K Number
K142396

Validate with FDA (Live)

Date Cleared
2014-10-09

(43 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Chemtrue® BUP/TCA Single/Multi-Panel Drug Screen Dip Card /Cup/Cassette Tests are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine (BUP) and/or Tricyclic Antidepressants (TCA) drugs in human urine. The tests are intended for prescription and Over-The-Counter (OTC) use. The tests provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to distinguish between prescription use or abuse of Buprenorphine and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for Buprenorphine and Tricvclic Antidepressants in urine.

The Chemtrue® Multi-Panel Drug Screen Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Marijuana, Morphine, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates, Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The tests are intended for prescription and Over-The-Counter (OTC) use. The tests provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

The Chemtrue® Multi-Panel Drug Screen Dip Card with OPI 2000 Tests are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Marijuana. Opiates 2000. Methamphetamine. Phencyclidine. Benzodiazepines. Barbiturates. Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The tests are intended for prescription and Over-The-Counter (OTC) use. The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

Device Description

The Chemtrue® Drug Screen Tests are colloidal gold based lateral flow immunoassays for the rapid, qualitative detection of drugs of abuse in human urine. The tests are single-use, in vitro diagnostic devices, which come in the formats of dip card, cup, or cassette, as indicated by the test name.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the Chemtrue® BUP/TCA Single/Multi-Panel Drug Screen Dip Card/Cup/Cassette Tests, based on the provided text.

Acceptance Criteria and Device Performance

The acceptance criteria for this device are demonstrated through its performance in various studies, primarily focusing on reproducibility (precision) and method comparison (accuracy) against GC/MS (or LC/MS for some lay-user studies). The device aims to provide qualitative detection of drugs of abuse in human urine.

Acceptance Criteria (Implicit from Performance Studies):
The device is expected to:

  • Consistently produce negative results for samples with no drug present or drug concentrations significantly below the cutoff (e.g., 50% of cutoff).
  • Consistently produce positive results for samples with drug concentrations significantly above the cutoff (e.g., 125%, 150%, 200% of cutoff).
  • Show predictable behavior around the cutoff, with a mix of positive and negative results when concentrations are at or near the cutoff (e.g., 75% of cutoff, cutoff).
  • Demonstrate high agreement with confirmed analytical methods (GC/MS or LC/MS).
  • Be free from interference from common substances/drugs.
  • Maintain performance across a range of urine pH and specific gravity.
  • Have a stable shelf-life.
  • Be easily understood and used by lay-users (for OTC versions).

Table of Acceptance Criteria and Reported Device Performance (BUP & TCA only, as they are the new analytes):

Study TypePerformance MetricAcceptance Criteria (Implicit)Reported Device Performance (Example for BUP Dip Card)
Reproducibility (Precision)Negative Samples: 100% negative results significantly below cutoff. Positive Samples: 100% positive results significantly above cutoff. Near Cutoff: Demonstrates consistent performance, with an expected increase in positive results as concentration approaches and exceeds cutoff.High consistency in results across different operators, lots, and days, especially for samples significantly below and above the cutoff.BUP Dip Card: - Negative (30 samples): 0 (+) / 30 (-) - 50% Cutoff (30 samples): 0 (+) / 30 (-) - 75% Cutoff (30 samples): 0 (+) / 30 (-) - Cutoff (30 samples): 10 (+) / 20 (-) - 125% Cutoff (30 samples): 30 (+) / 0 (-) - 150% Cutoff (30 samples): 30 (+) / 0 (-) TCA Dip Card: - Negative (30): 0 (+)/30 (-); 50% cutoff (30): 0 (+)/30 (-); 75% cutoff (30): 0 (+)/30 (-); Cutoff (30): 1 (+)/29 (-); 125% cutoff (30): 30 (+)/0 (-); 150% cutoff (30): 30 (+)/0 (-)
SpecificityCross-reactivity: No significant false positives from non-target compounds. Known Cross-reactants: Identified and quantified cross-reactivity for structurally similar compounds.Low or no cross-reactivity with common substances/drugs at typical concentrations, and expected cross-reactivity with known related compounds.Buprenorphine (BUP): Norbuprenorphine 10 ng/mL (100% cross-reactivity). Tricyclic Antidepressants (TCA): Nortriptyline 1000 ng/mL (100% cross-reactivity), Amitriptyline 1000 ng/mL (100%), Desipramine 300 ng/mL (333%), Imipramine 50 ng/mL (2000%), etc. Interference: Over 100 potential interferents tested and found not to cross-react at 100 µg/mL at ±25% of BUP and TCA cut-off.
Method Comparison (Accuracy)Agreement with GC/MS: High percentage agreement for both positive and negative results when compared to the gold standard GC/MS. Discordant Results: Limited number of discordant results, primarily near the cutoff concentration.High sensitivity and specificity demonstrated by high agreement (e.g., >95-100%) with GC/MS for samples well above or below the cutoff.BUP Dip Card: - Agreement for Positives: 100% (based on 33 samples: 2 near-cutoff neg, 6 near-cutoff pos, 25 GC/MS pos) - Agreement for Negatives: 96.3% (based on 52 samples: 40 no drug, 7 GC/MS neg, 5 near-cutoff neg) TCA Dip Card: - Agreement for Positives: 100% (based on 34 samples: 0 near-cutoff neg, 8 near-cutoff pos, 26 GC/MS pos) - Agreement for Negatives: 100% (based on 55 samples: 40 no drug, 6 GC/MS neg, 9 near-cutoff neg) Discordant: 2 discordant results for Buprenorphine (9.5 and 9.8 ng/mL GC/MS value, but device read positive).
Effect of Urine pH & Specific GravityConsistent performance across physiological ranges.Test performance unaffected by urine pH 3.0 to 8.5 and specific gravity 1.002 to 1.030.
Lay-user Studies (OTC)Agreement with LC/MS: High percentage agreement with LC/MS across various drug concentrations. Ease of Use: High percentage of users find instructions easy to follow.High accuracy when used by lay-users; clear and understandable instructions.BUP Lay-user (Dip Card): 100% agreement across all tested concentrations (no drug, 50%, 75%, 125%, 150%, 200% of cutoff). TCA Lay-user (Dip Card): 100% agreement except for 75% of cutoff (1 positive/29 negative, 96.7% agreement). Ease of Use: 99% of lay-users found the instructions easy to follow. Flesch-Kincaid reading grade level < 7.

