Search Results
Found 104 results
510(k) Data Aggregation
(178 days)
CFR §882.5805 / OBP
Secondary Regulation and Product Code:
21 CFR §882.5802 / QCI
21 CFR §882.1870
/ OBP | Same. |
| Secondary regulation(s) / product code(s) | 21 CFR §882.5802 / QCI21 CFR §882.1870
The Nexstim Navigated Brain Stimulation (NBS) 6 System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode, as well as an adjunct for the treatment of adult patients suffering from Obsessive Compulsive Disorder (OCD).
The Nexstim NBS 6 System is a non-invasive, repetitive transcranial magnetic stimulation (rTMS) system that delivers repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat depressive episodes without inducing seizure in adult patients with Major Depressive Disorder (MDD) who have failed antidepressant medication, as well as to provide adjunct treatment to adult patients suffering from Obsessive Compulsive Disorder (OCD).
The Nexstim NBS 6 System is used for patient treatment by prescription only and must be operated by a trained medical professional. It can be used in both inpatient and outpatient settings including physician's offices and clinics, psychiatric hospitals, and general medical/surgical hospitals with psychiatric units.
The Nexstim NBS 6 System consists of a group of devices designed to localize the stimulation site in the brain and deliver rTMS stimulation using controlling and interpretive software. Operational control of the Nexstim NBS 6 System is provided by the software.
The Nexstim NBS 6 System combines magnetic resonance imaging-based (MRI-based), three dimensional (3D) localization of cortical motor areas of the brain with non-invasive TMS and simultaneous electromyography (EMG) measurement to locate areas of the brain that are capable of evoking muscle responses when stimulated, and to locate the target areas for treatment of MDD and adjunct treatment of OCD.
The software used in the Nexstim NBS 6 System is essentially the same software used in the Reference device Nexstim NBT System 2 cleared by FDA in 510(k) submissions K171902 and K182700 except for a new workflow-oriented graphical user interface (GUI) designed for improved ease of use. This GUI is new to the Nexstim NBS 6 System, and allows the user to select which functionality the software will present, including:
• Access to the same functions as in NBT System 2 for:
N/A
Ask a specific question about this device
(234 days)
and Product Code:** 21 CFR §882.5805 / OBP
Secondary Regulations and Product Codes:
- 21 CFR §882.1870
----|-----------------------------------|
| Secondary regulation(s) / product code(s) | 21 CFR §882.1870
NBS 6 is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
For Adolescents: NBS 6 is indicated as an adjunct for the treatment of MDD in adolescent patients (age 15-21).
The Nexstim NBS 6 System is a non-invasive, repetitive transcranial magnetic stimulation (rTMS) system that delivers repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat depressive episodes without inducing seizure in adult patients with Major Depressive Disorder (MDD) who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode, and as an adjunct for the treatment of MDD in adolescent patients, 15 to 21 years of age.
The Nexstim NBS 6 System is used for patient treatment by prescription only and must be operated by a trained medical professional. It can be used in both inpatient and outpatient settings including physician's offices and clinics, psychiatric hospitals, and general medical/surgical hospitals with psychiatric units.
The Nexstim NBS 6 System consists of a group of devices designed to localize the stimulation site in the brain and deliver rTMS stimulation using controlling and interpretive software. Operational control of the Nexstim NBS 6 System is provided by the software.
The Nexstim NBS 6 System combines magnetic resonance imaging-based (MRI-based), three dimensional (3D) localization of cortical motor areas of the brain with non-invasive TMS and simultaneous electromyography (EMG) measurement to locate areas of the brain that are capable of evoking muscle responses when stimulated, and to locate the target areas for treatment of MDD.
The software used in the Nexstim NBS 6 System is essentially the same software used in the reference device Nexstim NBT System 2 cleared by FDA in 510(k) submissions K171902 and K182700 except for a new workflow-oriented graphical user interface (GUI) designed for improved ease of use. This GUI is new to the Nexstim NBS 6 System, and allows the user to select which functionality the software will present, including:
• Access to the same functions as in NBT System 2 for:
N/A
Ask a specific question about this device
(125 days)
Re: K252842
Trade/Device Name: SafeOp 3: Neural Informatix System
Regulation Number: 21 CFR 882.1870
Classification Name: Stimulator, Electrical, Evoked Response
Regulation Number: 21 CFR 882.1870
PDQ, ETN | Identical to predicate, SafeOp 3 System (K234092) |
| Regulation Number (21 CFR) | §882.1870
, §882.1320, §882.1350, §890.1375, §874.1820, §874.1820 | §882.1870, §882.1320, §882.1350, §890.1375,
The SafeOp 3: Neural Informatix System is intended for use in monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, extracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures.
SafeOp 3 Accessories: The SafeOp Accessories are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.
The SafeOp™ 3: Neural Informatix System (SafeOp 3 System), consists of the SafeOp patient interface with power supply and IV pole mount, the Alpha Informatix Tablet with docking station and power supply and a data transfer USB cable. Associated disposable accessories consists of an electrode harness, surface and/or subdermal needle electrodes, MEP Activator, Cranial Hub, Delta Dilators and Stimulating Probe and Delta Clip contained in various kits.
The subject device is intended for use by trained healthcare professionals, clinical neurophysiologists/technologists and appropriately trained non-clinical personnel. The subject device is intended for use in operating room environments of hospitals and surgical centers. System setup may be performed by both clinical and trained non-clinical personnel.
The provided FDA 510(k) clearance letter and summary for the SafeOp 3: Neural Informatix System (K252842) primarily describe the device's substantial equivalence to a predicate device (K234092) and outline a Predetermined Change Control Plan (PCCP) for future modifications. However, it does not include a detailed study proving the device meets specific acceptance criteria in the context of clinical performance or AI algorithm validation for the initial clearance.
The "Performance Data" section explicitly states: "Determination of substantial equivalence is not based on an assessment of clinical performance data." and refers to nonclinical performance testing (functional, system, software requirements) and usability testing.
Therefore, the following information is extracted and synthesized based on the available document, but a comprehensive study proving acceptance criteria for all aspects of neural informatic performance is not present. The document focuses on the future PCCP for algorithmic changes, which will have performance criteria.
Acceptance Criteria and Reported Device Performance
The core of the documentation for this 510(k) clearance is about demonstrating substantial equivalence to an existing predicate device (SafeOp 3™: Neural Informatix System, K234092). Therefore, the "acceptance criteria" for the cleared device itself are primarily its ability to match the predicate's performance and specifications, which is shown through a direct comparison.
The document does not report specific quantitative performance metrics for the SafeOp 3: Neural Informatix System for its initial clearance, as it relies on substantial equivalence.
However, the Predetermined Change Control Plan (PCCP) outlines future modifications to the SSEP Baseline Classification Algorithm and defines acceptance criteria for these future algorithmic changes. These are the only explicitly stated performance acceptance criteria in the document.
Table 1: Acceptance Criteria and Reported Device Performance (from PCCP for future algorithmic changes)
| Acceptance Criteria (for future SSEP algorithm modifications) | Reported Device Performance (for future SSEP algorithm modifications) | Notes |
|---|---|---|
| Modification #1 & #4 (SSEP Baseline Classification Algorithm Retraining/Maintenance): | No reported performance data for the current cleared device. | The document states these are criteria for future modifications of the SSEP algorithm, not for the initial clearance of the SafeOp 3 System. |
| Overall Positive Predictive Value (PPV) ≥ 90% | To be demonstrated with verification testing of modified algorithm. | These criteria are goals for future changes. |
| Overall Accuracy ≥ 85% | To be demonstrated with verification testing of modified algorithm. | |
| Overall F1-score ≥ 85% | To be demonstrated with verification testing of modified algorithm. | |
| Performance of each stimulation site-specific classifier to be demonstrated as non-inferior to the current SafeOp 3 System at the 95% confidence level. | To be demonstrated with verification testing of modified algorithm. | |
| Modification #2 (Removal of Recalculating Active Trace (RAT) State): | No reported performance data for the current cleared device. | These are criteria for a future modification involving signal processing. |
| Average noise content for the reconstructed Ensemble Average (EA) following restart to be demonstrated as non-inferior to that of the preceding EA(s) with 95% confidence. | To be demonstrated with internal testing and retrospective analysis of intraoperative data. | These criteria are goals for future changes. |
Study Details (Based on the document, for the initial clearance)
Because the 510(k) is based on substantial equivalence and explicitly states "Determination of substantial equivalence is not based on an assessment of clinical performance data," there is no detailed "study" proving the device meets acceptance criteria in the way a de novo or PMA submission might. The "study" mentioned below refers to the verification and validation activities conducted to demonstrate substantial equivalence and safety/effectiveness for the initial clearance, as described in the "Performance Data" section.
