Phencyclidine25 ng/mLHigh agreement (e.g., >95%)98.9% (Dip Card), 98.9% (Cassette)
Overall->98.9% total correlation98.9% total correlation

Note: The document explicitly states "Correlation studies produced a 98.9% total correlation when compared to the GC/MS methodology" as the overarching performance metric, suggesting this was the primary acceptance criterion. The individual agreement percentages for each drug further support the overall claim.


Study Details

  1. Sample size used for the test set and the data provenance:

    • Sample Size: The study evaluated 200 OTC (Over-the-Counter) lay-users. The tables provided show results for various drug concentrations, including "no drug present", "GC/MS Negative (-50% to -25% cutoff)", "Near cutoff negative (-25% cutoff to cutoff)", "Near cutoff positive (cutoff to +25% cutoff)", "GC/MS Positive (+25% to +50% cutoff)", and "GC/MS Positive (+50% to 200% cutoff)". For each of these categories and for each drug, there were typically 30 samples, indicating a total of 180 samples per drug, and thus 1080 samples across all 6 drugs for each device type (Dip Card and Cassette). The document states "blind-labeled spiked urine correlation study," implying the urine samples were prepared and then tested.
    • Data Provenance: The study used "spiked urine" samples, meaning the drug concentrations were artificially introduced into urine. The samples were tested by lay-users from "three (3) sites". It's not explicitly stated if these sites were in a specific country, but given the company's address (San Diego, CA, USA) and the FDA submission, it's highly probable the study was conducted in the USA. The study design is prospective in the sense that new tests were run according to a predefined protocol.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of experts: Not applicable in the traditional sense of human expert consensus. The ground truth was established by a laboratory method.
    • Qualifications: The gold standard for establishing ground truth was Gas Chromatography / Mass Spectrometry (GC/MS) methodology. This is a highly accurate analytical chemistry technique performed by trained laboratory personnel.
  3. Adjudication method for the test set:

    • Adjudication: Not applicable. The ground truth was determined by GC/MS, so there was no need for human expert adjudication to resolve discrepancies in ground truth. The study compared the device's reading to the GC/MS result directly.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, an MRMC comparative effectiveness study was not performed in the context of human readers improving with or without AI assistance. This device is a rapid, visually-read immunoassay, not an AI-powered diagnostic tool requiring interpretation by experts in an MRMC setting. The study assessed the ability of "200 OTC lay-users" to correctly interpret the visual results of the device.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: No, a standalone algorithm-only performance was not done. The device is a "visually-read" immunoassay where human users (specifically "lay-users" for OTC use) interpret the presence or absence of colored lines. The reported performance includes the human interpretation component.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Type of Ground Truth: The ground truth was established using an analytical reference method: Gas Chromatography / Mass Spectrometry (GC/MS). This is a highly accurate chemical method for identifying and quantifying substances in a sample. The document also mentions Liquid Chromatography / Mass Spectrometry (LC/MS) as an alternative confirmatory method.
  7. The sample size for the training set:

    • Training Set Sample Size: The document does not explicitly state a training set size. For this type of immunoassay device, there isn't typically a "training set" in the machine learning sense. The device's performance characteristics (e.g., analytical sensitivity, precision, specificity, stability) were established and likely optimized during product development, but this process isn't described as using a distinct "training set" in the filing. The "K102203" submission is referenced for "Other performance data" such as analytical sensitivity and precision, which would have been part of the internal development and validation, not a separate training set as understood in AI/ML.
  8. How the ground truth for the training set was established:

    • Training Set Ground Truth: As there's no explicitly defined "training set" in the context of this immunoassay's submission, the method for establishing its ground truth is not detailed via this document. However, the foundational analytical sensitivity and specificity of immunoassays are typically established by running known concentrations of analytes and interferents and confirming these with reference methods like GC/MS or LC/MS during the assay's development and internal validation.

