K Number
K111322

Validate with FDA (Live)

Date Cleared
2011-09-14

(126 days)

Product Code
Regulation Number
862.3170
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Chemtron Biotech, Inc.'s Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of up to six of the following drugs in a variety of combinations in human urine. The designed cutoff concentrations and the calibrators used for these drugs are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration
BenzodiazepinesBZOOxazepam300 ng/mL
BarbituratesBARSecobarbital/Pentobarbital300 ng/mL
EcstasyMDMA/XTCd,I-Methylenedioxymethamphetamine500 ng/mL
MethadoneMTDMethadone300 ng/mL
OpiatesOPI/MORMorphine2000 ng/mL
OxycodoneOXYOxycodone100 ng/mL

The Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are intended for the qualitative detection of drugs of abuse for in vitro diagnostic and prescription use ONLY. They are not intended for point-of-care settings or over the counter use. These assays provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods.

Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

Device Description

The Drugs of Abuse (DOA) Screen Panels are one-step lateral flow immunoassays in which chemically modified drugs (drug-protein conjugates) compete for limited antibody binding sites with drugs that may be present in urine. The test device consists of up to six test strips placed into separate panels of a plastic holder. On each test strip, a drug-protein conjugate is striped on the test band of the membrane - known as the test region (T) and the anti-drug antibody-colloidal gold conjugate pads are placed at the forward end of the membrane.

AI/ML Overview

The Chemtrue® Single/Multi-Panel Drug Screen Tests are qualitative immunoassays designed to detect up to six drugs of abuse in human urine. The acceptance criteria for this device are defined by its performance in detecting specific drugs at or above their established cutoff concentrations, as compared to a reference method (GC/MS).

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the "Overall Agreement" percentages with the GC/MS reference method. While explicit pass/fail thresholds are not stated here, the reported high agreement percentages indicate that the device meets an acceptable level of performance for clinical utility.

TestAcceptance Criteria (Implied)Reported Device Performance (Dip Card)Reported Device Performance (Cassette)
BZOHigh agreement with GC/MS98.5%98.5%
BARHigh agreement with GC/MS98.4%98.4%
MDMAHigh agreement with GC/MS98%98%
MTDHigh agreement with GC/MS99%99%
OPI(MOR)2000High agreement with GC/MS100%100%
OXYHigh agreement with GC/MS99.1%99.1%

2. Sample Sizes Used for the Test Set and Data Provenance

The studies used clinical specimens for the test set.

  • BZO: 203 samples (Dip Card and Cassette)
  • BAR: 190 samples (Dip Card and Cassette)
  • MDMA: 101 samples (Dip Card and Cassette)
  • MTD: 104 samples (Dip Card and Cassette)
  • OPI(MOR)2000: 106 samples (Dip Card and Cassette)
  • OXY: 106 samples (Dip Card and Cassette)

The data provenance is from clinical specimens, implying a retrospective study design where samples were collected and then tested. The document does not specify the country of origin.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The ground truth was established by Gas Chromatography/Mass Spectrometry (GC/MS), which is an analytical chemical method, not human expert consensus.

4. Adjudication Method for the Test Set

This information is not applicable as the ground truth was established by an objective chemical method (GC/MS), not human adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This is not applicable. The device is an immunoassay for drug screening, not an AI-assisted diagnostic tool for human readers. Therefore, an MRMC study and analysis of human reader improvement with AI assistance are outside the scope of this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the performance study (method comparison) represents a standalone evaluation of the device. The Chemtrue® cassettes and dip cards were compared directly against the GC/MS reference method, acting as an "algorithm only" in the sense that its output (positive/negative) was assessed independently of human interpretation influence beyond reading the test lines.

7. The Type of Ground Truth Used

The ground truth used was analytical/objective data from Gas Chromatography/Mass Spectrometry (GC/MS). The document states: "Samples with drug concentration above the cut-off level were considered presumptive positive and concentration below the cut-off were considered negative." and "The product performance characteristics of were evaluated with the GC/MS confirmed values in a blind-labeled clinical specimen correlation study."

8. The Sample Size for the Training Set

The document does not provide information regarding a distinct training set. The device is a lateral flow immunoassay, a chemical sensing technology, and not typically developed using machine learning models that require explicit training sets in this context.

9. How the Ground Truth for the Training Set Was Established

As no training set is described for this type of device, this information is not applicable. The device's performance characteristics (such as antibody specificity and cutoff concentrations) are established during the assay development and manufacturing process, not through a 'training set' in the machine learning sense.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "chemtron" in a stylized font. The "c" is enclosed in a partial circle, and there is a line above the "em" portion of the word. The text is in black and the background is white. The font appears to be a serif font.