-
Sample size used for the test set and the data provenance:
- Not explicitly stated for clinical performance data for the initial clearance. The document mentions "nonclinical performance testing" for functional, system, and software requirements, and "usability testing." These typically involve controlled test cases and user interactions rather than a large clinical test set.
- For the future PCCP modifications of the SSEP algorithm, it mentions "an independent, firewalled dataset" and "additional labeled data" for re-training and testing, but no sample size or provenance is specified within this document.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not explicitly stated for the initial clearance. Since the clearance is based on substantial equivalence and does not assess clinical performance data, detailed expert ground truth establishment for a clinical test set is not described.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not explicitly stated for the initial clearance. Not applicable given the reliance on substantial equivalence and lack of clinical performance data assessment.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. The document does not describe an MRMC study. The device is a neural informatic system, and while it involves signal processing, it's not positioned as an AI-assisted diagnostic tool that human readers would interpret with assistance in the traditional sense of an MRMC study. The SSEP algorithm mentioned in the PCCP appears to be a standalone classification, not a human-in-the-loop assistance tool subject to such a study.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- For the initial clearance, the document refers to "nonclinical performance testing" demonstrating the system meets functional, system, and software requirements. This implies standalone testing of device functions.
- For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), the acceptance criteria (PPV, Accuracy, F1-score) clearly indicate standalone algorithm performance is intended to be evaluated. However, no such standalone performance data is provided for the initially cleared device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not explicitly stated for the initial clearance. For the "nonclinical performance testing" and "usability testing," ground truth would typically be established based on engineering specifications and validated operational procedures.
- For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it refers to "labeled data," which implies a ground truth established by experts or other definitive methods, but the specific type is not detailed in this document.
-
The sample size for the training set:
- Not explicitly stated for the initial clearance. No AI algorithms are reported as being evaluated for the initial clearance.
- For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it mentions the algorithm "will be re-trained using site-specific data" and "periodically re-trained as additional labeled data become available," but no sample size is specified within this document.
-
How the ground truth for the training set was established:
- Not explicitly stated for the initial clearance.
- For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it refers to "labeled data," but the method of labeling (establishing ground truth) is not detailed in this document.
Ask a specific question about this device
(153 days)
Rapid2 Magnetic Stimulators (Magstim Rapid2, Magstim Super Rapid2, Magstim Super Rapid2 Plus1) are intended to stimulate peripheral nerves for relief of chronic intractable, post-traumatic and post-surgical pain for patients 18 years or older.
The Magstim Rapid2, Magstim Super Rapid2, Magstim Super Rapid2 Plus1 (herein collectively referred to as "Rapid2 Magnetic Stimulators") are computerized, electromechanical medical devices that provide brief and focused magnetic pulses in order to non-invasively stimulate peripheral nerves for relief of chronic intractable, post-traumatic and post-surgical pain for patients 18 years or older. The subject device is intended to be used in hospitals and clinics such as pain management clinics.
Rapid2 Magnetic Stimulators are integrated systems consisting of a combination of hardware, software, and accessories. Rapid2 Magnetic Stimulators are offered in multiple configurations:
- Rapid2
- Super Rapid2
- Super Rapid2 Plus1
All three configurations have identical intended use/indications for use, common specifications, equivalent performance characteristics and equivalent composition to each other. Specifically, Rapid2 and Super Rapid2 have received prior clearance under K051864 for Peripheral Nerve Stimulation (Product Code: GWF, Regulation 21 CFR 882.1870). All Rapid2 Magnetic Stimulators are made up of components that have received prior clearance under K051864 (e.g., the 3190-00, 3192-00 and 3193-00 coils) and components which have received prior clearance under K051864 but have received modifications due to aspects like obsolescence (Mainframe, Power Supply etc.).
All Rapid2 Magnetic Stimulators are composed from the following main components:
- Stimulating Unit & Power Supply
- User Interface
- Stimulating Coil
- System and Stimulating Coil Cart and Holding Arm
Rapid2 Magnetic Stimulators include temperature monitoring via two independent temperature sensors to ensure surfaces of the coils do not reach unacceptable levels. The cut-off is set to act at 40°C at which point the system will automatically be disabled. Over-temperature conditions are also communicated on the User Interface (UI) via a temperature gauge and alarm system. Rapid2 Magnetic Stimulators also includes the 3910-00 air-cooled coil to further mitigate any temperature conditions. The 3910-00 air-cooled coil comes with all 3 configurations (Rapid2, Super Rapid2 and Super Rapid2 Plus1) as standard.
The provided document is an FDA 510(k) Clearance Letter for the Rapid2 Magnetic Stimulators. It does NOT contain information about a study proving the device meets acceptance criteria related to its performance in pain relief. Instead, it focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing of technical characteristics, safety standards compliance, and physical properties.
The document does not describe an AI/ML-driven device, nor does it present results from a clinical study with patients or human readers using AI. The acceptance criteria and performance metrics described are related to physical and electrical characteristics of the magnetic stimulator, not diagnostic accuracy or clinical effectiveness in a traditional sense of "performance" as one might expect for an AI diagnostic device.
Therefore, many of the requested items (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, MRMC study, standalone performance, ground truth type) cannot be answered based on the provided text, as they pertain to clinical or AI/ML performance evaluation, which is not the subject of this 510(k) summary.
However, I can extract information related to the technical acceptance criteria and how they align with the device's measured performance as described in the summary:
Acceptance Criteria and Device Performance (Based on Technical and Safety Equivalence)
The acceptance criteria for the Rapid2 Magnetic Stimulators are primarily based on demonstrating substantial equivalence to the predicate device (MagVenture Pain Therapy) in terms of technical characteristics, safety, and effectiveness for the stated indications for use. The "performance" reported is adherence to these characteristics and safety standards.
| Acceptance Criteria Category/Characteristic | Subject Device Performance (Rapid2 Magnetic Stimulators) | Predicate Device Performance (MagVenture Pain Therapy) | Evaluation / Proof of Meeting Criteria |
|---|---|---|---|
| Indications for Use | Stimulate peripheral nerves for relief of chronic intractable, post-traumatic and post-surgical pain for patients 18 years or older. | Stimulate peripheral nerves for relief of chronic intractable, post-traumatic and post-surgical pain for patients 18 years or older. | Identical. Meets criteria by having the same intended use. |
| Anatomical Sites | Any area, such as hand, arm, waist, buttock, thigh, calf, back and lower back etc. | Any area, such as hand, arm, waist, buttock, thigh, calf, back and lower back etc. | Identical. |
| Treatment Facilities | Hospitals & Clinics | Hospitals & Clinics | Identical. |
| Treatment Time | 13 minutes per session (800 seconds) | 13 minutes per session (800 seconds) | Identical. |
| Pulse Frequency | Rapid2: 0.1 – 50 Hz (pps); Super Rapid2 and Super Rapid2 Plus1: 0.1 – 100 Hz (pps) | MagPro R30 & MagPro R30 with MagOption: 0.1 – 30 Hz (pps); MagPro X100 & MagPro X100 with MagOption: 0.1 – 100 Hz (pps) | Similar range. Subject device's range covers or extends slightly beyond predicate, but the recommended protocol (0.5Hz) is well within both. |
| Pulse Amplitude | 0 – 100% | 0 – 100% | Identical. |
| On-cycle duty period | 2-800 Seconds (0.5 Hz and up to 400 pulses) | 2-800 Seconds (0.5 Hz and up to 400 pulses) | Identical. |
| Off-cycle rest period | N/A | N/A | Identical. |
| Maximum Repetition Rate | Rapid2: 50Hz; Super Rapid2: 100Hz; Super Rapid2 Plus1: 100Hz | MagPro R30 & MagPro R30 with MagOption: 30 pulses per second; MagPro X100 & MagPro X100 with MagOption: 100 pulses per second | Upper limit identical compared to predicate. Substantial equivalence demonstrated despite differences in how maximum output is achieved (explained in "SE Note 1"). |
| Pulse Width | Biphasic (300-425 µs) | Biphasic (280-320 µs) | Similar range. Differences deemed not to raise new safety/effectiveness questions due to compensating factors (see "SE Note 2"). |
| Pulse Mode | Standard | Standard | Same. |