Study Details

  1. Sample size used for the test set and the data provenance:

    • Reproducibility (Precision) Studies:
      • Test Set Size: For each analyte (BUP and TCA) and each format (Dip Card, Cup, Cassette), there were 6 concentration levels (Negative, 50% cutoff, 75% cutoff, Cutoff, 125% cutoff, 150% cutoff). Each level was tested in 10 replicates across 3 lots and 3 operators.
      • Total Samples per Analyte-Format Combination: 6 concentrations * 10 replicates = 60 samples. Since 3 lots were used, this implies 180 tests per concentration level total (60 per lot). The tables show n=30 per concentration level, which suggests 30 replicates per concentration (10 replicates x 3 lots total, assigned to 3 operators).
      • Data Provenance: The samples were "GC/MS confirmed drug spiked samples." This implies they were prepared in a lab, not directly from patients, but confirmed by a reference method. The country of origin of the data is not specified, but the submission is to the US FDA.
    • Method Comparison Studies (Accuracy):
      • Test Set Size: A total of 174 clinical specimens were used for BUP/TCA combined across all formats. Given three formats (Dip Card, Cup, Cassette) and two analytes (BUP, TCA), this number of specimens would have been divided among these. The individual tables for each format show ~50-60 samples per analyte per format for calculation of agreement percentages across different concentration categories (e.g., for BUP Dip Card, agreement for negative was based on 52 samples, and for positive on 33 samples).
      • Data Provenance: "blind-labeled clinical specimen correlation study". This suggests retrospective clinical samples, though the country of origin is not specified.
    • Lay-user Studies (OTC):
      • Test Set Size: A total of 300 people participated. Each lay-user was given "up to two blind labeled samples and test device(s)".
      • Total Samples: The tables show 30 replicates for each of 6 concentration levels for both BUP and TCA and for each of the three device formats. This implies 6 levels * 30 replicates = 180 total tests for BUP, and 180 for TCA, for each device format in the lay-user study.
      • Data Provenance: "LC/MS confirmed urine samples...by spiking drugs into drug-free urine pool." This means the samples were prepared in a lab and confirmed by LC/MS, not directly from patients. The study was conducted at "three (3) intended user sites," implying these were likely in a controlled setting for observational use. Country of origin not specified.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • For the Reproducibility (Precision) Studies and Method Comparison Studies, the ground truth was established by GC/MS (Gas Chromatography/Mass Spectrometry). This is a highly accurate and generally accepted reference method for drug confirmation. The document does not specify human experts for establishing GC/MS results or their qualifications.
    • For the Lay-user Studies, the ground truth was established by LC/MS (Liquid Chromatography/Mass Spectrometry). Similar to GC/MS, this is a highly accurate analytical method, and human expert qualifications for interpreting these results are not specified.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • No adjudication method (like 2+1 or 3+1 expert review) is mentioned for establishing the ground truth from GC/MS or LC/MS. These instrumental methods provide quantitative results that are then categorized as positive or negative based on the cutoff concentration. The "agreement" between the device and the GC/MS/LC/MS result serves as the primary metric.
    • For the reproducibility study, three operators performed the testing, suggesting a comparison of their readings rather than an adjudication process for ground truth.
    • For the method comparison study, "Three operators performed the testing. Each operator tested one device format with one unique set of blind coded samples." This points to individual operator readings against the GC/MS ground truth, not an adjudication process among operators to define the ground truth.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This device is a rapid lateral flow immunoassay (a diagnostic test kit), not an AI-assisted diagnostic device. The "readers" are the individuals interpreting the visible lines on the dip card/cup/cassette. The "lay-user study" involves human interpretation, but it's not framed as a comparative effectiveness study with vs. without AI assistance.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This is a diagnostic test kit with a visual readout, where a human interprets the presence or absence of lines. Therefore, there is no "algorithm only" or "standalone" performance in the context of AI. The device itself is standalone in that it provides a result without additional complex equipment beyond the collection of urine. However, the interpretation is human-in-the-loop.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The primary ground truth used was analytical confirmation by Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS). These are considered the gold standard for confirming the presence and concentration of drugs in urine.
  7. The sample size for the training set:

    • The document describes performance studies for new analytes (BUP and TCA) and mentions that existing analytes were cleared under previous 510(k)s. It does not explicitly state a "training set" size in the context of machine learning or algorithm development. These are immunoassay test kits, and their "training" is in their design, manufacturing, and calibration processes. The presented data represents validation studies, which are analogous to a test set in machine learning.
  8. How the ground truth for the training set was established:

    • As there isn't a "training set" in the machine learning sense for this type of device, the concept of establishing ground truth for it doesn't directly apply. The immunoassay is designed based on known chemical reactions and antibody-antigen binding. The validation studies (reproducibility and method comparison) then verify its performance against the analytically confirmed presence or absence of drugs (GC/MS/LC/MS as ground truth).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 9, 2014

CHEMTRON BIOTECH, INC. JANE ZHANG QA/RA DIRECTOR 9245 BROWN DEER ROAD, SUITE B SAN DIEGO CA 92121

Re: K142396

Trade/Device Name: Chemtrue® BUP/TCA Single/Multi-Panel Drug Screen Dip Card/Cup/Cassette Tests: Chemtrue® Multi-Panel Drug Screen Dip Card Tests; Chemtrue® Multi-Panel Drug Screen Dip Card with OPI 2000 Tests Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: II Product Code: DJG, LFG Dated: August 27, 2014

Received: August 28, 2014

Dear Ms. Jane Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K142396

Device Name

Chemtrue® BUP / TCA Single/Multi-Panel Drug Screen Dip Card / Cup / Cassette Tests Chemtrue® Multi-Panel Drug Screen Dip Card Tests Chemtrue® Multi-Panel Drug Screen Dip Card with OPI 2000 Tests

Indications for Use (Describe)

Chemtrue® BUP / TCA Single/Multi-Panel Drug Screen Dip Card / Cup / Cassette Tests:
The Chemtrue® BUP/TCA Single/Multi-Panel Drug Screen Dip Cassette Tests are rapid lateral flow
immunoassays for the qualitative detection of Buprenorphine (BUP) and/or Tricyclic Antidepressants (TCA) drugs in
human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
BuprenorphineBUPBuprenorphine10
Tricyclic AntidepressantsTCANortriptyline1000

The single and multi-panel tests are available in Dip Card, Cup and Cassette formats. The tests are intended for prescription and Over-The-Counter (OTC) use. The tests provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to distinguish between prescription use or abuse of Buprenorphine and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for Buprenorphine and Tricyclic Antidepressants in urine.

Chemtrue® Multi-Panel Drug Screen Dip Card Tests:

The Chemtrue® Multi-Panel Drug Screen Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Marijuana, Morphine, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates, Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
BuprenorphineBUPBuprenorphine10
Tricyclic AntidepressantsTCANortriptyline1000
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMP/METd-Methamphetamine1000
MorphineMORMorphine300
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ9-THC9-COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue® Multi-Panel Drug Screen Dip Card Tests panel can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a

{3}------------------------------------------------

confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

Chemtrue® Multi-Panel Drug Screen Dip Card with OPI 2000 Tests:

The Chemtrue® Multi-Panel Drug Screen Dip Card with OPI 2000 Tests are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Marijuana, Opiates 2000, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates, Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration (ng/mL)
BuprenorphineBUPBuprenorphine10
Tricyclic AntidepressantsTCANortriptyline1000
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMP/METd-Methamphetamine1000
OpiatesOPIMorphine2000
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-A9-THC9-COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue® Multi-Panel Drug Screen Dip Card with OP1 2000 Tests can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

510(k) Summary

AS REQUIRED BY 21 CFR 807.92(c)

A. SUBMITTER: Chemtron Biotech, Inc. 9245 Brown Deer Road, Suite B, San Diego, CA

TEL: 858-450-0044

FAX: 858-450-0046

Contact Person: Jane Zhang, Director of QA/RA

Official FDA Correspondent 9245 Brown Deer Road, Suite B San Diego, CA 92121 Office: (858) 450-0044; FAX: (858) 450-0046 Email: jane@uschemtronbio.com

Date Prepared: October 08, 2014

B. DEVICE

Device Name: Chemtrue® BUP/TCA Single/Multi-Panel Drug Screen Dip Card/Cup/Cassette Tests Chemtrue® Multi-Panel Drug Screen Dip Card Tests Chemtrue® Multi-Panel Drug Screen Dip Card with OPI 2000 Tests

Common Name: Multi-Drug Urine Test Panel

Regulatory Class: Class II

Regulatory Information:

Drug of AbuseProduct CodeDevice PanelRegulation Number
Buprenorphine (BUP)DJGToxicology 9121CFR 862.3650
Tricyclic Antidepressants (TCA)LFGToxicology 9121 CFR 862.3910

This 510(k) has been submitted for clearance of new BUP/TCA drugs of abuse cup devices and to incorporate BUP/TCA analytes to previously FDA cleared Chemtrue Single/Multi-Panel Drug Screen Dip Card/Cassette Tests (K102203, K111322, K121339, and K123080). These tests have been previously cleared for prescription use (K102203 and K111322) and over-the-counter use (K121339 and K123080) and already contain the following drugs of abuse analytes:

Drug of AbuseProduct CodePanelRegulation Section
AmphetamineDKZToxicology 9121CFR 862.3100, Amphetamine Test System
CocaineDIOToxicology 9121 CFR 862.3250, Cocaine and metabolites Test System
MethamphetamineLAFToxicology 9121 CFR 862.3610, Methamphetamine Test System
OpiatesDIGToxicology 9121 CFR 862.3650, Opiate Test System
PhencyclidineLCMToxicology 91Unclassified, Enzyme immunoassay Phencyclidine
MarijuanaLDJToxicology 9121 CFR 862.3870, Cannabinoids Test System
BenzodiazepinesJXMToxicology 9121 CFR 862.3170, Benzodiazepines Test System

{5}------------------------------------------------

BarbituratesDISToxicology 9121 CFR 862.3150, Barbiturates Test System
Ecstasy (MDMA)DICToxicology 9121 CFR 862.3610, Methamphetamine Test System
MethadoneDJRToxicology 9121 CFR 862.3620, Methadone Test System
OxycodoneDJGToxicology 9121 CFR 862.3650, Opiate Test System
MorphineDNKToxicology 9121 CFR 862.3640, Morphine Test System

C. PREDICATE DEVICE:

510(k): K061718. Innovacon Spectrum II Test Card and Innovacon Spectrum II Test Card with Integrated Cups

D. INDICATONS FOR USE:

Device Name: Chemtrue® BUP / TCA Single/Multi-Panel Drug Screen Dip Card / Cup / Cassette Tests

Chemtrue® Multi-Panel Drug Screen Dip Card Tests

Chemtrue® Multi-Panel Drug Screen Dip Card with OPI 2000 Tests

Indications for Use:

Chemtrue® BUP / TCA Single/Multi-Panel Drug Screen Dip Card / Cup / Cassette Tests:

The Chemtrue® BUP/TCA Single/Multi-Panel Drug Screen Dip Card /Cup/Cassette Tests are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine (BUP) and/or Tricyclic Antidepressants (TCA) drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration
BuprenorphineBUPBuprenorphine10 ng/mL
Tricyclic AntidepressantsTCANortriptyline1000 ng/mL

The single and multi-panel tests are available in Dip Card, Cup and Cassette formats. The tests are intended for prescription and Over-The-Counter (OTC) use. The tests provide only a preliminary result. A more specific alternative chemical must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to distinguish between prescription use or abuse of Buprenorphine and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for Buprenorphine and Tricvclic Antidepressants in urine.