{0}------------------------------------------------

510(k) Summary

MAY 1 7 2012

AS REQUIRED BY 21 CFR 807.92(c)

The Assigned 510(k) number is __KI 21339

Date of Summary: May 02nd, 2012

Common Name: Drugs of Abuse Screening Tests

Classification Name: Immunoassay for the detection of drugs of abuse

Trade/Proprietary Name:

Chemtron Biotech, Inc.'s Chemtrue® Single / Multi-Panel Drug Screen Dip Card / Cassette Tests, contain 1 to 6 of the following DOA test(s) in each device:

    1. Amphetamine test strip
  • Cocaine test strip 2.
    1. Marijuana (THC) test strip
    1. Methamphetamine test strip
    1. Opiates (Morphine) cut-off: 300ng/ml test strip
    1. Phencyclidine test strip

Owner:

Ellen Meng President, Chemtron Biotech, Inc. . 8370 Juniper Creek Lane, #1-2 San Diego, CA 92126 TEL: (858) 530-2868 FAX: (858) 530-2878

Contact Person:

Jane Zhang, Director of QA/RA Official FDA Correspondent 8370 Juniper Creek Lane, #1-2 San Diego, CA 92126 Office: (858) 530-2868 Mobile: (858) 997-7472 FAX: (858) 530-2878

{1}------------------------------------------------

Substantial Equivalency

The Chemtrue® Single/Multi-Panel Drug Screen Test is substantially equivalent to other tests currently on the market.

Predicate Device Name Predicate Device 510(k)# K102203 Chemtrue® Single/Multi-Panel Drug Screen Test

The predicate kit package insert is enclosed in ATTACHMENT A, page 41 of this submission.

Proposed Labeling or Promotional Material for the Device:

Description of the device can be found in the attached proposed labeling, including an explanation of how the device functions, technical principle and concepts that form the basis for the device, as well as the physical and performance characteristics of the device design, materials used and physical properties. In accordance with FDA labeling requirements 21 CFR 809.10, enclosed are the draft copies of product labeling including copies of the technical product inserts. See ATTACHMENT B and C, page 42 and 46 of this submission.

Intended Use

The Chemtrue® test device is intended for the qualitative detection of drugs of abuse, for Over-the-Counter (OTC) and in vitro diagnostic use.

The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods.

Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

Technological Characteristics and Science Principles

The Drugs of Abuse (DOA) Screen Panels are one-step lateral flow immunoassays in which chemically labeled drugs (drug-protein conjugates) compete for limited antibody binding sites with drugs that may be present in urine. The test device consists of up to six test strips placed into separate panels of a plastic holder. On each test strip, a drug-protein conjugate is striped on the test band of the membrane - known as the test region (T) and the drug antibody-colloidal gold conjugate pads are placed at one end of the membrane (opposite in morphine). In the absence of drugs in the urine, the solution of the colored antibody-colloidal gold conjugates move along with the sample solution upward chromatographically by capillary action across the immobilized drug-protein conjugate zones on the test band region. The colored antibody-gold conjugates then complexes with the drug-protein conjugates to form visible lines. Therefore, the formation of the visible precipitant in the test band occurs when the test urine is negative for the drug. If any drug is present in the urine, the

{2}------------------------------------------------

drug/metabolite antigen competes with drug-protein conjugates on the test band region for the limited antibody on the colored drug antibody-colloidal gold conjugate pad. When a sufficient amount of drug is present in the urine, the drug will saturate the limited antibody binding sites and the colored antibody -colloidal gold conjugate cannot bind to the drug-protein conjugate at the test region of the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result.

A control band with a different antibody reaction is added to the membrane strip at the control region (C) to indicate that the test has performed properly. This control line is manufactured as a builtin internal control of the test device and should always appear regardless of the presence of drug or metabolite. If the control line does not appear the test cassette should be discarded. The presence of this colored band in the control region also serves 1) as verification that adequate specimen volume is added (flooding, if too much urine is added, or no flow, due to insufficient urine volume), 2) the test device IS properly functioning, and 3) as reagent control.

