K Number
K102203

Validate with FDA (Live)

Date Cleared
2011-01-20

(169 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Chemtron Biotech, Inc.'s Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are rapid chromatographic immunoassays for the qualitative detection of up to six of the following drugs in a variety of combinations in human urine. The designed cutoff concentrations and direct calibrator for these drugs are as follows:

AnalyteAbbreviationCalibratorCutoff
Concentration
AmphetamineAMPd-Amphetamine1000 ng/ml
CocaineCOCBenzoylecgonine300 ng/ml
MarijuanaTHC11-nor- $\Delta^9$ -THC9-COOH50 ng/ml
MethamphetamineMETd-Methamphetamine1000 ng/ml
OpiatesOPI/MORMorphine300 ng/ml
PhencyclidinePCPPhencyclidine25 ng/ml

The Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card are intended for qualitative detection of drugs of abuse for health care professionals, in vitro diagnostic use and prescription use ONLY. It is not intended for point-of-care settings or over the counter use.

This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods.

Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

Device Description

Chemtrue® Single / Multi-panel Drug Screen Cassette and Dip Card tests are one-step, lateral flow, colloidal gold based chromatographic immunoassays for the rapid, qualitative detection of Amphetamine, Methamphetamine, Marijuana (THC), Phencyclidine, Cocaine and Opiates (Morphine) 300 in human urine. It is intended for Professional /Prescription Use ONLY.

This assay provides a preliminary test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory result. Clinical and professional judgment must be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Chemtrue® Single/Multi-panel Drug Screen Tests, extracted from the provided text:

Acceptance Criteria and Device Performance

Test AnalyteCut-off (ng/ml)Acceptance Criteria (from predicate kit reference)Reported Device Performance (Overall Agreement with GC/MS)
Amphetamine1000>90% Analytical Accuracy (from predicate kit)99.1%
Cocaine300>90% Analytical Accuracy (from predicate kit)100%
Opiates300>90% Analytical Accuracy (from predicate kit)98.3%
Methamphetamine1000>90% Analytical Accuracy (from predicate kit)98.1%
Phencyclidine25>90% Analytical Accuracy (from predicate kit)100%
Marijuana (THC)50>90% Analytical Accuracy (from predicate kit)100%

Note on Acceptance Criteria: The document explicitly states: "Chemtron Biotech, Inc. has reviewed the requirements of Section 514 of the Act, which states that to date no performance standards has been established for drug screen test systems by the FDA." The acceptance criteria are inferred from the predicate device's reported accuracy of ">90%" and the general claim of substantial equivalence. The overall performance goal for the clinical specimen correlation study was "> 98% total correlation when compared to the GC/MS methodology."


Study Details

1. Sample size used for the test set and the data provenance:

  • Test Set (Clinical Specimen Correlation Study):
    • Total Samples: 673 clinical specimens.
    • Per Analyte Breakdown (from Table 1):
      • AMP: 114
      • COC: 115
      • OPI(MOR)300: 119
      • MET: 106
      • PCP: 114
      • THC: 106
    • Provenance: "Clinical specimens." The country of origin is not specified, but the submission is to the US FDA, implying relevance to a US context or similar regulatory standards. The study is retrospective, as it used "blind-labeled clinical specimen correlation study" with "confirmed GC/MS values" already established.

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable in the context of this device. The ground truth was established by Gas Chromatography/Mass Spectrometry (GC/MS), not by human expert interpretation of the test device itself. GC/MS is a laboratory analytical method.

3. Adjudication method for the test set:

  • Not applicable, as the ground truth was established by GC/MS, which is a definitive analytical method, not requiring adjudication of subjective interpretations.

4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic assay, not an AI-assisted diagnostic tool that requires human reader improvement assessment.

5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Yes, a standalone performance study was done. The "accuracy (Method comparison) study" directly compared the Chemtrue® device's results to GC/MS values, without human-in-the-loop assistance for interpretation. The test device itself provides a visual result (colored lines) which is then interpreted by a human, but the study focused on the device's analytical accuracy against the gold standard.