| Temperature Control | Automatic disable at 40°C; includes air-cooled coil; UI communication of over-temperature. | Automatic disable at 43°C. | Comparable/Better. Subject device has a lower cutoff and additional cooling/reporting features. |
| Peak Magnetic Field at Coil Surface | 1.0-1.5T | 1.15-2.6T | Substantially equivalent. Subject device is a subset of the predicate's range. Differences explained in "SE Note 2" as not raising new safety/effectiveness concerns. |
| Peak Magnetic Field Gradient (dB/dt) at 20mm from Coil Center | 9-12kT/s | 9-24kT/s | Substantially equivalent. Subject device is a subset of the predicate's range. Differences explained in "SE Note 2" as not raising new safety/effectiveness concerns. |
| Waveform | Biphasic, Biphasic Burst | Biphasic, Monophasic, Biphasic Burst, Halfsine (combinations vary by predicate configuration) | Substantially equivalent. Subject device's waveform is within the range available in the predicate. |
| Software/Firmware Control | Yes | Yes | Identical. Verified per IEC 62304. |
| Power Supply Type | Power Supply via dedicated power supply modules each using a separate input mains line cord. | Power Supply via Isolation Transformer. | Similar. |
| Power Consumption | 230/240V Systems – 3000VA peak per input; 115V Systems – 2300VA peak per input | Maximum 2700VA | Similar. |
| User Interface | LCD Capacitive Touchscreen | LED Display | Similar. A difference in display technology, but performs the same function. |
| Housing Material Construction | Stimulator: PUR, Stainless/Galvanized Steel; Coils: PC, PUR | Stimulator: Aluminum, Aluzinc; Coils: PVC, ABS, PA, POM | Similar. Different specific materials but serve the same function. |
| Applied Parts (Coils) | Various, including previously cleared (K051864, K080499, K130403) and new coils (4150-00, 4170-00, 4189-00, 4190-00, 4510-00). | Various, all previously cleared. | Substantially equivalent coil range. New coils are evaluated for safety and function to be equivalent (see "SE Note 2"). |
| Applied Part Area | Butterfly Coils: 152mm – 191mm; Circular Coils: 124.5mm | Butterfly Coils: 150mm; Circular Coils: 110-126mm; Special Coils: 160x80 mm | Substantially equivalent. (see "SE Note 2"). |
| Sterilization | Non-sterile when used. | Non-sterile when used. | Identical. |
| Electrical Safety | Complies with IEC 60601-1 Ed. 3.2 | Complies with IEC 60601-1 Ed. 3.1 | Meets/Exceeds. Complies with a newer edition of the standard. |
| Mechanical Safety | Complies with IEC 60601-1 Ed. 3.2 | Complies with IEC 60601-1 Ed. 3.1 | Meets/Exceeds. Complies with a newer edition of the standard. |
| Thermal Safety | Complies with IEC 60601-1 Ed. 3.2 | Complies with IEC 60601-1 Ed. 3.1 | Meets/Exceeds. Complies with a newer edition of the standard. |
| Radiation Safety | No radiation generated. | No radiation generated. | Same. |
| Biocompatibility | Complies with ISO 10993-1, -5, -10; materials tested for Skin Irritation, Cytotoxicity, Skin Sensitization. | Complies with ISO 10993. | Same/Exceeds. Detailed compliance with relevant parts of the standard. |
| Standards Compliance | ISO 13485 (company); IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62366-1 (device). | EN ISO 13485 (company). | Same/Equivalent. Device-specific standards compliance indicated. |
| Acoustics | Tested per 60601-1 type testing and in-house, demonstrating substantially equivalent acoustic output. Labeling requires earplugs (30dB noise reduction). | (Not specified beyond "similar") | Comparable. Demonstrated and mitigated with user instructions. |
| E-Field Decay & Linearity of Output | Performance data showed "very similar" E-Field decay, linearity, and electric/magnetic field spatial distributions. | (Not explicitly detailed, but implied to be baseline for comparison) | Comparable. Results demonstrate equivalent effects at 0-2cm from coil surface. |
Regarding the specific questions that cannot be answered from the provided text:
- Sample sizes used for the test set and the data provenance: Not applicable or provided. The "test set" here refers to non-clinical testing of device characteristics, not a clinical study on patients or data.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable or provided. Ground truth in this context would implicitly be engineering specifications, laboratory measurements, and standard compliance testing, not expert clinical assessment of patient data.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable or provided. No adjudication process detailed for establishing technical specifications.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device, nor is it a diagnostic device being evaluated for reader performance.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. Not an AI/ML algorithm.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this device's performance evaluation is based on engineering specifications, direct physical measurements (e.g., of magnetic fields, temperatures, electrical properties), and compliance with international safety and performance standards (e.g., IEC 60601 series, ISO 10993).
- The sample size for the training set: Not applicable or provided. This device does not use a "training set" in the machine learning sense.
- How the ground truth for the training set was established: Not applicable or provided.
In summary, the provided document details the 510(k) clearance process for a non-AI/ML magnetic stimulator for pain relief. The "acceptance criteria" and "performance" are framed around demonstrating substantial technical and safety equivalence to a legally marketed predicate device, rather than clinical efficacy data from patient studies or AI algorithm performance metrics.
Ask a specific question about this device
(30 days)
Road Middleton, Wisconsin 53562
Re: K243982
Trade/Device Name: Nicolet EDX Regulation Number: 21 CFR 882.1870
|
| Classification Name: | Evoked Response Electrical Stimulator21 CFR 882.1870
The Nicolet EDX is intended for the acquisition, display, analysis, storage, reporting, and management of electrophysiological information from the human nervous and muscular systems including Nerve Conduction (NCS), Electromyography (EMG), Evoked Potentials (EP), Autonomic Responses and Intra-Operative Monitoring including Electroencephalography (EEG).
Evoked Potentials (EP) includes Visual Evoked Potentials (VEP), Auditory Evoked Potentials (AEP), Somatosensory Evoked Potentials (SEP), Electroretinography (ERG), Electrooculography (EOG), P300. Motor Evoked Potentials (MEP) and Contingent Negative Variation (CNV). The Nicolet EDX with Natus Elite Software may be used to determine autonomic responses to physiologic stimuli by measuring the change in electrical resistance between two electrodes (Galvanic Skin Response and Sympathetic Skin Response). Autonomic testing also includes assessment of RR Interval variability. The Nicolet EDX with Natus Elite Software is used to detect the physiologic function of the nervous system, for the location of neural structures during surgery, and to support the diagnosis of neuromuscular disease or condition.
The listed modalities do include overlap in functionality. In general, Nerve Conduction Studies measure the electrical responses of the nerve; Electromyography measures the electrical activity of the muscle and Evoked Potentials measure electrical activity from the Central Nervous System.
The Nicolet EDX with Natus Elite Software is intended to be used by a qualified healthcare provider.
The Nicolet EDX system is designed for the acquisition, display, analysis, reporting, and management of electrophysiological information from the human nervous and muscular systems. The system is designed to perform Nerve Conduction (NCS), Electromyoaraphy (EMG), Evoked Potentials (EP), and Autonomic Responses. Nicolet EDX system provides a variety of tests spanning the various modalities.
The Nicolet EDX system consists of the following major components:
- Base unit
- Amplifier (2- or 8-channel)
- Control panels
- Computer- laptop or desktop (with display, keyboard and mouse)
- Display Monitor (for desktop system)
- Application Software (Natus Elite)
The Nicolet EDX optional accessories/ components consist of the following:
- Audio stimulators (Headphones or other auditory transducers)
- Visual stimulators (Natus Visual Stimulator, LED goggles or stimulus monitor)
- Electrical stimulators (RS10 comfort probe, WR50 comfort probe plus)
- Cart and associated accessories (such as arms, mounts and isolation transformer
- Miscellaneous options and accessories such as Patient Response button. Single/Triple footswitch, Tendon (Reflex) hammer, Natus photic stimulator, temperature probe, ultrasound device, bone conductor, printer, etc.
The electrodiagnostics system is powered by a connection to mains.
The entire user interface of Nicolet EDX system consists of two major elements:
- The primary means to interact with the system is via a personal computer (PC) running Natus Elite.
- The second means of interaction is the user interface elements on the hardware.
The Nicolet EDX is intended to be used by a qualified healthcare provider. This device does not provide any diagnostic conclusion about the patient's condition to the user. The intended use environment is in a professional healthcare facility environment.
This FDA 510(k) summary for the Nicolet EDX device does not contain the detailed information necessary to fully address all aspects of your request regarding acceptance criteria and study design.
However, based on the provided text, here's what can be extracted and inferred:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" for the device's clinical performance with quantitative metrics. Instead, it compares the technological characteristics of the Nicolet EDX to predicate devices, implying that meeting or being substantially equivalent to these characteristics demonstrates acceptable performance.