Chemtrue® Multi-Panel Drug Screen Dip Card Tests:

The Chemtrue® Multi-Panel Drug Screen Div Card Tests are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Marijuana, Morphine, Methamphetamine, Phencyclidine, Benzodiazepines, Barbiturates,

{6}------------------------------------------------

AnalyteAbbreviationCalibratorCutoff Concentration(ng/mL)
BuprenorphineBUPBuprenorphine10
Tricyclic AntidepressantsTCANortriptyline1000
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300
MethamphetamineMAMP /METd-Methamphetamine1000
MorphineMORMorphine300
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor-Δ'-THC9-COOHરેણ
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd.l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

The Chemtrue® Multi-Panel Drug Screen Dip Card Tests panel can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

Chemtrue® Multi-Panel Drug Screen Dip Card with OPI 2000 Tests:

The Chemtrue® Multi-Panel Drug Screen Dip Card with OPI 2000 Tests are rapid lateral flow immunoassays for the qualitative detection of Buprenorphine, Amphetamine, Cocaine, Marijuana. Opiates 2000. Methamphetamine. Phencyclidine. Benzodiazepines. Barbiturates. Ecstasy, Methadone, Oxycodone and Tricyclic Antidepressants drugs in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteAbbreviationCalibratorCutoff Concentration(ng/mL)
BuprenorphineBUPBuprenorphine10
Tricyclic AntidepressantsTCANortriptyline1000
AmphetamineAMPd-Amphetamine1000
CocaineCOCBenzoylecgonine300

{7}------------------------------------------------

MethamphetamineMAMP /METd-Methamphetamine1000
OpiatesOPIMorphine2000
PhencyclidinePCPPhencyclidine25
MarijuanaTHC11-nor- $\Delta^9$ -THC9-COOH50
BenzodiazepinesBZOOxazepam300
BarbituratesBARSecobarbital/Pentobarbital300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
MethadoneMTDMethadone300
OxycodoneOXYOxycodone100

The Chemtrue® Multi-Panel Drug Screen Dip Card with OPI 2000 Tests can consist of any combination of the above listed drug analytes. The tests are intended for prescription and Over-The-Counter (OTC) use.

The tests provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

The tests are not intended to differentiate between drugs of abuse and prescription use of Benzodiazepines, Barbiturates, Buprenorphine, Oxycodone and Tricyclic Antidepressants. There are no uniformly recognized cut-off concentration levels for these drugs in urine.

E. DEVICE DESCRIPTION

The Chemtrue® Drug Screen Tests are colloidal gold based lateral flow immunoassays for the rapid, qualitative detection of drugs of abuse in human urine. The tests are single-use, in vitro diagnostic devices, which come in the formats of dip card, cup, or cassette, as indicated by the test name.

F. SUBSTANTIAL EQUIVALENCE INFORMATION

Comparison with the predicate devices is outlined below:

ItemCandidate DevicesPredicate
Indication(s) for useFor qualitative detection of drugs of abuse in human urineSame
Specimen TypeHuman urineSame
Methodology/TechnologicalCharacteristicsLateral flow, competitive binding immunoassay based onthe principle of antigen and antibody immunochemistry.Same
ResultsQualitativeSame
Cut OffBuprenorphine 10 ng/mLTricyclic Antidepressants 1000 ng/mLAmphetamine 1000 ng/mLCocaine 300 ng/mLSame, except thepredicate also has a300 ng/mL cut off

{8}------------------------------------------------

Methamphetamine 1000 ng/mLMorphine 300 ng/mLOpiates 2000 ng/mLPCP 25 ng/mLTHC 50 ng/mLBenzodiazepines 300 ng/mLBarbiturates 300 ng/mLEcstasy (MDMA) 500 ng/mLMethadone 300 ng/mLOxycodone 100 ng/mLand a 500 ng/mLcut off, forAmphetamine andMethamphetamine,respectively.
ConfigurationsDip Card, Cup (for BUP/TCA tests only) and CassetteTest Card and testcard with integratedcup
Intended UsePrescription and OTC UsePrescription Useonly

G. TEST PRINCIPLE

The Chemtrue® Drug Screen Tests are rapid lateral flow immunoassays in which chemically modified drugs (drug-protein conjugates) compete with drugs that may be present in urine. On each test strip, a drug-protein conjugate is striped on the test band of the membrane - known as the test region (T) and the anti-drug antibody-colloidal gold conjugate pads are placed at the forward end of the membrane. If target drugs are present in the urine specimen below its cut-off concentration, the solution of the colored antibody-colloidal gold conjugates moves along with the sample solution by capillary action across the membrane to the immobilized drug-protein conjugate zone on the test band region. The colored antibody-gold conjugates then complexes with the drug-protein conjugates to form visible lines. Therefore, the formation of the visible precipitant in the test band indicates a negative result. If the target drug level exceeds its cut-off concentration, the drug/metabolite antigen competes with drugprotein conjugates on the test band region for the limited antibody on the colored drug antibody-colloidal gold conjugate pad. The drug will saturate the limited antibody binding sites and the colored antibody-colloidal gold conjugate cannot bind to the drug-protein conjugate at the test region of the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result.