6. The type of ground truth used:

  • Pathology/Outcomes Data: The primary ground truth for the accuracy study was Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS). These are described as "the preferred confirmatory methods" and "specific alternative chemical method[s]". The study states it evaluated performance "with the confirmed GC/MS values".

7. The sample size for the training set:

  • Not explicitly stated for a "training set." This type of in-vitro diagnostic device (immunoassay) is not typically "trained" in the same way a machine learning algorithm is. The device's performance is based on its chemical and biological design.
  • However, various other studies were conducted on the device's inherent characteristics:
    • Analytical Sensitivity Study Data: 180 samples per control level (0, -50%, -25%, Cut-off, +25%, +50% of cut-off), totaling 1080 samples for each drug assay's sensitivity testing.
    • Analytical Precision Study Data (Reproducibility): For each drug (AMP, COC, MAMP, OPI, PCP, THC), 30 samples were tested per control level (Negative, -50%, -25%, Cut-off, +25%, +50% of cut-off) across three operators and three batches for 10 days, accumulating significant data for reproducibility.
    • Analytical Specificity Study: Data summarized in a section, implying tests with various interfering substances/cross-reactants.
    • Stability Study: Three lots of each product format (Dip Card and Cassette) were included.

8. How the ground truth for the training set was established:

  • Again, the concept of a "training set" doesn't directly apply here. For the various analytical performance studies (sensitivity, precision, specificity), the "ground truth" or reference values were established using commercial urine controls that were "GC/MS confirmed" and "blind coded". This implies that external, verified standards were used to assess the device's performance under controlled conditions.

{0}------------------------------------------------

510(k) Summary

AS REQUIRED BY 21 CFR 807.92(c)

JAN 2 0 2011

The Assigned 510(k) number is _ _ 102203

Date of Summary: August 3rd, 2010

Common Name: Drugs of Abuse Screening Tests

Classification Name: Immunoassay for the detection of drugs of abuse

Trade/Proprietary Name:

Chemtron Biotech, Inc.'s Chemtrue® Single / Multi-panel Drug Screen Dip Card / Cassette Tests, contain 1 to 6 of the following DOA test(s) in each device:

    1. Amphetamine test strip
    1. Cocaine test strip
    1. Marijuana (THC) test strip
    1. Methamphetamine test strip

с

    1. Opiates (Morphine) cut-off: 300ng/ml test strip
    1. Phencyclidine test strip

Owner:

Ellen Liu President, Chemtron Biotech, Inc. 8370 Juniper Creek Lane, #1-2 San Diego, CA 92126 TEL: (858) 530-2868 FAX: (858) 530-2878

Contact Person:

Jane Zhang, Director of QA/RA Official FDA Correspondent

8370 Juniper Creek Lane, #1-2 San Diego, CA 92126 Office: (858) 530-2868 Mobile: (858) 997-7472

{1}------------------------------------------------

FAX: (858) 530-2878

Substantial Equivalency:

The Chemtrue® Single/Multi Drug Screen Test is substantially equivalent to other tests currently on the market.

Test AnalytePredicate Device NamePredicate Device510(k) #
AmphetamineACON (INNOVACON) One Step Drug Screen Test CardK020771
CocaineACON (INNOVACON) One Step Drug Screen Test CardK020771
MethamphetamineACON (INNOVACON) One Step Drug Screen Test CardK020771
THCACON (INNOVACON) One Step Drug Screen Test CardK020771
Opiates (Morphine)300ACON (INNOVACON) One Step Drug Screen Test CardK020771
PCPACON (INNOVACON) One Step Drug Screen Test CardK020771

The predicate kit package insert is enclosed in ATTACHMENT A of this submission.

Proposed Labeling or Promotional Material for the Device:

Description of the device can be found in the attached proposed labeling, including an explanation of how the device functions, technical principle and concepts that form the basis for the device, as well as the physical and performance characteristics of the device design, materials used and physical properties. In accordance with FDA labeling requirements 21 CFR 809.10, enclosed are the draft copies of product labeling including copies of the technical product inserts. See ATTACHMENT B, C and D.

Intended Use

The Chemtrue® test device is intended for qualitative detection of drugs of abuse, for professional use, in vitro diagnostic use and for prescription use ONLY.