CHEMTRON BIOTECH, INC 8370 Juniner Creek Lane, 8 AN DIEGO. CA 92126 (858) 530-2868 AX (858) 530-2878 ww.uschemtronbio.co

SEP 1 4 2011

510(k) Summary

AS REQUIRED BY 21 CFR 807.92(c)

The Assigned 510(k) number is K111322

Date of Summary: September 11th, 2011

Common Name: Drugs of Abuse Screening Tests

Classification Name: Immunoassay for the detection of drugs of abuse

Trade/Proprietary Name:

Chemtron Biotech, Inc.'s Chemtrue® Single / Multi-Panel Drug Screen Dip Card / Cassette Tests, contain 1 to 6 of the following DOA test(s) in each device:

    1. Benzodiazepines (BZO) test strip
    1. Barbiturates (BAR) test strip
    1. Ecstasy (MDMA/XTC) test strip
    1. Methadone (MTD) test strip
    1. Opiates /Morphine (OPI/MOR/MOP) cut-off: 2000ng/mL test strip
    1. Oxycodone (OXY) test strip

Owner:

Ellen Liu President, Chemtron Biotech, Inc. 8370 Juniper Creek Lane, #1-2 San Diego, CA 92126 TEL: (858) 530-2868 FAX: (858) 530-2878

Contact Person:

Jane Zhang, Director of QA/RA Official FDA Correspondent 8370 Juniper Creek Lane. #1-2 San Diego, CA 92126 Office: (858) 530-2868 Mobile: (858) 997-7472 FAX: (858) 530-2878

{1}------------------------------------------------

Substantial Equivalency:

The Chemtrue® Single/Multi-Panel Drug Screen Test is substantially equivalent to other tests currently on the market.

Test AnalytePredicate Device NamePredicate Device510(k) #
BenzodiazepinesACON (Alere) One Step Drug Screen Test CardK061718
BarbituratesACON (Alere) One Step Drug Screen Test CardK061718
Ecstasy (MDMA)ACON (Alere) One Step Drug Screen Test CardK061718
MethadoneACON (Alere) One Step Drug Screen Test CardK061718
Opiates (Morphine)2000ACON (Alere) One Step Drug Screen Test CardK061718
OxycodoneACON (Alere) One Step Drug Screen Test CardK061718

The predicate kit package insert is enclosed in ATTACHMENT A of this submission.

Proposed Labeling or Promotional Material for the Device:

A description of the device can be found in the attached proposed labeling, including an explanation of how the device functions, technical principle and concepts that form the basis for the device, as well as the physical and performance characteristics of the device. such as the device design, materials used and physical properties. In accordance with FDA labeling requirements (21 CFR 809.10), enclosed are the draft copies of product labeling including copies of the technical product inserts. See ATTACHMENT B. C and D.

Intended Use

The Chemtrue® test device is intended for the qualitative detection of drugs of abuse, for in vitro diagnostic use and for prescription use ONLY.

The test provides only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods.

Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

Technological Characteristics and Scientific Principles

The Drugs of Abuse (DOA) Screen Panels are one-step lateral flow immunoassays in which chemically modified drugs (drug-protein conjugates) compete for limited antibody binding sites with drugs that may be present in urine. The test device consists of up to six test strips placed into separate panels of a plastic holder. On each test strip, a drug-protein

{2}------------------------------------------------

conjugate is striped on the test band of the membrane - known as the test region (T) and the anti-drug antibody-colloidal gold conjugate pads are placed at the forward end of the membrane. In the absence of drugs in the urine, the solution of the colored antibodycolloidal gold conjugates moves along with the sample solution by capillary action across the membrane to the immobilized drug-protein conjugate zone on the test band region. The colored antibody-gold conjugates then complexes with the drug-protein conjugates to form visible lines. Therefore, the formation of the visible precipitant in the test band occurs when the test urine is negative for the drug. If any of the subject drugs of abuse is present in the urine, the drug/metabolite antigen competes with drug-protein conjugates on the test band region for the limited antibody on the colored drug antibody-colloidal gold conjugate pad. When a sufficient amount of drug is present in the urine, the drug will saturate the limited antibody binding sites and the colored antibody -colloidal gold conjugate cannot bind to the drug-protein conjugate at the test region of the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result.

A control band with a different antigen/antibody reaction is added to the membrane strip at the control region (C) to indicate that the test has performed properly. This control line is manufactured as a built-in internal control of the test device and should always appear regardless of the presence of drug or metabolite. If the control line does not appear the test should be discarded. The presence of this colored band in the control region also serves: 1) as verification that adequate specimen volume is added, or no flow, due to insufficient urine volume, 2) to indicate that the test device is properly functioning, and 3) as an internal quality control of the system.