Comparison with Predicate

The Chemtron Biotech, Inc.'s Chemtrue® Single/Multi-panel Drug Screen Cassette and Dip Card tests are similar or the same as the previously cleared predicate(s) in the following ways: test principle, indication for use, used in a professional, point-of-care setting, OTC use, read time and sample matrix. The candidate device and the predicates are both visually-read single use devices. All of these products are based on the same technological characteristics, scientific principle and similar testing procedures. The similarities and differences between these tests are summarized as follows:

SIMILARITIES
ItemChemtrue® DevicePredicate Kit
Indications For UseA rapid qualitative lateral flow immunoassayfor the detection of potential abuse of one ormore drugsSame
SpecimenUrineSame
Technological Characteristics andPrincipleOne-Step lateral flow competitiveImmunoassaySame
Positive result1 colored lineSame
Device Design/Negative result2 colored linesSame
PerformanceDetection reagentColloidal goldSame
AccuracyAssessmentConfirm with GC/MS reference methodSame
Cut-offAmphetamine 1000 ng/mlMethamphetamine 1000 ng/mlCocaine 300 ng/mlMarijuana (THC) 50 ng/mlPhencyclidine 25 ng/mlOpiates (Morphine) 300 ng/mlSameSameSameSameSameSame

{3}------------------------------------------------

Safety and PrecautionAll urine specimens should be consideredpotentially hazardous and handled in the samemanner as infectious agent.Same
Read time5 minutesSame
Storage2 - 30 ℃ (36 - 86°F)Same
DIFFERENCES
ItemChemtrue® DevicePredicate Kit
Intended useFor Over-the-Counter (OTC) UseFor Prescription Use Only

DISCUSSION AND CONCLUSION:

Based on the technological characteristics/principle, features of the device designs, test specimen matrix, test method and performance characterizations, as the set forth above, it can be concluded that Chemtrue® Single/ Multi-panel Drug Screen tests are substantially equivalent to the predicate kit and the other products that are manufactured by Alere and Alfa Scientific Design Inc. presently distributed commercially.

Performance Data:

Chemtron Biotech, Inc. has reviewed the requirements of Section 514 of the Act, which states that to date no performance standards has been established for drug screen test systems by the FDA.

However, the studies listed in the notification were conducted according to "Assessing the Safety and Effectiveness of Home-Use in vitro Diagnostic Devices (IVDs): Draft Points to Consider Regarding Labeling and Premarket Submissions (Text Only). Center for Devices and Radiological Health. October 1988", "The Draft Guidance for Industry and FDA Staff" - " Premarket Submission and Labeling Recommendations for Drugs of Abuse Screening Tests, issued on: December 2, 2003", including the design of draft labeling and package inserts.

Chemtrue® Single / Multi-panel Drug Screen Cassette and Dip Card tests are one-step, lateral flow, colloidal gold based chromatographic immunoassays for the rapid, qualitative detection of Amphetamine, Methamphetamine, Marijuana (THC), Phencyclidine, Cocaine and Opiates (Morphine) 300 in human urine. It is intended for Over-the-Counter (OTC) Use.

This assay provides a preliminary test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory result. Clinical and professional judgment must be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

The product performance characteristics of Chemtrue® Single/Multi-panel Drug Screen cassette and Dip Card were evaluated by 200 of OTC lay-users from three (3) sites with the confirmed GC/MS

{4}------------------------------------------------

values in a blind-labeled spiked urine correlation study. The results of these studies demonstrate Chemtrue® Single/ Multi-panel Drug Screen Test to be substantially equivalent to the performance characteristics of GC/MS methodology. Correlation studies produced a 98.9% total correlation when compared to the GC/MS methodology.

Chemtrue® Single /Multi-Panel Drug Screening Tests vs GC/MS Value Analysis

Samples with drug concentration above the cut-off level were considered presumptive positive and concentration below the cut-off were considered negative.