This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods.

Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

Technological Characteristics and Science Principles

The Drugs of Abuse (DOA) Screen Panels are one-step lateral flow immunoassays in which chemically labeled drugs (drug-protein conjugates) compete for limited antibody binding sites with drugs that may

{2}------------------------------------------------

be present in urine. The test device consists of up to six test strips placed into separate panels of a plastic holder. On each test strip, a drug-protein conjugate is striped on the test band of the membrane - known as the test region (T) and the drug antibody-colloidal gold conjugate pads are placed at one end of the membrane (opposite in morphine). In the absence of drugs in the urine, the solution of the colored antibody-colloidal gold conjugates move along with the sample solution upward chromatographically by capillary action across the immobilized drug-protein conjugate zones on the test band region. The colored antibody-gold conjugates then complexes with the drug-protein conjugates to form visible lines. Therefore, the formation of the visible precipitant in the test band occurs when the test urine is negative for the drug. If any drug is present in the urine, the drug/metabolite antigen competes with drug-protein conjugates on the test band region for the limited antibody on the colored drug antibody-colloidal gold conjugate pad. When a sufficient amount of drug is present in the urine, the drug will saturate the limited antibody binding sites and the colored antibody -colloidal gold conjugate cannot bind to the drug-protein conjugate at the test region of the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result.

A control band with a different antigen/antibody reaction is added to the immunochromatographic membrane strip at the control region (C) to indicate that the test has performed properly. This control line is manufactured as a built-in internal control of the test device and should always appear regardless of the presence of drug or metabolite. If the control line does not appear the test cassette should be discarded. The presence of this colored band in the control region also serves 1) as verification that adequate specimen volume is added (flooding, if too much urine is added, or no flow, due to insufficient urine volume), 2) the test device IS properly functioning, and 3) as reagent control.

Summary of Device Similarities and Differences

Chemtron Biotech, Inc.'s Chemtrue® Single/Multi-panel Drug Screen Cassette and Dip Card tests are similar to other FDA-cleared devices for the qualitative detection of drugs of abuse tests: Amphetamine, Cocaine, Marijuana, Opiates, Methamphetamine and Phencyclidine test from other manufacturers such as ACON (Current INNOVACON), Branan Medical, etc. All of these products are based on the same technological characteristics, science principle and similar procedures. The similarities and differences between these tests are summarized as follows:

SIMILARITIES
ItemChemtrue® DevicePredicate Kit
Intended Use/Indications for useQualitative detection of drugs-of- abuse in urine for Professional, Prescription, In VitroSame

{3}------------------------------------------------

Diagnostic Use Only
SpecimenUrineSame
Technological Characteristics andPrincipleOne-Step lateral flow competitiveImmunoassaySame
Positive result1 colored lineSame
Device Design/Negative result2 colored linesSame
PerformanceDetectionreagentColloidal goldSame
AccuracyAssessmentConfirm with GC/MS reference methodSame
Amphetamine 1000 ng/mlSame
Methamphetamine 1000 ng/mlSame
Cut-offCocaine 300 ng/mlSame
Marijuana (THC) 50 ng/mlSame
Phencyclidine 25 ng/mlSame
Opiates(Morphine) 300 ng/mlSame
Safety and PrecautionAll urine specimens should be consideredpotentially hazardous and handled in the samemanner as infectious agent.Same
Read time5 minutesSame
Storage2 - 30 °C (36 - 86°F)Same
DIFFERENCES
ItemChemtrue® DevicePredicate Kit
Pre-treatment for urinespecimenDoes not require pre-treatmentfor urine specimen. It is betterthan the Predicate kit.Urine specimen needs to be centrifuged,filtered, or allowed to settle to obtainclear specimen for testing.
Time Frame of ResultStabilizationDo not read after 8 minutes.Results remain stable for up to 4 hoursafter test initiation.

DISCUSSION AND CONCLUSION:

Based on the technological characteristics/principle, features of the device designs, test specimen matrix, test method and performance characterizations, as the set forth above, it can be concluded

{4}------------------------------------------------

that Chemtrue® Single/ Multi-panel Drug Screen tests are substantially equivalent to the predicate kit ACON's (INNOVACON) One Step Single and Multi Drug Screen test card product presently distributed commercially.