Summary of Device Similarities and Differences

Chemtron Biotech, Inc.'s Chemtrue Single/Multi-Panel Drug Screen Cassette and Dip Card tests are similar to other FDA-cleared devices for the qualitative detection of drugs of abuse: Benzodiazepines, Barbiturates, Ecstasy (MDMA), Opiates (Morphine) 2000, Methadone and Oxycodone test from other manufacturers such as ACON (Alere), Alfa Scientific Designs, Inc. All of these products are based on the same technological characteristics, scientific principle and similar procedures. The similarities and differences among these tests are summarized as follows:

SIMILARITIES
ItemChemtrue® DevicePredicate Kit
Intended Use/Indications for useQualitative detection of drugs-of- abuse in urinefor Prescription, In Vitro Diagnostic Use OnlySame
SpecimenUrineSame
Technological Characteristics andPrincipleOne-Step lateral flow competitiveImmunoassaySame

{3}------------------------------------------------

Positive result1 colored lineSame
Device Design/PerformanceNegative result2 colored linesSame
Detection reagentColloidal goldSame
AccuracyAssessmentConfirm with GC/MS reference methodSame
Cut-offBenzodiazepines 300 ng/mLSame
Barbiturates 300 ng/mLSame
Ecstasy (MDMA) 500 ng/mLSame
Methadone 300 ng/mLSame
Opiates(Morphine) 2000 ng/mLSame
Oxycodone 100 ng/mLSame
Safety and PrecautionAll urine specimens should be consideredpotentially hazardous and handled in the samemanner as infectious agent.Same
Read time5 minutesSame
Storage2-30°C (36-86°F)Same
DIFFERENCES
ItemChemtrue® DevicePredicate Kit
Pre-treatment for urinespecimenDoes not require pre-treatment forurine specimen. It is ease of use.Urine specimen needs to be centrifuged,filtered, or allowed to settle to obtain clearspecimen for testing.
Time To Read ResultDo not read after 8 minutes.Results remain stable for up to 4 hoursafter test initiation.

DISCUSSION AND CONCLUSION:

Based on the technological characteristics/assay principle, features of the device designs, test specimen matrix, test method and performance characterizations, as the set forth above, it can be concluded that Chemtrue® Single/ Multi-Panel Drug Screen tests are substantially equivalent to the predicate kit ACON's (INNOVACON/Alere) One Step Single and Multi Drug Screen test card product which is currently distributed commercially.

The Chemtrue® Drug Screen Tests do not require pre-treatment for urine specimen, thereby making it much simpler and convenient for the intended user. With regarding to the difference of Time to read result, since the device labeling clearly indicates the required "Read Time", it does not impact the result interpretation by the intended user.

{4}------------------------------------------------

Performance Data:

Chemtron Biotech, Inc. has reviewed the requirements of Section 514 of the Act, which states that to date no performance standards has been established for drug screen test systems by the FDA.

However, the studies listed in the notification are conducted according to "The Draft Guidance for Industry and FDA Staff' - " Premarket Submission and Labeling Recommendations for Drugs of Abuse Screening Tests, issued on: December 2, 2003", including the design of draft labeling and package inserts.

Chemtrue® Single / Multi-Panel Drug Screen Cassette and Dip Card tests are one-step, lateral flow, colloidal gold based immunoassays for the rapid, qualitative detection of Benzodiazepines, Barbiturates, Ecstasy (MDMA), Opiates (Morphine) 2000, Methadone and Oxycodone in human urine. They are intended for Professional Use and Prescription Use ONLY.

For BAR or BZO test, it may vield preliminary positive results when BAR or BZO is ingested at or above therapeutic doses. There are no uniformly recognized drug levels for barbiturate and benzodiazepine in urine. The Chemtrue® Single / Multi-Pane drugs of abuse urine test device shows the drug was or was not present at the cutoff level.

The assay provides a preliminary test result. A more specific alternate chemical method must be used to obtain a confirmed assay result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory result. Clinical and professional judgment must be applied to any drugs of abuse test result, particularly when preliminary positive results are obtained.

Chemtrue® Single /Multi-Panel Drug Screening Tests vs GC/MS Reference Method

Samples with drug concentration above the cut-off level were considered presumptive positive and concentration below the cut-off were considered negative.