A band should form in the control region (C) of the devices regardless of the presence of drug in the sample to indicate that the test has been performed properly.

Monoclonal anti-drug antibodies are used on the BUP/AMP/COC/MET/MOR/OPV PCP/THC/BAR/MDMA/MTD/OXY Test devices which are derived from mouse. The polyclonal anti-drug antibodies are used on TCA/BZO Test devices which are derived from sheep/mouse.

{9}------------------------------------------------

H. PERFORMANCE CHARACTERISTICS

Performance data is only provided for BUP and TCA, as the new analytes. All other drugs of abuse analytes of the candidate devices (AMP, COC, MAMP, OPI, MOR, PCP, THC, BZO, BAR, MDMA, MTD, and OXY) were previously cleared under K102203, K111322, K121339, and K123080. Test strips, sample matrix, test format, and cut-off concentrations for the other drugs of abuse analytes are identical to those cleared under K102203, K111322, K121339, and K123080. See K102203, K111322, K121339, and K123080 for additional precision, specificity, interference, method comparison, and lay-user study information.

1. Reproducibility (Precision) Studies:

The precision study was conducted by three (3) Operators with three (3) lots of each device format in replicates of 10 devices/lot at each concentration level of Negative, 50%, 75%, cutoff, 125% and 150% of the cutoff. The study was conducted over a six (6) day period using GC/MS confirmed drug spiked samples. The samples were blind coded according to a random table and randomly distributed to three operators by the project Manager. The data is analyzed and summarized in the tables below:

ConcentrationLevelnTOTAL
+-
Negative30030
50% of cutoff30030
75% of cutoff30030
Cutoff301020
125% of cutoff30300
150% of cutoff30300

Table 1a. BUP Dip Card Test: Cutoff: 10 ng/mL

Table 1b. TCA Dip Card Test: Cutoff: 1000 ng/mL

ConcentrationLevelnTOTAL
+-
Negative30030
50% of cutoff30030
75% of cutoff30030
Cutoff30129
125% of cutoff30300
150% of cutoff30300

Table 1c. BUP Cup Test: Cutoff: 10 ng/mL

ConcentrationLevelnTOTAL
+-
Negative30030
50% of cutoff30030

{10}------------------------------------------------

75% of cutoff30129
Cutoff30525
125% of cutoff30300
150% of cutoff30300

Table 1d. TCA Cup Test: Cutoff: 1000 ng/mL

ConcentrationnTOTAL
Level+-
Negative30030
50% of cutoff30030
75% of cutoff30030
Cutoff30129
125% of cutoff30300
150% of cutoff30300

Table 1e. BUP Cassette Test: Cutoff: 10 ng/mL

ConcentrationLevelnTOTAL
Negative30030
50% of cutoff30030
75% of cutoff30030
Cutoff30921
125% of cutoff30291
150% of cutoff30300

Table 1f. TCA Cassette Test: Cutoff: 1000 ng/mL

ConcentrationLevelnTOTAL
+-
Negative30030
50% of cutoff30030
75% of cutoff30228
Cutoff301812
125% of cutoff30291
150% of cutoff30300
    1. Specificity Study: These studies were conducted by adding various drugs, drug metabolites, and other structurally-similar compounds likely to be present in the actual urine specimen. The following structurally-related compounds were tested for cross-reactivity and found to be positive if the levels were greater than the following listed concentrations:

Table 2a. Buprenorphine (BUP):

{11}------------------------------------------------

SubstancesConcentration (ng/mL)% Cross Reactivity
Buprenorphine10100
Norbuprenorphine10100

Table 2b. Tricyclic Antidepressants (TCA):

SubstancesConcentration (ng/mL)% Cross Reactivity
Nortriptyline1000100
Amitriptyline1000100
Desipramine300333
Doxepin HCl2,00050
Imipramine502000
Protriptyline475021.1
Trimipramine2,00050

3. Interference:

Over one hundred of potential interferents were tested and found not to cross-react when tested at concentrations of 100 µg/mL at ±25% of the drug (BUP and TCA) cut-off concentrations.