One of the advantages of the Chemtrue® Drug Screen Tests is no urine specimen pre-treatment, which is better than the Predicate kit, because it is much simple and convenience for the intended user(s).

With regarding to the result read time frame difference, since the device labeling clearly indicates the required "Read Time", it does not impact the result interpretation by the intended user(s). The device safety and effectiveness won't be affected either. For detailed discussion, refer to the section "Substantial Equivalence Discussion" of this submission.

Performance Data:

Chemtron Biotech, Inc. has reviewed the requirements of Section 514 of the Act, which states that to date no performance standards has been established for drug screen test systems by the FDA.

However, the studies listed in the notification are conducted according to "The Draft Guidance for Industry and FDA Staff" - " Premarket Submission and Labeling Recommendations for Drugs of Abuse Screening Tests, issued on: December 2, 2003", including the design of draft labeling and package inserts.

Chemtrue® Single / Multi-panel Drug Screen Cassette and Dip Card tests are one-step, lateral flow, colloidal gold based chromatographic immunoassays for the rapid, qualitative detection of Amphetamine, Methamphetamine, Marijuana (THC), Phencyclidine, Cocaine and Opiates (Morphine) 300 in human urine. It is intended for Professional, Prescription Use ONLY.

This assay provides a preliminary test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory result. Clinical and professional judgment must be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

The product performance characteristics of Chemtrue® Single/Multi-panel Drug Screen cassette and Dip Card were evaluated with the confirmed GC/MS values in the blind-labeled clinical specimen correlation study. The results of these studies demonstrate Chemtrue® Single/ Multi-panel Drug Screen Test to be substantially equivalent to the performance characteristics of GC/MS methodology. Correlation studies, using clinical specimens, produced a > 98% total correlation when compared to the GC/MS methodology. Also the accuracy study data of Chemtrue® devices demonstrates the substantial equivalency to the data in the package insert of the predicate kit – ACON (INNOVACON) One Step Single and Multi Drug Screen test card.

{5}------------------------------------------------

Chemtrue® Single /Multi-Panel Drug Screening Tests vs GC/MS Value Analysis

Samples with drug concentration above the cut-off level were considered presumptive positive and concentration below the cut-off were considered negative.

TestPositive AgreementNegative AgreementOverall Agreement
AMP57/57 = 100%56/57 = 98.2%113/114 = 99.1%
COC61/61= 100%54/54= 100%115/115= 100%
OPI(MOR)30055/56= 98.2%62/63 = 98.4%117/119 = 98.3%
MET54/56 = 96.4%50/50 = 100%104/106 = 98.1%
PCP52/52= 100%62/62= 100%114/114 = 100%
THC56/56 = 100%50/50= 100%106/106= 100%

Table 1. The data was summarized from method comparison (Accuracy) study against the GC/MS values of the clinical specimens

Conclusion:

Compared to the GC/MS values of clinical urine specimens (Total of 673 specimens), the results of the Accuracy (Method comparison) study demonstrate the substantial equivalency between the Chemtrue® Single/ Multi-Panel Drug Screen tests and the GC/MS reference method. It is also equivalent to the predicate kit, when compared to the data provided in the package insert of the predicate kit [ACON (INNOVACON) One Step Drug Screen test card]. The results demonstrated that Chemtrue® Single/ Multi-panel Drug Screen cassette and Dip card is safe and effective in detecting Amphetamine, Cocaine, Methamphetamine, Marijuana, Opiates (Morphine) 300 and Phencyclidine in human urine.

Other non-clinical performance data, such as analytical sensitivity (Cutoff characteristics), precision (Reproducibility), specificity (Cross-reactivity and interference) and stability studies are summarized in Section "Performance Characteristics", and the raw data is enclosed in ATTACHMENT E and ATTACHMENT G of this submission.

510(k) Summary was prepared By: Jane Zhang on August 3rd, 2010

PREMARKET NOTIFICATION

TRUTHFUL AND ACCURATE STATEMENT

(As required by 21 CFR 807.87(j))

{6}------------------------------------------------

I certify that in my capacity as QA/RA Director of Chemtron Biotech, Inc., I believe to the best of my knowledge that all data and information submitted in the premarket notification are truthful and accurate and that no material fact has been omitted.