Here's a table summarizing the technological characteristics described as identical or similar to predicate devices, which serve as an implicit benchmark for acceptance:
| Characteristic | Nicolet EDX (This Submission) | Primary Predicate (K120979) | Reference Predicate (K243495) | Discussion of differences (from document) |
|---|---|---|---|---|
| 1. General | ||||
| Product Code | GWF, IKN | GWF | GWF | Identical to predicate. |
| Indications for Use | Unchanged from primary predicate | Same as Nicolet EDX | Same as Nicolet EDX | Identical to predicate. |
| Warnings | Items related to off label use or misuse. | Same | Same | Identical to predicate |
| Contraindications | Not for use in flammable anesthetics or with defibrillation. | Same | Same | Identical to predicate |
| Target Population | Patients with neuromuscular diseases (pediatric and adult) | Same | Same | Identical to predicate |
| Environment of Use | Professional healthcare facility environment. | Same | Same | Identical to predicate |
| 2. General Design | ||||
| General Systems Approach | Computer based equipment with dedicated hardware peripherals/components. | Same | Same | Identical to predicate |
| User Input Device | Windows mouse/keyboard driven graphic interface with dedicated control panel. | Same | Same | Identical to predicate |
| User Output Device | Digital color display and commercial printers. | Same | Same | Identical to predicate |
| Patient Inputs | 1 to 8 channel amplifier, isolated | 1 to 8 channel amplifier, isolated | 1 to 4 channel amplifier, isolated | Identical to primary predicate. Similar to reference predicate. |
| Signal Acquisition | Analog to digital conversion at 48 kHz per sample rate | Same | Same | Identical to primary predicate. |
| Trigger Input (synchronization to external events) | Yes | Yes | Yes | Identical to predicate |
| Trigger Output (synchronization for external devices) | Yes | Yes | Yes | Identical to predicate |
| Footswitch | Yes | Yes | Yes | Identical to predicate |
| Use of standard software platform (Operating System) | Yes, Microsoft Windows | Same | Same | Identical to predicate |
| Customization of protocols | Via storage / retrieval of user defined settings | Same | Same | Identical to predicate |
| Application flexibility / expandability | Via software update | Same | Same | Identical to predicate |
| Patient circuitry isolation | Optic/transformer | Same | Same | Identical to predicate |
| System Components | EDX base console, 2 electrical stimulators, auditory stimulator, trigger I/O, LED goggle interface; Control panel; Amplifier; Computer, monitor, keyboard, mouse, printer | Same | Base console, 1 electrical stimulator, auditory stimulators, trigger I/O, LED goggle interface; Control panel (integrated); Amplifier; Computer; Monitor; Keyboard; Mouse; Printer | Identical to predicate. Similar to the reference predicate. |
| System computer interface | USB | USB | USB | Identical to predicate |
| System power supply | Mains (100 -240VAC) thru an isolation transformer depending on system configuration | Same | Same | Identical to predicate |
| 3. Design - Acquisition | ||||
| Number of channels | 1 to 8 | 1 to 8 | 1 to 4 | Identical to primary predicate. Similar to reference predicate. |
| CMRR | >110 dB, 115 dB typical | >110 dB, 115 dB typical | >120 dB | Identical to primary predicate. Similar to reference predicate. |
| Noise | <0.6 uV RMS (from 2Hz to 10kHz) | <0.6 uV RMS (from 2Hz to 10kHz) | <0.4 uV RMS (from 2Hz to 10kHz) | Identical to primary predicate. Similar to reference predicate. |
| Input Impedance | >1000 MΩ | >1000 MΩ | >1000 MΩ | Identical to primary predicate. |
| Low Filter | 0.05 Hz to 5 kHz | 0.05 Hz to 5 kHz | 0.05 Hz to 5 kHz | Identical to primary predicate. |
| High Filter | 30 Hz to 20 kHz | 30 Hz to 20 kHz | 30 Hz to 20 kHz | Identical to predicate. |
| Notch Filter | 50/60 selectable | 50/60 selectable | 50/60 selectable | Identical to primary predicate. |
| A/D conversion | 24 bit | 24 bit | 24 bit | Identical to primary predicate. |
| Sampling rate (cumulative) | 384 kHz | 384 kHz | 192 kHz | Identical to primary predicate. Similar to reference predicate. |
| Trigger mode | Free-run, internal, external | Same | Same | Identical to primary predicate. Similar to reference predicate. |
| Signal delay (pre/post) | -90% to +90% of sweep, depending on time base | Same | Same | Identical to primary predicate. |
| Impedance meter | 500Ω to 480 kΩ | 500Ω to 480 kΩ | 1kΩ to 1,000 kΩ | Identical to primary predicate. Similar to reference predicate. |
| 4. Design - Stimulators | ||||
| 4.1 Electrical Stimulator | ||||
| Type | Constant Current or Constant Voltage | Same | Constant Current | Identical to primary predicate. Similar to reference predicate. |
| Number | 1 or 2 | 1 or 2 | 1 | Identical to primary predicate. Similar to reference predicate. |
| Maximum Output | 100mA or 400V | 100mA or 400V | 100mA | Identical to primary predicate. Similar to reference predicate. |
| Duration | 0.01 to 1 ms | 0.01 to 1 ms | 0.01 to 1 ms | Identical to primary predicate. |
| Mode | Single or Train | Single or Train | Single or Train | Identical to primary predicate. |
| Biphasic | Yes | Yes | Yes | Identical to primary predicate. |
| 4.2 Auditory Stimulator | ||||
| Type | Click, Pip, Burst | Same | Same | Identical to primary predicate. |
| Intensity | 0 to 139 dB pSPL | Same | Same | Identical to primary predicate. |
| Polarity | Condensation, Rarefaction, Alternating | Same | Same | Identical to primary predicate. |
| Tone Frequency | 125 to 8000 Hz | Same | Same | Identical to primary predicate. |
| Click Duration | 0.05 to 1 ms | Same | Same | Identical to primary predicate. |
| Side | Left, Right, Both | Same | Same | Identical to primary predicate. |
| Transducers | Ear Phones, Inserts, Bone Vibrator | Same | Same | Identical to predicate. |
| 4.3 Visual Stimulator | ||||
| LED Goggles | Yes | Yes | Yes | Identical to predicate |
| 5. EMG Application Modules | (All "Yes") | (All "Yes") | (All "Yes") | Identical to predicate |
| 6. Evoked Potential Application Modules | (All "Yes") | (All "Yes") | (All "Yes") | Identical to primary predicate |
| 7. Other Application Modules | (All "Yes") | (All "Yes") | (All "Yes") | Identical to primary predicate |
| 8. Additional Features | ||||
| Automatic Report Narrative Generation | Yes | Yes | Yes | Identical to primary predicate |
| Electrical Stimulus Automation | Yes | Yes | Yes | Identical to primary predicate |
| 9. Image Display and Control Interface | ||||
| Display and control of noninvasive third party imaging modality (e.g., Ultrasound) | Yes, Integrated (concurrent) ultrasound display and control using Sonoscanner Ultrasound System (K232285) with Natus Elite software. | No, Supports ultrasound by supplying a third-party device which operates in a nonintegrated mode. Using the Sonoscanner Ultrasound System (K232285) | Yes, Integrated (concurrent) ultrasound display and control using Sonoscanner Ultrasound System (K232285) with Natus Elite software. | Identical to reference predicate. |
The study proving the device meets the acceptance criteria is detailed by the statement: "Verification and validation activities were conducted to establish the performance and safety characteristics of the Nicolet EDX. The results of these activities demonstrate that the Nicolet EDX is safe, effective, and performance is substantially equivalent to the predicate devices."
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The 510(k) summary focuses on "substantial equivalence" based on technological characteristics and general verification/validation, rather than a clinical study with a specific test set, patient data, and provenance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided as a detailed clinical study with ground truth established by experts is not described in this summary. The device is for electrophysiological information acquisition and analysis, where "ground truth" often refers to the actual physiological signals or clinical diagnoses, which would typically be assessed by qualified healthcare providers using established medical guidelines.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided as a formal clinical study with adjudication of a test set is not detailed.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study is not mentioned in this summary. The device description explicitly states: "This device does not provide any diagnostic conclusion about the patient's condition to the user." It is a diagnostic aid for acquiring and analyzing physiological information, not an AI-driven interpretive tool that would typically undergo MRMC studies to assess human reader improvement.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
A standalone algorithm performance study is not mentioned. The device is an electrodiagnostic system with software, but the summary does not present it as an autonomous AI system requiring standalone performance evaluation in the classification sense. Rather, it emphasizes its role as a tool for a "qualified healthcare provider."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Since a specific clinical study with a detailed test set is not described, the type of ground truth used is not specified. For a device like this, the "ground truth" for performance validation would likely involve:
- Technical Benchmarking: Verification that the electrical signals acquired and processed meet specified technical parameters by comparing them to known or simulated physiological signals, or against established high-fidelity measurement systems.
- Clinical Utility Confirmation: Ensuring that the output provides information that is consistent with expected physiological responses and is interpretable by healthcare professionals for supporting diagnosis.