AcetaminophenEstroneOxolinic Acid
AcetanilideEthanolOxymetazoline HCl
6-Acetylmorphineß-EstradiolOxymorphone
N-AcetylprocainamideFenoprofenPapaverine HCl
4-AminoantipyrineGlucosePenicillin G
AmpicillinL-GlutaminePentazocine
(±) AmphetamineHydrochlorothiazidePhenacetin
R(-)AmphetamineHemoglobin HumanPhenalzine Sulfate
S(+)AmphetamineHydralazinePheniramine Maleate
R (-) Apomorphine HClHydrocodoneR- Phenylephrine
AspirinHydrocortisonePrednisolone
L-Ascorbic AcidHydromorphonePrednisone
Atropine3-Hydroxytyramine hydrobromideProtryptiline
BenzocaineImipramine(+) Quinidine
Benzoic AcidIsoprenaline HCl(-) Quinine
BenzoylecgonineKetamineRanitidine
BilirubinLabetolol HClRiboflavin
Bromopheniramine MaleateLoperamide HClSalicylic Acid
Buprenorphine(+/-) MethadoneSerotonin HCl
CannabidiolMethoxyphenamineSodium Chloride

Table 3. The following compounds do not interfere with the tests:

{12}------------------------------------------------

ChloramphenicolMorphineSulfamethazine
ChlorothiazideNaloxoneTemazepam
CholesterolNalidixic AcidTetracycline
Cholic AcidNaproxenTetrahydrozoline HCl
CortisoneNiacinamideΔ°-Tetrahydrocannabinol
(-)-CotinineNicotinamideThiamine HCl
11-DeoxycorticosteroneNifedipineTrifluoperazine HCl
Desalkyl FlurazepamNorcodeineTrimethoprim
DiazepamNordiazepamTryptamine
Diclofenac Sodium Salt19-NorethindroneTyramine
DiflunisalNormorphineDL-Tyrosine
2,5 Dihydroxybenzoic(Gentisic)AcidNorsertralineL-Tryptophan
DopamineNoscapine HClUric Acid
Doxepin HClNortriptylineVerapamil HCl
Duloxetine HCl(+/-) OctapamineZomepirac sodium salt
EDDP PercholateOxazepam
    1. Effect of Urine pH and Specific Gravity Studies: The testing results demonstrate that the urine pH ranges from 3.0 to 8.5 at ±25% and ±50% drug (BUP and TCA) cut-off concentrations do not affect the test performance. The specific gravity (SG) ranges of 1.002, 1.010, 1.015, 1.020, 1.025 and 1.030 at ±25% and ±50% drug (BUP and TCA) cut-off concentrations do not affect the test results.
    1. Stability Study: To establish and support the shelf life and expiration date, stability studies were conducted under accelerated temperature (at 60°C and real time (25°C±2°C) with three (3) lots of each device format. The stability study results support two (2) years shelf-life of the products at (2 to 30°C). The real time stability study is still on-going.
    1. Method Comparison Studies:

Chemtrue® BUP/TCA Single /Multi-Panel Tests were compared to the GC/MS Reference Method. The accuracy of the Chemtrue® Test devices were evaluated against the confirmed GC/MS values in this blind-labeled clinical specimen correlation study (a total of 174 clinical specimens). Three operators performed the testing. Each operator tested one device format with one unique set of blind coded samples. Each blind labeled sample was randomly distributed to each operator by the Clinical Research Cooperator. The results are summarized in the tables below:

Table 6a. Summary from method comparison (Accuracy) study of Chemtrue® Drug Screen Dip Card Test results versus GC/MS

Chemtrue®(-)(+)% Agreement
----------------------------------

{13}------------------------------------------------

Drug ScreenDip CardNo drugpresentGC/MSNegative(≤ 50% of thecutoff)Near cutoffnegative(75% of theC/O to cutoff)Near cutoffpositive(Cutoff to 125%of the C/O)GC/MSPositive(≥150% of thecutoff)with GC/MSvalues
BUP (+)002625100%
(-)40750096.3%
TCA (+)000826100%
(-)406900100%

Table 6b. Summary of a method comparison (Accuracy) study of Chemtrue® Drug Screen Cup Test results versus the GC/MS

Chemtrue®DrugScreen CupNo drugpresent(-)(+)% Agreementwith GC/MSvalues
GC/MSNegative(≤ 50% of thecutoff)Near cutoffnegative(75% of the C/O tocutoff)Near cutoffpositive(Cutoff to 125% ofthe C/O)GC/MSPositive(≥150% of thecutoff)
BUP (+)002625100%
BUP (-)40750096.3%
TCA (+)000826100%
TCA (-)406900100%