(Signature), (Date)

Jane Zhang

Director of QA/RA, Chemtron Biotech, Inc.

FDA Official Correspondent

K

510(k) Number

Executive Summary

Description of the Device

Chemtrue® Single / Multi-panel Drug Screen Cassette and Dip Card tests are one-step, lateral flow, colloidal gold based chromatographic immunoassays for the rapid, qualitative detection of

{7}------------------------------------------------

Amphetamine, Methamphetamine, Marijuana (THC), Phencyclidine, Cocaine and Opiates (Morphine) 300 in human urine. It is intended for Professional /Prescription Use ONLY.

This assay provides a preliminary test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory result. Clinical and professional judgment must be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

Concise Summary of Performance Testing:

Control LevelNo. ofTestedNo. of Negative
AMPn=30COCn=30THCn=30OPI300n=30METn=30PCPn=30No. ofNeg.No. ofPreliminary
01803030303030301800
-50% of the cutoff1803030303030301800
-25% of the cutoff1803030303029291782
Cut-off18018161716151411268
+25% of the cutoff18014312112168
+50% of the cutoff1800000000180
    1. Analytical Sensitivity Study Data
      Conclusion: The results support the claimed cut-off levels for each drug analyst. It also demonstrates that the performance of both Dip card and cassette product formats is consistent and equivalent.

2. Analytical Precision Study Data:

The precision study was conducted by three Operators with three batches per product format in replicates of 10 devices/lot at each control level of Negative, -50%, -25%, cut-off, +25% and +50% cutoff. The study was conducted in non-consecutive ten days using GC/MS confirmed commercial urine controls. The controls are blind coded according to a random table and evenly distributed to three operators. Every device was tested and interpreted by the same operator. The total of ten days run data is grouped into three sets: 3-day, 3-day and 4-day. A concise summary is in Table below:

Control Leveln3-Day3-Day4-DayTOAL
+-+-+-+-
Negative300801309030
-50% of cutoff3001205013030
-25% of cutoff3011101008129

AMP Test: Cutoff: 1000 ng/ml (Three lots were tested by three Operators)

{8}------------------------------------------------

CAMBARLARIAL FREELLAR RANGE LEAR BE RESEARCEHERARLARILA: A ARAGELAR A COUTEST TET TET TET TENTER TO THE STCALLER OF A RESEARCE A.C. 1988 CHARACT CHANAGE CHANAGE CHANA1Adless dead.HEREAS ASSESSMENT LANDAL P & STORES TT TENTERS TOStation Article of Children------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
status (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1)------------------------------------------------------------------------------------------------------------------------------------------------------------------------------CHARALELA LE ARRELL And versearAND NOT OFF COURSE FOR WATE-------------PAGELLEA CHARLES COLORIES OF CONSULERS OF CARE LEASE --------------------------------------------------------------------------------------------------------------------------------------------------------------------
A-4-4-4-4-4- S - S - P P + t + t ========================================================================================================================================================

Cocaine Test: Cutoff: 300 ng/ml (Three lots were tested by three Operators)

Control Leveln3-Day3-Day4-DayTOAL
+-+-+-+-
Negative3001009011030
-50% of cutoff3001108011030
-25% of cutoff3001007013030
Cutoff304644481218
+25% of cutoff3030131103264
+50% of cutoff3010080120300

MAMP Test: Cutoff: 1000 ng/ml(Three lots were tested by three Operators)

Control Leveln3-Day3-Day4-DayTOAL
+-+-+-+-
Negative3008011011030
-50% of cutoff3001108011030
-25% of cutoff3001019010129
Cutoff306634471317
+25% of cutoff306070161291
+50% of cutoff3070110120300

OPI (MOR) Test: Cutoff: 300 ng/ml (Three lots were tested by three Operators)

Control Leveln3-Day3-Day4-DayTOAL
+-+-+-+-
Negative3001107012030
-50% of cutoff3001201107030
-25% of cutoff300408117129
Cutoff305632771515
+25% of cutoff3071110101282
+50% of cutoff3080120100300

