Search Results
Found 1 results
510(k) Data Aggregation
(129 days)
The EXPEDIUM Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.
The EXPEDIUM Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion in skeletally mature patients.
The EXPEDIUM Spine System is a 5.5mm rod-based and platebased system offered in both titanium and stainless steel. The system consists of monoaxial screws, polyaxial screws, uni-planar screws, reduction screws, reduction hooks, hooks, extended tab implants, sacral extenders, lateral connectors, washers, fixed bolts, polyaxial bolts, closed screws, slotted connectors, plates, nuts, washers, drop-entry connectors, modular cross connectors, transverse rod connectors, and wires.
The proposed addition to the EXPEDIUM 5.5mm Spine System is an increased favored angle polyaxial screw. The polyaxial screw head opens posteriorly for placement of a rod which is secured to a screw by either a dual inner or unitized set screw. This additional component is available in various geometries and sizes to accommodate patient anatomy. It will be provided non-sterile.
Here's a breakdown of the acceptance criteria and study information for the K102249 submission, based on the provided text:
Acceptance Criteria and Device Performance
The provided document describes a Special 510(k) submission for an addition to an existing spinal system, the EXPEDIUM® Spine System. The acceptance criteria and "device performance" (in this context, demonstrating that the new component meets the same performance standards as the predicate device) are based on mechanical non-clinical testing.
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Static cantilever beam in accordance with ASTM F1798-97 | Not explicitly stated (implied to be meeting the standard's requirements for strength/stability) | "The acceptance criteria was/were met." |
Static axial slip in accordance with ASTM F1798-97 | Not explicitly stated (implied to be meeting the standard's requirements for resistance to slip) | "The acceptance criteria was/were met." |
Dynamic cantilever beam in accordance with ASTM F1798-97 | Not explicitly stated (implied to be meeting the standard's requirements for fatigue life/durability) | "The acceptance criteria was/were met." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact sample size (number of devices/components tested) for each mechanical test. It only states that the tests were conducted in accordance with ASTM F1798-97.
- Data Provenance: The data is from non-clinical mechanical testing performed by the manufacturer, DePuy Spine, Inc., presumably in their labs. This is not patient-derived data, so terms like "country of origin" or "retrospective/prospective" are not applicable.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This question is not applicable to this submission. The "ground truth" for mechanical testing is established by the specified ASTM standard (F1798-97), which outlines the methodology and performance metrics. There are no "experts establishing ground truth" in the sense of clinical interpretation or diagnosis for this type of non-clinical mechanical evaluation.
4. Adjudication Method for the Test Set
This question is not applicable. Adjudication typically refers to the process of resolving discrepancies in expert opinions, often in clinical studies. For mechanical testing against an ASTM standard, the results are objectively measured and compared to the predefined criteria of the standard.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission explicitly states: "No clinical tests were performed."
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This question is not applicable. This submission is for a physical medical device (spinal implant components), not a software algorithm or AI-driven system. Therefore, standalone algorithm performance is not relevant.
7. The Type of Ground Truth Used
The "ground truth" for demonstrating the device meets acceptance criteria is adherence to established mechanical testing standards (ASTM F1798-97). The performance of the new components was compared to these standards to ensure they are equivalent to the predicate devices.
8. The Sample Size for the Training Set
This question is not applicable. This submission is for a physical medical device, not a machine learning model. Therefore, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
This question is not applicable. As there is no training set for a machine learning model, there is no ground truth for a training set to be established.
Ask a specific question about this device
Page 1 of 1