Table 6c. Summary of a method comparison (Accuracy) study of Chemtrue® Drug Screen Cassette Test results versus the GC/MS

Chemtrue®DrugScreenCassetteNodrugpresent(-)GC/MSNegative(≤50% of thecutoff)(-)Near cutoffnegative(75% of the C/Oto cutoff)(+)Near cutoffpositive(Cutoff to 125%of the C/O)(+)GC/MSPositive(≥150% of the cutoff)% AgreementwithGC/MSvalues
BUP(+)002625100%
(-)40750096.3%
TCA(+)000826100%
(-)406900100%

Table 6d. DISCORDANT RESULTS: There are two (2) discordant results for each format:

Cutoff Value (ng/mL)Analyte assay (POS/NEG)Drug/Metabolite GC/MS value (ng/mL)
Drug AnalyteGC/MS Value (ng/mL)
Buprenorphine 10+Buprenorphine9.5
Buprenorphine 10+Buprenorphine9.8
  1. OTC Lay-user Studies:

A total of 300 people participated in an OTC lay-user study at three (3) intended user sites with LC/MS confirmed urine samples at the following concentrations: negative, 50%, 75%, 150% and 200% of the cutoff by spiking drugs into drug-free urine pool. Each sample was aliquotted into an individual blind-labeled container. Each lay-user was provided with a package insert, up to two blind labeled samples and test device(s). The results are summarized below:

{14}------------------------------------------------

DrugCut Off(ng/mL)ResultsDrug Concentrations (Per LC/MS Values)
AnalyteNo Drugpresent50% of thecutoff75% ofthe cutoff125% ofthe cutoff150% ofthe cutoff200% ofthe cutoff
BUP10# of positive000303030
# of Negative303030000
Total303030303030
Agreement100%100%100%100%100%100%

Table 7a. Chemtrue® BUP/TCA Drug Screen Dip Card Test vs LC/MS Value Analysis

Table 7b. Chemtrue® BUP/TCA Drug Screen Dip Card Test vs LC/MS Value Analysis

DrugAnalyteCut Off(ng/mL)Drug Concentrations (Per LC/MS Values)
ResultsNo Drugpresent50% of thecutoff75% ofthe cutoff125% ofthe cutoff150% ofthe cutoff200% ofthe cutoff
TCA1000# of positive001303030
# of Negative303029000
Total303030303030
Agreement100%100%96.7%100%100%100%

Table 7c. Chemtrue® BUP/TCA Drug Screen Cup Test vs LC/MS Value Analysis

DrugCut OffDrug Concentrations (Per LC/MS Values)
Analyte(ng/mL)ResultsNo Drug present50% of the cutoff75% of the cutoff125% of the cutoff150% of the cutoff200% of the cutoff
BUP10# of positive000303030
# of Negative303030000
Total303030303030
Agreement100%100%100%100%100%100%

Table 7d. Chemtrue® BUP/TCA Drug Screen Cup Test vs LC/MS Value Analysis

DrugAnalyteCut Off(ng/mL)ResultsDrug Concentrations (Per LC/MS Values)
No Drugpresent50% of thecutoff75% ofthe cutoff125% ofthe cutoff150% ofthe cutoff200% ofthe cutoff
TCA1000# of positive000303030
# of Negative303030000
Total303030303030
Agreement100%100%100%100%100%100%

Table 7e. Chemtrue® BUP/TCA Drug Screen Cassette Test vs LC/MS Value Analysis

DrugAnalyteCut Off(ng/mL)ResultsDrug Concentrations (Per LC/MS Values)
No Drugpresent50% of thecutoff75% ofthe cutoff125% ofthe cutoff150% ofthe cutoff200% ofthe cutoff
BUP10# of positive000303030
# of Negative303030000
Total303030303030
Agreement100%100%100%100%100%100%

Table 7f. Chemtrue® BUP/TCA Drug Screen Cassette Test vs LC/MS Value Analysis

DrugAnalyteCutOff(ng/mL)ResultsDrug Concentrations (Per LC/MS Values)
No Drugpresent50% of thecutoff75% of thecutoff125% of thecutoff150% of thecutoff200% of thecutoff

{15}------------------------------------------------

TCA1000# of positive000303030
# of Negative303030000
Total303030303030
Agreement100%100%100%100%100%100%

Lay-users were also given surveys on the ease of understanding the package insert instructions. 99% of the lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on the package insert and the score revealed a reading grade level of less than 7.

I. CONCLUSION:

Based on the test principle and performance characteristics of the proposed device, it is concluded that the candidate devices are substantially equivalent to the predicate device.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).