PCP Test: Cutoff: 25 ng/ml (Three lots were tested by three Operators)

Control Leveln3-Day3-Day4-DayTOAL
+-+-+-+-
Negative300809013030
-50% of cutoff3001307010030
-25% of cutoff3009010011030

{9}------------------------------------------------

CONSULTIVE CONSTITUTION COLLEGION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTIстрания выпарамент продуктивные из композиции протегентали и пристемента на пристемати в малази в мести при при при при при при при при при при при при при при при при при пр
меншеншенше жетен менен менен менен менен менен менен мен мен мен мен мен мен мен мен мен мен мен мен мен мен мен мен мен мен мен мен мен мен кенен мен кенен керек мен кезен+25% of cutoffA . GLAA ARAA LILA ALL A PART PER TELL LET TE LE BE TE TERRETARABARA ISLARA ALA USURE ARA ALL A STANTA A CONTRACT THEMANAGE A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
50% of cutoff10000I--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Children & Life on A------------------------------------------------------------------------------------------------------------------------------------------------------------------------------CARANA ALLA
Control Leveln3-Day3-Day4-DayTOAL
+-+-+-+-
Negative3003010017030
-50% of cutoff3007012011030
-25% of cutoff300906015030
Cutoff303742591218
+25% of cutoff3013112121273
+50% of cutoff3011070120300

THC Test: Cutoff: 50 ng/ml (Three lots were tested by three Operators)

The study raw data are enclosed in the ATTACHMENT E for reference.

Conclusion: The results of the reproducibility study demonstrate that the Chemtrue® Drug Screen tests are able to produce consistent results from lot-to-lot, Operator-to-Operator and dayto-day runs, during the repeated assays and measurements. It also demonstrates an equivalent assay performance between Dip Card and Cassette formats

    1. Analytical Accuracy Study: The accuracy of the Chemtrue Drug Screen Test device was evaluated with the confirmed GC/MS values in the blind-labeled clinical specimen correlation study (a total of 673 clinical specimens). The results of these studies demonstrate Chemtrue® Single/ Multi-panel Drug Screen Test to be substantially equivalent to the performance characteristics of GC/MS methodology. Correlation studies, using clinical specimens, produced a > 98% total correlation when compared to the GC/MS methodology. Also the accuracy study data of Chemtrue® devices demonstrates the substantial equivalency to the data in the package insert of the predicate kit – ACON (INNOVACON) One Step Single and Multi Drug Screen test card. For details, refer to the "Method comparison (Accuracy) study, section Performance Characteristics" of this submission.
    1. Analytical Specificity Study: The data is summarized in Section "Performance Characteristics, Specificity" and the raw data is enclosed in ATTACHMENT E of this submission.
    1. Stability study: Three lots of each product format: Dip Card and Cassette were included in the stability study, and accelerated stability at 60°C and 40°C data supported two years shelf-life of the products. Accelerated 40°C and real-time stability studies are still on-going. The copy of the . raw data is enclosed in ATTACHMENT G.

Substantial Equivalency:

{10}------------------------------------------------

Chemtron Biotech, Inc.'s Chemtrue® Single/Multi-panel Drug Screen Cassette and Dip Card tests are similar to other FDA-cleared devices for the qualitative detection of drugs of abuse tests: Amphetamine, Cocaine, Marijuana, Opiates, Methamphetamine and Phencyclidine test from other manufacturers such as ACON (Current INNOVACON), Branan Medical, etc. All of these products are based on the same principle, same technological characteristics and similar procedures.