8. The sample size for the training set
This information is not provided. This summary does not indicate that the device uses machine learning or AI models that require a "training set" in the conventional sense. The "Natus Elite Software" is application software for data acquisition, display, and analysis, not necessarily a predictive AI model requiring a distinct training dataset.
9. How the ground truth for the training set was established
As no training set is mentioned, the method for establishing its ground truth is not applicable/provided.
Ask a specific question about this device
(30 days)
Wisconsin 53562
Re: K243495
Trade/Device Name: Natus Ultrapro S100 (982A0594) Regulation Number: 21 CFR 882.1870
|
| Product Code/Regulation Number: | GWF | 21CFR §882.1870
The UltraPro is intended for the acquisition, display, analysis, storage, reporting, and management of electrophysiological information from the human nervous and muscular systems including Nerve Conduction (NCS), Electromyography (EMG), Evoked Potentials (EP), Autonomic Responses and Intra-Operative Monitoring including Electroencephalography (EEG).
Evoked Potentials (EP) includes Visual Evoked Potentials (VEP), Auditory Evoked Potentials (AEP), Somatosensory Evoked Potentials (SEP), Electroretinography (ERG), Electrooculography (EOG), P300. Motor Evoked Potentials (MEP) and Contingent Negative Variation (CNV). The UltraPro with Natus Elite Software may be used to determine autonomic responses to physiologic stimuli by measuring the change in electrical resistance between two electrodes (Galvanic Skin Response and Sympathetic Skin Response). Autonomic testing also includes assessment of RR Interval variability. The UltraPro with Natus Elite Software is used to detect the physiologic function of the nervous system, for the location of neural structures during surgery, and to support the diagnosis of neuromuscular disease or condition.
The listed modalities do include overlap in functionality. In general, Nerve Conduction Studies measure the electrical responses of the nerve; Electromyography measures the electrical activity of the muscle and Evoked Potentials measure electrical activity from the Central Nervous System.
The UltraPro with Natus Elite Software is intended to be used by a qualified healthcare provider.
The UltraPro S100 system is designed for the acquisition, display, analysis, reporting, and management of electrophysiological information from the human nervous and muscular systems. The system is designed to perform Nerve Conduction (NCS). Electromyography (EMG), Evoked Potentials (EP), and Autonomic Responses. UltraPro S100 system provides a variety of tests spanning the various modalities.
The UltraPro S100 system consists of the following major components:
- Main unit (also known as base unit or main base unit) with integrated control panel; ●
- Amplifier (3- or 4-channel);
- . Computer- laptop or desktop (with keyboard and mouse)
- Display Monitor (for desktop system)
- . Application Software (Natus Elite)
The UltraPro S100 has the following optional accessories/ components:
- Audio stimulators (Headphones or other auditory transducers)
- Visual stimulators (LED goggles or stimulus monitor)
- . Electrical stimulators (RS10 probes, stimulus probe with controls)
- Cart and associated accessories when using cart such as isolation transformer
- Miscellaneous accessories such as Patient Response button, Triple footswitch, Reflex hammer, temperature probe and adapter, ultrasound device, printer, etc.
The electrodiagnostics system is powered by a connection to mains.
The entire user interface of UltraPro S100 system consists of two major elements:
- The primary means to interact with the system is via a personal computer (PC) running ● Natus Elite.
- The second means of interaction is the user interface elements on the hardware.
The UltraPro S100 is intended to be used by a qualified healthcare provider. This device does not provide any diagnostic conclusion about the patient's condition to the user. The intended use environment is in a professional healthcare facility environment.
The provided text is a 510(k) Summary for the Natus Ultrapro S100 device. While it describes the device's indications for use and compares its technological characteristics to predicate devices, it does not contain information about the acceptance criteria or the specific study that proves the device meets those criteria, such as a clinical performance study with defined metrics like sensitivity, specificity, or accuracy. This document focuses on demonstrating substantial equivalence to a predicate device primarily through technical specifications and intended use.
Therefore, I cannot provide a table of acceptance criteria, reported device performance, sample sizes used for test/training sets, data provenance, number or qualifications of experts, adjudication methods, or details about MRMC or standalone studies based on the provided text. The document is primarily a comparison of features and intended use.
The "Conclusion" section on page 14 states: "Verification and validation activities were conducted to establish the performance and safety characteristics of the UltraPro S100. The results of these activities demonstrate that the UltraPro S100 is safe, effective, and performance is substantially equivalent to the predicate devices." However, it does not elaborate on what these activities entailed or the specific criteria and results.
Ask a specific question about this device
(90 days)
, North Carolina 28269
Re: K242345
Trade/Device Name: Delphi Stimulator Regulation Number: 21 CFR 882.1870
Classification Code: Device: Stimulator, Electrical, Evoked Response Product Code: GWF Regulation No: 21 CFR 882.1870
|
| Regulation Number | 882.1870
| 882.1870
The Delphi Stimulator is intended to be used for stimulation of peripheral nerves for diagnostic purposes.
Delphi stimulator is a Magnetic stimulator used for Magnetic stimulation of peripheral nerves for diagnostic purposes. Magnetic stimulation is a non-invasive technique for stimulating neural tissue. Application areas of magnetic stimulation are a sub-set of the application areas for current stimulation. The Delphi stimulator is connected to a Magnetic Coil which transfers the magnetic stimulation to the tissue.
The stimulator consists of power electronics to generate the magnetic field in the Magnetic Coil. It is controlled via a user interface installed on a PC workstation connected to the stimulator. enabling the operator to overview all functions, stimulus sequences, controls, status and measured data. The magnetic pulse is Biphasic waveform and the stimulator can stimulate with a frequency of up to 3 pulses per second (pps).
The provided text does not contain information about acceptance criteria and a study that proves the device meets those criteria in the format requested. The document is an FDA 510(k) summary for the "Delphi Stimulator," which focuses on demonstrating substantial equivalence to a predicate device.
The "Performance Data" section (Section 10) lists various tests performed, but these are primarily bench tests for substantial equivalence, electrical safety, EMC, and software validation. It does not present specific acceptance criteria (e.g., minimum sensitivity or specificity targets) or a clinical study with human subjects to evaluate the device's diagnostic performance against such criteria.
The information provided is as follows:
1. A table of acceptance criteria and the reported device performance:
- No explicit table of acceptance criteria with corresponding device performance for diagnostic accuracy (e.g., sensitivity, specificity) is provided.
- The document mentions "Stimulation output comparison to Predicate" where the results "demonstrated substantial equivalency." This is a comparison for substantial equivalence, not a performance metric against a set acceptance criterion.
2. Sample size used for the test set and the data provenance:
- Not applicable. No clinical test set involving patient data is described. The performance data refers to nonclinical bench tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. No clinical test set requiring expert-established ground truth is described.
4. Adjudication method for the test set:
- Not applicable. No clinical test set requiring adjudication is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a stimulator for diagnostic purposes, not an AI-powered diagnostic system that assists human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a stimulator, not an algorithm.
7. The type of ground truth used:
- Not applicable. For the bench tests performed, the "ground truth" would be the expected electrical and magnetic outputs, and adherence to safety and software standards, which were verified directly rather than against a clinical ground truth.
8. The sample size for the training set:
- Not applicable. This document describes a medical device, a stimulator, not a machine learning or AI model that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable. As above, this document does not refer to a machine learning or AI model.
Summary of available "Performance Data" from the document:
| Test | Test Method Summary | Results |
|---|---|---|
| Stimulator Output Comparison to Predicate | Stimulation output was measured in SMT units (Electrical field) and compared to the predicate (per FDA Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems) | The obtained results demonstrated substantial equivalency |
| System verification testing | The device was operated at various intensities to verify it functions as intended | No failures were observed; test passed. |
| ASCA test for basic safety and essential performance | Testing per IEC 60601-1 and IEC 60601-1-2 | Pass |
| Software testing | Software testing per the requirement of the software life cycle as defined in IEC 62304:2006 Amendment 2015 - Medical device software - Software life cycle processes. | Pass |
| Cybersecurity assessment | Evaluation of threat per: - FDA Guidance, Post-market Management of Cybersecurity in Medical Devices - FDA Guidance, Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions. - AAMI TIR57:2016 Principles for Medical Device Security-Risk Management - FDA Guidance, Select Updates for the Premarket Cybersecurity Guidance: Section 524B of the FD&C Act | No vulnerabilities were found; labeling includes all required elements addressing cybersecurity |
In conclusion, the document describes non-clinical bench tests and adherence to safety and software standards to support the substantial equivalence claim, rather than a clinical study evaluating diagnostic performance against specific acceptance criteria.
Ask a specific question about this device
(115 days)
92008
Re: K234092
Trade/Device Name: SafeOp 3: Neural Informatix Systeem Regulation Number: 21 CFR 882.1870
|
| Classification Name: | Stimulator, Electrical, Evoked Response(21 CFR 882.1870
andreference devices |
| RegulationNumber (21CFR) | §882.1870
| §882.1870
| §882.1870
The SafeOp 3: Neural Informatix System is intended for use in monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures.