Table 1. The data are a summary from a method comparison (Accuracy) study of
Chemtrue® Drug Screen Dip Card Test results versus GC/MS values of the
clinical specimens
TestPositive AgreementNegative AgreementOverall Agreement
BZO116/118 = 98.3%84/85 = 98.8%200/203 = 98.5%
BAR103/105 = 98.1%84/85 = 98.8%187/190 = 98.4%
MDMA40/41 = 97.6%59/60 = 98.3%99/101 = 98%

{5}------------------------------------------------

MTD44/44 = 100%59/60 = 98.3 %103/104 = 99%
OPI(MOR)200041/41= 100%65/65= 100%106/106= 100%
OXY47/47 = 100%58/59= 98.3%105/106= 99.1%

Table 2. The data are a summary of a method comparison (Accuracy) study of Chemtrue® Drug Screen Cassette Test results versus the GC/MS values of the clinical specimens

TestPositive AgreementNegative AgreementOverall Agreement
BZO116/118 = 98.3%84/85= 98.8%200/203 = 98.5%
BAR103/105= 98.1%84/85= 98.8%187/190= 98.4%
MDMA40/41= 97.6%59/60 = 98.3%99/101 = 98%
MTD44/44 = 100%59/60 = 98.3 %103/104 = 99%
OPI(MOR)200041/41= 100%65/65= 100%106/106= 100%
OXY47/47 = 100%58/59= 98.3%105/106= 99.1%

Conclusion:

The product performance characteristics of were evaluated with the GC/MS confirmed values in a blind-labeled clinical specimen correlation study. The results of these studies demonstrate Chemtrue® Single/ Multi-Panel Drug Screen Test to be substantially in agreement with the GC/MS assigned values.

Additionally, the Chemtrue® Single/ Multi-Panel Drug Screen tests were shown to be equivalent, when compared to the data in the package insert of the predicate kit ACON (INNOVACON/Alere) One Step Drug Screen test card. These results demonstrated that Chemtrue® Single/ Multi-panel Drug Screen cassette and Dip Card are safe and effective in detecting Benzodiazepines, Barbiturates, Ecstasy (MDMA), Methadone, Opiates (Morphine) 2000 and Oxycodone in human urine.

Other non-clinical performance data, such as cutoff characteristics, precision (Reproducibility), specificity (Cross-reactivity and interference) and stability studies are summarized in the Section labeled "Performance Characteristics" and the raw data are enclosed in ATTACHMENT E and ATTACHMENT H of this submission.

510(k) Summary was revised By: Jane Zhang on September 110, 2011

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/6/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The eagle is composed of three curved lines that suggest feathers. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.

Chemtron Biotech. Inc. c/o Jane Zhang 8370 Juniper Creek Lane Suite 1-2 San Diego, CA 92126

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

SEP 1 4 2011

Re: k!11322 Trade/Device Name: Chemtrue Single/Multi-Panel Drug Screen Casseue and Dip Card Tests Regulation Number: 21 CFR 862.3170 Regulation Name: Benzodiazepine test system Regulatory Class: Class H Product Code: JXM. DIS. DJC. DJR. DJG Dated: August 18. 2011 Received: August 19. 2011

Dear Ms Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class 111 (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{7}------------------------------------------------

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Parl 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its liternet address http://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm.

Sincerely yours,

jz

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use

510(k) Number (if known): K111322

Device Name: Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests

Indications for Use:

The Chemtron Biotech, Inc.'s Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of up to six of the following drugs in a variety of combinations in human urine. The designed cutoff concentrations and the calibrators used for these drugs are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration
BenzodiazepinesBZOOxazepam300 ng/mL
BarbituratesBARSecobarbital/Pentobarbital300 ng/mL
EcstasyMDMA/XTCd,I-Methylenedioxymethamphetamine500 ng/mL
MethadoneMTDMethadone300 ng/mL
OpiatesOPI/MORMorphine2000 ng/mL
OxycodoneOXYOxycodone100 ng/mL

The Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are intended for the qualitative detection of drugs of abuse for in vitro diagnostic and prescription use ONLY. They are not intended for point-of-care settings or over the counter use. These assays provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods.

Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Asso

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K111322

§ 862.3170 Benzodiazepine test system.

(a)
Identification. A benzodiazepine test system is a device intended to measure any of the benzodiazepine compounds, sedative and hypnotic drugs, in blood, plasma, and urine. The benzodiazepine compounds include chlordiazepoxide, diazepam, oxazepam, chlorzepate, flurazepam, and nitrazepam. Measurements obtained by this device are used in the diagnosis and treatment of benzodiazepine use or overdose and in monitoring levels of benzodiazepines to ensure appropriate therapy.(b)
Classification. Class II (special controls). A benzodiazepine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).