Product AttributeChemtrue® Drug Screen TestsPredicate KitSubstantiallyEquivalent
AnalytesAmphetamine, Cocaine,Methamphetamine, Opiates (MOR),PCP and THCAmphetamine, Cocaine,Methamphetamine, Opiates(MOR/MOP) and THC, as well assix other drug analytesYes
Intended UseQualitative detection of drugs-of-abuse in urineQualitative detection of drugs-of-abuse in urineYes
Specimen TypeurineurineYes
Assay MethodologyLateral flow immunochromato-graphic assayLateral flow immunochromato-graphic assayYes
Assay Cut-offAmphetamine 1000ng/mLCocaine 300ng/mLMethamphetamine 1000ng/mLOPI (MOR) 300 ng/mLPCP 25 ng/mLTHC 50ng/mLAmphetamine 1000ng/mLCocaine 300ng/mLMethamphetamine 1000ng/mLOPI (MOR) 300 ng/mLPCP 25 ng/mLTHC 50ng/mLYes
Time to read5 minutes5 minutesYes
Indications of UsePrescriptionPrescriptionYes
Analytical Accuracy>96%>90%Yes

Device Comparison:

DIFFERENCES
AttributeChemtrue® DevicePredicate Kit
Pre-treatment for urine specimenDoes not require pre-treatment for urine specimen. It is better than the Predicate kit.Urine specimen needs to be centrifuged, filtered, or allowed to settle to obtain clear specimen for testing.
Time Frame of Result StabilizationDo not read after 8 minutes.Results remain stable for up to 4 hours after test initiation.

DISCUSSION AND CONCLUSION:

Based on the technological characteristics/principle, features of the device designs, test specimen matrix, test method and performance characterizations, as the set forth above, it can be concluded that Chemtrue® Single/ Multi-panel Drug Screen tests are substantially equivalent to the predicate

{11}------------------------------------------------

kit ACON's (INNOVACON) One Step Single and Multi Drug Screen test card product presently distributed commercially.

The difference of the read time claims between Chemtrue® device and the predicate kit does not impact the result interpretation by the intended user(s), since the device labeling clearly indicates the required "Read Time". The device safety and effectiveness won't be affected either. For details, refer to the section "Substantial Equivalence Discussion" of this submission.

Proposed Labeling

  1. Package Labeling - Copies of the final draft package labeling for the Chemtron Biotech, Inc.'s Chemtrue® Single/ Multi-panel Drug Screen Cassette and Dip Card tests are in Attachment B and C. It is our intention that final package labeling will be substantially similar to the draft labeling attached.

  2. Package Insert – A copy of the final draft package insert is enclosed in ATTACHMENT D. The final package insert will be substantially similar to that enclosed.

  3. Promotional Material - Promotional material and advertising materials, when prepared, will be consistent with the product package insert and this 510(k) pre-market notification submission.

Summarized By: Jane Zhang

Date: August 3rd, 2010

{12}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Chemtron Biotech, Inc. c/o Jane Zhang Official FDA Correspondent 8370 Juniper Creek Lane Suite 1-2 San Diego, CA 92126

JAN 20 201

Re: Re: K102203 Trade/Device Name: Chemirue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests
Requlation Number 21 CFD 860 2140 Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine Test System Regulatory Class: II Product Code: DKZ, DIO, DJC, DNK, LCM, LDJ Dated: January 10, 2011 Received: January 10, 2011

Dear Ms. Zhang,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical in microadcon in micrisate conninents, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, or and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, RDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

{13}------------------------------------------------

Page 2 –

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

G.C.

Courmey Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{14}------------------------------------------------

Indications for Use

510(k) Number (if known): K102203

Device Name: Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests

Indications for Use:

The Chemtron Biotech, Inc.'s Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are rapid chromatographic immunoassays for the qualitative detection of up to six of the following drugs in a variety of combinations in human urine. The designed cutoff concentrations and direct calibrator for these drugs are as follows:

AnalyteAbbreviationCalibratorCutoff
Concentration
AmphetamineAMPd-Amphetamine1000 ng/ml
CocaineCOCBenzoylecgonine300 ng/ml
MarijuanaTHC11-nor- $\Delta^9$ -THC9-COOH50 ng/ml
MethamphetamineMETd-Methamphetamine1000 ng/ml
OpiatesOPI/MORMorphine300 ng/ml
PhencyclidinePCPPhencyclidine25 ng/ml

The Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card are intended for qualitative detection of drugs of abuse for health care professionals, in vitro diagnostic use and prescription use ONLY. It is not intended for point-of-care settings or over the counter use.

This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods.

Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Divisions Sign Off

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K102203

Page 1 of 1

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).