Table 1. Method comparison (OTC Accuracy) study result summary between the Chemtrue® Dip Card
Test and the GC/MS values
Chemtrue® DrugScreen Dip Card Test(-)(+)%AgreementwithGC/MSvalues
No drugpresentGC/MSNegative(-50% to-25% cutoff)Near cutoffnegative(-25% cutoff tocutoff)Near cutoffpositive (cutoffto +25% cutoff)GC/MSPositive(+25% to +50%cutoff)GC/MSPositive(+50% to 200%cutoff)
AMP(+)001303030100%
(-)30302900098.9%
COC(+)000303030100%
(-)303030000100%
MET(+)000303030100%
(-)303030000100%
OPI300(+)000303030100%
(-)303030000100%
PCP(+)001303030100%
(-)30302900098.9%
THC(+)000303030100%
(-)303030000100%

Table 2. Method comparison (Accuracy) study result summary between the Chemtrue® Cassette Test and the GC/MS values

(-)(+)%
Chemtrue® DrugScreen Cassette TestNo drug presentGC/MS Negative(-50% to -25% cutoff)Near cutoff negative(-25% cutoff to cutoff)Near cutoff positive (cutoff to +25% cutoff)GC/MS Positive(+25% to +50% cutoff)GC/MS Positive(+50% to 200% cutoff)Agreement with GC/MS values
AMP(+)000303030100%
(-)303030000100%
COC(+)000303030100%
(-)303030000100%
MET(+)000303030100%

{5}------------------------------------------------

Chemtron Biotech, Inc. 510(k) Submission: Chemtrue® Drug Screen Dip Card/Cassette Tests for OTC Use

(-)303030000100%
OP1300(+)ﻢ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ0303030100%
30302900098.9%
РСР(+)00303030100%
30302900098.9%
THC+000303030100%
3030300. 00100%

CONCLUSION:

Compared to the GC/MS values of the spiked urine samples that were tested by 200 OTC Lay-users from separate three (3) sites, the OTC accuracy study results demonstrate that the intended lay-users can perform the testing and interpret the results correctly with equal to and greater than 98.9% agreement with the GC/MS assayed values. The results also showed the substantial equivalency between the Chemtrue® Single/ Multi-Panel Drug Screen tests of both Dip Card and Cassette formats and the GC/MS reference method. Based on the study results, it can conclude that Chemtrue® Single/ Multi-panel Drug Screen cassette and Dip card is safe and effective in detecting Amphetamine, Cocaine, Methamphetamine, Marijuana, Opiates (Morphine) 300 and Phencyclidine in human urine.

Other performance data, such as analytical sensitivity (Cutoff characteristics), precision (Reproducibility), specificity (Cross-reactivity and interference) and stability studies were established in K102203.

510(k) Summary was prepared By: Jane Zhang on May 02ª, 2012

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the seal is a stylized image of a human figure with outstretched arms.

Food and Drug Administration

10903 New Hampshire Avenue Silver Spring, MD 20993

Chemtron Biotech, Inc. c/o Jane Zhang 8370 Juniper Creek Lane Suite 1-2 San Diego, CA 92126

MAY 1 7 2012

Re: K121339

Trade/Device Name: Chemtrue Single/Multi-Panel Drug Screen Cassette and Dip Card Tests

Regulation Number: 21 CFR 862.3100

Regulation Name: Amphetamine test system

Regulatory Class: Class II

Product Code: DKZ, DIO, DJC, DNK, LCM and LDJ

  • Dated: May 02, 2012
    Received: May 03, 2012

Dear Ms Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts , 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{7}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you deall office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 proxio note are regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrig's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance...

You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours,

Signature

Couriney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use

510(k) Number (if known): ___K I乙1339

Device Name: Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests

Indications for Use:

The Chemtron Biotech, Inc.'s Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of up to six of the following drugs in a variety of combinations in human urine. The designed cutoff concentrations and the calibrators used for these drugs are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration
AmphetamineAMPd-Amphetamine1000 ng/mL
CocaineCOCBenzoylecgonine300 ng/mL
MarijuanaTHC11-nor-Δ9-THC9-COOH50 ng/mL
MethamphetamineMETd-Methamphetamine1000 ng/mL
OpiatesOPI/MORMorphine300 ng/mL
PhencyclidinePCPPhencyclidine25 ng/mL

The Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are intended for the qualitative detection of drugs of abuse for health care professional, in vitro diagnostic and Over-the-Counter (OTC) use. These assays provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods.

Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _____X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Concurrence of CDRL

Division Sign-Off

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) 121339

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).