SafeOp 3 Accessories: The SafeOp Accessories are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.
The SafeOp™ 3: Neural Informatix System (SafeOp 3 System), consists of the SafeOp patient interface with power supply and IV pole mount, the Alpha Informatix Tablet with docking station and power supply and a data transfer USB cable. Associated disposable accessories consists of an electrode harness, surface and/or subdermal needle electrodes, MEP Activator, Cranial Hub, PMAP Dilators and stimulating probe or clip contained in various kits.
The subject device is intended for use by trained healthcare professionals, clinical neurophysiologists/technologists and appropriately trained non-clinical personnel. The subject device is intended for use in operating room environments of hospitals and surgical centers. System setup may be performed by both clinical and trained non-clinical personnel.
The subject device records the following modalities:
- Somatosensory evoked potentials (SSEP)
- Motor evoked potentials (MEP),
- . Train-of-four neuromuscular junction (TO4),
- Triggered electromyography (tEMG) and ●
- . Free run electromyography (sEMG)
The provided text does not contain detailed information about specific acceptance criteria for the device's performance, nor does it describe a study that rigorously proves the device meets such criteria through a clinical validation or similar performance evaluation.
The document is a 510(k) premarket notification summary for the "SafeOp 3: Neural Informatix System." Its primary purpose is to demonstrate substantial equivalence to a previously cleared predicate device (SafeOp2: Neural Informatix System, K213849, and reference device Cascade IOMAX Intraoperative Monitor, K162199), rather than to present a full clinical performance study with defined acceptance criteria and detailed results.
Here's a breakdown of what the document does say, and what it lacks in relation to your request:
What the document provides:
- Device Name: SafeOp 3: Neural Informatix System
- Intended Use/Indications for Use: Monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ) during various surgical procedures.
- Technological Comparison: A table comparing the SafeOp 3 System to predicate and reference devices, focusing on technical specifications like monitoring modalities, amplifier channels, stimulation parameters (voltage, current, pulse duration, repetition rate), and filter ranges. This comparison primarily aims to establish that the differences in technology do not raise new questions of safety or effectiveness.
- Performance Data (Non-clinical): Mentions that "Nonclinical performance testing demonstrates that the subject SafeOp 3 System meets the functional, system, and software requirements." It also states "EMC and Electrical Safety Testing... was performed to ensure all functions... are electrically safe, and comply with recognized electrical safety standards." Usability testing was also performed.
- Clinical Information Disclaimer: Explicitly states, "Determination of substantial equivalence is not based on an assessment of clinical performance data."
What the document lacks significantly for your request:
- A table of acceptance criteria and reported device performance: This is the most significant omission for your request. The document details technical specifications and comparisons but does not provide quantitative performance metrics (e.g., accuracy, sensitivity, specificity, or specific error rates) against pre-defined acceptance thresholds for any of its functionalities (MEP, SSEP, EMG, NMJ). The performance data mentioned are non-clinical (functional, system, software, EMC, electrical safety, usability), not clinical performance metrics.
- Sample size used for the test set and data provenance: Since specific clinical performance studies are not detailed, this information is not provided.
- Number of experts used to establish ground truth and qualifications: Not applicable as a clinical ground truth establishment process for performance evaluation is not described.
- Adjudication method for the test set: Not applicable.
- MRMC comparative effectiveness study: No such study is mentioned or detailed.
- Standalone (algorithm only) performance: While the device is an "algorithm only" in a sense (it processes physiological signals), its performance isn't quantified in a standalone clinical evaluation or comparative study.
- Type of ground truth used: No clinical ground truth is described for performance evaluation.
- Sample size for the training set: Not applicable, as this is related to AI/ML development and training, which is not described. The device is a neuromonitoring system, not explicitly stated to be an AI/ML device in the context of this submission.
- How the ground truth for the training set was established: Not applicable.
Why this information is missing:
The FDA 510(k) pathway for "substantial equivalence" often relies on demonstrating that a new device is as safe and effective as a legally marketed predicate, without necessarily requiring new clinical trials or detailed performance studies if the technological differences are minor and well-understood. The focus is on showing that any differences do not introduce new safety or effectiveness concerns.
In summary, based solely on the provided text, I cannot complete the table of acceptance criteria or describe a study that proves the device meets these criteria in a clinical performance context. The document focuses on demonstrating substantial equivalence through technical comparison and non-clinical testing, rather than presenting clinical performance metrics.
Ask a specific question about this device
(94 days)
Device Name: MEGA-IOM system with Neuro-IOM.NET software (32/B - 32/S - 16/S) Regulation Number: 21 CFR 882.1870
|
| Classification andProduct Code: | 21 CFR 882.1870
Neuro-IOM system with Neuro-IOM.NET softwaremodels -32/B -32/S -16/SNeurosoft Ltd21 CFR 882.1870
| ClassificationNumber | 21 CFR 882.1870
| 21 CFR 882.1870
MEGA-IOM system with Neuro-IOM.NET software is a medical device intended for intraoperative neurophysiologic monitoring: the device provides information to assess a patient's neurophysiological status.
The system allows to monitor the functional integrity and/or mapping of central and peripheral nervous system including motor and sensory pathways.
It is provided in three different configurations:
I. 32/B
II. 32/S
III. 16/S
The system ensures the following IOM modalities: free-run EMG (electromyography), direct nerve stimulation including pedicle screw test, SSEP (somatosensory evoked potential), MEP (motor evoked potential), EEG (electroencephalography), AEP (auditory evoked potential), VEP (visual evoked potentials), direct cortical stimulation. Also the train-of-four (TOF) stimulation is performed.
The system is not intended to measure vital signs. It records the data to be interpreted by the neuromonitoring specialist.
MEGA-IOM system is intended for use in adults age 18 years and older.
The MEGA-IOM system with Neuro-IOM.NET software is intended for intraoperative neurophysiological monitoring (IOM).
The system is intended for monitoring the functional integrity and/ or mapping of central and peripheral nervous system including motor and sensory pathways.
Soterix Medical is assuming manufacturing of the predicate product cleared in K190703. This 510(k) submission includes the same hardware cleared with the predicate device system (no changes) and an updated version of the software compared to the predicate device.
Below, a high-level summary of the changes from the predicate are provided:
Software: The primary changes in this 510(k) submission are related to the software. The changes are troubleshooting updates for greater usability such as user interfaces and bug fixes. Detailed descriptions of these software changes per version are provided below:
Version 1.1.52.7: Fixes related to the stimulation stop after a few hours, adding a stimuli limit for double train stimulation, allowing pulse duration (200 microseconds) entry, masking noise limitation per IEC 60601-2-40 (Particular requirements for the basic safety and essential performance of electromyographs and evoked response equipment), reporting correct status of multichannel operation, prevent skipping of traces, and check for incompatible stimulation parameter entry.
Version 1.1.52.8: Fixes related to facilitating quick turning off stimulation, real-time stimulus change control, scrolling in spectral setup window, constraining window within pattern printout to screen, ensuring modality window description is unchanged, incorrect Ch 17 display in emulation mode, video filters error during installation, ensuring all text is displayed within visibility area, incorrect auditory evoked potentials (EPs) average calculation in emulation mode.
Version 1.1.52.9: Fixes related to exam saved message, flexibility to change test template sequence, sleep mode prevention, rendering speed, interpulse interval display correction, allowing comments in the event window, automatic creation of user template, template list update, impedance measurement window rendering, parameter display correction in EMG trigger window and Group mode, ensuring non-disappearance of red line around patient's birth date, enabling stimulation via local menu in Scoliosis test template, frame switching error during video recording, displaying correct device status.
Version 1.1.53.7: Fixes related to addition of video clip creation button, accidental stimulation activation protection, display of preliminary train in double train mode, voltage and charge value display in stimulation panel, allowing repetitive stimulation in Double Train and Train+Pulse modes, option to view amplitude values up to 10 microA, option to select scales in X axis, option to delay auto saving if stimulation in progress, option to titrate stimulation amplitude using mouse wheel, option to select grid type, confirmation on layout uninstallation, addition of the Pause button in Stimulation programs, option to select motor response search interval and displaying the motor response search interval, addition of measured impedance visualization, offering first the windows already available when adding a new window, preventing closure of window if input row is active in derivation test, enable display of additional TOF parameters, option to playback trace fragments, addition of Take a screenshot command, parameter trend export, pulse duration interval display in TES train, option to enter multiple-line comments, option to re-name layout, option to hide command list, ability to save only responses beyond a certain level, stimulator list scrolling, RMS noise value display, test template cloning, motor response onset marker, previous recorded traces display, layouts during new test reset, stimulator reminder mod, video file duration based on frame rate, saving of the number of traces, test template rewriting, group trace visualization display, layout switching, stimulation resumption post coagulator off, video playback if another program is run.
