K Number
K070387

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-09-18

(221 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EXPEDIUM™ Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion in skeletally mature patients.

Device Description

The subject EXPEDIUM™ Spine System components are designed to accept a 5.5mm rod and are available in various geometries and sizes.

AI/ML Overview

The provided text describes a 510(k) submission for the EXPEDIUM™ Spine System, a medical device for spinal fixation. It details the device's intended use, materials, and a summary of performance data. However, it does not explicitly state acceptance criteria or a comprehensive study report with the specific details requested in your prompt regarding device performance, sample sizes, ground truth establishment, or expert involvement.

The document focuses on demonstrating substantial equivalence to predicate devices for regulatory approval, rather than a detailed clinical or standalone performance study.

Here's what can be extracted and what cannot be provided given the input:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Not explicitly stated"Performance data per ASTM F 1798 were submitted to characterize the subject EXPEDIUM™ Spine System components addressed in this notification."
  • Explanation: The document mentions that performance data per ASTM F 1798 was submitted, implying the device met the requirements of this standard for characterizing spinal system components. However, the specific acceptance criteria (e.g., minimum load to failure, fatigue life) from ASTM F 1798 and the actual measured performance values of the EXPEDIUM™ Spine System against those criteria are not detailed in this summary.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not specified.
  • Data Provenance: Not specified. The performance data is stated to be "per ASTM F 1798," which is a standard for mechanical testing of spinal instrumentation, not typically involving patient data or clinical trials in the context of this 510(k) summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. The "ground truth" in this context refers to mechanical performance data tested against a standard, not expert interpretation of medical images or clinical outcomes.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. This is not a study assessing diagnostic performance or clinical outcomes requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is a spinal fixation system, not an AI-powered diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Yes, in a sense, a "standalone" performance evaluation was done. The device components were mechanically tested "per ASTM F 1798" to characterize their performance independently. This is a standalone mechanical test, not a standalone AI algorithm performance test.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The "ground truth" for the performance data would be the physical properties and mechanical behavior of the device components as measured by standardized mechanical tests outlined in ASTM F 1798. This typically involves measurements such as static and dynamic compressive strength, fatigue life, torsional strength, etc., defined by the standard.

8. The sample size for the training set

  • Not applicable. This is a mechanical device, not an AI model requiring a training set.

9. How the ground truth for the training set was established

  • Not applicable. (See #8)

Summary of what the document provides regarding acceptance criteria and performance:

The document states that the EXPEDIUM™ Spine System's components were characterized by "Performance data per ASTM F 1798". This indicates that the device was subjected to the mechanical testing protocols specified by the ASTM F 1798 standard. For a 510(k) submission, meeting the requirements of such a recognized standard serves as evidence of safety and effectiveness, and by extension, fulfilling implied acceptance criteria for mechanical performance. The specific quantitative results or detailed acceptance thresholds required by DePuy Spine, Inc. for their device under ASTM F 1798 are not present in this summary.

{0}------------------------------------------------

5. 510(K) SUMMARY

Submitter:DePuy Spine, Inc. 325 Paramount Drive Raynham, MA 02767
-----------------------------------------------------------------------

SEP 1 8 2007

K020387

Contact Person:Christopher Klaczyk
Regulatory Project Manager
Voice:(508) 828-2852
Fax:(508) 828-3797
E-Mail:cklaczyk@dpyus.jnj.com

February 8, 2007 Date Prepared:

Device Class: Class III

Classification Name: Pedicle screw spinal fixation per 21 CFR §888.3070

Spinal interlaminar fixation orthosis per 21 CFR §888.3050

Spinal intervertebral body fixation orthosis per 21 CFR §888.3060

Classification Panel: Orthopedics

FDA Panel Number: 87

Product Code(s):NKB, MNH, MNI, KWP, KWQ
Proprietary Name:EXPEDIUM Spine System
Predicate Devices:EXPEDIUMTM 5.5mm Spine System (K033901)EXPEDIUMTM 6.35mm Spine System (K062174)VIPER Spine System (K041801, K061520)MOSS® Miami Spine System (K011182)Ulrich TangoRS (K052385)Aesculap macsTL (K032059)
Device Description:The subject EXPEDIUMTM Spine System components aredesigned to accept a 5.5mm rod and are available in various

geometries and sizes.

{1}------------------------------------------------

Traditional 510(k) Submission -- EXPEDIUM Spine System

Intended Use:The EXPEDIUM™ Spine System is intended fornoncervical pedicle fixation and nonpedicle fixation for thefollowing indications: degenerative disc disease (defined asback pain of discogenic origin with degeneration of the discconfirmed by history and radiographic studies);spondylolisthesis; trauma (i.e., fracture or dislocation);spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/orlordosis); tumor; pseudarthrosis; and failed previous fusionin skeletally mature patients.
Materials:Manufactured from ASTM F 136 implant grade titaniumalloy.
Performance Data:Performance data per ASTM F 1798 were submitted tocharacterize the subject EXPEDIUM™ Spine System

components addressed in this notification.

:

:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DePuy Spine, Incorporated % Mr. Christopher Klaczyk Regulatory Project Manager 325 Paramount Drive Raynham, Massachusetts 02767-0350

SEP 1 8 2007

Re: K070387

Trade/Device Name: EXPEDIUM™ Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle Screw Spinal System Regulatory Class: Class III Product Code: NKB, MNI, MNH, KWP and KWQ Dated: June 20, 2007 Received: June 21, 2007

Dear Mr. Klaczyk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Christopher Klaczyk

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark A. Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE STATEMENT 4.

510(k) Number (if known):

Device Name: EXPEDIUM Spine System

Indications For Use:

The EXPEDIUM™ Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion in skeletally mature patients.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark A. Mellman

(Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number

DePuy Spine, Inc., a Johnson & Johnson Company

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.