Version 1.1.53.8: Fixes related to allowing multiple image view within the image window, allowing EEG artifact recording, showing baseline in the Overlay mode, option to disable program, option to save screen shots to a specific folder, stimulation program panel addition, correction of autoincrement window bug, correction of DSA rendering.
Version 1.1.53.9: Fixes related to custom scale adjustment in Freerun window, randomly changing pulse intervals in multichannel mode, event trace copy to report, layout panel updating post deletion, sound, threshold, and color parameter display during new site addition, EMG channels in template during long-term monitoring, site visibility during stimulation in presence of temporary markers, auto sweep resets, temporary markers display, trace export to EDF, layout non-save post video window opening, type conversion in trace window.
Version 1.53.10: Fixes related to video window copying over to report, trace scale reset, appearance of average trace display in case of no averages due to artifacts, site name display in TOF window, saving of initial study state in autosave mode, incomplete installation of video components, volume increment of auditory stimulator signal volume.
Version 1.1.53.11: Fixes related to active probe switch-off during monitoring, limiting maximum increment adjustment of electrical stimulation amplitude to 1 mA and 5V, stimulus duration set-up when stimulation amplitude is increased, stimulation disablement after power supply resume, stimulus sound notification, text visualization in dialogue box, non-active stimulator panel, maximum stimulus artifact width display, minimum and maximum MEP trace markers.
Version 1.1.53.12: Fixes related to pulse width change during stimulation in multichannel mode, MS Office word report, TeamViewer, logo, and translation corrections.
Version 1.1.53.13: Fixes related to removal of sites containing TOF and logo change.
Version 1.1.53.14: Fixes related to start monitoring during video camera switch off, video recording in some interface languages, sound disablement during impedance measurement, printing test template, USB driver-related Windows 10 OS bugs.
In summary, the aforementioned software changes are related to resolving software bugs and increasing user convenience, but the changes do not impact the intended use, indications for use statement, contraindications, or warnings. They also don't increase the likelihood that the device will be used by a broader or different group of users and do not raise any new risks.
Labeling: Modifications have been made to labeling and these changes are related to the new device trade name and change in manufacturer. Instances of "Neuro-IOM" were changed to "MEGA-IOM" in the user manual and on the device labels. Soterix Medical is named the manufacturer and point of contact in the labeling including for all complaints and service requests.
Waveforms: The MEGA-IOM system has the same waveforms as the predicate device. No change to the waveforms.
The provided text is a 510(k) Premarket Notification from the FDA for the "MEGA-IOM system with Neuro-IOM.NET software". This document primarily focuses on demonstrating substantial equivalence to a predicate device (K190703 Neuro-IOM system with Neuro-IOM.NET software) rather than detailing extensive clinical studies with acceptance criteria for new device performance.
The submission is for a device that has primarily undergone software updates and a change in manufacturer. Therefore, the "study that proves the device meets the acceptance criteria" is in the form of software verification and validation, and comparison to the predicate device's established performance.
Here's an analysis based on the provided text:
1. A table of acceptance criteria and the reported device performance:
The document doesn't provide explicit "acceptance criteria" in the traditional sense of a clinical trial (e.g., target sensitivity/specificity for a diagnostic AI). Instead, the acceptance criteria are implicitly demonstrating that the updated software, in conjunction with the unchanged hardware, performs identically or equivalently to the predicate device. The performance is assessed by confirming that all functional and technical specifications remain the same.
| Parameter | Acceptance Criteria (Implicit - Identical to Predicate) | Reported Device Performance (MEGA-IOM) | Comment |
|---|---|---|---|
| Device Name and Model | Neuro-IOM 16S, 32S, 32B with Neuro-IOM.NET software | MEGA-IOM 16S, 32S, 32B with Neuro-IOM.NET software | New trade name for the subject device. Functionally identical. |
| Classification Name | Stimulator, Electrical, Evoked Response | Stimulator, Electrical, Evoked Response | Identical |
| Product Codes | GWF (primary), GWE, GWJ, OLT, PDQ | GWF (primary), GWE, GWJ, OLT, PDQ | Identical |
| Regulatory Class | Class II | Class II | Identical |
| Classification Number | 21 CFR 882.1870 | 21 CFR 882.1870 | Identical |
| Indications for Use (IFU) | Matches predicate (detailed in document) | Matches predicate (detailed in document) | Similar; new IFU includes new device name. All other wording identical. |
| Intended User | Trained personnel only | Trained personnel only | Identical |
| Device Hardware Setup | Connected to PC, not standalone | Connected to PC, not standalone | Identical (hardware is unchanged) |
| Electrical Safety Standards | Conforms to specified standards (older versions acceptable for predicate) | Conforms to updated versions of same standards | Subject device conforms to updated versions of standards. |
| Workflow, Menu | PC-controlled | PC-controlled | Identical |
| Interface to Computer | USB | USB | Identical |
| Channels | 16/32 | 16/32 | Identical |
| 1.5 mm touch-proof input jacks | Same | Same | Identical |
| Cable Length | 5m | 5m | Identical |
| Input Impedance | >1000 MOhm | >1000 MOhm | Identical |
| Common Mode Rejection (CMRR) | >90 dB | >90 dB | Identical |
| Low Frequency Filters | 0.2 Hz - 2000 Hz | 0.2 Hz - 2000 Hz | Identical |
| High Frequency Filters | 10 Hz - 4 kHz | 10 Hz - 4 kHz | Identical |
| Notch Filter | 50/60 Hz | 50/60 Hz | Identical |
| Sample Rate | 50 kHz | 50 kHz | Identical |
| Sensitivity | 0.05 µV/division to 20 mV/division | 0.05 µV/division to 20 mV/division | Identical |
| Noise Level | <0.6 µV (<9.5 nV/√Hz) | <0.6 µV (<9.5 nV/√Hz) | Identical |
| Artifact Rejection | Independent for each channel | Independent for each channel | Identical |
| Number of Channels | 16/12/4/4 | 16/12/4/4 | Identical |
| Max Intensity (Stimulator) | 200 mA | 200 mA | Identical |
| Duration (Stimulator) | 0.02 - 5 ms | 0.02 - 5 ms | Identical |
| Stimulus Type | Mono-/biphasic | Mono-/biphasic | Identical |
| Electrical Modes | Single, repetitive, trains | Single, repetitive, trains | Identical |
| TcMEP (Transcranial Electrical stimulator) | 4 channels, 1000V, 0.04-0.2ms duration | 4 channels, 1000V, 0.04-0.2ms duration | Identical |
| Low Current Stimulator | 3/2/1 channels, 20mA, 0.05-0.5ms duration | 3/2/1 channels, 20mA, 0.05-0.5ms duration | Identical |
| Stimulation Type (Auditory) | Click, tone, noise | Click, tone, noise | Identical |
| Rate (Auditory) | 0.01 - 100 Hz | 0.01 - 100 Hz | Identical |
| Intensity (Auditory) | 120 dB nHL | 120 dB nHL | Identical |
| Polarity (Auditory) | Condensation, rarefaction, alternating | Condensation, rarefaction, alternating | Identical |
| Transducers (Auditory) | Insert earphone EAR-3A-10 Ohms | Insert earphone EAR-3A-10 Ohms | Identical |
| Waveform Visualization (Examples) | Freerun, SSEP, AEP, VER, MEP, EEG, CSA, DSA, Trending, Log Book, History, Report Template | Identical visual representations | Identical |
| SSEP, MEP, TcMEP, BAEP, VEP, EMG, EEG, Multimodality support | Yes | Yes | Identical |
| Predefined test templates | Yes | Yes | Identical |
| Creation/editing of templates | Yes | Yes | Identical |
| Generation of neuromonitoring report | Yes | Yes | Identical |
| Image review from microscope/other sources | Yes | Yes | Identical |
| Trending | Yes | Yes | Identical |
| ESU Detection | Yes | Yes | Identical |
| Power Supply | 220/230 V AC, 50/60 Hz | 220/230 V AC, 50/60 Hz | Identical |
| Visual Stimulation | LED goggles | LED goggles | Identical |
| Type of electrodes | Any legally marketed (in the U.S.) probes and surface or needle electrodes with standard lead wire. | Any legally marketed (in the U.S.) probes and surface or needle electrodes with standard lead wire. | Identical |
2. Sample size used for the test set and the data provenance:
The document states that the changes are primarily software-related (bug fixes, usability improvements) and do not impact the device's fundamental performance characteristic or intended use. Therefore, instead of a clinical "test set" with patient data in the typical sense for AI/diagnostic devices, the "testing" described is focused on software verification and validation.
- Sample size for test set: Not applicable in the sense of clinical cases or patient samples. The "test set" consists of software functionalities and user interface elements.
- Data provenance: Not directly applicable to patient data. The provenance relates to the software development process and testing environment. It is internal software validation data.
- Retrospective or prospective: Not applicable to patient data. Software verification and validation are typically performed prospectively during development and after changes.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This is not applicable as the "ground truth" for this submission is the established, cleared performance of the predicate device. The changes are bug fixes and usability enhancements, not a new diagnostic algorithm requiring expert-established ground truth. The "experts" would be the software developers and quality assurance personnel performing the verification and validation, ensuring the software performs as designed and intended, consistently with the predicate's established functions.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. Software verification and validation follow structured testing protocols (e.g., unit testing, integration testing, system testing, regression testing) and risk management processes. There is no "adjudication" in the sense of reconciling clinical interpretations.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No. This submission is not for a new AI/CAD device requiring an MRMC study. It is for an established device platform with software updates. The device itself is an "Evoked Response Electrical Stimulator" and "Intraoperative Neurophysiologic System" that records data for interpretation by a neuromonitoring specialist; it does not explicitly feature AI assistance for interpretation that would necessitate an MRMC study.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This device is an instrumentation system for intraoperative neuromonitoring, where the data is recorded to be interpreted by the neuromonitoring specialist. It is inherently a human-in-the-loop system. The software updates are "troubleshooting updates for greater usability such as user interfaces and bug fixes."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this submission is the documented and previously cleared performance of the predicate device. The new device demonstrates substantial equivalence by showing that its software updates do not alter the fundamental safety or effectiveness of the system, which is achieved by confirming that hardware specifications remain identical, and software functions operate as expected (i.e., fixing bugs, improving usability) without introducing new risks or changing the device's core output.
8. The sample size for the training set:
Not applicable. This device is not an AI/machine learning device that requires a "training set" in the context of learned algorithms. The "software updates" are bug fixes and usability improvements to a traditional software application, not a learning algorithm.
9. How the ground truth for the training set was established:
Not applicable, as there is no "training set" in the context of machine learning. The "ground truth" for the software's correct operation is defined by its functional requirements and the established performance of the predicate device.
Ask a specific question about this device
(28 days)
Headboxes, ISIS Neurostimulator, ISIS Xpert Plus, ISIS Xpert, ISIS Xpress Regulation Number: 21 CFR 882.1870
| |
| ClassificationRegulation: | 21 CFR §882.1870
The product is used for intraoperative monitoring and testing during surgical procedures
· to examine neuronal tissue (central and peripheral nervous system) by recording and stimulation
The product can be used during surgical procedures that justify non-therapeutic clinical use of the following modalities or their combinations:
- · Measurement:
- o Auditory evoked potentials (AEP)
- o Electroencephalography (EEG)
- o Electrocorticography (ECoG)
- o Electromyography (EMG)
- o Somatosensory evoked potentials (SSEP)
- o Motor evoked potentials (MEP)
- o Train of Four (TOF)
- · Stimulation:
- o Transcranial electrical stimulation (TES)
- o Direct cortical and subcortical stimulation (DCS)
- o Direct nerve stimulation (DNS)
- o Transcutaneous intraoperative nerve stimulation (TINS)
- o Direct muscle stimulation (DMS)
The ISIS Headboxes and the ISIS Neurostimulator constitute multimodal intraoperative neuromonitoring systems called ISIS IOM Systems. These systems consist of custom stimulation and recording hardware, a standard laptop or desktop personal computer running an off-the-shelf operating system, and operating software called NeuroExplorer. As an option, these systems mount on device carriers or housings tailored for intraoperative use.
The ISIS IOM Systems support the following measurement modalities:
- -Auditory Evoked Potentials
- -Transcranial and cortical Motor Evoked Potentials
- Somatosensory Evoked Potentials -
- Free-running and triggered Electromyography -
- Electroencephalography -
- Train of Four -
The provided document is a 510(k) summary for the ISIS Headboxes and ISIS Neurostimulator, which are intraoperative neuromonitoring systems. It focuses on demonstrating substantial equivalence to a predicate device (K212166) rather than establishing novel acceptance criteria and proving performance against them in a clinical study.
The document states that no additional clinical testing was performed for the subject devices. Instead, the submission relies on bench testing against established standards and internal requirements to demonstrate safety, effectiveness, and performance that is "as well as or better than" the predicate device.
Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of a de novo AI/ML device (which often involves clinical performance metrics like sensitivity, specificity, etc.) are not directly applicable or available in this document.
Here's an attempt to extract the relevant information based on the provided text, acknowledging the limitations:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a submission demonstrating substantial equivalence to a predicate device through bench testing rather than a de novo AI/ML device with pre-defined performance metrics, the acceptance criteria are primarily related to adherence to international standards and internal requirements for electrical safety, EMC, and software validation.
| Acceptance Criteria Category | Specific Criteria (Implicit or Explicit in Document) | Reported Device Performance |
|---|---|---|
| Biocompatibility | No patient contact materials | Not applicable |
| Software | Conformance to: | Demonstrated compliance with predetermined specifications, applicable guidance documents, and standards. |
| - FDA Guidance: Content of Premarket Submissions for Device Software Functions, Jun 14, 2023 | ||
| - FDA guidance: Off-the-shelf software use in medical devices, August 11, 2023 | ||
| - FDA guidance: General principles of software validation: Final guidance for industry and FDA staff, Jan 02, 2002 | ||
| - FDA guidance: Content of premarket submissions for management of cybersecurity in medical devices, Oct 02, 2014 | ||
| - IEC 62304:2006/AMD1:2015, Medical device software - Software life cycle processes | ||
| Electrical Safety | Conformance to: | Test results demonstrate compliance with applicable standards. |
| - IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (or IEC 60601-1: 2012 reprint) | ||
| - IEC 80601-2-26:2019 (Electroencephalographs) | ||
| - IEC 60601-2-40:2016 (Electromyographs and evoked response equipment) | ||
| Electromagnetic Compatibility | Conformance to: | Test results demonstrate compliance with applicable standards. |
| - IEC 60601-1-2:2014 | ||
| Performance Testing – Bench | Fulfillment of internal requirement specifications for: | Products successfully underwent bench testing, confirming fulfillment of requirements. |
| - Electrical medical systems | ||
| - System carrier | ||
| - Amplifier (ISIS Headboxes) and stimulator (ISIS Neurostimulator) modules | ||
| - Operating software (NeuroExplorer) incl. firmware | ||
| - Accessories (adaptor boxes) | ||
| - Custom Microsoft® Windows 10 image | ||
| Human Factors | Demonstrates safety and no need for further UI improvement | Testing confirms products are safe to use. |
| Overall Performance | As safe, as effective, and performs as well as or better than the legally marketed predicate. | Demonstrated. |
2. Sample Size Used for the Test Set and the Data Provenance
- Sample Size: Not applicable in the traditional sense of a clinical test set with patient data. The testing was primarily bench and software validation. The "sample size" would refer to the number of units tested, which is not specified, but implied to be sufficient for type testing.
- Data Provenance: The data refers to the results of internal bench tests and software validation activities, conducted by the manufacturer (inomed Medizintechnik GmbH) in Germany. It is entirely retrospective in the sense that it's performed on a developed product to meet predefined standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- This information is not provided because no clinical test set with expert-established ground truth was used. The ground truth for functional verification would be the expected output or behavior according to engineering specifications and regulatory standards.
4. Adjudication Method for the Test Set
- Not applicable as there was no clinical test set requiring expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC comparative effectiveness study was done. The device is a neuromonitoring system, not an AI-assisted diagnostic or interpretive tool that would inherently involve "human readers" in that sense. The submission explicitly states: "No additional clinical testing was performed".
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- The document implies that the "software validation" and "bench testing" constitutes a standalone performance evaluation of the device's functional integrity as an algorithm/system. The software (NeuroExplorer) and hardware components were tested to ensure they meet their predetermined specifications and comply with relevant standards independently of human interpretation of clinical outcomes. However, this is not an "AI algorithm only" type of standalone performance, but rather functional performance of medical device software/hardware.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For the bench and software testing, the "ground truth" would be established by the engineering specifications, international performance standards (e.g., IEC 60601 series), and documented functional requirements of the device. This is a technical (rather than clinical/biological) ground truth.
8. The sample size for the training set
- This device is not described as an AI/ML device that requires a "training set" in the context of machine learning model development. Therefore, this question is not applicable. The software development follows a traditional software lifecycle process, not a machine learning training paradigm.
9. How the ground truth for the training set was established
- Not applicable, as there is no mention of an AI/ML training set.
Ask a specific question about this device
Page 1 of 11